Quality Control Testing of GE Digital Mammography Systems John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI john.sandrik@med.ge.com GE Healthcare 2007 Day 2: Overview • • General considerations The QC tests – – GE Healthcare Tests also done by Rad. Tech. • Flat Field and Phantom Image Quality • MTF and CNR • AOP and SNR Tests that need no introduction • kVp Accuracy and Reproducibility • Beam Quality Assessment (HVL) • Radiation Output • Mammo. Unit Assembly Evaluation 2007 Day 2: Overview (cont’d.) • The QC tests – GE Healthcare FFDM-specific, FFDM-modified tests • Collimation Assessment / “Film-less” • Evaluation of Focal Spot Performance • Sub-system MTF • Ent. Exposure, Avg. Glandular Dose • Artifact Eval., Flat Field Uniformity • View Conditions Check and Setting • Monitor Calibration • Image Quality–SMPTE Pattern • Display Screen Uniformity 2007 General Considerations GE Healthcare 2007 Protecting the Detector • Repeated imaging of a stationary object may lead to formation of “ghost” image. (2.5 mm thick) – • E.g., HVL, kVp tests. Cover detector with steel & lead plate (provided) when images not needed. 235 mm 233 mm GE Healthcare 2007 Order of Performing Tests 1. Flat Field test (RT Test, Chap. 1) – but not Phantom IQ tests. 2. Artifact Evaluation; Flat Field Uniformity (MP Test, Chap. 2) 3. Remaining tests – including Phantom IQ tests, Chap. 1 Minimize artifacts caused by testing. GE Healthcare 2007 Analyzing Images • Images displayed after acquisition are in logarithmic format. – • Labeled “Proces” in Browser. Quantitative tests must be performed on “Raw” images. – – – CNR Test MTF Measurements AOP Mode and SNR Check Automatic image selection for RT Tests on Seno DS and Essential GE Healthcare 2007 Raw or Processed Image? Annotation Level “Full” or “Partial” Processing Description • “PROC_0,” “PROC_1,” etc., then Processed • None, then Raw GE Healthcare 2007 Seno DS & Essential: FineView From the QC Manual: “… a physicist’s measurements (e.g. MTF and noise) performed using methods other than those described in this manual can be affected when FineView processing is applied.” • FineView compensates for the detector MTF. • With FineView enabled, measurement procedures differing from the QC Manual or QAP Tool may give unexpected results. • Disable Fine View for Sub-system MTF. Check FineView before doing independent tests. GE Healthcare 2007 The QC Tests Tests also done by Rad. Tech. GE Healthcare 2007 Flat Field Test Browser Window Then Select Flat Field from list of tests. GE Healthcare 2007 Flat Field Test • Image 25 mm thick PMMA plate – • • • • To avoid false results, plate must be clean and free from imperfections. Automated parameter selection and analysis Set up instructions on-screen Make two exposures Results, Action Limits, and Pass/Fail posted at end of test GE Healthcare 2007 Flat Field Test • • • • Image sampled using 2 x 2 cm2 ROIs at 1 cm intervals Mean, Std. Dev. measured for each ROI Test five image quality measures Brightness Non-Uniformity (BNU) – • (Max ROI Mean – Min ROI Mean) expressed as % of Average ROI Mean. High Frequency Modulation (HFM) – GE Healthcare Max (ROI Std. Dev. / ROI Mean) expressed as percent. 2007 Flat Field Test • SNR Uniformity – – – Second image subtracted from first and std. dev. calculated for each ROI of difference image. SNRs calculated from ROI Means of first image and ROI Std. Devs. of difference image. (Max ROI SNR – Min ROI SNR) expressed as percent of Average ROI SNR. GE Healthcare 2007 Flat Field Test • Bad Pixel Each pixel signal compared to its ROI Mean. – |Pixel signal – ROI Mean| > threshold means “bad pixel.” – • Bad ROI – ROI with ≥ 2 bad pixels = “bad ROI.” GE Healthcare 2007 Phantom Image Quality Breast Support Phantom Chest wall edge • Acquire image in Medical Application • 19 x 23 centered FOV for Essential • Use specified manual techniques • Score Processed phantom image on acquisition WS and printer – Also on review WS for 2000 D GE Healthcare 2007 Change of CNR Test: 2000 D • • • Use image from phantom IQ test. Analyze largest mass in Raw image of phantom. Measurement compared to baseline established from average of five measurements over five days. GE Healthcare 2007 Change of CNR Test: 2000 D Second ROI centered between largest mass and group of largest specks. 2 1 • Difference of means measures contrast. • Std. Dev. of background measures noise. First ROI centered over largest mass ∆CNR must not exceed 0.2. GE Healthcare 2007 MTF Measurement: 2000 D • Check of detector MTF – • Bar pattern on surface of Bucky – – • • • • Minimal magnification Para. or perp. to chest wall edge. To avoid false results, the pattern must be clean and free from scratches. Remove compression paddle. Set specified parameters. Acquire image in Medical Application Analyze Raw image. GE Healthcare 2007 MTF Measurement: 2000 D ••• ROI to measure mean of “space” material. 3.93 2.09 1.23 ROI to measure std. dev. of “4 lp/mm” pattern. 1.11 1.0 ROI to measure mean of “bar” material. ••• ••• ROI to measure std. dev. of “2 lp/mm” pattern. Std. Dev. × 222 % MTF = Meanspace − Meanbar GE Healthcare 2007 CNR & MTF: DS, Essential • • • • • Image of Image Quality Signature Test (IQST) tool. Automated parameter selection and analysis Select QAP button. Then select CNR and MTF test. Set up instructions on-screen GE Healthcare 2007 CNR & MTF: DS, Essential IQST – Image Quality Signature Test Automates MTF and CNR Tests GE Healthcare 2007 CNR & MTF: DS, Essential Resolution Uniformity Noise Power Spectrum MTF Contrast GE Healthcare 2007 CNR & MTF: DS, Essential • Measurement compared to baseline established from average of five measurements over five days. – • Comparison not valid until after fifth measurement. Results, Action Limits, and Pass/Fail posted at end of test. – GE Healthcare If failed, ensure that FineView has not been changed from previous setting. 2007 AOP Mode and SNR Check Check for • Correct selection of – – – – kVp, anode track, filter, and mAs and Correct level of SNR using Automatic Optimization of Parameters (AOP) when varying phantom thickness. • GE Healthcare 2007 AOP Mode and SNR Check For 2000 D • • • • • • • • Acquire images in Medical App. Select the STD AOP mode. Acquire images. Record parameters selected. Open the Raw images for analysis. Set the Zoom to “True Size.” Set ROI, read mean and std. dev. Calculate SNR. GE Healthcare 2007 AOP Mode and SNR Check For DS, Essential • Select QAP button. • Then select AOP and SNR Check. • Set up instructions on-screen. • AOP mode selected automatically. • Acquire images. • Parameters compared to specs. • Results, Action Limits, and Pass/Fail posted automatically at end of test. GE Healthcare 2007 AOP Test Plates 20 x 20 cm2 Seno 2000 D GE Healthcare 10 cm 21 cm Seno DS, Essential 2007 AOP Test Plate Change Improved simulation of breast shape GE Healthcare 2007 AOP Change & Phantom IQ • Seno 2000 D – – – • Mo / Mo, 26 kVp, 125 mAs Simulates AOP CNT mode “film-like” Seno DS and Essential – – – GE Healthcare Rh / Rh, 29 kVp, 56 mAs Simulates AOP STD mode “digital” 2007 AOP Change Predicted Track / Filter Combination Use Rh/Rh, 79% Mo/Rh, 20% Mo/Mo, 1% Optimized for consistent CNR rather than detector exposure N.Shramchenko, P.Blin, C.Mathey and R.Klausz. Optimized exposure control in digital mammography. Medical Imaging 2004: Physics of Medical Imaging, Proc. SPIE 5368, 445-456 (2004) GE Healthcare 2007 The QC Tests Tests that need no introduction GE Healthcare 2007 QC Tests for the Med. Phys. • • • • kVp Accuracy and Reproducibility Beam Quality Assessment (HVL) Radiation Output Mammographic Unit Assembly Evaluation Tests not unique to FFDM. No procedures provided in the QC plan. GE Healthcare 2007 The QC Tests FFDM-specific, FFDM-modified tests GE Healthcare 2007 Collimation Assessment Seno 2000 D • Only 24 x 30 film and cassette identified in equipment list – Will be generalized in next revision Seno DS and Essential • Identify “auxiliary image receptor” – 24 x 30 cassette – smaller cassette rotated and/or elevated – general rad. screen-film or CR cassette – multiple, small, distributed detectors GE Healthcare 2007 Collimation Assessment Check both Mo and Rh anode tracks Seno Essential • Acceptance, after a major repair (MEE) • – – – – • 24 cm x 30.7 cm 19 cm x 23 cm, centered 19 cm x 23 cm, offset right 19 cm x 23 cm, offset left Annual QC survey – GE Healthcare 24 cm x 30.7 cm 2007 Collimation Assessment Moving and sizing FOV for Essential • Two buttons on back of collimator – – • lower sets field size upper sets field position Setting size – – – First press of size button turns on light. Second press changes FOV to next size. Press repeatedly for desired size. GE Healthcare X-ray tube head, rear view FOV Position FOV Size 2007 Collimation Assessment Moving and sizing FOV for Essential X-ray tube head, • Setting position rear view – – – First press of position button turns on light. Second press moves FOV to next position. Press repeatedly for desired position. FOV Position FOV Size GE Healthcare 2007 Collimation Assessment Sliding paddle on Essential • Paddle position must agree with FOV position to enable exposure. • To slide paddle, press either paddle release button. – • Release button once paddle has slid from its initial position. At pre-defined position, paddle automatically locks. GE Healthcare 2007 Collimation Assessment • • • Paddle release for Essential Turn paddle unlocking knob clockwise until it points to unlocked padlock symbol on the paddle holder. Release knob, slide paddle away. – – Knob then returns to its initial position. For sliding paddle, not necessary to push sliding release button. GE Healthcare 2007 Collimation Assessment Suggested procedure for Essential • Set gantry at 0 degrees. • Set 19 x 23 FOV size (lower button). • Set FOV position (upper button). • Mark the light field edges. • Install sliding 19 x 23 paddle to match FOV position. • Make the exposures. GE Healthcare 2007 “Filmless” Collimation Test Use of GAFCHROMIC® XR-QA Light field 3.2 mm Aluminum plate 42 chips per sheet GE Healthcare 10" x 12" sheet 4 x 4 cm2 chip 2007 “Filmless” Collimation Test Set up • White side up • Better contrast of light field Exposure • 30 kVp, 200 mAs, Mo/Mo and Rh/Rh • Self-develops “instantaneously” Attenuator • 3.2 mm aluminum • Useful contrast on film at exposure level used • Avoid detector saturation and “ghost” images GE Healthcare 2007 “Filmless” Collimation Test GAFCHROMIC® XR-M XR-M film strip placed white side up Light field Orange White side side (case marked) White paper 3.2 mm Aluminum plate International Specialty Products http://www.ispcorp.com/products/dosimetry/content/gafchromic/index.html GE Healthcare 2007 “Filmless” Collimation Test GAFCHROMIC® XR-M light field x-ray field detector Light – X-ray deviation = 5 mm X-ray – Detector dev. = 4 mm in film plane. Then scale to detector plane. GE Healthcare 2007 Focal Spot Performance • • • • Film-based Essentially the same as screen-film test. Magnification tested at 1.5X. Recommend to replace with SubSystem MTF test GE Healthcare 2007 Focal Spot Perf., Essential For test of small focal spot, cassette will not lie flat on detector cover GE Healthcare 2007 Focal Spot Perf., Essential Support cassette on 3 cm of AOP test plates or do Sub-System MTF Test GE Healthcare 2007 Sub-System MTF • Bar pattern at 4.5 cm above breast support – • • • Film-less measurement Option to replace detector MTF test and (film-based) Eval. of Focal Spot Perf. Physicist must provide two bar patterns – – • Tests focal spot plus detector sub-system. “low freq.” with 2.09 and 3.93 lp/mm “high freq.” with 5 and 8 lp/mm Fine View must be disabled (DS & Ess.). GE Healthcare 2007 Sub-System MTF Contact Configuration • • • • 4.5 cm acrylic block on Bucky Bar pattern on block Mo and Rh targets Perp. and Para. to Anode-Cathode Axis GE Healthcare 2007 Sub-System MTF Small focal spot • Tested at 1.8X. • For Seno 2000 D, – – • edge of bar pattern within 1 cm of chest wall edge of image receptor. 9 cm x 9 cm FOV For DS & Ess., – – edge of bar pattern 50 mm from chest wall surface of mag. stand. • Larger value, more reliable measurement • Less sensitive to positioning error 13 cm x 18 cm FOV GE Healthcare 2007 Sub-System MTF Magnification Configuration • • • • • Bucky removed • 9x9 FOV for 2000 D, 13x18 for DS, Ess. 4.5 cm acrylic block on Mag. Stand at 1.8X Bar pattern on block Mo and Rh targets Perp. and Para. to Anode-Cathode Axis GE Healthcare 2007 Sub-System MTF Test Measure Mean and Std. Dev. in ROIs 3 4 1.23 1 1 2 5 2.09 1.37 1 1.52 1.69 1 1.88 2.09 2.32 2 3.93 2.09 and 3.93 lp/mm for Contact GE Healthcare 2 8 2 2.58 2.87 3.19 3.54 3.93 4.37 4.86 8 10 11 12 13 14 15 16 17 1819 20 1.11 4 5 1.0 3 5 and 8 lp/mm for Magnification 2007 Sub-System MTF Calculation • Presence of acrylic block reduces signal compared to detector MTF test. • Noise correction added to improve accuracy. GE Healthcare 2007 Bar Pattern Accuracy Bar Pattern Frequencies (lp/mm) Pattern Group Pattern A Avg. Freq. Pattern B Avg. Freq. 5 5.0 4.5 6 6.0 5.5 7 7.0 6.5 8 7.9 7.5 9 9.0 8.5 10 10.0 9.4 11 10.7 10.5 12 11.8 11.2 13 13.2 12.5 GE Healthcare Def’n. of Avg. Freq.: 4.5 line pairs 0.9 mm 4.5 lp 0.9 mm = 5.0 lp/mm Bar pattern calibration is recommended. 2007 Exposure, Dose, Repro. Dose in Automatic Optimization of Parameters (AOP) mode measured by • Acquiring image of accreditation phantom and recording parameters, • Replacing phantom by ion chamber, • Setting AOP-selected parameters in Manual mode, • Measuring exposure for selected parameters. GE Healthcare 2007 Exposure, Dose, Repro. AOP parameter selection based on • Most attenuating “breast” tissue in AOP sense window. – • May reject highly attenuating object. Breast composition – Based on attenuation and compression paddle height For accurate dose measurement • No high attenuators in sense window • No objects on phantom top GE Healthcare 2007 Dose Measurement for AOP Phantom rotated 180º. Wide edge of wax insert frame 2000 D & DS / Essential 23 / 31 cm X-ray field AOP Sense Window 16 / 23 cm 19 / 24 cm 3.5 cm 14 / 19 cm Ion chamber outside AOP Sense Window GE Healthcare Phantom Chest wall edge No protruding screws or disk on phantom 2007 Exposure, Dose Summary • • • • • AOP parameter selection is more consistent with phantom rotated 180°. Acrylic disk on phantom increases dose. Disk and raised thumbscrews cause thickness estimate error. Consistent compression force of 5 daN improves AOP consistency. ACR method – uniform phantom – is only for accreditation application, not QC. GE Healthcare 2007 Artifact Eval., Flat Field Unif. • • • • • • • • Image of uniform acrylic plate Mo/Mo, Mo/Rh, Rh/Rh at lowest clinical kVp Large focal spot with grid in Small focal spot without grid Review Raw image Set window width to 400 - 450 Visual evaluation of uniformity No artifact or non-uniformity that is expected to mimic or obscure clinical information GE Healthcare 2007 Artifact Eval., Flat Field Unif. Bands in small FS raw image Anode cut off, ~ 16 cm from chest wall Collimator edge for 13 cm field chest wall Signal boosted by gain map, not radiation. Area masked in processed image. Gain Depth of Gain Map 19 cm Gain map (schematic) to produce flat 19 cm field Position GE Healthcare 2007 Viewing Conditions Check and Setting • • • • • • Set intended illuminance for interpretation of mammograms. Darken displays. Measure illuminance at display screens. For the RWS, recommend reducing ambient illuminance upper limit from 50 to 20 lux. For Seno Advantage, upper limit is 20 lux. Provide “map” enabling RT to reproduce illuminance conditions of your measurement. GE Healthcare 2007 Monitor Calibration • Not a calibration, but a check. Field Engineer sets black / white levels. – • FE performs perceptual linearization. – • Uses internal photometer. FE records 5 baseline luminance levels. – • Uses external, calibrated photometer. Uses internal photometer. During QC, physicist compares lum. levels to baseline. – Log Luminance • L180 L255 L120 L60 L10 0 50 100 150 200 250 Digital Driving Level Uses internal photometer. GE Healthcare 2007 Monitor Calibration • • • For “RWS,” background set to DDL = 180. For Seno Advantage, background set to DDL that gives luminance ~ 20% of luminance for DDL = 255. Encourage QC RT to ensure that Field Engineer leaves baseline luminance values after re-calibration. GE Healthcare 2007 Image Quality–SMPTE Pattern • Be sure to use SMPTE pattern from graphics display driver not Browser or elsewhere. – • Center of pattern must have bar pattern not black and white squares. Examine image for the listed features. GE Healthcare 2007 Display Screen Uniformity • • • • Screen set to DDL = 255. Examine screen for artifacts. During system acceptance When necessary to isolate system artifacts. – • Not an annual test No artifact or non-uniformity that is expected to mimic or obscure clinical information GE Healthcare 2007 Thank you GE Healthcare 2007 Appendix GE Healthcare 2007 DS & Essential: Auto. RRA Automated Repeat / Reject Analysis (RRA) • User classifies each image • Accept (default) • Repeat (extra dose to patient) • Reject (no intention to keep, e.g., QC test) • User assigns cause to each Repeat and Reject • System maintains data base • Repeat, Reject rates posted at user’s request • Export database for off-line analysis Replace paper-based record keeping, hand calculations. GE Healthcare 2007 Automated RRA – Input Classify image • Accept • Repeat • Reject Assign cause Set all images to same class and cause GE Healthcare Enter unlisted cause 2007 Automated RRA – Output Cause # Repeats % Repeats # Rejects % Rejects Exposure Total Repeat, Reject Totals Repeat, Reject Rates GE Healthcare 2007