Series MPM-815H 150W PFC MEDICAL ATX POWER SUPPLY FEATURES l +5V Stand-by and Remote ON/OFF l 1U, 150W convection cooled with ATX l Input Active PFC for medical purpose Power good / Power fail signal l Thermal protection l EN 60601-1, 3rd edition l RoHS compliant l DESCRIPTION The ATX power supply MPM-815H is a 150W medical device Short Circuit Protection: The PSU will go into hiccup mode against with active PFC for computerised medical applications. This short circuit or over load conditions and will auto-recovery while faulty power supply is EN 60601-1, 3rd edition approved and has a conditions are removed very high quality standard and thermal protection. The device Power ON/OFF: The PSU will be turned on when the power ON/OFF utilizes a thermally efficient U channel chassis design and pin is connected to secondary GND convection cooling. Over Voltage Protection: If for some reason the PSU fails to control itself, the build-in over voltage protection circuit will shut down the outputs to prevent damaging external circuits. The trigger point is about SPECIFICATION 5.8-6.4V at +5V. If the OVP occur, PSU cannot be recovered. Input Voltage: 90 ~ 264VAC, Continuous input range Over Temperature Protection: When the PSU is operating over the Input Frequency: AC Input 47-63Hz temperature or over load limit, the device will be shut down Hold-up Time: 16ms, Nominal AC input voltage automatically to protect itself. The protection point is at the temperature (115VAC, rated load) of the HS1 over 110°C. After temperature of HS1 going down, the Inrush Current: Max. 30/60A; Nominal AC input voltage power supply will restart automatically. (115VAC/230VAC) one cycle at 25°C Relative Humidity: Min. 5% / Max. 95% RH, non-condensing Input Current: Max. 4/2A; Nominal AC input voltage Altitude: Operating: Max. 10K Feet, non-operating: Max. 40K (115VAC/230VAC), rated load Operating Temperature Efficiency: Nom. 75%, rated load, 115VAC (without the optional top cover ie open frame): Varies with distribution of loads among output Derate linearly above 50°C by 2.5% per deg C to a max. temperature Input Protect: Non-user serviceable internally located of 70°C. At 100% Load: Max. +50°C - At 50% Load: Max. +70°C AC input line fuse Operating Temperature (with the top cover): 30°C at 100% rated Total Regulation: On condition of a setting output voltage, load, 40°C at 90% rated load and 50°C at 70% rated load as derating input voltage from 90-264VAC and output from 0W to 400W curve. (Please see Note 5 below.) Remote Voltage Sense: Compensates for wire drop Warranty: 24 months Turn-on Delay: Min. 0.3 sec. / Max. 6 sec.; Time required for initial output voltage stabilization Storage Temperature: Min. -20°C / Max. +70°C www.hitekpower.com Series MPM-815H 150W PFC MEDICAL ATX POWER SUPPLY SPECIFICATION Approvals: IEC 60601-1, 3rd edition CB approved EN 60601-1, 3rd edition TUV approved UL 60601-1, 1st edition, 2006-04-26 UL approved CAN/CSA-C22.2 No. 601.1-M90, 2005 cUL approved EMI: EN 60601-1-2:2001 Class B PFC: EN 61000-3-2: 2000 & EN 610003-3:2001 Class D EMS: IEC 61000-4-2:2004 8kV air discharge, 6kV contact discharge Level 3 IEC 61000-4-3:2002 3m/V Level 3 IEC 61000-4-4:2004 2kV line & PE Level 3 IEC 61000-4-5:2001 1kV line to line, 2kV line to PE Level 3 IEC 61000-4-6:2004 3V/m Level 3 IEC 61000-4-8:2001 3A/m Level 3 IEC 61000-4-11:2004 Output Voltage / Current Rating Chart: Measured at output power connector Model Output Voltage Min. Output Current Rated Output Current Max. Output Current Line Regulation Load Regulation Ripple & Noise Initial Setting Accuracy (Note 2) (Note 3) (Note 1) MPM-815H +5V 1A 11A 14A +/- 1% +/- 2% 50mV 5.05V to 5.15V MPM-815H +12V 0A 5A 10A +/- 1% +/-4% 100mV 11.6V to 12.6V MPM-815H -12V 0A 0.5A 1A +/- 1% +/-5% 150mV -11.4V to 12.6V MPM-815H +3.3V 0A 7.5A 12A +/- 1% +/- 4% 50mV 3.2V to 3.4V MPM-815H +5Vsb 0A 0.75A 1.5A +/- 1% +/- 4% 100mV 4.8 to 5.2V NOTE: Total Output Power: 150W at 50°C environment temperature (Note 4) 1) The maximum total combined output power on the +3.3V and +5V rails is 90W 2) Measured by a 20MHz bandwidth limited oscilloscope, each output is connected with a 10ìF electrolytic capacitor and a 0.1ìF ceramic capacitor. 3) Initial setting accuracy is at input 110VAC and all output at 60% rated load. 4) The total DC continuous power shall be kept with 150W at input voltage at 110-264VAC. With input voltage 90-109VAC the total DC continuous power shall be kept with 120W max. The maximum total combined output power on the +3.3V and +5V rails is 90W. On condition of with the option cover, the maximum 150W is at 30°C environment temperature (Please see operating temperature). 5) To order unit with top cover, please add suffix -C to the required model number, eg MPM-815H-C www.hitekpower.com Series MPM-815H 150W PFC MEDICAL ATX POWER SUPPLY Operating Temperature Curve (without the option cover (open frame) Operating Temperature Curve (with the top cover version) Connector: CN1 - AC Input CN3 - DC Output CN4 - Fan Output CN5 - PG/PF CN6 - PS ON/OFF CN7 - +5Vsb output 3 positions terminal blocks 8 positions terminal blocks Molex 5045-02A or equivalent Molex 5045-02A or equivalent Molex 5045-02A or equivalent Molex 5045-02A or equivalent Pin Assignment: CN1 CN3 CN4 CN5 CN6 Pin Pin Pin Pin Pin 1. L 1. -12V 1. +12V 1. +5V 1. +5Vsb 2. N 2. GND 2. GND 2. GND 2. GND 3. GND 3. +3.3V 4. GND 5. +5V 6. +5V www.hitekpower.com 7. +12V 8. GND Series MPM-815H 150W PFC MEDICAL ATX POWER SUPPLY Mechanical Drawings: 198 (L) x 97 (W) x 40.5 (H)mm, Tolerance +/- 0.4mm (U-Channel) CN3 8 7 6 5 4 3 2 1 CN5 1 1 1 CN6 CN7 CN4 CN1 1 3 2 1 Use M3 screw Protrusion limlt 2.5mm x 6 Measuring the screw protrusion f SCRE W Customer system chassis www.hitekpower.com irst: Series MPM-815H 150W PFC MEDICAL ATX POWER SUPPLY Cable: 866-815H www.hitekpower.com Series MPM-815H 150W PFC MEDICAL ATX POWER SUPPLY Cable: 866-815H-24 www.hitekpower.com GERMANY Joh.-Friedr.-Boettger-Str. 21 63322 Roedermark Phone +49 6074 69285-0 Fax: +49 6074 69285-10 E-Mail: sales.de@hitekpower.com Revision 10/2013 www.hitekpower.com