Air Quality Monitoring Systems Continuous monitoring of ambient Nitrogen Dioxide AS32M Analyzer The European Directive 2008/50/EC requires the measurement of nitrogen dioxide (NO2) in all urban areas of more than 100 000 inhabitants. Currently, sites located in the vicinity of roads, directly impacted by vehicle emissions are most likely to experience poor air quality. Large touch-screen with animated real-time synoptic flow diagram optionally Designed to alleviate potential disadvantages* of the chemiluminiscence method and to provide the most accurate value for NO2 concentration measurements, the AS32M analyzer provide fast and accurate measurements. EXCLUSIVE FEATURES: Accurate, precise and direct continuous monitoring of ambient NO2 Innovative technology plus touchscreen convenience (optionally) Visible (450 nm) absorption measurement using patented Cavity Attenuated Phase Shift (CAPS) technology Direct measurement of the sample - no chemical conversion required Measurement of ambient concentrations to 1000 ppbv Essentially interference-free Insensitive to presence of varying levels of nitric oxide, aerosols, humidity and other trace atmospheric species No Toxic Gas Emissions Real-time synoptic flow diagram display Extremely compact, easy to use with minimal maintenance (periodic change of particle filter) Linear Response (0-1000 ppbv) Built-in USB port and serial interface (RS 232 / RS 422) Full remote emulation of the analyzer US EPA equivalent method applied NEW: on board web server compatible with any internet browser. Es@cloud™ user interface with on-line help for the display, configuration, maintenance, diagnostics or software updating of the analyser, remotely, from any PC, tablet or iPhone. APPLICATIONS : es@cloud™ TCP/IP remote control with dynamic, multilingual interface, featuring intuitive navigation by pictograms. Ambient air monitoring Indoor air quality continuous monitoring Mobile and fixe monitoring laboratories Traffic stations NO2 measurement according to the European Directive 2008/50/CE www.environnement-sa.com Direct NO2 measurement Ambient Direct Nitrogen Dioxides Analyzer AS32M PRINCIPLE OF OPERATION : Repetability standard deviation at zero: 0.05 ppb Repetability standard deviation at a level of the hourly limit value (200µg/m3 ): 0.2 ppb Short zero drift: 0.75 ppb (24h) Short span drift: 1.5 ppb Response time (rise): 16 s Response time (fall): 16 s Difference betwen rise time and fall time: <1s Lack of fit (linearity)(residual from the linear regression function): <4% Lack of fit (linearity) at zero (residual at zero): 0.6 ppb (24h) Sensitivity coefficient of surrounding temperature (0 to 30°C): <0.2 ppb/K Authonomy: up to one year of NO2 measurements (on 1/4 h sampling rate) Valves block for selection of external zero and span gas Sample dryer Ethernet network connection Integrated web-server with full remote emulation of the analyser The Cavity Attenuated Phase Shift nitrogen dioxide (AS32M) monitor as deployed here operates as an optical absorption spectrometer, utilizing a blue light-emitting diode (LED) as a light source, a sample cell incorporating two high reflectivity mirrors centered at 450 nm and a vacuum phototube detector. Its efficacy is based on the fact that nitrogen dioxide (NO2) is a broadband absorber of light in the visible region of the spectrum. The configuration of the monitor is quite simple and is shown below in the schematic form. The air sample enters the monitor and passes through a disposable filter cartridge which removes all particulates in order to prevent contamination of the mirror. The sample then proceeds through PFA tubing into a stainless steel sample cell at one end and out the other through tubing where it is directed to a 1 l/min diaphragm pump. The high reflectivity mirrors are directly attached to the ends of the sample cell, forming the optical cavity which provides for the concentration measurement, The LED, filter and appropriate focusing optics are attached directly to the sample cell. The light emanating from the cell is directed into a vacuum phototube where the resultant signal is integrated, digitized and sent to the internal computer where all subsequent data processing takes place. The sample cell contains both pressure and temperature sensors which allow for both accurate correction of the nitrogen dioxide absorption coefficient and baseline subtraction. The pressure reading is also used to check on the proper operation of the particle filter. Sample Echantillon 450 Drier Phototube Filtre optique Bandpass filter Vacuum (détecteur) Phototube Pompe Pump 10 nm OPTIONS: Large touch screen permitting one-touch control of the system as well as pre-configuration of analysis tasks ESTEL electronic board (1 or 2) with: 4 independent analog inputs 4 independent analog outputs 4 remote control inputs 6 dry contacts outputs SOREL electronic board with: 4 dry contacts outputs 4 dry contacts inputs Built-in permeation bench with NO2 tube Digital Electronics Mélange Mixer fréquence + PreAmp Préampli *The chemiluminescence - reference method for the measurement of nitrogen dioxide (NO2) and oxides of nitrogen (NOx) is based on the reaction of nitric oxide (NO) with Ozone (O3) and requires high maintenance and has some negative aspects. Potential errors can arise when measuring the concentrations of NO and NO2 depending on the duration of the measurement cycle. If part of the light energy produced by chemiluminescence is lost in collisions with other molecules present in the sample, such as moisture, under-estimation of the sample may occur. Conversely, a risk of overestimating of NO2 concentrations if other species are converted to NO2 in the NO catalytic oven can occur. Finally there is a degree of error in the conversion efficiency of the furnace. Standard version: analyzer equipped with LCD screen 111 BD Robespierre 78304 POISSY - France Tel. : +33(0)1 39 22 38 00 Web: www.environnement-sa.com E-mail: info@environnement-sa.com Distributed by: Preliminary version for imediate release. Typical spécifications subject to changes without prior notice. Measurement Range: 0-1 ppm Detection limit (2s): 0.1 ppb AS32M-EN_PSO - InDes -06/12 SPECIFICATIONS: