BIOSENSORS Applications and Detection Created by the faculty at Morehouse College © 2006-7 ACKNOWLEDGMENTS The authors would like to thank the following people for their advice and support in the development of this module: 2 FOREWORD 3 TABLE OF CONTENTS Acknowledgments Foreword Introduction What is a Biosensor? 2 Types of Biosensors 3 Brief History of Biosensors Trends of the Past 4 Decades 6 7 Scientific Principles The Components of a Biosensor Manufacturing of Biosensors “Not-So-Common” Applications of Biosensors References Resources Teacher Demonstrations Demonstration Materials & Equipment Grid Demonstration #1 Chemiluminescence (Luminol Reaction) Demonstration #2 Testing Know Concentrations of Peroxide Lab Activities Laboratory Materials Equipment Grid Was it really a poppy seed bagel? Is it a boy or a girl? Low carbs or no carbs? Bioluminescence Experiment #5 (TBD) Experiment #6 (TBD) Design Project Build-Your-Own Electrochemical Glucose Monitor Biosensor Quiz Assessments Glossary 4 INTRODUCTION Module Objective: The objective of this module is to identify and characterize the fundamental properties of biosensors and to recognize the impact that biosensors have on human societies today. Key Concepts Covered: Types of sensors and their associated technologies The history of biosensors Basic understanding of signal transduction The fundamental elements of biosensor devices and their design Technical definitions including, but not limited to: calibration, selectivity, sensitivity, reproducibility, detection limits, response time, immobilization Prerequisites: A familiarity with the following concepts would be helpful in understanding the information in contained within this module. Fundamental knowledge of organic chemistry An understanding of interparticle forces (e.g. van der Waals) The nature of the chemical bond (both ionic and covalent bonds) Basic laboratory technique / laboratory safety precautions Placement in Curriculum: This module, as presented, could be used at the college level in either an introductory or advanced chemistry course, as well as any general physical science course. 5 WHAT IS A BIOSENSOR? Put simply, a biosensor is an analytical tool. It consists of a biologically active material used in close conjunction with a device that will convert a biochemical signal into a quantifiable electrical signal. Compared to other types of sensors, biosensors have many advantages. For example, biosensors are hailed for both simple and low-cost instrumentation, fast response times, minimum sample pretreatment, and high sample throughput. Despite these inherent advantages, relatively few biosensors have been commercialized, though market demands are bringing biosensors closer to both field testing and commercialization in the United States, Europe, and Japan. To maintain the current push towards broader implementation of biosensors devices in the marketplace, research in this area continues to demand the development of both new and novel materials, new and better analytical techniques, and new and improved biosensors.7-11 In general, a biosensor has two key components: a receptor and a detector. The receptor is responsible for the selectivity of the sensor. Examples of common receptors include enzymes, antibodies, nucleic acids, and lipids. The detector, which plays the role of the transducer, translates the physical or chemical change occurring at the receptor by recognizing the analyte and relaying it through an electrical signal. As a rule, then, the detector component of a Potential Applications of Biosensors: 1-6 Agricultural, horticultural and veterinary analysis Pollution, water, and microbial contamination analysis Clinical diagnosis and biomedical applications Fermentation analysis and control Industrial gases and liquids Mining and toxic gases Explosives and military arena Flavors, essences and pheromones 6 Figure 1. A biosensor configuration that allows measurement of a target analyte without using reagents. biosensor is not selective. Examples transducer elements range from pH and oxygen electrodes to piezoelectric crystals. Figure 1 describes a typical biosensor configuration that allows measurement of a target analyte without using reagents. The device incorporates a biologicalsensing element with a traditional transducer. The biological-sensing element selectively recognizes a particular biological molecule through a reaction, specific adsorption, or other physical or chemical process, and the transducer converts the result of this recognition into a usable signal, which can be quantified. Common transduction systems are optical, electro-optical, or electrochemical; this variety offers many opportunities to tailor biosensors for specific applications.1-6 For example, the glucose concentration in a blood sample can be measured directly by a biosensor (which is made specifically for glucose measurement) by simply dipping the sensor into the sample. 7 TYPES OF BIOSENSORS In this section we will discuss the various types of possible biosensors. We will analyze the working mechanism of each one of them. Resonant Biosensors In this type of biosensors an acoustic wave transducer is coupled with antibody (bioelement). When the analyte molecules (antigen) get attached to the membrane, the membrane mass changes, resulting in a subsequent change in the resonant frequency of the transducer. This frequency change is measured out [19]. Optical-Detection Biosensors The output transduced signal that is measured is light signal for this type of biosensors. The biosensors can be made based on optical diffraction or electrochemilluminence. In optical diffraction based devices, a silicon wafer is coated with a protein via covalent bonds. The wafer is exposed to UV light through a photomask and the antibodies made inactivated in the exposed regions. The diced wafer chips when incubated in analyte antigen-antibody binding is formed in active regions, thus creating diffraction grating. This grating produces diffraction signal when illuminated with a light source such as laser. This signal can be measured or can be further amplified before measuring for improving sensitivity [19]. 8 Thermal-Detection Biosensors This type of biosensors are constructed combining enzymes with temperature sensors. When the analyte comes in contact with the enzyme, the heat reaction of the enzyme is measured and is calibrated against the analyte concentration [19]. Ion-Sensitive Biosensors These are basically semiconductor FETs having ion-sensitive surface [7, 19]. The surface electrical potential changes when the ions and the semiconductor interact. This potential change can be measured. The Ion Sensitive Field Effect Transistor (ISFET) can be constructed by covering the sensor electrode with a polymer layer. This polymer layer is selectively permeable to analyte ions. The ions diffuse through the polymer layer and in turn cause a change in the FET surface potential. Fig. 11 shows an ISFET having enzyme enzyme layer Figure 10: Specificity of Enzymes, Source : [5] placed on it; also called ENFET (Enzyme Field Effect Transistor) [7]. This type of biosensor are primarily used for pH detection. Electrochemical Biosensors Electrochemical biosensors are mainly used for detection of hybridised DNA, DNAbinding drugs, glucose concentration, etc. The underlying principle for this class of biosensors is that many chemical reactions produce or consume ions or electrons which in turn cause some change in the electrical properties of the solution which can be sensed out and used as measuring parameter [1, 19]. The electrochemical biosensors can be classified based on the measuring electrical parameters as : (1) conductimetric, (2) amperometric and (3) potentiometric [1]. A comparative discussion of these: 9 Conductimetric Biosensors The measured parameter is the electrical conductance/resistance of the solution. When ectrochemical reactions produce ions or electrons the overall conductivity/resistivity of the solution changes. This change is measured and calibrated to a proper scale. Conductance measurement has relatively low sensitivity. The electric field generated using sinusoidal voltage (AC) which helps in minimizing undesirable effects such as Faradaic process, double layer charging and concentration polarization [1]. Amperometric Biosensors This high sensitivity biosensor cab detect electroactive species present in biological test samples. Since the biological test samples may not be intrinsically electro-active, enzymes needed to catalyze production of radio-active species. In this case, the measured parameter is current [1]. Potentiometric biosensors In this type of sensors the measured parameter is oxidation/reduction potential (of an electrochemical reaction). The working principle of that when a ramp voltage is applied to an electrode in solution the current flow occurs because of electrochemical reaction. The voltage at which these reactions occurs indicate a particular reaction and particular species [1]. Piezoelectric Sensors In this mode, sensing molecules are attached to a piezoelectric surface – a mass to frequency transducer – in which interactions between the analyte and the sensing molecules set up mechanical vibrations that can be translated into an electrical signal proportional to the amount of the analyte (Figure 3). Example of such a sensor is quartz crystal micro or nano balance. 10 Potentiometric biosensors (Continued) Optical Sensors In optical biosensors, the optical fibers allow detection of analytes on the basis of absorption, fluorescence or light scattering. Since they are non-electrical, optical biosensors have the advantages of lending themselves to in vivo applications and allowing multiple analytes to be detected by using different monitoring wavelengths. The versatility of fiber optics probes is due to their capacity to transmit signals that reports on changes in wavelength, wave propagation, time, intensity, distribution of the spectrum, or polarity of the light. In general, acquisition of the signal from these devices is accomplished through flexible cables, which can transmit light to the biological component. Optrodes use fiber optics for performing optical measurement away from the measuring locations (e.g., intra–arterial determination using FIA systems). A powerful and sensitive analytical methodology has been constructed based on the luciferin/luciferase bioluminescence reaction. Electrochemical Sensors In this configuration, sensing molecules are either coated onto or covalently bonded to a probe surface. A membrane holds the sensing molecules in place, excluding interfering species from the analyte solution. The sensing molecules react specifically with compounds to be detected, sparking an electrical signal proportional to the concentration of the analyte. The bio-molecules may also respond to an entire class of compounds such as opiates and their metabolites. The most common detection method for electrochemical biosensors involves measurement of current, voltage, conductance, capacitance and impedance. 11 A BRIEF HISTORY OF BIOSENSORS His invention of the oxygen electrode, and its subsequent modification with enzymes, allow us to clearly identify Professor Leland C. Clark, Jr. as the father of the biosensor concept. In 1956, Clark published his definitive paper on the oxygen Photo Credit Needed electrode. Based on this experience and addressing Figure 2. Professor Leland C. Clark his desire to expand the range of analytes that could be measured in the body, he made a landmark address in 1962 at a New York Academy of Sciences symposium. In this he described "how to make electrochemical sensors (pH, polarographic, potentiometric or conductometric) more intelligent" by adding "enzyme transducers as membrane enclosed sandwiches." The concept was illustrated by an experiment in which glucose oxidase was entrapped at a Clark oxygen electrode using a dialysis membrane. The decrease in measured oxygen concentration was proportional to glucose concentration. In the published paper, Clark and Lyons coined the term enzyme electrode, which many reviewers have mistakenly attributed to Updike and Hicks, who expanded on the experimental detail necessary to build functional enzyme electrodes for glucose. Figure 3. Taken from Professor Clark’s notebook 12 The biosensor, essentially invented by Clark, was the basis of numerous variations on the basic design and many other (oxidase) enzymes were immobilised by various workers as a result. Indeed, Clark's basic design was so successful that many research biosensors and at least one commercial biosensor are still produced using the original concept of oxygen measurement. However, nowadays the preferred alternative is to measure hydrogen peroxide. Most notable of the commercially-available biosensors today is probably the range of biosensors sold by the Yellow Springs Instrument Company (YSI, Ohio, USA). Their glucose biosensor was successfully launched commercially in 1975 and was based on the amperometric detection of hydrogen peroxide. This was the first of many biosensor-based laboratory analysers to be built by companies around the world. Guilbault and Montalvo were the first to detail a potentiometric enzyme electrode. They described a urea sensor based on urease immobilised at an ammonium-selective liquid membrane electrode. The use of thermal transducers for biosensors was proposed in 1974 and the new devices were christened thermal enzyme probes and enzyme thermistors, respectively. Lubbers and Opitz coined the term optode in 1975 to describe a fibre-optic sensor with immobilised indicator to measure carbon dioxide or oxygen. They extended the concept to make an optical biosensor for alcohol by immobilising alcohol oxidase on the end of a fibre-optic oxygen sensor. Commercial optodes are now showing excellent performance for in vivo measurement of pH, pCO2 and pO2, but enzyme optodes are not yet widely available. The biosensor took a further fresh evolutionary route in 1975, when Divie suggested that bacteria could be harnessed as the biological element in microbial electrodes for the measurement of alcohol. This paper marked the beginning of a major research effort in Japan and elsewhere into biotechnological and environmental applications of biosensors. 13 In 1976, Clemens et al. incorporated an electrochemical glucose biosensor in a bedside artificial pancreas and this was later marketed by Miles (Elkhart) as the Biostator. Although the Biostator is no longer commercially available, a new semi-continuous catheter-based blood glucose analyser has recently been introduced by VIA Medical (San Diego). In the same year, La Roche (Switzerland) introduced the Lactate Analyser LA 640 in which the soluble mediator, hexacyanoferrate, was used to shuttle electrons from lactate dehydrogenase to an electrode. Although this was not a commercial success at the time, it turned out in retrospect to be an important forerunner of a new generation of mediated biosensors and of lactate analysers for sports and clinical applications. A major advance in the in vivo application of glucose biosensors was reported by Shichiri et al. who described the first needle-type enzyme electrode for subcutaneous implantation in 1982. Companies are still pursuing this possibility, but no device for general use is available as yet. The idea of building direct immunosensors by fixing antibodies to a piezoelectric or potentiometric transducer had been explored since the early seventies, but it was a paper by Liedberg et al. that was to pave the way for commercial success. They described the use of surface plasmon resonance to monitor affinity reactions in real time. The BIAcore (Pharmacia, Sweden) launched in 1990 is based on this technology. It was during the 1980s, however, that large-scale commercial success was first achieved. YSI had built up a steady and thriving business, but it was not in the same league as the success which had been predicted and, indeed, widely expected. The basic problem lay largely with the cost of producing the biosensors of the time, which made them uncompetitive with the other technologies widely used in the massive rapid testing sector. It was within this sector that the hopes were pinned, since it was (and still is) a huge market. 14 In 1984, a much cited paper on the use of ferrocene and its derivatives as an immobilised mediator for use with oxidoreductases was published. These were crucial components in the construction of inexpensive enzyme electrodes, and formed the basis for the screen-printed enzyme electrodes launched by MediSense (Cambridge, USA) in 1987 with a pen-sized meter for home blood-glucose monitoring. The electronics have since been redesigned into popular creditcard and computer-mouse style formats, and MediSense's sales showed exponential growth, reaching $175 million per annum by 1996, when they were purchased by Abbott. Boehringer Mannheim (now Roche Diagnostics), Bayer and LifeScan now have competing mediated biosensors and the combined sales of the four companies dominate the world market for biosensors and are rapidly displacing conventional reflectance photometry technology for home diagnostics. Academic journals now contain descriptions of a wide variety of devices exploiting enzymes, nucleic acids, cell receptors, antibodies and intact cells, in combination with electrochemical, optical, piezoelectric and thermometric transducers. Within each permutation lies a myriad of alternative transduction strategies and each approach can be applied to numerous analytical problems in healthcare, food and drink, the process industries, environmental monitoring and defence and security. Some defining events in the history of biosensor development show that the 1980s was a very inventive decade, with commercialization being the theme of the 1990s. There appears to be no sign of the latter theme changing as we enter the new millennium. This demonstrates that, in some areas, biosensors have become a mature technology. It must be remembered, however, that this commercial success and maturity is limited to a small number of applications and that it 15 came as a result of a great deal of research and development and has only really taken place where market size/share justified significant financial investment. 16 TRENDS OVER THE PAST FOUR DECADES (TIMELINE?) 17 SCIENTIFIC PRINCIPLES In very nonspecific terms, a sensor is any device that can be used to measure the presence or absence of an object. For example, a sensor can be used to detect the presence of an object as large as a person walking into a supermarket equipped with automatic doorways. A sensor can also be used to detect the presence of objects as small as molecules. In this way, a carbon monoxide detector can be used to denote the presence, at levels of harm to humans, of the gas, carbon monoxide (Molecular formula CO). For our purposes here, a biosensor is a device that combines a biological component (i.e. enzyme, antibody, DNA) with a physicochemical detector component, the combination of which can be used specifically for the detection of any object, large or small. Despite its narrow definition, the biosensor is becoming increasingly important as a tool in everyday life, but especially in the field of medicine. In fact, the most widespread example of a commercial biosensor is the blood glucose monitor, which uses an enzyme to break blood glucose down. In doing so it transfers an electron to an electrode and this is converted into a measure of blood glucose concentration. The high market demand for such sensors has fueled development of associated sensor technologies. Recently, arrays of many different detector molecules have been applied in so called electronic nose devices, where the pattern of response from the detectors is used to fingerprint a substance. A canary in a cage, as used by miners to warn of gas could be considered a biosensor. Many of today's biosensor applications are similar, in that they use organisms which respond to toxic substances at a much lower level than us to warn us of their presence. Such devices can be used both in environmental monitoring and in water treatment facilities. Below we learn some of the scientific principles involved in the preparation and utilization of biosensors. The Components of a Biosensor A basic biosensor consists of at least three (3) parts: A sensitive biological element A transducer A detector element To gain a greater appreciation/understanding of each of these three components, let’s explore each of them in some detail. Biorecognition Systems Biocatalysis-based biosensors for environmental applications depend universally on the use of enzymes. These enzyme-based biosensors primarily rely on two operational mechanisms. The first mechanism involves the catalytic transformation of a pollutant (typically from a nondetectable form to a detectable form). The second mechanism involves the detection of pollutants that inhibit or mediate the enzyme's activity. For environmental applications, the first mechanism involving catalytic transformation of the pollutant, shows certain advantages and limitations. They are simple in design and operation. Examples of this format include the use of tyrosinase for the detection of phenols11-12 and the use of organophosphate hydrolase for the detection of organophosphorus pesticides.13 These biosensors can be configured to operate continuously and reversibly. They can also be configured such that the only required reagent is the analyte of interest Inherent limitations for this type of biosensor are primarily those imposed by the nature of the enzyme itself and include the limited number of environmental pollutants which happen to be substrates for the enzyme and the relatively high detection limits (compared to those required by many environmental monitoring applications) for environmental pollutants. The detection limits 19 for these sensors are determined by the enzyme's catalytic properties and are defined by Km and Vmax values. Biosensor formats have been devised which substantially reduce these inherent limitations. For example, in the case of the tyrosinase enzyme electrode, catalytic cycling of the enzyme intermediate between the quinone and catachol oxidation states can significantly amplify the biosensor response. This results in lower detection limits for phenols than expected from Km and Vmax values.12 Mechanisms for the tyrosinase biosensors involve the detection of phenols either through the electrochemical reduction of quinone intermediates or through oxygen consumption (O2 is a co-substrate) using a Clark electrode.14 In another example, a biosensor was designed to increase both sensitivity and the range of substrates typically measured using the tyrosinase enzyme electrode. With this biosensor, a wide range of chlorinated phenols are detected as oxidation-reduction mediators in a glucose oxidase electrode system.15 This is accomplished by chemically oxidizing the chlorophenols which then cycle between the quinone and hydroquinone oxidation states. In this case, the quinone acts as an electron shuttle for the glucose oxidase-catalyzed oxidation of glucose. This configuration results in low nM detection limits for a number of chlorinated phenols. The other primary mechanism used in biocatalytic biosensors for environmental applications is inhibition of the enzyme by the pollutant. Inherent advantages for these formats involve the larger number of environmental pollutants, usually of a particular chemical class, that inhibit the enzyme and the low concentrations needed to affect the enzyme activity. Detection limits for biosensors based on irreversible inhibitors are usually within the range required for a variety of environmental applications.16 For example, detection limits for cholinesterase biosensors are reported to be in the µg/l to ng/l range for compounds such as aldicarb, carbaryl, carbofuran, and dichlovos.17 20 There are several inherent limitations for biosensors based on enzyme inhibition. In addition to the analyte of interest, these assay formats require the use of substrates and in some cases, cofactors and mediators. Further, for a number of these assays, pollutants must be chemically oxidized to metabolic intermediates to show maximum sensitivities. The irreversible nature of many analyte-enzyme interactions that result in increased sensitivity also renders the biosensor inactive after a single measurement. This may not be a problem, however, for those systems that can be reactivated or that employ disposable sensing elements. Another potential limitation to this type of biosensor mechanism involves the sometimes diverse classes of pollutants that inhibit a specific enzyme. In most circumstances, this would not be expected to present problems. In some cases e.g., the co-contamination of environmental samples with organics and heavy metals, such interferences may lead to unexpected results. The interface of these devices to the contaminated matrix is critical to the successful application of biocatalytic-based biosensors, particularly for in situ applications. These matrices may range from pristine drinking water to highly contaminated industrial sludges. Although a considerable amount of work has been done with respect to use of biosensors in biological matrices (such as blood and fermentation media), little work has focused on the development of membrane barriers for the direct sampling of groundwater or pore water in sludges and saturated sediments. Bioaffinity-based biosensors for environmental applications primarily depend on the use of antibodies because of the availability of monoclonal and polyclonal antibodies directed toward a wide range of environmental pollutants as well as the relative affinity and selectivity of these recognition proteins for a specific compound or closely related groups of compounds.18-19 In addition to the wide range of antibodies directed toward different environmental pollutants, a 21 range of assay formats has also been demonstrated with virtually every type of reported signal transducer. Because most small molecular weight organic pollutants in the environment have few distinguishing optical or electrochemical characteristics, the detection of stoichiometric binding of these compounds to antibodies is typically accomplished using competitive binding assay formats. Competitive immunosensor formats rely on the use of an antigen-tracer which competes with the analyte for a limited number of antibody binding sites. For affinity-based biosensors, this is typically accomplished in one of several ways. In one type of format, antigen-tracer competes with analyte for immobilized antibody binding sites (Figure 2). This format is often used in fluorescence-based systems. In another commonly used format, the antigen is immobilized to the signal transducer (operationally becoming the analyte-tracer) while free binding sites on the antibody, which has been previously exposed to the analyte, bind to the surface-immobilized antigen (Figure 2). Because the antibody is a relatively large molecule, its binding to the surface can be detected by signal transduction methods such as surface plasmon resonance, acoustic systems, and optical systems that measure changes in refractive index and thus, do not require an optical tag. The third commonly used format requires an indirect competitive assay and relies on the use of an enzyme-labeled antigen-tracer (Figure 2). In this format, the assay is completed in two steps. First, the enzyme-tracer competes with analyte for immobilized antibody binding sites. Then, after removal of the unbound tracer (by means of a washing step), a non-detectable substrate is catalytically converted to an electrochemically or optically detectable product. This assay format is used almost universally with electrochemical signal transduction. 22 Immunosensors are becoming the most frequently reported type of biosensor for environmental applications.19 Rather than expanding the envelope of fundamental understanding, however, immunosensors (and biosensors in general) for the most part represent technological advances for existing bioanalytical assays. It is important to address the issue of whether or not a biosensor shows the potential to improve the characteristics of a particular assay with respect to known or anticipated applications. Because of the wide variety of scientifically established and commercially available immunoassays (test kits), this is particularly relevant in the area of immunosensors. Although there are a variety of ways to group immunosensors based on signal transducers or format considerations, one functionally useful discriminator involves classification based on reusable/regenerable or disposable format configurations. Because immunosensors (particularly those using disposable formats) are most closely related to immunoassay test kit technology, issues which become important for this comparison are more practical in nature and involve the potential for multi-analyte capability, format versatility, assay time, assay sensitivity, system cost, assay cost, shelf life, reproducibility, ruggedness, etc. In contrast to the disposable formats, the multi-use immunosensors, which can be recharged or regenerated, offer certain advantages, particularly for use as detectors for chromatographic and flow injection analysis systems.20 For example, continuous flow and fiber optic immunosensors that have been reported for detection of explosive residues in ground water use multi-assay formats and perform comparably to chemical and immunoassay test kit methods. Initial cost estimates suggest that for a limited number of assays (e.g., < 400), the test kits are more cost-effective. For long-term monitoring projects, however, the biosensor methods which 23 require an initial instrument investment but offer lower cost per assay would be more costeffective.21 Nucleic acid-based affinity biosensors which may potentially be developed for environmental applications have recently been reported. Application areas for these biosensors include the detection of chemically induced DNA damage22 and the detection of microorganisms through the hybridization of species-specific sequences of DNA.23 Although results from these reports are still preliminary, they appear to offer promising avenues for further investigation. Microorganism-based biosensors tend to use one of three primary mechanisms. For the first mechanism, the pollutant is a respiratory substrate. Biosensors that detect biodegradable organic compounds measured as biological oxygen demand (BOD) are the most widely reported of the microorganism-based biosensors using this mechanism. Several of these devices are commercially available from vendors including: Nisshin Electric Co. Ltd., Tokyo; Autoteam, GmbH, Berlin; Prufgeratewrk, Medingen GmbH, Dresden; and Dr. Lange, GmbH, Berlin. The use of these devices has been incorporated into industrial standard methods in Japan.24-25 Biological oxygen demand is widely used as an indicator of the amount of biodegradable organic compounds found in industrial waste water. The standard procedure (termed BOD5) involves a 5 day incubation of the environmental or industrial water sample with an inoculum of microorganisms typically present in the waste treatment system to yield an endpoint measurement for oxygen consumption.25 The use of indicator microorganisms interfaced to signal transducers allows the measurement of the rate of organic compound metabolism rather than an endpoint; thus, data can be acquired in a significantly shorter time frame (e.g., 15 min to 1 hr), rendering this technology highly advantageous for process control applications. Although these biosensors appear to work well for in situ monitoring of industrial waste waters that result 24 in high BOD values, they currently require improvements in several areas. The primary limitations for these methods involve the variability encountered in calibration of the biosensor response to BOD5 values. This arises from the fact that specific microbial species (used in biosensors) have characteristic substrate spectra which may or may not correspond well with the spectrum of compounds present in the sample. Additional variability results from the presence of polymers (such as protein, starch, and lipid) which must be broken down to monomers before they can be metabolized; this changes the linearity of the response over time and can make interpretation of the result problematic. Current progress on these technologies involves several areas. These areas include: the acid-induced breakdown of biological polymers prior to biosensor analysis, the selection of microorganisms with broad substrate spectra, and the introduction of novel transduction techniques. In addition, a recent report exploring the feasibility of disposable BOD sensors suggests considerable promise for advancement in this area.26 Another mechanism used for microorganism-based biosensors involves the inhibition of respiration by the analyte of interest. Microbial respiration and its inhibition by various environmental pollutants have been measured both optically27-28 and electrochemically.29 Inherent advantages of these techniques apply primarily to the use of microorganisms as compared to isolated enzymes.24 Microorganism-based biosensors are relatively inexpensive to construct and can operate over a wide range of pH and temperature. General limitations involve the long assay times including the initial response and return to baseline. These characteristics are primarily determined by the cellular diffusion characteristics that can be modified by using genetically engineered microorganisms. The broad specificity of these biosensors to environmental toxins may be an advantage or disadvantage depending on the intended 25 application. In this respect, these devices might be most applicable for general toxicity screening or in situations where the toxic compounds are well defined, or where there is a desire to measure total toxicity through a common mode of action. Biosensors have also been developed using genetically engineered microorganisms (GEMs) that recognize and report the presence of specific environmental pollutants. The microorganisms used in these biosensors are typically produced with a constructed plasmid in which genes that code for luciferase or -galactosidase are placed under the control of a promoter that recognizes the analyte of interest. Because the organism's biological recognition system is linked to the reporting system, the presence of the analyte results in the synthesis of inducible enzymes which then catalyze reactions resulting in the production of detectable products. With respect to environmental applications, the primary disadvantage for this type of biosensor is the limited number of GEMs which have been constructed to respond to specific environmental pollutants. Nevertheless, reported advances include the development of GEMs involving a variety of bioremediation pathways and mechanisms. GEMs that could report both the metabolic condition of the relevant microorganisms as well as the rates of pollutant breakdown could be very useful. Physico-chemical transducers Biosense Detectors 26 Manufacturing of Biosensors Biosensor •Specific•Robust •Cheap •Portable •Simple •Easy to use 1) Low incentive to use (effects are long term) 2) Cost high 3) What do you do with the information? 4) Learnt nothing from the Glucose industries development. Integration •Sensor systems •Integration of several steps •Multiple analytes •Expensive •Lab environment •Trained users Miniaturisation •Making integrated systems smaller •Mass production •Cheaper components 27 Common Applications of Biosensors The driving force behind advances in sensor technology came from health care area, where it is now generally recognized that measurements of blood gases, ions and metabolites are often essential and allow a better estimation of the metabolic state of a patient. In intensive care units for example, patients frequently show rapid variations in biochemical levels that require an urgent remedial action. Also, in less severe patient handling, more successful treatment can be achieved by obtaining instant assays. At present, the list of the most commonly required instant analyses is not extensive. In practice, these assays are performed by analytical laboratories, where discrete samples are analyzed, frequently using the more traditional analytical techniques. Many of the technologies first made useful in the hospital setting have now been commercialized into units for home use. A few examples of such devices follows: Blood Glucose Monitors ClearBlue Home Pregnancy Test Cholesterol monitoring 28 “Not-So-Common” Applications of Biosensors Remote Sensing Of Airborne Bacteria Universal Detection Technology (UDTT) has licensed a spore detection technology from NASA's Jet Propulsion Laboratory (JPL) and has developed a first of its kind real-time continuous detection device capable of identifying abnormal levels of bacterial spores in the air, an indication of a possible anthrax attack. Other “Not-So-Common” Uses for Biosensors Remote detection of pathogens Determining levels of toxic substances before and after bioremediation Detection and determining of organophosphate Routine analytical measurement of folic acid, biotin, vitamin B12 and pantothenic acid as an alternative to microbiological assay Determination of drug residues in food, such as antibiotics and growth promoters, particularly meat and honey Drug discovery and evaluation of biological activity of new compounds 29 Activity 4(JW) Bioluminescence Background Bioluminescence (bī´´ōl´´mĭnĕs´ns), the production of light by living organisms. Organisms that are bioluminescent include certain fungi and bacteria that emit light continuously. The dinoflagellates, a group of marine algae, produce light only when disturbed. Bioluminescent animals include such organisms as ctenophores, annelid worms, mollusks, insects such as fireflies, and fish. The production of light in bioluminescent organisms results from the conversion of chemical energy to light energy. In fireflies, one type of a group of substances known collectively as luciferin combines with oxygen to form an oxyluciferin in an excited state, which quickly decays, emitting light as it does. The reaction is mediated by an enzyme, luciferase, which is normally bound to ATP (see adenosine triphosphate) in an inactive form. When the signal for the specialized bioluminescent cells to flash is receive, the luciferase is liberated from the ATP, causes the luciferin to oxidize, and then somehow recombines with ATP. Different organisms produce different bioluminescent substances. Bioluminescent fish are common in ocean depths; the light probably aids in species recognition in the darkness. Other animals seem to use luminescence in courtship and mating and to divert predators or attract prey. Objectives Upon completion of this exercise, students should • Know aseptic techniques. • Understand some of the biochemistry of bioluminescence. • Be familiar with some of the natural history of Vibrio fischeri. Time Requirements This kit can be completed in two or three consecutive class periods. Materials Materials for this kit are sufficient for a class of 30 students working in 6 groups of 5. The materials have been supplied for the sole use of this lab exercise. Carolina Biological Supply Company disclaims all responsibility for any other use of these materials. Included in the kit Vibrio fischeri culture tube 6 metal inoculating loops 7 bottles photobacterium agar 40 sterile disposable petri dishes autoclave disposal bag Needed, but not provided 30 marking pen a disinfectant such as a 10% bleach solution or 70% alcohol Bunsen burner or alcohol lamp water bath or pan of boiling water Discussion A few species of bacteria exhibit bioluminescence, the extraordinary ability of living organisms to produce light. Many of these bacteria live in marine habitats, some as parasites of saltwater fish. Most bioluminescent bacteria give off a blue-green light, while some give off a yellow light. None of these species are harmful. Vibrio fischeri is a bioluminescent marine bacterium that commonly inhabits fish. This bacterium has a Gram-negative cell wall, is motile by means of flagella, and the cell shape is a curved rod. It requires some salt in the medium in order to grow, and it is cultured on photobacterium agar. Photobacterium agar provides the nutrients necessary for good growth and bioluminescence by Vibrio fischeri. Various species of fish, insects, fungi, and microbes can express bioluminescence. Bacteria of the Vibrio genus have been found in association with squid, nematodes, microscopic organisms, and with insects that feed on nematodes. Some gather in the pockets of fish in a symbiotic relationship. The fish are luminous because of the photobacteria. The purpose, if any, of the many various vibrio pigments is still a mystery. However, researchers do know that the occurrence of luminescence is due to the bacterial cells’ electron transport system, involving an aldehyde, an enzyme, oxygen, and an altered form of riboflavin. Though Vibrio isolates are facultative anaerobes, they are bioluminescent only when O2 is present. Several components are needed for bacterial bioluminescence: the enzyme luciferase, a long-chain aliphatic aldehyde, flavin mononucleotide (FMN), and O2. The primary electron donor is NADH, and the electrons pass through FMN to the luciferase. The reaction can be expressed as FMNH2 + O2 + RCHO ➝FMN + RCOOH + H2O + light (Luciferase) The light-generating system constitutes a bypass route for shunting electrons from MNH2 to O2, without involving other electron carriers such as quinones and cytochromes. The enzyme luciferase shows a unique kind of regulatory synthesis called autoinduction. The bioluminous bacteria produce a specific substance, the autoinducer, which accumulates in the culture medium during growth, and when the amount of this substance has reached a critical level, induction of the enzyme occurs. The autoinducer in V. fischeri has been identified as N-β ketocaproylhomoserine lactone. Thus cultures of bioluminous bacteria at low cell density are not bioluminous, but become bioluminous when growth reaches a sufficiently high density so that the autoinducer can accumulate and function. Because of the autoinduction phenomenon, it is obvious that the free-living bioluminescent bacteria in seawater will not be bioluminescent because the autoinducer could not accumulate, and bioluminescence only develops when conditions are favorable for the development of high population densities. Although it is not clear why bioluminescence is density dependent in free-living bacteria, in symbiotic strains of bioluminescent bacteria, the rationale for density-dependent bioluminescence is clear: bioluminescence only develops when sufficiently high population densities are reached in the light organ of the fish to allow a visible flash of light. Much new information about bioluminescence has emerged from studies of the genetics of this process. Several lux operons have been identified in bioluminescent Vibrio species, and the key structural genes cloned and sequenced. The luxA and the luxB genes code for the a and b 31 subunits, respectively, of bacterial luciferase. The luxC, luxD, and the luxE genes code for polypeptides that function in the bioluminescence reaction and in the generation and activation of fatty acids for the bioluminescence system. Light-emitting strains of Escherichia coli have been constructed by inserting lux genes cloned from a Vibrio species into cells of E. coli and then placing their expression under the control of specific and easily manipulable E. coli promoters. This has served to accelerate the pace of research on the regulatory aspects of bioluminescence because many E. coli genetic tools can be employed to probe aspects of lux genes’ control. In addition, cloned lux genes have stimulated biotechnological exploitation of bioluminescence in clinical diagnostics and other biomedical fields. Procedure Note: The culture of Vibrio fischeri you received will probably have passed its prime of bioluminescence. Therefore, it is necessary to transfer the bacteria upon receipt of this kit to successfully demonstrate their bioluminescent ability. However, the culture itself will remain viable for at least 2 weeks at room temperature storage. Part 1. Medium Preparation 1. Slightly loosen the bottle caps and set the bottles in a boiling water bath (or pan of boiling water) to melt the agar. Make sure the water level is even with the agar level. Using an oven mitt, swirl the bottles gently to be sure that all of the agar has melted. 2. Cool the agar to 45°C (the bottle should feel comfortably hot to the touch) by cooling the water bath to that temperature or by letting the bottles sit for several minutes at room temperature. 3. Wipe down the work surfaces with a disinfectant such as 70% alcohol or 10% bleach solution. Wash your hands. 4. Swirl the bottle of agar, remove the cap, flame the mouth over a Bunsen burner for a few seconds, and begin pouring five petri dishes: • Lift the lid of a dish just enough to pour in the molten agar. • Pour just enough agar to cover the bottom of the petri dish. • Replace the lid immediately to prevent contamination. 5. Repeat step 4 with the other photobacterium agar bottles, using one bottle for five dishes. 6. Allow about 2 hours for the agar to solidify. The agar dishes may be stored for up to 1 week at room temperature or can be stored in the refrigerator for up to 3 weeks. Part 2. Preparing Vibrio fischeri Stock Cultures 32 1. Set aside 6 (1 for each student group) of the photobacterium agar dishes for inoculation with the original Vibrio fischeri tube culture and store the other 34 of the 40 dishes for the students to use for their transfers. 2. Before transferring Vibrio fischeri, observe the culture tube in total darkness. Allow at least 5 minutes for your eyes to adjust to the dark in a room with no light leaks. (Note: A room with the lights out and shades drawn does not provide enough darkness to observe bioluminescence). You may or may not observe any growth depending on the bioluminescent phase the bacteria has reached and the time that has elapsed since the initial transfer. Locate an area in the classroom where you and the students can incubate the cultures in an undisturbed, totally dark area (e.g., a cleaned, disinfected cabinet or covered box at room temperature. 3. Transfer from the bioluminescent areas, if any were apparent. To transfer bacteria, place the stock tube in the palm of one hand. With the other hand, flame an inoculating loop until it turns red (about 1 min). Remove the cap with the “loop” hand and flame the mouth of the tube. Insert the cooled loop into the stock tube and pick up a loopful of the culture. Flame the mouth of the tube, replace the cap, and set the tube aside. Gently raise the cover of the petri dish. Touch the inoculating loop to the top of the dish and streak from side to side all the way to the bottom edge. Lower the cover and flame the loop between each transfer. Flame the mouth of the tube and replace the cap. 4. Repeat for the remaining dishes (each of the six plates of transferred culture needs to provide enough growth for 5 students to obtain a sample for their secret message). 5. After inoculating the six plates, incubate cultures at room temperature for 18 to 24 hours in the dark. 6. When checking the cultures for bioluminescence, again allow at least 5 minutes for your eyes to adjust to the dark in a room with no light leaks. Note: A room with the lights out and shades drawn does not provide enough darkness to observe bioluminescence. Bioluminescence is most vivid 18 to 24 hours after subculture. Thereafter, the amount of light emitted decreases inversely with time elapsed. 7. To retain a bioluminescent culture, subculture from bioluminescent areas every 3 to 4 days. If this is not practical, a new subculture should be made 18 to 24 hours before any observations are to be made. Part 3. Student Activity 1. Divide into six groups of five people. Each group needs the following materials: a Vibrio fischeri culture, five photobacterium agar plates, a metal inoculating loop, and a gas (Bunsen) burner or alcohol lamp. 33 2. Disinfect the work surface with a disinfectant, such as 70% ethanol or 10% sodium hypochlorite (see above for preparation). Wash your hands. 3. Each student should label the underside of a petri dish with initials and the date. 4. Before transferring Vibrio fischeri, each group should observe the culture plate in total darkness. Allow at least 5 minutes for your eyes to adjust to the dark in a room with no light leaks. Note: A room with the lights out and shades drawn does not provide enough darkness to observe bioluminescence. Bioluminescence is most vivid 18 to 24 hours after subculture. Thereafter, the amount of light emitted decreases inversely with time elapsed. 5. Light the burner as directed by your instructor. 6. Begin the transfer procedure one at a time. Use aseptic technique, and flame-sterilize the wire end of the metal inoculating loop (heat to redness). 7. After the inoculating loop has cooled, use aseptic technique and obtain a small portion of the bacteria from the prepared culture dish and quickly replace its cover. 8. Inoculate a dish by printing a secret message with the inoculating loop across the surface of the medium (do not penetrate or stab the surface of the agar, or the message may become unreadable). Replace the cover. 9. Flame-sterilize the inoculating loop to remove any remaining bacteria. The next student should begin at step 6 and continue until all transfers are complete. 10. Invert all petri dishes and incubate at room temperature (20–26° C) for 18 to 24 hours in total darkness. 11. After 18–24 hours, locate your petri dish and observe the secret message. Observe the culture in total darkness. Allow at least 5 minutes for your eyes to adjust to the dark. 12. Bioluminescence is most vivid at 18–24 hours and will then begin to decrease in light intensity. At or around hour 96 (the third day), there may be little or no bioluminescence. Try to observe your culture as many times as possible to observe this phenomenon. 13. Observe the message in a lighted area and note the difference. 14. After all observations have been made, dispose of the culture dishes in a designated autoclave disposal bag. (More Background if needed) Bioluminescence is simply light produced by a chemical reaction which originates in an organism. It can be expected anytime and in any region or depth in the sea. Its most common occurrence to the sailor is in the often brilliantly luminescent bow wave or wake of a surface ship. In these instances the causal organisms are almost always dinoflagellates, single-cell algae, often numbering many hundreds per liter. They are mechanically excited to produce light by the 34 ship's passage or even by the movement of porpoises and smaller fish. The deep-sea fish Aristostomias has more than one light organ. Bioluminescence is a primarily marine phenomenon. It is the predominant source of light in the largest fraction of the habitable volume of the earth, the deep ocean. In contrast, bioluminescence is essentially absent (with a few exceptions) in fresh water, even in Lake Baikal. On land it is most commonly seen as glowing fungus on wood (called foxfire), or in the few families of luminous insects. Bioluminescence has evolved many times in the sea as evidenced by the several distinct chemical mechanisms by which light is emitted and the large number of only distantly related taxonomic groups that have many bioluminescent members. Bioluminescent bacteria occur nearly everywhere, and probably most spectacularly as the rare "milky sea" phenomenon, particularly in the Indian Ocean where mariners report steaming for hours through a sea glowing with a soft white light as far as the eye can see. Spectrum In the sea, bioluminescent light is concentrated in the blue window of greatest optical transparency of seawater. Most organisms emit between 440 nm and 479 nm. Some cnidarians have green fluorescent proteins that absorb an initially blue emission and emit it shifted towards the green (~505 nm). One remarkable fish has a similar mechanism to shift the initial emission into the red for use in viewing prey in the near infrared with its red-sensitive eyes. Measurements in situ at various depths confirm emission clustering in the blue to green region of the spectrum. Intensity The luminescence of a single dinoflagellate is readily visible to the dark adapted human eye, as the demonstration will show. Most dinoflagellates emit about 6e8 photons in a flash lasting only about 0.1 second. Much larger organisms such as jellyfish emit about 2e11 photons per second for sometimes tens of seconds. The intensity of luminescence by photosynthetic dinoflagellates is strongly influenced by the intensity of sunlight the previous day. The brighter the sunlight the brighter the flash. Kinetics Some organisms emit light continuously, but most emit flashes of durations ranging from about 0.1 s to 10 s. Some dinoflagellates can respond repetitively to excitation over a short period. In most multicellular species luminescence is neurally controlled. Thus in some fish the sympathetic nervous system controls luminescence by way of the neurotransmitter nor-adrenaline. In fireflies the transmitter is glutamate. In most marine invertebrates the transmitters are unknown. In such forms the "trigger" to luminescence is some detected behaviorally significant event. In single cell organisms like dinoflagellates or radiolarians luminescence is triggered by deformation of the 35 cell surface by minute forces(1 dyne per square cm). Mechanical deformation causes an action potential sweeping over the vacuole membrane and this is thought to induce light emission by admitting protons from the acidic vacuole into contact with the cellular elements that contain the light emission chemistry. In a some instances in marine invertebrates with eyes or other light receptors, light emission can be induced by photic excitation, even by another luminescing organism. Called "empathetic" luminescence, this phenomenon has as yet undemonstrated potential to enhance the luminescence generated by a moving source by photic transfer from the luminescent organisms mechanically triggered by the moving source. 36 Activity 5(JW) Are We Expecting?? The search for hCG using biosensors Background HUMAN CHRIONIC GONADOTROPIN (hCG) The term “pregnancy test” is actually is a misnomer since most pregnancy tests do not actually determine pregnancy but rather the level of hCG. The pregnancy test detects only the unique betasubunit of hCG. Since hCG is only produced in the developing embryo (except in rare cases of secretion of hCG by hydatidiform moles or choriocarcinoma), and since it is detectable within a few days of implantation, it is a very specific and early test for pregnancy. HCG is a glycoprotein hormone composed of two dissimilar units, a and b, joined noncovalently. There are multiple forms of hCG in serum and urine samples in early HCG polypeptide structure pregnancy, including intact hCG (molecular weight- 36700) and its free subunits free a(molecular weight- 14500) and free b (molecular weight- 10000) free b Unit is degraded by macrophage enzymes and in the kidney to make a b subunit core fragment is the principal hCGrelated molecule present in urine during pregnancy(1) intact hCG comprises approximately 70% amino acids and 30% sugar residues. The human chorionic gonadotropin (hCG) test is done to measure the amount of the hormone hCG in blood or urine to see whether a woman is pregnant. HCG is made by the placenta during pregnancy. HCG may also be made abnormally by certain tumors, especially those that come from an egg or sperm (germ cell tumors). HCG levels are generally tested in a woman who may have abnormal tissue growing in her uterus, a molar pregnancy, or a cancer in the uterus (choriocarcinoma) rather than a normal pregnancy. Several hCG tests may be done after a miscarriage to be sure a molar pregnancy is not present. In a man, hCG levels may be measured to help see whether he has cancer of the testicles. A biosensor extends the range of human senses, making more precise observations and measurements possible. A biosensor has two basic functions-recognition of the analyte, the substance being detected, and production of a signal. If you were making a peroxide biosensor, you might use the enzyme peroxidase because it recognizes the molecule peroxide. But peroxidase alone does not produce a visible signal. You would need to use an indicator molecule, such as sodium luminol, along with peroxidase to provide a visible signal. Thus, by taking sample from the flask and adding peroxidase and sodium luminol to them, you could determine which flask contains which liquid. In this biosensor, the amount of light produced 37 correlate with the amount to peroxide in each flask. (Here we won’t measure concentration, just presence) In this design, the enzyme peroxidase recognized the analyte to be measured (peroxide) and catalyzes the reaction that oxidizes the indicator molecule (sodium luminol). The oxidized indicator molecule gives off light, signaling the presence of peroxide. In this activity, you’ll be making and testing a similar peroxide biosensor, using peroxidase and the indicator molecule 4CN. ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA) Enzyme linked immunosorbent assay (ELISA) tests were originally developed for antibody measurement. These immunoassays have also been adapted to successfully detect samples that contain antigens. This ELISA simulation experiment has been designed to detect a hypothetical patient's circulating level of hCG. ELISAs are done in microtiter plates or strips which are generally made of polystyrene or polyvinyl chloride. The plates or strips contain many small wells which are somewhat transparent and in which liquid samples are deposited. First, the antibody against hCG is added to the wells where some remain adsorbed by hydrophobic association to the walls after washing away the excess. There is no specificity involved in antigen or antibody adsorption although some substances may exhibit low binding to the microwell walls. In certain cases the antigens can be covalently cross-linked to the plastic using UV light. After washing away unadsorbed material, the unoccupied sites on the walls of the plastic wells are blocked with proteins, typically gelatin or bovine serum albumin. In this simulation experiment, hCG present in the urine sample will bind to the adsorbed antibody in the well and remain there after washing. If hCG has remained bound to the anti-hCG antibody in the well, then the secondary antibody will bind to another antigenic determinant on the hCG. This complex will remain attached after washing. The second antibody to hCG is purified and covalently cross linked to an enzyme with a high turn over number such as horseradish peroxidase. This modification does not significantly affect the binding specificity and affinity of the antibody or the enzymatic activity of the peroxidase. After washing the well to remove unbound secondary antibody, a solution containing hydrogen peroxide and aminosalicylate is added to each well. Peroxidase possesses a high catalytic activity and can exceed turnover rates of 106 per second. Consequently, amplification of a positive sample can occur over several orders of magnitude. Many hydrogen donor co-substrates can be used by peroxidase. These co-substrates include aminodiansidine, aminoantipyrine, aminosalicylic acid and numerous phenolic compounds that develop color upon oxidation. The substrate solution added is nearly colorless. Peroxidase converts the peroxide to H2O + O2 using the salicylate as the hydrogen donor. It should be noted that polyclonal antibody preparations to a given antigen can have variable binding affinities due to differences in the immunological responses between animals. Different immunizations with the same antigen in the same animal can also produce variable binding affinities. The use of monoclonal antibodies directed against a single epitope eliminates this variability and makes the ELISA highly specific of hCG detection. In this experiment, students will use the ELISA test to determine pregnancy of four hypothetical patients. 38 EXPERIMENT OBJECTIVE: The objective of this simulation experiment is to introduce concepts and experimental procedures involved with enzyme linked immunosorbent assays (ELISA) in the context of testing for pregnancy. LABORATORY SAFETY Gloves and goggles should be worn routinely as good laboratory practice. GENERAL INSTRUCTIONS AND PROCEDURES Labeling the Microtiter Strip: • Place the microtiter strip as shown in Figure 1. • Carefully mark the strip with your initials or lab group number and number the wells 1-4 down the side. • Do not separate the reaction wells. Labeling the Plastic Transfer pipets: Label 5 transfer pipets as follows: • ( - ) (negative) • (+ ) (positive) • USP 1 (Urine Sample Patient 1) • USP 2 (Urine Sample Patient 2) • PBS (Phosphate Buffered Saline) Use the appropriately labeled plastic transfer pipet for sample additions, removals, and washes as outlined in the experimental procedures. If using transfer pipets to add reagents to the wells, label 3 additional pipets "hCG Ab", "hCG Ab2" (hCGAb-HRP conjugate), and "substrate". INSTRUCTIONS FOR ADDING LIQUIDS AND WASHING WELLS Adding Reagents to Wells: For adding reagents to the wells, use the labeled transfer pipets or use an automatic micropipet and disposable tips. Liquid Removal and Washes: When instructed in the experimental procedures, remove liquids with the appropriately labeled transfer pipet, and then wash the wells as follows: A. Use the transfer pipet labeled "PBS", to add PBS buffer to each of the wells. Add PBS buffer until each well is almost full. 39 The capacity of each well is approximately 200 l. Do not allow the liquids to spill over into adjacent wells. B. With the appropriately labeled transfer pipet, remove all the liquid (PBS buffer) from each of the wells. Dispose the liquid in the beaker labeled "waste". ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA) 1. To all 4 wells, add 50 l or 3 drops of “hCG Ab” (human chorionic gonadotropin antibody). If using a transfer pipet, use the one labeled "hCG Ab". 2. Incubate for 5 minutes at room temperature. 3. Remove all the liquid (hCG antibody) with a transfer pipet. 4. Wash each well once with PBS buffer as described above ("Liquid Removal and Washes"). In research labs, following this step, all sites on the microtiter strip are saturated with a blocking solution consisting of a protein mixture, such as BSA. We have designed this experiment to eliminate this step to save time. 5. To Test Patient 1, add reagents as outlined below: Remember to use the appropriately labeled transfer pipets provided with this experiment for adding a new reagent. If you are using automatic micropipets, use a clean micropipet tip for each reagent. • Add 50 l or 3 drops of PBS Buffer to the first well. (This is the negative control.) • Add 50 l or 3 drops of "+" (positive) to the second well. (This is the positive control.) • Add 50 l or 3 drops Urine Sample from Patient 1 “USP 1” to the third well. • Add 50 l or 3 drops Urine Sample from Patient 2 “USP 2” to the fourth well. 6. Incubate at 37oC for 15 minutes. 7. Remove all the liquid from each well with the appropriately labeled transfer pipet. 8. Wash each well once with PBS buffer (as described under "Instructions for Adding Liquids and Washing Wells" on page 8). -hCG peroxidase conjugate (hCG Ab2) to all 4 wells of each strip. If using a transfer pipet, use the one At this time you can obtain the substrate to be used in step 13. Since the substrate must be prepared just prior to use, your instructor will prepare it towards the end of the incubation in step 10. 40 11. Remove all the liquid from each well with the appropriately labeled transfer pipet. 12. Wash each well once with PBS buffer (as described under "Liquid Removal and Washes"). r pipet, use the one labeled substrate". 15. Remove the strip for analysis. LABORATORY NOTEBOOK RECORDINGS: Address and record the following in your laboratory notebook or on a separate worksheet. Before starting the experiment: • Write a hypothesis that reflects the experiment. • Predict experimental outcomes. During the Experiment: • Record (draw) your observations, or photograph the results. Following the Experiment: • Formulate an explanation from the results. • Determine what could be changed in the experiment if the experiment were repeated. • Write a hypothesis that would reflect this change. STUDY QUESTIONS Answer the following study questions in your laboratory notebook or on a separate worksheet. 1. Why is the ELISA reaction so sensitive? 2. Why is it important to have a positive control? 3. What is being detected in the standard home pregnancy test? 4. What is the difference between polyclonal and monoclonal antibodies 41 Activity 6(EA) USE OF BIOTECHNOLOGY TO ASSESSING ENVIRONMENTAL IMPACT OF BIODEGRADATION ON ECOSYSTEM TOXICITY Reporter genes or screenable markers have proven very useful in biotechnology research as they provide a visual means to identify and monitor the effects of harmful compounds released into the environment using genetically engineered cells in organisms such as, transgenic plants, animals and microorganisms. This technology has been proven useful in assessing the harmful effects of degradable materials, agricultural and industrial runoffs into the environment, as well as assessing the adverse effect of degrading commonly used household items. Cells transformed with the green fluorescence protein gene exhibit bright fluorescence under ultraviolet or blue light. The proposed experiment is one that uses bacterial cells containing the pGLO plasmid that was genetically engineered to express the gene for green fluorescence protein (GFP) from jellyfish. This gene was placed next to the promoter that regulates expression of bacterial genes needed to break down, or digest, the simple sugar arabinose. This promoter is activated in the presence of arabinose but not in its absence. Bacterial cells containing this plasmid, pGLO, when grown in the presence of arabinose, the GFP gene is turned on and the bacteria fluoresce a brilliant green color under long-wave ultraviolet light as shown below. QuickTime™ and a TIFF (Uncompressed) decompressor are needed to see this picture. Obtained from BIO-RAD pGLO can also be used to introduce students to the concept of genes and their most basic function — to code for proteins. Content Focus 42 Harmful compounds are released into the environment from degradable materials, agriculture, industry and even from our own households. Importance of detecting and predicting the effects of compounds released from this process on the environment. Released in small amounts that are disperse into soil and water, therefore difficult assess long-term effect. Use of harmless bacteria as biological monitors to determine whether a compounds/chemicals released from degradation are toxic to the environment. Suitable field-test and cost effective when compared to tests done with mice, rabbits, and rats. Activities 1. Bioluminescence. The use of biological engineered bacteria to produce green fluorescence protein (GFP) to produce light. QuickTime™ and a TIFF (Uncompressed) decompressor are needed to see this picture. Obtained from Carolina Biological Demonstrate potential mutagenic/toxic effects of biodegradable compounds in using genetically engineered bioluminescent bacteria Recording the level of light produced when cells are grown in the presence of a test compound Protocol Culture GFP expressing bacteria overnight in presence of arabinose and ampicillin (This can be done in10ml test tubes or a flask) Aliquot samples of overnight culture into 96-well microtiter plate containing of nutrient broth with arabinose and ampicillin (Help select for GFP expressing bacteria) Add varying concentrations of compound in the lane of wells for each compound tested a. Here, students will perform a serial dilution of a control compound know to cause mutations and affect cell growth, positive control. 43 b. There will be two control lanes, one lane containing the control compound and another containing no compound to serve as a negative control. c. Other lanes can be used to test material from biodegradation. Incubate overnight at 370C. Examine the effect of each compound on the growth and fluorescence by recording degree of fluorescence detected in each lane for each compound tested or counting the number of cells/ml of culture after plating cells on selective agar medium. Obtained from BIO-RAD GFP Quantification. The fluorescence of GFP-producing cells that are grown in culture can be quantified by exposing culture to long-wave ultraviolet light. The mutagenic/toxic effect of compounds can be determined by one of two methods. 1. Visually accounting for the lost of fluorescence when cells are exposed to compounds tested. 2. The process can be quantified by plating cells from cultures illustrating a visual effect of the compound(s) being tested, i.e., lost of fluorescence. Under these 44 conditions, a serial dilution can be performed on cultures showing a lost in fluorescence to, (1) determine toxicity of the compound (lost in number of cells/ml) and/or (2) mutagenicity of the compound (lost in number of cells fluorescing). This can be accomplished by: Performing a serial dilution of the culture Plate and incubate cells overnight Count number of cells per plate and multiply by dilution factor to determine number of cells/ml in culture Demonstrating the effect of the compound on cell growth by grafting effect of increasing concentration of compound on number of cells in culture. 2. Assessment the toxicity/mutagenicity of compounds released from biogradation and/or commonly used household products. Serial dilutions of compounds and household products to determine the concentration needed to be toxic/mutagenic. Use of light producing bacteria to assess toxicity of compounds released from degradation process. Use light producing bacteria to determine toxic effect of commonly used household products. Test effect (toxicity/mutagenicity) of compounds based on lost or intensity of luminescence from bacterial exposure to household compounds or compounds resulting from biodegradation. Effect of compound on cell growth could also be determined by performing a serial dilution of cells and plating an aliquot of the culture on selective agar plates a counting the number of cells on plate as a result of exposure to degradable compound Students develop ability to use statistical analysis, graphing, and tabulation of findings Discussions Points: Scientific method Biological molecules (monomers, polymers, synthesis and hydrolysis, enzymes) 45 Microbiology (prokaryotes – structure, metabolism) Ecology (pollution) Mutagenicity/toxicity Cancer 46 Demonstration #1(JW) Chemiluminescence Although exothermic reactions usually liberate energy as heat, there are occasional reactions in which energy is emitted as light at or below room temperature. These chemiluminescent processes can occur when the energy given off in a reaction is capable of producing molecules in an electronically excited state. As these excited molecules drop to the ground state, the light emitted is sometimes visible. Many such reactions occur in nature (for example, in the firefly), where they are referred to as bioluminescence. In this experiment, luminol (3aminophthalhydrazide) will be prepared and oxidized under basic conditions causing the emission of light. Although the mechanism of this chemiluminescent process in aqueous solution is not entirely understood, the singlet (all electrons paired) excited state of the 3aminophthalate ion is know to be the light-emitting species. NO2 NO2 O COOH + COOH 3-Nitrophthalic Acid NH2 NH2 NH -H2O NH O 3-Nitrophthalhydrazide Na2S2O4 NH2 O NH NH O Luminol 47 A. 3-Nitrophythalhydrazide Pour 15 mL of water into a small beaker or flask and heat it on a steam bath or hot plate. Place 1.00 g (4.74 mmol) of 3-nitrophthalic acid and 2 mL of an 8% (by weight) solution of aqueous hydrazine into a large test tube (20 x 150 mm or 25 x 200 mm). Add a boiling chip, mount the tube at a slight angle on a support track, and heat the mixture with a small burner to dissolve the solid. Add 3 mL of triethylene glycol and insert a thermometer so it rest on the bottom of th test tube. Boil the solution to remove the excee water, allow the temperature to rise rapidly (over several minutes) to 215oC, and maintain a temperature of 210-22- oC for 3-4 min. Let the reaction mixture cool to 100 oC, add 15 mL of the hot water prepared previously, and cool the test tube in running water. Collect the light yellow granular product and use it without purification or drying in the next step. B. 3-Aminophthalhydrazide (Luminol) Rinse the test tube used in Part A and place the wet 3-nitrophthalhydrazide into it, together with 5 mL of 3M sodium hydroxide and 3.0 g (14 mmol) of sodium hydrosulfite dihydrate (also called sodium dithionite), Na2S2O4.2H2O. Heat the mixture to boiling, stir, and maintain boiling 5 minutes. Cool the reaction mixture and in a hood slowly add 2 mL of glacial acetic acid. Collect the precipitated luminol by suction filtration, wash with a little water and allow it to dry or dry it with gentle heating (heat lamp for above). The moist material can be used in the chemiluminescence step next. C. Chemiluminescence Mix 50 mg of luminol with 5 mL of 3 M sodium hydroxide in a 250 mL Erlenmeyer flask and dilute the solution to about 150 mL. Add 10 mL of 3% hydrogen peroxide solution. Read ahead to be certain you will have everything you will need and then go to an area that can be darkened for the next part of the experiment. (Complete darkness in not necessary.) Add approximately 250 mg of small crystalline chunks of potassium hexacyanoferrate (III) also know as potassium ferricyanide to the solution, swirl it and note what happens. Pour some of the luminescent solution into a small beaker and add some 3 M hydrochloric acid. Record the result. Put one or two pellets of sodium hydroxide into another small beaker and add some of the luminescent. Record what happens in the vicinity of the pellets. To each solution, place in a warm water bath and report your observations. 48 Proposed Emission of Light (h) Intersystem crossing S1 T1 Emission -h Peroxide -N2 Luminol Dianion + O2 Photon absorption +h S0 3-Aminophthalate dianion Fluorescence emission spectrum of the 3-aminophthalate dianion Intersystem crossing S1 T1 Emission -h Peroxide -N2 Luminol Dianion + O2 3-Aminophthalate dianion Photon absorption +h S0 49 Demonstration #2(JW) Testing Know Concentrations of Peroxide In this activity, you will see whether or not your peroxide biosensor can distinguish different concentrations of peroxide. Then you’ll use your biosensor to try to identify unknown concentrations of peroxide. Predictions: You will be testing samples that contain different concentrations of peroxide. You will test 1% and 0.00003% peroxide solutions and six other peroxide concentrations of your choice. Predict what will happen when you test each sample. How will the results differ for different concentrations? Record your predications in a data table. Make a data table with the following categories: The concentrations of peroxide your are testing Your predictions about what will happen to the sensor strip placed in each concentration Your observations of what actually happened to the sensor strip placed in each concentration Your quantification of the data Gather these materials: 8 different concentrations of peroxide (including 1% and 0.00003%) 8 test-tube caps (or other small containers) Self-stick labels Graduated droppers Peroxidase enzyme solution Tweezers Sensor strips Paper towels Centimeter ruler 50 Procedure 1) Get 1 mL of each of the peroxide concentrations you will test 2) Label a container for each of the peroxide concentrations you will be testing 3) Using a graduated dropper, measure 0.1 mL of peroxidase into each container 4) Use another dropper to aid 0.3 mL of the appropriate peroxide concentrations to each container. To avoid contamination of the sample, star with the least concentrated sample and continue I order to the most concentrated sample. Gently tap each container to mi the contents 5) Use tweezers to add a sensor stripe to each container 6) When the entire strip is wet, use tweezers to remove it from the solution and place it on a paper towel. Record your observations for each concentration in you data table. 51 Glossary (ALL) Aerobe - An organism requiring free atmospheric oxygen for normal activity. Anaerobe - An organism that derives energy for life from chemical changes that do not require oxygen. Some anaerobes cannot tolerate the presence of free oxygen. Bacteria - A group of microscopic one-celled organisms. Bioluminescence - An emission of light by living organisms. Chemiluminescence – is the emission of light (luminescence) without emission of heat as the result of a chemical reaction. Electron - A stable elementary particle which is the negatively charged constituent of ordinary matter. Enzyme - A protein that acts as a catalyst. Exothermic Reactions – A reaction in which heat is released. Flagella - Whip-like projections of cytoplasm used in locomotion by certain simple organisms. Fluorescence – is a luminescence that is mostly found as an optical phenomenon in cold bodies, in which the molecular absorption of a photon triggers the emission of another photon with a longer wavelength Luciferase - An enzyme that catalyzes the oxidation of luciferin. Luciferin - A species-specific pigment in many bioluminescent organisms that emits heatless light when combined with oxygen. Mutagenicity – the act by which a physical or chemical agent changes the genetic information (usually DNA) of an organism and thus increases the frequency of mutations above the natural background level. Photobacteria - Bacteria that can synthesize light. Plasmid – A plasmid is a DNA molecule separate from the chromosomal DNA and capable of autonomous replication. It is typically circular and double-stranded Protein – Large organic compounds made of amino acids arranged in a linear chain and joined together by peptide bonds between the carboxyl and amino groups of adjacent amino acid residues Symbiotic - The living together of two or more different organisms in close association. 52 53 Activity #5: Are we expecting? 54 Activity Outline: Electrochemical Glucose Sensor (BL) Introduction Background information about Type I/II Diabetes: Insulin dependence and why? Feedback mechanisms: Insulin/Somatostatin Teams and Materials Needed Sugars (Sucrose, Glucose, Fructose, Starch, Cellulose) Common beverages (milk, soda, beer, etc.) Electrochemical cell components (most likely microscale) Still looking for some “self-sampling” technique that doesn’t involve finger pricking, although this would be an interesting option The Concepts Needed Electrochemistry o Nernst Equation o Redox Reactions: Balancing of half-cell reactions Polymers o Synthetic (isotactic, syndiotactic, and atactic) o Natural (isoprene rubber) Organic polymers (basic concepts of sugar polymerization) o Biopolymers and Biocolloids—Cellulose, Starch, Collagen, Insulin, Mucus, Lipids Electricity concepts/Circuitry o Basic understanding of voltage, current, and resistance o Kirchoff’s Laws Colligative properties o Ionization o Conductivity of solutions Procedure Something to do with testing of different mono- and disaccharides as well as cellulose and starch from pure compounds to mixtures of liquids with KNOWN glucose concentrations. o Accuracy of sugar analysis—Blood glucose metering o Selectivity of sugar analysis o Tolerance Affects of conditions on accuracy of measurements (pH, temperature, concentration) Analysis Something to do with analysis of accuracy of detecting glucose present different mono- and disaccharides as well as cellulose and starch Conclusions and Considerations Accuracy of detection v. insulin dosing How to make a “better” (i.e. prickles) monitor 55