Introduction to HPLC High Performance/Pressure Liquid Chromatography Prepared By: Dr. Radwan Almofti What is HPLC? A separation technique based on the differential affinity of sample components between the solid stationary phase and the liquid mobile phase For the purposes of identification, and/or quantification of these components DEFINITIONS Chromatography: a technique used to separate the components of mixtures of chemical compounds based on different rates of travel through a stationary phase resulting from the differential affinities of these components of mixtures for the stationary and for the mobile phases. The aim is to identify and/or quantify these components. Differential Affinities: The difference between the attractions of the components of a mixture to the mobile phase and to the stationary phase and the rate of the migration of these components between the two phases. Attraction examples are: 1- Polarity 2- Solubility 3- Ion Exchange 4- Size exclusion Chromatograph: An equipment that uses chromatography Chromatogram: The visual pattern of separated components of a mixture (sample) obtained by chromatography History of HPLC Liquid Chromatography (LC) was firstly defined in 1906 by the work of M.S. Tswett on the separation of leaf pigments. The revolutionary High Pressure LC (HPLC) was firstly described by Prof. C. Horvath in 1970. In 2004, Ultra High Performance LC (UPLC) was introduced. Advantages of HPLC HPLC is the primary technique used in Pharmaceutical Quality Control lab because it is: Sensitive Selective Low sample consumption Moderate analytical conditions Superior capability and reproducibility Automation Components of HPLC Detector Column Pump Mobile phase Injector Drain Data processor Degasser MOBILE PHASE (ELUENT) Could be polar (water, buffer, methanol, acetonitrile), non-polar (hexane and benzene) or mixture of both Purity (HPLC grade) Detector Compatibility (undetectable) Solubility of Sample (to carry it throughout the system) Low Viscosity (to avoid building pressure) Chemical Inertness (Analytes stability) Reasonable price (HPLC uses large volume) DEGASSER To remove dissolved air in the mobile phase Dissolved air causes: Unstable delivery by pump More baseline noise and drift There are two main types: Offline degasser: Using ultrasonic along with decompression by aspirator before connecting it to HPLC system Online degasser: Using helium purge or gas-liquid separation membrane while mobile phase is connected to HPLC system PUMP The role of the pump is to force the mobile phase through the column. Two types: 1) Constant pressure (fluctuation flow, less use) 2) Constant flow (most common) because of its ability to deliver: - Constant mobile phase “ISOCRATIC” - Varying mobile phase “GRADIENT” INJECTOR Sample is introduced to the HPLC system through the injector. There are two types of injectors: Manual injectors Autosamplers Manual Injector From pump Loop Loop From pump To column LOAD To column INJECT Autosampler From pump To column To column From pump Needle Sample vial Measuring pump LOAD INJECT COLUMN Type 316 Stainless Steel Inner diameter range 2.6 – 5 mm Length 10 – 30 cm Packed with the stationary phase (usually silica gel or its derivatives) of particle size range of 3 to 10 µm and pore size range of 60 to 100 A° FRITS Frits are filters and have two main types: 1.Column frits: one at the beginning (inlet frit) and one at the end (outlet frit) of the column. Made of Stainless Steel 316 grade to be inert, resistant to corrosion and withstand high pressure. pore size range from 0.5 to 2 um (depending on the stationary phase particle size) Primary functions are: Filter particulate contaminants (microbial and inorganic) to protects the column Retain packing (stationary phase) in column Complete distribution of sample uniformly into column Provide uniform flow of sample into column 2) Frit for purge valve: Made of Polytetrafluoroethylene (PTFE) polymer. Filter particulate contaminants See this youtube: https://www.youtube.com/watch?v=mt3wcpEv1Wc Types of Stationary Phases 1) Normal Phase: Silica gel - Polar 2) Reversed Phase: Silica derivatized - Non-polar 3) Ion Exchange: Silica or resin derivatized to have anionic or cationic active sites 4) Size Exclusion: To separate polymers per their MWs NORMAL PHASE Polar Stationary Phase Silica gel: -Si-OH or its derivatives such as: Non-Polar Mobile Phase • • Amino type: -Si-CH2CH2CH2NH2 Basic solvent (hexane, benzene) Additive solvent to adjust polarity to improve separation (ethanol, methanol). Used for Non or Low Polar Analytes The highest polar analytes have the highest affinity to polar stationary phase therefore they exit the column the last REVERSED PHASE Non or Low Polar Stationary Phase • • Most common is silica gel bond to aliphatic chain Most common is (ODS) silica bond to OctaDecyl (-C18H37) Si -O-Si CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH3 C18 (ODS) Polar Mobile Phase • • • Water has the highest polarity Less polar solvents commonly used: methanol, acetonitrile, tetrahydrofuran Buffer is added if pH is important for separation REVERSED PHASE (Continue) Used for Polar Analytes Most drugs are polar salts soluble in water. Therefore, reserves phase HPLC is the most common in pharma QC. The highest polar analytes have the lowest affinity to the nonpolar stationary phase thus exit the column the first and vice versa. C18 (ODS) OH Weak Strong CH3 NORMAL PHASE VS REVERSED PHASE Phase Type Stationary Mobile Phase (SP) Phase (MP) Separation - Affinity Reversed Phase Non-polar Polar •Polar analytes have weak affinity to SP & strong affinity to MP thus exit early •Non-polar analytes have strong affinity to SP & weak affinity to MP thus exit late Normal Phase Polar Non-polar •Polar analytes have strong affinity to SP & weak affinity to MP thus exit late •Non-polar analytes have weak affinity to SP & strong affinity to MP thus exit early DETECTOR 1) Ultraviolet-Visible (UV-VIS): Based on the principle that the emission absorbance is proportional to the concentration of the absorbing analytes Detection cell C: Concentration Ein A Eout l A = e·C·l = –log (Eout / Ein) (A: absorbance, E: absorption coefficient) C DETECTOR (Continue) Ultraviolet-Visible (UV-VIS) has three types: A- Fixed one wavelength B- Variable wavelength Ability to select the wavelength at which the analyte has highest absorbance C- Photo Diode Array (PDA) Ability to detect the absorbance across the entire UV spectrum Variable wavelength and PDA detectors are the most common used ones in pharma QC. DETECTOR (Continue) 2) Florescence: For analytes that once excited emit florescence proportionally to their concentration 3) Conductivity: For ion-exchange chromatography 4) Electrochemical: Detect the current generated from the oxidation/reduction reaction of the analytes. Data Collection and Processing Using Software (Empower for Waters HPLCs) Collect the output from the detector (HPLC: 1 to 5 data points per second) Plot the detector response (UV absorbance) over time (Chromatogram) Process data collected and present the required results (identity, concentration) Data are stored in local computers, servers or over network (Lab Information Management System “LIMS”) Intensity of detector signal (mAU) CHROMATOGRAM tR Peak t0 t0: Non-retention time h A Time Injection tR: Retention time A: Peak area h: Peak height CAPACITY (K’) The ratio of the elution time for the analyte (tR) to that for the non-retained one (t0). Its value depends on the affinity of the analyte to the stationary phase compared to that to the mobile phase Calculated as follows: Detector Response K’ = (tR – t0)) / t0 tR t0 Time tR: Retention time t0: Non-retention time SELECTIVITY (α) The ability of HPLC stationary phase to separate two components in a sample Calculated as the ratio of the capacity factors k’ α = k2’/k1’ = (tR2 - to) / (tR1 - to) EFFICIENCY (N) The width (sharpness) of the peak that represents the suitability of HPLC stationary phase to perform separation Defined in terms of Number of Theoretical Plates (N) and calculated as: N = 16 X (tR / Wb)2 (tR is retention time and Wb is peak width at baseline) RESOLUTION (R) The most important HPLC performance measurement Defined as the extent two compounds can be separated Calculated as: R = 2 X (tR1 – tR2) / (Wb1 + Wb1) Required to be ≥ 1.5 tR1 tR2 Wb1 Wb2 PEAK TAILING A measure of peak symmetry Calculated as: T = W0.05 / (2 X f ) (W0.05 is peak tail width and f is peak front width both at 0.05 height) T value for ideal symmetrical peak is 1 Acceptable range 0.9 – 1.2 SYSTEM SUITABILITY To verify that HPLC system is adequate for the analysis Two parts: 1) Performance Measurements: To ensure that capacity, resolution and tailing values as determined by test method are adequate to ensure successful analysis 2) System Reproducibility: • The ability of HPLC system to reproduce the desired injection volumes • Relative standard deviation (RSD) ≤ 2% for 5 replicate injections of primary standard (USP 621) • For multiple samples, a check standard after 6 sample injections with a recovery of 100.0 ± 3.0% (USP 621) Chromatographic Troubleshooting Baseline fluctuation/noise Baseline drifting Ghost peaks Peak doublets Peak broadening Peak tailing Peak fronting Negative peaks Baseline Fluctuation/Noise Causes for baseline noise are: Detector problem (dirty flow cell, air bubbles, lamp failing) Method (mobile phase change, insufficient degassing, system not washed properly between runs, dirty solvent) Pump (mixing issue, pulse) Baseline Drifting Could be upward or downward drifting Causes for baseline drifting are: Gradient elution issue Unstable temperature Mobile phase contamination Mobile phase not in equilibrium with column System bleed Upward drifting Ghost Peaks Peaks appears even no sample injection is made. Causes for ghost peaks are: Dirty mobile phase Dirty column System is not washed properly between runs Peak Doublet All peaks doublets, possible causes are: Void volume in column Partially blocked frit One peak doublet, possible cause is: Co-eluting components Peak Broadening All peaks broadening, possible causes are: Loss column efficiency Void volume in column Large injection volume High mobile phase viscosity Some peaks broadening, possible cause is: Late elution from previous sample High MW (Polymer) Peak Tailing Small peak eluting on tail of larger peak Residual silanol interaction Column contamination Bad column Peak Fronting Usually due to sample overload Negative Peak Usually due to the absorbance of the sample being less than the mobile phase Pharmaceutical Quality Control Guidelines Canada: Food and Drug Regulations Part C, Division 2 (GMP guidelines) Sections: C-02-009/010/016-019 USA: Code of Federal Regulations, Title 21 Food and Drugs, part 211 (cGMP guidelines) Subpart I (Laboratory Control) & Subpart J (211-194 Laboratory Records) The United State Pharmacopeia (USP), European Pharmacopeia (EP) and Japanese Pharmacopeia (JP) are approved guidelines by the regulators Pharmaceutical Quality Control Guidelines (continue) Canadian GMP and US cGMP have similar main guidelines for quality control 1) The timely and full permanent documentations of all materials, equipment and procedures used, test results, name of analyst and date of each entry 2) The timely and full documented qualification and calibration of all equipment used (as applicable) THANK YOU…