S5. DESIGN OF A SPECTROPHOTOMETRIC ANALYSIS PURPOSE 1. To determine the appropriate series of standards for a visible absorption calibration graph. 2. To determine the concentration of a sample by absorption spectrophotometry. INTRODUCTION To determine the concentration of an unknown solution by spectroscopy involves analysing a series of solutions of known concentration, the standards, and comparing the value obtained for the unknown. Beer’s Law is generally linear over the range of 0.1 – 1.0 absorbance units hence it is advisable to have all solutions fall within this range. When the appropriate standards are not already prepared the chemist will need to devise a procedure that will provide standards that fall within the appropriate range. The following method is one way of determining the standards required (Note the method uses the premise that the absorptivity coefficient is a constant as per Beer’s Law): Example A 1000mg/L standard solution has an absorbance of 3.811. Diluted 10 times to 100 mg/L, it gives an absorbance of 1.128. Step 1- Determining the A/c relationship Example Calculate the constant (k=ab) from the absorbance and concentration of the standard. The absorbance of this solution must be less than 2.0 k = 1.128/100 = 1.128 x 10-2 k = A/c Step 2 – Determining the Concentration of the 0.2 Absorbance standard Calculate the concentration of a solution of the analyte with an absorbance of 0.2 using the value of k calculated in Step 1. c = 0.2/k c = 0.2 / 1.128 x 10-2 = 17.73 mg/L Step 3 – Determining a suitable concentration range Conc (mg/L) 17.73 20 for standards Round the concentration from step 2 to a manageable 40 value (nearest 5 or 10). The other standards are 2, 3 and 4 60 times the concentration 80 Approx Abs 0.2 0.4 0.6 0.8 Step 4 – Preparation of the standards The 0.2, 0.4, 0.6 and 0.8 absorbance standards can be most conveniently and accurately prepared by dilution of 5, 10, 15 and 20 mL aliquots, respectively, of a more 5 mL of X mg/L diluted to concentrated stock solution (X). 100 mL of 20 mg/L. To calculate the concentration of X, assume that 100 mL of the final standards are being prepared. Therefore, 5 X = 20 x 20 mg/L mL of X is diluted to produce 100 mL of 0.2 Abs std. = 400 mg/L Using C1V1 = C2V2 then C2 = 100 x conc 0.2 std 5 = 20 x conc. of 0.2 std Step 5 – Preparation of X If X is greater than 500 mg/L, than it can be prepared directly. Otherwise, it should be prepared by diluting a more concentrated standard. Using pipettes, the most convenient and accurate dilutions are 2 (50 100), 4(25 100), 5 (20 100), 10 (10100) or 20 (5100). A burette could, however, be used. You need at least 100 mL of stock standard ( 5+10+15+20). 400 mg/L is too low to prepare directly (unless you make at least 500mL of it). Given that 1000mg/L standards are very often available, a suitable method would involve a 2.5 dilution of the 1000 mg/L stock would give 400 mg/L. 40 mL of 1000 mg/L diluted to 100mL of 400 mg/L PROCEDURE 1. Obtain two coloured solutions: a 1000 mg/L stock solution and a sample solution of unknown concentration. 2. Using a measuring cylinder, dilute the stock solution to 100mg/L by taking 10 mL of the 1000 mg/L and making it up to 100 mL. Run an absorption spectrum of this solution. 3. Determine the wavelength of maximum absorbance, and the absorbance value for the solution prepared in Step2. 4. Using the data obtained in step 3, determine the range of concentrations suitable for preparation of a calibration graph. Your final dilutions should use 5, 10, 15 and 20 mL aliquots diluted quantitatively to 100 mL. 5. Measure the absorbance of your sample. Determine the factor by which it needs to be diluted to bring the absorbance into the range of your standards. 6. Prepare, quantitatively 100 mL of diluted unknown sample. 7. Record the absorbance of your standards and the diluted sample. CALCULATIONS 1. Graph the absorbance of the standards against their concentration. . 2. Determine the concentration of the diluted sample. 3. Determine the concentration of the concentrated unknown sample. QUESTIONS 1. What is the optimum range of absorbance for Beer’s Law calibration graphs? 2. What errors arise at absorbance below the optimum range? 3. What errors arise at absorbance above the optimum range? 4. When determining the absorbance of the stock solution, why could you carry out dilution of the solution approximately? 5. Why is it convenient to prepare the calibration graph standards using the 5, 10, 15 and 20 mL series of aliquots? S5. RESULTS SHEET Date of analysis Identity of instrument: Sample solution code: Wavelength of maximum absorbance: Absorbance of solutions Solution Maximum absorbance 100 mg/L (from stock) Sample solution Step 1. Determining the A/c constant Calculation 2. Determining the concentration of 0.2 Abs standard 3. Determining a suitable concentration range 4. Preparation of the standards 5. Preparation of the sample Dilution factor for unknown: Solution Blank Standard 1 Standard 2 Standard 3 Standard 4 Diluted sample Conc (mg/L) 0 Abs Have you? Completed the instrument log Completed the sample register Completed the standard register Teachers signature Date Signature