Implementing 21 CFR Part 11 in Analytical Laboratories Part 5, The

advertisement
Regulatory Issues
Wolfgang Winter and
Ludwig Huber
T
he first four parts of this series gave
an overview of the requirements of
the FDA rule (21 CFR Part 11) on
electronic signatures and records (1).
We focused on data security, data integrity,
long-term archiving, and ready retrieval of
data (2–5). We demonstrated how access to
the system and critical functions could be
limited to authorized personnel. We also
demonstrated how the integrity of data can
be assured at the time of data analysis and
evaluation and how creation, modification,
and deletion of records are logged in a
computer-generated audit trail. And we
Implementing 21 CFR Part 11
in Analytical Laboratories
Part 5, The Importance of
Instrument Control and Data Acquisition
The time for compliance with
21 CFR Part 11 is now. Bringing
different laboratory instruments
into compliance takes planning.
The key strengths and
weaknesses of different levels of
control and feedback for
analytical instruments and data
transfer systems are highlighted in
this fifth installment of our series.
Wolfgang Winter is worldwide product
manager for networked data systems and
corresponding author Ludwig Huber is worldwide
product marketing manager, HPLC, at Agilent
Technologies GmbH, PO Box 1280 D-76337,
Waldbronn, Germany, +49 7243 602 209, fax
+497243 602 501, ludwig_huber@agilent.com,
www.agilent.com.
showed the best method for archiving data
and accurately retrieving it after several
years.
In those first four articles, we focused on
compliance of data generated by the system.
Frequently, the question comes up whether
computers that just control analytical
instruments — those that do not acquire data
— must comply with Part 11. The answer is
simply, “Yes, if FDA has ever looked at or
asked for paper printouts of the parameters.”
Without proper documentation of the
instrument control parameters, it is difficult
to prove that a given result was generated
according to the appropriate procedure or
protocol. If a computer was used in the
procedure, and if the control parameters are
stored on a durable storage device (typically
the computer’s hard disk or a storage card for
the instrument itself), then Part 11 applies.
Levels of Instrument Control
Analytical laboratories typically operate
with a diverse base of instruments, often
from a variety of manufacturers for
historical or strategic reasons. Because most
modern instruments are computer
controlled, the instrument control
parameters have to be treated as the rest of
the metadata are treated. (See the “Learning
21 CFR Part 11 Language” box for words
in italics.)
Level 1. Instrument control can be
implemented at differing levels of
sophistication and complexity (Table 1).
Often, instrument parameters are set
manually using the instrument’s own panel
and keyboard, with the signal recorded by an
analog–digital converter (level 1). This is
frequently the approach chosen to integrate
an instrument into a system from a different
manufacturer. In such cases, it is often
impossible to obtain a printout of the
instrument set points used during an
analysis. Analysts are forced to document
instrument parameters manually.
Furthermore, analog–digital converters do
not always support binary coded decimal
(BCD) or bar code input from an
autosampler, which could be used to
positively correlate an injection with a given
sample using the sample name or vial
number. In agreement with other authors, we
view BCD communication with an
autosampler as essential to ensuring sample
continuity (6).
Level 2. Many systems implement a
rudimentary level of instrument control
obtained through reverse engineering:
recreating the design of the communication
protocols for another manufacturer’s
instrument by analyzing the final product
(level 2). This method supports the basic
parameters of an instrument (such as solvent
composition, flow, oven temperature, or
detector wavelength). If the control codes are
not officially disclosed by the particular
instrument manufacturer, it may be more
difficult to obtain an officially supported
solution. Also, additional effort should be
planned for performing qualification and
other validation tasks on such a system.
Because the manufacturer of the original
instrument may neither be aware of nor
responsible for the implementation of the
communication, instrument firmware
updates may result in nonfunctional
communication with the data system. Error
handling and logging are typically weak at
this level. When selecting a system that will
control instruments from other
manufacturers, it is therefore important to
verify that the control codes are officially
obtained from the manufacturer of the
instrument and not from reverse-engineering.
Regulatory Issues
Level 3. In most cases, manufacturers
achieve full instrument control for their own
systems (level 3). That makes it easier to
create a complete set of raw and metadata
and the proper documentation. At this level,
the error reporting and handling are quite
sophisticated, which makes it easier to
verify that analyses were completed without
technical failures and to diagnose errors
when they occur.
Going one better. Some manufacturers have
implemented an additional level of
instrument capability that can be controlled
from within the data system. A data system
controls those functions that execute
detailed and sophisticated instrument
diagnostics and other service functions. Also
under this instrument control are provisions
for preventive maintenance and early
maintenance feedback (EMF), a technique
first used in the aeronautics industry (to alert
technical personnel to perform maintenance
jobs proactively) and subsequently
implemented by companies such as Agilent
Technologies (Palo Alto, CA).
Systems implemented at this level
provide sophisticated support for tracking
instrument or module serial numbers and for
firmware revisions. Such information is
handy for inventory tracking, and it also
performs some of the function checks
required by Part 11.
Level 4. In level 4 instrument control, all
communications (including commands and
data transfer) are performed using a
handshake. A handshake requires the
recipient of a data record to actively
acknowledge to the sender that the record
has been received. For example, the
controller sends a command like “START” to
the device, the device interprets the
command and acknowledges “OK, START.” If
the device is unable to execute the
command, it sends a negative receipt like
“NOT OK, NO START.” This approach prevents
situations in which the controller thinks it
sent the instructions to the device, but the
device never executes them.
Protocols for Data Integrity
Understanding the strengths and weaknesses
of some widely used instrument
communication protocols will help ensure the
data integrity and traceability required by 21
CFR Part 11. One example can be drawn from
the legacy world: a general-purpose interface
bus (GPIB), the well known and widely
Table 1. Levels of instrument control, the parameters that make up that level, and the
consequences of that level of control in complying with 21 CFR Part 11
Level
Compliance
with Part 11
Parameters
Level 1. Parameter set up on
Start/Stop (no digital
the instrument, synchroniza- instrument control or
tion using external contacts data acquisition)
to start and stop an
analysis, analog signal
acquisition
Metadata: Instrument
parameters must be
documented manually;
Device checks: Positive
ID of sample vials may
not be available (using
bar codes or BCD input)
Level 2. Rudimentary digital
instrument control (such as
a LAN, RS232, or GPIB)
Audit trail: Typically no
instrument error
information available,
requires additional
inspections to determine
the validity of the
measurements;
Validation: Could be more
difficult to support and
validate if reverse
engineered
Basic instrument parameters,
such as the flow rate of
an HPLC pump or the
wavelength of an HPLC
detector
Level 3. Full digital instrument All control parameters
control (for example through including injector program
a LAN, RS232, or GPIB)
and method sequencing;
wavelength calibration;
error recording
Audit trail and Metadata:
Full documentation of
instrument parameters
used to generate a result
Level 4. Advanced functions
Advanced error
prevention and
detection; Validation:
facilitates the execution
of instrument qualification
and preventive
maintenance; qualifies for
device checks required
by the rule; guaranteed
and reproducible
execution of data
acquisition independent
of the current data
system load (facilitates
the qualification of data
integrity and traceability)
Handshake protocol between
controller and device (active
acknowledgment of correct
receipt); self-diagnostics and
early maintenance feedback
(EMF); automatic tracking of
serial and product numbers,
electronic instrument log book;
supports advanced tagging of
components, such as columnid tags; instrument performs
real-time data acquisition and
synchronization independent
of the computer
spread “IEEE 488” standard.
Contrast the GPIB to state-of-the-art
networking protocols like the well known
and ubiquitous TCP/IP protocol used for
intra- or Internet communication. (Tables 2
and 3 provide a detailed list of the strengths
and weaknesses of these two communication
systems and recommendations for avoiding
some weaknesses.) We are not providing a
detailed technical description of the
technology itself. Many publications cover
those aspects accurately and in technical
detail (7–9).
Instrument communication using GPIB. GPIB is a
parallel communication interface that can
connect up to 15 devices on a common bus.
All communications using GPIB, including
commands and data, use a hardware
handshake for every byte. All devices
connected to the bus participate in that
handshake. As a consequence, every device
on the bus can influence the ongoing
communication or cause severe
communication problems like bus “hangups” or data corruption. The reason for that
can be a firmware error or a hardware failure
in one of the participating devices (such as
the printer). But powering a seemingly
“idle” GPIB device on or off during ongoing
communication also can cause such
problems. Even though the electrical
specifications of GPIB do not prohibit the
actions that lead to those scenarios, the
combination of chip-set implementation,
firmware, and application software often
leads to that failure.
Regulatory Issues
Table 2. Strengths and weaknesses of instrument communication using GPIB, and
recommendations for meeting 21 CFR Part 11 with this level of instrument control
Strength
Weakness
Recommendation
Fast enough for highspeed data acquisition
(for example spectral
data from diode array or
mass-selective detectors)
Limited number of
devices on the same
bus (15 maximum)
Not applicable
GPIB allows sending of
bidirectional, addressed
information (allows for the
implementation of device
checks according to
21 CFR Part 11)
The standard does not
define power on/off on
“idle” devices during ongoing
communication. Because all
devices must participate
electrically in the handshake,
powering a device off during
communication may hang the
entire system, which can lead
to data loss.
Avoid powering “idle” GPIB
devices down while other
devices are acquiring data
to minimize the risk of
data loss or data corruption.
Not applicable
Monitoring correct instrument
functions from remote devices
requires a computer
(instrument controller) that
functions as an “information
broker” next to the instrument.
Requires separate
operational qualification of
the instrument controller
unaffected by the addition or removal of
“idle” devices in the network. In contrast to
most GPIB implementations, TCP/IP
supports (without risking data loss) the
safety procedures of analytical laboratories
that require turning off instruments not
currently in use.
Selecting a Good System
Table 3. Strengths and weaknesses of instrument communication using TCP/IP, and
recommendations for meeting 21 CFR Part 11 with this level of instrument control
Strength
Weakness
Recommendation
Suited for high-speed data
acquisition (for example
spectral data from
diode array or massselective detectors)
Not applicable
Proactively plan network
infrastructure, especially in
large laboratories with many
instruments
Inherent mechanisms for
Not applicable
error detection and correction
(an important measure for data
integrity and the basis for
device checks according to
21 CFR Part 11)
Reliable instrument communication
should be verified during system
qualification; turn off “idle”
instruments not currently in use
Remote status information is
available directly from the
instrument and does not
necessarily require an extra
instrument controller as
“information broker”
Covered by system qualification
Not applicable
LAN Communication Using TCP/IP
Local area network (LAN) communication
using the transmission control protocol/
Internet protocol (TCP/IP), often somewhat
casually described as the “language of the
Internet,” enables devices to exchange
information over a network. The central
idea of TCP/IP is the breaking of
information into pieces or packets. The
packets are specifically structured to allow
error detection and correction by using
redundancy mechanisms like checksums.
Redundancy mechanisms are the major
difference between more advanced systems
and the majority of GPIB implementations.
Checksums are, in principle, a running total
of all transmitted bytes attached to the packet
and are used by the recipient to back-calculate
and compare with the original checksum
provided by the sender. If a mismatch is
detected, a retransmission is requested. This
technique guarantees error-free data transport
and is excellent for implementing the “device
checks” and “system checks” mandated by 21
CFR Part 11.
Communication in a TCP/IP environment
is, by definition, highly dynamic. Ongoing
communication between other participants is
When selecting and setting up instrument
control and data acquisition systems, the
following recommendations should help.
1. Assess the level of instrument used
currently in your laboratory (level 1, 2, 3,
or 4).
2. Find out what levels of control are
available from the different manufacturers
of that hardware.
3. Write a protocol documenting the
instrument parameters for instruments not
directly controlled by the data system
(level 1).
4. Determine whether the communication
protocols were obtained with the approval
and support of the instrument manufacturer
for instruments that claim to be controlled.
5. Plan additional qualification and
acceptance tests to obtain a high degree
of certainty that the control and
communications are accurate and reliable
for instruments for which communication
protocols were apparently developed by
reverse engineering.
6. Adapt your company’s internal
procedures to take advantage of additional
diagnostics, maintenance, and tracking
functions that validate, maintain, and
document the system and the measurements
obtained from those functions.
7. Define test cases for boundary
conditions: Does the system reliably
synchronize all the devices required for an
analysis? Could a contact closure problem
allow it to go unnoticed that a device, such
as a detector, did not start? If the instrument
has a local user interface, does it track
parameter changes from the local interface
while data are being acquired from the
computer? Alternatively, is the local
interface “locked” while data is acquired?
Does the system quickly detect power
failures of a connected device, or are data
lost until a “time out” occurs? These are the
kinds of questions to ask when bringing
your laboratory instrument control and data
acquisition systems into compliance with 21
CFR Part 11.
Regulatory Issues
Learning 21 CFR Part 11 Language
Bus is a collection of wires through which
data travel within a computer. In this
context, bus means an interface and
communication system for peripheral
devices (such as connections, cables,
and the communication protocol).
Byte is an abbreviation for binary term. It
is a storage unit capable of holding eight
bits or the space required for a single
letter or number, a single character.
peripheral. Most devices, whether
peripheral or not, require a program
called a device driver that acts as a
translator, converting general
commands from an application into
specific commands that the device
understands.
Durable storage device is typically the
computer’s hard disk or a storage card
for a particular instrument.
disruptive and therefore requires careful
planning and appropriate migration
support from the manufacturer.
Local-area networks (LANs) are networks
with computers geographically close
together (that is, in the same building),
and wide-area networks (WANs) have
computers farther apart and connected
by telephone lines or radio waves.
Metadata is complete data with
Checksums are a running total of all
Early maintenance feedback (EMF) is a
transmitted bytes that are attached to a
packet and are used by the message
recipient to back-calculate and compare
with the original checksum provided by
the message sender. If a mismatch is
detected, a retransmission is requested.
This technique facilitates detection of
data transport errors.
technique that automatically alerts
technical personnel to perform
maintenance jobs proactively.
Networking: A group of two or more
computer systems linked together.
Firmware is a combination of hardware
Packet: A piece of a transmitted message
Chip-sets are a number of integrated
circuits designed to perform one or more
related functions. For instance, the
integrated circuit components of a
specific GPIB interface card.
Device is any machine or component that
attaches to a computer, such as disk
drives, printers, mice, and modems.
Those particular devices fall into the
category of peripheral devices because
they are separate from the main
computer. Display monitors and
keyboards are also devices, but
because they are integral parts of the
computer they are not considered
and software written in read-only
memory.
Handshake requires the recipient of a data
record to actively acknowledge to the
sender that the record has been
received.
IEEE is the Institute of Electrical and
Electronic Engineers that develops
standards for computers and the
electronics industry.
Legacy systems are hardware and software
applications in which a company has
already invested considerable time and
money. Legacy systems typically perform
critical operations in companies for many
years even though they may no longer
use state-of-the-art technology.
Replacing legacy systems can be
Looking Ahead
In the next article, we will discuss
biometrics devices for system access and
electronic signatures. In computer security,
biometrics are authentication techniques that
rely on measurable physical characteristics
that can be automatically checked, such as
fingerprints, retinas and irises, voice
patterns, facial patterns, and hand
measurements for system access and
electronic signature.
References
(1) Office of Regulatory Compliance, Code of
Federal Regulations, Food and Drugs:
(2)
(3)
(4)
(5)
Electronic Records; Electronic Signatures, Title
21, Part 11 (U.S. Government Printing Office,
Washington, DC), issued March 2000. Available
at www.fda.gov/ora/compliance_ref/ part11.
L. Huber, “Implementing 21 CFR Part 11 in
Analytical Laboratories: Part 1, Overview and
Requirements,” BioPharm 12(11), 28–34
(1999).
W. Winter and L. Huber, “Implementing 21
CFR Part 11 in Analytical Laboratories: Part 2,
Security Aspects for Systems and
Applications,” BioPharm 13(1), 44–50 (2000).
W. Winter and L Huber, “Implementing 21
CFR Part 11 in Analytical Laboratories: Part 3,
Ensuring Data Integrity in Electronic
Records,” BioPharm 13(3), 45–49 (2000).
L. Huber and W. Winter, “Implementing 21 CFR
Part 11 in Analytical Laboratories: Part 4, Data
©Reprinted from BIO PHARM, September 2000
AN
ADVANSTAR ★ PUBLICATION
processing parameters and audit trail
logs.
that contains both the data and the
destination address. In TCP/IP
networking, packets are called
datagrams. When you send an email
message, the message can be broken
into several packets, each packet can
be transmitted separately, each packet
may travel different routes, and all the
packets can be put back together at the
recipient’s site.
Reverse engineering: Recreating the design
of hardware or software by analyzing the
final product and working backward.
TCP/IP: Transmission control
protocol/Internet protocol, enables
devices to exchange information over a
network.
(6)
(7)
(8)
(9)
Migration and Long-Term Archiving for Ready
Retrieval,” BioPharm 13(6), 58–64 (2000).
R. D. McDowall, “Chromatography Data
Systems: Part 1, The Fundamentals,” LCGC
North America 18(1), 56–67 (2000).
W. Winter, “Dynamic Interprocess
Communication between a Spectrophotometer
and a Spreadsheet,” diploma thesis and
presentation for faculty for physical electronics,
University of Karlsruhe (31 July 1989).
M. F. Arnett et al., “Understanding Basic
Network Concepts,” Inside TCP/IP (New
Riders Publishing, Indianapolis, 1994)
pp. 51–54.
ANSI/IEEE Std. 488.1-1987: Standard Digital
Interface for Programmable Instrumentation
(The Institute for Electrical and Electronics
Engineers, New York, 1987). BP
Printed in U.S.A.
Copyright Notice Copyright by Advanstar Communications Inc. Advanstar Communications Inc. retains all rights to this article. This article may only be viewed or printed (1) for personal use. User may not actively
save any text or graphics/photos to local hard drives or duplicate this article in whole or in part, in any medium. Advanstar Communications Inc. home page is located at http://www.advanstar.com.
Agilent Publication Number
5988-0946EN