X-Ray Microscopy System with an Electronic Zooming Tube K. Shinohara1, 2, A. Ito3, H. Nakano2, T. Honda4, K. Yada5 1 Radiation Research Institute, Faculty of Medicine, The University of Tokyo, Tokyo 113, Japan, E-mail: kshino@m.u-tokyo.ac.jp 2 Department of Radiation Research, Tokyo Metropolitan Institute of Medical Science, Tokyo 113, Japan 3 Department of Nuclear Engineering, School of Engineering, Tokai University, Kanagawa 259-12, Japan 4 Department of Image Science, Faculty of Engineering, Chiba University, Chiba 263, Japan 5 Aomori Public College, Aomori 030, Japan Abstract. An X-ray microscopy system is presented using an electronic zooming tube as a two dimensional X-ray detector based on previous works on the X-ray contact microscopy of dried human cells in a wide wavelength range of X-rays and on the X-ray holographic microscopy of dried human cells. The system is designed to work for a contact microscopy and for a holographic microscopy, and is arranged as a user-friendly system. The system was tested for a resolution, X-ray attenuation curve by aluminum, and efficiency of the detection of photons at the energy of 2472.7 eV. The results showed that the resolution was about 1 µm, that photon counting method was suitable for a wide dynamic range, and that the efficiency of the system was 2.6x10-3. Imaging phosphorus in a cell was successful by subtraction of images at the energy on (2152.8 eV) the peak in XANES profile of DNA and below (2145.3 eV) K-absorption edge of phosphorus. The present results suggest that the system is successful but remains to be improved. 1 Introduction Electronic zooming tube is a two dimensional detector of soft X-rays with a resolution higher than 0.5 µm [1]. The detector may be directly applicable to an X-ray microscopy system as an X-ray contact microscopy or an X-ray holography with the following features: (1) No optical elements are required for imaging specimens at the resolution of the detector; (2) Images obtained with different wavelengths can be directly compared; and (3) the detector is applicable to a wide wavelength range of X-rays even shorter than 1 nm down to 0.1 nm [1]. With an electronic zooming tube, we have succeeded in X-ray contact microscopy of dried human cells with various wavelengths of X-rays from 1.5 - 10 nm which include K-absorption edges of carbon, nitrogen and oxygen, and L-absorption edges of iron, calcium and sulfur [2]. The data were analyzed for the relative fraction of elements in local areas of cells. The results suggest that the system may be applicable to imagines of chemical natures in biomolecules using a specific absorption peak in I - 170 K. Shinohara et al. XANES profiles. It has been demonstrated that the detector is also applicable to an Xray holographic microscopy in combination with coherent X-rays [3]. In the present communication, a new type of X-ray microscopy system with an electronic zooming tube is presented based on the experimental results mentioned above. The system is designed to work for an X-ray contact microscopy and for an Xray holographic microscopy, and is arranged as a user-friendly system. 2 X-Ray Microscopy System The major modifications had been performed at the photocathode part of an electronic zooming tube. Figure 1 illustrates a structure of a specimen holder and holder heads (one for contact microscopy and the other for holographic microscopy). The photocathode was arranged to be easy mounting system as a part of the specimen holder. Fig. 1. Specimen holder (a), and holder heads for contact microscopy (b) and holography (c). X-Ray Microscopy System with an Electronic Zooming Tube I - 171 3 Test of the System 3.1 Resolution Resolution of the present system was studied at the photon energy of 2472.7 eV with the edge pattern of EM-grid made of copper (Fig. 2). The resolution defined as the distance to show 10 - 90 % in the intensity profile of edge pattern was 0.98 µm when 100 nm thick gold film was used as photocathode and the magnification of the system was 210 corresponding to 0.246 µm/pixel. Fig. 2. Resolution of the present system. Profile of relative photon intensity along the gray line in the inserted picture of EM-grid was presented. The distance between two vertical white lines indicates the resolution. 3.2 Efficiency of the Detection of Photons Beam size was restricted by a pinhole of 0.1 mm in diameter. Then the total photons were detected by a silicon photodiode (AXUV-100, International Radiation Detectors). In this condition, total beam area was covered by the photocathode in the electronic zooming tube. Therefore, the total photons obtained by the electronic zooming tube was estimated by summing up the photons detected in each pixel. The measurement was performed at the photon energy of 2472.7 eV. For example, total photons detected by the zooming tube operated at photon counting mode was 2.09 photons/mA ring current/sec under the condition that incident X-rays were attenuated by aluminum of 14 µm thickness. At the same time, total current in photodiode was 8.81x10-5 nA/mA ring current corresponding to 807.4 photons/mA ring current/sec because the conversion factor of photodiode was 1 electron/3.63 eV from IRD (International Radiation Detector) data sheet. Therefore, the efficiency was 2.6x10-3. Similar efficiency was obtained when different photon fluxes were used by inserting aluminum of various thickness. I - 172 K. Shinohara et al. 3.3 X-Ray Attenuation by Aluminum Figure 3 shows the relative intensity of transmitted X-rays at the energy of 2472.7 eV with respect to the thickness of aluminum. The results by photodiode were well coincided with those by the electronic zooming tube when the data were obtained by the photon counting method. On the other hand, the data was removed from the expected line of attenuation curve within 2 orders of magnitude when the integration mode was used. The results strongly recommended the use of photon counting method for the analysis of data with a wide dynamic range. Fig. 3. Attenuation curve of X-rays by aluminum sheet. Transmitted X-rays through 1 mmφ or 0.1 mmφ pinhole were measured . Intensity was normalized at that of aluminum. 4 Imaging Phosphorus in a Cell Phosphorus distribution in a cell was obtained by dividing a cellular image on the absorption peak in XANES of DNA at the phosphorus K-absorption edge by that below the absorption peak. XANES profile of DNA is shown in another paper by us in this volume. HeLa cells were cultured on the reverse side of SiN thin film and dried. On the front side of SiN Au was then coated with the thickness of 100 nm. Photon energies were adopted 2152.8 eV for the energy on the absorption peak and 2145.3 eV for that below the absorption peak. Panel (a) and (b) in Fig. 4 show these images of cells. Phosphorus distribution was obtained as a ratio image shown in panel (c). This picture clearly indicated the preferential distribution in nuclear areas in cells. 5 Conclusions 1) Operation of the system, and handling of specimens and photocathodes were greatly improved. 2) Phosphorus mapping of cells shows the preferential location of phosphorus in a nucleus. 3) The system needs further improvement of dynamic range, resolution and efficiency to study detailed difference in the local distribution of elements in a cell. X-Ray Microscopy System with an Electronic Zooming Tube I - 173 Fig. 4. Images of HeLa cells. (a) On the absorption peak (2152.8 eV), (b) below the peak (2152.3 eV), (c) Ratio image. Acknowledgments This work was performed under the approval of the Photon Factory Advisory Committee (Proposal nos. 93G319, 95G283 and 95G284), and partly supported by a Grant-in-Aid for Scientific Research (A) from the Ministry of Education, Science, Sports and Culture. References 1 K. Kinoshita, T. Matsumura, Y. Inagaki, N. Hirai, M. Sugiyama, H. Kihara, N. Watanabe, and Y. Shimanuki, Proc. SPIE 1741, 287 (1992). 2 A. Ito, K. Shinohara, H. Nakano, T. Matsumura, and K. Kinoshita, J. Microsc. 181, 54 (1996). 3 K. Shinohara, A. Ito, H. Nakano, I. Kodama, T. Honda, T. Matsumura, and K. Kinoshita, J. Synchrotron Rad. 3, 35 (1996).