Wednesday Case of the Day Category: Physics Sara C. Gavenonis MD, Andrew D.A. Maidment PhD Hospital of the University of Pennsylvania, Philadelphia, PA History: 45 year-old female enrolled in an imaging research study. A B C 1. There is an increase in noise between A and B. What radiographic factor would most directly account for this difference? 2. There is a decrease in subject contrast between B and C. What radiographic factors could account for this difference? Question #1 Discussion: The x-ray tube current (mAs) directly controls the noise seen in an x-ray or CT image. A B Figure 2: These two images show the effect of reducing the tube current (mAs) on image quality. A) This image was acquired at 28 kVp with a molybdenum x-ray target and a rhodium filter. The image was acquired at 48 mAs. B) This image was acquired at the same kVp, target and filter as image A. However, the image was acquired at 16 mAs, or one-third the dose. Digital breast tomosynthesis will be discussed in this exhibit. Please note that as of 11-9-2010, digital breast tomosynthesis is not FDA approved. Question #1 Discussion: The noise in an image is determined by the dose. The most direct determinant of radiographic dose is the x-ray tube current or mAs. 28 kVp 8 mAs 0.13 mGy 28 kVp 70 mAs 1.13 mGy 28 kVp 220 mAs 3.39 mGy Figure 3: A mammographic phantom is shown imaged at 3 different mAs values. As the mAs (and hence dose) increases, the number of objects visible in the phantom increases too. The typical diagnostic technique is 70 mAs. Question #1 Discussion: Reduction of tube current will adversely affect the ability to detect fine linear structures such as fibrils, and micro-calcifications (arrow). In general, large objects are not affected (double arrow) Figure 4: A normal mAs (left) and low mAs (right) image of a calcification cluster. Question #1 Discussion: This patient was enrolled in a research study in which digital mammography and digital breast tomosynthesis exams were obtained. A B A) Digital mammography image. B) Digital breast tomosynthesis source projection image . Please note that as of 11-9-2010, digital breast tomosynthesis is not FDA approved. Discussion: The source projection images used to create tomosynthesis image reconstructions are acquired at low dose. The total dose of a tomosynthesis image set is comparable to the dose of a mammogram. Since 10-25 source images are used in the reconstruction, each source image is acquired at 1/10th to 1/25th the mammographic dose Figure 5: A tomosynthesis image is acquired as a series of low dose projection images. Here 9 source positions are shown. Discussion: While each tomosynthesis projection image is noisy, the tomosynthesis reconstruction process reduces the noise in the reconstructed image Figure 6: The tomosynthesis source projection image (left) is noisy, limiting the detection of the calcifications. However, the tomosynthesis reconstruction combines the information from many source projection images. Therefore, the resulting reconstruction (right) has low noise and more calcifications are visible. Question #2 Discussion: The contrast of an image is determined by the kVp and target/filter combination. B C Figure 7: These two images show the effect of changing the target material, filtration, and kVp on image quality. B) This image was acquired at 28 kVp with a molybdenum x-ray target and a 0.025 mm thick rhodium filter. B) This image was acquired at 49 kVp with a rhodium x-ray target, a 0.025 mm thick rhodium filter and a 0.3 mm thick copper filter. The dose to the breast in these two images is approximately the same. Question #2 Discussion: Changing the kVp or target/filter combination will alter the contrast of an image. 24 kVp 185 mAs 1.2 mGy 28 kVp 71 mAs 1.2 mGy 34 kVp 30 mAs 1.2 mGy Figure 8: A mammographic phantom is shown imaged at 3 different kVp values. As the kVp increases, the mAs was decreased to keep dose constant. The optimal image is at 28 kVp. The 24 kVp image is noisy because the number of x-rays transmitted through the phantom is reduced. The 34 kVp image is noisy because the x-ray subject contrast is lower; to get sufficient display contrast, the window width was reduced which amplifies the appearance of the noise. Discussion: High energy (high kVp) x-ray images of the breast are being used in clinical trials for contrast-enhanced digital mammography and contrast-enhanced digital breast tomosynthesis. HE HE 10000 1800000 1600000 1000 1400000 K-edge 1000000 Iodine 800000 10 600000 Breast HE spectrum 1 Pre contrast 1200000 100 400000 Post contrast 200000 0.1 0 0 10 20 keV 30 40 50 Time Figure 9: The high energies emphasize the attenuation of iodine, increasing the contrast of enhancing lesions. However, high energy images do not preserve breast tissue contrast. # Photons Mass Attenuation [cm2/g] Iodine contrast Summary: The x-ray spectrum determines the contrast and noise seen in radiographic and tomographic images, as well as the dose to the patient. The x-ray spectrum is controlled by the tube current (mAs), tube voltage (kVp) and target material, and the filtration added to the x-ray beam. References: 1) JT Bushberg, et al. The Essential Physics of Medical Imaging, 2nd Edition. 2) LT Niklason, et al. Digital tomosynthesis in breast imaging. Radiology 1997; 205:399-406. 3) SC Chen, et al. Initial clinical experience with contrast-enhanced digital breast tomosynthesis. Academic Radiology 2007; 14: 229-238. Acknowledgements: The clinical images shown in this exhibit were processed and reconstructed using the AdaraGPU™ and Briona™ software from Real Time Tomography, LLC. (Villanova, PA) and have been provided courtesy of Susan Ng.