UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE Valid for: 2 years from approval Table of Contents Page 1. 2. 3. 4. 5. 6. 7. 8. 9. Scope Objective Definitions Responsibility Procedure 5.1. Diagram of KERN® 120-4NM Electronic Precision Balance 5.2. Installation 5.3. Configuration 5.4. Operation 5.5. Calibration 5.6. Maintenance 5.7. Troubleshooting 5.8. Flow charts Precautions References Appendices Revision History Page 1 of 13 2 2 2 3 3 3 4 4 4 5 5 6 7 12 12 12 12 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE Valid for: 2 years from approval 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students making use of the KERN® 120-4NM Precision Balance in the laboratories of the Pharmacy Department, University of Malta. 2. Objective To describe the procedure for the installation, configuration, operation, calibration, maintenance and troubleshooting of the KERN® EW 120-4NM Precision Balance. 3. Definitions 3.1. Calibration: A procedure performed to check a measuring instrument against an accurate standard to determine any deviations and correct for errors. 3.2. Configuration Parameters: A number of different options that are used to set the balance according to the user’s preferences (see SOP/PD/310_02/A1). 3.3. F Key: To choose the desired function within the configuration parameters. 3.4. Foot Screws: Circular screws situated at the bottom of the balance that are manipulated to level the balance out. 3.5. On/Off Key: To switch the balance on or off. 3.6. Print Key: To print weight value when balance is connected to an external printing device. 3.7. S Key: To save the chosen function in an individual configuration parameter during balance configuration. 3.8. Tare: Allows the balance to display the weight of the sample only and not the combined weight of the sample and its container. Page 2 of 13 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE 3.9. Valid for: 2 years from approval TARE Key: Used to tare the weight of the container during weighing and also to change the function in an individual configuration parameter. 3.10. Weighing Pan: The area on top of which the material to be weighed is placed. 3.11. Weighing Range: Bar graph divided into 20 graphic cuboids with none being displayed when there is no weighing value, 10 displayed when balance is loaded up to one half of its weighing range and all 20 cuboids and F are displayed when balance is loaded to full capacity. 4. Responsibility 4.1. The members of the Department of Pharmacy (staff and students) are responsible for following this SOP. 4.2. The designated Laboratory Officer or Laboratory Assistant is responsible for ensuring that this SOP is followed. 5. Procedure 5.1. Diagram of KERN® 120-4NM Electronic Precision Balance Draft Shield Weighing Pan Spirit Level Display Keyboard Foot Screw Page 3 of 13 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE 5.2. Valid for: 2 years from approval Installation 5.2.1. Place balance on a firm level surface, away from any vibrations. 5.2.2. Manipulate the foot screws at the bottom of the balance until the air bubble in the spirit level is in the displayed circle. 5.2.3. Attach the mains adaptor to the balance and then to the mains supply if using the electrical supply to power the balance. 5.2.4. Insert suitable rechargeable batteries to power the balance if an electrical supply is not available. 5.2.5. Switch on the balance from its main supply. 5.2.6. Ensure that [*] is displayed to indicate that balance is being supplied with power via the mains adaptor. 5.2.7. Press the On/Off Key to switch on balance. 5.2.8. Allow the balance to warm up for about 10 minutes. 5.2.9. Ensure that a [0] stability sign appears on the display to indicate that balance is in a stable condition. Ensure that all draft shields are closed if [0] does not appear. 5.2.10. Wait until balance displays [0.000] to start using it. 5.2.11. Press the TARE Key to reset the balance to zero when [0.000] is not being displayed inspite of having an empty weighing pan. 5.3. Configuration 5.3.1. 5.3.2. 5.3.3. 5.3.4. 5.4. Press the F key for about 4 seconds until [FUNC] appears. Press the F Key repeatedly until the desired configuration parameter is displayed (see SOP/PD/310_02/A1). Press the TARE Key repeatedly until the desired number, which corresponds with the desired function, is displayed. Press the S Key to save the chosen function and return balance to weighing mode. Operation 5.4.1. 5.4.2. 5.4.3. Ensure that the desired weighing unit is displayed. If this is not so, use the configuration parameters menu to change it. Place sample onto the weighing pan, close all draft shields and read off value from the LCD display. If using the tare function, place empty weighing container on the weighing pan, press the TARE key and then place the sample in the weighing container Page 4 of 13 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE 5.4.4. 5.4.5. 5.5. Close all draft shields and read off the tared value from the display. Use the tare function repeatedly if necessary. Calibration 5.5.1. 5.5.2. 5.5.3. 5.5.4. 5.5.5. 5.5.6. 5.5.7. 5.5.8. 5.6. Valid for: 2 years from approval Choose an appropriate calibration weight. Press and hold the F Key until [CAL] is displayed. Press the F Key while holding down the TARE Key and then subsequently release both keys simultaneously. Place the desired calibration weight on the weighing pan. Wait until the value of the calibration weight is displayed to remove the weight from the weighing pan. Repeat the calibration procedure if [-Err] is displayed due to a calibration error or the use of a wrong calibration weight. Wait until balance automatically returns to weighing mode before further use. Repeat calibration process with a different calibration weight if necessary. Maintenance 5.6.1. 5.6.2. 5.6.3. 5.6.4. Disconnect the balance from the mains supply before attempting any cleaning procedures. Use a clean fine brush to remove fine particles and powders. Use a moistened cloth with a mild cleaning agent for any spilt material. Ensure that the air bubble in the spirit level is in the centre of the displayed circle. Page 5 of 13 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE 5.7. Valid for: 2 years from approval Troubleshooting Fault Weight display not illuminated - Weight display changes continually - Weight result seemingly incorrect 5.7.1. 5.7.2. - Possible Cause/s Balance not switched on Mains supply cable not properly plugged or is faulty Interruption in power supply Draught / air movements Table / floor vibrations Weighing pan in contact with foreign matter Electromagnetic fields / static charging in the vicinity Balance display not set to zero Calibration is outdated Great fluctuations in temperature Electromagnetic fields / static charging in the vicinity Switch off the balance if other error messages appear and then switch it on again. Contact the manufacturer if the error message re-appears again after switching on. Page 6 of 13 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE 5.8. Valid for: 2 years from approval Flow Charts 5.8.1. Installation Start Place balance on firm level surface, away from any vibrations Manipulate foot screws at bottom of balance until air bubble in spirit level is in the displayed circle No Using electrical power supply Insert rechargeable batteries Yes 1 Attach mains adaptor to balance and then to mains supply Switch on mains supply No Check adaptor is correctly inserted into balance and mains supply [*] displayed Yes 1 Press On/Off Key to turn on balance Allow balance to warm up for about 10 minutes No [0] displayed Close all draft shields Yes Wait until balance displays [0.000] No [0.000] displayed Yes End Page 7 of 13 Press TARE Key to reset balance UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE 5.8.2. Configuration Start Press F Key for about 4 seconds until [FUNC] appears Press F Key repeatedly until desired configuration parameter is displayed (see SOP/PD/310_02/A1) Press TARE Key repeatedly until desired number, corresponding to a particular function, is displayed Press S key to save chosen function and return balance to weighing mode End Page 8 of 13 Valid for: 2 years from approval UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE 5.8.3. Valid for: 2 years from approval Operation Start No Desired weighing unit displayed 1 Press F Key for about 4 seconds until [FUNC] appears Yes No Using Tare function Place sample onto weighing pan, close all draft shields and read off value Yes Press F Key repeatedly until [S.u.] is displayed Press TARE Key repeatedly until the number that corresponds with the desired weighing unit is displayed Place empty weighing container on weighing pan Press S key to save and return balance to weighing mode Press TARE Key 1 Place sample in weighing container Close all draft shields and read off tared value Use tare function repeatedly if necessary End Page 9 of 13 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE 5.8.4. Valid for: 2 years from approval Calibration Start Choose an appropriate calibration weight 1 Press and hold F Key until [CAL] is displayed Press and hold F Key while holding down TARE Key Subsequently release both keys simultaneously Place desired calibration weight on weighing pan No No Calibration weight value displayed Yes [-Err] displayed Wait until calibration weight value is displayed Remove calibration weight 2 Yes 2 Remove calibration weight from weighing pan Wait until balance returns back to weighing mode Wait until balance returns to weighing mode before further use 1 No Repeat calibration with different standard weight End Yes 1 Page 10 of 13 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE 5.8.5. Valid for: 2 years from approval Maintenance Start No Balance requires cleaning 1 Yes Disconnect balance from mains supply No Any surrounding powder Yes Remove using a clean fine brush No Any spilt material Yes Remove using a moistened cloth with a mild cleaning agent No Air bubble in spirit level in the centre of displayed circle 1 Yes End Page 11 of 13 Manipulate foot screws at bottom of balance until this is achieved UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE Valid for: 2 years from approval 6. Precautions 6.1. Avoid extreme heat as well as temperature fluctuations caused by installing next to a radiator or from direct sources of sunlight. 6.2. Protect the balance from air draughts due to open windows and doors. 6.3. Keep weighing pan clean and dry to prevent inaccuracies. 6.4. Do not move balance during weighing. 6.5. Ensure that the maximum tolerated weight of 120g is not exceeded during weighing procedures and also when using the tare function. 6.6. When transferring balance to a significantly warmer environment, allow it to acclimatise for approximately 2 hours at room temperature after it has been disconnected from the mains. 6.7. If using rechargeable batteries to operate balance, charge these for at least 8 hours, using the mains adaptor, before using it for the first time. 6.8. Do not leave a permanent load on the weighing pan. 6.9. Only use balance if spirit level is correct. 6.10. Adjust the air bubble in the spirit level each time the balance is to be moved to another location. 6.11. Avoid static charging of the material to be weighed, the weighing container and the draft shield. 6.12. Do not place in a location where there are possible electromagnetic fields, static charging and / or an unstable power supply since these may all cause major display deviations. 7. References KERN.Operating Instruction Precision balances KERN® EW/EG-N/EWB. Balingen: KERN and Sohn GmbH; 2006. 8. Appendices SOP/PD/310_02/A1 – Configurations Parameter Table 9. Revision History Version Number 01 02 Page 12 of 13 Amendments/ Reasons for change Initial Release Inclusion of subheading titles of Procedure section in Table of Contents Update of entire SOP to version 2 UNIVERSITY OF MALTA FACULTY OF MEDICINE & SURGERY PHARMACY DEPARTMENT Ref. No. SOP/PD/310_02 STANDARD OPERATING PROCEDURE KERN® EW 120-4NM PRECISION BALANCE Valid for: 2 years from approval SOP/PD/310_02/A1 – Configuration Parameters Table Configuration Parameter Display Symbol Function Bar Graph (to display weighing range) b.G. [0] Turn off bar graph display [1] Turn on bar graph display Zero Alignment A.O [0] To perform no zero correction [1] To activate automatic zero correction point Automatic shut-off after 3 mins when using rechargeable batteries A.P. [0] Automatic shut-off deactivated [1] Automatic shut-off activated Display Speed rE. Ranging from sensitive and fast [1] for very tranquil installation locations to non-sensitive but slow [5] for very unsettled installation locations Vibratory Filter S.d. Ranging from sensitive and fast [1] for very tranquil installation locations to non-sensitive but slow [5] for very unsettled installation locations Weighing unit S.u. Baud Rate b.L. [01] gram (g) [14] carat (ct) [15] ounce (oz) [16] pound (lb) [17] troy ounce (ozt) [18] penny weight (dwt) [19] grain (►) [1A] Hong Kong tael (tlH) [1b] Malaysian tael (tlM) [1C] Taiwan tael (tlT) [1d] momme (mom) [1E] Indian tola (to) [1] 1200bps [2] 2400 bps [3] 4800 bps [4] 9600 bps Parity PA. [0] No parity [1] Odd parity [2] Even parity Page 13 of 13