Reminder: These notes are meant to supplement, not replace, the lab manual. Gas Chromatography notes Application: Gas Chromatography (GC) provides a quick, accurate and relatively inexpensive way to analyze very small samples for composition. Simple GC machines may cost in the area of $5,000i. This is fairly inexpensive as far as analytical instrumentation goes. They can be installed in a variety of laboratories designs including mobile testing units. GC’s are easy to operate and provide reliable results. 1. Here is some terminology related to this experiment: Chromatography is a technique in which compounds to be separated are distributed between a mobile phase and a stationary phase. Different distributions or partitioning between the two phases give rise to the separation. Volatile- Easily converted into a gas. Evaporating rapidlyii. Gas Chromatography (GC) is useful for separating and analyzing volatile liquid or gas mixtures. GC does not work on many solids. The stationary phase in gas chromatography (GC) is almost always a relatively nonvolatile liquid, like a wax. This liquid is coated on either solid particles or the inside of a capillary tube. The stationary phase used in this experiment is carbowax coated on small inert particles. The mobile phase in gas chromatography is an inert (nonreactive) gas, usually helium or nitrogen. Nitrogen will be used in this lab. This mobile phase is often called the carrier gas. The retention time is the time (in seconds or minutes) between the injection of a sample and the middle of when the sample reaches the detector. The retention time is measured on a chromatogram between the start mark and the top of a peak. These times are used like Rf values in TLC, to establish the identity of the compound giving the peak by comparison. To elute a compound means to have it emerge from or come off of a column. 2. The fact that the mobile phase is a gas usually makes it necessary to use a more elaborate setup to confine the gas and direct its flow. Nevertheless, this is still chromatography; there is a sample, a mobile phase, and a stationary phase. In GC the stationary phase is a liquid and the mobile phase is a gas. Three different types of Gas Chromatographs (GCs) may be used in this lab. The first two types (Shimadzu and Gow-Mac) are the most common. A typical set up is drawn below. The components are not drawn to scale. Each has an external pressurized gas cylinder attached. In addition, new Vernier mini-GC’s may be used. These mini-GC’s use air from the room and do not need a separate pressurized gas tank. The carrier gas is typically kept in a large metal cylinder under high pressure. The gas flows through a valve into a regulator, which reduces the pressure to a lower value the rest of the system can use. (Scuba divers use regulators for the same purpose). The gas flows past the injection port, where samples are injected using syringes. The injection port is heated, and the samples evaporate there. The sample and carrier gas mixture are carried into an oven that contains the chromatography column. (The heat prevents components of the sample from condensing). The components partition and separate in the column. The separated components of the mixture pass through the detector, which sends a signal to the strip chart recorder or integrator. 3. The separation occurs by different partitioning or distribution of compounds between the stationary (non-volatile liquid) phase and the mobile (gas) phase. What physical characteristic describes the energy needed to move a compound from the liquid to the gas phase? A compound that spends more of its time in the gas phase (mobile phase) would have a shorter retention time than a compound that spends more of its time in the liquid phase (stationary phase). The boiling point of compounds is the primary means of predicting the order of elution from a GC. The lower the boiling point, the shorter the retention time (RT). Secondarily to boiling point is polarity. If a mixture of similar compounds with known boiling points is injected into a GC, the identity of the peaks can be made by correlating the retention time to boiling point. The lowest boiling point material will spend a higher portion of its time in the gas phase and will have the shortest retention time. 4. To validate a peak identification, a pure authentic sample of the suspected compound is injected into the same GC at the exact same conditions (temperature, flow rate, column). The Rt of the authentic sample is then compared with the Rt of the suspected peak. If the Rt’s are different, they are different compounds. If the Rt’s are the same, a positive identification has been made. In this way, Rt values of GC are very similar to Rf values of TLC. 5. A photo of one of the Gow-Mac GC’s found in the organic lab is shown. 6. The technique of gas chromatography is widespread in academic, industrial, and governmental labs. It is used to separate the components of volatile liquid and gas mixtures, identify the compounds (by comparison of retention times), and determine the amounts of the compounds present. 7. Samples for GC analysis must be able to be vaporized into the gas phase in order to pass through the machine. Non-volatile liquids and non-volatile solids cannot be analyzed using GC. The sample size necessary for GC is quite small. A typical injection volume ranges between 1/10 and 5 microliters (l). One microliter is equal to 1/1000 of a milliliter or 1/1,000,000 or 1x 10 -6 liters. If a drop of liquid is equal to 1/20 of a milliliter, then one drop has enough volume for ten 5 l injections. 8. The size of a peak (area) is proportional to the amount of compound present. The relative size (area) of two peaks will indicate the relative composition of those two materials in the sample. If one peak has an area of 2 cm 2 and another peak has an area of 3 cm2, then the sample is composed of (2/(2+3) x 100%) or 40 % of the first compound and 60 % of the second compound. 9. The volume of sample injected is not critical (within reasonable limits). It doesn’t matter if 3, 4, 5, or 6 l of sample is used. It is not reasonable to inject 2 mL of a sample into a GC. If double the volume is injected (3l vs 6 l) , the absolute size of the peaks will double, but the ratio of the sizes of the two peaks will remain the same, because the percent composition will remain the same. The Rt will also not change. 10. If a sample containing a mixture of three compounds is injected multiple times on the same machine, with the same column, the same flow rate and the same temperature of the column, the retention times of the three compounds will be the same in each injection. If the relative composition of the samples are different, the relative sizes of the peaks will change, but the retention time will remain the same. 11. Here are the structures of compounds which make up the mixtures used in the 2230L experiment. The exact % composition of each in the mixture will be determined using the GC. Each unknown is made up of these two materials. o Ethyl acetate, B. P. 77 C o Butyl Acetate, B. P. 126 C Both of these compounds are esters of acetic acid (acetate). They differ by two CH2 groups. The addition of a single CH2 group raises the boiling point by over 20oC, so the addition of two methylene groups raise the boiling point by 49°C. Most esters have a pleasant scent and a moderately low toxicity. These are no exceptions to this rule. Which compound is expected to have the highest retention time? 12. Safety considerations for this experiment include: Syringes are sharp and can easily prink skin. If a syringe is left too long in the injection port without depressing plunger, the liquid will volatilize and expand, shooting the plunger out at an unsuspecting student. The injection port is hot and can easily burn skin. Be sure to wear goggles while operating GC. Cylinders of gases under high pressure must be securely fastened to solid, permanent furniture or walls. If such cylinders fall and break, the escaping gas transforms them into potentially deadly missiles. The organic materials analyzed in this experiment are highly flammable. 13. Changes in the procedure can, as always, have a profound effect on the outcome of the experiment. Here are some possible changes in procedure and their effects. Faster carrier gas flow rates and higher column temperatures give shorter retention times, but peaks may not be resolved as well. Slower flow rates and lower temperatures give longer retention times and better resolution. The primary characteristic which determines which compound will have shorter retention times (elute quicker) is the boiling point. The more volatile compounds have lower boiling points and will elute the column faster and hence have shorter retention times. Secondarily to boiling point, is polarity. Less polar compounds generally have shorter retention times (there are exceptions to this). Stationary phases should not evaporate at the temperatures of the experiment, and should not react with the compounds under study. Stationary phases have a significant impact on the retention time. To change a stationary phase means to physically remove the column (with wrenches) from the GC machine and replace it with another column packed with a different stationary phase. This is not done frequently. 14. Gas chromatography can be used for both qualitative (identification) and quantitative (amounts) analyses. The RT is used for identification. The area of the peak is used to determine the amount of each component present. The area of a peak in a chromatograph is proportional to the amount of compound present and various instrument factors. Integrators connected to gas chromatographs provide RTs, peak area, and percent area. The percent area of each peak is the same as the percent of each material present. This is read directly off the integrator output. A sample output is below. PK 1 2 3 4 RT 0.23 1.14 2.47 2.78 AREA 7054 642298 3205001 1816737 TYPE BB PB BP BP AREA % 0.12439 11.32583 56.51473 32.03505 There are four peaks and hence four compounds in the above sample. The first peak at 0.23 can be disregarded due to its minute size and insignificant presence (0.12439%). The third peak with a retention time of 2.47 minutes is the major component with 56.51 %. The fourth peak at 2.78 minutes is the second most prevalent material and the peak at 1.14 minutes is present in 11.32%. 15. If an integrator is not available, the output of a GC detector is then recorded using a plotter or strip chart recorder. The peak area will then have to be measured manually. The method for manually determining peak area is described below. A typical plotter or strip chart recorder output looks like the image below. The area of each peak is defined by the height of the peak times the width of the peak at half of the maximum height. (H x W 1/2H) Area = Height (cm) x Width at Half maximum peak height (cm) = cm2 For the below chromatogram, the area of peak A is 4.68 cm times the width of the peak at a height of 2.34 cm. This equals 4.68 cm x 0.18 cm = 0.84 cm2. For the below chromatogram, the area of peak B is 6.50 cm times the width of the peak at a height of 3.25 cm. This equals 6.50 cm x 0.17 cm = 1.1 cm 2. For this chromatogram, the percent of each is calculated as follows. The percent of each compound in the mixture is the area of that peak, divided by the sum of all of the areas times 100%. 16. If an integrator is not available, the retention time can easily be calculated. This is done by knowing the chart speed of the printer or strip chart recorder in cm per minute (cm/min) and the distance from the injection to the center point of the peak. For peak B, the distance from the injection point at 0.0 to the peak of peak B is 8.05 cm. If the chart speed was 0.50 cm/min then the retention time for peak B is calculated as following: ( i ) Shimadzu Instrumentation Price http://www.shimadzu.com/an/gc/index.html (January 22, 2012) ii Grant, J. Hach’s Chemical Dictionary,McGraw-Hill, 1969, p 717 Revised October 13, 2014, S. L. Weaver