alarming device

advertisement
ALARMING DEVICE
FOR DETECTION OF GAMMA RADIATION
AND TOXIC VAPORS PM2010M
OPERATING MANUAL
CONTENTS
1 INTRODUCTION ……………………………………………………………………….3
2 DESCRIPTION AND OPERATION OF THE DEVICE .........................................4
2.1 Application of the device ...................................................................................4
2.2 Delivery kit ........................................................................................................5
2.3 Specifications ....................................................................................................6
2.4 Design and theory of operation .........................................................................8
2.4.1 Design of the device ................................................................................8
2.4.2 Principle of the device operation ..............................................................8
2.4.3 Operation modes .....................................................................................10
2.4.3.1 Test mode ......................................................................................11
2.4.3.2 Calibration mode.............................................................................11
2.4.3.3 TSM blowing mode .........................................................................11
2.4.3.4 Mode of detection of exceeding the preset thresholds of gamma
radiation and concentrations of toxic substance vapors ............................11
2.4.3.5 Mode of communication with PC ....................................................11
2.5 Marking and sealing ..........................................................................................12
2.6 Packing .............................................................................................................12
3 USE OF THE DEVICE .........................................................................................13
3.1 Preparation for use............................................................................................13
3.1.1 General information .................................................................................13
3.1.2 Safety precautions ...................................................................................13
3.1.3 Preparation for operation .........................................................................13
3.1.4 Device operation check up.......................................................................13
3.2 Device operation ..............................................................................................14
3.2.1 Turning on, testing and calibration of the device.......................................14
3.2.2 Operation in the mode of detection of exceeding the set thresholds
of gamma radiation and concentrations of toxic substance vapors....................14
3.2.3 Parameters setting....................................................................................15
3.2.4 Operation in the mode of communication with PC ....................................15
3.2.5 Control over the battery voltage................................................................15
3.2.6 Use of external power supply unit.............................................................16
3.2.7 Turning off the device ..............................................................................16
4 MAINTENANCE...................................................................................................17
5 TROUBLESHOOTING.........................................................................................20
6 STORAGE AND SHIPPING.................................................................................21
7 UTILIZATION ......................................................................................................22
8 WARRANTY ........................................................................................................23
APPENDIX A Copy of certificate of toxic substance detection module (TSM)........24
2
1 INTRODUCTION
The present Operating Manual is intended to describe the design and operation of the
alarming device for detection of gamma radiation and toxic vapors PM2010M (hereinafter
referred to as device). The Operating Manual includes general description of the device,
specifications, operation instructions, instructions for its maintenance, as well as some other
information necessary for the proper operation of the device and full realization of its possibilities.
The name of the device in the documentation is written as follows: “Alarming device
for
detection
of
gamma
radiation
and
toxic
vapors
PM2010M
Type 100345122.042-2004”.
During manufacturing of the device some changes may be introduced in its electrical
scheme, construction, appearance and software that do not influence the specifications and
metrological parameters and, therefore, may be not specified in this manual.
3
2 DESCRIPTION AND OPERATION OF THE DEVICE
2.1 Application of the device
2.1.1 The device is designed for
- alarming about exceeding a set threshold of the ambient dose equivalent rate !*(10) of
photon radiation (DER) relative to set threshold;
- alarming about exceeding vapor concentration in air of toxic substances organophosphorus substances (sarin, soman, V-gases) and arsenic-containing substances
(lewisite) relative to the threshold of toxic substances vapors concentration.
The device operation history is stored in its non-volatile memory and can be transferred
to a personal computer (PC) via an infrared (IR) interface of data transmission using IrDA
interface.
Note! The device encloses a toxic substance vapor detection module (TSM),
containing 63Ni beta radiation source.
The device may be operated both indoor and outdoor by a wide range users involved in
detecting of ionizing radiation sources and vapors of toxic substances.
The device may be used by staff of emergency services, by border guards, customs
officers and other services.
2.1.2 The device is to be used under environmental conditions as follows:
- temperature
from –20 oC to + 60 oC;
o
- relative humidity at +25 C
and lower temperature
up to 98 %;
- atmospheric pressure
from 84 kPa to 106.7 kPa;
- air dust content, no more than
0.2 mg/l.
4
2.2 Delivery kit
2.2.1 The delivery kit of the alarming device is presented in the Table 1.
Table 1
Item
Alarming device for detecting of gamma
radiation and toxic vapors PM2010M
Operating manual
Case
CD with software
Passport of radioactive source
1 piece
1 piece
1 piece
1 piece
1 piece
Certificate of toxic substance module
1 piece
Battery Energizer L91BP – 2 AA 1)
IR adapter (ACT-IR220L or IR-210B) 2)
Power supply for РМ2010M including2:
transformer
cable
Set of spare parts and tools (SET-O):
Filter
Rod
Screwdriver
Filtering tissue element
Filtering element
Bottle with filter filler
Coarse calico strip
Imitator 2)
Rotameter 2)
Fitting
Filter
2 pieces
1 piece
1)
2)
Quantity
Note
included in operating
manual, Appendix B
included in operating
manual, Appendix A
1 piece
1 piece
1 piece
1 piece
1 piece
4 pieces
4 pieces
1 piece
0.4 m
1 piece
1 piece
1 piece
4 pieces
other batteries with similar parameters may be used;
available as option upon request
5
2.3 Specifications
2.3.1
2.3.2
2.3.3
2.3.4
2.3.5
2.3.6
2.3.7
2.3.8
2.3.9
2.3.10
2.3.11
2.3.12
Modes of operation
- test mode;
- calibration mode;
- TSM blowing mode;
- mode of detection of exceeding the set
thresholds of gamma radiation and
concentrations of toxic substance vapors;
- mode of communication with PC
Range of the DER threshold
from 0.001 to 999 mSv/h
Step of threshold setting
1 µSv/h
Main relative error of a comparison of a ± (30+K/!) %,
registered DER value at 137Cs radiation where ! – DER value, mSv/h,
with the set DER threshold value
K – coefficient equal 0.01 mSv/h
Gamma radiation energy range
from 0.06 to 3.0 MeV
Energy response relative to 137Cs, no
more
30 %
Time of the device response in case of a
step DER increase from a value of
10 µSv/h to a value of no less than
10 s
100 µSv/h, no more than
The device should produce audio and
visual signals in case when the set
thresholds are exceeded and record in
non-violate memory the events of set
DER threshold exceeding and history of
DER registration in case of supply
from 2.5 to 3.2 V
voltage changes
Threshold of detection (with a toxic
substance module) at contents higher
than threshold values at normal climatic
conditions (NCC): temperature 20±5 oC;
atmospheric
pressure
101.3(+5.4;
-17,3) kPa; relative humidity 60(+20;
-30) % should be no more than
- vapors of toxic organophosphorus
5⋅10-5 mg/l for 15 s
substances
- arsenic-containing substance
2⋅10-4 mg/l for 5 s
Variation of the threshold of detection
(with a toxic substance module) at limit
temperature values compared to the
-50 %
threshold value at NCC, no more than
Variation of the threshold of detection
(with a toxic substance module) at
relative air humidity (98 % at 25 oC and
lower) compared to the threshold value
-400 %
at NCC, no more than
Residual effect of a toxic substance
module at operation temperature range,
30 s
no more than
6
2.3.16
Instability of comparison of registered
DER value with threshold value for 24 h
of continuous operation no more than
Device should not produce visual and
audible alarms when substances as
follow are present in air:
- gasoline at vapor concentration no
more than
- diesel fuel (used both in summer and
winter) at vapor concentration no more
than
- exhausts of isolated internalcombustion engines (gasoline or diesel)
at distance to exhaust tube more than
Battery lifetime at normal conditions and
audio alarm sounding no more than
5 min/24 h
Device is resistant to
2.3.17
Device is resistant to
2.3.13
2.3.14
2.3.15
2.3.18 Device is resistant to RF electrical and
magnetic fields (under IEC 61000-4-3:1995):
- in the frequency range of 80 to 1000 MHz
- in the frequency range of 800 to 960 MHz
and of 1.4 to 2.0 GHz (generated by digital
mobile phones)
2.3.19 Device is resistant to electrostatic
discharges (under IEC 61000-4-2:1995)
2.3.20 Device in a transportation case is proof
against
2.3.21
2.3.22
2.3.23
2.3.24
2.3.25
5%
1.5 mg/l
0.5 mg/l
15 m
40 h
- temperature of ambient air from -20 to
+60 oC;
- relative air humidity up to 98 % at 25 oC;
- atmospheric pressure from 84 to 106.7 kPa
- sinusoidal vibration in a frequency range
of 5 – 35 Hz and a bias amplitude for
frequencies lower than the transition
frequency 0.75 mm;
- repeated shocks with acceleration of
100 m/s2, duration of shock pulse of
2-50 ms, at shock rate of 60-180 shocks per
minute
10 V/m (severity degree 3)
30 V/m (severity degree 4)
8 kV
- temperature from – 20 up to + 60 oC;
- humidity up to 100 % at 35 oC;
- shocks with an acceleration of 147 m/s2, a
duration of 10 ms;
- vibrations with a frequency of 5-35 Hz and
a bias amplitude 0.75 mm
IP31
0,55 kg
1,2 kg
60х35х230 mm
Protection degree of the case
Device weight, no more than
Device weight in a case, no more than
Device dimensions, no more than
Reliability parameters:
- average time to first failure
no less than 20 000 h
- average service life
no less than 10 years
- average time of recovery
no more than 120 min
7
2.4 Design and theory of operation
2.4.1 Design of the device
2.4.1.1 The device consists of two independent modules:
- gamma radiation detection module (GMM);
- toxic substance vapor detection module (TSM).
2.4.1.2 The device is designed as a monoblock consisting of a toxic substance detection
module (TSM), a Geiger-Muller module (GMM), a printed circuit board of a microprocessor
controller (MPC) and a housing involving an audio alarming device (AAD) and batteries. The
input connecting pipe 2 of TSM (Fig. 1) with a cap 16 is located on the top cover of the device.
When the device is operated the cap is taken off the input connecting pipe 2 and fixed adjacent
to the connecting pipe at a plug 1. During the device calibration, the filter 15 involved in the set
of spare parts and tools (SET-O) is screwed on the connecting pipe 2. The Red-LED Ph 5, RedLED Arc 6, Yellow-LED 4, Green-LED 7 , ON/OFF button 9, window 8 of the IR interface and
device serial number 3 are located on the front panel of the device.
The clip 14 is placed on the back side of the device. The Fig. 1 shows how to mount and
remove the clip.
The Geiger-Muller module (GMM) is located in the bottom part of the device. The
housing 12 involving batteries and the audio alarming device (AAD) 13, as well as the cap 11 of
the output connecting pipe are fixed to the lower cover of the device.
The GMM effective center 10 is shown in Fig. 1, and an arrow points a direction of
calibration.
2.4.2 Principle of the device operation
2.4.2.1 The principle of the device operation is based on comparison of the registered
DER value with the set DER threshold and the registered concentration value of toxic substance
vapors with the set threshold value of concentration level.
The dose equivalent rate is registered in the mode of detection of exceeding the set
thresholds of gamma radiation using GMM based on the Geiger-Muller counter.
Exceeding of air concentrations of toxic substances vapors relative to set threshold level
is detected by using a TSM based on an ionizing chamber with 63Ni beta-source. An analyzed
air is pumped through the chamber using a micropump.
The detection modules are operated by a microprocessor controller with a built-in
analogue-digital converter (ADC).
The device is powered by two AA batteries with a nominal voltage of each battery 1.5 V,
or by 220V AC through a 12V/1,1A power supply unit and transformer or by 9-36V car battery
by using a transformer from the delivery set of the device.
8
1 - plug for fixing the cap of the input connecting pipe of TSM; 2 – input connecting pipe of TSM;
3 – window with the device serial number; 4 – Yellow-LED; 5 –Red-LED Ph; 6 – Red-LED Arc;
7- Green-LED; 8 – input window of the IR interface; 9 – ON/OFF button; 10 – GMM geometric
center; 11 - cap of the output connecting pipe; 12 – housing with batteries; 13 – window for the
audio alarming device; 14 – clip; 15 – filter, 16 – cap; 17– filter cover
Fig. 1
9
2.4.2.2 A block diagram of the device is presented in Fig. 2.
The device consists of:
- gamma radiation detection module (GMM);
- toxic substance vapor detection module (TSM);
- microprocessor controller (MPC);
- audio alarm device (AAD);
- red light-emitting diode (R-LED Ph) for alarming about exceeding a threshold of
organophosphorus substance vapor concentration;
- red light-emitting diode (R-LED Arc) for alarming about exceeding a threshold of
arsenic-containing substance vapor concentration;
- yellow light-emitting diode (Y-LED) for alarming about exceeding a threshold of the
gamma radiation DER;
- green light-emitting diode (G-LED) for indicating the device ON state and the battery
voltage;
- infrared interface (IRI);
- ON/OFF button (BN);
- batteries (B);
- micropump (MP).
РМ 2010М
AAD
BN
Fig. 2 - Block diagram of the device
B
TSM
MPC
GMM
IRI
MP
IrDA
IrD
R-LED
Ph
R-LED
Arc
СЗ
G-LED
Y-LED
Fig.2 – Block diagram of the device
The device operation algorithm provides for a continuous measurement, statistical
processing of measurement results, fast adaptation to variations of the radiation intensity (a
time of measurement is inversely dependent on the radiation intensity). An infrared interface (IrI)
is used for an exchange of information between MPC and PC.
The device is equipped with a non-volatile memory used to accumulate and store
information.
2.4.3 Operation modes
The device has the following operation modes:
- test mode;
- calibration mode;
- TSM blowing mode;
- mode of detection of exceeding the set thresholds of gamma radiation and
concentrations of toxic substance vapors;
- mode of communication with PC.
10
In any operation mode the device controls over the battery voltage (see clause 3.2.5).
2.4.3.1 Test mode
The device enters this mode every time when it is turned ON or the batteries are
replaced. While in this mode the microprocessor controller provides for testing and diagnostics
of the main device modules. If all tests are passed successfully the device enters the detection
or calibration mode.
2.4.3.2 Calibration mode
The device enters this mode automatically after the test mode is finished if the batteries
were replaced before its turning ON or after the long (more than 3 s) pressing of the ON/OFF
button.
While in this mode, the device performs the calibration of the TSM channel (blow through
the ionization chamber, autocalibration of "zero" of the TSM channel). After calibration is
finished, the device automatically goes into the mode of detection of exceeding the preset
thresholds of gamma radiation and concentrations of toxic substance vapors.
2.4.3.3 TSM blowing mode
The device can enter this mode in case that the TSM calibration process takes more
than 2 minutes because of the effect of high toxic substance vapor concentration or the ionizing
chamber is contaminated. In this mode the ionizing chamber is blown out. If the device stays in
this mode more than 5 minutes it is required to clean the gas systems and the ionizing chamber.
In case that blowing passes successfully the device returns automatically to the calibration
mode.
2.4.3.4 Mode of detection of exceeding the preset thresholds of gamma radiation and
concentrations of toxic substance vapors
While in this mode the device performs a continuous and independent control of the
radiation and chemical conditions.
While in this mode the device produces light and audible signals as follows:
- "Chemical alarm" (Red-LED Ph or Red-LED Arc is flashing) when the threshold value of
the TSM is exceeded;
- "Radiation alarm" (Yellow-LED is flashing) when the preset DER threshold is exceeded.
The light alarms are accompanied by audible alarms if these are enabled in the mode of
communication with PC.
The device provides for a possibility to record and store the history of operation (DER
values with set time intervals, alarms, i.e. exceeding of the set threshold value of DER and
concentration of TSM), and values of preset parameters in the non-volatile memory.
2.4.3.5 Mode of communication with PC
The device performs an exchange of information with PC using WINDOWS
98/2000/NT/XP through the infrared channel. An IR adapter and an exchange protocol
compatible with the IrDA interface are used.
While in the mode of communication with PC the following operations may be performed:
- the device can be registered to a user;
- to store a time when the device was received and returned by a user;
- to read information from the device memory;
- to change the following device parameters:
- the set DER threshold;
- to disable audio alarms;
- the time interval to store the history of operation;
11
-to store and transmit to PC the history of the measurement of the photon radiation DER,
records about exceeding the preset thresholds of gamma radiation and concentrations of toxic
substance vapors.
2.5 Marking and sealing
The logo and name of the manufacturer, name of the device, IP code, and the device
serial number are inscribed on the device. The "Radiation danger" sign with the source type and
its activity is marked on the bottom side of the device housing.
A seal that prevents an unauthorized opening of the device is placed in the faceplate of
the device in the screw groove.
2.6 Packing
The device is put into a pasteboard box together with its documents and SET-O.
12
3 USE OF THE DEVICE
3.1 Preparation for use
3.1.1 General information
3.1.1.1 When buying the device please check the delivery kit according to clause 2.2.1
and perform the device operation check up according to clause 3.1.4.
The device should not be subjected to severe mechanical shocks, any strong chemicals
and solvents and open fire sources.
3.1.2 Safety precautions
3.1.2.1 During the device assembling, calibration, repair and maintenance when
radioactive sources are used, all the requirements and regulations ensuring the radiation and
sanitary safety should be observed.
3.1.2.2 The beta radiation source IBIRRZN-63-IIa, 7016-056-41805307-2002 which is
included in the ionization chamber contains the 63Ni radionuclide of 0.6 GBq activity. This
source is located inside of the ionizing chamber in the device housing.
3.1.2.3 The beta radiation DER does not exceed the natural background level at any
point of the outer surface of the package with the device inside. The device design excludes the
possibility of the access to the source during device operation. Therefore it is not necessary to
take any additional safety precautions actions for radiation security.
3.1.3 Preparation for operation
3.1.3.1 Before using the device, please read this operating manual carefully.
3.1.3.2 Take the device from the package.
3.1.3.3 Insert the batteries into the battery compartment 12 (Fig. 1).
3.1.3.4 Remove cap 16 from the connecting pipe of the ionizing chamber 2 and the
output connecting pipe 11.
3.1.3.5 Screw cap 16 on the connecting pipe at a plug 1, and put cap 10 of the output
connecting pipe into SET-O.
3.1.3.6 When starting the operation after the long storage of the device, it is required to clean
the gas system and the ionizing chamber in according to clause 4.4 of the present Manual.
3.1.3.7 Calibrate the device according to clause 3.2.1.2. After calibration is finished the
device is ready for operation.
3.1.4 Device operation check up
3.1.4.1 After device calibration according to clause 3.2.1 it is necessary to check up the
device operation. After calibration is finished, the Green-LED should light constantly and a
sound of the working micropump should be heard.
To check up the chemical channel operation, bring the Phosphorous imitator from the
SET-O (set of the spare parts and tools) to the input of TSM. The light alarm (Red-LED Ph)
should flash and audio alarm will be heard. Remove the imitator. The light and audio alarms will
be switched OFF in no more than 30 s. After this bring the Arsenic imitator from the SET-O to
the input of TSM. The light alarm (Red-LED Arc) should flash and audio alarm will be heard.
Remove the imitator. The light and audio alarms will be switched OFF in no more than 30 s.
Caution! When checking up the chemical channel operation with the imitator from
the SET-O in enclosed space it is necessary to provide the blowout from the output
connecting pipe outdoors (exhaust ventilation). Before the iterative checking up it is
necessary to blow out the gas system through the TSM filter for 10-20 minutes.
To check up the radiation channel, bring a radioactive source (its energy range is from
0.06 to 3.0 MeV) close to the GMM geometrical center (Fig. 1). The light alarm (Yellow-LED) will
flash and audio alarm will be heard. Remove the source. The light and audio alarms will be
switched OFF.
13
To turn the device OFF, press the button 9. The Green-LED and the micropump should
be switched OFF.
3.2 Device operation
3.2.1 Turning on, testing and calibration of the device
3.2.1.1 To turn the device ON, press the ON/OFF button 9 (Fig. 1). The device enters the
test mode after it is turned ON or after the batteries are replaced. While the device is in this mode,
Red-LED Ph, Red-LED Arc and Yellow-LED are flashing in turn. After completion of the test mode
the device can go either to the calibration mode (if batteries were replaced), or to the mode of
detection of exceeding the preset thresholds. The device is ready for operation, when Red-LED Ph,
Red-LED Arc and Yellow-LED are switched OFF and Green-LED is continuously ON.
3.2.1.2 While in the calibration mode, Red-LED Ph, Red-LED Arc and Yellow-LED are
ON simultaneously. The device enters this mode if the batteries were replaced or the ON/OFF
button was pressed for a long time. The calibration of TSM is performed in this mode. To
calibrate the TSM, mount the filter 15 (Fig. 1) (included in SET-O) to the input connecting pipe 2
and turn its cover 17 to the “Open” position. The calibration will be completed in no more than 2
minutes. The calibration completion is indicated by Red-LED and Yellow-LED switching OFF.
If after 2 minutes the device goes to the TSM blowing mode, one need to wait for not more than
5 minutes while completion of this mode. If the device stays in the TSM blowing mode more
than 5 minutes, it is required to clean the gas systems and the ionizing chamber of the device
according to clause 4.4 and to turn on the device once again. After the calibration is finished,
turn the filter cover 17 to the “Close” position, remove the filter 15 and bring it into the right
position in the SET-O. The device automatically enters the mode of detection of exceeding the
preset thresholds of the gamma radiation DER and concentrations of toxic substance vapors.
Caution! If the calibration process was interrupted (the device was turned OFF
before the simultaneous flashing of the Yellow-LED and the Red-LED was stopped), then
in the mode of detection of exceeding of the set thresholds of gamma radiation and
concentrations of toxic substance vapors the Red-LED will be flashing with 10 s
intervals. In this case to ensure detection thresholds, perform calibration until both the
R-LED and the Y-LED are off.
3.2.2 Operation in the mode of detection of exceeding the set thresholds of gamma
radiation and concentrations of toxic substance vapors
3.2.2.1 While in the mode of detection of exceeding the preset thresholds of the gamma
radiation DER and concentrations of toxic substances vapors, the device can be used for the
simultaneous detection of toxic substance vapors (TSV) in air and the detection and location of
gamma radiation sources (GRS).
The device operates with the parameter values set by the manufacturer (see clause
3.2.3.1). To use other parameters, it is necessary to set the other values (see clause 3.2.4).
The device automatically enters the mode of detection of exceeding the preset thresholds
of GRS and TSV after the completion of the calibration mode or after completion of test mode.
3.2.2.2 Detection of GRS and TSV
To achieve the maximum sensitivity during the GRS detection, the device should be
placed in such a position that the calibration direction indicated in Fig. 1 coincides with a
direction to a scanned object.
The detection efficiency is dependent on the proximity of the device to an examined
object (thing, car, person, etc) and velocity of its movement along the object.
To achieve the maximum sensitivity during the TSV detection, the input connecting pipe
of the toxic substance module should be completely opened.
3.2.2.3 Location of GRS
When a gamma radiation source is detected, it can be located.
14
To locate the source, the device should be at no more than 10 cm distance from the
scanned object. The velocity of the device movements along the scanned object should not
exceed 1cm/s. As the device moves closer to the gamma radiation source, a rate at which
Yellow LED is flashing and audio tone is repeated will increase.
When the maximum rate of audio signals is achieved, a continuous audio tone is
produced. The source can not be located better.
N o t e - The location of GRS is only possible in the only case when DER emitted by a
source at a distance of 10 cm is higher than 1 µSv/h, as the minimum settable threshold equals
to 1 µSv/h.
3.2.3 Parameters setting
3.2.3.1 The preset parameters
The device is shipped with the preset parameters as follows:
- time intervals for storage of current photon radiation DER
values in the non-volatile memory of the device
- threshold photon radiation DER value
- threshold for detection of vapors of
- organophosphorus substances (sarin, soman, V-gases, etc. )
- arsenic-containing substances (lewisite)
- audio alarm device
60 min
1 µSv/h
5⋅10-5 mg/l
2⋅10-4 mg/l
enabled
3.2.4 Operation in the mode of communication with PC
3.2.4.1 The device is in the mode of communication with PC during 15 s after it is turned
ON. To repeat a communication, it is necessary to turn the device OFF and then to turn it ON
again.
To use the mode of communication with PC is possible only after the application program
supplied on CD is installed on PC.
An IR adapter and the application program PM2010M (supplied on CD) should be used
to operate with the device in the mode of communication with PC.
Minimum requirements to a computer and its software:
- Pentium;
- 16 Mb RAM;
- Windows 98/NT/2000 XP (small font);
- 2 Mb free on the hard disk and an adequate space available to install the database;
- 800x600 resolution of a monitor.
Connect the cable of the IR adapter to a communication port of the personal computer or
use the PC with a built-in IR adapter.
To install the application program, use the CD with software included in the delivery kit.
The Help file describes how to run the application program.
To communicate with PC, turn the device ON and place it for 15 s near the IR adapter at
a distance of 10-20 cm, the IR interface window should face the IR adapter. The information
exchange between the device and PC will be carried out in some seconds. Follow the
instructions given in the application program for further activities.
3.2.5 Control over the battery voltage
3.2.5.1 In any mode of operation, the device continuously controls over the batteries
voltage, that is indicated by a solid light of Green-LED. If the batteries voltage drops below the
permissible level, Green-LED flashes. The batteries should be replaced. If the batteries are not
replaced, the device microprocessor switches OFF the micropump and the device will operate in
the only mode of detection of gamma radiation sources, and the TSM will not operate (GreenLED is flashing). If the batteries voltage drops even lower, the device will be turned OFF.
15
3.2.6 Use of external power supply unit
To use the external power supply of PM2010M:
- unscrew the housing with batteries 12 with the help of screwdriver (Fig.1);
- place the converter 3 (Fig.3) observing the polarity;
- fix the converter with a screw.
To power the device from car battery in a range of 9 – 36V, use cable 2 with red positive
clamp. To power the device from the 220V AC, use 12V/1,1A power supply 1 from the delivery
set.
1 – power supply; 2- connecting cable; 3- converter
Fig. 3
3.2.7 Turning off the device
3.2.7.1 To turn the device OFF, press the button 9 (Fig. 1). After all LEDs are OFF put
the cap 16 to the input connecting pipe of TSM 2 and the cap of the output connecting pipe 11
in its place.
16
4 MAINTENANCE
4.1 General guidelines
4.1.1 The device maintenance is performed by specially trained personnel who have
finished the training and have the experience in maintenance.
4.1.2 The maintenance is to be carried out in due time to provide the device readiness for
operation and maximum service life.
4.1.3 The maintenance is carried out:
- when the device is delivered by the manufacturer;
- in the course of operation.
4.2 Safety measures during maintenance
4.2.1 The TSM module includes the beta radiation source. The TSM module design
ensures the operation safety of personnel without using special radiation shields. The radiation
background level on the device surface is within the natural background values.
WARNING! It is forbidden to disassemble the ionization chamber and to take out
the beta radiation sources without a special need.
4.2.2 After the beta radiation source service life, indicated in source passport (Appendix
B) is expired, it should be replaced by a new one. The radioactive source replacement should
be performed by the specially trained personnel of the manufacturer.
4.2.3 The utilization of used sources should follow a determinate order observing the
requirements of handling radioactive materials.
4.3 Maintenance procedure
4.3.1 When the device is put into service, as well as during operation the scheduled
maintenance should be provided. A list of activities and the time intervals between their
fulfillment are given in the Table 2.
Table 2
List of activities
Time intervals between activities
1 Cleaning of gas systems and
when putting into service;
ionizing chamber:
- detection of
organophosphorus
1 year;
substances
- detection of arsenic –
1 year
containing substances
2 Replacement of filtering
elements of filter
2.5 years
3 Replacement of radioactive
source
5 years
4 Decontamination of device
as needed, but at least once a year
5 Replacement of the batteries
as required
6 Measurement of output rate of
After the works described in item 1
the micropump
7 TSM operation check up by imitator
1 year
17
4.4 Cleaning of gas system and ionizing chamber
4.4.1 To clean the input connecting pipe of the TSM:
- turn the device OFF, put the device in an upright position, filter 15(Fig. 1) up, take the
filter out;
- take the rod and a coarse calico strip, 8-10 mm wide, 10-15 mm long (from the
SET-O), pass the strip through the rod slot, moisten it with alcohol and squeeze to
remove drops;
- insert the rod into the TSM input connecting pipe to a depth of 25-30 mm and wipe
the rod hole with rotator movements;
- renew the strip and repeat the procedure until the tissue is quite clean.
4.4.2 To clean the ionizing chamber:
- insert the rod with tissue moistened with alcohol into the ionizing chamber through the
cleaned input connecting pipe;
- wipe the chamber internal surface with rotator movements.
Repeat the procedure 2-3 times renewing the tissue strip.
When the cleaning is completed the device should be held OFF without filter for one
hour.
4.4.3 To blow out the gas system, it is necessary to turn the device ON (the filter cover
open) and hold it ON for one hour. False “chemical alarms” may occur. After an hour turn the
device OFF and close the filter cover.
CAUTION! When the maintenance of the gas system and ionizing chamber is
performed, alcohol drops and tissue fibers should not be left in the ionizing chamber and
connecting pipe.
4.5 Replacement of the filtering elements of the TSM filter
4.5.1 Unscrew the filter 15 (Fig. 1).
Figure 4 shows the filter structure. The replacement of the filtering elements of the TSM
filter is performed as follows:
- open the cover 9;
- unscrew the nut 8 from the body 1 using a screwdriver (included in SET-O);
- take off spring 7, washer 6, filtering element 5 and gauze screen 3;
- remove carbon filler 4;
- take off screen 3, filter 11 and filtering tissue 2.
- replace filters 2, 5 and 11 by new ones from the SET-O;
- put filters 2, 11 and screen 3 into the body 1;
- using the measuring jar from the SET-O fill up the body 1 with a carbon from the SETO and tamp it by light tips;
- fill up the body 1 hole with carbon 4 (included in SET-O);
- put screen 3, filter 5, washer 6, spring 7;
- screw in nut 8 until spring 7 is compressed;
- close the filter cover 9.
18
1- filter body, 2 – filter tissue, 3 and 3’ – gauze screen, 4 – carbon filler (included in SET-O), 5 –
filtering element, 6 – washer, 7 – spring, 8 – nut, 9 – cover, 10 – washer, 11 - filter
Fig. 4
4.6 Device deactivation
4.6.1 The maintenance involves the external examination, dust and dirt removal and
decontamination in case when radioactive substances are fallen on the device housing.
Decontamination is carried out by wiping the device housing with a soft tissue moistens with
ethyl alcohol (State Standard GOST 18300-87).
4.7 Replacement of the batteries
To replace the batteries:
- unscrew and remove the housing with the batteries;
- remove the batteries from the batteries compartment;
- place the new batteries observing the polarity;
- place back the housing and fix it using the screw and the screw driver.
4.8 Measurement of the output rate of the micropump
To measure the output rate of the micropump it is necessary:
- to put rotameter upright so that the input connecting pipe is up;
- to take away the filter and to connect the input pipe of the rotameter to the input
connecting pipe of the TSM 2, Fig.1;
- to switch on the instrument. The blast passing through the rotameter lifts the float of the
rotameter. Determine the output rate of the rotameter by using the scale. The air consumption
should be within 400-600 ml/min.
4.9 TSM efficiency control by using imitator
TSM efficiency control by using imitator is performed as follows:
- switch the instrument ON. 120 s after the switching bring a container with the imitator to
the input connecting pipe (at 5-10 mm distance), the cap of the container should be preliminary
opened;
- not more than in 10-15 s both audible and visual alarms of the TSM channel of the
device should be activated;
- put the TSM filter with the opened cover on the input connecting pipe of the device;
- blow off the TSM channel within 10 min, switch OFF the device and close the filter cover.
N o t e - It is prohibited to touch the input connected pipe by the container with the
imitator as well as ingress of imitator drops inside the ionizing chamber.
19
5 TROUBLESHOOTING
5.1 Possible problems and their solutions are given in Table 3.
Table 3
Problem
1 Device is not turned ON;
Possible cause
the There are no batteries, the
batteries are discharged or
inserted improperly.
when powered by the car Power adapter is installed not
properly.
battery;
when powered
batteries;
by
The cable is broken.
when powered by 220V No supply-line voltage.
AC
Faultiness of the power supply
unit
2 Green-LED is flashing
Batteries are discharged
3 Device does not produce Cap of the ionizing chamber is
“Chemical alarm” when the not removed.
imitator is put to the input Imitator cover is not removed.
TSM module is failed.
connecting pipe of TSM
Contamination of gas systems
and ionizing chamber
4 Red-LED Ph and Red- Contamination of gas systems
LD Arc are flashing in turn and ionizing chamber
Solution
Replace batteries or insert
them properly.
Install power adapter
observing the polarity.
Check the cable and the
polarity of connection to
the battery.
Check the supply-line
voltage.
Check the operation of the
power supply
Replace batteries
Remove the cap.
Remove the cover.
Send device to the
manufacturer.
Clean gas systems and
ionizing chamber
according to clause 4.4
Clean gas systems and
ionizing chamber
according to clause 4.4 if
the flashing continues
more than 5 min.
20
6 STORAGE AND SHIPPING
6.1 Devices are to be stored in the manufacture’s package at the air temperature from 20 to 60 °C and relative humidity up to 95 % at temperature of +35 °C. Storage duration should
not exceed the average service life, i.e. 10 years.
Devices without package are to be stored at the air temperature from 10 to 35°C and
relative humidity of 80 % at temperature of 25 °C.
The storage place should be free of dust, vapors of strong chemicals and other
substances that may cause corrosion.
6.2 Devices in package should be fastened in a vehicle. They must be arranged and
fastened in transport so as to ensure their stable position and to avoid shocks to each other and
the walls of a vehicle.
When carried by see, devices in package should be placed in hermetic plastic bags with
silica gel.
When carried by air, the devices in package should be placed in hermetic compartments.
The environmental conditions during transportation should be within the limits as follow:
- air temperature from - 20 to 60 °C;
- relative humidity up to 100% at temperature 35 °C.
21
7 UTILIZATION
7.1 The device contains the beta radiation source type IBIRZN-63-IIa,
7016-056-41805307-2002 and belongs to the radioisotope devices of the group 3. The
utilization of the used source should be performed according to the requirements of regulations,
and safety measures for the radioactive source handling should be observed.
7.2 Precious metal data contented in the device are not quoted as their weight is not
higher than the values stated in the Sate Standard GOST 2.608-78
22
8 WARRANTY
8.1 The manufacturer guarantees this device to meet the requirements of the Design
Documentation provided that the user observes the guidelines of the device operation, shipping
and storage described in the operating manual.
8.2 The warranty period is 18 months since the date of putting into service.
8.3 The warranty storage period is 6 months since the date of the device acceptance by
a quality control department officer.
8.4 Warranty and after-warranty repairs are carried out by the manufacturer or by a
company authorized by the manufacturer.
8.5 Warranty does not cover devices:
- without an operating manual;
- which were subjected to the user’s service (evidences of the device opening);
- with mechanical damages, if the requirements of operation and storage were not
satisfied;
- which warranty period is expired.
8.6 The warranty period is prolonged for a period of warranty repair.
8.7 Warranty does not cover batteries. The battery replacement is not considered as a
warranty repair.
23
APPENDIX A
COPY OF CERTIFICATE OF TOXIC SUBSTANCE DETECTION MODULE (TSM)
serial No _________, batch No ___________
A.1. ACCEPTANCE CERTIFICATE
A.1.1 The TSM is manufactured to meet the requirements of the design specifications
and is accepted for operation.
A.2 SPECIFICATIONS
A.2.1 The TSM is designed for detection of vapors of organophosphorus substances
(sarin, soman, V-gases) and arsenic-containing substances (lewisite) in air.
A.3 DELIVERY KIT
A.3.1 The TSM delivery kit includes:
Toxic substance detection module
1 piece
Quality certificate (passport) of radioactive source (Appendix B of the 1 piece
present Manual)
SET-O (set of spare parts and tools)
1 set
N o t e - A batch of TSM delivered to one address is supplied with 3 operating manuals
per 10 pieces of TSM.
A.4 SERVICE LIFE, WARRATY
A.4.1 The TSM service life is no less than 10 years.
A.4.2 The warranty period is 2 years.
24
Download