CORRELATION BETWEEN LEAKAGE CURRENT AND NOISE Talk by Edmund WIDL (HEPHY Vienna) presentation can be downloaded from: wwwhephy.oeaw.ac.at/u3w/e/ewidl/www/talks/NOISE.ppt INTRODUCTION For detectors measuring absorbed energies it is useful to quantify the noise in terms of Equivalent Noise Charge (ENC) This ENC corresponds to the amount of charge that would have to traverse the detector to yield a signal as big as the actual noise: noise = ε . ENC ε ... calibration constant AMPLIFIER NOISE • main noise source: input transistor at the amplifier • contributions of further stages usually neglected • load capacitance plays an important role due to the integrating nature In a simple approach, the amplifier noise can be described by the sum of: • a constant value (parallel noise) and • a part which scales with the load capacitance C (series noise) ENC APV = ENC APV, parallel + C ⋅ ENC APV, series 250 + 36 pF −1 (peak ) ENC APV [e] = −1 + 400 60 pF (deconvolut ion) ADDITIONAL NOISE SOURCES equivalent network for a single strip & readout: Ileak Rpoly C Rstrip TP ... fraction of the detector leakage current seen by one strip ... polysilicon resistor ... strip capacitance ... line resistance of one strip ... peaking time numerical noise equations: ENCleak [e] = 106 ⋅ Ileak [nA] ⋅ TP [µs] TP [µs] ENCpoly [e] = 758 ⋅ Rpoly [MΩ] R strip [Ω] ENCRS [e] = 0.395 ⋅ C ⋅ TP [µs] The total noise figure is the square sum of the individual contributions, since the individual sources are uncorrelated: ENC2 = ∑ ENCi2 CORRELATION FOR OB2-MODULES OB2-sensors: • Rpoly = 1.5 MΩ • C = Cint + Cback w strip = 0 .8 + 1 .6 pF cm−1 pstrip ⇒ C = 11.5 pF • Rstrip: wstrip=1.15 ⋅ wimplant= 52.6 µm dstrip≈1.5µm Lstrip=9.44 cm ⇒ R strip = ρ Al L strip = 31.7 Ω w strip ⋅ dstrip OB2-modules: • 2 electrically daisy-chained sensors • Rpoly of each sensor in series • C of each sensor parallel • APV: TP = 0.05 µs This yields: 1066 (peak ) ENC APV [e] = 1759 (deconv.) ENCpoly [e] = 138 ENCRS [e] = 319 ⇒ (561.8 ⋅ Ileak [nA ] + 1 254 949 )1 2 (peak ) ENC(Ileak ) [e] = 12 ( ) ⋅ + 561 . 8 I [ nA ] 3 214 805 (deconv.) leak RESULTS In order to correlate the noise from different sensors the relative noise of each sensor is used: noise (Ileak ) ENC (Ileak ) = noise (Ileak = 0 ) ENC (Ileak = 0 ) left side of the equation: data from 57 OB2 modules • Ileak for each strip from the tracker database • noise(Ileak= 0) is taken as the noise from daisychained strips where the sum of Ileak is less than 2 nA (minimal error – less than 0.5‰) right side of the equation: theoretical relative noise following strips were not included: • strips #1 & #128 of each APV (are known to have higher noise) • strips with an Ileak-entry in the database higher than the total detector current at 400V • unbonded strips 28623 strips remaining leakage current distribution: (in bin 0-10 nA) leakage current vs. strip noise (peak mode): theoretical correlation leakage current vs. strip noise (peak mode): theoretical correlation leakage current vs. strip noise (deconv.): theoretical correlation leakage current vs. strip noise (deconv.): theoretical correlation MODULES WITH CM-PROBLEMS Several modules showed drastically increased noise on single APVs, obviously caused by single, extreme leaky strips. None of these strips showed an abnormal behavior prior to assembly: