Dose Considerations in DR S. Jeff Shepard, MS, DABR, FAAPM Imaging Physics Department Diagnostic Imaging Division The University of Texas M. D. Anderson Cancer Center Houston, Texas AAPM Annual Meeting 2007 MDACC Imaging Physics Acknowledgements: A. Kyle Jones, PhD Rai Polman Outline Dose Considerations • Noise, not optical density for appropriateness • Exposure creep • Exposure Indices • Rules for repeats • TG116 and IEC WG43 MDACC Imaging Physics 2 Dose Considerations Recognizing Bad Images • In the filmfilm-screen world, underunder- and overover- exposures were easily recognized by the appearance of the recorded image – Repeat at a higher or lower technique based on optical density MDACC Imaging Physics 3 1 Dose Considerations Recognizing Bad Images • With Digital Radiography, underunder- and overover-exposures are not so easily recognized – Adequate images of a much wider range of exposure – Excellent dynamic range may have downdown-sides MDACC Imaging Physics 4 Dose Considerations MDACC Imaging Physics 5 Dose Considerations MDACC Imaging Physics 6 2 Dose Considerations “Exposure creep” creep” • UnderUnder-exposure – Higher noise – Detail visibility suffers • OverOver-Exposure – Lower noise (“ (“Pretty” Pretty” – improved SNR) – High patient dose • Radiologists may complain about noise, not usually about overover-exposure – Technologists learn quickly how to avoid criticism – If no one is paying attention, exposures will “creep” creep” up. MDACC Imaging Physics 7 Dose Considerations Widely known problem that’ that’s been around for a long time • Freedman M, Pe E, Mun SK, Lo SCB, Nelson M, “The potential for unnecessary patient exposure from the use of storage phosphor imaging systems,” systems,” SPIE 1897:4721897:472-479 (1993). • Gur D, Fuhman CR, Feist JH, Slifko R, Peace B, “Natural migration to a higher dose in CR imaging,” imaging,” Proc Eighth European Congress of Radiology, Vienna Sep 1212-17, 154 (1993). MDACC Imaging Physics 8 Dose Considerations Solution – Exposure Index • Numerical indicator related to dose to detector (noise content in the image) • QC programs based on exposure indices are successful – Seibert JA, Shelton DK, and Moore EH, “Computed Radiography XX-ray Exposure Trends,” Trends,” Academic Radiology 3, 313313-318 (1996). • Not all manufacturers provide indices. • No standardization MDACC Imaging Physics 9 3 Dose Considerations Approximate Exposure Indicator Values vs. Receptor Exposure Manufacturer Symbol 50 µGy 100 µGy 200 µGy REX 50 100 200 Canon (Brightness = 16, Contrast = 10) IDC (ST = 200) f# -1 0 1 Philips EI 200 100 50 Fuji, Konica S 400 200 100 Kodak (CR, STD) EI 1700 2000 2300 Seimens EI 500 1000 2000 10 MDACC Imaging Physics Dose Considerations Approximate Exposure Indicator Values vs. Receptor Exposure Manufacturer Symbol 50 µGy 100 µGy 200 µGy REX 50 100 200 Canon (Brightness = 16, Contrast = 10) IDC (ST = 200) f# -1 0 1 Philips EI 200 100 50 Fuji, Konica S 400 200 100 Kodak (CR, STD) EI 1700 2000 2300 Seimens EI 500 1000 2000 MDACC Imaging Physics 11 Dose Considerations AAPM TG116 • Recommendation for a standard detector exposure index for all digital radiography • Index (KIND) is the air Kerma that the detector would have received under standard beam conditions for the same raw pixel value • Standard beam – – – 70 + 4 kVp 0.5 mmCu + [0 – 3] mmAl filtration 6.8 mm (+ (+ 0.2 mm) Al HVL • Calculated for every image from the forfor-processing pixel value in the VOI recognized by the system • Final approval expected by RSNA MDACC Imaging Physics 12 4 Dose Considerations AAPM TG116 also calls for a Deviation Index DI = 10 × Log10{KIND/KTGT(b,v)} • KTGT(b,v) (b,v) is a table of target KIND values stored by body part (b) and view (v) DI = 0 is a perfect exposure DIL = +1 means exposure was high/low by about 28% (one density or mAs step) • KTGT tables to be customized for each site • Format: – TG116: Decimal string with one decimal place – WG43: Integer • Both indices saved in the DICOM header (Tag TBD) • Both indices change with VOI modification by the tech 13 MDACC Imaging Physics Dose Considerations Number of Pixels Both indices change with VOI modification by the tech KIND and DI calculated from this pixel value KIND = KTGT DI = 0.0 Values of Interest Pixel Value 14 MDACC Imaging Physics Dose Considerations VOI recognition algorithm fails Number of Pixels • Gonadal shields, prosthetics, surgical mods • False DI reported KIND ≠ KTGT KIND and DI incorrectly calculated from this pixel value DI = -1.3 Correct Values of Interest Incorrect Values of Interest MDACC Imaging Physics Pixel Value 15 5 Dose Considerations Tech returns VOI to proper position manually Number of Pixels KIND and DI calculated from this pixel value KIND = KTGT DI = 0.0 Correct Values of Interest Incorrect Values of Interest Pixel Value 16 MDACC Imaging Physics Dose Considerations International Electrotechnical Commission (IEC) standard • • • Subcommittee 62b, WG43 Definition only Current draft parallels AAPM TG116 – EI: A unitless value (KREC * 100 µGy-1) • Intent is to avoid confusion with patient exposure – Similar standard beam – Deviation Index (DI) same as DI (integer value vs decimal string with one decimal) • Committee draft for vote (CDV) expected this fall 17 MDACC Imaging Physics Dose Considerations AAPM TG116 IEC WG43 (SC 62b) Exposure Index AirAir-kerma at the receptor KIND = Krec (uGy) uGy) EI = Krec *100 µGy-1 (unitless) unitless) Calibration Energy RQARQA-5 Equivalent RQARQA-5 Equivalent 70 + 4 kVp 70 + 4 kVp Calibration Filtration RQARQA-5 Equivalent RQARQA-5 Equivalent 0.5 mm Cu (+ 00-3 mmAl) mmAl) 6.8 + 0.2 mm A1 HVL or 21 mm Al 0.5 mm Cu + 2 mm Al or 21 mm Al 6.8 + 0.3 mm A1 HVL Deviation Index Deviation Index DI = 10*log10(KIND/KTGT) Deviation Index DI = 10*log10(EI/ETGT) Index format Signed decimal string with 3 significant Fig’ Fig’s Unspecified 6 Dose Considerations What the index is NOT for: • Patient dose estimation – Need beam HVL, output, pt. thickness, SSD, SID grid atten, atten, AEC pickup atten, atten, detector input atten – If you have all these, you don’ don’t need KIND! • System intercomparisons – Index says nothing about detector energy dependence or efficiency MDACC Imaging Physics 19 Dose Considerations Rules for repeats • Little clinical information on reject thresholds in literature – Van Metter and Yorkston • Chest and abdominal imaging • Wide range of patient thicknesses • Most AEC controlled images fell within the range of DI = ± 1.2. MDACC Imaging Physics 20 Dose Considerations Rules for repeats • Emulating film/screen limits ⎛E ⎞ ⎛E ⎞ ∆OD = λ * Log10 ⎜ 2 ⎟ and ∆DI = 10 * Log10 ⎜ 2 ⎟ E ⎝ 1⎠ ⎝ E1 ⎠ ⎛ ∆OD ⎞ Combining: ∆DI = 10 * ⎜ ⎟ ⎝ γ ⎠ For an OD range of ± 0.3 OD (0.6 OD total) and γ = 2.5 : ⎛ 0.6 ⎞ ∆DI = 10 ∗ ⎜ ⎟ ⎝ 2.5 ⎠ = 2.4 or DI = ±1.2 MDACC Imaging Physics 21 7 Dose Considerations KIND DI Action Between -1.0 and +1.0 79% < exp < 126% < -1.0 < 79% of target Between +1.0 and +3.0 126% – 200% of target > 3.0 > 200% of target Check for VOI recognition failure, accept if ok. Check for noise and consult with radiologist on need for repeat, investigate cause. Check for VOI recognition failure, accept if ok. Investigate cause. Check for saturation and consult with radiologist on need for repeat, investigate cause 22 MDACC Imaging Physics Dose Considerations Exposure Index monitoring • Collect KIND or DI for every image and analyze – – – – – – by tech technique factors by xx-ray system by plate scanning unit (CR) by processing unit (CR) by anatomical view • Longitudinal studies – Track performance over time – Mean & Standard Deviation – Watch for trends upward (“ (“Creep” Creep”) and increasing std. dev. 23 MDACC Imaging Physics S-Number Histogram DI-D Dose Considerations Exposure Index Tracking 300 800 250 700 • Use histogram analysis to illustrate problem areas Frequency 600 Mean = 187.5 Std Dev = 158.5 500 150 400 Mean 300 Std Dev 100 200 100 S MDACC Imaging Physics 00 00 00 50 30 10 0 0 80 90 0 0 70 0 Nov-06 Month 60 50 0 0 0 30 Oct-06 40 20 0 Sep-06 0 0 0 50 10 Index Value 200 Dec-06 24 8 Dose Considerations Exposure Index monitoring • Many systems have no index at all • Many systems do not log the indices • For those that do, – multiple machines with logs that are not accessible over the network – inconsistent file formats • Must be performed manually – Paper logs – Manual electronic log downdown-load – Manual conversion to common file format Impractical! Too much work! MDACC Imaging Physics 25 Dose Considerations Exposure Index monitoring • HomeHome-made solution: – A batch file on every IIP runs every night and copies the Fuji SS-number log file to a location that is accessible to their FTP service – Unix program contacts each IIP and extracts the log file (ftp), cleans up the data and parses it into a database. MDACC Imaging Physics 26 Dose Considerations Data from any month can be downdownloaded over the intranet for analysis MDACC Imaging Physics 27 9 Dose Considerations AEC calibration and technique charts are still important in controlling patient exposure • AEC may need to be adjusted for CR and addadd-on DR – Energy dependence may not be the same as GdO2S • Technique charts are important for portables and extremities MDACC Imaging Physics 28 Summary Dose Considerations • Noise, not density • Exposure creep • Exposure indices – AAPM TG116 – IEC WG43 • Exposure index logs • Rules for repeats AAPM Radiographic and Fluoroscopic SubSub-Committee is always open to new ideas MDACC Imaging Physics 29 THANK YOU! jshepard@di.mdanderson.org 10 References: Fredman M, Pe E, Mun SK, Lo SCB, Nelson M, The potential for unnecessary patient exposure from the use of storage phosphor imaging systems, SPIE 1897:4721897:472-479 (1993). Gur D, Fuhman CR, Feist JH, Slifko R, Peace B, Natural migration to a higher dose in CR imaging, Proc. Eighth European Congress of Radiology, Vienna,Sep 1212-17, 154 (1993). Yorkston J. FlatFlat-panel DR detectors for radiography and fluoroscopy. In: Specifications, Performance Evaluations, and Quality Assurance of Radiographic and Fluoroscopic Systems in the Digital Era, Goldman LW and Yester MV eds. Madison, WI: Medical Physics Publishing (2004)177(2004)177-228. MDACC Imaging Physics 31 References: Willis CE, Thompson SK and Shepard SJ. Artifacts and Misadventures in Digital Radiography. Applied Radiology 33(1):1133(1):11-20, January 2004. R.E. Alvarez, J.A. Seibert, and S. K. Thompson, Comparison of dual energy detector system performance, Medical Physics31(3), 556556-565 (2004). JA Seibert, DK Shelton, and EH Moore, Computed Radiography XX-ray Exposure Trends, Academic Radiology 3, 313313-318 (1996). J A Seibert, et al, AAPM Report #93, “Acceptance Testing and Quality Control of Photostimulable Storage Phosphor Imaging Systems: Report of AAPM Task Group 10.” 10.” AAPM (2006) MDACC Imaging Physics 32 References: Richard S., Siewerdsen J. H., Jaffray D. A., Moseley D. J. and Bakhtiar B., Generalized DQE analysis of radiographic and dualdual-energy imaging using flatflatpanel detectors, Med. Phys. 32 (5), May 2005, 1397 – 1415 Lehman L. A., Alvarez R. A., Macovski A., Brody W. R., Pelc N. J., Reiderer S. J., and Hall A., Generalized image combinations in dual KVP digital radiography, Med. Phys. 8 (5), Sept/Oct 1981, 659 – 667 MDACC Imaging Physics 33 11