Short Increment Sensitivity Index (SISI) - Quick Guide

advertisement
Short Increment Sensitivity Index (SISI)
Page 1 of 2
Short Increment Sensitivity Index (SISI) Quick Guide
Description
SISI is designed to test the ability to recognise 1dB increases in intensity during a series of bursts of pure
tones presented 20dB above the pure tone threshold for the test frequency. It can be used to differentiate
between cochlear and retro cochlear disorders, as a patient with a cochlear disorder will be able to perceive
the increments of 1dB, whereas a patient with a retro cochlear disorder will not.
Required Items
•
•
Headphones or insert phones
A response button
Test Procedure
1. Press and hold the Tests button and select SISI: Short increment sensitivity index using the scroll
wheel on the audiometer.
2. Select Tone or Warble in channel 1. If cross hearing is suspected, masking should be presented in
channel 2.
3. Select the desired test frequency and set the input level 20dB above threshold using the scroll wheel
on the audiometer.
4. In the most common type of SISI test, the incremental steps are set to 1dB. By pressing and holding
the function button AM, it is possible to change the increments. Prior to presenting the actual test of
1dB increments, a trial of 2 or 5dB steps can be employed to ensure that the patient understands the
test procedure and what will be required. If you are uncertain about the responses from the patient,
0dB can be used.
D-0113810 - A – 2016/06
Short Increment Sensitivity Index (SISI)
Page 2 of 2
5. Explain to the patient that they will now hear a series of tones. If suddenly there is a change in the
loudness of the presentation of the tone, the response button should be pushed.
6. Start the test by pressing
. The system will automatically count the number of reactions
from the patient. Note that the system needs 20 presentations to calculate a SISI score.
7. Repeat the test for all desired test frequencies.
8. Press
to end the test.
9. Save the SISI test by clicking Save Session.
SISI Results
The SISI test should be conducted at 20dB SL for all frequencies tested. If the patient does not manage to
get a high score on the SISI test, this could be indicative of retro cochlear damage.
D-0113810 - A – 2016/06
Download