Series IR7000 TELEDYNE ANALYTICAL INSTRUMENTS

TELEDYNE
ANALYTICAL INSTRUMENTS
Series IR7000
Non-dispersive Infrared Gas Analyzers
T
eledyne’s Series IR7000 features excellent linearity and stability.
Using advanced microprocessor technology, these instruments
make monitoring a process gas stream quick and accurate.
Optical Bench
The optical bench consists of the IR source, sample cell, and patented
detector. The source control circuit provides a current square wave
to the infrared source. No mechanical chopper / motor is used, thereby
avoiding moving parts. The closed sample path between source and
detector eliminates interfering absorption from ambient air. Infrared energy
from the source passes through the sample cell where energy at
specific wavelengths is absorbed by the sample.
The patented detector consists of two chambers in optical series with
a sensitive flow transducer to measure the relative infrared energy
absorbed. The advantage of this detector over others is that the signals
are balanced to improve sensitivity and selectivity, thereby minimizing
background process gas interference.
Front Panel Ease
All operations are selected through the four front panel buttons. The
operator can control calibration gas concentration, automatic calibration
frequency, active chart well as manual zero and span calibrations.
All operator settings are protected by battery backup.
Standard Features
• User selected auto-zero, auto-span calibration
• Linearized instrument range, eliminating need for separate instruments
to achieve full scale span
• Closed sample path is not exposed to ambient air eliminating the need
for periodic purging
• 4 selectable chart recorder ranges
• Modular design for easy maintenance
• Configured to accommodate an optional oxygen channel
• Low power requirements. No heat build up and therefore no
premature failure of components.
• No moving parts associated with the optical bench.
• No “tuning” required with optics unlike competitive NDIR methods
• The signal from the detector is digitized making signal processing
reliable and flexible
• Thorough self-diagnostics software package
• Easy set up and interface
• CE marked
Applications
• Chemical and petrochemical processes
• Combustion and flue gas processes
• Pulp and paper
• Vapor recovery systems
• Air separation
• Metals, ceramics and heat treating atmospheres
• Land fill gas power stations
• Carbon dioxide scrubber efficiency
• CO / CO2 / C2H4 monitoring in oxyhydrochlorination process in EDC
manufacturing
• Cement plant applications (CO, CO2, O2)
• CO / CO2 measurement in medical gas, medical air
Model Designations
IR7000 - Panel / 19” rack mount, CE mark approved
IR7010 - Split architecture (analysis unit - explosion proof)
IR7000P - Portable
IR7000T - Trace gas
IR7000D - Dual bench
IR7000B - Wall mount
IR7000DB - Wall mount dual bench
NOTE: Hazardous area configurations, X/Z purge for Div I & II, Zone
1, 2 CENELEC certified
Built for Reliability and Performance
TELEDYNE
ANALYTICAL INSTRUMENTS
Series 7600
Non-Dispersive
Infrared
Gas Analyzer
T
eledyne’s Series 7600 Infrared (IR) Gas Analyzer is
capable of detecting up to four chosen, individual IR
absorbing components (i.e. NO, SO2, CO2, CO, and
CH4) on a continuous basis.
Conveniently packaged in either a 19” rack mount or
NEMA-4 wall mount enclosure, the Series 7600 can also
be supplied with an oxygen sensor, providing the operator
with a space-saving, five in one, cost-effective design.
The NEMA-4 enclosure can be X or Z-purged to satisfy
hazardous area installation requirements.
A high-sensitivity mass flow type twin detector is utilized
for infrared measurements. By utilizing a single beam,
double path design in conjunction with a serial dual-layer
transmission detector, the Series 7600 delivers long
term, drift-free performance. The oxygen concentration
in the sample gas can be detected by either a built-in
paramagnetic sensor or an externally mounted, in-situ based
zirconium oxide sensor.
The concentration of the desired gases is displayed on a
large, easy-to-read back-lit LCD. The user interface is very
intuitive and the menu / mode selection buttons, which are
readily accessible, provide the operator with dynamic control
and extensive diagnostic capabilities.
• Slide-out, chassis design to facilitate any optical or
maintenance adjustments required to fine tune analyzer
performance (7600A)
• In-depth, valuable analyzer functions attainable from the
front-panel user interface buttons
- Follow & Hold output signal control (during calibration)
- Remote range change contro
- Low / Hi limit alarms
- Range ID signals
- Auto-calibration with user adjustable frequency and
gas flow time setting programming capabilities
- Remote auto-calibration initiation
- Auto-calibration status contacts
- Instrument or calibration error contact outputs
- Extra functions included such as average value
computation, O2 conversion
APPLICATIONS
OPTIONS
The Series 7600 is ideally suited for multi-parameter gas
analysis requirements for:
• Percent O2 detector – Paramagnetic (built-in) or ZrO2
(externally installed), user preference
• Combustion control within the power, pulp and paper, steel,
and cement industries
• Heat treating / Inert gas blanketing atmosphere control
• O2 correction (the conversion of measured CO and SO2
readings into values at standard O2 concentration).
Consult factory for more detail of functionality
• Bulk-gas impurity analysis within the air separation industry
• Communication functions:
• Anaerobic digester / Bio-gas / Land-fill gas analysis
- RS-232C (9 pins D-Sub connector)
• Vent gas analysis of oxyhydrochlorination reactors (EDC)
- Half-duplex bit serial
• Off-gas analysis on PTA and Maleic Anhydride reactors
- Modbus protocol
• Fluid Catalytic Cracker (FCC) regeneration gas analysis
• Ammonia / Fertilizer process gas stream analysis
• Continuous Emissions Monitoring Systems (CEMS)
FEATURES
• Simultaneous measurement of up to five components
• Excellent long-term stability
• Large, easy to read LCD display showing all simultaneous
measurements and computations
SPECIFICATIONS
Principle of measurement: Non-dispersive infrared (NDIR)
absorption method. Single source, dual
beam, dual-layer transmission design.
Power consumption: 250 VA max
Operating conditions: -5 to 45º C, 95% RH max
Enclosure:
Model 7600A: Steel casing (19” rack design for indoor use)
• Dimensions: 8.66” H x 19.0” W x 26” D (220 x 483 x 661 mm)
Measurable gas components and measuring range:
NO
SO2
CO2
CO
CH4
Min range
0 - 100 ppm
0 - 100 ppm
0 - 20 ppm
• Weight: Approx 48 lbs (configuration dependent)
Model 7600B: NEMA-4 enclosure (wall mount design)
• Dimensions & Weight: Application dependent
• X and Z-Purge configurations available for hazardous area
installations
lower ranges
available upon
request
0 - 50 ppm
0 - 500 ppm
0 - 100 vol%
0 - 100 vol%
• Sample cell: SUS304 / Neoprene Rubber
0 - 5 vol%
0 - 25 vol%
• Internal tubing: Application dependent (Std = Toaron tubing)
O2
(built in)
O2
(external
zirconia)
N2O
Max range
0 - 5000 ppm
0 - 10 vol%
0 - 100 vol%
Wetted parts:
• Inlet / Outlet fittings: SUS304 NPT 1/4 internal thread or Rc 1/4
• Cell windows: CaF2
Standard sample gas conditioning requirements:
0 - 5 vol%
0 - 25 vol%
0 - 200 ppm
0-100%
• Flow rate: 0.5L / min, ±0.2 L / min
• Gas temperature: 0 to 50º C
• Pressure: 10 kPa or less (1.5 psig); sample gas should vent to
stable atmospheric pressure
• Maximum five components measurement including oxygen
• Dust: 100ug / Nm3 or less in particle size of 1um or less
• 1 or 2 measuring ranges per component
• Moisture: Below the point at which saturation occurs at ambient temperature (non-condensing)
• Measuring range ratio: ≤ 1:5 (built-in O2); ≤ 1:20 (except
built in O2)
A maximum of five components and two ranges are
selectable: includes oxygen measurement. Other gases /
ranges can be measured with the Series 7600. Please contact
the factory for details.
Display: Digital indication (4 digits – back-lit
LCD)
Analog output signal: 4-20 mADC or 0-1 VDC; 550 ohms
max for 4-20 mADC and 100k ohms
for 0-1 VDC
Analog input signal: For externally mounted ZrO2 percent
O2 sensor (purchased separately)
Relay contacts: 250 VAC / 2 Amp; resistive load; all
relay contacts are isolated from internal
circuits
Contact input: Non-voltage contact (On / 0V; Off / 5V,
5mA flowing at ON). Contact inputs are
not isolated from one another.
Note: Only M3.5 screw terminals are used for signal inputs
and outputs
Power supply: 100 VAC – 240 VAC; 50/60 Hz (3-pin
inlet terminal used)
• No corrosive components such as HCL, CL2, HF, etc. (must be below 1 ppm max)
EC DIRECTIVE COMPLIANCE
The product conforms to the requirements of the Low Voltage
Directive 73/23/EEC and EMC directive 89/336/EEC (as
amended by Directive 92/31/EEC), both as amended by
Directive 93/68/EEC. It conforms to following standards for
product safety and electromagnetic compatibility:
EN61010-1: 2001 Safety requirements for electrical equipment
for measurement, control and laboratory use.
“Installation Category II”
“Pollution Degree 2”
EN61326-1: 1997, AI: 1998, A2: 2001
Electrical equipment for measurement, control and laboratory
use — EMC requirements.
SERIES 7600 - INFRARED GAS ANALYZER
PERFORMANCE
Repeatability: ±0.5% of full scale
Linearity:
±1% of full scale
Zero drift:
±1% of full scale / week
Span drift:
±1% of full scale / week
Response time (for 90% FS response): Within 60 seconds
including replacement
time of sample gas
(when gas flow rate is
0.5L / min)
Model 7600B serves as the explosion proof
configuration. The NEMA-4 enclosure can be X
or Z-purged to satisfy hazardous area installation
requirements.
Principle diagram of NDIR type
measurement
(for NO, CO2, CO, CH4, SO2)
TELEDYNE
ANALYTICAL INSTRUMENTS
A Teledyne Technologies Company
16830 Chestnut Street
City of Industry, California 91748, USA
TEL: 626-934-1500
FAX: 626-934-1651
TOLL FREE: 888-789-8168
Visit our website at:
www.teledyne-ai.com
© 2006 Teledyne Analytical Instruments, A Teledyne Technologies Company.
06/06LD
All rights reserved. Printed in the USA.
Warranty
Instrument is warranted for 1 year against defects in material
or workmanship
NOTE: Specifications and features will vary with application. The above are established and
validated during design, but are not to be construed as test criteria for every product. All
specifications and features are subject to change without notice.
TELEDYNE
ANALYTICAL INSTRUMENTS
Series 7500
Non-Dispersive
Infrared
Gas Analyzer
T
eledyne’s Series 7500 Infrared (IR) Gas
Analyzer is capable of detecting up to
three individual IR absorbing components
(i.e. CO, CO2, CH4, SO2) on a continuous
basis. Packaged in either a 19” rack mount or
NEMA-4 wall mount enclosure, the Series 7500
can also be supplied with an oxygen sensor,
providing the operator with a space-saving,
four in one, cost-effective design. The NEMA4 enclosure can be X or Z-purged to satisfy
hazardous area installation requirements.
o Remote range change control
o Range ID signals
o Auto-calibration with user adjustable frequency
and gas flow time setting programming capabilities
o Remote auto-calibration initiation
o Auto-calibration status contacts
o Instrument or calibration error contact outputs
OPTIONS
• Percent O2 detector – Paramagnetic
(built-in) or ZrO2 externally installed), user
preference
• O2 correction (the conversion of
measured CO and SO2 readings into
values at standard O2 concentration).
• Communication functions:
A high-sensitivity mass flow sensor is utilized
for infrared measurements. By using a single
beam, single path design in conjunction with
a serial dual-layer transmission detector, the
7500 delivers long-term, drift-free performance.
The oxygen concentration in the sample
gas can be detected by either a built-in
paramagnetic sensor or an externally mounted,
in-situ-based zirconium oxide sensor.
The concentration of the desired gases is
displayed on a large, back-lit LCD. The user
interface is intuitive and the menu / mode
selection buttons, which are readily accessible,
provide the operator with dynamic control and
extensive diagnostic capabilities.
FEATURES
• Simultaneous measurement of up to four
components
• Excellent long-term stability
• Slide-out, chassis design to facilitate any
optical or maintenance adjustments required
to fine tune analyzer performance (7500A)
• In-depth, valuable analyzer functions
attainable from the front-panel user interface
buttons
o Follow & Hold output signal control (during calibration)
o RS-232C (9 pins D-Sub connector)
o Half-duplex bit serial
o Modbus protocol
APPLICATIONS
The Series 7500 is ideally suited for multi-parameter
gas analysis requirements for:
• Combustion control within the power, pulp and
paper, steel, and cement industries
• Heat treating / Inert gas blanketing atmosphere
control
• Bulk-gas impurity analysis within the air separation
industry
• Anaerobic digester / Bio-gas / Land-fill gas analysis
• Vent gas analysis of oxyhydrochlorination reactors
(EDC)
• Off-gas analysis on PTA and Maleic Anhydride
reactors
• Fluid Catalytic Cracker (FCC) regeneration gas
analysis
• Ammonia / Fertilizer process gas stream analysis
• Continuous emissions monitoring
SPECIFICATIONS
Model 7500A - Steel casing (19” rack design for indoor use)
Principle of measurement: Non-dispersive infrared
(NDIR) absorption method. Single source, dual-layer
transmission design.
Measurable gas components and measuring range:
• Dimensions: 7” H x 19.0” W x 19.4” D
177 x 483 x 493 mm
• Weight:
Approx 25 lbs (configuration dependent)
Model 7500B - NEMA-4 enclosure (wall mount design)
CO2
Min range
0 - 100 ppm
Max range
0 - 100 vol%
CO
0 - 200 ppm
0 - 100 vol%
CH4
0 - 1000 ppm
0 - 100 vol%
SO2
0 - 1000 ppm
0 - 5000 ppm
• Inlet / Outlet fittings: SUS304 NPT ¼ internal thread or Rc 1/4
O2 (built in)
0 - 5 vol%
0 - 100 vol%
• Sample cell: SUS304 / Neoprene Rubber
O2 (external
• Cell windows: CaF2
0 - 5 vol%
0 - 25 vol%
• Internal tubing: Application dependent (Std = Toaron
tubing)
zirconia)
• Maximum four components measurement
including oxygen
• Dimensions & Weight: Application dependent
• X & Z-Purge:
Configurations available for hazardous
area installations
Wetted parts
Standard sample gas conditioning requirements
• Flow rate: 1L / min, ±0.5 L / min
• Measuring range ratio:
≤ 1:5 (except built-in
O2)
≤ 1:20
(built in O2)
• Gas temperature: 0 to 50º C
• Pressure: 10 kPa or less (1.5 psig); sample gas
should vent to atmospheric pressure.
A maximum of four components and two ranges
are selectable including an oxygen measurement.
Other gases can be measured with the Series 7500.
Please contact the factory for details.
• Dust: 100ug / Nm3 or less in particle size of
0.3um or less.
• Moisture: Below the point at which saturation
occurs at ambient temperature.
• 1 or 2 measuring ranges per component
Accuracy: ±1% of FS range
Display: Digital indication (4 digits – back-lit LCD)
Analog output signal:
4-20 mADC or 0-1 VDC; 550 ohms max
for 4-20 mADC and 100k ohms for 0-1
VDC
Analog input signal: For externally mounted ZrO2 percent O2
sensor (purchased separately)
Relay contacts: 250 VAC / 2 Amp; resistive load; all
relay contacts are isolated from internal
circuits
Contact input: Non-voltage contact (On / 0V; Off / 5V,
5mA flowing at ON). Contact inputs are
not isolated from one another.
Note: Only M3.5 screw terminals are used for signal
inputs and outputs
EC DIRECTIVE COMPLIANCE
The product conforms to the requirements of the
Low Voltage Directive 73/23/EEC and EMC directive
89/336/EEC (as amended by Directive 92/31/
EEC), both as amended by Directive 93/68/EEC. It
conforms to following standards for product safety
and electromagnetic compatibility:
EN61010-1: 2001 Safety requirements for electrical
equipment for measurement, control and laboratory
use.
“Installation Category II”
“Pollution Degree 2”
Power supply: 100 VAC – 240 VAC; 50/60 Hz
Power consumption: 70 VA max
EN61326-1: 1997, AI: 1998, A2: 2001
Operating conditions: -5 to 45º C
Electrical equipment for measurement, control and
laboratory use — EMC requirements.
Series 7500 Infrared Gas Analyzers
PERFORMANCE
Repeatability: ±0.5% of full scale
Linearity:
±1% of full scale
Zero drift:
±2% of full scale / week
Span drift:
±2% of full scale / week
Response time (for 90% FS
response):
One or two component measurement:
within 15 seconds including replacement
time of sample gas.
More than three components
measurement: within 30 seconds
including replacement time of sample
gas.
Model 7500B serves as the explosion proof configuration
of the series. The NEMA-4 enclosure can be X or Z-purged
to satisfy hazardous
area installation
requirements.
TELEDYNE
ANALYTICAL INSTRUMENTS
A Teledyne Technologies Company
16830 Chestnut Street
City of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168
FAX: 626-934-1651 EMAIL: ask_tai@teledyne.com
www.teledyne-ai.com
© 2006 Teledyne Analytical Instruments, A Teledyne Technologies Company.
06/06LD
All rights reserved. Printed in the USA.
Principle diagram of NDIR type
measurement (for CO2, CO, CH4, SO2)
Warranty
Instrument is warranted for 1 year against defects in material
or workmanship
NOTE: Specifications and features will vary with application. The above are established
and validated during design, but are not to be construed as test criteria for every product. All
specifications and features are subject to change without notice.
TELEDYNE
ANALYTICAL INSTRUMENTS
7320
Infrared Gas Analyzers
7300A
• Expanded measurement capability
• Patented detector design
Teledyne’s7300seriesofnon-dispersiveinfrared
(NDIR)gasanalyzersfeaturesfastresponse,high
accuracy,sensitivity,stability,andexcellentlinearity.The
costcompetitivemeasurementsolutionsaredueinpart
totheuniquelydesignedandpatentedIRdetector.
The7300Seriescanbeprovidedinavarietyof
configurations.
•7300A-Flushpanelorrackmount
•7300B-NEMA4wallmount (suitableforDiv2orEex(p)purge)
•7320-Fullyexplosionproofdesign (ClassI,Division1,GroupB)
Theheartofthesemicroprocessor-basedanalyzers
isatemperaturecompensated,hermeticallysealed,
steadystatethermopiledetectorintegratedintoanIR
photometricbench.
Thisdesigneliminatesthetraditionalmotordriven
chopperwheel,signalconditioningcircuitry,and
complexopticsresultinginacompactandrugged
analyzer.Theopticaldetectionbenchhasbeen
qualifiedforcriticallifesupportsystemsinspace
suits,shuttles,andspacestationFreedom.
TheSeries7300NDIRAnalyzersaresuppliedwith
thefollowingstandardfeatures:
•Threeprogrammablerangeswithauto-ranging
capabilities
•Auto-calibrationsoftware
•4-20mADCisolatedand0-1VDCoutputs
•RS-232Cbi-directionalserialinterface
•Dualconcentrationalarmsandsystemfailure
alarm
•Self-diagnosticcheckofsystemelectronics
Built for Reliability and Performance
Series 7300
Specifications
Applications
• Chemical and petrochemical processes
• Metals, ceramics and heat treating atmospheres
• Combustion and flue gas processes
• Landfill gas power stations
• Pulp and paper
• Emissions testing (part of the mobile stations)
• Vapor recovery systems
• Carbon dioxide scrubber efficiency
• Enhanced oil recovery
• CO / CO2 / C2H4 monitoring in oxyhydrochlorination process in EDC manufacturing
• Food, agriculture, medical
Specifications
Accuracy:
±2% of full scale at constant temperature
or better
Linearity:
±1% of full scale
Repeatability: ±1% of full scale
Drift:
±1% of full scale per week at constant temperature
Response time: 90% of full scale in less than 5 seconds
Span stability:
Less than ±1% of full scale change per month
Noise: ±1% of full scale
Analysis ranges:
Typical Gas Analysis Applications
CO2
0-1% to 0-100%
CO
0-5% to 0-100%
CH4
0-10% to 0-100%
C2 to C5 0-5% to 0-100%
(For lower ranges and other gases, contact factory.)
Ranges:
Three user programmable ranges with
auto-ranging
Ratio of 4:1 for low to high range (i.e. 0 - 5 up to 0 - 20% CO)
Range ID:
Via range ID contacts
TELEDYNE
ANALYTICAL INSTRUMENTS
A Teledyne Technologies Company
16830 Chestnut Street
City of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168
FAX: 626-934-1651 EMAIL: ask_tai@teledyne.com
www.teledyne-ai.com
© 2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.
All rights reserved. Printed in the USA.
07/05LD
Outputs:
4-20 mADC iso or 0-1 VDC negative ground
Serial output: RS-232C bi-directional user interface
Auto-calibration: User programmable auto-cal
Alarms: software standard
Dual alarms + system failure alarm
Self-diagnositics: Self-check of analyzer electronics
Operating temperature:
5 to 45°C (41 to 113°F)
Ambient temp: 5 to 45°C (41 to 113°F)
Power supply : 110 VAC or 230 VAC, 50 Hz or 60 Hz (Specify at the time of order)
Wetted parts: Application dependent
Area Classification / Dimensions
7300A:
Non-hazardous
7.5” H x 10.8” W x 13.7” D
7300B:
Non-hazardous (can also be purged to meet Div 2 or Eex(p) areas)
11.81” W x 20.28” H x 8.9” D
7320:
Class I, Division 1, Groups B, C, D
25.63” H x 5.25” W x 11” D
Warranty
Instrument is warranted for 1 year against defects in material
or workmanship
NOTE: Specifications and features will vary with application. The above are established
and validated during design, but are not to be construed as test criteria for every product.
All specifications and features are subject to change without notice.
TELEDYNE
ANALYTICAL INSTRUMENTS
MODEL GFC-7002E
Gas Filter Correlation
N2O Analyzer
• Bidirectional RS-232 for remote operation
• Adaptive signal filtering optimizes response time
PRINCIPLE OF OPERATION
Teledyne’s GFC-7002E measures low ranges of nitrous oxide
by comparing infrared energy absorbed by a sample to that
absorbed by a reference according to the Beer-Lambert law. This
is accomplished with a Gas Filter wheel which alternately allows
a high energy light source to pass through a N2O filled chamber
and a chamber with no N2O present.
The light path then travels through the sample cell, which has a
folded path of 2.5 meters. The energy loss through the sample
cell is compared with the zero reference signal provided by the
gas filter to produce a signal proportional to concentration, with
little effect from interfering gases within the sample.
ADVANCED PERFORMANCE
This design produces excellent zero and span stability and a
high signal to noise ratio, allowing extreme sensitivity. Multitasking software gives real time indication of numerous operating
parameters and provides automatic alarms if diagnostic limits are
exceeded.
Built-in data acquisition and internal memory allows logging
multiple parameters, including average and instantaneous
values, calibration data, and operating parameters. Up to
8,000 records of stored data may be retrieved via the RS-232
port or from the front panel. The Model GFC-7002E features
rugged construction and is designed to perform with a minimum
of attention. In the event maintenance is required, modular
construction makes service a simple operation.
APPLICATIONS
The GFC-7002E is ideally used as a safety monitor for
accumulation of solubalized N2O in oxygen and breakthrough
of N2O from particular mole sieves in the air supply associated
with cryogenic air separation equipment that features particular
reboiler designs. It can also be used to monitor the N2O
used for food processing propellant purposes as well as for
semiconductor manufacturing oxidizer use (i.e. high purity
applications).
Built for Reliability and Performance
Model GFC-7002E Specifications
Ranges:0-100 ppm N2O full scale, user selectable; dual
ranges and autoranging; 0-10 ppm and lower ranges
available
Units: ppm, mg/m3
•Internal pump
•Bi-directional RS-232
Span noise: <1% (RMS) of reading
Lower detectable limit (LDL): 0.5 ppm (RMS)
Zero drift: (24 hours) < 0.25 ppm* / (7 days) <0.5 ppm *
Span drift: (7 days) <1% reading*
Rise and fall time:<30 seconds to 95%
Linearity: 1% full scale
Precision: 0.5% of reading
Sample flow rate: 2000 cc / min ±20%
Operating temperature
range: 41 to 104°F (5 - 40°C)
Weight: •Selectable voltage
•10 isolated digital status outputs
Zero noise: <0.25 ppm (RMS)
Dimensions: CONFIGURATIONS
7” H x 17” W x 25” D (178 x 432 x 635 mm)
50 lbs (22.7 kg)
•2 optoisolated alarm outputs
•Supported voltages:
100V / 50Hz 100V / 60Hz
115V / 60Hz 220V / 50Hz
220V / 60Hz 230V / 50Hz (CE)
240V / 50Hz
•Output voltage:
10V
1V
•Particulate filter:
5V
100mV
47mm
OPTIONS
Power: Selectable voltage: 100V / 115V / 220-240V
Selectable frequency: 50 Hz or 60 Hz
• “U” version - 0-10 ppm and lower ranges
Approvals: CE
•Rack mount only
Analog outputs: Bi-polar, 10V, 5V, 1V, 100 mV, selectable
•Isolated 0-20mA or 4-20mA output
Recorder offset: ±10%
•Multi-drop RS-232 connection
RS-232: •Calibration valves
Bi-directional
•Rack mount (19”) with chassis slides
Status (digital): 10 outputs from optoisolator, included with
standard configuration
•Stainless steel valves for selection of customer-supplied zero and
span gas
Alarm output: ACCESSORIES
Hi, Hi-Hi or Lo-Hi, 2 optoisolated output
Current output: 0-20 mA or 4-20 mA isolated output (optional)
Interferences: H2O, CO2; negligible
* At constant temperature and voltage
TELEDYNE
ANALYTICAL INSTRUMENTS
A Teledyne Technologies Company
16830 Chestnut Street
City of Industry, California 91748, USA
TEL: 626-934-1500
FAX: 626-934-1651
TOLL FREE: 888-789-8168
Visit Our Web Site at:
www.teledyne-ai.com
© 2006 Teledyne Analytical Instruments, A Teledyne Technologies Company.
06/06LD
All rights reserved. Printed in the USA.
•RS-232 cable
•Expendables kit
•Spare parts kit
Warranty
Instrument is warranted for 1 year against defects in material
or workmanship
NOTE: Specifications and features will vary with application. The above are established
and validated during design, but are not to be construed as test criteria for every product.
All specifications and features are subject to change without notice.
TELEDYNE
ANALYTICAL INSTRUMENTS
MODEL GFC-7001E
Trace CO Analyzer
• Adaptive signal filtering optimizes response time
• GFC wheel guaranteed against leaks for 5 years
• Temperature and pressure compensation
• Internal data logging with 1 minute to 24 hour
averages
• Advanced remote operation software
PRINCIPLE OF OPERATION
Teledyne’s GFC-7001E Gas Filter Correlation CO analyzer
measures low ranges of carbon monoxide by comparing infrared
energy absorbed by a sample to that absorbed by a reference
gas according to the Beer-Lambert law. This is accomplished
with a gas filter wheel which alternately allows a high energy light
source to pass through a CO filled chamber and a chamber with
no CO present.
The light path then travels through the sample cell, which has a
folded path of 14 meters. The energy loss through the sample
cell is compared with the zero reference signal provided by the
gas filter to produce a signal proportional to concentration, with
little effect from interfering gases within the sample.
ADVANCED PERFORMANCE
This design produces excellent zero and span stability and a
high signal to noise ratio allowing extreme sensitivity. Multitasking software gives real time indication of numerous operating
parameters and provides automatic alarms if diagnostic limits are
exceeded.
Built-in data acquisition and internal memory allows logging
multiple parameters including average and instantaneous values,
calibration data and operating parameters. Up to 8000 records
of stored data can be retrieved via the RS-232 port or from the
front panel. The GFC-7001E features rugged construction and is
designed to perform with a minimum of attention. If maintenance
is required, modular construction makes service a simple
operation.
APPLICATIONS
The GFC-7001E, typically applied for combustion efficiency and
emissions monitoring purposes, is also ideal for the continuous
detection of ultra trace levels of CO in ultra-high purity bulk gases
(N2, Ar, O2 and He) used in the manufacturing of semiconductor
and optical fiber related components. The GFC-7001E is also
ideally applied for the detection of trace carbon monoxide (0-5
ppm) in medical grade oxygen and medical grade air.
Built for Reliability and Performance
Model GFC-7001E Specifications
Ranges:0-1000 ppm full scale, user selectable; dual ranges
and auto ranging supported; 0-1 ppm full scale
optional range available
Alarms: 3 x energizable alarm relays can be provided for CO
Concentration, Cal in Process, Range Change or for
System Failure indication purposes (option)
Units:
ppb, ppm, µg/m3, mg/m3
Zero noise:
<0.02 ppb (RMS)
Span noise:
<0.5% of reading above 5 ppm (RMS)2
Lower detectable limit (LDL):
Zero drift:
0.04 ppm
(24 hrs): <0.1 ppm*; (7 days): 0.2 ppm*
CONFIGURATIONS
•Auto ranging and dual ranges
•47mm particulate filter
•8 isolated digital status outputs
•6 isolated digital inputs
•Bi-directional RS-232
•Advanced remote operation software
•Supported voltages:
100V, 115V, 220 - 240V
• Supported frequencies: 50 Hz, 60 Hz
•Selectable output voltage: 10V, 5V, 1V, 100mV
Span drift: (24 hrs) < 0.5% reading* (7 days) 1% of reading*
Lag time:
10 seconds
Rise & fall time: <60 seconds to 95%
Linearity:
1% full scale
Precision:
0.5% of reading
Sample flow rate:800 cc / min ±10%
Operating temperature
range:
41 to 104°F (5 - 40°C)
OPTIONS
• “U” version - 0-1 ppm full scale range
•Rack mount (19”) with chassis slides
•Rack mount only
•Isolated 0-20mA or 4-20mA output
•Multi-drop RS-232 connection
•Calibration Valves:
Power:Selectable voltage: 100V / 115V / 220-240V
Selectable frequency: 50 Hz or 60 Hz
Internal Zero Span (IZS) Stainless steel auto-cal valves (mounted
within the GFC-7001E) for selection of customer-supplied zero
and span gas
Shut-off valve and flow control for external span gas cylinder
Internal zero air scrubber
Analog outputs: Bi-Polar, 10V, 5V, 1V, 100 mV, selectable
•RS-485 Communications
Recorder offset: ±10%
ACCESSORIES
RS-232:
Standard, DB-9 connector
RS-485:
Optional, DB-9 connector
•RS-232 cable
•Expendables kit
•Spare parts kit
Dimensions:
7” H x 17” W x 25” D (178 x 432 x 635 mm)
Weight:
40 lbs (18.2 kgs)
Status (digital): 8 outputs and 6 inputs (opto-isolated)
Current output:0-20mA or 4-20mA isolated output (optional)
Approvals:
USEPA, CE
* At constant temperature and voltage
TELEDYNE
ANALYTICAL INSTRUMENTS
A Teledyne Technologies Company
16830 Chestnut Street
City of Industry, California 91748, USA
TEL: 626-934-1500
FAX: 626-934-1651
TOLL FREE: 888-789-8168
Visit Our Web Site at:
www.teledyne-ai.com
© 2006 Teledyne Analytical Instruments, A Teledyne Technologies Company.
06/06LD
All rights reserved. Printed in the USA.
Warranty
Instrument is warranted for 1 year against defects in material
or workmanship
NOTE: Specifications and features will vary with application. The above are established
and validated during design, but are not to be construed as test criteria for every product. All specifications and features are subject to change without notice.
TELEDYNE
ANALYTICAL INSTRUMENTS
MODEL GFC-7000E
Ultra Trace CO2 Analyzer
produce an output proportional to concentration,
with little effect from interfering gases within the
sample.
ADVANCED PERFORMANCE
• Adaptive signal filtering optimizes response time
• Temperature and pressure compensation
• Internal data logging with 1 minute to 24 hour
averages
• Advanced remote operation software
PRINCIPLE OF OPERATION
The GFC-7000E measures carbon dioxide
(CO2) by comparing infrared energy absorbed
by a sample to that absorbed by a reference
gas according to the Beer-Lambert law. This is
accomplished by using a Gas Filter Wheel which
alternately allows a high energy infrared light
source to pass through a CO2 filled chamber and
a chamber with no CO2 present.
The light then travels through the sample cell,
which has a folded path. The energy loss
through the sample cell is compared with the
zero reference signal provided by the gas filter to
This design produces superior zero and span
stability and a high signal-to-noise ratio, allowing
excellent sensitivity. Multi-tasking software
gives real time indication of numerous operating
parameters and provides automatic alarms if
diagnostic limits are exceeded.
Built-in data acquisition and internal memory allow
logging multiple parameters including average
and instantaneous values, calibration data, and
operating parameters. Up to 8,000 records of
stored data can be retrieved via the RS-232 port or
from the front panel.
APPLICATIONS
The GFC-7000E is ideal for detecting the
breakthrough of ppb CO2 levels from molecular
sieves which are applied on the incoming air used
for cryogenic air separation purposes. In addition,
the GFC-7000E can also analyze ultra trace
CO2 levels which are dissolved in liquid oxygen
found within the main condensers of cryogenic air
separation plants.
Built for Reliability and Performance
Model GFC-7000E Specifications
Ranges:
Alarms: Units:
0 -10 ppm to 0-100 ppm full scale, user selectable;
dual ranges and auto ranging supported; 0-100 ppb
optional range available
3 x energizable alarm relays can be provided for CO2
Concentration, Cal in Process, Range Change or for
System Failure indication purposes (option)
ppm, mg/m3
<2.5 ppb (RMS)
Span noise: <0.5% of reading (RMS)2
Lower detectable
limit (LDL): <5 ppb
Zero drift: (24 hrs): 10 ppb*; (7 days): 20 ppb*
Span drift: (7 days): 1% of reading*
Lag time:
10 seconds
Rise & fall time: <90 seconds to 95%
Linearity:
Precision: • Auto ranging and dual ranges
• 47mm particulate filter
• 12 isolated digital status outputs
• 3 isolated digital inputs
• Bi-directional RS-232
Zero noise: CONFIGURATIONS
1% full scale
0.5% of reading
Sample flow rate:800 cc / min ±10%
Operating temperature
range:
41 to 104°F (5 - 40°C)
Dimensions:
7” H x 17” W x 25” D (178 x 432 x 635 mm)
Weight:
40 lbs (18.2 kgs)
Power:
electable voltage: 100V / 115V / 220-240V
S
Selectable frequency: 50 Hz or 60 Hz
• Advanced remote operation software
• Supported voltages / frequencies:
100V / 50Hz 100V / 60Hz
115V / 60Hz 220V / 50Hz
220V / 60Hz 230V / 50Hz (CE)
240V / 50Hz
• Selectable output voltage: 100mV, 1V, 5V, 10V
OPTIONS
• 0-100 ppb range
• Rack mount (19”) with chassis slides
• Rack mount only
• Isolated 0-20mA or 4-20mA output
• Multi-drop RS-232 connection
• Calibration Valves:
Recorder offset: ±10%
Internal Zero Span (IZS) Stainless steel auto-cal valves (mounted
within the GFC-7000E) for selection of customer-supplied zero
and span gas
Shut-off valve and flow control for external span gas cylinder
Internal zero air scrubber
RS-232:
• RS-485 Communications
Analog outputs: Bi-Polar, 10V, 5V, 1V, 100 mV, selectable
Standard, DB-9 connector
Status (digital): 12 outputs and 3 inputs (opto-isolated)
Current output: 0-20mA or 4-20mA isolated output, optional
Approvals: CE
* At constant temperature and voltage
TELEDYNE
ANALYTICAL INSTRUMENTS
A Teledyne Technologies Company
16830 Chestnut Street
City of Industry, California 91748, USA
TEL: 626-934-1500
FAX: 626-934-1651
TOLL FREE: 888-789-8168
Visit Our Web Site at:
www.teledyne-ai.com
© 2006 Teledyne Analytical Instruments, A Teledyne Technologies Company.
06/06LD
All rights reserved. Printed in the USA.
ACCESSORIES
• R
S-232 cable
• Expendables kit
• Spare parts kit
Warranty
Instrument is warranted for 1 year against defects in material
or workmanship
NOTE: Specifications and features will vary with application. The above are established
and validated during design, but are not to be construed as test criteria for every product. All specifications and features are subject to change without notice.
Process Photometers
PUV3402 and PIR3502
Continuous Measurements
VistaNET™ Connectivity
Multiple Component Samples
Measures Vapor or Liquid Samples
Operates in IR, NIR, UV and VIS Regions
Fiber Optic Option for NIR Applications
Multiple Interference Compensation Capability
AnalyzeIT
Field
IT
Control
IT
Engineer
IT
Field
IT
Inform
IT
Operate
IT
Power
IT
Industrial
IT
UV/VIS/NIR Fiber Optics
Option
Cell-In-Oven Sample
System
The PFO3502 Multiwave™ offers a fiber optic option
for applications that require remote sampling capability. With this option, light is transmitted via one
waveguide to the remote sample cell. A second
waveguide is used to return the sample-modified
light to the detector.
The Multiwave™ Cell-In-Oven design
provides a simple and effective package for high temperature applications.
The major benefit of this design is that
the sample cell and sample handling
components are together in one common temperature controlled enclosure.
The elimination of heat trace requirements for sampling lines minimizes
the risk of “cold spots”. This design
also offers excellent accessibility for
fast cell removal during maintenance.
This Fiber Optics option is effective in
UV/Visible/NIR applications where:
The sample stream is highly toxic.
Corrosive products are analyzed.
Fast response time is required.
The sample is at high pressure.
Performance Specifications
Precision:
± 1 % of full scale
Noise:
± 1 % of full scale at 0.02 absorption units
± 0.5 % of full scale at 0.20 absorption units
Linearity:
Standard ± 2 % of full scale
Zero Drift:
± 0.5 % of full scale per day for IR;
± 1.0 % of full scale per/day for UV
Response Time:
Programmable
Ambient Electronic Stability:
± 1 % of full scale for 18°F (10°C) in 4 hours
Electrical Power To Analyzer:
100/115/220/230 VAC, 45–66 HZ,
150W, Maximum Power Consumption
600W With Electric Heat
Instrument Air Supply for Enclosure Purge
(Safety):
Pressure 40–80 psi (3–6 bars)
Flow Rate 0.5 CFM (15L/min.)
Optical Purge, typically nitrogen:
Pressure @ 15–30 psi (1–2 bar)
Flow Rate 10–15 cc/min
General Installation
Protect the instrument from direct sunlight and
rain at operating temperature between 32°–113° F
(0–45°C)
Operating Specifications
Wavelength Range:
Ultraviolet 200–400 nm
Visible 400–800 nm
Near Infrared 800–2500 nm
Fundamental Infrared 2.50–14.50 um.
Ambient Temperature Range:
32° to 113°F (0–45°C)
Max. Cell Heat:
302°F (150°C)
Electric Cell Heat:
Power Consumption 450 watts maximum
(temperature is application dependant)
Sample Flow Rate:
Typical 20–500 cc/min for vapors;
5–120 cc/min for liquids
Sample Pressure:
0–500 PSIG (0–34 bar)
Voltage Input Variation:
10 % fluctuation without causing an
output variation of 0.05 % of full scale
Dimensions
Weight:
approximately 80 lbs (36.28 kg)
Height:
13.5 inches (342.9 mm)
Depth:
10.5 inches (266.7 mm)
Width:
10 inches (254 mm) per module
Overall Length:
(based on cell pathlength)
Minimum
26.8 inches (681.4 mm) for 0.5 to 16 mm cell
Maximum
65.5 inches (1665 mm) for 1 meter cell
Tube Fittings
Sample Inlet/Outlet:
Metallic Cells
Teflon Cell
Size:
1/4"
1/8"
Brand: Standard Gyrolok Gallek or Fluorocarbon
(Swagelok available)
Material: 316 SS
Teflon
Hastelloy “C”, Monel
Purge Inlet and Outlet:
Metal Cell Size:
1/4" NPT-F
Brand:
Gyrolok
Material:
316 SS
Power:
Size: 18 AWG, 3/4" Conduit Hub
Type: 3 conductors each
Output Signals
Analog Outputs:
4 ea. 4-20 mA isolated into 600 ohms max.
Contact Closures:
2 ea. Relay, 3W at 0.25 A or 28 VDC, 5 ea.
isolated solid state. Both relay and solid state
contact closures N.O. or N.C.
Digital Inputs:
8 ea., 2 ea. are dedicated
Digital Outputs:
4 ea., 110 VDC, 25 watts ea., max. standard,
110 VDC, 24 VDC optional.
Area Classification
NEC Class I, Groups B, C, D, Division 2 without
enclosure purge; Division 1 with type Y-Purge;
CSA Class I, Groups B, C, D, Division 2 with type
Z-Purge; Division 1 with type X-Purge (CSA 155493)
CE Zone 2, II3G; EEx pz IIB+H2 T4 to T2;
CE Zone 1, II2G; EEx pd [ib] IIB+H2 T4 to T2
(LCIE 03 ATEX 6007X)
Process Photometers
PUV3402 and PIR3502 ...
The Next Generation of On-Line
Process Photometers.
Multiwave™ Process Photometers
are designed to provide on-line measurements of gas or liquid components,
in simple or complex process streams,
for process control, product quality
assurance, safety, catalyst protection
and area monitoring.
The Multiwave™ is a fixed filter photometer that utilizes optical filters to
make continuous measurements. The
single beam, dual wavelength concept
used in the Multiwave™ compensates
for source and detector aging and the
obstruction of cell windows, while
allowing the sample cell to be isolated
from the electronics.
The Multiwave™ design takes the single
beam, dual wavelength concept one
important step further, by adding up
to eight filters to the filter wheel. This
improved method provides more measurement solutions than a conventional
single component process photometer.
The use of multiple wavelengths enables the Multiwave™ to compensate
for numerous interferences and perform multiple component applications.
The new Multiwave™ Process Photometers offer even more performance,
operating efficiency and versatility to
the user. The Multiwave™ line features
two basic models, each offering a wide
range of applications capability.
PIR3502 Multiwave™ Photometers
can be applied to Infrared and Near
Infrared measurements.
PUV3402 Multiwave™ Photometers
are designed for the Ultraviolet and
Visible spectral regions.
Optical Schematic
Multiwave™ Photometers can be connected to ABB Analytical’s VistaNET™
Process Analyzer Network, enabling
data interchange from the analyzer to
the DCS, seamless connectivity to the
Plant LAN and remote user access.
With the addition of VistaNET™, all
operator functions may be performed
at the analyzer or from a remote PC.
Multiwave™ Process Photometers Leading Applications*
PIR3502
Multiwave™ Spectral Regions
Infrared (IR) 2.50–14.5 um
Near Infrared (NIR) 800–2500 nm
Visible (VIS) 400–800 nm
Ultraviolet (UV) 200–400 nm
Infrared:
Isocyanate in chloroaromatic solvent
Ambient air monitoring
Multicomponent monomers in polymer process
Near Infrared:
Measurement of Hydrogenic Compounds (C-H, N-H, O-H)
Water in organics, such as Ethylene Dichloride (EDC)
Caustic in Acid Gas Scrubbers
PUV3402
VistaNET™
Connectivity
The VistaNET™ Process Analyzer
Network is a local area network that
supports data interchange from
process analyzers to the Distributed
Control System (DCS), in a dedicated
and secured manner. VistaNET™ also
provides seamless connectivity to the
plant operating system (Plant LAN).
Through the Multiwave™ Remote
User Interface, the user can configure,
operate, or troubleshoot the operation
of the analyzer from a remote PC.
With this remote access capability all
operator functions may be performed.
Visible:
Ultraviolet:
Color Measurements (ASTM, APHA, Saybolt Color Units)
%T @ Multiple Wavelengths in Glycol Process
Chlorine in Phosgene
*Consult our Multiwave™ Applications Brochure
for more details.
Remote configuration and observation of the Multiwave™ photometer,
enabling technicians to use their
time more efficiently.
Graphic display of Absorbance vs.
Concentration data, allowing the
user to check the validity and
linearity of calibration samples.
Remote maintenance via modem –
multiple users can view data
simultaneously and work together
in troubleshooting activities.
Reports and Tables can be printed
and reviewed at remote locations.
And VistaNet™ is designed for use
with widely used and accepted PC
hardware and operating systems,
including Windows™ ‘95, ‘98 and NT,
TCP/IP and other protocols. Connecting to other network operating systems is simple and very economical*.
*Consult our VistaNET™ Brochure for more details.
MultiwaveTM Process Photometer
Designed for Reliability
and Performance
PUV3402 and PIR3502 Multiwave™ Process Photometers are designed to meet all the significant
challenges of the process environment, including
ambient temperature variation, moisture, corrosive
and explosive stream compositions, electrical
hazards, dust and vibration.
Accessibility – Every major component may
be easily removed and replaced, significantly
reducing maintenance time.
Solid State Detectors – The PIR3502 and PUV3402
use solid state detectors. These detectors are
thermally stable and insensitive to vibration.
They require no mechanical adjustment and
minimal interconnecting circuitry. They provide
excellent linearity and long service life.
Brushless Chopper Motor – Provides good
mechanical reliability and long service life.
Linearizer Circuit with Up to 8 Points – Ensures
a ± 2 % of full scale linearity and allows the user
to select the best accuracy for a specific range.
Isolated Sample Cell – Prevents flammable and
corrosive streams from contacting electronics,
permits cell heating to optimum sample conditions, allows easy access and optimum optical
pathlength selection.
MultiwaveTM Process Photometer
Self-Aligning Optics – Saves time during
maintenance operations.
Temperature Controlled Filter Assembly –
Eliminates temperature effects on filters,
improving long term stability.
Electric Cell Heating Concept – The first ever
Division 1/Zone 1 electrically heated sample cell.
Employs a unique heat pipe concept that reduces
temperature gradients across the cell, providing
a more stable output and accurate temperature
control.
Data validation to ensure the reliability of
Multiwave™ data.
Self-diagnostics to aid in troubleshooting
operations.
Functional User Interface – Direct user interface
at the analyzer is simple and easy using the
Multiwave™'s front panel keyboard and display
screen. And with built-in VistaNET™ capability,
all operator functions may also be performed
from a remote PC.
© 2005 ABB Inc. BUASC 3-2-105
IT
The Industrial wordmark and all mentioned product names
IT
in the form XXXXXX are registered or pending trademarks of ABB.
ABB has Sales & Customer Support expertise in over
100 countries worldwide.
www.abb.com/analytical
ABB Inc.
843 N. Jefferson Street
Lewisburg
WV 24901
USA
Office: 304.647.4358
Fax: 304.645.4236
The Company’s policy is one of continuous
product improvement and the right
is reserved to modify the information
contained herein without notice.
Series IR7000 Non-dispersive Infrared Gas Analyzers
Specifications
Weight:
Rack mounted:
Portable:
Wall mount:
38 lbs (17.2kg)
19 lbs (8.6kg)
43 lbs (19.5kg)
Measuring method:
NDIR single beam
Gas measured:
User specified
Measuring range:
Per application
Standard Ranges - IR7000 / IR7000P
Response time:
3 seconds (application and flow dependent)
Gas*
Display:
Vacuum fluorescent
Alarms:
High and low limit, user selectable, 1 A, 60 VDC, 30
VAC
Analog output:
Selectable 1,5, or 10 volts; optional isolated 4-20 mA
current loop
CO
CO2
SO2
CH4
Propane
NO
Max. load impedance: 4-20 mA isolated output 500 ohms
Analog ranges:
Cal valve actuation:
Typical Low Range*
0-100 to 0-1000 ppm
0-30 to 0-300 ppm
0-50 to 0-500 ppm
0-300 to 0-3000 ppm
0-100 to 0-1000 ppm
0-100 to 0-1000 ppm
Typical Hi Range*
0-1% to 0-100%
0-1% to 0-100%
0-1% to 0-100%
0-1% to 0-100%
0-1% to 0-100%
0-1% to 0-100%
4, each with adjustable full scale; selectable auto
range
Resolution: 0.1% of full scale
Repeatability:
± 0.1% of full scale
Isolated triac control. Rated maximum load: 0.6 amp
at instrument voltage
Noise:
± 0.1% of full scale
Drift:
± 0.3% of full scale per week determined on
maximum range only and is absolute for all other
ranges
± 1% of full scale
Power source:
120 / 240 VAC, 50 / 60Hz (Portable model includes
rechargeable batteries - minimum 6 hours continuous
operation)
Linearity:
Max. power
110 VAC supply - 2 amps consumption: 230 VAC
supply - 1 amp
High Sensitivity Ranges - IR7000T
Power consumption:
50 watts / channel
Gas*
Materials in sample flow path: Glass, gold, buna-n, lexan, epoxy, sapphire,
304 stainless steel
Sample flow:
0.2 to 2.0 liters / minute (Trace gas - 5.0 - 10.0 liters/
minute) (Portable - internal sample pump)
Sample temp:
Range
Range
CO
0-5 to 0-50 ppm
0-20 to 0-200 ppm
CO2
0-2 to 0-20 ppm
0-20 to 0-200 ppm
For lower ranges of other gases, contact factory.
Resolution: 0.01 ppm
-10° to +50°C
Repeatability:
± 0.25% of full scale
Sample condition:
Non-condensing, particulate free
Noise:
± 0.1% of full scale
Warm-up time:
Usable in 60 mins, optimum operation in 3 hours
Drift:
Ambient conditions:
Operating: -10° to 50°C (non-condensing) Storage: -10° to 80°C (0° to 50°C for O2 sensor)
± 1% of full scale per week determined on max
range only; is absolute for all other ranges; auto zero
per day recommended
Linearity:
± 2% of full scale
Dimensions:
Rack mounted 571mm L x 447mm W x 133mm H (for 19” rack)
Portable: 20.0” L x 8.5” W x 5.25” H
508mm L x 216mm W x 133mm H
Wall mount:
22.5” L x 17.1” W x 5.25” H
24.0” L x 20.0” W x 6.0” D
609mm L x 508mm W x 152mm D
TELEDYNE
ANALYTICAL INSTRUMENTS
A Teledyne Technologies Company
16830 Chestnut Street
City of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168
FAX: 626-934-1651 EMAIL: ask_tai@teledyne.com
www.teledyne-ai.com
© 2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.
12/05LD
All rights reserved. Printed in the USA.
* Four outputs are available from minimum to maximum: Example CO offers 0-5, 0-10,
0-25, 0-50 ppm, all represented as 0-1 V output
* At maximum range and constant ambient conditions. Other gases
available by request
Options
• 0-25% Electrochemical oxygen channel
• RS-232 interface
Warranty
Instrument is warranted for 1 year against defects in material
or workmanship
NOTE: Specifications and features will vary with application. The above are established
and validated during design, but are not to be construed as test criteria for every product.
All specifications and features are subject to change without notice.