Proposal of the Application-Specific Integrated Circuit and Radio

advertisement
World Applied Sciences Journal 32 (9): 1962-1969, 2014
ISSN 1818-4952
© IDOSI Publications, 2014
DOI: 10.5829/idosi.wasj.2014.32.09.1143
Proposal of the Application-Specific Integrated Circuit and
Radio Transceiver Choice for the Wireless Endoscopic Capsule
Mikhaylov Dmitry, Zhukov Igor, Starikovskiy Andrey, Khabibullin Timur and Konev Vladimir
Engineering Centre of the National Research Nuclear University “MEPhI”,
Kashirskoye Highway, 31, Moscow, Russian Federation
Abstract: This paper overviews advantages and disadvantages of CMOS and CCD matrixes used in endoscopic
capsules, specifies requirements that should be met by the microcontroller of the device for the improvement
of its effectiveness and concludes on the need of developing the application-specific integrated circuit (ASIC)
for the wireless endoscopic capsule. The ASIC microcontroller designed especially for the wireless capsule
endoscope will improve the specifications of the device in terms of the recorded images` size and the size of
transmitted data as well as reduce the energy consumption. The article also deals with radio transceiver choice
as it is essential in terms of energy consumption reduction and the data transmission speed increase. The main
ISM protocols are compared, namely NRF, Chipcon, Wi-Fi and Bluetooth. The article also provides the power
estimation of the proposed ASIC.
Key words: ASIC Microcircuit Controller
MICS standard ISM
Power consumption
INTRODUCTION
The wireless capsule endoscopy is a non-invasive
way to record images of the digestive tract for use in
medical examination and diagnosis [1-3]. A typical
capsule is the size and shape of a large pill and
contains a miniature camera. After a patient swallows
the capsule, the camera takes images as it traverses
down the gastrointestinal tract of the patient and the
capsule transmits them wirelessly to recorder. The
primary use of capsule endoscopy is to examine
areas of the small intestine that cannot be seen by
other types of endoscopy, such as colonoscopy or
esophagogastroduodenoscopy [4].
After the battery of the capsule is exhausted, or the
capsule leaves the digestive tract in a natural way,
recorder is connected to the gastroenterologist’s personal
computer and all the images are transferred by special
software on PC's hard drive to be further analyzed. The
block-diagram of the capsule is shown in Figure 1 [5].
During the work on this project the first generation of
wireless endoscopic capsules was developed. It was
based on stock components – image sensor, FPGA
Corresponding Author:
Image compression
Radio transceiver
(field-programmable gate array), SDRAM (Synchronous
Dynamic Random Access Memory) memory, MCU
(Multipoint Control Unit) and radiofrequency transceiver.
[6-8].
During the development of the capsule several global
problems that cannot be solved without an ASIC
(application-specific integrated circuit) were faced,
namely:
High power consumption (because of the size
constraints, the CR1/3N batteries with capacitance at
2 mA and current of 160 mA were used);
Very small endoscopic capsule`s size;
CMOS (Complementary Metal Oxide Semiconductor)
sensor does not have the scaler and internal memory,
so the logic should act very fast to read the frame,
scale it in FPGA and store it in external memory.
Of course, no compression was applied because of
the power and size constraints.
All these problems resulted in low image resolution
(120x160 pixels) and low speed of image taking (2 frames
per second).
Mikhaylov Dmitry, Kashirskoye Shosse 31, Moscow, 115409, Russian Federation.
1962
World Appl. Sci. J., 32 (9): 1962-1969, 2014
Fig. 1: Wireless endoscopic capsule`s scheme
Since capsule endoscopy is a growing medical
sphere it was decided to develop the second generation
of wireless endoscopic capsule eliminating disadvantages
by using ASIC providing among other things data
compression. Moreover, it is important to choose the
correct radio transceiver for data transmission from the
capsule to portable reader to reduce energy consumption
and increase the speed of data transfer.
Comparison of Cmos and Ccd Matrixes: To visualize the
mucosa of gastrointestinal tract of a patient the elemental
base of the wireless endoscopic capsule should include
photosensitive matrix such as CMOS or CCD (Charge
Coupled Device) with an object lens comprising a single
lens or a lens system providing the viewing angle of at
least 120 degrees.
To select the type of photosensitive matrix for the
capsule endoscopy it is proposed to compare the
specifications of CCD and CMOS matrixes from the point
of effectiveness and appropriateness of their use in the
endoscopic examination of the digestive tract.
CMOS matrix is characterized by low power
consumption (up to two orders of magnitude lower than
CCD), reading circuits` compactness (simple external
circuits or integrated with one circuit with matrix) and
"inherited" light-striking resistance of adjacent pixels.
CMOS sensor also has a high operating speed (up to 500
frames per second), but it is quite sufficient in CCD.
Moreover, CMOS matrix is able to read the image part, but
the use of this property (except, perhaps, lighting
correction) has not been found yet [9].
CCDs are characterized by the best image quality
(but CMOS has photographic quality up to several
megapixels that is sufficient for use in the wireless
endoscopic capsule), lesser noise, higher sensitivity and
dynamic range.
Despite the existing opinion there is no significant
difference in price between the two types of matrixes with
all other features being equal.
With respect to the wireless endoscopic capsule the
most important features are energy saving and the matrix
size. In this case CMOS matrix is more efficient. The
smaller sensitivity (in comparison to CCD) is balanced out
by the LED assembly and small spaces inside the human
body.
Therefore, it is proposed to use CMOS image sensor
in the wireless endoscopic capsule being developed.
However, the CMOS sensor has a disadvantage in
the form of a visual defect called rolling shutter (lines
displacement during the frame freeze). It occurs when a
line or a group of lines is read out while all other lines on
the sensor continue to be exposed [10].
It means that the image should be read quickly. The
radio channel for data transmission is quite narrow, so the
quick reading is often either impossible or leads to a
substantial increase in the chip cost. Therefore, the image
should be buffered directly on the capsule.
Micro Controllers: To improve the performance of the
endoscopic capsule of first generation it is necessary to
compress the images because the original image will have
a heavy weight [11]. For example, a frame with a resolution
of 100x100 pixels will weigh 20 kilobyte. It is difficult to
perform the transmission through a narrow channel of
even 20 kilobytes since the transfer of each byte is a time
and energy-consuming process.
For this end, the capsule of second generation
should contain the path of fast buffering and
simultaneous compression of image.
Several works on image compressor of the
endoscopic capsule have been reported [12-17]. They
show good results and present original ideas; however,
the possible disadvantages of these proposals may be
large capsule power consumption limiting the examination
time, limitations of matrixes specifications, etc. As distinct
from them, the proposed ASIC is able to perform two
types of compression: scale (proportional image
shrinking) and compression (data amount reduction).
1963
World Appl. Sci. J., 32 (9): 1962-1969, 2014
Compression and buffering logic is fast-acting, so
one of the most energy-consuming. In the shooting mode
of two images per second the camera takes up to 5 mA
and the transmitter – less than 10 mA (can be reduced to
2 mA). The rest power is consumed by the central path of
buffering and compression.
The implementation of this path is performed on
runoff microcontrollers with memory of at least 32
kilobytes (big memory for microcontrollers), or with high
eigen clock frequency of the core, around 100 MHz, or
with built-in interfaces DCMI or DMA.
DCMI (Digital Camera Interface) is the interface via
which the CMOS sensor produces an image; DMA (Direct
Memory Access) is the universal channel of direct
memory access without the involvement of the central
processing unit.
The implementation of a single peripheral is useful
because core clock frequency does not really matter. Only
the rate at which interfaces DCMI or DMA work is
important. This is a significant advantage, since the more
clock frequency the core has (controller that executes
commands) the greater power consumption of the
controller (exponential dependence) is.
However, if the microcircuit has a specific element
responsible, for example, for frame buffering, then this
area will consume less. This dependence has been tested
on the controllers STM 32F205 and STM 32F207 [18].
The STM 32F20x family is based on the highperformance ARM® Cortex™-M3 32-bit RISC core
operating at a frequency of up to 120 MHz. The STM
32F205xx and STM 32F207xx constitute the STM 32F20x
family whose members are fully pin-to-pin, software and
feature compatible, allowing the user to try different
memory densities and peripherals for a greater degree of
freedom during the development cycle. [18]
STM 32F205 is a two-bit general purpose
microcontroller in a small 3x4 mm package that is very
convenient for the developed endoscopic capsule of
second generation. However, the microcontroller has a
significant disadvantage. Although STM 32F205 has the
DMA interface that can be configured to accept the
image via the DCMI interface and immediately send them
to the microcontroller memory, the consumption of the
microcontroller is about 40 mA. It means that the total
consumption of the wireless endoscopic capsule is about
50 mA (in this case, the charge of the batteries will last for
about one hour of capsule operation).
STM 32F207 has a built-in DCMI interface and
other specifications quite similar to those of STM 32F205.
However, STM 32F207 is not available in a miniature
package.
Fig. 2: Block diagram of the proposed controller.
Characteristics of the Proposed ASIC for the Wireless
Endoscopic Capsule: Thus, the implementation of the
wireless endoscopic capsule of the second generation on
the runoff logic does not make sense. Therefore, it is
necessary to develop the special ASIC.
ASIC is an integrated circuit designed for a particular
rather than for general-purpose use. The proposed ASIC
chip will take the image via the DCMI interface, compress
the data and send them to its own memory.
The block diagram of the proposed microcontroller
for the wireless endoscopic capsule being developed is
presented in Figure 2.
The developed ASIC for the wireless capsule
endoscope should cover the most energy- and sizeconsuming elements, including data processing and meet
the following requirements:
1964
Ihe ASIC should be very energy-efficient – it is
possible to provide 10 mA in active mode, but the
lower - the better;
World Appl. Sci. J., 32 (9): 1962-1969, 2014
Receiving frames from CMOS sensor (in the
developed endoscopic capsule OmniVision OV7675
and OV7690 sensors are used);
Scaling frames;
Compressing frames;
Storing frames into ASIC buffer;
Transmitting frames to the radiofrequency
transceiver (in the developed endoscopic capsule
Nordic NRF solution is used);
Operate the LED flash (using PWM GPIO (GeneralPurpose Input/Output));
Control the sensor (via SCCB interface);
Control the transceiver (via SPI interface);
The DCMI and SCCB interface should be compatible
with OV7675 and OV7690 sensors;
The SPI should operate up to 4MHz.
If the ASIC microcircuit will have the general-purpose
programmable outputs this will expand its functionality
(LED notification, camera and transceiver preset, etc.).
To achieve the requirements mentioned above it is
proposed to replace the most energy-intensive path of the
wireless endoscopic capsule for the component of our
own design. Matrix and the transceiver are not replaced
intentionally due to the fact that the CMOS sensors are
quickly updated (energy consumption is reduced, the
sensitivity is increased and the image becomes clearer and
sharp).
In case the microcircuit implements DCMI interface
that is standard for CMOS matrix the matrix may be
replaced if necessary.
If the controller supports the standard interfaces of
communication between chips (UART, I2C or SPI) and has
programmable general-purpose outputs the wireless
endoscopic capsule will be able to work with any
industrial radio transceiver. So, to increase the flexibility
of the integrated circuit it is proposed to include UART
and I2C interfaces.
To add the programmable logic, it is worth adding
8051 core to operate the GPIO, control the sensor,
transceiver and LEDs.
Radio Transceiver Choice: According to the international
law, implantable medical devices should comply with the
standard for the Medical Implantable Communications
Service, MICS [10]. MICS is the unlicensed standard with
very low power that is used for data transmission to
perform diagnostic or therapeutic functions associated
with the implantable medical devices. MICS has
limitations in the frequency range 402-405 MHz and a
signal output 25 mW.
However, capsule endoscopes do not have to meet
the MICS standard because MICS as a standard was
developed for devices which stay in human body for a
long time (10-15 years) such as cardiac pacemakers,
implantable defibrillators, hearing aids, automatic drug
delivery systems [19]. In turn, the wireless endoscopic
capsule stays in human body for not more than 48 hours
and the capsule radiation is significantly smaller than the
radiation received from external sources such as Wi-Fi, so
it can be ignored. Moreover, 25 mW provide a very low
data rate, so it is inappropriate to stick to MICS in the
process of the endoscopic capsule development.
In this regard, the standard device of ISM (Industrial,
Scientific and Medical) band will be used. ISM that does
not require licensing opens for use at certain powers
frequencies 433 MHz, 868 MHz, 902 MHz and 2.4 GHz. [20]
The most common wireless standards of ISM band are
Bluetooth, Wi-Fi, 802.15.4 and Zigbee. Standards can
have an open protocol such as Bluetooth and Wi-Fi, or be
proprietary (closed-protocol).
The only requirement for products development in
ISM-band is compliance with standards set by regulatory
authorities for a particular part of frequency spectrum.
The rules are different in different countries. In the United
States the rules are set by the Federal Communication
Commission, FCC and in Europe – by the European
Telecommunications Standards Institute, ETSI. [21]
Consider the specifications of radio transceivers
operating on proprietary protocols of Texas Instruments
Company – Chipcon technology (Chip Connect, ÑÑ) and
Nordic (NRF), as well as on open protocols Wi-Fi and
Bluetooth (Table 1) [22-25].
As the Table 1 shows, the radio transceivers that
exchange data via Bluetooth and Wi-Fi protocols have
high energy consumption so are not considered for use in
the wireless endoscopic capsule of second generation.
The Chipcon and NRF devices have similar
specifications but for the endoscopic capsule radio
transceiver NRF has been selected, as it has a higher
speed of data transfer with similar energy consumption as
Chipcon.
Power Estimation
Calculation Method: Power estimation for ASIC has been
performed using the power information of its components.
The following memories have been considered:
MCU8051:
256 Byte internal data memory (DATA_MEM, single
port);
1965
World Appl. Sci. J., 32 (9): 1962-1969, 2014
Table 1: Comparison of the specifications of different ISM radio transceivers
NRF
CC
Wi-Fi
Bluetooth
Speed
Upto 1 Mbit/s
Up to 76,8 kbit/s
Up to 2 Mbit/s
Up to 4 Mbit/s
Current, mAReception Transmission
Voltage, V
Size, mm
10.518
1.9 - 3.6
33x16x11
9.18.6
2.7-3.6
-
38120
3.0 - 3.7
27x18x3.1
10.518
1.7 - 4.8
-
Table 2: Different variants for PCLK and MCUCLK
PCLK [MHz]
MCUCLK [MHz]
Peak power [mW]
Pavg [mW]
Average Idd [mA]
Peak Idd [mA]
24
12
12
20
20
8
113.47
56.74
56.74
12.23
12.23
12.23
6.8
6.8
6.8
63.04
31.52
31.52
512 Byte data memory (DATA_MEMx, single port);
4 kByte program memory (PROG_MEMx, single port);
64kByte frame buffer (FRAME_BUFFER_MEM, dual
port).
correstion factor =
2kByte LUMI_MEM;
2kByte CHROM_MEM.
Data Cycle and Average Power Consumption: The whole
data cycle from data acquisition to read out of the frame
buffer by the MCU can be divided into three phases
(Figure 3):
JPEG Encoder:
16kByte R2B_MEM;
320Byte R2B_OFIFO0_MEM+ R2B_OFIFO1_MEM;
272Byte DCT0_MEM + DCT1_MEM;
256Byte VLE_HUFFM_TAB0_MEM;
256Byte VLE_HUFFM_TAB1_MEM;
128Byte QUANT_TAB0_MEM;
128Byte QUANT_TAB1_MEM;
768Byte VLC_FIFO0_MEM+VLC_FIFO1_MEM.
Phase 1: Transfer of image data from sensor to the scaler,
scaling, compression and storing the compressed data
into the frame buffer. The following components of the
ASIC are active during this phase: scaler with memories,
JPEG encoder with memories and frame buffer. The
preferred pixel clock is 12 MHz.
Phase 2: Read out of the frame buffer by MCU and
transfer data to the RF-Chip by SPI. The following
components of the ASIC are active during this phase:
frame buffer, MCU with memories. The preferred MCU
clock is 20 MHz.
MCU8051 has 3.9mW@8MHz for a 0.18µm CMOS
technology.
The dynamic power consumption for a NAND2
(1 gate equivalent) is 0.02232 µW/MHz. The power
consumption has been calculated using the following
formula:
(1)
where PGE – power for one NAND2 equivalent; GC – gate
count of IP; fclk – clock frequency.
The leakage power is in the range of pW and has
been neglected.
At first the power consumption of the MCU8051 has
been calculated using formula (1). Then a correction factor
has been calculated using the calculated power
consumption for the MCU and the power consumption
specified in the documentation:
(2)
The correction factor is 0.72. It means that the
calculated power using the gate-count-method is too
high. So the power consumption for the scaler and the
JPEG encoder has been calculated using the gate-countmethod and corrected by the factor 0.72.
Scaler:
PIP = PGE * GC * fclk
Pdoc
Pcalculated
Phase 3: Scaler, encoder, frame buffer and MCU are in
power-down state.
Because of the small data transmission rate of the RFChip (only 2 Mbit/s) the maximum possible frame rate is 4
fps. Therefore, the shortest duration of one frame is
250ms.
During data processing the digital part of the ASIC
consumes power only in phase 1 and phase 2. Therefore,
the average power consumption is less than the peak
power consumption in the active phases. Nevertheless,
the clock frequencies for both phases should be as low as
possible to reduce the peak power in the active phases,
because the integrated voltage regulator has to support
this peak power.
1966
World Appl. Sci. J., 32 (9): 1962-1969, 2014
Fig. 3: Phases of the data cycle
The power consumption has been calculated for each phase separately using the assigned clock frequency. The
resulting power consumption of the digital part has been calculated by averaging the separate power values over the
duration of one frame.
where FB – byte count of frame buffer; PCLK – pixel
clock; CR – compression ratio; MCUCLK – clock for the
µC; tframe – duration of one frame.
The compression ratio CR is calculated using the
following formula:
(6)
The amount of input data is 614400 bytes for one
image with resolution of 640 x 480 pixels, because one
color pixel needs two byte (YCbCr). The necessary
compression ratio is 9.375 in order to prevent frame buffer
overflow.
Using a pixel clock (PCLK) of 12 MHz, a MCU clock
(MCUCLK) of 20 MHz this calculation results in a
power consumption of the digital part of 12.23 mW.
The average operating current of the digital core will be
6.8mA at a core voltage of 1.8V. The current consumption
of the analogue part has been neglected. The peak power
occurs at phase1 with 56.54 mW. Therefore, the voltage
regulator has to support a minimum current of about
32 mA.
The following table shows some variants for different
values of PCLK and MCUCLK.
will improve the specifications of the wireless capsule
endoscope of second generation being developed in
terms of the size of recorded images and their
transmission (for the same time the larger size frame can
be transmitted) and reduce energy consumption
increasing the duration of the capsule operation.
The studies on further ASIC testing as well as
capsule components` construction, reliability and
functionality improvements are underway.
REFERENCES
1.
2.
3.
4.
CONCLUSIONS
The
proposed
implementation
of
ASIC
microcontroller in combination with NRF radio transceiver
1967
Faigel, D.O. and D.R. Cave, 2008. Capsule
Endoscopy.
Saunders Elsevier, Amsterdam,
Netherlands.
Moglia, A. Menciassi and P. Dario, 2008.
Recent patents on wireless capsule endoscopy.
Recent Patents on Biomedical Engineering,
1(1): 24-33.
Swain, P., 2008. The future of wireless capsule
endoscopy. World Journal of Gastroenterology,
14(26): 4142-4145.
Kukushkin Alexander, Mikhaylov Dmitry, Zhukov
Igor, Starikovski Andrey, Konev Vladimir, Tolstaya
Anastasia, Khabibullin Timur, 2013. Recognition of
polyps in the images of wireless capsule endoscopy.
International Journal of Emerging Trends &
Technology in Computer Science (IJETTCS),
2(2): 485-488.
World Appl. Sci. J., 32 (9): 1962-1969, 2014
5.
6.
7.
8.
9.
10.
11.
12.
13.
Zhukov, I.Yu., D.M. Mikhaylov, V.N. Konev,
E.D. Fedorov, E.V. Ivanova and T.R. Khabibullin,
2013. Wireless controlled endoscopic capsule for
gastrointestinal tract examination. Scientific and
technical Journal “Specialized machinery and
communication”, 3. Moscow 2013, pp: 49-53.
Sychev, N.V., J.A. Kazymova and A.V. Trifonenkov,
2013. The components of the developed capsule
endoscopic complex “Landish” of the second
generation and the requirements applied to its
reliability. Scientific and technical Journal
“Specialized Machinery and Communication”,
4: 18-20.
Aristov, M.S., S.I. Sivkov, D.A. Vaganov, A.K.
Krevskiy and E.A. Koshevnikov, 2013. Development
of the capsule of the endoscopic complex Landish
of the second generation. Scientific and technical
Journal “Specialized machinery and communication”,
4. Moscow 2013, pp: 21-24.
Froimson, M.I., Î.I. Shishin and A.S. Pronichkin, 2013.
Development of the reader of the endoscopic
complex “Landish” of the second generation.
Scientific and technical Journal “Specialized
machinery and communication”, ¹4. Moscow 2013,
pp: 25-27.
Dave Litwiller, 2001. CCD vs CMOS: facts and
fiction , January 2001 issue of PHOTONICS
SPECTRA © Laurin Publishing Co. Inc.
Electronic Shuttering, 2011. Motion Video Products,
2011-12-22. URL: http://www.motionvideoproducts.
com/MVP%20papers/Rolling%20vs%20Global%20
Shutter.pdf.
Zhukov, I., E. Fedorov, D. Mikhaylov, E. Ivanova, A.
Kukushkin, A. Starikovski and A. Tolstaya, 2012.
General Research on Automatic Image Shrinking in
the Wireless Capsule Endoscopy. International
Journal of Image, Graphics and Signal Processing
(IJIGSP), ISSN Print: 2074-9074, ISSN Online: 20749082, 4(3), April 2012.
Tareq, Hasan Khan and Khan A. Wahid, 2011.
Lossless and Low-Power Image Compressor for
Wireless Capsule Endoscopy. VLSI Design, Volume
2011 (2011), Article ID 343787, 12 pages. URL:
http://www.hindawi.com/journals/vlsi/2011/343787.
Wahid, K., S.B. Ko and D. Teng, 2008. Efficient
hardware implementation of an image compressor for
wireless
capsule
endoscopy
applications.
Proceedings of the International Joint Conference on
Neural Networks, June 2008. pp: 2761-2765.
14. Xiao Wen Li, Xinkai Chen, Xiang Xie, Guolin Li, Li
Zhang and Zhihua Wang, 2007. Pre-processing and
vector quantization based approach for CFA data
compression in wireless endoscopy capsule. 4th
IEEE International Symposium on Biomedical
Imaging: From Nano to Macro, 2007. ISBI 2007,
pp: 1172-1175.
15. Xiao Wen Li, Xiang Xie, Xinkai Chen, Guolin Li, Li
Zhang, Zhihua Wang and Hong Chen, 2007. Design
and Implementation of a Low Complexity Nearlossless Image Compression Method for Wireless
Endoscopy Capsule System. IEEE International
Symposium on Circuits and Systems, 2007. ISCAS
2007, pp: 1321-1324.
16. Turcza, P. and M. Duplaga, 2007. Low-power image
compression for wireless capsule endoscopy.
Proceedings of the IEEE International Workshop
on Imaging Systems and Techniques, May 2007,
pp: 1-4.
17. Lin, M., L.R. Dung and P.K. Weng, 2006. An ultralow-power image compressor for capsule endoscope.
BioMedical Engineering Online, 5(14): 1-8.
18. STM32F205xx and STM32F207xx Specification
Datasheet, 2012. URL: http://www.st.com/web/en/
resource/technical/document/datasheet/CD0023739
1.pdf.
19. Savci, H.S., A. Sula, Zheng Wang, N.S. Dogan and
E. Arvas, 2005. MICS transceivers: regulatory
standards and applications [medical implant
communications service]. SoutheastCon, 2005.
Proceedings. IEEE. URL: http://ieeexplore.ieee.org/
xpl/articleDetails.jsp?tp=&arnumber=1423241&quer
yText%3DMICS+Transceivers%3A+Regulatory+St
andards+and+Applications+%5BMedical+Implant+
Communications+Service%5D.
20. Batler, P., A. Sotnikov and O. Kharni, 2008. Intelligent
wireless modem for reliable wireless data
transmission in ISM bands (433 MHz, 868 MHz and
902 MHz). Journal “Wireless technologies” ¹2, 2008.
URL: http://www.wireless-e.ru/articles/diapasons/
2008_2_32.php.
21. Petrov, D., 2010. Wireless communication standards
of ISM band. Journal “Electronic Components” ¹10,
2010. P. 28-32. URL: http://www.russianelectronics.
ru/leader-r/review/2187/doc/53409.
22. Wi-Fi module. Specification RN-171 802.11 b/g
Wireless LAN Module. Version 3.21 10/2/2012, URL:
http://ww1.microchip.com/downloads/en/DeviceDo
c/rn-171-ds-v3.21.pdf.
1968
World Appl. Sci. J., 32 (9): 1962-1969, 2014
23. Bluetooth module. Specification CC2560 Bluetooth.
2010.
URL:
http://www.ti.com/pdfs/wtbu/
cc2560_slyt377.pdf.
24. Chipcon module. Specification CC1010, Single Chip
Very Low Power RF Transceiver with 8051Compatible
Microcontroller.
PRELIMINARY
Datasheet
(rev.
1.0)
2002-09-18,
URL:
http://www.keil.com/dd/docs/datashts/chipcon/cc1
010_ds.pdf.
25. NRF module. Specification Single chip 2.4 GHz
Transceiver nRF2401. Revision: 1.1, June 2004. URL:
http://www.keydemon.com/files/nRF2401.pdf.
1969
Download