MQSA Survey Considerations Gary T. Barnes Department of Radiology UAB Medical Center University of Alabama at Birmingham X-Ray Imaging Innovations Birmingham, Alabama OUTLINE • MQSA Regs — Pluses & Minuses • Workflow Considerations • Auditing Processor Performance • Dose Creep • Random Site Survey Observations MQSA Regulations — Strengths • Mandates Routine Quality Control • Assesses Clinical Image Quality • Mandates Physics Audits • Mandates Phantom Image Quality • Limits Breast Dose • Mandates Random Audits MQSA Regs — Weaknesses • Processor Performance Not Addressed • Inadequate AEC Perform Requirements • Inadequate Cassette Exp Uniformity Performance Requirements • Digital QC and Phys Performance Evaluations Left to the Manufacturer • Mandates Annual FDA Audits Inadequate AEC Perform Requirements: • AEC Exp Reproducibility can be ± 5% (should be < 2%) • AEC Tracking is Required for 2 to 6 cm Phantom Thicknesses (should be 2 to 8 cm) Inadequate Cassette Exposure Uniformity Performance Requirements: • Cass Exposure Uniformity is ± 0.30 OD (should be ± 0.15 OD) OUTLINE • MQSA Regs — Pluses & Minuses • Workflow Considerations • Auditing Processor Performance • Dose Creep • Random Site Survey Observations Audit Workflow Considerations • AEC Exposure Reproducibility • Cassette Exposure Uniformity • AEC Tracking vs Phantom Thickness and kVp OUTLINE • MQSA Regs — Pluses & Minuses • Workflow Considerations • Auditing Processor Performance • Dose Creep • Random Site Survey Observations Rationale Subpar processing can result in: • Reduced film and image contrast • Reduced film speed & increased dose • Artifacts and increased image noise Rationale (cont.) • Processing and Processors are a major problem area in mammography • Processing artifacts are addressed in the ACR Mammography QC Manual • Processor performance — film speed and contrast is not addressed by MQSA or ACR Measuring film speed and film contrast (ave gradient) and comparing these values to ref values for the same type of film is the key consideration This can best be done by using a calibrated box of film The ref values can be the averages of the sites audited If one has sites using different types of film, a box is needed for each type Film speed is conventionally defined at net density of 1.00 • This corresponds to gross density of 1.20 • In mammography the important background density range is from to 1.50 to 2.00 • Film speed should be defined at a net density of 1.25 or 1.50 Film ave gradient is conventionally defined between net densities of 0.25 and 2.00 • This corresponds to gross densities of 0.45 and 2.20 • In mammography the important density range is from to 1.00 to 2.50 • Film ave gradient should be defined between net densities of 0.75 and 2.25 Additional information and insight can be obtained by measuring: • Developer & fixer specific gravity (both processor & replen tanks) • Developer & fixer pH • Hypo retention and dark room fog Processor Audit Form Screen-Film System: Processor Mfr . Model . Chemistry Mfr Processing Time . Audit Film . Ref Film Speed . Ref Film Ave Grad .Process Dev Sp Grav Replen Dev Sp Grav . Emulsion No. Site Film Speed Site Film Ave Grad Processor Audit Form (cont.) Hypo Retention Dark Rm Fog . Comments: . . . . . . Auditing Processor Performance: • Allows one to make specific recommendations rather than vague observations OUTLINE • MQSA Regs — Pluses & Minuses • Workflow Considerations • Auditing Processor Performance • Dose Creep • Random Site Survey Observations • Ten years ago ACR phantom doses were typically in the 125 to 175 mrad range • Currently phantom doses are typically in the 175 to 225 mrad range Two principle reasons for this dose creep are: 1. ACR Phantom and clinical film densities are higher today than ten years ago 2. Ten years ago deep tank processors were commonly used for mammography. Today shallow tanks processors are more common ACR Phantom Dose vs Film Density 350 Increasing ACR Phantom Film Density from 1.60 to 2.00 Increases Phantom Dose by > 20% 250 150 50 1.0 1.5 2.0 Film Optical Density 2.5 3.0 Deep vs Shallow Tank Processor Speed 105 100 Deep Tank Processors are 10% faster than Shallow Tank Processors Deep Tank Shallow Tank 95 90 85 0 1 2 3 Sites 4 5 6 Additional Factors that have been found to contribute to Dose Creep at individual sites • • • • • Subpar chemistry performance Higher than acceptable grid Bucky factors Slower than typical emulsion speed kVp may be low and result in increased mAs Low rad output and loss of film speed due to reciprocity law failure at long exposure times Kodak RP vs Brand X Chemistry (Kodak Shallow Tank Processors) 100 Kodak RP Developer Soln is 18% faster than Brand X 90 Kodak RP Dev Brand X Dev 80 70 0 1 2 3 Sites 4 5 Bucky Factors of 15 Grids 3 Range 2.24 — 2.83 Average 2.62 Std Dev 0.15 2.5 2 18 x 24 grids 24 x 30 grids Emulsion Speed Variations • Emulsion speed variation of ± 5 % are typical • Greater speed variations can occur Effect of kVp on Dose • Decreasing the kVp by 1 kV increases ACR Phantom Dose by 6.5% • If the measured kVp is 1 kV less than the nominal 25 kVp, ACR Phantom Dose is 7% greater than it would be if there was no error in the measured kVp Effect of Radiation Output on Dose • Depending on the mammo unit, radiation output varies from 1000 to 2000 mR/sec, i.e., a factor of 2.0x in exposure time • A factor of 2.0x longer exposure time will increase ACR Phantom Dose by 4% due to film reciprocity failure with Min-R 2000 Dose Creep • • Dose creep occurs in mammography • ACR Phantom doses 250 mrad need to be investigated by the medical physicist to determine the cause or causes One or more factors can result in undesirably high ACR Phantom Doses OUTLINE • MQSA Regs — Pluses & Minuses • Workflow Considerations • Auditing Processor Performance • Dose Creep • Random Site Survey Observations ACR Random Site Audits: • ACR mandated to perform a limited number of surveys each year • All states visited in a cycle • Audit team consists of a radiologist, medical physicist and ACR staff mammo technologist • Purpose is primarily educational and the goal is to be of assistance to the sites Physicist Role in Random Site Audits: • Review physicist’s reports • Review technologist’s QC records • Acquire phantom image with TLD • Evaluate phantom image quality • Review technologist’s phantom scoring • Review results with site General Classification of Problems: • Evaluation of unit • Discrepancies in report • Artifacts • Image quality • Technologist related Unit Evaluation: • Unit assembly evaluation - Compression paddle loose at site visit, no comment in report - Grid assembly loose at site visit, no comment in report - Breast thickness off by > 0.5 cm in report, not reflected in summary - Power off not tested - Auto compression release not tested Unit Evaluation (cont.): • Collimation Assessment - Not in report - Values in report not in regulatory compliance, not reflected in summary • Half Value Layer (HVL) - HVL not checked for all targets - HVL not checked over range of use Unit Evaluation (cont.): • Automatic Exposure Control (AEC) - Insufficient check of AEC modes - 8 cm not checked - Significant density difference observed between 18 x 24 and 24 x 30, no comment - Modes used for physics evaluation are different than used in clinical practice Artifacts: • Grid artifacts noticeable on phantom image obtained during site visit, no comment • Artifacts present on phantom QC images, no comment • Rh target artifact noted in data sheets, no comment in summary Image Quality: • Technique of 28 kVp for 2 cm breast • Magnification technique, long exposure time • Physics phantom technique different than used at site • Phantom images noisy • Marginal phantom image quality • Bad phantom Physics Report: • Unit identification not noted • Viewbox luminance not noted • Summary of recommendations not given • Problems indentified in data sheets are not reflected in summary • Discrepancies present between data sheets and summary observations/recommendations Technologist QC Program - Phantom: • Improper scoring of phantom • Phantom artifacts present - no follow up • Disk density difference outside of ± 0.05 • Difference in phantom and clinical technique • Marginal phantom images • Phantom OD too low or too high Technologist QC Program - Processor: • Crossover not in records • Error in performing crossover • QC ref values not reset when necessary • QC program spotty • Dark room fog excessive • Relative processor speed significantly low Technologist Related: • Noticeable dust present • Cassette ID not clear • Cassette storage • Densitometer not functioning correctly • OD of dense glandular areas < 1.0 OD OUTLINE • MQSA Regs — Pluses & Minuses • Workflow Considerations • Auditing Processor Performance • Dose Creep • Random Site Survey Observations Conclusions • Physicist have made and continue to make important contributions • Processor performance should be audited • More care should be taken in conducting audits and preparing reports • Problems found should be communicated • Greater care should be taken in reviewing the technologist’s QC program