AbstractID: 7821 Title: On The Accuracy Of Current Methods For... HDR Brachytherapy Sources

advertisement

AbstractID: 7821 Title: On The Accuracy Of Current Methods For Calibrating Ir-192

HDR Brachytherapy Sources

The accuracy of the averaging procedure proposed by Goetsch et al. (Med. Phys. 18(3), (1991) 462-467) for calibrating

192

Ir HDR brachytherapy sources is investigated. Quantities required for obtaining air kerma calibration factors N k

for a Farmer ionization chamber, irradiated by 250 kV x-ray beams and

137

Cs,

192

Ir and

60

Co photons, have been calculated using the EGSnrc simulation system. Use of a medium filtered x-ray beam (E ave

= 146 keV) and removing the influence of A wall by averaging over N k

A wall

predicts an N k

value for the

192

Ir HDR brachytherapy source, which is 1.1% higher than the Monte Carlo value. This method agrees within

0.2% with the N k

Monte Carlo estimate for a heavily filtered x-ray beam (E ave

= 205.6 keV). A simpler averaging of just the calibration factor N k

has been found to improve the accuracy to 0.7% and 0.02% for the medium and the heavily filtered x-ray beams and this also avoids use of the invalid linear extrapolation technique of determining A wall

. The energy dependence of the calibration factor N k

is mainly governed by the wall to air mass energy absorption coefficient ratios rather than by the correction factor for attenuation and scatter in the chamber walls A wall

. An air-kerma weighted averaging technique of the chamber response (Verhaegen et

al. Phys. Med. Biol. 37(11), (1992) 2071-2082) produces an N k

calibration factor that is within 0.02% of the Monte Carlo estimate. If the average is taken over the calibration factor, the agreement is still excellent (0.03%).

Download