Electronic Brachytherapy Quality Assurance Jessica Hiatt MS ACMP, Virginia Beach, VA May 2, 2009 Outline: • • • • • • Overview and intro to EB Dose characteristics Acceptance, commissioning and QA Breast Trial Endometrial Trial Future directions Electronic Brachytherapy: • Internal radiation therapy that involves placing a miniature x-ray source inside the patient close to or in the tumor. X-Ray Tube Xoft Tube (2006) HV Cable X-ray Source • Tube diameter = 2.25 mm • Requires a cooling catheter – Assembly diameter = 5.4 mm • Nominal dose rate: – 0.6 Gy/min at 3 cm in water Source Characteristics: Comparison of Dose Rate vs. Depth in Water for Various Sources 1.E+03 50 kV 1.E+02 50 kV MCNP5 Pd 103 I 125 Dose Rate (cGy/min) 1.E+01 Ir 192 1.E+00 1.E-01 1.E-02 1.E-03 1.E-04 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 Radius (cm) Dose rate curve is quite similar to 192Ir over region of interest Graph courtesy of Xoft Intracavity APBI: MammoSite HDR 192Ir Xoft 50kV Dickler et al. Brachytherapy, 2007 Axxent Controller: Touch Screen Display Barcode Reader Controller Pullback Arm USB Port Well Chamber SI Max 4000 Electrometer Breast Applicator: Balloon Inflation Valve Multi-lumen Flexible Shaft Drainage Holes Radiation Source Lumen Drainage Port Valves Vaginal Applicator Set: • FDA clearance received in May 2008 for a set of reusable vaginal applicators • Available in four diameters – 20, 25, 30 and 35 mm • Each set contains four vaginal cylinders, four source channels and a board & clamp assembly • Enhanced visibility with CT and fluoroscopic imaging Vaginal Cylinder: EB IB EB for Skin: • Applicator development of 10mm, 20mm, 35mm, 50 mm • Stainless Steel: • Easily sterilizable • Applicator cone and source channel • Flattening filter integrated in cone Slide provided by Xoft EB for Skin: • Developed Axxent Applicator compatible with system for typical lesions (up to 50 mm diameter of target area with 4 sizes) • Circular • Reusable Scalp treatment with Leipzig applicator 10, 20, 30 mm sizes Slide provided by Xoft Commissioning and QA: AAPM TG-182 Recommendations on Electronic Brachytherapy Quality Management Chair: Bruce Thomadsen IntraBeam Axxent Clinical Michael Ringor Chris Stacey Vendor/Vendor Clinical Frank Weigand Gene Cardarelli Mark Rivard Jessica Hiatt Randall Holt Ben Smith Calibration/QA Tina Pike Larry DeWerd Acceptance and Commissioning: • • • • • • • • Well-chamber constancy Beam stability Source positional accuracy Output stability Timer linearity Marker/source position coincidence Controller functionality/safety interlocks Treatment planning data verification Well-chamber Constancy: • Intercomparison - Well-chamber/Electrometer – SI HDR Plus Well Chamber – SI Max 4000 electrometer – 4 mg Ra-eq Cs-137 source Well Chamber Constancy Check RIH Unit Readings (pA) 41.928 41.892 41.885 Percent Difference: -1.89 41.944 41.967 41.976 Percent Difference -0.73 Xoft Unit Readings (pA) Beam Stability: • Simulated treatment data extracted from controller and plotted. Source Positional Accuracy : QA Test Fixture Digitized image of film positioned w/in fixture during exposure Output Stability : Source Stability Measurements Rate (nA) Trial Number Dwell Position (cm) 1 2 3 4 5 Ave. SD Rel. SD 24.5 0.006 0.006 0.006 0.006 0.006 0.006 0.000 0.00 23.5 0.022 0.022 0.022 0.022 0.022 0.022 0.000 0.00 22.5 0.110 0.112 0.111 0.111 0.111 0.111 0.001 0.01 21.5 0.190 0.199 0.199 0.200 0.199 0.197 0.004 0.02 20.5 0.077 0.079 0.080 0.080 0.080 0.079 0.001 0.02 Total Rate (nA) 0.405 0.418 0.418 0.418 0.418 0.416 0.006 0.01 Timer Linearity: Marker/Source Position Coincidence: Controller Functionality and Safety Interlocks: • • • • • • • Setup and delivery Status indicator light Emergency-off Treatment recovery Applicator symmetry Pullback force error No travel through bent applicator Treatment Planning Data Verification: Jessica Hiatt Daily QA: Daily QA: Pre-treatment checklist: Monthly QA: EB QA Conclusion: • EB offers the potential to profoundly impact brachytherapy practice • As of this moment, institutions need to form their own QA procedures • AAPM TG-182 will brainstorm recommendations for EB QA