• Ohm’s Law ce ric d. co m ICD Electronics – Building Blocks • ICD Shocking / Pacing Outputs .p a • Sensing w w w • Battery Technology 1 ICD Electronics – April 2007 - wks ce ric d. co m Ohm’s Law Voltage = Current x Resistance • V (Volts) (V lt ) = I (A (Amps)) x R (ohms) ( h ) .p a Voltage is the “pressure” pressure that pushes the current through the circuit. It is also called electromotive force (emf). w Current is the flow of charge through the circuit. w w Resistance R i t i a measure off the is th amountt off currentt you will ill get based on a certain Voltage (pressure). Resistance is basically the availability of charge carriers – the fewer charge carriers, the higher the resistance 2 ICD Electronics – April 2007 - wks ce ric d. co m Ohm’s Law 1 Amp = 1,000 milliAmps = 1,000,000 microAmps • 1 milliAmp = 1,000 microAmps .p a Sensing currents are typically 5 – 20 microamps w w Pacing currents are typically 1 – 20 milliamps w Defibrillation currents are typically 1 – 20 Amps 3 ICD Electronics – April 2007 - wks ce ric d. co m Power and Energy Power = Voltage x Current • P(watts) = V(Volts) x I(Amps) .p a Energy = Power x Time (sec) w w E(watt-sec) E(watt sec) = V x I x Time w E (joules) = V x I x Time 4 ICD Electronics – April 2007 - wks ce ric d. co m Power and Energy • .p a Energy is the ability to do work. 1 Joule = 1 Volt x 1 Amp x 1 sec 1 Joule = 1 Watt x 1 sec w w w If the ball has a mass of 1 kilogram (2 (2.2 2 pounds) and it was dropped 1 meter, its kinetic energy gy would be approximately 10 joules 5 ICD Electronics – April 2007 - wks ce ric d. co m .p a w w w 6 ICD Electronics – April 2007 - wks ce ric d. co m .p a w w w 7 ICD Electronics – April 2007 - wks ce ric d. co m .p a w w w 8 ICD Electronics – April 2007 - wks ce ric d. co m Ohm’s Law – Pacing During pacing, the current flow will be between the cathodal tip and the anode ring or shocking electrode. Typical impedance is 400 - 1200 ohms w .p a • w w + 9 ICD Electronics – April 2007 - wks ce ric d. co m Ohm’s Law – Defibrillation + .p a • w + w w - 10 IIn an active ti can ICD system, t current flow will be between the ICD metal can and the shocking electrode(s) in the heart. Typical impedance is 20 – 80 ohms When the shock is delivered delivered, sufficient current density much be achieved to depolarize a critical iti l mass off cardiac di cells. ll ICD Electronics – April 2007 - wks ce ric d. co m Ohm’s Law – Series Circuit Pacing Switch • R - Lead Pacing Capacitor R @ Lead Tip / Tissue Interface w w w Current Flow .p a + Series Circuit – when the pacing switch closes, current flows from the pacing capacitor through the resistance of the lead and through the resistance at the tip of the lead. All of the current goes through the same path, so the resistances ADD. 11 ICD Electronics – April 2007 - wks ce ric d. co m Ohm’s Law – Series Circuit Pacing Switch • R - Lead Pacing Capacitor R @ Lead Tip / Tissue Interface w w w Current Flow .p a + NUMBERS – If the resistance in the lead is 50 ohms and the resistance at the tip / tissue interface is 550 ohms, the series resistance is the summation of the resistances or 550 + 50 = 600 ohms. Since the current must go through each resistance in series, the current or flow “sees” the total of all the individual resistances. 12 ICD Electronics – April 2007 - wks ce ric d. co m Ohm’s Law – Parallel Circuit Pacing Switch • RLead Pacing Capacitor w w w .p a + Current Flow 13 RLead R @ Lead Tip / Tissue Interface R@ Lead Tip / Tissue Interfac e Parallel Circuit – when the pacing switch closes, l currentt flows fl from the pacing capacitor through the resistance of the first lead and through the resistance at the tip of the lead lead. Also Also, current goes through the second lead that is in parallel to the first lead. This is an example of BiV pacing the RV and LV with a tied output device. ICD Electronics – April 2007 - wks ce ric d. co m Ohm’s Law – Parallel Circuit Pacing Switch • RLead Pacing Capacitor w w w Current Flow .p a + 14 R @ Lead Tip / Tissue Interface NUMBERS – If the resistance in the first lead is 50 ohms and the resistance of its tip / tissue interface is 550 ohms, the series resistance is the summation of the resistances or 550 + 50 = 600 ohms. If the resistance of the second lead is 100 ohms and the resistance of the tip / tissue interface is 900 ohms, the series resistance is the summation of the resistances or 100 + 900 = 1000 ohms. So we now have 600 ohms in parallel with 1000 ohms ICD Electronics – April 2007 - wks NUMBERS – So we now have 600 ohms in parallel with 1000 ohms. The current will travel down both the 600 ohm path and down the parallel path of 1000 ohms ohms. The total resistance would then be the parallel path of 1000 ohms and 600 ohms. Mathematically, the resulting total resistance of both paths in parallel will always be LOWER than either path single path path. Sort of like pouring water down two long tubes in parallel to each other – you get more total water flow (the resistance to water flow has gone down). .p a • ce ric d. co m Ohm’s Law – Parallel Circuit w w Here is the math to calculate the total resistance when you have two parallel paths p p ... w Total Resistance = 600 ohms x 1000 ohms = 15 375 ohms 600 ohms + 1000 ohms ICD Electronics – April 2007 - wks ce ric d. co m Shocking Outputs Programmable Stored Energy - The energy stored in the .p a output capacitors is programmable. programmable The pulsewidth of the output may also be programmable. Typically 10 0 –15% 5% o of tthe e sto stored ed e energy e gy iss lost ost due to ccircuitry cu t y in the ICD and losses in the waveform. Programmable Delivered Energy - The energy that will be w w w delivered to the shocking lead system is programmable and the AICD stores sufficient energy to ensure the delivered output. 16 ICD Electronics – April 2007 - wks ce ric d. co m Shocking Outputs .p a During ICD charging, the battery generates low Voltage g and high g current which is transformed to a high Voltage and a low current and this current then charges up the capacitors it tto specific ifi voltage lt llevels l depending on programmed energy w w w During energy delivery, current flows from the output capacitors similar to how water empties from a column. column Initially, Initially the current flow is high, but it decreases as the voltage decreases – the capacitor empties 17 ICD Electronics – April 2007 - wks ce ric d. co m Shocking Outputs Once a specific Voltage is reached, energy delivery d l is stopped Monophasic pulse w w w .p a Voltage and current start high and then decrease as energy in the capacitors flows out 18 ICD Electronics – April 2007 - wks Typically, biphasic DFTs are 30 – 40% lower than monophasic DFTs w w w .p a ce ric d. co m Shocking Outputs 19 Biphasic pulse – Polarity of waveform is reversed during the shock delivery ICD Electronics – April 2007 - wks ce ric d. co m Sensing and Detection • w .p a Amplitude is Typically > 5 mV i the in th ventricle ti l w Slew Rate = Amplitude / Time = dV/dT w NOTE: “Far-field potentials arise from electrical activity distant from the electrode and include contralateral ventricular activation, skeletal muscle potentials, and external electromagnetic interference (EMI). “ 20 ICD Electronics – April 2007 - wks • ce ric d. co m Sensing and Detection Pacemakers P k typically i ll use programmable bl fixed fi d sensitivity ii i circuits for detecting intrinsic activity. w w w .p a 2.5 mV 21 Blanking Programmable Refractory ICD Electronics – April 2007 - wks • ce ric d. co m Sensing and Detection ICDs use auto-adjusting sensitivities or auto-adjusting amplification for detecting intrinsic activity. activity w w w .p a 2.5 mV 22 Blanking Refractory ICD Electronics – April 2007 - wks NSR Lead Signal w .p a AICD Filtered Signal ce ric d. co m Sensing and Detection w w R f Refractory t 23 ICD Electronics – April 2007 - wks NSR Lead Signal 2 mV R-wave and 4 mV T-wave w .p a AICD Filtered Signal ce ric d. co m Sensing and Detection w w Refractory 24 ICD Electronics – April 2007 - wks ce ric d. co m Sensing and Detection .p a Rate Sensing Electrogram Showing Sensing Post-Pacing Rate Sensing EGM w w w Markers 25 ICD Electronics – April 2007 - wks ce ric d. co m Battery Technology Since 1972, lithium batteries have been the primary power source for pacemakers. Today, Lithium-Iodine batteries are used d in i virtually i t ll allll pacemakers. k • w .p a Other chemistries that have been used include Lithium cupric sulfide, Lithium thionyl chloride, Lithium lead and Lithium silver chromate. The type of chemistry determines the battery’s voltage and discharge characteristics. w w • 1958 - Rechargeable R h bl batteries b tt i • 1958 through 1960 - zinc-mercury batteries. Pacemaker longevity of about 2 years. years • Nuclear Batteries - half-life of 87 years. 26 ICD Electronics – April 2007 - wks ce ric d. co m Battery Technology - LiI • Voltage curve .p a BOL = 2.8 V ERT w w w ERT Impedance is yp y 5K to 10K ohms typically 27 Impedance curve ICD Electronics – April 2007 - wks BOL = 3.2 V • ce ric d. co m Battery Technology - Lithium Silver Vanadium Oxide Voltage curve ERT .p a ERT Impedance is typically 0.5 to 0.6 ohms w w Impedance curve w QUESTION: What typically happens to the charge time for a high energy shock as an AICD approaches ERT? 28 ICD Electronics – April 2007 - wks ce ric d. co m Very Highly Accelerated Battery Run Down Highly Accelerated Test Time 2 Years .p a • w w w October 1990 29 ICD Electronics – April 2007 - wks ce ric d. co m Highly Accelerated Test Time - 2 Years Charge Cycles Every TWO months (4 charge cycles per test) 3.5 3 • 2.5 Charge g Time 18 seconds 2 .p a Vo olts / Charge T V-drop 1.5 V(CHARGING) Charge Time / 10 seconds w w 1 V(MONITORING) 0 0 w 0.5 200 400 600 800 1000 1200 1400 1600 1800 Battery Capacity 30 ICD Electronics – April 2007 - wks ce ric d. co m Accelerated Test Time - 3 Years Charge Cycles Every THREE months (4 charge cycles per test) 35 3.5 3 • 2 1.5 .p a Vo olts / Charge tim e 2.5 3 times V (MON) per year V(CHRG) per year Charge T w 1 Charge g Time 18 seconds 0 50 w w 05 0.5 200 00 350 520 5 0 675 6 5 825 8 5 950 1100 00 1250 50 1375 3 5 1525 5 5 Battery Capacity 31 ICD Electronics – April 2007 - wks ce ric d. co m “Non-Accelerated” Test Time - 5 Years Charge Cycles Every SIX months (4 charge cycles per test) 3.5 • 3 V-drop 2 .p a Vo olts / Charge T 2.5 Charge Time 30 seconds 1.5 V(CHRG) Charge Time / 10 seconds w w 1 V(MONITORING) 0 0 w 0.5 200 400 600 800 1000 1200 1400 1600 1800 Battery Capacity 32 ICD Electronics – April 2007 - wks