PROACT Smoothline

advertisement
PROACT
Smoothline Laryngoscope Handles & Blades
Smoothline™
Fibre Optic
- Compatible with ISO 7376-3 Green System Fibre Optic
Instructions For Use
WARNINGS
PRE-USE CHECKS
Tracheal intubation requires specialist training. Only
competent trained personnel should use this device.
Do not use this device unless the following checks
have been performed.
•Avoid exposure to ultraviolet or direct sunlight.
•Ensure that used batteries are disposed of in accordance with appropriate national and local laws, regulations and procedures.
regulations and procedures.
•This device is not suitable for use in a Magnetic Resonance Imaging (MRI) environment.
•Remove batteries from handles when not in use.
•Check device for completeness, sharp edges, damage, flaws or loose components.
•Engage and disengage laryngoscope twice prior to use to confirm operation, ensure consistent light when force is exerted on blade tip in all directions.
•Insufficient, flickering or absence of light may indicate that batteries require replacing or a problem with the bulb or handle contact.
•Inspect optic light source for clarity. Light output may degrade over time from repeated autoclaving. Quality is guaranteed for 300 cycles.
•In event of pre-use check failure return product to supplier for inspection.
DIRECTIONS FOR USE
CLEANING INSTRUCTIONS
These directions are general guidelines intended for use
by qualified medical personnel. Any instructions and
contraindications given are not exhaustive and it is the
responsibility of the clinician to ensure the safe, correct
use of this product.
•Select the appropriately sized device for patient.
•Unscrew cap at base of handle to insert fresh batteries observing correct polarity (see diagram).
•Attach laryngoscope blade to compatible handle.
•Ensure blade is fitted firmly onto handle and clicked into place.
•To switch on, pull blade up and lock into position. Light will activate automatically.
•Examine / intubate trachea following current accepted medical guidelines.
•To switch off, move blade down. Light will automatically turn off.
Battery Polarity
Infection from patient to patient can potentially be
transmitted by any piece of medical equipment.
Sufficient measures must be taken to clean, disinfect and
sterilise (where appropriate) this device between uses.
Laryngoscope Blades
• Clean externally with a damp cloth using soap, alcohol or disinfectant solution to remove any deposits.
• Observe manufacturer’s usage instructions for all cleaning substances.
• pH neutral cleaning solutions recommended. Avoid caustic ingredients.
• Device may be autoclaved up to a maximum of 138°C for periods in compliance with national recommendations. Guaranteed for 3,000 5-minute cycles.
• Suitable for sterilisation by Ethylene Oxide (ETO), Steris or Sterrad procedures.
• Do not use abrasive cleaning techniques.
• Do not flash autoclave.
• Do not ultrasonically clean.
Laryngoscope Handles
• Remove batteries.
• Clean externally with a damp cloth using soap, alcohol or disinfectant solution to remove any deposits.
• Observe manufacturer’s usage instructions for all cleaning substances.
• pH neutral cleaning solutions recommended. Avoid caustic ingredients.
• Ensure that fluid does not seep into the electrical contact area.
• Suitable for sterilisation by Ethylene Oxide (ETO), Steris or Sterrad procedures.
• Ensure that handle is dry both inside and out prior to storage.
• Do not immerse handles in disinfectant or washing solutions.
• Do not use abrasive cleaning techniques.
• Do not autoclave.
• Do not ultrasonically clean.
Manufactured for:
PROACT Medical Ltd
9-13 Oakley Hay Lodge, Great Folds Road, Oakley Hay Business Park,
Corby, Northants, NN18 9AS, UK
Tel: +44 (0)1536 461981 E-mail: sales@proactmedical.co.uk
PROACT Medical SAS
Parc d’Activitiés des Béthunes, 8 Avenue du Fief Lot No 9,
BP 20424 Saint Ouen l’Aumône, 95005 Cergy Pointoise Cedex, France
Tel: +33 (0) 1.34.42.95.32 E-mail: ventes@proactmedical.fr
IFU 203 - v1
1/2
TECHNICAL INFORMATION AND RE-ORDERING CODES
wC
All Smoothline Laryngoscopes comply with ISO 7376:2009 where applicable.
Smoothline Laryngoscope Blades, Green System Fitting, Polished Finish
Smoothline Laryngoscope Handles, Green System Fitting
Part Number
H9020P
H9021P
H9022P
H9023P
H9024P
H9025P
H9029P
H9030P
H9031P
H9032P
H9033P
Part Number
H4551
H4552
H4553
Style
Macintosh
Macintosh
Macintosh
Macintosh
Macintosh
Macintosh
Miller
Miller
Miller
Miller
Miller
Size
0
1
2
3
4
5
00
0
1
2
3
Blade Length
58.0mm
72.0mm
88.0mm
112.0mm
131.0mm
148.0mm
45.0mm
55.0mm
80.0mm
130.0mm
170.0mm
Blade Width
11.0mm
10.0mm
13.5mm
14.5mm
14.5mm
15.5mm
10.0mm
10.0mm
10.0mm
13.0mm
13.0mm
Size
Adult
Paediatric
Stubby
Batteries Required
2xC
2 x AA
2 x AA
Smoothline Laryngoscope Blades, Green System Fitting, Matt Finish
Part Number
H9020
H9021
H9022
H9023
H9024
H9025
H9029
H9030
H9031
H9032
H9033
HGSKESS3
HGSKESS4
HGSLEFTHMAC3
Style
Macintosh
Macintosh
Macintosh
Macintosh
Macintosh
Macintosh
Miller
Miller
Miller
Miller
Miller
Kessel
Kessel
Left-Handed
Macintosh
HGSLEFTHMAC4 Left-Handed
Macintosh
HGSMAGILL0
Magill
HGSMAGILL1
Magill
HGSMAGILL2
Magill
HGSMAGILL3
Magill
HGSPOLIOMAC3 Polio
HGSPOLIOMAC4 Polio
HGSRSHAW0
Robertshaw
HGSRSHAW1
Robertshaw
HGSSEWARD1
Seward
HGSSEWARD2
Seward
HGSWISC0
Wisconsin
HGSWISC1
Wisconsin
HGSWISC2
Wisconsin
Size
0
1
2
3
4
5
00
0
1
2
3
3
4
3
Blade Length
58.0mm
72.0mm
88.0mm
112.0mm
131.0mm
148.0mm
45.0mm
55.0mm
80.0mm
130.0mm
170.0mm
112.0mm
131.0mm
111.5mm
Blade Width
11.0mm
10.0mm
13.5mm
14.5mm
14.5mm
15.5mm
10.0mm
10.0mm
10.0mm
13.0mm
13.0mm
14.5mm
14.5mm
14.5mm
4
130.0mm
15.0mm
0
1
2
3
3
4
0
1
1
2
0
1
2
74.0mm
83.0mm
101.0mm
153.0mm
115.0mm
135.0mm
70.0mm
90.0mm
83.0mm
108.0mm
54.0mm
80.0mm
88.0mm
10.5mm
12.0mm
12.0mm
13.5mm
15.0mm
15.0mm
9.5mm
10.5mm
9.5mm
13.5mm
10.7mm
10.7mm
12.5mm
Manufactured for:
PROACT Medical Ltd
9-13 Oakley Hay Lodge, Great Folds Road, Oakley Hay Business Park,
Corby, Northants, NN18 9AS, UK
Tel: +44 (0)1536 461981 E-mail: sales@proactmedical.co.uk
PROACT Medical SAS
Parc d’Activitiés des Béthunes, 8 Avenue du Fief Lot No 9,
BP 20424 Saint Ouen l’Aumône, 95005 Cergy Pointoise Cedex, France
Tel: +33 (0) 1.34.42.95.32 E-mail: ventes@proactmedical.fr
IFU 203 - v1
2/2
Download