UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE Valid for: 2 years from approval Table of Contents Page 1. 2. 3. 4. 5. 6. 7. 8. 9. Scope Objective Definitions Responsibility Procedure 5.1. Installation 5.2. Configuration 5.3. Operation 5.4. Calibration 5.5. Troubleshooting 5.6. Maintenance 5.7. Flowcharts Precautions References Appendices Revision History 2 2 2 3 3 3 4 4 4 5 6 7 12 12 12 12 Page 1 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE Valid for: 2 years from approval 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the AA® Series AA-160 Precision Balance in the laboratories of the Pharmacy Department, University of Malta. 2. Objective To describe the procedure for the installation, configuration, operation, calibration, maintenance and troubleshooting of the AA® Series AA-160 Precision Balance. 3. Definitions 3.1. Adjustable Feet: Circular screws situated at the bottom of the balance that are manipulated to level the balance out. 3.2. Auto CAL Key: To initiate automatic calibration using internal weights. 3.3. Calibration: A procedure to check a measuring instrument against an accurate standard to determine any deviations and correct for errors. 3.4. Calibration Variance: The corrections made to the calibration values to compensate for differences between the actual weight being used and the desired value. 3.5. Configuration Parameters Set-Up Menu: Allows the user to adapt the balance to different weighing situations (see SOP/PD/315_02/A1). 3.6. Message Display: Part of the dual LCD display that uses text to provide additional information. 3.7. Number Display: Part of the dual LCD display that continuously shows the weighing results. 3.8. On/Off Key: To turn the display ON or OFF or to enter the desired option in the Set-Up menu. 3.9. Print Key: To transmit sample weight to printer or computer. Is also used to scroll through the different options in the Set-Up Print menu. Page 2 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE Valid for: 2 years from approval 3.10. Select Key: To scroll through the configuration parameters of a particular option in the Configuration Parameters Set-Up menu. 3.11. Set-Up Key: To access the Configuration Parameters Set-Up menu and then to cycle through its options. Is also used to access the Set-Up Print menu. 3.12. Tare: Allows the balance to display the weight of the sample only and not the combined weight of the sample and its container. 3.13. Tare Key: To set the Numeric Display to zero. Is also used to exit SetUp menu and return the balance to normal weighing mode. 3.14. U symbol: An unstable indicator that is automatically displayed whenever the balance is not stable and then disappears when the balance becomes stable. 3.15. Weighing Pan: The area on top of which the material to be weighed is placed. 4. Responsibility 4.1. The members of the Department of Pharmacy (staff and students) are responsible for following this SOP. 4.2. The designated Laboratory Officer or Laboratory Assistant is responsible for ensuring that this SOP is followed. 5. Procedure 5.1. Installation 5.1.1. 5.1.2. 5.1.3. 5.1.4. 5.1.5. Place the balance on a stable surface, away from any vibrations. Turn the adjustable feet at the bottom of the balance clockwise or anticlockwise until the air bubble in the spirit level is in the centre of the marked circle. Insert power supply connector in the plug on the rear of the balance and plug the power supply into the mains supply. Press the On/Off key to switch on the balance. Wait until the start-up message and then 0.0000 are displayed before starting to operate the balance. Page 3 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE 5.2. Configuration 5.2.1. 5.2.2. 5.2.3. 5.2.4. 5.3. Valid for: 2 years from approval Press the Set-Up key briefly but repeatedly until the desired option is displayed (see SOP/PD/315_02/A1 – Configuration Parameters Set-Up Table). Press the On/Off key to enter into the desired option. Press the SELECT key briefly but repeatedly until the desired configuration parameter is displayed. Press the TARE key to save changes and return the balance to normal weighing mode. Operation 5.3.1. 5.3.2. 5.3.3. 5.3.4. Place the material that needs to be weighed on the weighing pan. Close the draft shield. Wait until the [U] symbol disappears. Read off the value from the display. If using the tare function: 5.3.5. 5.3.6. 5.3.7. 5.3.8. 5.3.9. Place an empty weighing container on the weighing pan. Close the draft shield. Wait until the [U] symbol disappears. Press the Tare key. Open draft shield and place the material to be weighed in the weighing container. 5.3.10. Close draft shield. 5.3.11. Wait until the [U] symbol disappears. 5.3.12. Read off the value from the display. 5.4. Calibration 5.4.1. Auto Calibration 5.4.1.1. 5.4.1.2. 5.4.1.3. 5.4.1.4. Press the On/Off Key to switch on the balance. Wait until the [U] symbol disappears from the message display. Press the TARE key. Press the AUTO CAL key. Page 4 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE 5.4.1.5. 5.4.2. Valid for: 2 years from approval Wait until the message display changes from [CAL] to [CAL-100-*] to [TARING] and finally to [GRAMS] to indicate that the auto-calibration process is complete. External Calibration 5.4.2.1. 5.4.2.2. Press the On/Off Key to switch on the balance. Wait until the [U] symbol disappears from the message display. 5.4.2.3. Press the TARE key. 5.4.2.4. Place the standard calibration weight on the weighing pan. 5.4.2.5. Wait for the [U] symbol to disappear and the calibration weight value together with [GRAM] to be displayed. 5.4.2.6. Press the AUTO CAL key. 5.4.2.7. Wait until [XXX.0000] and [CAL-XXX-] are displayed. 5.4.2.8. Remove the standard weight. 5.4.2.9. Perform the calibration procedure once again if it is to be repeated with a different standard weight. 5.4.2.10. See SOP/PD/315_02/A2 – Calibration Variance Table if calibration variance needs to be modified. 5.5. Maintenance 5.5.1. 5.5.2. 5.5.3. 5.5.4. Disconnect the balance from its power supply before attempting to clean and leave it in its normal operating position. Use a clean fine brush to remove any surrounding powder. Use a moistened cloth with a mild cleaning agent to clean and remove any spilt material. Ensure that the air bubble in the spirit level on the front end of the scale is in the centre of the marked circle. Page 5 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE 5.6. Valid for: 2 years from approval Troubleshooting Fault - Both displays are blank [OVER] is displayed Possible Cause/s Display previously turned off by the On/Off key - Possible Solution/s Press the On/Off Key - Power cord not connected properly - Attach the power cord accordingly - Power cut or an improper voltage supply present - Check if a proper power supply is present - Temporary fault - Exceeded weighing capacity - Pan is not properly installed Disconnect power cord, wait for at least 5 seconds and reconnect Reduce the weight - Install pan properly [UNDER] is displayed - Pan is obstructed - - Air currents present around balance - - Sample not stationary - [GRAM U] continues to be displayed [NO CAL] is displayed during calibration process Balance capacity is exceeded Page 6 of 14 Move balance to a better location Move balance to a better location or close all draft shields Ensure sample is made stationary Use a 50g, 100g or 150g standard calibration weight to complete calibration UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE 5.7. Flow Charts 5.7.1. Installation Start Place balance on a stable surface, away from any vibrations Turn adjustable feet until air bubble in spirit level is in the centre of the marked circle Insert power supply connector in the plug on rear of balance and plug power supply into mains supply Press On/Off key to switch on balance Wait until start-up message and 0.0000 are displayed before use End Page 7 of 14 Valid for: 2 years from approval UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE 5.7.2. Configuration Start Press Set-Up key briefly but repeatedly until desired option is displayed (see SOP/PD/315_02/A1) Press On/Off key to enter desired option Press SELECT key briefly but repeatedly until desired configuration parameter is displayed Press TARE key to save changes and return balance to normal weighing mode End Page 8 of 14 Valid for: 2 years from approval UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE 5.7.3. Valid for: 2 years from approval Operation Start No Using tare function Yes Place empty weighing container on weighing pan Close draft shield Wait until [U] symbol disappears Press Tare Key Open draft shield and place material to be weighed in weighing container Close draft shield Wait until [U] symbol disappears Read off value from display End Page 9 of 14 Place material on weighing pan UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE 5.7.4. Valid for: 2 years from approval Calibration Start Press On/Off key to switch on balance Wait until [U] symbol disappears Press TARE Key No Perform auto calibration by pressing the AUTO CAL key Performing external calibration Yes Wait until message display changes from [CAL] to [CAL-100-*-] to [TARING] and finally to [GRAMS] Place standard weight on weighing pan Wait for [U] to disappear and calibration weight value together with [GRAM] to be displayed Press AUTO CAL key Wait until [XXX.000] and [CAL-XXX-] are displayed Remove standard weight Repeat calibration with different standard weight Yes Page 10 of 14 No End UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE 5.7.5. Valid for: 2 years from approval Maintenance Start No Balance needs cleaning Yes Disconnect balance from power supply Leave balance in its normal operating position and do not tilt sideways No Any surrounding powder Yes Remove using a clean fine brush No Any spilt material Yes Remove using a moistened cloth with a mild cleaning agent No Air bubble in spirit level is in centre of marked circle Yes End Page 11 of 14 Manipulate adjustable feet until this is achieved UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE Valid for: 2 years from approval 6. Precautions 6.1. 6.2. 6.3. 6.4. 6.5. Do not install balance near extremes of heat and cold. Protect balance from dust and air draughts. Keep weighing pan clean and dry to prevent inaccuracies. Do not move balance during weighing. Ensure that the maximum tolerated weight of 160g is not exceeded during weighing procedures and also when using the tare function. 6.6. Do not leave any items on the balance when this is not in use. 6.7. Only use the balance if the spirit level is correct. 6.8. Adjust the air bubble in the spirit level each time the balance is moved to a different location. 6.9. Avoid static charge buildup of the material to be weighed and of the weighing container. 6.10. When transferring balance to a significantly warmer environment, allow it to acclimatise for approximately 2 hours at room temperature after it has been disconnected from the mains. 6.11. Do not unplug balance as this is designed to be continuously plugged in. 6.12. Ensure that an audible beep is heard whenever one of the keys is pressed if the beeper option is activated. If this beep is not heard, press the key again until the beep is heard. 7. References Denver Instrument Company. AA® Series Electronic Analytical Balances Operating Instructions. 8. Appendices SOP/PD/315_02/A1 – Configuration Parameters Set-Up Table SOP/PD/315_02/A2 – Calibration Variance Table 9. Revision History Version Number 01 02 Page 12 of 14 Amendments/ Reasons for change Initial Release Inclusion of subheading titles of Procedure section in Table of Contents Update of entire SOP to Version 2 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE Valid for: 2 years from approval SOP/PD/315_02/A1 – Configuration Parameters Set-Up Table Option Function CURRENT To display the current selected configuration parameters DEFAULT To reset the balance to the default factory configuration parameters FILTER To choose the desired filter settings from [FT FAST] in ideal conditions, [FT NORMAL] in standard laboratory conditions and [FT SLOW] in areas with drafts and vibrations AUTO ZERO To choose to automatically compensate for zero drift [A-Z ON] or to turn off this correction feature [A-Z OFF] BEEPER To choose between short beeps [BP SHORT], long beeps [BP LONG] or else to turn off the beeper function [BP OFF] Page 13 of 14 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/315_02 STANDARD OPERATING PROCEDURE AA® SERIES AA-160 PRECISION BALANCE Valid for: 2 years from approval SOP/PD/315_02/A2 – Calibration Variance Table Procedure Message Display Press On/Off key to switch on balance Press AUTO CAL key CAL Press SET UP key repeatedly until calibration value to be modified is displayed for e.g. for 100g CAL-100- Press On/Off key V + 00.0 Press SELECT key repeatedly until desired + or – sign is displayed V + 00.0 or V – 00.0 Press On/Off key Press SELECT key repeatedly to change first digit (between 0 and 1) Press On/Off key when desired digit is displayed Press SELECT key repeatedly to change next digit (between 0 and 9) Press On/Off key when desired digit is displayed Repeat the last 4 steps for subsequent digits Page 14 of 14 V +/- x0.0 V +/- 1x.0