Original Article Evaluation of the Effect of Gain on the Meal Response of an Automated Closed-Loop Insulin Delivery System Antonios E. Panteleon, Mikhail Loutseiko, Garry M. Steil, and Kerstin Rebrin A continuous closed-loop insulin delivery system using subcutaneous insulin delivery was evaluated in eight diabetic canines. Continuous glucose profiles were obtained by extrapolation of blood glucose measurements. Insulin delivery rate was calculated, using a model of -cell insulin secretion, and delivered with a Medtronic MiniMed subcutaneous infusion pump. The model acts like a classic proportional-integral-derivative controller, delivering insulin in proportion to glucose above target, history of past glucose values, and glucose rate of change. For each dog, a proportional gain was set relative to the open-loop total daily dose (TDD) of insulin. Additional gains based on 0.5 ⴛ TDD and 1.5 ⴛ TDD were also evaluated (gain dose response). Control was initiated 4 h before the meal with a target of 6.7 mmol/l. At the time of the meal, glucose was similar for all three gains (6.0 ⴞ 0.3, 5.2 ⴞ 0.3, and 4.9 ⴞ 0.5 mmol/l for 0.5 ⴛ TDD, TDD, and 1.5 ⴛ TDD, respectively; P > 0.05) with near-target values restored at the end of experiments (8.2 ⴞ 0.9, 6.0 ⴞ 0.6, and 6.0 ⴞ 0.5, respectively). The peak postprandial glucose level decreased significantly with increasing gain (12.1 ⴞ 0.6, 9.6 ⴞ 1.0, and 8.5 ⴞ 0.6 mmol/l, respectively; P < 0.05). The data demonstrate that closed-loop insulin delivery using the subcutaneous site can provide stable glycemic control within a range of gain. Diabetes 55:1995–2000, 2006 I ntegration of glucose-sensing and insulin delivery technologies can potentially lead to a fully ambulatory closed-loop insulin delivery system or “artificial pancreas.” Such a system may achieve glucose and insulin profiles near those observed in individuals with normal glucose tolerance while at the same time decreasing the incidence of hypoglycemia and minimizing the complications of diabetes. Glucose-sensing technology has been rapidly advancing (1– 6), and fast-absorbing insulin analogs (7,8) have been available for several years; nonetheless, little data exist demonstrating the feasibility of closed-loop insulin delivery using the subcutaneous route (overview in 9). Early studies in the diabetic canine, using intravenous insulin (10,11), showed promising results, delivering insulin in proportion to glucose above or below target and its rate of From Medtronic MiniMed, Northridge, California. Address correspondence and reprint requests to Garry Steil, PhD, Medtronic MiniMed, 18000 Devonshire Rd., Northridge, CA 91325. E-mail: garry.steil@medtronic.com. Received for publication 15 October 2005 and accepted in revised form 30 March 2006. AUC, area under the curve; TDD, total daily dose. DOI: 10.2337/db05-1346 © 2006 by the American Diabetes Association. The costs of publication of this article were defrayed in part by the payment of page charges. This article must therefore be hereby marked “advertisement” in accordance with 18 U.S.C. Section 1734 solely to indicate this fact. DIABETES, VOL. 55, JULY 2006 change, and later work extended these results with rapidacting insulin analogs and alternate sites of delivery (12– 14). The current study was undertaken with three objectives. The first was to evaluate the ability of a previously proposed multiphasic model of -cell secretion (15) to provide adequate control when the relative amount of insulin in each phase is adjusted to compensate for the delay in subcutaneous insulin delivery. The second was to assess the stability of the system as the closed-loop gain is increased or decreased. The third objective was to assess how well the in vivo closed-loop responses could be described using standard metabolic models. RESEARCH DESIGN AND METHODS Experiments were performed on eight diabetic dogs (26.8 ⫾ 1.7 kg, range 20.3–35.3) housed at Medtronic MiniMed animal facilities. Diabetes was induced by partial pancreatectomy. Dogs were treated with continuous subcutaneous insulin infusion for at least 1 week before experiments to determine the total daily dose (TDD) of insulin required for open-loop glucose control. On the day before experiments began, a new insulin infusion catheter was inserted. Closed-loop experiments were performed using Humalog. Dogs were housed under controlled kennel conditions and used for experiments only if judged to be in good condition (physical appearance, hematocrit, white blood cell count, body temperature, and weight stability). The experimental protocol was approved by the appropriate animal research committee. Insulin delivery algorithm. Insulin delivery was calculated by an algorithm that responds to glucose with three components: a proportional component (P) that reacts to the difference between plasma glucose and basal glucose (the putative -cell set point for fasting glucose), an integral component (I) that adapts to persistent hyper- or hypoglycemia, and a derivative component (D) that acts in response to the rate of change of blood glucose. We have previously shown the model to describe insulin secretion during hyperglycemic clamps performed in humans with normal glucose tolerance (16). The model is identical to the well-known proportional-integral-derivative algorithm widely used in engineering (17). It is described by: PID(t) ⫽ P(t) ⫹ I(t) ⫹ D(t); PID(t) ⱖ 0 P(t) ⫽ KP(G ⫺ GB) 冕 t KP I(t) ⫽ (G ⫺ GB) dt; I(0) ⫽ IDB TI (1) 0 D(t) ⫽ KPTD dG dt where the proportional gain KP (units/h per mg/dl) determines the rate of insulin delivery in response to glucose above the set point (GB; mmol/l), TI (integral time; min) determines the rate at which the underlying basal rate adapts, and TD (derivative time; min) determines the relative amount of insulin delivered in response to the rate of change of glucose. GB was fixed at 6.7 mmol/l, and KP was set in relation to the TDD of insulin under open-loop treatment: KP ⫽ TDD ⫻ reference KP reference TDD (2) Reference KP was 0.675 units/h per mg/dl, and reference TDD was 1.0 units 䡠 kg⫺1 䡠 day⫺1. TD and TI were fixed at 66 and 150 min, respectively. These 1995 AUTOMATED CLOSED-LOOP INSULIN DELIVERY FIG. 1. Blood glucose (A), plasma insulin (B), and algorithm output (C) during 10 h of closed-loop insulin delivery (n ⴝ 8) for 0.5 ⴛ TDD (f), TDD (䡺), and 1.5 ⴛ TDD (E). values were based on a computer simulation (computer model in 18) and preliminary experiments performed in dogs during the development of the system. Using these nominal gains, we performed a dose response with KP increased by 50% (1.5 ⫻ TDD) or decreased by 50% (0.5 ⫻ TDD). At the time closed-loop control was initiated (approximately 8:00 A.M.), all preprogrammed basal rates were set to 0, and the initial condition of the integral term (IDB) (Eq. 1) was set to the 8:00 A.M. basal rate. Setting the initial condition to the 8:00 A.M. basal rate allowed the animals starting closed-loop control at target glucose [P(t) ⫽ 0] and stable rate of change of glucose [D(t) ⫽ 0] to transfer onto closed-loop control with no change in insulin delivery. For animals starting closed-loop control above target, the initial condition (IDB) was also set to the 8:00 A.M. value basal rate, and the animals received an insulin bolus calculated by the algorithm: (KP ⫻ TD) ⫻ (glucose ⫺ GB). If PID(t) was calculated to be ⬍0, insulin delivery was suspended, and no changes were made in the integral component. Continuous glucose concentration was calculated by a linear extrapolation of the measured blood glucose samples (referred to in engineering literature as a first-order hold) (19). Insulin delivery was calculated with a discrete form of Eq. 1 and delivered with a Medtronic MiniMed Paradigm 511 pump. Experimental details. After an overnight period, during which euglycemia was achieved with continuous subcutaneous insulin infusion, a blood-sampling catheter was inserted into a peripheral vein. Blood samples were obtained every 20 min until noon, at which time the dog was fed a standard meal of commercially available dog food (LabDiet). After the meal, blood samples were drawn every 10 min for 3 h and every 20 min thereafter for an additional 3 h. Experiments on the same dog were separated by at least 5 days, and the order was randomized. Insulin kinetic analysis. The plasma insulin concentration [Ip(t)] in response to a single subcutaneous bolus of insulin was assumed to follow a biexponential curve [h(t)] with the response to multiple boluses linearly summing: 1996 冕 t Ip (t) ⫽ PID(t) ⫻ h(t ⫺ )d 0 [ t t h(t) ⫽ B ⫻ e ⫺1 ⫺ e ⫺2 ] (3) Eq. 3 is the solution to a standard two-compartment (subcutaneous and plasma) insulin model (20). Parameters were identified by nonlinear least squares (Civilized Software, Silver Spring, MD). From the parameters B, 1, and 2, the time to peak insulin concentration after a subcutaneous bolus and insulin clearance rate were calculated as ln(2/1)/(1/1 ⫺ 1/2) and 1/B/(1 ⫺ 2), respectively. Statistical analysis. Data are reported as the means ⫾ SE. Differences in parameters were evaluated with either paired Student’s t tests or one-way ANOVA (GraphPad Prism; GraphPad Software, San Diego, CA). Assays. Blood glucose values were immediately ascertained, using a Chiron 865 automated analyzer (Chiron, Emeryville, CA). Additional blood was drawn into heparinized tubes (⬃1 ml) and centrifuged, and the plasma was stored at ⫺20°C. Insulin was measured in duplicate, using an enzyme-linked immunosorbent assay kit (ALPCO Diagnostics, Windham, NH). RESULTS Closed-loop glucose control. Preprandial glucose levels were similar for the different algorithm gains (6.0 ⫾ 0.3, 5.2 ⫾ 0.3, and 4.9 ⫾ 0.5 mmol/l for 0.5 ⫻ TDD, TDD, and 1.5 ⫻ TDD, respectively; P ⬎ 0.05) (Fig. 1A). Peak postprandial glucose significantly decreased as gain increased (12.1 ⫾ 0.6, 9.6 ⫾ 1.0, and 8.5 ⫾ 0.6 mmol/l; P ⬍ 0.05). Maximum plasma insulin levels were similar (224.5 ⫾ 50.8, DIABETES, VOL. 55, JULY 2006 A.E. PANTELEON AND ASSOCIATES FIG. 3. Blood glucose concentration (A), plasma insulin delivery (B), and insulin concentration (C) before consuming a standard meal and 6 h after the meal, using closed-loop insulin delivery gains calculated with 0.5 ⴛ TDD, TDD, and 1.5 ⴛ TDD (n ⴝ 8). FIG. 2. Postprandial incremental AUC for glucose (A), postprandial incremental AUC for insulin (B), and incremental area under the insulin curve during the first 2 h after the meal (C). Results based on closed-loop insulin delivery gain KP calculated as 0.5 ⴛ TDD, TDD, and 1.5 ⴛ TDD, respectively (n ⴝ 8 for all). *P < 0.05. 187.3 ⫾ 27.8, and 194.2 ⫾ 42.6 pmol/l; P ⬎ 0.05) (Fig. 1B) as gain increased. Early-phase insulin delivery was substantially higher as gain increased (peak values 2.5 ⫾ 0.9, 4.2 ⫾ 0.8, and 10.1 ⫾ 3.5 units/h, respectively; P ⬍ 0.05) (Fig. 1C). Insulin delivery 6 h postmeal was elevated at the 0.5 ⫻ TDD level compared with TDD and 1.5 ⫻ TDD (2.1 ⫾ 0.3 vs. 0.6 ⫾ 0.2 and 0.7 ⫾ 0.4 units/h; P ⬍ 0.05). The lower peak glucose levels achieved with increasing gain resulted in dramatically lower postprandial incremental area under the glucose curve (⌬AUCGLC) for the 6-h period after the meal (2,009 ⫾ 143, 1,182 ⫾ 158, and 1,000 ⫾ 106 mmol/l ⫻ min, for 0.5 ⫻ TDD, TDD, and 1.5 ⫻ TDD, respectively; P ⬍ 0.05) (Fig. 2A), particularly as the gain was increased from 0.5 ⫻ TDD to TDD. Insulin delivery over the same period tended to increase with increasing gain, but this did not achieve statistical significance (13.3 ⫾ 1.5, 14.0 ⫾ 0.9, and DIABETES, VOL. 55, JULY 2006 15.5 ⫾ 1.1 units, respectively; P ⬎ 0.05). All gains resulted in similar incremental insulin AUC (Fig. 2B), but the initial rise in plasma insulin was more rapid with high gain leading to significant differences in the 2-h incremental AUC (P ⬍ 0.01) (Fig. 2C). Hypoglycemia (⬍3.3 mmol) occurred in one experiment for 0.5 ⫻ TDD, three experiments for TDD, and two experiments for 1.5 ⫻ TDD. For all gains, insulin concentration and insulin delivery remained substantially elevated 6 h after the meal, even though glucose levels had, with the exception of the low gain response, returned to preprandial basal levels by t ⫽ 600 min (8.2 ⫾ 0.9, 6.0 ⫾ 0.6, and 6.0 ⫾ 0.5, respectively; P ⬎ 0.05) (Fig. 3). Insulin kinetic model analysis. Plasma insulin kinetics were well described by the insulin kinetic model (Eq. 3) for all gains (r2 ⫽ 0.77, 0.82, and 0.78 for 0.5 ⫻ TDD, TDD, and 1.5 ⫻ TDD, respectively; correlation calculated between measured and model predicted insulin levels) (Fig. 4). No differences were observed in insulin clearance (44.6 ⫾ 6.5, 48.3 ⫾ 5.0, and 46.9 ⫾ 6.1 ml 䡠 min⫺1 䡠 kg⫺1) or kinetic time constants (1 ⫽ 48.6 ⫾ 7.0, 58.4 ⫾ 5.5, and 1997 AUTOMATED CLOSED-LOOP INSULIN DELIVERY FIG. 4. Insulin kinetic model analysis and residuals for 0.5 ⴛ TDD (A and B), TDD (C and D), and 1.5 ⴛ TDD (E and F) gain (n ⴝ 8). Plasma insulin (F) is shown together with model fit (in solid line) and algorithm output (gray shaded area). Residual plots are shown for each level of gain in subpanels. 53.4 ⫾ 7.5 min; 2 ⫽ 33.0 ⫾ 6.9, 31.3 ⫾ 6.0, and 36.2 ⫾ 6.1 min) as gain increased (P ⬎ 0.05 for all), with predicted time-to-peak insulin concentration after a single bolus to be ⬃40 – 44 min at all gains. DISCUSSION The current study has three main conclusions. First, automated closed-loop glucose control, based on a model of the -cell, is achievable across a wide range of gains, 1998 using subcutaneous insulin delivery. Second, after meals, high plasma insulin levels are required well past the time when plasma glucose is normalized. Third, plasma insulin kinetics during closed-loop insulin delivery are well described by a simple two-compartment insulin model. The -cell model/closed-loop algorithm used here (Eq. 1) reproduces the well-established pattern of -cell insulin secretion in response to a hyperglycemic clamp (16,21). DIABETES, VOL. 55, JULY 2006 A.E. PANTELEON AND ASSOCIATES Although the model used here has been shown to describe hyperglycemic clamp data, other models of -cell secretion have been proposed (22–30). The models are similar insofar as each describes glucose-induced insulin secretion as the sum of components that react immediately to a change in glucose, have a delayed reaction to a change in glucose, and/or react to the rate of change of glucose. Of these models, the one proposed by Cerasi et al. (29) is of particular interest because it also produces a slowly rising second phase without using an integral term per se. Previous closed-loop studies in diabetic dogs have used phases analogous to P(t) and D(t), with intravenous insulin delivery, to show that meal excursions could be obtained that were actually lower than those observed in normal dogs (10,11). More recent studies using insulin analogs have evaluated other administration routes (12– 14). The main difference between the algorithms in these earlier studies (10 –14) and the algorithm used here is that the proposed -cell model has an integral component [I(t)] (Eq. 1) that can theoretically adjust insulin delivery (using transient periods of glucose concentration above or below set point) to restore steady-state glucose in the case of changes in insulin sensitivity or endogenous glucose production (result shown using computer simulations based on the Bergman minimal model) (16). Similar adaptation of the -cell was demonstrated in rats undergoing chronic glucose infusion, whereby an increase in basal insulin secretion was observed without change in glucose (31). The hypothesis is also consistent with earlier observation in humans that adiposity affects fasting insulin concentration independent of fasting glucose (32). In the PID algorithm, the integral component creates the slow rise in second-phase insulin observed during hyperglycemic clamps (21). In the current study, the integral component maintained an elevated insulin delivery rate after meals, even though near normoglycemia was established (Figs. 1 and 3). The need for elevated insulin delivery was likely caused by a prolonged rate of appearance of glucose after the meal, which has been shown in canines receiving basal insulin alone to lead to elevated plasma glucose levels for 16 –20 h (33). Although the maintenance of target glycemia during periods of elevated glucose appearance is desirable, it can be expected that once glucose appearance returns to normal, the elevated value of the integral component will result in a plasma glucose excursion below target (a consideration when deciding to use a high 6.7-mmol/l target rather than a more physiological 5.5 mmol/l). The reference gain used here was set in direct proportion to open-loop total daily insulin requirement (TDD) in an attempt to keep the product of KP and insulin sensitivity constant. Underlying this approach is the hypothesis that the -cell has a similar relationship between insulin secretion and insulin sensitivity—the so-called disposition index (34,35). It is not clear, however, whether the “reference KP” was optimal or whether TDD is a suitable surrogate for insulin sensitivity. The 50% increase in KP did produce a lower meal excursion without postmeal hypoglycemia (Fig. 1), and it could be argued to be a more optimal gain. However, at this gain, postprandial glucose had a tendency to undershoot/overshoot in some individual experiments, suggesting that a further increase in gain might lead to an oscillatory glucose response. Also affecting stability was the choice of TD and TI. During a hyperglycemic glucose clamp, TD determines the amount of rapidly released first-phase insulin, and TI DIABETES, VOL. 55, JULY 2006 determines the rate at which the second-phase response linearly increases (16). For human subjects, we have estimated -cell secretion to be well described with values of ⬃40 and 100 min (TD and TI, respectively) (16). Here, the amount of insulin delivered as first phase was substantially increased (TD ⫽ 66 min), and the second-phase rate of increase was substantially slower (TI ⫽ 150 min). Although these changes were made to compensate for the delays inherent in subcutaneous insulin delivery, they may not be optimal. Optimization of controller parameters (KP, TI, and TD), can likely only be achieved using a metabolic model (18). We attempted to describe the gain dose response with the minimal model of insulin action (36) and meal models proposed by Hovorka et al. (37) and by Lehmann et al. (38). However, the large change in AUCGLC observed as the gain was increased from 0.5 ⫻ TDD to TDD, versus the much smaller change observed as the gain was further increased to 1.5 ⫻ TDD (Fig. 1), could only be fit by introducing nonlinearities, either in insulin-dependent gastric carbohydrate extraction or in minimal model metabolic parameters (data not shown). Support for both mechanisms exists in the literature. Splanchnic glucose uptake (39) and the amount of ingested carbohydrate that is extracted by the gut (37,40) have both been argued to be affected by plasma insulin. Conversely, nonlinearities in insulin sensitivity and glucose effect have also been used in different metabolic models (rev. in 41). Differentiating between these two mechanisms and validating the resulting model describes the reference gain (TDD in this study) response, and perturbations in gain (0.5 ⫻ TDD and 1.5 ⫻ TDD) will be essential if a model-based optimization of control parameters (KP, TI, and TD) is to be performed. To differentiate between the two mechanisms, it is likely that closed-loop experiments, using glucose tracers, will need to be performed. Although we were not successful in describing the complete closed-loop dose response, plasma insulin kinetics were well described by the simple two-compartment approximation (r2 ⬎0.77) (Fig. 4). Time constants estimated from this approach suggest peak absorption times between 40 and 44 min, similar to the peak time observed in humans (42,43). Insulin clearance was not affected by changes in insulin delivery (P ⬎ 0.05) (see RESULTS), and no systematic residuals were observed in the fit of the simple two-compartment model, suggesting that a more complicated insulin absorption model (20,44,45) is unnecessary. The ability to describe plasma insulin concentration during closed loop allows for an insulin feedback term to be added to the PID -cell model. This would allow the putative inhibition of -cell insulin secretion by plasma insulin (46 – 48) to be added to the artificial -cell algorithm used here. In summary, automated closed-loop insulin delivery, based on a simple PID model of -cell secretion combined with subcutaneous insulin delivery, results in stable meal response for a wide range of gain. This suggests that the proposed closed-loop system can adjust to day-to-day variability in metabolic processes. Further optimization of algorithm tuning, either by adjusting the relative ratios of the different insulin delivery phases or by inclusion of insulin feedback, may lead to improved meal responses. Optimization is likely to be facilitated by using a more comprehensive metabolic model together with glucose tracer information. 1999 AUTOMATED CLOSED-LOOP INSULIN DELIVERY ACKNOWLEDGMENTS This study was supported by National Institutes of Health Research Grants DK57210 (to K.R.) and DK64567 (to G.M.S.). The authors thank Natalie Kurtz, Vera Stafekhina, and Gayane Voskanyan for their contribution to the experiments and subsequent laboratory analysis. REFERENCES 1. Feldman B, Brazg R, Schwartz S, Weinstein R: A continuous glucose sensor based on wired enzyme technology: results from a 3-day trial in patients with type 1 diabetes. Diabetes Technol Ther 5:769 –779, 2003 2. Kovatchev BP, Gonder-Frederick LA, Cox DJ, Clarke WL: Evaluating the accuracy of continuous glucose-monitoring sensors: continuous glucoseerror grid analysis illustrated by TheraSense Freestyle Navigator data. Diabetes Care 27:1922–1928, 2004 3. Tansey MJ, Beck RW, Buckingham BA, Mauras N, Fiallo-Scharer R, Xing D, Killman C, Tamborlane WV, Ruedy KJ: Accuracy of the modified continuous glucose monitoring system (CGMS) sensor in an outpatient setting: results from a diabetes research in children network (DirecNet) study. Diabetes Technol Ther 7:109 –114, 2005 4. Kerssen A, de Valk HW, Visser GH: The continuous glucose monitoring system during pregnancy of women with type 1 diabetes mellitus: accuracy assessment. Diabetes Technol Ther 6:645– 651, 2004 5. Steil GM, Rebrin K, Mastrototaro J, Bernaba B, Saad MF: Determination of plasma glucose during rapid glucose excursions with a subcutaneous glucose sensor. Diabetes Technol Ther 5:27–31, 2003 6. Steil GM, Rebrin K, Hariri F, Jinagonda S, Tadros S, Darwin C, Saad MF: Interstitial fluid glucose dynamics during insulin-induced hypoglycaemia. Diabetologia 48:1833–1840, 2005 7. Howey DC, Bowsher RR, Brunelle RL, Woodworth JR: [Lys(B28), Pro(B29)]-human insulin: a rapidly absorbed analogue of human insulin. Diabetes 43:396 – 402, 1994 8. Reynolds NA, Wagstaff AJ: Insulin aspart: a review of its use in the management of type 1 or 2 diabetes mellitus. Drugs 64:1957–1974, 2004 9. Hovorka R: Continuous glucose monitoring and closed-loop systems. Diabet Med 23:1–12, 2006 10. Rebrin K, Fischer U, von Woedtke T, Abel P, Brunstein E: Automated feedback control of subcutaneous glucose concentration in diabetic dogs. Diabetologia 32:573–576, 1989 11. Fischer U, Freyse EJ, Salzsieder E, Rebrin K: Artificial connection between glucose sensing and insulin delivery: implications of peritoneal administration. Artif Organs 16:151–162, 1992 12. Shimoda S, Nishida K, Sakakida M, Konno Y, Ichinose K, Uehara M, Nowak T, Shichiri M: Closed-loop subcutaneous insulin infusion algorithm with a short-acting insulin analog for long-term clinical application of a wearable artificial endocrine pancreas. Front Med Biol Eng 8:197–211, 1997 13. Sekigami T, Shimoda S, Nishida K, Matsuo Y, Ichimori S, Ichinose K, Shichiri M, Sakakida M, Araki E: Comparison between closed-loop portal and peripheral venous insulin delivery systems for an artificial endocrine pancreas. J Artif Organs 7:91–100, 2004 14. Matsuo Y, Shimoda S, Sakakida M, Nishida K, Sekigami T, Ichimori S, Ichinose K, Shichiri M, Araki E: Strict glycemic control in diabetic dogs with closed-loop intraperitoneal insulin infusion algorithm designed for an artificial endocrine pancreas. J Artif Organs 6:55– 63, 2003 15. Steil GM, Panteleon AE, Rebrin K: Closed-loop insulin delivery: the path to physiological glucose control. Adv Drug Deliv Rev 56:125–144, 2004 16. Steil GM, Rebrin K, Janowski R, Darwin C, Saad MF: Modeling beta-cell insulin secretion–implications for closed-loop glucose homeostasis. Diabetes Technol Ther 5:953–964, 2003 17. D’Azzo JJ, Houpis CH: Linear Control Systems: Analysis and Design. New York, McGraw-Hill, 2002 18. Steil GM, Rebrin K: Closed-loop insulin delivery: what lies between where we are and where we are going? Expert Opin Drug Deliv 2:353–362, 2005 19. Proakis JG, Manolakis DG: Digital Signal Processing: Principles, Algorithms, and Applications. Upper Saddle River, NJ, Prentice-Hall, 1996 20. Nucci G, Cobelli C: Models of subcutaneous insulin kinetics: a critical review. Comput Methods Programs Biomed 62:249 –257, 2000 21. Elahi D: In praise of the hyperglycemic clamp: a method for assessment of -cell sensitivity and insulin resistance. Diabetes Care 19:278 –286, 1996 22. Breda E, Toffolo G, Polonsky KS, Cobelli C: Insulin release in impaired glucose tolerance: oral minimal model predicts normal sensitivity to glucose but defective response times. Diabetes 51 (Suppl. 1):S227–S233, 2002 23. Breda E, Cavaghan MK, Toffolo G, Polonsky KS, Cobelli C: Oral glucose tolerance test minimal model indexes of -cell function and insulin sensitivity. Diabetes 50:150 –158, 2001 24. Mari A, Camastra S, Toschi E, Giancaterini A, Gastaldelli A, Mingrone G, 2000 Ferrannini E: A model for glucose control of insulin secretion during 24 h of free living. Diabetes 50 (Suppl. 1):S164 –S168, 2001 25. Mari A, Schmitz O, Gastaldelli A, Oestergaard T, Nyholm B, Ferrannini E: Meal and oral glucose tests for assessment of beta-cell function: modeling analysis in normal subjects. Am J Physiol 283:E1159 –E1166, 2002 26. Mari A, Tura A, Gastaldelli A, Ferrannini E: Assessing insulin secretion by modeling in multiple-meal tests: role of potentiation. Diabetes 51 (Suppl. 1):S221–S226, 2002 27. Cretti A, Lehtovirta M, Bonora E, Brunato B, Zenti M, Tosi F, Caputo M, Caruso B, Groop LC, Muggeo M, Bonadonna RC: Assessment of beta-cell function during oral glucose tolerance test by minimal model of insulin secretion. Exp Clin Endocrinol 31:405– 416, 2003 28. Hovorka R, Chassin L, Luzio SD, Playle R, Owens DR: Pancreatic beta-cell responsiveness during meal tolerance test: model assessment in normal subjects and subjects with newly diagnosed noninsulin-dependent diabetes mellitus. J Clin Endocrinol Metab 83:744 –750, 1998 29. Cerasi E, Fick G, Rudemo M: A mathematical model for the glucose induced insulin release in man. Eur J Clin Invest 4:267–278, 1974 30. Nesher R, Cerasi E: Modeling phasic insulin release: immediate and time-dependent effects of glucose. Diabetes 51 (Suppl. 1):S53–S59, 2002 31. Steil GM, Trivedi N, Jonas JC, Hasenkamp WM, Sharma A, Bonner-Weir S, Weir GC: Adaptation of beta-cell mass to substrate oversupply: enhanced function with normal gene expression. Am J Physiol 280:E788 –E796, 2001 32. Bagdade JD, Bierman EL, Porte D Jr: The significance of basal insulin levels in the evaluation of the insulin response to glucose in diabetic and nondiabetic subjects. J Clin Invest 46:1549 –1557, 1967 33. Goriya Y, Bahoric A, Marliss EB, Zinman B, Albisser AM: Responses to mixed meals in pancreatectomized dogs deprived of postprandial insulin. Am J Physiol 242:E335–E342, 1982 34. Bergman RN, Ader M, Huecking K, Van Citters G: Accurate assessment of beta-cell function: the hyperbolic correction. Diabetes 51 (Suppl. 1):S212– S220, 2002 35. Kahn SE, Prigeon RL, McCulloch DK, Boyko EJ, Bergman RN, Schwartz MW, Neifing JL, Ward WK, Beard JC, Palmer JP, et al.: Quantification of the relationship between insulin sensitivity and -cell function in human subjects. Evidence for a hyperbolic function. Diabetes 42:1663–1672, 1993 36. Bergman RN, Finegood DT, Ader M: Assessment of insulin sensitivity in vivo. Endocr Rev 6:45– 86, 1985 37. Hovorka R, Canonico V, Chassin LJ, Haueter U, Massi-Benedetti M, Orsini FM, Pieber TR, Schaller HC, Schaupp L, Vering T, Wilinska ME: Nonlinear model predictive control of glucose concentration in subjects with type 1 diabetes. Physiol Meas 25:905–920, 2004 38. Lehmann ED, Deutsch T, Roudsari AV, Carson ER, Sonksen PH: Validation of a metabolic prototype to assist in the treatment of insulin-dependent diabetes mellitus. Med Inform (Lond) 18:83–101, 1993 39. Basu R, Basu A, Johnson CM, Schwenk WF, Rizza RA: Insulin doseresponse curves for stimulation of splanchnic glucose uptake and suppression of endogenous glucose production differ in nondiabetic humans and are abnormal in people with type 2 diabetes. Diabetes 53:2042–2050, 2004 40. Dalla MC, Caumo A, Cobelli C: The oral glucose minimal model: estimation of insulin sensitivity from a meal test. IEEE Trans Biomed Eng 49:419–429, 2002 41. Steil GM, Clark B, Kanderian S, Rebrin K: Modeling insulin action for development of a closed-loop artificial pancreas. Diabetes Technol Ther 7:94 –108, 2005 42. Mudaliar SR, Lindberg FA, Joyce M, Beerdsen P, Strange P, Lin A, Henry RR: Insulin aspart (B28 asp-insulin): a fast-acting analog of human insulin: absorption kinetics and action profile compared with regular human insulin in healthy nondiabetic subjects. Diabetes Care 22:1501–1506, 1999 43. Bolli GB, Di Marchi RD, Park GD, Pramming S, Koivisto VA: Insulin analogues and their potential in the management of diabetes mellitus. Diabetologia 42:1151–1167, 1999 44. Wilinska ME, Chassin LJ, Schaller HC, Schaupp L, Pieber TR, Hovorka R: Insulin kinetics in type-I diabetes: continuous and bolus delivery of rapid acting insulin. IEEE Trans Biomed Eng 52:3–12, 2005 45. Serne EH, IJzerman RG, Gans RO, Nijveldt R, De Vries G, Evertz R, Donker AJ, Stehouwer CD: Direct evidence for insulin-induced capillary recruitment in skin of healthy subjects during physiological hyperinsulinemia. Diabetes 51:1515–1522, 2002 46. Baynes C, Anyaoku V, Johnston DG, Elkeles RS: Feedback inhibition of insulin secretion in type 2 diabetes. Clin Sci (Lond) 81:685– 690, 1991 47. DeFronzo RA, Binder C, Wahren J, Felig P, Ferrannini E, Faber OK: Sensitivity of insulin secretion to feedback inhibition by hyperinsulinaemia. Acta Endocrinol (Copenh) 98:81– 86, 1981 48. Elahi D, Nagulesparan M, Hershcopf RJ, Muller DC, Tobin JD, Blix PM, Rubenstein AH, Unger RH, Andres R: Feedback inhibition of insulin secretion by insulin: relation to the hyperinsulinemia of obesity. N Engl J Med 306:1196 –1202, 1982 DIABETES, VOL. 55, JULY 2006