A WIRELESS ELECTRICAL STIMULATION SYSTEM FOR WOUND HEALING THERAPY WITH BIPHASIC HIGH-VOLTAGE PULSED CURRENT OUTPUT by DANIEL STEVEN HOWE Submitted in partial fulfillment of the requirements For the degree of Doctor of Philosophy Dissertation Adviser: Dr. Steven Garverick Department of Electrical Engineering and Computer Science CASE WESTERN RESERVE UNIVERSITY May, 2013 CASE WESTERN RESERVE UNIVERSITY SCHOOL OF GRADUATE STUDIES We hereby approve the thesis/dissertation of Daniel Steven Howe candidate for the Doctor of Philosophy (signed) Steven L. Garverick degree *. (chair of the committee) Kath M. Bogie Pedram Mohseni Christian A. Zorman (date) 03/15/2013 *We also certify that written approval has been obtained for any proprietary material contained therein. ii To my wife and family. iii Contents List of Tables .............................................................................................................. 10 List of Figures ............................................................................................................. 11 Acknowledgements ..................................................................................................... 16 List of Abbreviations and Common Signal Names .................................................... 19 Abstract ....................................................................................................................... 21 1 Introduction ........................................................................................................ 23 1.1 Clinical Motivation ..................................................................................... 23 1.2 Electrotherapy and Wound Healing ............................................................ 24 1.3 Previous Wound Healing Electrotherapy Studies ....................................... 25 1.3.1 Current research limitations .................................................................... 26 1.4 Existing electrical stimulation therapies for wound care ............................ 27 1.4.1 Microcurrent devices ............................................................................... 27 1.4.2 TENS/ pain management devices ........................................................... 28 1.4.3 Research Device: DS4 ............................................................................. 28 1.5 Previous work ............................................................................................. 30 1.5.1 Surface Stimulation Device (SSD) .......................................................... 30 1.6 2 Research Objective ..................................................................................... 31 Modular Surface Stimulator for Rodent Studies (MSS) .................................... 33 2.1 Research Study Overview ........................................................................... 33 iv 2.1.1 Battery review ......................................................................................... 36 2.1.2 VDD Step-up converter ........................................................................... 37 2.1.3 Microcontroller........................................................................................ 38 2.2 Boost Converter Circuit .............................................................................. 38 2.2.1 Battery Limitations .................................................................................. 39 2.2.2 Operational Considerations ..................................................................... 39 2.2.3 Critical Component Selection ................................................................. 40 2.2.4 Output capability ..................................................................................... 41 2.2.5 Controller firmware ................................................................................. 42 2.3 Stimulator Circuits ...................................................................................... 45 2.4 Experimental Results .................................................................................. 46 2.4.1 Modular Surface Stimulator Version 1(MSS) ......................................... 47 2.4.2 Stimulator Performance........................................................................... 48 2.4.3 Boost converter efficiency ...................................................................... 50 2.5 3 Summary ..................................................................................................... 54 Application-Specific Integrated Circuit ............................................................. 55 3.1 Functional Requirements ............................................................................ 55 3.2 Foundry Process Features ........................................................................... 56 3.3 High-Voltage Boost Converter ................................................................... 56 3.3.1 Step-up Topology .................................................................................... 57 v 3.3.2 Control loop operation ............................................................................ 59 3.3.3 Inductor current sense ............................................................................. 61 3.3.4 Current sense amplifier ........................................................................... 64 3.3.5 Current sense comparator ........................................................................ 65 3.3.6 NMOS Switch ......................................................................................... 69 3.3.7 Output Rectifier ....................................................................................... 69 3.3.8 Output capacitor ...................................................................................... 70 3.3.9 Output Voltage Sense .............................................................................. 71 3.4 Stimulator .................................................................................................... 72 3.4.1 Architecture ............................................................................................. 72 3.4.2 High-side switches .................................................................................. 73 3.4.3 Low-side DAC pair ................................................................................. 77 3.5 Support circuits ........................................................................................... 80 3.5.1 Bias generation and distribution .............................................................. 80 3.5.2 SPI Serial Port ......................................................................................... 81 3.5.3 Pad Frame, ESD Protection, and Guard rings ......................................... 82 3.6 ASIC Test Results ....................................................................................... 83 3.6.1 Introduction ............................................................................................. 83 3.6.2 MSS ASIC Test Setup ............................................................................. 85 3.6.3 Integrated Circuit Edits Using Focused Ion Beam .................................. 85 vi 3.6.4 Bias circuit............................................................................................... 86 3.6.5 Boost converter operation ....................................................................... 87 3.6.6 Stimulator ................................................................................................ 89 3.6.7 Summary ................................................................................................. 95 4 Wireless Stimulation Bandage and System for a Large Animal Model ............ 97 4.1 Large Infected Wound Study ...................................................................... 97 4.2 Bandage Substrate ..................................................................................... 100 4.2.1 Bandage Electrode characterization ...................................................... 101 4.3 Battery Selection ....................................................................................... 103 4.4 Stimulator Module PCB ............................................................................ 104 4.4.1 Wireless communication protocol and processor selection .................. 104 4.5 Firmware ................................................................................................... 105 4.5.1 Stimulation State machine ..................................................................... 106 4.5.2 Bluetooth Interface ................................................................................ 109 4.6 Device assembly ....................................................................................... 111 4.6.1 MSS3 Stimulator Module Assembly..................................................... 111 4.6.2 Device Sterilization ............................................................................... 113 4.6.3 Intra-Operative Bandage Assembly ...................................................... 114 4.7 Wireless Base Station ............................................................................... 115 4.7.1 Base station construction ....................................................................... 116 vii 4.7.2 Base station operation ........................................................................... 118 4.8 Benchtop Testing ...................................................................................... 119 4.8.1 ASIC validation ..................................................................................... 120 4.8.2 Wireless Communication ...................................................................... 123 4.8.3 Operating current test ............................................................................ 124 4.9 Clinical Validation .................................................................................... 124 4.9.1 Method .................................................................................................. 124 4.9.2 Bandage conductivity test ..................................................................... 125 4.9.3 Acute stimulation test ............................................................................ 126 4.10 5 Summary ................................................................................................... 128 Conclusions and Future Work .......................................................................... 129 5.1 Achievements ............................................................................................ 129 5.2 Future work ............................................................................................... 131 Appendix A: Supplement Information for Chapter 2 (MSS2)................................. 133 A.1: MSS2 Schematic ........................................................................................... 133 Appendix B: Supplemental Information for Chapter 3 (ASIC) ................................ 134 B.1 MSS ASIC Test Fixture Schematic ................................................................ 134 Appendix C: Supplemental Information for Chapter 4 (MSS3) ............................... 135 C.1: MSS3 Module Schematic .............................................................................. 135 C.2 Stimulation Bandage GATT profile ............................................................... 136 viii C.3 MSS3 Parameter programming procedure using TI BLE Device Monitor .... 137 C.4 Base station Arduino programming procedure ............................................... 139 C.5 Base station BLE112 programming procedure .............................................. 140 C.6 MSS3 CC2541 programming procedure ........................................................ 141 References ................................................................................................................. 142 ix List of Tables Table 1-1: Specifications of a portable electrical stimulation unit. ............................ 29 Table 2-1: Design requirements for small-animal device ........................................... 35 Table 2-2: Review of battery chemistries in coin/button form factor ......................... 36 Table 3-1. Functional requirements for ASIC. ........................................................... 55 Table 3-2. Transistor sizes for current sense comparator circuit. ............................... 66 Table 3-3. Comparison of on-chip and off-chip boost converter switch components. 69 Table 3-4. Key characteristics of the BAT46W schottky diode [35]. ........................ 70 Table 3-5. Device sizes for bias circuit. ...................................................................... 81 Table 3-6. List of MSS ASIC bond pads and their function....................................... 84 Table 3-7: Comparison of simulated and measured bias circuit voltages. ................. 86 Table 3-8: Measured performance of the boost converter current sense circuits. ...... 88 Table 3-9: Summary of MSS ASIC circuit functionality ........................................... 96 Table 4-1: Clinical Requirements for Infected Wound Study Device ........................ 98 Table 4-2: Survey of high-capacity batteries for long-duration devices................... 103 Table 4-3: A review of microcontrollers with built-in 2.4 GHz radios. ................... 105 Table 4-4: Stimulator control parameters available through Bluetooth wireless link. ......................................................................................................................................... 111 Table 4-5: Stimulation Parameters used in MSS3 stimulator demonstration. .......... 122 Table 4-6: Measured operating current and projected device lifetime for several compliance voltage levels. .............................................................................................. 124 Table 5-1: Electrotherapy Device Summary............................................................. 131 x List of Figures Figure 1-1. Example of a chronic wound on a foot [7]. .............................................. 23 Figure 1-2: The Vomaris Procellera(R) microcurrent wound dressing. ....................... 27 Figure 1-3. Example of a TENS portable stimulator (from [15]). .............................. 28 Figure 1-4: DigiTimer Biphasic stimulator for benchtop testing................................ 29 Figure 1-5: Surface Stimulation Device (SSD), from [16]. ........................................ 30 Figure 1-6: SSD attached to rabbit ear ........................................................................ 31 Figure 2-1: Block diagram of the proposed MSS consisting of a flexible printed bandage and a reusable stimulator module. ...................................................................... 33 Figure 2-2: Rat wound model used with MSS ............................................................ 34 Figure 2-3. Graphical representation of the programmable stimulation waveform parameters that cover a range of potentially therapeutic levels. ....................................... 35 Figure 2-4: Comparison of battery energy density for rechargeable and nonrechargeable chemistries (from [19]). ............................................................................... 37 Figure 2-5: Schematic diagram of the boost converter circuit using discrete components for HV circuitry and microcontroller for the control loop............................ 40 Figure 2-6: Plot of the charging time vs. output voltage setpoints for several battery voltages. ............................................................................................................................ 44 Figure 2-7. Schematic diagram showing the high-voltage stimulator circuit. ............ 45 Fig. 2-8. Schematic diagram of the high-side driver that controls PMOS switches with a constant VSG across a 90 V supply range. ...................................................................... 46 Figure 2-9: Modular Surface Stimulator (MSS1) Version 1 Device .......................... 47 xi Fig. 2-10. Plot of output pulses for several load resistors. .......................................... 49 Figure 2-11: Measured maximum pulse frequency vs. voltage setpoint with various supply voltages.................................................................................................................. 50 Figure 2-12. Measured MSS1 Boost Converter Charging Efficiency for Selected Supply Voltages. ............................................................................................................... 51 Figure 2-13: Measured MSS stimulator efficiency for selected boost capacitor voltages. ............................................................................................................................ 52 Fig. 2-14. Measured battery lifetime versus stimulation voltage for selected output resistances. Pulse rate is 25 Hz, pulse width is 200 µs, and stimulation interval is 10 min/hr. ............................................................................................................................... 53 Figure 3-1: NDMOS floating transistor in I2T100 process (modified from [26]). .... 56 Figure 3-2: Dickson Charge Pump Voltage Conversion Technique (from [27]) ....... 57 Figure 3-3. Representative schematic of the flyback converter architecture. ............. 58 Figure 3-4. Boost converter topology. ........................................................................ 59 Figure 3-5. Idealized inner-loop current control for the boost converter. .................. 60 Figure 3-6. MSS ASIC-based boost converter circuit. ............................................... 61 Figure 3-7. Inductor current sense method across shorting NMOS............................ 62 Figure 3-8. Inductor current sense method using battery-side PMOS. ....................... 63 Figure 3-9. Inductor current sense using off-chip resistor. ......................................... 63 Figure 3-10. Differential Amplifier used as inductor current sense amplifier. ........... 64 Figure 3-11. Schematic diagram of current sense comparator circuit. ....................... 66 Figure 3-12. Schematic diagram of boost converter output voltage sense circuit. ..... 72 Figure 3-13. High-level architecture of the stimulator circuit. ................................... 73 xii Figure 3-14. Stimulator High-side circuits. ................................................................ 74 Figure 3-15. Schematic diagram of floating CMOS reference circuit. ....................... 76 Figure 3-16. High-voltage level shift circuit............................................................... 77 Figure 3-17. Stimulator low-side DAC. ...................................................................... 78 Figure 3-18. Schematic diagram of current DAC. ...................................................... 79 Figure 3-19. Schematic diagram of folded-cascode amplifier used in stimulator circuits. .............................................................................................................................. 80 Figure 3-20. Schematic diagram of bias voltage generation circuit. .......................... 81 Figure 3-21. MSS SPI Serial Port Configuration. ...................................................... 82 Figure 3-22: Die microphotograph of MSS ASIC. ..................................................... 83 Figure 3-23. The MSS ASIC wirebonded in QFN40 surface mount package............ 84 Figure 3-24. ASIC test PCB with QFN-packaged DUT . ........................................... 85 Figure 3-25. Example of circuit edit using focused ion beam (FIB). ......................... 86 Figure 3-26: Test setup for differential amplifier and comparator circuits. ............... 87 Figure 3-27: MSS ASIC Boost Converter DMOS switching behavior with 50-ohm load.................................................................................................................................... 88 Figure 3-28: Boost converter charging performance using on-chip DMOS switch. .. 89 Figure 3-29: Test setup used to measure negative voltage regulator circuit............... 90 Figure 3-30: Negative voltage regulator output for HVDD between 5 V and 25 V. .. 91 Figure 3-31: Stimulator High-side PMOS test circuit ................................................ 91 Figure 3-32: Switching behavior of stimulator high-side PMOSFET with HVDD = 20 V, f = 1 kHz, and 500 Ω resistor to GND. ........................................................................ 92 Figure 3-33: ASIC low-side stimulator test circuit. .................................................... 93 xiii Figure 3-34: Stimulator low-side IDAC current transfer characteristic with HVDD= 5 V. ..................................................................................................................... 93 Figure 3-35: Measured IDAC_SENSE voltage driving gate of NMOS cascode across IDAC code range with HVDD =5 V................................................................................. 94 Figure 3-36: Stimulator compliance voltage sensitivity for several HVDD values and RL = 100 Ω. ....................................................................................................................... 95 Figure 4-1: Approximate wound locations on porcine model for infected wound study (2 wounds per side) [45]. .................................................................................................. 98 Figure 4-2: Stimulation bandage concept for large-wound studies. ......................... 100 Figure 4-3. Top (left) and bottom (right) views of bandage substrate ...................... 101 Figure 4-4: Test fixture and setup for substrate verification..................................... 101 Figure 4-5: Measured bandage substrate electrode impedance. ............................... 102 Figure 4-6: Block diagram of MSS3 Stimulator PCB .............................................. 104 Figure 4-7: Block diagram of boost converter controller. ........................................ 107 Figure 4-8: Generation of stimulation pulse timing signals using hardware timers . 108 Figure 4-9: Bluetooth Protocol Stack (from [57]) .................................................... 109 Figure 4-10: Communication scheme for transferring parameters and measurements between base station and bandages. ................................................................................ 110 Figure 4-11: MSS3 microcontroller programmer connection using PCB probe. ..... 111 Figure 4-12: Snap connectors mounted on bottom side of MSS3 PCB module....... 112 Figure 4-13: MSS3 Stimulator Module PCB Assembled with Battery. ................... 113 Figure 4-14: Bandage substrates packaged for sterilization. .................................... 114 xiv Figure 4-15: Communication network between stimulator modules, base station, and internet data server. ......................................................................................................... 116 Figure 4-16: Wireless base station hardware PCB stack. ......................................... 117 Figure 4-17: Procedure for programming bandage stimulation parameters. ............ 118 Figure 4-18. MSS3 printed circuit board .................................................................. 120 Figure 4-19: Battery-powered MSS3 boost converter output charging waveform (VBATT = 4.15 V) .............................................................................................................. 121 Figure 4-20: Demonstration of the MSS3 stimulator delivering a biphasic pulse. ... 122 Figure 4-21: Self-reported received signal strength (RSSI) from CC2541 radio as a function of distance from USB dongle. .......................................................................... 123 Figure 4-22: Photograph of bandage with adhesive hydrocolloid material. ............. 125 Figure 4-23: Measured MSS3 substrate electrode impedance measured in an acute test on living pig skin. ..................................................................................................... 126 Figure 4-24: Acute test schematic for measuring current flowing through tissue. ... 127 Figure 4-25: Electrode voltage and current waveforms measured in acute stimulation test. .................................................................................................................................. 128 xv Acknowledgements I would like to thank my research advisor Dr. Garverick for his guidance and instruction during both my undergraduate and graduate education at CWRU. I have enjoyed the opportunity to design circuits at every level from the transistor to the embedded system for both industrial and medical applications, and I likely would not have had such a diverse experience elsewhere. I appreciate the respect he shows each student and the encouragement he provides through the toughest challenges. I could not have completed this program without Dr. Bogie, my biomedical research advisor. She not only obtained the funding to carry out this research, but she granted me the opportunity to extend my biomedical training far beyond that of a traditional electrical engineer. I gained invaluable hands-on research experience by participating in the animal experiments that will help me design better medical devices in the future. She has been advisor both personally and professionally, and I truly appreciate all she has done for me. I also owe a debt of gratitude to committee members Dr. Zorman and Dr. Mohseni for their valuable guidance in the design and testing of the devices presented in this work. Dr. Zorman also provided me access to his lab and equipment to manufacture numerous bandages for the animal studies. Dr. Mohseni has funded and maintained the Cadence EDA server used to design integrated circuits for all the design groups in the department, and without his support my work could not have continued. I also thank Dr. Saab and xvi James Cox for providing me technical support and access to the Mentor Graphics EDA tools used to complete the design verification steps of the ASIC design. Jeremy Dunning has been instrumental to the success of the research by developing fabrication methods for the substrates, and he has spent countless hours manufacturing them with me. I also thank Jen Graebert for the practical training in animal care and her friendship during many long surgery days and experiments. I have also learned a lot from other members of the PES study team, including Dr. Kristi Henzel, Bruce Kinley, Danli Lin, and the staffs of the VA ARF and CWRU ARC facilities. I thank the staff of the APT Center for maintaining an organization that supports collaborations between clinicians and researchers from a wide range of technical backgrounds and diverse organizations. In particular, I thank Ron Triolo and Suzana Iveljic for their efforts establishing the APT Center and providing the initial seed funding for our research. I also thank Brad Boggs for his training in industry best practices and for supporting this project. I thank Reza Moshtaghin, Nan Avishai, and Amir Avishai of the Swaglok Center for the Surface Analysis of Materials for training me to use the Scanning Electron Microscope and helping me obtain the results needed to complete my work. I have enjoyed working part-time on SBIR-funded projects with Dr. Walt Merrill for the past five years, and I learned enormously from this experience. I am also grateful for the time I had working with the late David Hiscock during his time with Haric and Scientific Monitoring. I would like to thank my colleagues and friends in both the MSIC group and the EECS and BME departments. I thank my labmate Steve Majerus for the all the technical, xvii professional, and personal discussions we have shared throughout our time together in the APT Center and Scientific Monitoring, Inc. Thank you to Bobby Lu, Dan Goff, Chris Roberts, Chia-Wei Soong, Kanokwan Limnuson, Srihari Rajgopal, Xinyu Yu, Paras Samsukha, Amita Patel, Masoud Roham, David Tian, Erik Peterson, Brian Murphy, and Christa Moss for your friendship and personal, physical, and professional encouragement during my time in graduate school. Mom and Dad, thank you for all the sacrifices you made throughout my childhood to give me a strong education and for teaching me to be a honest and caring person. Betsy, thank you for being a supportive sister and for always setting a high standard for me to follow. Dad Walter, thank you for raising a wonderful daughter, accepting me wholeheartedly as your son-in-law, and supporting us through our years as “professional students.” No words can express my gratitude to my wife Kristen Walter for her love, patience, support, encouragement, friendship, and guidance along the rocky road of graduate school. I am so fortunate to have met my perfect complement in a partner early in life, and you have helped me grow as a person in so many ways. We have made it through an extended long-distance relationship, and I am excited to begin the next chapter of our lives in a place we can pursue our dreams together. This work has received funding and support from the Cleveland VA APT Center and research grants from the Department of Veterans Affairs RR&D (# F7129R) and STERIS Corporation. xviii List of Abbreviations and Common Signal Names ADC: Analog to Digital Converter BLE: Bluetooth Low-Energy wireless protocol CC2540: “System on a Chip” microcontroller with built-in Bluetooth radio CHSEL: Channel Select signal (MSS ASIC). CMOS: Complementary Metal Oxide Semiconductor DAC: Digital to Analog Converter DC: Direct Current DUT: Device Under Test DVDD: Digital Voltage supply (MSS ASIC) ES: Electrical Stimulation FIB: Focused Ion Beam GND: Ground level for power supplies. HV: High voltage HVDD: High voltage supply for stimulator circuits HVL: Logic low level for circuits operating from HVDD supply (MSS ASIC). IDAC_SENSE: stimulator compliance voltage feedback signal (MSS ASIC) ISSD: Surface Stimulator Device (first-generation stimulator) LCP: Liquid Crystal Polymer LSB: Least Significant Bit MCU: Micro-Controller Unit MOSFET: Metal Oxide Semiconductor Field Effect Transistor xix MSS1: Modular Surface Stimulator (second-generation stimulator) MSS3: Modular Surface Stimulator (third-generation stimulator) NBIAS: NMOS current mirror bias signal (MSS ASIC) NCAS: NMOS cascode transistor bias signal (MSS ASIC) NMOS: N-type Metal Oxide Semiconductor transistor PFM: Pulse Frequency Mode PIC: Peripheral Interface Controller (Microcontroller) PMOS: P-type Metal Oxide Semiconductor transistor PTAT: Proportional To Absolute Temperature PWM: Pulse Width Modulation RTOS: Real Time Operating System SAR ADC: Successive Approximation Register analog to digital converter SPI: Serial Peripheral Interface TENS: Transcutaneous Electrical Nerve Stimulation VDD: Power supply voltage for digital circuits xx Abstract A Wireless Electrical Stimulation System For Wound Healing Therapy With Biphasic High-Voltage Pulsed Current Output Abstract by DANIEL STEVEN HOWE In this research, two wearable surface stimulation systems have been developed for use in wound electrotherapy. These self-contained, battery-powered bandages have been demonstrated in-vivo using both a rat chronic wound model and a pig infected wound model and provide a new means to investigate the physiological mechanisms of wound healing. The first stimulation bandage was designed for use with a rat wound model and consists of a stimulator PCB module and plastic electrode bandage. The PCB is constructed from discrete COTS components and is powered by a small button cell battery providing at least seven days of continuous use. This voltage-mode device generates stimulation pulses that are 10 - 90 V in amplitude, 10 - 200 µs in width, and 12 – 25 Hz in frequency. Stimulation was typically applied to wounds for 10 minutes every hour for one week, and then the device and wound dressings were replaced. xxi An ASIC has been developed using the OnSemi 0.7-µm I2T100 process and is capable of operation up to 100 V. A high-gain, current-mode boost converter addresses challenges associated with efficient generation of the large amplitude compliance voltage (up to 90 V) from a small battery with limited output current capability. A biphasic current mode stimulator was demonstrated with ±21-mA output range, 0.33-mA resolution, and a voltage headroom requirement of 3.5 V at full scale output. The second stimulation bandage uses this ASIC and was designed for use on larger wounds, up to 6 cm in diameter. A rechargeable lithium polymer battery allows a single stimulator module to be used for an entire 28-day study. The disposable electrode bandage portion of the device is easily replaced in-situ by the clinician. Continuous monitoring of the delivered stimulation current while the device is in place on an animal is achieved using a microcontroller with a built-in Bluetooth Low Energy radio. Performance information from up to six devices is recorded to a wireless base station located outside the animal pen and may also be remotely accessed by research personnel. xxii 1 Introduction 1.1 Clinical Motivation Chronic wounds such as shown in Figure 1-1 are a major clinical challenge for a large clinical population (estimated to be 1 to 3 million people in the US alone [1]) with conditions such as diabetes, paraplegia, and other physical impairments. A wound is considered chronic if it does not heal in a reasonable period of time. A practical definition that is often cited is longer than 1-3 months [2, 3]. Such wounds are both difficult and costly to treat. The treatment costs average around $22,000 per patient [4] but varies widely with severity of the wound. The total cost of treatment within Medicare alone is over $1 billion per year [5]. In addition, these wounds cause significant physical and emotional stress for the patient [6]. Figure 1-1. Example of a chronic wound on a foot [7]. 23 In addition to the treatment of chronic wounds, there is an emerging need for new treatments for wounds infected with multidrug-resistant bacteria [8]. Electrotherapy may be beneficial in the treatment of infected wounds both to reduce the use of antibiotics (which leads to drug resistance) and to treat these drug-resistant infections. This concept will be tested in a study using a porcine infected wound model and a wearable stimulation device that is tailored for the animal. 1.2 Electrotherapy and Wound Healing Electrical stimulation has been proposed as an adjunctive therapy for chronic wounds, but little scientific evidence is available to support its practice. There are several theories supporting the use of electrical stimulation as a chronic wound treatment. The endogenous electrical currents found in normally-healing wounds are disrupted in chronic wounds [9], and it is believed that restoring these currents may encourage healing [10]. This “current of injury” is believed to facilitate the movement of various cells within wounds via galvanotaxis during each phase of wound healing. Another possibility is that this current upregulates fibroblasts, neutrophils, and other cells involved in the healing process. This injury current has been measured on the order of 35 µA/cm2 and flows out of the wound and within 2-3 mm of the wound margin. In chronic wounds, the injury current is disrupted possibly due to tissue dehydration and chronic inflammation. In addition to maintaining a moist wound environment, it is believed that restoring this injury current using low-level DC current from an externally applied source may promote wound healing. Another theory is that electrical stimulation has an antiseptic or antibacterial effect that kills micro-organisms infecting a wound [11]. It is suggested that this wound 24 disinfection reduces inflammation and subsequent healing phases may commence. In this approach, high-voltage pulsed current (HVPC) is applied to the tissue to induce changes in pH or temperature that kills micro-organisms. However, if the stimulation amplitude is too high, gas formation may result due to water electrolysis that damages the tissue. 1.3 Previous Wound Healing Electrotherapy Studies Kloth presents an extensive review of electrical stimulation wound healing studies [12] that cover in-vitro, animal, and human subjects. Two studies using HVPC stimulation are presented here to describe the range of waveform parameters. Jercinovic et al [13] performed a human study using two round electrodes either 50 or 75 mm in diameter positioned on healthy skin 3 cm from the wound. Biphasic chargebalanced pulses 250 µs in width were delivered at 40 Hz for 4 s followed by a 4 s pause for 2 hrs/day, 5 days/week, for 4 weeks. The current amplitude was increased to just below the muscle contraction threshold of 35 mA. In their study with spinal-cord injured (SCI) patients, they found healing improved with the application of ES. As a follow up to the study, the initial control group was offered the stimulation treatment and this new “crossover” group showed significant improvement as well. This is particularly compelling because the same wounds failed to heal using standard care but healed after ES treatment. Weber et al [14] used their ImpediStim device to deliver HVPC stimulation to healthy human subjects using a multi-electrode array. A twin-peaked waveform consisting of two 100-µs pulses at 100 Hz and up to 200 V in amplitude were used. The patients adjusted the amplitude to just below the threshold of pain. The acute experiment applied 25 stimulation for 30 minutes, then measured the skin temperature, TcPO2, and tissue impedance at 1-h and 6-h timepoints. The brief experiment did not show any change in the measured parameters that would indicate a physiological effect. However, erythema (redness of the skin) was observed under the electrodes. The authors suggest increasing the treatment duration or the stimulation amplitude may result in detectable differences between stimulated and non-stimulated skin. 1.3.1 Current research limitations In the reported research studies, ES therapy is limited to relatively brief sessions lasting no more than two hours per day and five days per week. This is generally due to researcher/caregiver time limitations and participant acceptance that allows treatment only during dressing changes as part of standard care. A wearable device will allow continuous ES therapy between dressing changes that may accelerate wound healing by maintaining a consistent dosage. Animal studies are particularly valuable for understanding the effects of electrical stimulation on the physiological mechanisms of wound healing. In particular, chronic wounds with consistent etiology, size, and shape can be studied in a study population with similar age, health, medication, and nutrition. Animal experiments also facilitate the study of control wounds in the same animal that rarely occur in a patient population. In a patient population, it is ethically required that ES be used only as an adjunctive therapy to the standard care although it may confound study results. For infection studies, it is preferable to inoculate wounds with a selected monoculture that is obviously restricted to animal studies. 26 1.4 Existing electrical stimulation therapies for wound care Currently there are two major types of electrical stimulation therapies in clinical use. Microcurrent devices attempt to restore the natural “current of injury” to the wound using low-voltage stimulation producing a DC current on the order of 10-100 uA. Other stimulators such as re-purposed TENS machines deliver stimulation current pulses up to 20 mA. 1.4.1 Microcurrent devices The Vomaris Procellera® (Figure 1-2) is a commercial wound dressing that generates micro-currents on the order of 10 µA and a voltage potential from 300 to 900 mV using printed “micro batteries” consisting of elemental silver and zinc. This product is indicated for use in the treatment of stage III and IV wounds but produces a much lower level of stimulation consistent with the galvanotaxic approach of wound healing. The silver ions may also contribute a bacteriostatic effect on the wound environment. Each dressing may remain in place for up to 7 days, and the product is indicated for no more than 28 days of continuous use. Figure 1-2: The Vomaris Procellera(R) microcurrent wound dressing. 27 1.4.2 TENS/ pain management devices The LGMedSupply TEC Elite portable TENS unit (Figure 1-3) is indicated for pain management and muscle therapy but the stimulator has similar output characteristics as the proposed device (Table 1-1). Many of the HVPC studies utilize bedside or portable TENS units such as this device. Figure 1-3. Example of a TENS portable stimulator (from [15]). 1.4.3 Research Device: DS4 The Digitimer DS4 shown in Figure 1-1 is a research tool used in short-term neurodynamics studies and has similar stimulation specifications as the proposed device, especially the compliance voltage. Note that it is not a stand-alone stimulator as it requires an external waveform generator to create stimulation pulses. 28 Figure 1-4: DigiTimer Biphasic stimulator for benchtop testing Table 1-1 shows a comparison between the specifications of the TENS and DigiTimer stimulators with the proposed wearable device. Table 1-1: Specifications of a portable electrical stimulation unit. Specification TEC-Elite Digitimer DS4 Proposed device Indication Approved for Human Use Non-human use only Animal use Power Supply 9V rechargeable battery (24 hr lifetime) AC Adapter Handheld Battery powered: 10 x GP123A (12V each) 40 mAh per cell 190 x 110 x 80 mm 500 g Battery powered 0-105 mA square biphasic pulse 4 TENS modes: Burst, constant, modulation, modulation 1A 3 Muscle Stimulation modes: Synchronous, Asynchronous, Delay 1-60 minutes ±10 µA to ±10 mA analog biphasic pulse Analog input (up to ±10V FS) ±20 mA N/A 1-60 minutes/hr for 1 week up to 5 kHz 12.5-25 Hz 5-40 µs rise time 10-200 µs ±48V output (battery-powered) ±15 V output (line-powered) 0-90 V Stimulator size Current Amplitude Treatment Period Pulse Frequency Pulse Width TENS mode: 0.5-150 Hz MS mode: 2-5 Hz, 90-130Hz 50-300 µs Compliance Voltage 0-50 V (1000 Ω load) Must fit on species used in research study Biphasic 29 1.5 Previous work 1.5.1 Surface Stimulation Device (SSD) The Surface Stimulation Device shown in Figure 1-5 was the first system developed in the current program to study wound healing. The device specifications, design, and performance are reported in [16]. The device consists of a PCB measuring 44.5 x 31.75 mm and a polyimide substrate with two platinum electrodes. The PCB includes an 8-bit microcontroller, an IrDA serial port, and a discrete boost converter stimulator powered by a 3-V Lithium coin battery. The device was hand-soldered using lead-free solder, then encapsulated in heat shrink and silicone. Figure 1-5: Surface Stimulation Device (SSD), from [16]. The device was tested on a rabbit ear wound model (Figure 1-6), and several key design issues were revealed. The rabbit ear was too flexible to support the weight of the device, and it was not well-tolerated by the animal. The electrode gels adhered poorly to the skin, and it was necessary to use adhesive Tegaderm® film to wrap the device in place. The infrared IrDA® serial port was difficult to use in a clinical setting, and there was otherwise no visual indication of the device operation. The folded-tab contacts used on the single-layer substrate were also prone to cracking. 30 Figure 1-6: SSD attached to rabbit ear 1.6 Research Objective The objective of this work is to develop a wearable stimulation system suitable for research studies that may be readily adapted for use in a range of applications and species. This device will enable research studies not previously feasible due to the need for tethers or device physical size and weight limitations. The following intermediate goals will lead to the realization of the main research objective: 1) Design a discrete-component device suitable for use with a pre-clinical small animal research studies (Chapter 2). 2) Develop an ASIC boost converter controller and current-mode stimulator circuit for stimulation up to 90 V or 21 mA (Chapter 3). 3) Develop a wearable, wireless system with an integrated circuit stimulator and provisions for simultaneous use of multiple devices, then validate device operation in an ex-vivo animal study (Chapter 4). 31 This research has been conducted as part of a research project sponsored by the US Department of Veterans Affairs focusing on the Physiological Mechanisms of Electrical Stimulation in ischemic wounds. 32 2 Modular Surface Stimulator for Rodent Studies (MSS) A stimulation bandage design is needed for use on a rat wound model. As compared to the SSD [16, 17], the new device needs to be smaller and lighter weight and must also accommodate a tissue interface that has a smaller radius of curvature that is subject to twisting and bending. The entirely re-designed device, named the Modular Surface Stimulator (MSS), has a smaller stimulator PCB module, lower power consumption, and longer run time. A high-level block diagram of the proposed device is shown in Figure 2-1. A reusable, battery-powered stimulator mounts to a disposable, flexible plastic “bandage.” The bandage adheres to the skin surrounding the wound, and two hydrogel electrodes conduct the stimulation current into the tissue. Figure 2-1: Block diagram of the proposed MSS consisting of a flexible printed bandage and a reusable stimulator module. 2.1 Research Study Overview The research study which uses this device seeks to identify the physiological mechanisms of electrotherapy in wound healing and identify the most effective stimulation waveform parameters. A previously-validated ischemic wound model in rats 33 [18] was adapted for electrotherapy studies using the proposed device. As shown in Figure 2-2, an ischemic tissue flap is created along the back of the rat according to the wound model procedure. Two ischemic wounds are created using a biopsy punch in this slowly-healing region. Two more wounds are created in the healthy tissue adjacent to the ischemic wounds. These control wounds do not receive stimulation but are used to normalize the rate of healing between subjects. Figure 2-2: Rat wound model used with MSS The MSS device is placed over the wounds on the back of the rat, then covered with additional protective dressings and a fabric jacket. Electrical stimulation is delivered to the wounds for 7 days, and then the device is replaced during weekly dressing changes. The stimulator PCB is removed from the disposable substrate, cleaned, and then refurbished for later use. The stimulator performance requirements listed in Table 2-1 include the expected range of effective stimulation waveform parameters and the required battery lifetime for use in in-vivo studies. The exceptionally low duty factor of the shortest pulse has a large 34 impact on the design of the high-voltage circuits, and severely limits their power efficiency. Table 2-1: Design requirements for small-animal device Design Requirement Specification Battery lifetime Stimulator weight Stimulator size Overall cost Compliance voltage Current Amplitude Pulse Width, Frequency Stimulation Period Up to 168 hrs Less than 50 g No more than 3 x 5 x 0.7 cm Less than $100 5-90 V 0-20 mA, biphasic 10-200 µs, 12-25 Hz 1-60 minutes/ hr The programmable timing parameters shown in Table 2-1 are represented graphically in Figure 2-3. The size and shape of ischemic wounds varies widely, so the device must provide stimulation with a wide range of current pulse amplitude and width. Tissue impedance also varies greatly, and typical skin preparation for electrode application cannot be performed on the compromised tissues surrounding a wound, so a compliance voltage of 90 V is required. Figure 2-3. Graphical representation of the programmable stimulation waveform parameters that cover a range of potentially therapeutic levels. 35 A key technical challenge of such a device is the generation of stimulation pulses with the required high voltage compliance in a controlled and power-efficient manner, and with a compact form factor. It is proposed that stimulation pulse shape is less critical to wound healing than the listed parameters, so this stimulator design maximizes output range and power efficiency at the expense of control of the pulse shape. 2.1.1 Battery review The Lithium battery chemistry used in the ISSD does not provide the required energy capacity within the size constraints of the MSS device, so other primary (nonrechargeable) battery chemistries were surveyed and the available types are presented in Table 2-2. Rechargeable cells were not included in the survey because their energy density is generally lower than non-rechargeable cells, as shown in Figure 2-4. Zinc AirLithium batteries use the only chemistry that produces an output voltage sufficient to directly power the microcontroller (>2.5V); other chemistries would require an intermediate voltage step-up converter. Among the button-cell chemistries, the Zinc-Air type provides the most appreciable increase in energy density compared to Lithium. Thus, Zinc-Air batteries were chosen for this device. Table 2-2: Review of battery chemistries in coin/button form factor Chemistry Type Nominal Voltage Maximum Cell Capacity Available Lithium Alkaline Silver Zinc-Air Coin Button Button Button 3.0V 1.50V 1.55V 1.40V 130 mAh 150 mAh 195 mAh 635 mAh Size of Max. Capacity Cell (Diameter x thickness) 16 x 32 mm 11.6 x 5.4 mm 11.6 x 5.4 mm 11.6 x 5.4 mm Approx. Energy Density mWh/cm3 61 394 530 1558 Discharge Curve Linear Linear Flat Flat 36 Figure 2-4: Comparison of battery energy density for rechargeable and nonrechargeable chemistries (from [19]). Zinc-Air batteries are safe for use with humans and are commonly used in hearing aid devices. The battery has perforations on the top of the case to allow air to enter the cell, and this poses two potential drawbacks. First, the open cell may be exposed to wound exudate or moisture. However, the battery is located on the top of the bandage away from the wounds, so the likelihood of this event is low. Second, Zinc-Air cells will self-discharge within one month [20] so they must be replaced frequently. This is also not an issue because the MSS devices will be replaced weekly during dressing changes. 2.1.2 VDD Step-up converter Though this appears inconvenient, a step-up converter would help avoid the battery voltage brownout condition experienced with the ISSD. In addition, a step-up converter allows any battery type to be discharged more deeply than would otherwise be limited by the microcontroller operating voltage. The discharge depth is limited by the the minimum input voltage of the low-voltage boost converter. An integrated boost 37 converter designed for battery operation as low as 0.8V [21] was selected for the MSS. The system voltage was selected to be 2.0 V, the minimum allowed for the microcontroller, to minimize energy consumption. 2.1.3 Microcontroller New microcontroller product lines released since the ISSD have moved to foundry processes with a smaller feature size, and both memory capacity and device functionality have increased. These processors are designed to run programs compiled a high-level language such as C, and the code development time is reduced. The 16-bit CPU and memory bus decreases the number of instructions required per computation, and the program execution time per pulse is reduced. Advancements in microcontroller power management now allow the clock frequency for the CPU and the peripherals to be individually controlled. The CPU is often the largest consumer of energy on the microcontroller, so throttling its clock rate while the communication and timer peripherals run at the oscillator frequency significantly reduces power consumption. For example, the IDD current drops more than two orders of magnitude (from 3 mA to 15 µA) when the clock is throttled from 16 MHz to 32 kHz [22]. The CPU may be switched back to normal operation when faster control is required. 2.2 Boost Converter Circuit A discrete boost converter with a novel controller is presented in this section. The following discussion assumes an understanding of the principles of boost converter operation [21] and will focus on the key differences with the presented circuit. 38 2.2.1 Battery Limitations The MSS uses a replaceable Zinc-Air button cell battery because it satisfies the device size and weight requirements and has higher energy density than rechargeable battery chemistries. A microcontroller and other logic circuits are directly powered from this battery. Lithium coin cell batteries have relatively limited output current capability, and this is a key restriction in the design and operation of the boost converter circuit. 2.2.2 Operational Considerations A voltage step-up circuit is required to generate the stimulator compliance voltage from the 1.4-V battery. The boost converter topology was selected over flyback and switched capacitor technologies because it doesn’t require a physically large transformer, it achieves a high step-up ratio in a single stage, and the controller is simple and robust. However, there is no monolithic boost converter IC presently available with programmable high-voltage capability and the required power efficiency for this pulsed stimulator. In a conventional voltage-mode boost converter circuit (Figure 2-5), the transistor switch is controlled by a pulse-width modulated (PWM) signal. The duty factor is modulated to maintain a specified output voltage in response to changes in the load current. When the load current is low, the controller goes into a pulse-skipping or PFM mode of operation. When the load current is too high, the duty factor is internally limited to protect the switch and source power supply [2]. 39 Figure 2-5: Schematic diagram of the boost converter circuit using discrete components for HV circuitry and microcontroller for the control loop. In this pulsed-output application, however, the boost converter only operates at the two load extremes. When the compliance voltage capacitor is not fully charged, the boost converter operates in a PWM mode at a maximum output current limited by the battery. The boost converter is constrained to operate in discontinuous mode to prevent the inductor current from increasing each cycle. After the capacitor is charged, the boost converter operates in pulse-skipping mode to maintain the desired level. 2.2.3 Critical Component Selection The inductor selection is the most critical to the boost converter operation, and it is constrained by the limited output current of the battery and the microcontroller clock frequency. The frequency determines the minimum possible PWM pulse width ∙ , and the average inductor current when the MOSFET switch is on is given by 40 ∙ ∙ , 2 ∙ (1) 2 where T is the PWM period, D is the duty factor, and ∙ is the on-time. Since the boost circuit operates in discontinuous mode, i.e. the inductor current ramps from zero to a maximum during each cycle. The time-average current drawn from the battery is , ∙ ∙ ∙ , (2) The inductor value must be chosen to limit the average inductor current to within the capability of the battery. Within these limitations, the smallest value inductor was selected to reduce package size and to minimize series resistance. The remaining boost converter components must also be selected with consideration for high-voltage yet efficient operation. The n-channel MOSFET switch and Schottky diode rectifier are rated for low reverse-bias leakage. The output storage capacitor is a ceramic X7R dielectric with a 100-V rating that achieves high capacitance in a small package. 2.2.4 Output capability Based on energy conservation, the steady-state output voltage of the converter, prior to activation of the stimulator was calculated as given in equation (2), where RFB is the resistance of the voltage divider used in the feedback loop, and is added to reflect the voltage step-up efficiency. 41 ∙ ∙ ∙ ∙ 2 (3) The formula has been found to be accurate in predicting the output voltage. For example, to achieve an output of 80 Volts from a 3-V battery, with L = 330 µH, RFB = 4 M, and measured = 20%, this boost converter uses T = 6 µsec and D = 0.71. This represents the maximum output voltage that can be achieved with the indicated parameters. The low efficiency value is expected since the step-up ratio is large and average output current is small [21, 23]. 2.2.5 Controller firmware The control loop for the compliance voltage is implemented in firmware. The output voltage is sensed with a resistor divider, and is then digitized using the microcontroller ADC. The sensed output voltage is compared to a threshold, and the controller operates at a fixed duty cycle (determined by the battery voltage to limit peak inductor current) when below the threshold and is completely turned off when above threshold. The feedback loop does not require compensation for stability. The high voltage step-up ratio limits the feed-forward gain, and the ADC conversion cycle intrinsically limits the control loop bandwidth. The boost converter generally operates in discontinuous conduction mode since the step-up ratio from the battery voltage to the compliance voltage is large. However, when first powered, the converter will operate in continuous conduction mode unless the peak inductor current is temporarily limited. If the average inductor current rises due to 42 continuous conduction, the high internal resistance of the battery will result in a significant voltage drop and potentially cause a system malfunction. Some monolithic converters can sense the inductor current directly, but the high voltage operation of this circuit makes such an approach impractical. Instead, a “soft start” feature is implemented in the controller firmware by reducing the PWM duty factor until the output voltage reaches a level where discontinuous operation can be assured. The voltage feedback divider continuously loads the converter output, so it is critical that the output be charged just prior to each stimulation pulse to minimize losses. For a given PWM duty factor, the charging time would tend to increase as the battery voltage drops. Some monolithic converters compensate for the drop in battery voltage using voltage feed-forward, by adjusting the slope of an analog sawtooth waveform used to set the PWM duty factor. However, this approach requires additional components and the circuit may be difficult to stabilize when the battery voltage fluctuates due to its high internal resistance. In this converter, the microcontroller periodically senses the battery voltage using the internal ADC and adjusts the PWM duty factor in firmware to achieve a consistent maximum inductor current. Since time to reach maximum current is inversely proportional to battery voltage, the energy drawn from the battery during one PWM pulse is independent of battery voltage. Thus the average power drawn from the battery is independent of battery voltage. In other words, 2 ∙ ∙ ∙ and ∙ 2 (4) 43 As shown in Figure 2-6, the measured charging time remains relatively constant for a given output voltage regardless of the battery voltage, although non-idealities have increasingly large effect at the higher set points. The consistent charging time saves energy since the boost converter is not activated until the last possible instant before each pulse. Figure 2-6: Plot of the charging time vs. output voltage setpoints for several battery voltages. The fitted curves in Figure 2-6 use a power-law relationship with an exponent of approximately 3/2. If components were ideal, the exponent would be 2 since energy stored on the output capacitor is proportional to voltage squared, and the time-average power supplied by the battery is approximately constant for all battery voltages and set points, as described above. The faster than expected charging can be explained, in part, by the behavior of the high-voltage capacitor, namely that incremental capacitance decreases as voltage increases [24]. 44 2.3 Stimulator Circuits A high-voltage stimulation circuit is needed to deliver pulses of current from the boost converter circuit to the tissue. The pulse timing is generated using hardware timers inside the microcontroller, so the stimulation circuit must accept an input at a logic level of 3 V or less. The stimulator circuit is shown in Figure 2-7. High-voltage MOSFETs M1 and M2 form a push-pull driver that applies the stimulation voltage stored on CBOOST to the electrodes through DC-blocking capacitor CBLOCK, which insures a biphasic output. M3 is used to DC-couple the stimulator to the electrodes during impedance measurements. Figure 2-7. Schematic diagram showing the high-voltage stimulator circuit. The high-side driver for PMOS switches is shown in Fig. 2-8. When the low-side NMOS is switched on, the Zener diode operates in reverse breakdown and the Zener voltage is applied to the VSG of the high-side PMOS pass switch. Since, the PMOS VSG is independent of VBOOST, the on-resistance will remain relatively constant as VBOOST decays during a stimulation pulse. When the LV control is turned off, the voltage across 45 the Zener diode collapses and the PMOS transistor is turned off. The turn-off time will be relatively slow, but this is not important since precise pulse shapes are not required, and because the duty factor of stimulation is very low. Fig. 2-8. Schematic diagram of the high-side driver that controls PMOS switches with a constant VSG across a 90 V supply range. The output current can be controlled indirectly by estimating the tissue resistance and charging the capacitor to the required voltage. Resistance is measured by charging the compliance voltage capacitor to a set voltage, then examining the discharge waveform for a fixed period of time. Resistance is then calculated in the microcontroller based on the slope of the discharge waveform, under the assumption of 1st-order RC discharge. Based on the voltage resolution (10 mV) and sample rate of the ADC (100 ksps), the computation has a resolution of 500 Ω. 2.4 Experimental Results The boost converter and stimulator circuits were fabricated as part of the MSS1 surface stimulation device prototype shown in Figure 2-9. The stimulator module measures 3.0 x 5.0 x 0.7 cm and weighs 7.2 grams. 46 2.4.1 Modular Surface Stimulator Version 1(MSS) As compared to the SSD stimulator reported in [16], the MSS1 was designed to fit on the back of a rat and for improved battery lifetime. The infrared serial port was replaced with a pushbutton and LED to simplify use during surgery. An EEPROM memory was added to increase the amount of data logger storage capacity. The device is programmed through a basic UART and pin header connection. Figure 2-9: Modular Surface Stimulator (MSS1) Version 1 Device The MSS1 PCB was soldered using solder paste stencils and a reflow oven, then cleaned using an ultrasonic bath. The encapsulation was changed from heat shrink and silicone to a deposited parylene coating to improve function and reliability. A new substrate was designed by another member of the project team. The material was changed from polyimide to liquid-crystal polymer (LCP) and the folded-tab PCB contacts were replaced with conductive adhesive tape. The platinum electrodes were replaced with hydrogel material designed for electrical stimulation applications. 47 The initial animal experiments using the MSS1 yielded several valuable design insights. The mechanical switch used to activate the stimulator would stop functioning after repeated use, possibly due to corrosion or other fouling. While useful during benchtop testing, the switch did not provide benefit to the clinicians since the stimulator is activated immediately after inserting the battery into the device. In later builds, the device begins stimulating automatically after powering up. The lithium battery used did have sufficient capacity for week-long usage, and a larger size cell could not be accommodated within the size constraints of the device. In addition, it is believed that the microcontroller would reset itself when the partiallydrained battery browned out during boost converter operation. The lithium cell was replaced with a Zinc Air battery and low-voltage boost converter circuit occupying the same PCB area. The microcontroller firmware for basic operation of the stimulation module consumed all available program and data memories, so it was not possible to expand functionality or otherwise improve performance of the device through code. In addition, the CPU architecture was not designed for high-level programming in the C language, and modifications to the assembly code would often result in misdirected data pointers. 2.4.2 Stimulator Performance A series of representative stimulation pulse waveforms are shown in Fig. 2-10 with pulse width parameter set to 200 μs. The pulse current drops as the compliance voltage capacitor decays, as expected for a relatively low value of resistance. A larger output 48 capacitor would improve output voltage compliance at the cost of charging time and overall power efficiency. The stimulator pulse rise time is approximately 10 μs and the fall time is negligible. The rise time is limited by the turn-on current within the high-side driver, which has been kept low to minimize current draw from the high-voltage boost capacitor. The turn-off transition is faster because the NMOS in the push-pull driver is driven directly by the low-impedance driver of the microcontroller. Fig. 2-10. Plot of output pulses for several load resistors. The maximum pulse rate for a specified voltage is shown in Figure 2-11 at several supply voltages and is limited by the boost capacitor charging time. The effect of the battery internal resistance is more apparent as the output voltage increases. The pulse 49 frequency could be increased by decreasing the fixed 32-ms interval devoted to system tasks during each period of stimulus. Maximum Pulse Frequency (Hz) 27 25 23 21 3V source 19 2.8V source 2.6V source 17 3V battery 15 0 20 40 60 Stimulation Voltage Setpoint (V) 80 Figure 2-11: Measured maximum pulse frequency vs. voltage setpoint with various supply voltages. 2.4.3 Boost converter efficiency Boost converter efficiency is the key factor in determining the battery lifetime, and the efficiency will vary with the stimulation settings. The overall efficiency is the product of two components: first, the boost converter efficiency in charging the output storage capacitor, and second, the efficiency in discharging the capacitor into the tissue. The boost converter draws high-frequency spikes of current that are challenging to accurately capture, so a low-pass filter circuit was used in the test circuit to average the 50 current drawn from the power source. Efficiency is defined as the energy stored on the capacitor after charging divided by the measured current integrated over the charging period. As shown in Figure 2-12, the efficiency is much lower than values typically found in datasheets for boost converters. This is not an unexpected result since the input current limit due to the battery and the high step-up ratio lead to large losses in the switching components. The efficiency is slightly lower when a battery is used instead of a supply because of the internal impedance of the cell. Figure 2-12. Measured MSS1 Boost Converter Charging Efficiency for Selected Supply Voltages. The second efficiency component is simply the ratio of energy discharged from the capacitor per stimulation pulse divided by the total energy stored on the capacitor. Any residual charge on the capacitor is lost between cycles through the voltage sense resistor 51 network. As shown in Figure 2-16, the efficiency decreases with increasing load resistance because the capacitor is discharged less per cycle. The theoretical calculation was made using the first-order equation for a capacitor discharging through a resistor: ∙ (5) Figure 2-13: Measured MSS stimulator efficiency for selected boost capacitor voltages. The energy capacity of coin cell batteries is heavily influenced by the discharge rate and other operating characteristics. The labeled capacity of Lithium batteries is typically measured under low, continuous current until the battery voltage drops to 2 V, but the actual capacity is heavily influenced by the detailed output current characteristics. The capacity of battery chemistries such as Zinc-air is generally reported for a set of conditions using a pulsed output current [4]. 52 Device run time was measured for a selection of stimulation parameters and load resistances. As shown in Fig. 2-14, the battery lifetime decreases with stimulation voltage but exceeds the 168-hour requirement for all tissue resistance values. The lifetime remains constant between 50 V and 80 V because the stimulator pulse rate is reduced for the highest output voltages. Device Lifetime (Hr) 300 250 200 150 5k 100 10k 20k 50 50k 0 0 10 20 30 40 50 60 70 80 Stimulation Voltage (V) Fig. 2-14. Measured battery lifetime versus stimulation voltage for selected output resistances. Pulse rate is 25 Hz, pulse width is 200 µs, and stimulation interval is 10 min/hr. The device power efficiency is defined as the ratio of the energy delivered to the stimulator output to that drawn from the battery, per cycle, including the energy consumed by the microcontroller, boost converter, and stimulator circuits. The measured device power efficiency ranges from 5% for the lowest output voltage to 20% for the highest. This relatively low efficiency is not surprising given the large voltage step-up ratio and intermittent operation. The main sources of energy loss in the boost converter are capacitive switching losses in the NMOS switch and diode and DC losses in the inductor as well as the feedback resistor divider. 53 2.5 Summary A 90-V boost converter and stimulator circuits have been developed for a wearable electrical stimulation system. The boost converter control algorithm has been designed to support the generation of high-voltage, low duty-cycle current pulses in a power-efficient manner. Voltage feed-forward and soft-start techniques are implemented in software to support operation from coin cell batteries with high output impedance. The battery lifetime of the fabricated device meets the 168-hour requirement for the experimental application. Future designs will incorporate the boost converter and stimulation circuitry in an integrated circuit to improve power efficiency and reduce overall device size. A currentmode stimulator will be implemented to achieve rectangular pulses. 54 3 Application-Specific Integrated Circuit An application-specific integrated circuit (ASIC) is necessary to achieve high system performance and move towards a single-chip electrical stimulation device. The goal of this first version of the stimulation ASIC is to improve the stimulator performance by implementing a new high-voltage boost converter controller and a biphasic current DAC. The logic for the stimulation waveform timing, memory, and the wireless communication circuits will remain on a microcontroller to maintain design flexibility to meet the needs of wound healing research studies. 3.1 Functional Requirements The functional requirements are presented in Table 3-1 and are derived from the clinical application specification. Table 3-1. Functional requirements for ASIC. Requirement Compliance voltage range Specification 10-90 V Stimulator Pulse current Current resolution Pulse width Pulse frequency (period) ±20 mA 0.5 mA 10-200 µs 12.5-25 Hz (40-80 ms) Load Tissue resistance 5- 50 kΩ [25] Boost converter These requirements present design challenges for the high-voltage stimulator that include efficiently generating a large compliance voltage from a high-impedance source, interfacing control signals between low-voltage analog/logic circuits and the high-voltage stimulator, and modulating the boost converter output based on the load impedance and stimulation variables. 55 3.2 Foundry Process Features This ASIC was fabricated using the ON Semi I2T100 mixed signal process. The process was specifically designed for 100-V operation and includes a “floating pocket” feature that electrically isolates a region of the substrate using a buried oxide layer between the epi silicon layer and bulk. As shown in Figure 3-1, a lightly-doped NTUB is also used to reduce the electric field between high and low voltage sections of the chip. A library of high-voltage diffusion MOSFETs having both analog and digital characteristics includes models that enable accurate design and simulation. Figure 3-1: NDMOS floating transistor in I2T100 process (modified from [26]). In the low-voltage sections of the chip, the transistors are designed for CMOS logic operating at 5 V. However, a low-threshold PMOS device is available for use in analog circuits. The circuits in this ASIC are designed to operate at 3.3 V using an off-chip regulator. 3.3 High-Voltage Boost Converter The variable compliance voltage (10-90 V) is generated from a small battery and requires a large step-up ratio capability and efficient operation at low output current. 56 3.3.1 Step-up Topology Several step-up circuit topologies were considered for this application as discussed here. The charge pump technique [27] can be generally described as storing, then transferring, charge between a series of capacitors in such a manner that the output voltage is increased at each stage. The Dickson topology [27] shown in Figure 3-2 requires a cascade of high-voltage capacitors and diode-connected MOSFETs that occupy a large silicon area on an integrated circuit. The theoretical voltage of the Dickson converter is (3-1) Typically the voltage is only increased 2 to 3 times per diode-capacitor stage, so numerous stages would be required to achieve a total voltage gain of 60. The step-up ratio is not easily modified. Depending on the specific implementation, high-voltage interface circuits may be required at each stage. Figure 3-2: Dickson Charge Pump Voltage Conversion Technique (from [27]) 57 The flyback boost converter shown in Figure 3-3 has been used to achieve large stepup ratios but requires a transformer. The turns ratio of the transformer is fixed, but the output voltage can be adjusted by the duty factor of switching. When the desired voltage gain is lower than the turns ratio of the transformer, large voltage spikes may appear at the output. Figure 3-3. Representative schematic of the flyback converter architecture. Ultimately, the boost converter topology shown in Figure 3-4 was selected since it provides the desired variable step-up ratio, has a low component count, and a wide range of control algorithms may be used to tailor the performance for the specific application requirements. This is the same boost converter topology used in the previous discrete versions of the device, but traditional PWM control is not used. Instead, new control methods are used to directly control the peak inductor current each cycle. 58 Figure 3-4. Boost converter topology. 3.3.2 Control loop operation Current mode control in boost converters refers to an inner control loop that sets the peak inductor current on a cycle-by-cycle basis [28]. This method accounts for large changes in the input voltage (such as in battery-operated systems) and achieves a similar benefit as input voltage feed-forward. Directly sensing the inductor current is preferred when the input voltage source has large internal resistance and the voltage may droop as the output current increases. Limiting the inductor current reduces the possibility of battery brown-out and increases the control system robustness. Two control loops govern the operation of the boost converter. The first (inner) current control loop limits the cycle-by-cycle peak inductor current. When the output is near the desired high-voltage level, the forward (open-loop) voltage gain is small so modulating the peak inductor current has little use. Therefore, the peak inductor current is a constant, independent of the output level. In this application, the boost converter switch should turn off when the inductor current reaches a maximum level and then 59 switch on again when the current approaches zero. The desired behavior is diagrammed in Figure 3-5. Figure 3-5. Idealized inner-loop current control for the boost converter. The second (outer) control loop regulates the output voltage by enabling and disabling the inner loop current-mode control. This loop operates at a rate determined by the specific stimulator implementation but will always be slower than the inner inductor current control loop. As shown in Figure 3-6, an attenuated and buffered version of the output voltage is driven off-chip to an ADC input on a microcontroller (MCU). The boost converter control loop state machine algorithm determines if the PWM operation should be enabled to increase the output voltage. Since the boost converter may require several milliseconds to initially charge the output capacitor, the microcontroller may sleep in a low-power mode until the output approaches the desired level. 60 Figure 3-6. MSS ASIC-based boost converter circuit. 3.3.3 Inductor current sense Several methods for sensing the inductor current were considered for this application. A common method is to measure the current flowing out of the inductor by sensing the drain-source voltage across the NMOS switch, as shown in Figure 3-7. The NMOS switch operates in the triode region during the inductor charging phase, so VDS increases approximately linearly with the ramping IDS (since (VGS -VT) >> VDS) and may be used as the feedback signal. However, the voltage at the drain node rises to a high voltage level during the second phase of the boost converter operation and would require additional circuitry to interface with logic-level feedback circuits. Additionally, these interface circuits would likely require a high-voltage supply and would consume energy from the supply output. 61 Figure 3-7. Inductor current sense method across shorting NMOS. A second method, shown in Figure 3-8, is to sense the current flowing into the inductor from the battery side of the inductor. In this case, a second MOSFET switch is inserted in series with the inductor and operated in the triode region. The VDS of the 2nd MOSFET may be sensed to approximate the inductor current, and the MOSFET provides the ability to disconnect the circuit from the power supply to further reduce energy consumption when the stimulator is not operating. High voltages do not appear on the battery-side of the inductor, so the feedback circuitry may be operated at logic levels. However, the VGS of the MOSFET must be carefully controlled to maintain a consistent IDS-VDS transfer characteristic. As previously described, the battery voltage is subject to large fluctuations when the internal resistance is high and the gate voltage would need to accurately track this level. 62 Figure 3-8. Inductor current sense method using battery-side PMOS. The third method, shown in Figure 3-9, was selected for this circuit and is simply a low-value resistor inserted in series with the inductor. The voltage across the resistor is proportional to the current flowing through the inductor, and again the low voltage feedback signal may be easily processed with circuitry operated at logic voltage levels. This component is kept off-chip so the peak inductor current may be easily modified to match the characteristics of the power supply available for the device in a particular study. Figure 3-9. Inductor current sense using off-chip resistor. 63 3.3.4 Current sense amplifier The peak voltage across the current sense resistor is kept small to minimize energy loss. The differential voltage VR across the resistor is amplified using the single-stage differential amplifier shown in Figure 3-10. Figure 3-10. Differential Amplifier used as inductor current sense amplifier. Since M4,5 are diode connected, voltage gain is dominated by transconductance and is approximately independent of bias current, as given by , , ≅ ⁄ ⁄ , , ∙ (3-2) Bias current was set to 30 µA to obtain adequately low noise floor and output impedance. A simple hand calculation using square-law models suggests a voltage gain of about 10, but the PMOS input pair is operating in weak inversion. SPICE simulation predicts a voltage gain of ~5. 64 The common-mode input range is 0 , to (3-3) 2.45 , , (3-4) where This easily accomodates the expected 0.8 V to 1.5 V range of many single-cell batteries. The common-mode output range of a single-ended output is 0.73 , ∙ to , (3-5) 1.72 (3-6) The maximum differential output is the difference of the single-ended extremes approximately ±1 V. 3.3.5 Current sense comparator The differential output of the inductor current sense amplifier drives both inputs of a comparator used to control the inductor switch MOSFET. The comparator uses a regenerative feedback circuit [29] that allows the hysteresis thresholds to be determined using device size ratios. The schematic diagram for the current sense comparator circuit is shown in Figure 3-11. The transistor sizes are provided separately in Table 3-2. 65 Figure 3-11. Schematic diagram of current sense comparator circuit. Table 3-2. Transistor sizes for current sense comparator circuit. Component Type W/L (µm / µm) M1 PMOSL1 224 / 2.8 M2a, M2b PMOSL 56 / 5.6 M3a, M3b NMOS 5.6 / 1.4 M4a, M4b, NMOS 4.2 / 1.4 M5a, M5b, M6a, M6b M7a,M7b PMOSL 11.2 / 1.4 M8a, M8b M9a, M9b PMOSL 56 / 2.8 1. PMOSL is a low-VTP device Normalized Ratio 80 10 4 3 8 20 The maximum and minimum inductor current levels are determined by the upper and lower comparator thresholds. First, ignoring PMOS devices M9a and M9b, the circuit is symmetric and the switching thresholds are equal and opposite [29] according to the equations: , where | | 1 1 (3-7) , , 66 The component sizes and bias levels were selected such that | | 245mV, so 35mV. Since the thresholds are equal and opposite in a symmetric circuit, device M9a is added to inject an offset current IOFFSET into the left side of the circuit. M9b does not conduct current but is included to improve ac symmetry. To determine the shifted switching thresholds for an offset current IOFFSET supplied by M9a, the following equations are solved. At the positive switching threshold, the current sourced by is equal to that sunk by , and is beginning to conduct: , ∙ and where . 2 Thus, 1 1 | 1 | When | 1 1 2 (3-8) 2 1 2 | 1 1 1 4/3 and | given that | | | 2 1/4, | ≅ 105 mV, | has a nominal value of 245 mV. 67 For the negative switching threshold, the current sourced by that sunk by , and and is equal to is beginning to conduct: 2 2 2 2 2 1 1 | 1 | | When In other words, the | 1 1 | 1 4/3 and 1/4, 3 14 1 7 | | | | ≅ 17mV, | 245 mV . term accounts for a ± threshold of ± term accounts for a positive shift of and (3-9) 1 1 1 | 2 2 2 | when 2 | | . The in the two cases. The previous calculations do not account for channel length modulation in the saturated MOSFETs. VSD9 will be greater than VSD1 , so the offset current will be slightly higher than . This increases both and but both thresholds are positive and proper switching behavior will be maintained. 68 3.3.6 NMOS Switch The boost converter switch is implemented using an integrated DMOS transistor rated for operation with VDS <= 100 V. The transistor was implemented from a library P-Cell (parameterized cell) and sized as wide as possible, given the die area constraints, to minimize on-resistance. The simulated characteristics of the on-chip DMOS are similar to the off-chip COTS component it replaces on previous discrete versions of the device [30]. Table 3-3. Comparison of on-chip and off-chip boost converter switch components. Size RDS,ON Vt 3.3.7 Off-chip switch: BSS123W-F Unknown (SC-70 package) 10 Ω 1.4 V nom. On-chip DMOS W/L = 15000/6 µm 7.2 Ω (simulated) 1.02 V Test condition (datasheet) VGS = 4.5 V, 0.17 A VDS=VGS, ID=1 mA Output Rectifier A key component of the boost converter is the output rectifier that prevents current from flowing backwards to the battery after the inductor current drops to zero. Synchronous rectification using a PMOS switch is growing in popularity for boost converter circuits [31, 32] because the lower voltage drop across the device increases efficiency. The PMOS also provides a means of disconnecting the output from the power supply when the backgate diode is disconnected. However, the PMOS must be switched at the precise time to prevent current flowing backwards after the inductor discharges. The PMOS gate must be driven to the high-voltage output level to turn it off and a low level to turn it on. In this application, the output voltage often exceeds the VSG rating of the PMOS, so the gate cannot be simply pulled to ground level. Thus, additional circuits would be necessary to control the device. These additional components would consume additional energy that would likely erase any performance benefit for the power 69 levels in this application. Therefore, an off-chip diode is used to implement the output rectification function. Traditionally a diode with fast recovery, low reverse-bias leakage, and low forwardbias voltage such as a Schottky is used to achieve this function. However, the Schottky diodes that may be fabricated in a standard CMOS process often suffer from high substrate leakage and high series resistance [33]. Higher-quality Schottky diodes can be fabricated with CMOS processes using additional mask steps but are not widely available [33]. Further, the aluminum interconnect and p-doped silicon used to form CMOS Schottky diodes with low forward voltage have similar work functions that result in a very small depletion region [34]. This application requires a rectifier that can withstand 100 V in reverse bias, further reducing the suitability of integrating an on-chip Schottky diode. The BAT46W [35] Schottky diode was chosen. Table 3-4. Key characteristics of the BAT46W schottky diode [35]. Mode Forward bias Reverse bias Parameter Forward Voltage Repetitive Peak Current Blocking Voltage Reverse Current Reverse Capacitance 3.3.8 Value 0.25 V 0.45 V 350 mA 100V 0.02 – 0.2 µA 20 pF Conditions IF = 0.1 mA IF = 10 mA Duty Cycle < 50% 25˚C, VR=10-100 V VR = 0V f = 1 MHz Output capacitor The selection criteria for the boost converter output capacitor were carefully investigated in the previous work [16]. In this application, a larger value capacitor may be used since the loading from the resistive feedback network in previous designs has 70 been mitigated. Increasing the capacitance reduces the compliance voltage droop during pulses and allows more energy to be delivered per pulse. 3.3.9 Output Voltage Sense The output (boost) voltage is regulated using the ADC provided by the microcontroller. It is necessary to attenuate the high-voltage output to a level within the conversion range (0-2.048 V) of the microcontroller. A fixed resistor voltage divider would continuously draw current from the high-voltage output between pulses and when no measurement is taking place. The boost voltage sense circuit shown in Figure 3-12 employs a high-side PMOS switch to disconnect the 1:50 resistor divider when no measurements are being taken. A high-voltage level translator is required to control the gate voltage and is presented in section 3.4.2.3. The resistance values used in the divider network are kept large to minimize loading of the high-voltage output, but this large output resistance is not suitable for driving the capacitive load of an off-chip SAR ADC input. A differential amplifier configured in unity-gain configuration and powered from the logic-level supply provides the output buffering required to charge the off-chip ADC input within 10 µs. The linear output range of the op-amp is 0.25 V to 2.5 V, so the boost voltage is accurately sensed between 12.5 V and 100 V. In practical application, the boost converter output will generally be charged to at least 15 V so this will not limit system performance. The amplifier is disabled when the feedback network is disconnected from the output to reduce energy consumption. The amplifier is discussed in Section 3.4.3.2. 71 Figure 3-12. Schematic diagram of boost converter output voltage sense circuit. 3.4 Stimulator The clinical application requires the stimulator to provide biphasic pulses up to 20 mA in amplitude with a resolution of 1 mA. The stimulator is designed for a compliance voltage up to 90 V. 3.4.1 Architecture The biphasic stimulator is implemented in an H-bridge architecture with the high-side section containing HV PMOS switches and the low-side section containing a programmable IDAC and current-steering circuitry (Figure 3-13). To facilitate testing, the two sections are joined together off-chip and then connected to the stimulation electrodes. 72 Figure 3-13. High-level architecture of the stimulator circuit. While the specified 20-mA range and 1-mA resolution could be covered in 21 discrete steps (including 0 mA) using a 5-bit code (32 steps), a 6-bit code is used to improve the resolution and to allow for calibration if the output current is found to be inaccurate due to circuit non-linearity or offset. When a 6-bit code is used, 64 steps are available. Each step will be set nominally to 0.33 mA so integer values of current can be produced. 3.4.2 High-side switches The stimulator high-side circuits are shown in Figure 3-14. A pair of high-side PMOS switches are used in conjuction with the low-side DAC in an H-bridge configuration to steer the stimulation current in either direction through the two electrodes. These PMOS transistors are the 100-V DMOS type and require gate drive signals that range from HVDD (Vsg = 0) to HVL (Vsg = HVDD-HVL). The transistor gate oxide thickness limits VSG to 15 V, so the gate voltage cannot be simply pulled to 73 ground level to turn on the device. A negative voltage regulator produces the HVL level, and voltage level converters translate the logic-level control signals to the PMOS gate drive level. Figure 3-14. Stimulator High-side circuits. 3.4.2.1 Negative voltage regulator The negative voltage regulator operates in a shunt configuration and uses a CMOS regulator circuit to set the HVL level. The low-voltage circuits are constructed in an isolated P-well with a floating substrate that operates at the HVL level. The Zener-like regulator circuit was designed to generate approximately 3.2 V using minimal bias current (100 µA) that is sunk from the circuit using a 100-V n-channel DMOS cascode switch connected to the logic-level circuitry. When the stimulator is not active, the regulator is shut down and the PMOS devices will turn off when the HVL supply collapses. 74 3.4.2.2 Floating CMOS Reference The floating CMOS reference circuit [36] shown in Figure 3-15 mimics the I-V characteristics of a discrete Zener diode. The circuit is fabricated in a HV pocket so lowvoltage devices may be used. However, care must be taken to prevent HVL from being shorted to a low-voltage level during operation. In steady state, the current through M2 and M3 is made equal and ∆ , , across R3 determines the bias current . This value is calculated as follows: , , ′ ′ (3-10) ′ Since . Thus, and ∆ With R3 = 9.5 kΩ, ′ 15μ / is 18.5 µA and ∆ , (3-11) and =40, 0.175 . , is given by Equation (3-12): The reference voltage 2 , | | ∆ , 2 2 (3-12) With | |= 1.05 V, and R1 = 4R3, VZ ≈ 3.0 V. SPICE simulations predict VZ = 3.15 V. 75 Figure 3-15. Schematic diagram of floating CMOS reference circuit. 3.4.2.3 Digital voltage level shift Efficient level shifting is critical to the low-power operation of the high-voltage circuitry. The capacitive level shift circuit [37] shown in Figure 3-16 is used to interface the logic-level control signals from the microcontroller to the HV level for driving the gates of HV-PMOS switches. Inverters I1-I3 are standard CMOS inverters constructed in the substrate biased at the system ground level. The CMOS inverters I4-I6 operating at the HV level are constructed using low-voltage devices in a floating pocket biased at the HVL voltage level. Capacitors C1 and C2 are charged to HVDD-DVDD and HVL-GND, so they must be designed to tolerate this high voltage. The capacitors must be sized to couple adequate charge to I4-I5. When a logic transition occurs at the In input, the bottom plates of C1 76 and C2 connected to I2 and I3 will simultaneously shift up/down by DVDD. If the capacitors hold their charge, then the top plates will also shift by DVDD. The crosscoupled inverter pair (I4 and I5) will toggle and the logic state will be latched at the highvoltage level. Inverter I6 buffers the output of I4 to drive the large gate of the HV PMOS output switch. The initial state of the output is indeterminate but cycling the input during system initialization will reset the voltage across each capacitor and set the output to a known state. Figure 3-16. High-voltage level shift circuit. 3.4.3 Low-side DAC pair The low side of the stimulator shown in Figure 3-17 consists of a SPI-controlled current DAC, a pair of HV NMOS current steering switches, and support circuitry. The IDAC uses a LSB reference current of 13 µA and converts a 6-bit digital code received from the SPI port into a reference current between 0-875 µA. A current mirror using NMOS devices operating in the triode region [38] scales the output current to the full 21– 77 mA range. Based on the state of CHSEL, one of two HV NMOS cascode transistors is driven to steer the output current to either the A- or B- pins. By operating the mirroring transistors in the triode region, power dissipation is reduced in the output stage and the minimum accurate output voltage is reduced. Unlike [38], the VDS of the triode mirroring devices are kept the same using a second op-amp control loop. Figure 3-17. Stimulator low-side DAC. 3.4.3.1 Current DAC (IDAC) The low-side current DAC (IDAC) is shown in Figure 3-18. The NCAS and NBIAS levels are locally regenerated from a 4.1-µA reference current from the global bias circuit to minimize inaccuracy due to device threshold mismatch across the die. A series of binary-weighted cascade current mirrors produce a scaled version (up to 110 µA) of the 78 desired output current. A cascaded PMOS current mirror reverses the current direction and multiplies the value by 8. Figure 3-18. Schematic diagram of current DAC. 3.4.3.2 Stimulator MOSFET Control Amplifiers The folded-cascode operational amplifier shown in Figure 3-19 is used both in the stimulator to control the VDS of the triode MOSFETs, and in the boost converter to buffer the output feedback voltage. The key requirements for this amplifier are an output swing of below VTn and as close to the upper rail as possible. The inputs must operate at a common-mode input as low as 0.1V. Additionally, the output must be able to charge the gate capacitance of the stimulator output stage NMOS and DMOS devices. 79 Figure 3-19. Schematic diagram of folded-cascode amplifier used in stimulator circuits. 3.4.3.3 Stimulator Control The stimulator control block uses the CHSEL signal input to direct the gate control voltage to one of the two cascode DMOS and forces off the other device. The STIM_EN signal also controls the DMOS gate operation and is used to control the stimulation pulse timing. 3.5 Support circuits 3.5.1 Bias generation and distribution The bias voltages and currents are generated using a traditional bandgap voltage reference circuit [39] shown in Figure 3-20. The MOSFET sizes are presented separately in Table 3-5 for clarity. VLOW is the reference voltage used with the op-amp control loops to set the desired VDS level of triode MOSFETs in the low-side stimulator circuit. The voltage is proportional to absolute temperature (PTAT) but the stimulator circuit does not require a precise value. 80 Figure 3-20. Schematic diagram of bias voltage generation circuit. Table 3-5. Device sizes for bias circuit. Subcircuit Startup Component M1 M2, M3, M4 Type PMOSL1 NMOS W/L (µm / µm) 2.8 / 4.8 2.8 / 2.8 Enable M5 PMOSL 28 / 2.8 Bandgap Reference M6 a, b M7 a, b M10 a, b M8 M9 PMOSL 224/2.8 NMOS 56 / 1.4 Vlow level M11 a, b PMOSL 112/2.8 Nbias, Ncas levels M12 a, b M13 M14 M15 PMOSL NMOS NMOS NMOS 64 / 2.8 16 / 2.8 32 / 2.8 IREF_SRC IREF_SRC4 IREF_SINK 3.5.2 M16 a, b PMOSL M17 a, b PMOSL M18 a, b NMOS 1. PMOSL is a low-VTP device 112 / 2.8 448 / 2.8 32 / 2.8 SPI Serial Port The SPI serial port accepts a 12-bit command in this version of the ASIC to facilitate future control register expansion. Bit assignment is summarized in Figure 3-21. Bits 81 D[11:8] are used to address the register (location 0h). Bits [7:6] are general-purpose outputs that are presently connected to control inputs STIM_EN and BIAS_EN to reduce I/O connections to the microcontroller. Bits [5:0] set the IDAC output current (1 LSB = 0.33 mA). D[11:8] Register address = 0h D[7] GP2 D[6] GP1 D[5:0] IDAC Current Figure 3-21. MSS SPI Serial Port Configuration. 3.5.3 Pad Frame, ESD Protection, and Guard rings A foundry-designed pad cell library was not available from MOSIS, so a custom library was developed based on existing designs [40]. Analog and digital input pins use a 800-Ω resistor to limit input current, and diode-connected PMOS and NMOS devices [41] clamp ESD transients and signals outside the voltage supply range. The digital output pins are also protected by both the diode-connected MOS protection devices to limit conduction in the body diodes of the output stage MOSFETs. The low-voltage circuit supplies use separate pins and traces for the analog and digital supplies. Special care was taken when routing high-voltage signals because the field threshold voltage is only 13 V for poly traces [26]. Therefore, the high-voltage circuits are placed in a separate region and the HVDD and HVL power traces are routed in top-level metal (Metal 3). All high-voltage cells are surrounded by p-type guard rings connected to ground, and substrate contacts are placed at most every 50 µm to avoid field inversion. Double guard rings are placed around the boost converter switch and the low-side stimulator output MOSFETs to avoid unwanted substrate current injection. 82 3.6 ASIC Test Results 3.6.1 Introduction The MSS ASIC was fabricated using the OnSemi I2T100 0.7-µm BCD process using the MOSIS service. The IC includes the boost converter, bias generation, serial SPI port, stimulator low-side IDAC, stimulator high-side switches, and HVL generation circuits. A die microphotograph is shown in Figure 3-22. The fabricated die measures 2.96 mm by 2.96 mm and uses 40 I/O pads, including 11 for test purposes. Figure 3-22: Die microphotograph of MSS ASIC. A list of the ASIC bond pads with I/O type and function is provided in Table 3-6. The die was packaged in a 40-pin QFN Open Cavity Package by the MOSIS service. The package is shown in Figure 3-23 and measures 6 mm x 6 mm x 0.8 mm. A leadless, small form-factor package was selected to minimize the stimulator size and facilitate 83 packaging. The packaged parts have the same pad orientation and as the bare die, so the same signal list in Table 3-6 is used. Table 3-6. List of MSS ASIC bond pads and their function. # Name Type Function # Name Type Function 1 AVDD Power Analog supply 21 HVSS Power Stimulator ground 2 AVSS Power Analog ground 22 ElecB- Analog Out Low stim. output B 3 VLOW Analog Out VDS level 23 ElecA- Analog Out Low stim. output A 4 NBIAS Analog Out Amplifier bias 24 ElecB+ Analog Out High stim. output B 5 NCAS Analog Out Amplifier bias 25 ElecA+ Analog Out High stim. output A 6 VBG Analog Out Reference 26 HVL Power High stim. gate drive 7 PBIAS Analog Out Amplifier bias 27 HVDD Power Stimulator supply 8 BIAS_EN Digital In Bias enable 28 VOUT Analog In Feedback divider input 9 D0 Digital Out GP output 29 LX Power Boost converter switch 10 NC - Unused 30 BVSS Power Boost converter ground 11 DVDD Power Digital supply 31 SHUNT Analog In Shunt amplifier neg. input 12 GND Power Digital ground 32 BATT Analog In Shunt amplifier pos. input 13 SCLK Digital In Serial clock 33 SENSE Analog Out Feedback divider output 14 SDI Digital In Serial data in 34 SHUNT_AMP Analog Out Shunt amplifier output 15 SS Digital In Serial select 35 BOOST_MODE Digital In Boost mode select 16 SDO Digital Out Serial data out 36 IMODE_PWM Digital Out Comparator output 17 STIM_EN Digital In Stim. Enable 37 EXT_PWM Digital In 18 CHSEL Digital In Channel select 38 BOOST_EN Digital In Boost converter PWM input Boost converter enable 19 IDAC_SENSE Analog 39 SENSE_EN Digital In 20 IDAC_MODE Digital In Low-side stim gate drive Current/voltage mode select 40 D1 Digital Out Feedback divider enable GP output Figure 3-23. The MSS ASIC wirebonded in QFN40 surface mount package. 84 3.6.2 MSS ASIC Test Setup The ASIC was tested using the evaluation board shown in Figure 3-24. Devices under test are installed using an adapter board with either a QFN solder pad footprint or a QFN socket for production device qualification. Parts packaged in a DIP40 may also be tested using this PCB. Figure 3-24. ASIC test PCB with QFN-packaged DUT . 3.6.3 Integrated Circuit Edits Using Focused Ion Beam Three issues were found during initial testing that could be corrected using a Focused Ion Beam (FIB) instrument. Repairs were simple and an updated CAD layout has been created and verified. A SEM micrograph of such an edit is shown in Figure 3-25. In step 1, ion beam milling is used to cut through one wire and expose the side of a second wire. In step 2, platinum is sputtered over the exposed wires to form a new connection. The test results presented in this section were collected from a repaired die. 85 Step 1: Cut vertical trace using ion etch in a region that exposes desired signal connection below. Step 2: Deposit platinum to electrically connect top vertical trace to bottom horizontal trace. Figure 3-25. Example of circuit edit using focused ion beam (FIB). 3.6.4 Bias circuit The bias circuit on the fabricated IC does not operate as predicted by simulation. A design issue was found in the startup circuit, and the bandgap reference does not settle to the desired operating point. When the startup circuit is removed, partial function is restored. The bias circuit voltages shown in Table 3-7 were measured on a repaired die using a multimeter. The measured values for NCAS and VLOW are outside the range of operation for the ASIC circuitry, so the PBIAS signal must be driven externally. The stimulator output current is proportional to a reference current generated in the bias circuit, so the PBIAS voltage level must be adjusted for each chip. Table 3-7: Comparison of simulated and measured bias circuit voltages. Level AVDD PBIAS VBG NCAS NBIAS VLOW Simulation 2.36 V 1.12 V 1.05 V 0.95 V 0.18 V Measurement (repaired) Measurement (PBIAS driven to 2.45 V) 3.30 V 2.10 V (2.45 V) 0.13V 0.60 V 0.29 V 1.02 V N/A (no external pin) 0.01 V 0. 05V 86 3.6.5 Boost converter operation 3.6.5.1 Current sense amplifier and comparator The current sense differential amplifier (Figure 3-10) and comparator (Figure 3-11) circuits were tested simultaneously using the test setup shown in Figure 3-26. The BATT pin is normally connected to the positive terminal of a battery, and so it is connected to a fixed DC voltage level. The SHUNT pin voltage normally decreases in amplitude relative to the BATT pin, and is driven by a function generator in this test. Figure 3-26: Test setup for differential amplifier and comparator circuits. The selected VBATT amplitudes represent the voltage range of a Zinc-air battery. The PBIAS level is set externally to 2.56 V using a resistor voltage divider. As shown in Table 3-8, the maximum switching frequency of the circuit is limited to 400 Hz and is too low for use in the boost converter circuit. A layout error was found in the power control switch that limits the differential amplifier bias current. This error is in a location that cannot be reliably corrected using FIB circuit editing techniques. 87 Table 3-8: Measured performance of the boost converter current sense circuits. BATT Level: 0.70 V 0.80 V 0.90V 1.10 V 1.30 V SHUNT Offset 0.60 V 0.75 V 0.85 V 1.05V 1.25V SHUNT amplitude 200 mVpp 100mVpp 100 mVpp 100mVpp 100 mVpp Max Switching Frequency 50 Hz 50 Hz 50 Hz 50 Hz 400 Hz 3.6.5.2 DMOS switch The boost converter DMOS switching speed and current capability was tested using a 50 Ω resistor connected between a 3 V supply and the LX pin. As shown in Figure 3-27, fast rise and fall times are achieved at the maximum expected switching frequency of 500 kHz. The switch is shown sinking 50 mA in this test, and the peak current during operation will be limited to below this value by the inductor size and the boost converter PWM on-time. The observed ringing is a result of inductance in the power supply test leads. Figure 3-27: MSS ASIC Boost Converter DMOS switching behavior with 50-ohm load. 88 3.6.5.3 Open-loop operation The open-loop boost converter operation operating using an external 1 kHz, 50% duty cycle PWM control signal to drive the on-chip DMOS switch is demonstrated in Figure 3-28. A low PWM frequency is used in this open-loop test to prevent the output voltage from exceeding the maximum VDS rating of the DMOS switch. The boost converter charges to the maximum rated value of 90 V within 500 ms and meets the compliance Boost Converter Output Voltage (V) voltage requirements for the stimulator circuits. 90 80 70 60 50 40 30 20 10 0 0 100 200 300 400 500 Time (ms) Figure 3-28: Boost converter charging performance using on-chip DMOS switch. 3.6.6 Stimulator 3.6.6.1 Negative Voltage Regulator The negative voltage regulator was tested using the circuit shown in Figure 3-29. HVDD was swept from 5 – 25 V using a benchtop power supply, and VZ = HVDD – HVL was measured. 89 Figure 3-29: Test setup used to measure negative voltage regulator circuit. As shown in Figure 3-30, the measured value of 1.3 V is lower than predicted in simulation (3.15 V). This result is likely caused by internal bias circuit drawing less current (20 µA) than required for proper operation (~40 µA). After an external 200 kΩ resistor was connected from HVL to HVSS, VZ increases linearly to 4.3 V when HVDD is 18 V. At this operating point, the external resistor draws an additional ~70 µA through the circuit, about twice the 40 µA bias current needed for proper operation. For HVDD values greater than 18 V, VZ remains constant at 4.3 V. The additional bias current supplied by the external resistor increases with HVDD to a maximum of 430 µA when HVDD = 90 V. This external resistor was used for the remainder of the ASIC benchtop testing. In the MSS3, the external resistor is connected to HVSS through a HV transistor so the external bias current is turned off between pulses and power consumption from the HVDD supply is reduced. 90 Figure 3-30: Negative voltage regulator output for HVDD between 5 V and 25 V. 3.6.6.2 High-side switches The performance of the fabricated high-side PMOS was tested using the circuit shown in Figure 3-31. This test also verifies the operation of the voltage level shifter and the floating HV logic circuits that drive the PMOS gates to HVDD and HVL. The saturation current of the high-side extended-drain PMOS switches is expected to be 100 mA when VSG = HVDD – HVL = 3.3 V, based on SPICE simulations. This is roughly 5 times the maximum stimulation current drawn by the stimulator IDAC. Figure 3-31: Stimulator High-side PMOS test circuit 91 HVDD and RL were selected to limit the current through the PMOS switch to 40 mA, about twice the maximum stimulation current. As shown in Figure 3-32, the PMOS turnon time is short and VDS across the PMOS switch is about 1 V. The voltage level shifter, floating HV logic, and HV PMOS circuits operate as expected. Figure 3-32: Switching behavior of stimulator high-side PMOSFET with HVDD = 20 V, f = 1 kHz, and 500 Ω resistor to GND. 3.6.6.3 Low-side DAC An open-circuit layout error was found in the IDAC_MODE signal path that was corrected using the FIB prior to circuit characterization. The DC current transfer function of the low-side DAC was measured using the circuit shown in Figure 3-33. The voltage applied to PBIAS was adjusted so the full-scale IDAC code produces a 21 mA current. In the first test, HVDD is fixed at 5 V and the digital IDAC code is swept across the range. HVDD is kept low to reduce on-chip power dissipation in the cascode NMOS. As shown in Figure 3-34, the output current range from 0 to 21 mA is achieved but linearity is poor. 92 Figure 3-33: ASIC low-side stimulator test circuit. Figure 3-34: Stimulator low-side IDAC current transfer characteristic with HVDD= 5 V. This behavior is partly explained by the plot of the NMOS cascode transistor gate drive voltage, IDAC_SENSE, shown in Figure 3-35. IDAC_SENSE should increase 93 linearly with the IDAC code. At both extremes of the operating range, inadequate performance of the replica control loop results in IDAC nonlinearity. As suggested by [38], IDAC linearity is less critical in biomedical applications and may be compensated using a lookup table in software. Figure 3-35: Measured IDAC_SENSE voltage driving gate of NMOS cascode across IDAC code range with HVDD =5 V. In the second test, the compliance voltage sensitivity is observed by repeatedly measuring the current transfer function for several HVDD values. Figure 3-36 shows the IDAC current is relatively insensitive to HVDD within a range that provides adequate voltage headroom but does not result in excessive on-chip power dissipation. The undervoltage condition is demonstrated in the HVDD = 4 V trace when the output current is reduced for large IDAC codes. In this case, the IDAC_SENSE voltage is near DVDD and indicates to the microcontroller that HVDD should be increased. For HVDD values ≥ 15 V, the IDAC code was swept from zero to the value when the calculated on-chip 94 power dissipation reached 50 mW. A low IDAC_SENSE voltage indicates that HVDD is too high and should be reduced. Figure 3-36: Stimulator compliance voltage sensitivity for several HVDD values and RL = 100 Ω. 3.6.7 Summary The functionality of the current version of the MSS ASIC is summarized in Table 3-9. The chip requires two FIB circuit edits and three external resistors to obtain the desired operation for the MSS3 device. The boost converter circuit must be operated in the external mode using a PWM signal generated by the microcontroller. 95 Table 3-9: Summary of MSS ASIC circuit functionality Circuit Functional status Required modifications Bias circuit: Partially functional Remove startup circuit signal connection using FIB. Use external resistor divider to adjust PBIAS level. Boost Converter: Internal current mode External PWM mode Non-functional Functional Use external PWM mode None Stimulator High Side: PMOS switches Level translator Negative Voltage Regulator Functional Functional Partially functional None None Connect 200 kΩ external resistor between HVL and HVSS. Partially functional Repair open circuit in IDAC_MODE signal using FIB. Stimulator Low Side: 96 4 Wireless Stimulation Bandage and System for a Large Animal Model The ASIC and supporting electronics have been developed for use in a wide range of animal and clinical studies. In this chapter, a stimulator module and bandage for studies involving large wounds similar to those found clinically in humans is presented. 4.1 Large Infected Wound Study Infection is a common issue in both chronic wounds [42] and in acute wounds. In particular, wounds suffered on the battlefield are particularly prone to infection [8, 43]. In addition, the rise in drug-resistant infections [44] has increased the need for alternative interventions. Electrotherapy has been shown to impede the growth of biofilms in wounds [11] and may be an effective non-pharmaceutical therapy for treating infection. A pilot study using electrical stimulation to treat infected wounds has been proposed [45] using a previously-validated infected wound model [46]. As shown in Figure 4-1, 2 wounds, 6 cm in diameter, will be created bilaterally (4 total) along the paraspinous region of a female Yorkshire pig. All four of the wounds are inoculated with a fluorescent strain of bacterium and allowed to colonize. Two of the wounds will be randomly selected to receive electrical stimulation therapy while the remaining two will serve as controls and receive non-stimulating bandages as standard care. The wound dressings and stimulation bandages are periodically removed over a 28-day period as required by the study protocol, then tissue samples are harvested from the wounds. 97 Figure 4-1: Approximate wound locations on porcine model for infected wound study (2 wounds per side) [45]. From a medical device engineering standpoint, the research experiment is summarized in Table 4-1. The stimulation requirements (6-8) are similar to previous studies, but a physically larger device is necessary to accommodate the increased wound size. Table 4-1: Clinical Requirements for Infected Wound Study Device # Design Requirement 1 2 3 4 5 6 7 8 Experiment duration Dressing change number, interval Bandage weight Bandage size Unit cost (not including ASIC) Pulse Amplitude Pulse Width Pulse Frequency 10 Number of bandages in simultaneous use 11 Secondary dressing requirements 12 Bandage feedback Specification 28 days 7 changes: D0, D1, D3, D5, D7, D14, D21 < 500 g No more than 10 cm x 10 cm Less than $100 0-21 mA, biphasic 0-200 µs 12-25 Hz (Max duty factor 0.5%) 3 stimulating bandages 3 non-stimulating bandages Each bandage must be self-contained and have no external connections. Bandage must indicate operation in non-visual manner The requirements of this animal study have implications that require a new stimulation bandage design. In particular, requirements (1, 2, and 10) result in a large number of bandages needed (4 wound locations * 7 changes = 28 bandages) during the 98 experiment. The stimulator modules may be refurbished, re-sterilized, and reused, but there is insufficient turn-around time during the first 7 days of the experiment to do so. Therefore, it would be preferable to re-use the stimulator portion of the device in-situ between changes and replace only the substrate portion of the bandage that is in contact with the skin. The bandage substrate presented in section 4.6.2 has been re-designed to fit the wound model and to facilitate in-situ replacement on the stimulator module. In order for the stimulator module to last the entire 28-day experiment, a battery with greater capacity is needed. Fortunately, the size of both the substrate and the animal is larger in this experiment, so different battery chemistries and packaging were evaluated, as presented in section 4.2. A method of communicating the device status to the clinical users is very important so clinicians can confirm the stimulator is delivering current into the skin. The devices will be obscured by secondary dressings and a protective jacket while in use, so a nonvisual indicator is needed. Low-power Bluetooth radios targeted for battery-powered applications became commercially available in 2011 and provide the desired capability. Based on these experimental requirements, the device concept shown in Figure 4-2 was created. Similar to previous design, 1 cm x 8 cm hydrogel electrodes are attached to the bottom side of a flexible plastic substrate. If desired, an observation window may be cut out from the substrate over the wound area. On the top side of the bandage, flexible conductive traces are painted to connect the stimulation electrodes to a snap connector that mates with the stimulator PCB module. The painted traces are connected to the hydrogel electrodes through 3 paint-filled vias made through the substrate and spaced 99 ~2.5 cm. A stimulator PCB with a battery is attached to the substrate via the snap connector and double-sided adhesive film. Figure 4-2: Stimulation bandage concept for large-wound studies. 4.2 Bandage Substrate The flexible bandage substrate shown in Figure 4-3 provides a convenient and repeatable method of positioning the stimulation electrodes over wounds. The substrate also functions as an occlusive wound dressing to maintain tissue hydration. Springloaded snaps [47] are used to electrically connect the stimulator module to the electrodes via painted conductive traces. The substrate is fabricated from Rogers Liquid Crystal Polymer (1 mil thickness). This material exhibits a lower rate of moisture absorption than polyamide substrates typically used in flex circuit applications. The complete substrate fabrication process was co-developed with Jeremy Dunning. 100 Figure 4-3. Top (left) and bottom (right) views of bandage substrate 4.2.1 Bandage Electrode characterization The surface resistivity of the painted traces is 55 Ω/sq (for a thickness of 50 µm) [48] and the width is 0.8 cm, so the resistance is expected to be ~70 Ω/cm length. The impedance to points along the length of the electrodes will not be uniform due to the impedance of the painted trace, hydrogel electrode, and the limited number of vias between the two. Electrode impedance was tested using the test fixture shown and diagrammed in Figure 4-4. The electrode is contacted by eight equally-spaced copper pads measuring 0.8 cm in width. The painted trace is contacted via the snap connector permanently mounted on the bandage. Figure 4-4: Test fixture and setup for substrate verification. 101 The electrode was hydrated using approximately 5 mL sterile saline, and then adhered to the test structure. Using an LCR meter, the magnitude of the complex impedance was measured between the snap connector and each of copper contact pads at frequencies of 100 Hz, 1 kHz, 10 kHz, and 100 kHz. As shown in Figure 4-5, the magnitude of the electrode impedance is less than 4 kΩ for frequencies above 1 kHz and is relatively insensitive to position. Given this value, up to 4 V of stimulator compliance voltage headroom will be lost within the electrode trace per 1 mA of stimulation current. The impedance decreases with increasing frequency to a value of about 3 kΩ at 10 kHz. The observed behavior is consistent with a series resistor-capacitor model and reflects the physical construction of a resistive carbon traces and capacitive hydrogel electrode, such as modeled in [49]. Figure 4-5: Measured bandage substrate electrode impedance. 102 4.3 Battery Selection The larger device required for this particular study can accommodate a larger battery that will permit larger stimulation current amplitudes and will allow the device to be immediately re-used between dressing changes without recharging the battery. The relaxed battery size requirement allows battery chemistries and form factors other than coin cells to be considered for this application. Lithium-chemistry cells offer high energy density and those shown in Table 4-2 were evaluated. Non-rechargeable, cylindrical lithium cells provide the largest capacity but the diameter would increase the overall device thickness. Prismatic lithium polymer cells provide the desired capacity in a thin form factor. Table 4-2: Survey of high-capacity batteries for long-duration devices. Type Chemistry Lithium Thionyl Chloride (Li-SOCl2) * Note: liquid cathode is toxic Lithium (Li/FeS2) Lithium (CxF-Li) Lithium Ion (MnO2 Li) Rechargeable Lithium Polymer Rechargeable Size Dimensions Weight AA Ø 14 x 50 mm 18g AA Ø 14 x 50 mm 14.5g AAA Ø 10 x 44 mm 8.55g 123A Ø 17 x 34 mm 16.5 g 2/3A Ø 17 x 34 mm 13.5g RCR-123A Ø 5 x 4 cm Prismatic 35 x 25 x 5 mm 9g 51 x 34 x 6 mm 22g Capacity mAh Connector Voltage 3.6 V Max. Continuous Current 100 mA 2100 mAh Solder Tab 3000 mAh Brand Tadiran Button 1.5V 3A Energizer 1200 mAh Button 1.5V 1.5 A Energizer 1500 mAh Button 3V 1.5 A Energizer 1450 mAh Button 3V 2.5 mA Panasonic 600 mAh Button 3.7V 400 mAh 2-pin plug 3.7V 2C = 800 mA Tenergy 1000 mAh 2-pin plug 3.7V 2C = 2000 mA Tenergy Pearstone 103 4.4 Stimulator Module PCB A block diagram of the MSS3 stimulator module PCB is shown in Figure 4-6. A single-cell Lithium Polymer battery supplies power to the PCB and to the high-voltage boost converter portion of the ASIC. A buck-mode switching regulator steps down the battery voltage to 3.3 V and supplies power to the Bluetooth SoC microcontroller and ASIC. A battery fuel gauge IC monitors the battery voltage and calculates the state of charge using a proprietary algorithm. The microcontroller coordinates all functions on the PCB and includes a radio for wireless communication. The ASIC boost converter generates the high-voltage supply and the biphasic current-output stimulation. The PCB connects to the bandage gel electrodes via mechanical snaps (not shown). Figure 4-6: Block diagram of MSS3 Stimulator PCB 4.4.1 Wireless communication protocol and processor selection A radio operating in a band near 2.4 GHz is desirable because a ¼-wavelength antenna is only 31.25 mm long. Inverted-F and fractal (chip) topologies were considered for this application because they consume the least PCB area and allow the antenna to be closely integrated on the PCB or as a surface-mount component. A printed meandered-F 104 antenna was ultimately selected because it occupies the least board area [50], does not require additional components, and has a generally omnidirectional radiation pattern [51] suitable for use on an ambulatory animal. The key selection criterion for the wireless protocol is the power consumption. The power consumption is determined by several factors including active power consumption, wake-up time, and maximum channel period. For this reason, Bluetooth Low-Energy was chosen. The ANT protocol has similar characteristics to Bluetooth Low-Energy but its proprietary status increases system development cost and limits its usage with COTS devices such as smart phones. Furthermore, the CC2541 System on a Chip (SoC) serves as a general-purpose microcontroller with integrated Bluetooth LE radio. Table 4-3: A review of microcontrollers with built-in 2.4 GHz radios. Protocol Reference Part Frequency Band Year Intro. TX Power Data rate Wakeup Time Notes: >100 ms Max Connection Interval 65 sec Bluetooth 2.0 CC256x [52] 2.4 GHz 1994 1-3 Mbps Zigbee (802.15.4) CC2530Fxx [53] [54] ANT CC2570 [55] Bluetooth Low-Energy (BLE) CC254x nRF51822 [56] 915 MHz 2.4 GHz 2003 39 mA @ 4 dBm 29 mA @ 1 dBm 250 kbps 15 ms 250-786 sec Proprietary specification Integrated 8051 MCU Proprietary Interface Integrated 8051 MCU (8 bit) 2.4 GHz 2005 2.4 GHz 2011 18.2 mA @ 0dBm 1 Mbps 3 ms 240 sec 2.4 GHz 2011 10.5 mA @0dBm 1 Mbps 3 ms 240 sec 1Mbps 500 ms Not available as SoC ARM M0 (32 bit) 4.5 Firmware In order to maintain the required timing for the Bluetooth protocol, the firmware is written within a Real-Time Operating System (RTOS) provided with the microcontroller 105 development kit [57]. In an RTOS, code is executed as a series of tasks in order of descending priority. This helps ensure that the most time-critical functions are always executed within the required time. In this case, the BLE protocol stack has strict timing requirements in order to maintain the connection with the host. After all tasks have been executed, the microcontroller enters a low-power mode for the remainder of the current cycle. 4.5.1 Stimulation State machine In the MSS2 design, a state machine activated the boost converter and stimulator sequentially to generate each pulse. In the MSS3 architecture, the boost converter runs independent of the stimulator and the output voltage is continuously adjusted to maintain the compliance voltage required to deliver the programmed stimulation current. 4.5.1.1 Boost converter controller A block diagram depicting the operation of the boost converter controller is shown in Figure 4-7. Recall from 3.4.3.3 that, in steady state, the voltage applied to the stimulator cascode devices should be approximately mid-scale, indicating there is sufficient compliance voltage headroom to force the commanded current. The analog voltages V(IDAC_SENSE) and V(HVDD_SENSE) produced by the MSS ASIC are converted to digital values by the microcontroller ADC. The V(IDAC_SENSE) signal is sampled at the end of each stimulation pulse, and V(HVDD_SENSE) is continuously sampled. Both digital values are averaged using 16-point FIR filters implemented in firmware. IDAC_SENSE is compared to a mid-scale threshold to produce an error value that is 106 ultimately used to control the HVDD level. The HVDD setpoint value is compared to the sensed HVDD value at a 100-Hz rate and the result determines the state of the BOOST_EN control signal sent to the ASIC. Figure 4-7: Block diagram of boost converter controller. 4.5.1.2 Stimulation controller Stimulation pulses are controlled by setting the DAC output of the ASIC to the appropriate value, then setting the STIM_EN and CHSEL pins to drive current in the selected direction. Although there are software timer functions available within the RTOS, the execution time cannot be precisely controlled. Given that a Bluetooth timeslot is only 625 µs, it is undesirable to stall the program execution for up to 400 µs total during each stimulation pulse pair. For accurate timing of the stimulation pulse, the pulse generation should not rely on or be interrupted by other RTOS tasks. The stimulation pulse timing signals, STIM_EN and CHSEL, are generated entirely within the microcontroller hardware peripherals using two hardware timers operating in tandem. The timers also trigger two CPU interrupts that call functions to set the stimulator current for positive and negative pulses. This technique uses the on-chip logic that is commonly used in infrared transceivers to modulate a carrier wave with communication symbols. The stimulation pulse control signal generation process is 107 shown in Figure 4-8. Timer 3 operates as an 8-bit modulo counter with a period of 250 µs. Timer 3 output compare channel 0 produces the clock input to Timer 1 and has a period of 250 µs. Timer 3 channel 1 sets the pulse width for the positive and negative stimulation pulses. Timer 1 operates as a 16-bit modulo counter with a period from 1 ms to 32,000 ms that sets the stimulation pulse interval. (The maximum pulse interval is limited in software to 1000 ms.) The pulse width of Timer 1 channel 1 is fixed to 500 µs, the maximum combined width of two stimulation pulses. Timer 1 channel 2 produces the CHSEL signal used to select either a positive or negative current pulse. Timer 1 channel 2 also produces an interrupt on the falling edge of the first STIM_EN pulse that sends the negative pulse amplitude value to the SPI serial port. Timer 1 channel 3 produces a complementary interrupt on the falling edge of the second STIM_EN pulse that sends the positive pulse amplitude value in preparation for the next pulse pair. If mono-phasic current pulses are desired, the negative pulse width may be set to zero in the Timer 3 channel 1 register and a zero amplitude value loaded in the SPI port register. Figure 4-8: Generation of stimulation pulse timing signals using hardware timers 108 The large-timescale stimulator timing (minutes per hour and total hours) is straightforward and is controlled within a 1-Hz RTOS task that implements a Real-Time Clock function. 4.5.2 Bluetooth Interface The Bluetooth wireless interface is used to program the stimulation parameters and to monitor the stimulator operation while the device is in use on an awake animal. The protocol stack shown in Figure 4-9 is provided by the manufacturer as a pre-compiled firmware library. The custom firmware written for this device is contained within the application layer, GAP (Generic Access Profile), and GATT (Generic ATTribute) profiles. The application layer contains the RTOS tasks that implement the boost converter and stimulator controller state machines. The GAP role profile determines how the device will function within a Bluetooth network and sets the communication timing constraints. GATT profiles are collections of device data (called characteristics) such as the stimulation parameters, battery status information, and device measurements. Figure 4-9: Bluetooth Protocol Stack (from [57]) 109 Bluetooth devices can operate in one of four GAP roles: broadcaster, observer, peripheral, or central. This device operates as a peripheral, meaning it will periodically advertise itself to the network and will allow other devices to pair with it. This mode is more power-efficient than the broadcaster role because data is broadcast only when another device is listening. To communicate with the stimulator module, a host device (such as a PC a BLE dongle or the base station presented in 4.7) operates in the central role, meaning it listens for advertising devices and then creates a connection (called pairing) to communicate with them. In traditional networking terms, a peripheral device operates as a slave and a central device as the master. Regardless of the GAP role used, each device within the network may operate as either a data client, data server, or both. As shown in Figure 4-10, the base station (operating in the central GAP role) is a client that reads from and writes data to a bandage server (operating in the peripheral GAP role). Figure 4-10: Communication scheme for transferring parameters and measurements between base station and bandages. For common types of data (such as the current time or a heart rate measurement), data structures called Services are pre-defined by the Bluetooth standard [58]. There is 110 currently no existing standard Service for a stimulator device, so a custom service was created to organize the stimulation parameters. The parameters and measurements available through the Bluetooth protocol are listed in Table 4-4. The GATT profile implementation of this service is included in Appendix C.3. Table 4-4: Stimulator control parameters available through Bluetooth wireless link. Index 1 2 3 4 5 6 7 8 9 10 Parameter Battery Voltage Battery SOC Positive Pulse Amp. Negative Pulse Amp. Positive Pulse Width Negative Pulse Width Pulse Period Stim Minutes per Hour Stimulation Hours Stimulation Enable Data Type Byte Byte Byte Byte Byte Byte Byte Byte Byte Byte LSB Unit 20 mV 0.5 % 0.33 mA 0.33 mA 1 µs 1 µs 500 µs 1 min 6 hr 1 = run 0 = stop Data Range 0–5V 0-99 % 21 mA 21 mA 200 µs 200 µs 1-128 ms 1-60 min 1-1000 hr 0-1 Client Access Read Read Read/ Write Read/ Write Read/ Write Read/ Write Read/ Write Read/ Write Read/ Write Read/ Write 4.6 Device assembly 4.6.1 MSS3 Stimulator Module Assembly The MSS3 stimulator module PCB is assembled using standard surface mount reflow techniques and lead-free solder paste. The microcontroller is programmed using the CC Debugger programming tool [59], SmartRF Flash Programmer [60] program, and a custom probe cable. The custom cable used to connect the programmer to the PCB programming pins is shown in Figure 4-11. Figure 4-11: MSS3 microcontroller programmer connection using PCB probe. 111 The packaged MSS ASIC is hand-soldered to the PCB after the other module components have been electrically tested to avoid potentially damaging this high-cost part. The plastic battery connector is installed, and female component of the electrode snaps is soldered to the bottom side of the PCB, as shown in Figure 4-12. The completely-assembled PCB is tested using the procedures described in Section 4.8. Figure 4-12: Snap connectors mounted on bottom side of MSS3 PCB module. The assembled and tested PCB is covered with a conformal coating of parylene to protect the components from moisture and other potential contamination in the wound environment. This step was performed by Jeremy Dunning of the APT Center. A thin film chemical vapor deposition process [61] is used to apply the layer of p-xylylene polymer (trade name Parylene®) with a target thickness of 25 µm. The PCBs are suspended in the reaction vessel to apply a uniform coating to all surfaces. The polymer coating is manually removed from the battery connector terminals and electrode snaps to re-expose these electrical connections. Electrical performance of the module is rechecked after the coating process. The MSS3 stimulator module PCB is integrated with the Lithium Polymer battery using clear heat shrink tubing [62] with a pre-shrunk diameter of 38.1 mm. The 112 arrangement of the PCB and battery is shown in Figure 4-13. This semi-permanent attachment method does not damage either component and provides an additional layer of protection from the wound environment. Figure 4-13: MSS3 Stimulator Module PCB Assembled with Battery. 4.6.2 Device Sterilization The MSS3 stimulator modules and substrates must be sterilized prior to use in animal or human studies. Medical instruments are commonly sterilized using steam in an autoclave at 121 ˚C -132 ˚C [63]. The LCP substrate [64] is rated for continuous operation at 150˚C and has low water absorption (0.04 % at 23˚C over a 24 hour period) and may be suitable for steam sterilization. However, the conductive paint used for the traces is not indicated for use at high temperature. Ethylene Oxide is a low-temperature, dry gas sterilization process that takes place at temperatures ranging from 30 ˚C - 60 ˚C. However, the process is hazardous and requires special validation after processing to ensure no carcinogenic compounds [65] remain on the substrates or devices before use. Ultimately, the STERRAD® process was selected for processing the MSS3 devices and substrates. The process uses hydrogen peroxide plasma and operates at a temperature 113 range from 47 ˚C -56 ˚C [66]. In addition, this process is specifically indicated for sterilizing Lithium polymer batteries, and this will permit the stimulator and battery to be sterilized simultaneously. The stimulator modules and substrates are individually packaged in Tyvek® pouches [67] specifically designed for low-temperature sterilization processes. The packed substrates are shown in Figure 4-14, and the stimulator module PCBs are packed in the same manner. Figure 4-14: Bandage substrates packaged for sterilization. 4.6.3 Intra-Operative Bandage Assembly The reusable MSS3 stimulator module is connected to a fresh bandage substrate during the initial surgery and each dressing change. Using sterile handling, the modules and bandages are removed from their packaging and delivered to the surgical field. The stimulator modules are sterilized with the battery unplugged, so this connection must be made prior to use. The adhesive strip on the top size of the bandage is exposed, then the 114 module mounted and the snap connections made. The device is now ready for application to the animal and activation using the radio interface. 4.7 Wireless Base Station In previous research studies using stimulation bandages, devices were occasionally damaged or removed by the animal. Future versions of the stimulation bandage may include additional sensors for monitoring the wound healing process, and this feedback could be provided to the clinician without creating a physical connection to the device on the awake animal. The stimulation bandage is controlled by manipulating the GATT profile values using either of two PC software programs that are currently available from Texas Instruments [68, 69]. These programs provide limited functionality and have three key limitations that reduce their effectiveness within the laboratory. First, there is no method to display the binary-coded data values read from the bandage in a meaningful representation to the clinical user. Second, neither program allows more than one simultaneous connection, so they cannot be used to monitor two operating bandages as required by the animal study. Third, the programs lack the ability to continuously log values read from the bandage (such as the battery voltage or the stimulator compliance voltage), and the microcontroller on the stimulation bandage has limited data memory to store measurement data. To address the limitations of the existing BLE interfaces, the wireless base station shown in Figure 4-15 is proposed. Each MSS bandage communicates using their Bluetooth radio to a central host radio. Although only two bandages will be active on an 115 individual animal, the Bluetooth network supports up to 6 simultaneous connections and there may be multiple animals in the facility at a given time. The base station host radio is controlled by an application processor that requests data from each of the connected bandages at regular intervals (typically 1 set of measurements every 30 seconds). The application processor then stores the received bandage data to a SD flash memory card for post-experiment analysis. The processor also transmits the measurements to an online data server through a WiFi radio connected to the wireless network within the facility. Figure 4-15: Communication network between stimulator modules, base station, and internet data server. 4.7.1 Base station construction The base station is constructed using components based on the open-source Arduino microcontroller development platform [70]. The bottom board in the stack is an Arduino 116 Mega 2560 microcontroller “board” containing an Atmel ATmega2560 8-bit microcontroller that executes the application firmware. Three “shields” stack on the microcontroller board: a WiFi shield [71] to connect to a wireless internet access point, a BLE shield [72] to communicate with the MSS3 stimulator modules, and a LCD shield [73] to provide display the four MSS3 battery voltages to investigators while in the facility. A piezo transducer provides audible cues to clinicians during device activation, and a fuel gauge module monitors the base station battery charge. The base station is powered from a PC through a USB cable during testing, and a 2200 mAh rechargeable Lithium Polymer battery pack when deployed in the animal facility. The base station hardware is shown in Figure 4-16. Figure 4-16: Wireless base station hardware PCB stack. The PCB stack is mounted in a plastic enclosure to protect the electronics while inside the animal facility. The base station is then placed outside the animal pen and requires no clinician interaction after initial activation of the bandages. 117 4.7.2 Base station operation The two main functions of the base station are to program the stimulation parameters into each bandage and to monitor performance during the experiment. The ability to program the devices wirelessly avoids the use of exposed signal pins on the stimulator module PCB that may be damaged during use on the animal. Continuous remote monitoring improves research data quality by helping to identify when bandages have been displaced from the wounds or the device battery charge is low. 4.7.2.1 Programming bandage stimulation parameters The bandage programming process is diagrammed in Figure 4-17. Device stimulation parameters are loaded from the PARAMS.CFG text file stored on a micro SD card installed on the WiFi shield. Users may modify stimulation parameters using a text editor on a PC, then transfer them to devices using simple prompts on the base station LCD display. The stimulation parameters are then written into each bandage according to the BLE bandage profile given in Appendix C.3. Figure 4-17: Procedure for programming bandage stimulation parameters. 4.7.2.2 Wireless bandage data transfer Each bandage transmits its performance information within the characteristic ID as part of the periodic BLE connection maintenance. The data samples for each device are 118 stored in separate files on the SD card. The samples are also uploaded to a secure internet server via the WiFi shield for continuous experiment monitoring while outside the animal facility. The base station reads measurements from the Bluetooth profile of each bandage every 30 seconds and stores the data in CSV text files on the micro SD card. These log files may be transferred to a PC using a SD card reader at the conclusion of the experiment. The base station uploads a measurement from each bandage to an internet data server every 10 minutes. 4.7.2.3 Base Station Clock Synchronization The Mega 2560 processor board is constructed using a 16-MHz resonator [71] with initial frequency tolerance of 0.5%. Over a period of 28 days, the accumulated time error could potentially exceed 200 minutes. The frequency error is compensated in two ways: first, a calibration value is loaded into a clock control register from EEPROM to reduce the timing error within the microcontroller. Second, the base station connects to the US Naval Observatory Master Clock time server [74] using the WiFi shield and downloads the correct time (within 1 second) once per hour. 4.8 Benchtop Testing The assembled stimulator module PCB shown in Figure 4-18 measures 31 mm tall by 38 mm wide and weighs 4.9 grams. The two-layer PCB is constructed using 0.062-in thick FR-4 laminate and 1-oz copper traces (equivalent to 1.4 mil thick [75]). The schematic drawings are provided in Appendix C.1 and the assembled PCB is shown in Figure 4-18. 119 Figure 4-18. MSS3 printed circuit board (actual size 38 mm x 31 mm) There are three main objectives in the stimulator module PCB benchtop testing. First, the function of the MSS ASIC when battery-powered and operated by the microcontroller was confirmed. Second, the wireless performance of the stimulator module PCB was measured. Third, the operating current of the device was measured under several operating modes and the battery lifetime was estimated. 4.8.1 ASIC validation 4.8.1.1 Boost converter The Lithium Polymer battery used in this version of the device has a low internal resistance, and the current-limiting boost converter circuits designed for coin batteries on the MSS ASIC are not necessary. More importantly, the nominal cell voltage of a Lithium Polymer battery is outside the designed input voltage range of the current sensing amplifier. Therefore, the boost converter on the MSS3 is operated using a PWM signal generated by the microcontroller. As shown in Figure 4-19, the boost converter output charges to about 35 V within a few ms when the MSS3 is powered by a Lithium Polymer battery using a PWM frequency of 125 kHz and duty factor of 50%. Large voltage ripple occurs when the output current is small because the boost converter outer 120 control loop code is executed once per 1 ms. In section 4.8.1.2, it will be shown that the variations in the boost converter output (HVDD) do not affect the current delivered to the electrodes. Figure 4-19: Battery-powered MSS3 boost converter output charging waveform (VBATT = 4.15 V) . 4.8.1.2 Stimulator Biphasic operation of the MSS3 stimulator is demonstrated in Figure 4-20 using a 1 kΩ resistor load and the stimulation parameters in Table 4-5. The STIM_EN signal controls the pulse width of the positive (cathodic) and negative (anodic) pulses. The CHSEL signal determines if the current flowing from ELECA to ELECB to positive (CHSEL = L) or negative (CHSEL = H). The stimulator outputs ELECA and ELECB remain biased at HVDD between pulses. The common mode level drops during the stimulation pulses due to trace impedance. The current delivered to the load is constant during each pulse despite voltage fluctuations on the HVDD supply. 121 Figure 4-20: Demonstration of the MSS3 stimulator delivering a biphasic pulse. Table 4-5: Stimulation Parameters used in MSS3 stimulator demonstration. Parameter Positive Pulse Current Positive Pulse Width Positive Pulse Current Positive Pulse Width Pulse Period HVDD Value 5 mA 100 µs 2.5 mA 200 µs 80 ms 35 V 122 4.8.2 Wireless Communication 4.8.2.1 Range test The Bluetooth radio is specified to operate at distances up to 10 m, but actual performance is extremely dependent on the PCB layout and device construction. In this application, the typical distance is only 3 m, and the radio transmitter and receiver power settings will be reduced automatically by the radio to extend battery life. In this experiment, the BLE USB dongle is attached via a USB cable to a PC and the MSS3 stimulator module is positioned at a variety of distances in an office environment. The stimulator module is powered from a battery attached to PCB with heat shrink in the same manner as the animal-ready device. The RSSI value reported by the SoC varied within a few codes, so repeated measurements (N=5) were taken at each distance to determine the average received signal strength (RSSI). The internal RSSI measurement range for the part is -79 dBm to -15 dBm [76] with an accuracy of ± 6 dBm. The plot shown in Figure 4-21 indicates that a signal of -65 dBm is received at a distance of 3.05 m. Figure 4-21: Self-reported received signal strength (RSSI) from CC2541 radio as a function of distance from USB dongle. 123 4.8.3 Operating current test The battery current is measured for each of the three operating modes of the device: stimulation, sleep, and communication. The stimulator current consumption was measured using a 4.1 V power supply in place of the LiPoly battery and the stimulation parameters previously listed in Table 4-5. When a 450 mAh LiPoly cell is used, the MSS3 will operate for 161 hours. To achieve a 1-week battery lifetime, the stimulation time should be reduced from 10 to 9 minutes per hour. Table 4-6: Measured operating current and projected device lifetime for several compliance voltage levels. Mode Standby Communication Stimulation Measured Current 300 µA 8.60 mA average 15 mA Total calculated average current consumption Estimated lifetime (450 mAh battery): % of operation time 50 min/ hr = 83% 15 ms/ 30 sec = 0.05% 10 min/ hr = 17% Contribution to Average Current 250 µA 4.3 µA 2.55 mA 2.8 mA 161 hr 4.9 Clinical Validation 4.9.1 Method An acute test of the surface stimulation bandage was performed to validate the electrical and mechanical function of the device. First, the infected wound model is created on an anesthetized pig according to the IRB-approved study protocol. The bandage is affixed to the animal skin using pre-cut pieces of sterile hydrocolloid dressing [77] as shown in Figure 4-22 . Light pressure is applied to the dressing for approximately 2 minutes (per the product instructions) to increase the bandage adhesion. Bonding is 124 confirmed by manually applying a small sheer stress load to the ventral edge of the bandage. Figure 4-22: Photograph of bandage with adhesive hydrocolloid material. 4.9.2 Bandage conductivity test The first test is to measure the skin impedance through bandage electrodes using a benchtop LCR meter connected to the electrode snaps. The stimulus voltage is a 1-V sine wave with frequency varying from 100 Hz – 100 kHz and the resulting current is below the threshold of sensation. Impedance was measured across the electrodes five minutes after the substrate was applied to the skin. Two substrates were tested using this method, and the magnitude of the measured impedance is lower than 6 kΩ above 10 kHz for both samples, as shown in Figure 4-23. Approximately 3 kΩ of the measured impedance is attributed to the substrate traces and electrodes. A trace on substrate 1 partially cracked during the test, and a large impedance magnitude of 64 kΩ was measured at 100 Hz. This test confirms the bandage is making reliable electrical contact with the skin and the impedance is within the range predicted by benchtop tests. 125 Figure 4-23: Measured MSS3 substrate electrode impedance measured in an acute test on living pig skin. 4.9.3 Acute stimulation test The second electrical test is to activate the bandage stimulator using the on-board wireless Bluetooth radio link and a laptop PC to verify the electrical operation of the device in-situ. The stimulator was configured to generate 50 µs pulses of 5 mA and -2.5 mA to the skin, and the circuit shown in Figure 4-24 was used to measure current flowing from each electrode. An oscilloscope was used to record the stimulation current pulse flowing into the skin (RSKIN) through two 10 Ω sense resistors (RTEST) mounted on the stimulator module PCB. These current sense amplifiers are unipolar and only measure positive current values. For this test, commercial electrodes [78] were used in place of the MSS3 substrate to connect the stimulator to the skin. 126 Figure 4-24: Acute test schematic for measuring current flowing through tissue. The measured electrode voltage and current waveforms are shown in Figure 4-25. Compared to the test using a resistor load (Figure 4-20), the acute test shows the effect of the complex impedance associated with the hydrogel electrodes and skin on the electrode voltage waveform. The positive and negative current pulses are delivered with the expected 50 µs width, but a third pulse is created between the two when the high-side PMOS devices switch state. This behavior is consistent with a series capacitive element in the electrodes or skin discharging back into the stimulator. A smaller cathodic current pulse is generated when t≈1500 µs and the PMOS switch to their original state. 127 Figure 4-25: Electrode voltage and current waveforms measured in acute stimulation test. 4.10 Summary Based on benchtop and acute in-vivo test results, the MSS3 device is suitable for use in long-term animal studies that require a programmable stimulator with wireless communication capability. Battery lifetime is dependent on the capacity of the LiPoly cell used and the selected stimulation parameters. A MSS3 that uses a 450 mAh battery to stimulate for 9 minutes per hour is expected to operate for 7 days before recharging is necessary. 128 5 Conclusions and Future Work Wound electrotherapy is currently applied in short sessions in a clinical setting, typically no more than 1 hour in duration, on a daily basis. Treatment duration is limited by caregiver resources and patient acceptance (e.g., restricted movement due to tethers). The objective of this research was to develop a wearable stimulation system suitable for research studies that could be readily adapted for use in a range of applications and species. The new medical devices presented in this work provide a pathway to new alternatives that will deliver continuous electrotherapy in a less-invasive manner. 5.1 Achievements The first bandage presented in this work was designed for use with a rat wound model and consists of a stimulator PCB module constructed from off-the-shelf components and is powered by a small button cell battery providing at least seven days of continuous use. This voltage-mode device generates stimulation pulses that are 10 – 90 V in amplitude, 10 – 200 µs in width, and 12 – 25 Hz in frequency. A disposable plastic electrode substrate was co-developed that transmits the stimulation current from the PCB to the skin. Stimulation was typically applied to wounds for 10 minutes every hour for one week, and then the device and wound dressings were replaced. An ASIC has been developed using the OnSemi 0.7-µm I2T100 process and is capable of operation up to 100 V. A high-gain, current-mode boost converter addresses challenges associated with efficient generation of the large amplitude compliance voltage (up to 90 V) from a small battery with limited output current capability. In particular, the control circuits limit the peak inductor current on a cycle-by-cycle basis to avoid battery brown-out. A biphasic current mode stimulator was also demonstrated with ±21-mA 129 output range, 0.33-mA resolution, and a voltage headroom requirement of 3.5 V at full scale output. The digitally-controlled current is programmed using a SPI serial interface. The second stimulation bandage uses this ASIC and is designed for use on larger wounds, up to 6 cm in diameter. A rechargeable lithium polymer battery allows a single stimulator module to be used for an entire 28-day study. The battery may be quickly recharged as needed during weekly dressing changes. The disposable electrode bandage portion of the device is easily replaced in-situ by the clinician. Continuous monitoring of the delivered stimulation current while the device is in place on an animal is achieved using a microcontroller with a built-in Bluetooth Low Energy radio. Performance information from up to six devices is recorded to a wireless base station located outside the animal cage and may also be remotely accessed by research personnel. A summary of the key features of the two new devices are presented in Table 5-1. 130 Table 5-1: Electrotherapy Device Summary. Target species Bandage1 Substrate Traces Electrodes Connector Stimulator PCB Dimensions Mass (w/o battery) Communication Battery Type LG TEC-ELITE [15] TENS device ISSD [16] Previous work MSS1 Discrete device MSS3 Integrated device Human Rabbit Rat Pig Individual Wire leads 3.8 cm x 3.8 cm plug Polyimide Platinum ECG gel, 1 cm x 2.5 cm Folded tab LCP Nickel under acrylic Hydrogel, 1 cm x 2.5 cm Conductive tape LCP Carbon under acrylic Hydrogel, 1 cm x 8 cm Snap (handheld) N/A 44 mm x 32 mm 6.8 g 25 mm x 25 mm 3.4 g 38 mm x 31 mm 4.9 g LCD Rechargeable 9V IrDA (infrared) Lithium (CR1632) Wired Serial Zinc-Air (675 size) 130 mAh @ 3.0 V < 7 days Yes Voltage RC 620 mAh @ 1.4 V 7 days Yes Voltage RC Bluetooth 4.0 (RF) Lithium Polymer (model 063048) 500 mAh @ 3.7 V 28 days (rechargeable) Yes Current (biphasic rectangular) Battery Capacity Lifetime Reusable Pulse Type: N/A 24 hrs Yes Current Biphasic square 1. Developed with Jeremy Dunning 5.2 Future work Further technological development is needed to adapt the existing device for additional pre-clinical studies using animal wound models and to prepare a human wound electrotherapy device suitable for widespread clinical use. In future studies using a small mammal (e.g. rodent) wound models, a stimulator module using the MSS ASIC could be designed to take full advantage of the currentmode boost converter circuitry using small batteries with limited output current capability. Given the small size of the animal, a reusable stimulator module with a flexible PCB would reduce unwanted skin irritation and increase animal comfort. 131 Future versions of the ASIC could include additional circuits to adaptively control the compliance voltage supplied by the boost converter based on a feedback signal either generated within the stimulator (such as IDAC_SENSE) or one based on a direct current measurement. If required by the electrotherapy study, the stimulator current resolution or range may be easily expanded. The stimulation pulse waveform timing signal generation could also be integrated on-chip with the addition of extra SPI serial port registers and logic circuits. This would reduce the microcontroller interaction and may reduce overall power consumption. The ASIC-based stimulator module with wireless communication capability can accommodate sensors to provide continuous information to researchers about the state of wound healing in a particular subject. For example, a temperature sensor would indicate the degree of inflammation, and an pH sensor might detect changes in tissue oxygenation. Although it was not the focus of this work, the design and manufacturing methods used to fabricate the electrode substrate portion of the devices need further refinement to enable large-scale production. A clinic-ready device will likely have the stimulator electronics assembled directly on a flexible bandage. 132 Appendix A: Supplement Information for Chapter 2 (MSS2) A.1: MSS2 Schematic 133 Appendix B: Supplemental Information for Chapter 3 (ASIC) B.1 MSS ASIC Test Fixture Schematic 134 Appendix C: Supplemental Information for Chapter 4 (MSS3) C.1: MSS3 Module Schematic 135 C.2 Stimulation Bandage GATT profile 136 C.3 MSS3 Parameter programming procedure using TI BLE Device Monitor Software: TI BLE Device Monitor V 1.1 or later. Equipment: TI CC2540 USB Dongle (included with CC2540 Mini Development Kit) loaded with the CC2540_USBdongle_HostTestRelease_All.hex firmware image. Step 1: Install USB dongle into PC and start the Device Monitor Application. A red LED on the dongle should be illuminated and the dongle address should appear in the BLE Network pane. Step 2: Make sure the MSS3 bandage battery is plugged in, then click scan in the BLE Network pane. The MSS3 device should appear under the Host device. 137 Step 3: Double-click on the MSS3 device, and the program will discover the characteristic parameters used to program the device. Step 4: To read a parameter value, double-click on the desired characteristic name (in the mnemonic column). To write a parameter value, double-click on the value column and enter the desired value. (If an entry box does not appear, then the parameter is read-only.) It is good practice to read back a parameter after writing a new value to verify the change has taken place. 138 C.4 Base station Arduino programming procedure Software: Arduino IDE (available from www.arduino.cc) Equipment: USB cable, Base Station hardware Step 1: Connect Arduino Mega 2560 processor board to computer using the USB cable. A serial port driver may automatically be installed. Step 2: Start the Arduino IDE software and open the base station project: File SketchbookSketchesbase_station. Step 3: Verify the correct serial port and board are selected: Tools Serial Port <COM port> Tools Boards Arduino Mega2560 or Mega ADK Step 4: Click the upload button (right arrow) to compile and upload the firmware to the Arduino Mega2560 board. It may take a few moments to compile the project before uploading begins. Step 5: If desired, open the serial monitor to verify the base station boots successfully. Otherwise, the base station may be disconnected from the computer. 139 C.5 Base station BLE112 programming procedure Software: BlueGiga BLE Update, bglib_master firmware image (project.bgproj). Equipment: CC Debugger with mini 10 pin ribbon cable (included with CC2540 Mini Development Kit), BLE Shield, Arduino processor board, USB cable. Note: The BLE112 module on the BLE Shield is controlled using commands issued from the BGLIB library on a host processor. This procedure should only be used to program a BLE112 on a newly-fabricated BLE shield. Step 1: Install the BLE shield on an Arduino microprocessor board to safely supply power to the BLE112. Connect the CC Debugger to the BLE shield using the ribbon cable (observe polarity markings for pin 1), Step 2: Connect the Arduino and CC Debugger to the PC using one standard USB cable and one “mini B” USB cable. Press the button the CC Debugger and observe the LED change to green, indicating a successful connection to the BLE112. Step 3: Start the BlueGiga BLE Update program and select the correct port and firmware image. Leave the license key field blank. Step 4: Click “Update” and wait for the CC Debugger to program the BLE112. The dialog window will turn green when the part is successfully programmed. Step 5: Unplug the ribbon cable from the BLE shield, then unplug the CC Debugger from the PC. Leave the Arduino board stack connected to the PC. Step 6: Use the serial monitor within the Arduino IDE or another terminal program to observe the base_station program serial output stream. Verify the BLE112 responds during the initialization routine. 140 C.6 MSS3 CC2541 programming procedure Software: TI Flash Programmer, IAR Embedded Workbench IDE. Equipment: CC Debugger with ribbon cable and custom 5 pin probe cable Step 1: Connect the CC Debugger to the PC using a USB cable. The LED should illuminate red, indicating power is present but no target device has been found yet. Step 2: Connect the MSS3 PCB to the CC Debugger using the ribbon cable and 5 pin probe. Apply power to the board using a battery or lab supply. Press the button the CC Debugger and observe the LED change to green, indicating a successful connection to the CC254x microcontroller. Step 3: Start the IAR Embedded Workbench IDE application. Load the project workspace, modify the code as necessary, then re-compile by pressing F7. Step 4: The on-chip debugger may be invoked by pressing Ctrl-D or selecting “Debug Download” from the menu. Start program execution by pressing F5, and terminate the debug session by pressing Ctrl-Shift-D. Step 5: To program a firmware image into the Flash memory, start the TI Flash Programmer application. Note that the debugger must not be active in the IDE in order to use the programmer. The CC254x device on the MSS3 should appear in the System-onChip window. Step 6: Select “Erase, Program, and Verify” option from the Actions menu, then click the “Perform Actions” button to program the part. A message will appear in the text box below the button indicating if the part was successfully programmed. 141 References [1] C. H. Lyder, "Pressure Ulcer Prevention and Management," Journal of the American Medical Association, vol. 289, no. 2, pp. 223-226, 2003. [2] T. Mustoe, "Dermal ulcer healing: Advances in understanding their pathogenesis," in Tissue repair and ulcer/wound healing:molecular mechanisms, therapeutic targets and future directions, Paris, 2004. [3] M. Stacey, "Wounds International: The chronic wound debate," 2 9 2010. [Online]. Available: http://www.woundsinternational.com/made-easys/the-chronic-wounddebate. [Accessed 11 10 2012]. [4] M. Plotzke, R. Subramanian and L. Olsho, "Cost Reductions Associated with Declines in Pressure Ulcers at MHA Hospitals," Abt Associates Inc., Cambridge, MA, 2011. [5] W. E. Staas and H. M. Cioschi, "Pressure sores--a multifaceted approach to prevention and treatment.," The Western Journal of Medicine, vol. 5, no. 154, pp. 539-544, 1991. [6] K. Woo, "Chronic Wound-associated Pain, Psychological Stress, and Wound Healing.," Surgery Technology International, vol. 9, no. 2012, 2012. [7] K. Brismar, "The diabetic foot- diagnosis of the foot at risk and managment of foot complications," 2008. [Online]. Available: http://www.ndteducational.org/brismarslide2008txt.asp. [Accessed 11 10 2012]. [8] J. H. Calhoun, C. K. Murray and M. M. Manring, "Multidrug-resistant Organisms in Military Wounds from Iraq and Afghanistan," in 2007 Meeting of the Musculoskeletal Infection Society, San Diego, 2008. [9] G. Talebi, G. Torkaman, M. Firouzabadi, M. Mofid, S. Shariat and S. Kahrizi, "Effects of micro-amperage direct current stimulation on injury potential and its relation to wound surface area in guinea pig," in Proceedings of the IEEE Engineering and Medicine in Biology Society, Lyon, 2007. [10] K. C. Balakatounis and A. G. Angoules, "Low-intensity Electrical Stimulation in Wound Healing: Review of the Efficacy of Externally Applied Currents Resembling the Current of Injury," Open Access Journal of Plastic Surgery, pp. 283-291, 16 May 2008. [11] B. McLeod, N. Wellman and S. M. Fortun, "Bacterial biofilms and the bioelectric effect.," Antimicrobial Agents and chemotherapy, vol. 40, no. 9, pp. 2012-2014, 1996. 142 [12] L. C. Kloth, "Electrical Stimulation for Wound Healing: A Review of Evidence From In Vitro Studies, Animal Experiments, and Clinical Trials," International Journal of Lower Extremity Wounds, vol. 4, no. 1, pp. 23-44, 2005. [13] A. Jercinovic, R. Karba, L. Vodovnik, A. Stefanovska, P. Kroselj, R. Turk, I. Dzidic, H. Benko and R. Savrin, "Low Frequency Pulsed Current and Pressure Ulcer Healing," IEEE TRANSACTIONS ON REHABILITATION ENGINEERING, vol. 2, no. 4, pp. 225-233, 1994. [14] S. A. Weber, P. A. Vonhoff, F. J. Owens, A. Byrne and E. T. McAdams, "Development of a Multi – Electrode Electrical Stimulation Device to Improve Chronic Wound Healing," in 31st Annual International Conference of the IEEE EMBS, Minneapolis, 2009. [15] LG Med Supply, ""LG-TEC ELITE" DIGITAL Dual COMBO TENS Unit and Muscle Stimulator with AC Adapter, Battery, Carrying Case, Electrodes," 25 November 2011. [Online]. Available: http://www.lgmedsupply.com/lgelteunandm.html. [Accessed 20 February 2013]. [16] D. S. Howe, "Electronics and Communication Technology for a Surface Stimulation Device," M.S. Thesis, Dept. Elect. Eng. and Comput. Sci., Case Western Reserve Univ., Cleveland, OH, 2009. [17] M. K. Henzel, D. S. Howe, J. Graebert and K. M. Bogie, "OPTIMIZATION OF A RAT ISCHEMIC WOUND MODEL FOR EVALUATION OF THE EFFECTS OF ELECTROTHERAPY ON WOUND HEALING," in Wound Healing Society, Atlanta, 2012. [18] L. J. Gould, M. Leong, J. Sonstein and S. Wilson, "Optimization and validation of an ischemic wound model," Wound Repair Regen, vol. 13, no. 6, p. 576, 2005. [19] I. Buchmann, "Battery University: Primary Batteries," [Online]. Available: http://batteryuniversity.com/learn/article/primary_batteries. [Accessed 2013 February 2013]. [20] D. Rohm, "Primary Zinc-Air: New Battery Option for Portable Device Design," in Wireless Portable Symposium & Exhibition, San Jose, 2001. [21] Maxim Integrated Products, "1-Cell to 2-Cell, Low-Noise,High-Efficiency, Step-Up DC-DC Converter," MAX1798 Datasheet, Jul. 1998. [22] Microchip Corporation, "PIC24F16KA102 Family Data Sheet," Microchip, Chandler, 2009. [23] Maxim Integrated Products, "1.5μA IQ, Step-Up DC-DC Converters in Thin SOT23-5," MAX1722/MAX1723/MAX1724 Datasheet, Jul. 2001. 143 [24] Novacap, "Technical Brochure," 6 November 2001. [Online]. Available: http://www.novacap.com/PdfFiles/tech_brochure.pdf. [Accessed 9 February 2012]. [25] Noraxon U.S.A. Inc., "The ABC of EMG: A Practical Introductionto Kinesiological Electromyography," Scottsdale, 2005. [26] AMI Semiconductor, "I2T100 Design and Layout Manual," Belgium , 2004. [27] L. Pylarinos, "Charge Pumps: An Overview," in IEEE International Symposium on Circuits and Systems, Bangkok, 2003. [28] A. I. Pressman, K. Billings and T. Morey, Switching Powering Supply Design, 3rd ed., New York: McGraw-Hill, 2009. [29] S. Majerus, "A Low-power wireless transceiver for deeply implanted biomedical devices," Dept. Elect. Eng. & Comput. Sci., Case Western Reserve Univ., Cleveland, OH, 2008. [30] Diodes Inc., "N-CHANNEL ENHANCEMENT MODE MOSFET BSS123W," April 2011. [Online]. Available: http://www.diodes.com/datasheets/ds30368.pdf. [Accessed 31 10 2012]. [31] Linear Technology, "Power Management for Portable Products," Product Solutions Manual, Vol 3, 2009. [32] Texas Instruments, "Power Management Selection Guide," 2009. [33] B. Rivera and J. Baker, "Design and layout of schottky diodes in a standard CMOS process," Research Presentation, Boise State University, Boise, 2001. [34] V. Milanovic, M. Gaitan, J. C. Marshall and M. E. Zaghloul, "CMOS Foundry Implementation of Schottky Diodes for RF Detection," IEEE Transactions on Electron Devices, vol. 43, no. 12, pp. 2210-2214, 1996. [35] Diodes Incorporated, "Surface Mount Schottky Barrier Diode," BAT46W Datasheet, 2008. [36] S. L. Garverick, Personal communication, 2011. [37] S. C. Tan and X.-W. W. Sun, "Low power CMOS level shifters by bootstrapping technique," Electronics Letters, vol. 38, no. 16, pp. 876-878, 2002. [38] M. Ghovanloo and K. Najafi, "A Compact Large Voltage-Compliance High OutputImpedance Programmable Current Source forImplantable Microstimulators," IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, vol. 52, no. 1, pp. 97-, 2005. 144 [39] K. E. Kuijk, "A Precision Reference Voltage Source," IEEE Journal of Solid-State Circuits, vol. 8, no. 3, pp. 222-226, 1973. [40] N. Weste and D. Harris, CMOS VLSI Design: A Circuits and Systems Perspective, 3rd ed., New York: Addison Wesley, 2004. [41] M.-D. Ker, "Whole-Chip ESD Protection Design with Efficient VDD-to-VSS ESD Clamp Circuits for Submicron CMOS VLSI," IEEE TRANSACTIONS ON ELECTRON DEVICES, vol. 46, no. 1, pp. 173-183, 1999. [42] S. J. Landis, "Chronic Wound Infection and Antimicrobial Use," Advances in Skin & Wound Care: The Journal for Prevention and Healing, vol. 21, no. 11, pp. 531-540, 2008. [43] K. Petersen, M. S. Riddle, J. R. Danko, D. L. Blazes, R. Hayden, S. A. Tasker and J. R. Dunne, "Trauma-related infections in battlefield casualties from Iraq," Annals of Surgery, vol. 245, no. 5, pp. 803-11, 2007. [44] S. Patel, "THE IMPACT OF MRSA ON WOUND HEALING," Wound Essentials, vol. 2, pp. 144-148, 2007. [45] K. M. Bogie, Personal communication, Cleveland, 2012. [46] J. Macknin, M. Cover and K. Bogie, "Development of a porcine model for acute infected wounds," Orthopaedic Journal, vol. 7, no. 1, pp. 62-66, 2010. [47] Prym Consumer USA, "Dritz Fasteners," 21 June 2007. [Online]. Available: http://www.dritz.com/tips/images/fasteners.pdf. [Accessed 12 February 2013]. [48] Bare Conductive, "Bare Paint Technical Datasheet," 9 May 2012. [Online]. Available: www.bareconductive.com. [Accessed 12 December 2012]. [49] E. Barsoukov and J. R. Macdonald, Impedance Spectroscopy Theory, Experiment, and Applications, Hoboken: John Wiley & Sons, Inc., 2005. [50] Texas Instruments, "Application Note AN058 (SWRA161B): Antenna Selection Guide," 5 October 2010. [Online]. Available: http://www.ti.com/lit/an/swra161b/swra161b.pdf. [Accessed 26 Feburary 2013]. [51] Texas Instruments, "Application Note AN043: Small Size 2.4 GHz PCB Antenna," 4 April 2008. [Online]. Available: http://www.ti.com/lit/an/swra117d/swra117d.pdf. [Accessed 26 February 2013]. [52] Texas Instruments, "CC256x Bluetooth Smart Ready Controller," 1 October 2012. [Online]. Available: http://www.ti.com/product/cc2560. [Accessed 20 February 2013]. 145 [53] Texas Instruments, "CC2530 Second Generation System-on-Chip Solution for 2.4 GHz IEEE 802.15.4 / RF4CE / ZigBee," [Online]. Available: http://www.ti.com/product/cc2530. [Accessed 26 February 2013]. [54] NewCircuits, "Zigbee Protocol," 2010. [Online]. Available: http://www.newcircuits.com/article.php?id=tut004. [Accessed 26 February 2013]. [55] Texas Instruments, "CC2570 Single Channel ANT RF Network Processor," [Online]. Available: http://www.ti.com/product/cc2570?DCMP=cc3000&HQS=cc3000-pr-pf3. [Accessed 26 February 2013]. [56] Nordic Semiconductor, "nRF51822: Bluetooth low energy and 2.4GHz proprietary SoC," [Online]. Available: http://www.nordicsemi.com/eng/Products/Bluetooth-Rlow-energy/nRF51822/(language)/eng-GB. [Accessed 20 February 2013]. [57] Texas Instruments, "SWRU271D: CC2540 Bluetooth® Low Energy Software Developer’s Guide v1.3," 13 July 2011. [Online]. Available: http://www.ti.com/lit/ug/swru271d/swru271d.pdf. [Accessed 9 October 2012]. [58] Bluetooth SIG, "Current Time Service (CTS)," 15 September 2011. [Online]. Available: http://developer.bluetooth.org/TechnologyOverview/Documents/CTS_SPEC.pdf. [Accessed 12 February 2013]. [59] Texas Instruments, "Debugger and Programmer for RF System-on-Chips," 23 February 2012. [Online]. Available: http://www.ti.com/tool/cc-debugger. [Accessed 28 February 2013]. [60] Texas Instruments, "SmartRF Flash Programmer," 20 January 2012. [Online]. Available: http://www.ti.com/tool/flash-programmer. [Accessed 28 February 2013]. [61] Specialty Coating Systems, "Parylene Deposition Process," 2012. [Online]. Available: http://scscoatings.com/what_is_parylene/parylene_deposition.aspx. [Accessed 28 February 2013]. [62] Digi-Key Corporation, "Part Number RNF-100-1-1/2-CL-SP datasheet," 7 March 1978. [Online]. Available: http://www.digikey.com/product-detail/en/RNF-100-11%2F2-CL-SP/RNF112C-25-ND/3854645. [Accessed 28 February 2013]. [63] W. A. Rutala and D. J. Weber, "Guideline for Disinfection and Sterilization in Healthcare Facilities," Centers for Disease Control and Prevention, Atlanta, 2008. [64] Rogers Corporation, "ULTRALAM® 3000 Liquid Crystalline Polymer Circuit Material Datasheet," Chandler, 2011. 146 [65] INTERNATIONAL AGENCY FOR RESEARCH ON CANCER, "IARC Monographs on the Evaluation of Carcinogenic Risks to Humans," WORLD HEALTH ORGANIZATION, Lyon, 1999. [66] Advanced Sterilization Products, "STERRAD 100NX Data sheet," Ethicon, Irvine, 2011. [67] Wipak, "STERIKING® LT-Blueline Self-sealable Pouches Product Specification," Bomlitz, Germany, 2008. [68] Texas Instruments, "BTool (within Bluetooth low energy software stack and tools)," 19 December 2012. [Online]. Available: http://www.ti.com/tool/blestack?DCMP=blestack&HQS=Ble-pr-tsw. [Accessed 26 February 2013]. [69] Texas Instruments, "BLE Device Monitor for Windows PC," 13 February 2013. [Online]. Available: http://www.ti.com/litv/zip/swrc258b. [Accessed 20 February 2013]. [70] J. Lahart, "Taking an Open-Source Approach to Hardware," The Wall Street Journal, 27 11 2009. [71] "Arduino - ArduinoWiFiShield," Arduino, 23 November 2012. [Online]. Available: http://arduino.cc/en/Main/ArduinoWiFiShield. [Accessed 17 January 2013]. [72] M. Kroll, "Bluetooth Low Energy (BLE) Shield for Arduino," 4 January 2013. [Online]. Available: http://www.mkroll.mobi/. [Accessed 17 January 2013]. [73] Adafruit Industries, "Adafruit I2C Controlled + Keypad Shield Kit for 16x2 LCD," [Online]. Available: http://learn.adafruit.com/downloads/pdf/rgb-lcd-shield.pdf. [Accessed 15 January 2013]. [74] US Naval Observatory, "US Naval Observatory Master Clock," [Online]. Available: http://tycho.usno.navy.mil/cgi-bin/timer.pl. [Accessed 3 February 2013]. [75] Daycounter Inc, "PCB Copper Thickness," 2004. [Online]. Available: http://www.daycounter.com/LabBook/PCB-Copper-Thickness.phtml. [Accessed 12 February 2013]. [76] Texas Instruments, "SWRS110C 2.4-GHz Bluetooth® low energy and Proprietary System-on-Chip," November 2012. [Online]. Available: SWRS084E. [Accessed 26 February 2013]. [77] Hollister Incorporated, Private Communication and custom samples, Chicago, 2013. [78] D. P. Kane, "Chronic wound healing and chronic wound management," in Chronic Wound Chare: A Clinical Source Book for Healthcare Professionals, 4th ed., 147 Malvern, PA, HMP Communications, 2007, pp. 11-23. [79] Q. Doan, Y. He, K. D. Peterson and W. J. Shi, "Techniques for sensing and adjusting a compliance voltage in an implantable stimulator device". European Union Patent EP1962950 A1, 3 September 2008. [80] E. Noorsal, K. Sooksood, H. Xu, R. Hornig, J. Becker and M. Ortmanns, "A Neural Stimulator Frontend With High-Voltage Compliance and Programmable Pulse Shape for Epiretinal Implants," IEEE Journal of Solid-State Circuits, vol. 47, no. 1, pp. 244-256, 2012. [81] . E. Barsoukov and J. R. Macdonald, Impedance spectroscopy: Theory, experiment, and applications, Hoboken: Wiley-Interscience, 2005. [82] Liquidware, "AMBI Ambient Light Sensor," 21 April 2010. [Online]. Available: http://www.liquidware.com/shop/show/SEN-LTE/AMBI+Light+Sensor. [Accessed 15 January 2013]. [83] A. I. Pressman et al., in Switching Power Supply Design, 3rd ed., New York, McGraw-Hill, 2009, pp. 31-40. [84] Energizer Holdings, "Product Datasheet- Energizer 675 Size- Mercury Free," [Online]. Available: http://data.energizer.com/PDFs/675.pdf. [Accessed 20 February 2013]. [85] D. Howe, K. Bogie and S. Garverick, "Battery-Powered, High-Voltage Stimulation Circuits for a Wound Healing Device," in submission to Transactions on Biomedical Engineering, 2013. [86] Energizer Holdings, Inc, "Energizer 522 Product Datasheet (9V)," 23 April 2013. [Online]. Available: http://data.energizer.com/PDFs/522.pdf. [Accessed 20 February 2013]. 148