Thermoluminescence TL – 200 / PMT

advertisement

Thermoluminescence

TL – 200 / PMT

Thermoluminescence TL-200/PMT consists of three components -

Control Unit, Thermoregulator and Measuring Chamber.

Thermoregulator

Control unit

Measuring chamber

Control Unit

The Control Unit executes the experiment according to protocols defined by the user in the FluorWin software. The protocol combines elemental commands, e.g., set temperature, set temperature ramp rate, set photomultiplier sensitivity, generate saturating pulse/pulses with defined duration and specifies the timing of their execution. The Control Unit communicates with the PC via the serial port. There are two Channel inputs in the Control Unit.

Channel 1 is used for the measurement of thermoluminescence signal; Channel 2 serves for the temperature measurement during the experiment. The measured curves are displayed in two formats:

1

Time/Temperature and Time/TL Signal, or Temperature/TL Signal. It is also possible to export the measured data to a *.txt file for an analysis in other software packages.

Temperature

TL signal time

TL signal temperature

Thermoregulator

The Thermoregulator controls the temperature of the sample disc in the range of -25°C to +70°C. The temperature can be set either manually or by the protocol (software). The actual temperature of the sample is displayed on the front panel of the Thermoregulator.

The temperature is read and recorded by the Control Unit. In the temperature ramp mode, the Thermoregulator increases the temperature of the sample linearly with a rate ranging between

0.1°C/sec - 2°C/sec.

Measuring

Chamber

The Measuring

Chamber consists of four key parts:

A/ Light source

The light flashes are emitted by a ring of 8 powerful Lumiled diodes ( λ max

=630nm) focused on the sample disc. The intensity of a single turnover flash is

Detector

LED

2

around 200 000 µmol(photons).m

-2 .s

-1 . The maximum flash duration is 150 µs. The light intensity as well as the flash duration are software controlled.

B/ Photomultiplier with an integrated power supply

The photomultiplier detects photons in the wavelength range 300-

900nm. A switch shuts the photomultiplier down when the measuring chamber opens so that the photomultiplier is protected from high ambient light. After the light flash the photomultiplier is switched on in 100 ms so that only delayed luminescence is detected.

C/ A/D converter

The electric current of the photomultiplier is amplified with a software controlled gain before digitization. The time response of the amplifier is fixed at 50ms, determining minimal sampling period to 100ms. The signal is digitalized directly in the measuring chamber.

D/ Sample disc with the cooling unit

Gold-plated copper

½

” Sample disc

The sample disc is made of gold-plated copper. The temperature sensor is situated right under the surface of the disc. The sensor is calibrated in the range of -25°C to

+70°C with an accuracy of ±0.2°C.

Suspensions of cells, chloroplasts or membranes ought to be measured on a filtration paper (12.7mm, 1/2inch in diameter) so that good thermal connection between the metal disc surface and the sample is maintained. The sample volume should be as low as possible

(<100 µ l) to minimize its heat capacity.

Built-in Peltier cell is responsible for cooling and heating of the sample disc. For operation between +10°C to +70°C, the Peltier cell is sufficiently cooled by the built-in fan. For temperatures lower than

3

+10°C an outer cooling water-circuit is required. The water source provided by the user is connected to two silicon hoses of the measuring chamber. For the highest cooling rate the water temperature should be optimally around +4°C. This allows that the cooling time from +20°C to -10°C is as short as 50 seconds. With tap water temperature between 15-20°C the cooling time increases to ca

70s.

Water pump

WARNINGS:

1) The Thermoluminescence TL – 200/PMT can also measure leaf segments. In this case, the dynamic thermal equilibrium between the sample metal disk and the leaf is harder to maintain and the thermometer reading must be calibrated by the user for the particular experimental setting.

2) The user also has to consider potential temperature-reading errors that can be caused by the phase change in the sample

(freezing, de-freezing). Because of the latent heat, the sample is arrested during the phase change at a constant temperature (the temperature rise is non-linear). This error will be small if the sample volume is low.

4

Technical specification :

Temperature control:

Temperature range

Modes of operation

Overheating protection

Sampling period

Control regimes

Sample:

Sample diameter

Sample disc material

Typical sample

Light source wavelength

Light source intensity

Detection system:

Sensor

Spectral response

Sampling period

Time response

Switch-on delay

Ambient light protection

A/D converter resolution

Control unit:

Custom defined protocols

Communication

Software

Electrical

-25°C - +70°C with water circuit

+10°C - +70°C without water circuit constant; linear change (0.1°C/sec – 2°C/sec)

80°C

10ms manual (constant temperature), protocol defined temperature profiles

½

” disc

Gold-plated copper algae, cyanobacteria on

½

” filtration paper

λ max=630nm up to 200 000 µmol(photons).m

-2

.s

-1

Photomultiplier with sensitivity control (software)

300nm-900nm

100ms

50ms

100ms mechanical switch

16bit

Variable timing, special language and scripts

Serial port

FluorWin 3.6

90V-240V

5

Download