PASSIVE SAMPLING OF ORGANIC CONTAMINANTS IN THE WATER PHASE Kees Booij, Evaline M. van Weerlee, Coen V. Fischer, José Hoedemaker Nederlands Instituut voor Onderzoek der Zee ©2000 This report is not to be cited without the acknowledgement of the source: Netherlands Institute for Sea Research (NIOZ) P.O. Box 59, 1790 AB Den Burg, Texel The Netherlands ISSN: 0923- 3210 PASSIVE SAMPLING OF ORGANIC CONTAMINANTS IN THE WATER PHASE Final report Kees Booij, Evaline M. van Weerlee, Coen V. Fischer, José Hoedemaker NETHERLANDS INSTITUTE FOR SEA RESEARCH NIOZ - RAPPORT 2000 - 5 SUMMARY The exchange kinetics between artificial sea water and three passive samplers (Semipermeable Membrane Devices (SPMDs), Low-Density PolyEthylene (LDPE) membranes, and silicone tubing) was studied in an ex­ perimental set-up using three controlled flow tech­ niques, and three temperatures between 2 and 30 °C. Aqueous concentrations of chlorobenzenes, PCBs and PAHs were maintained by continuous circulation of the water over a C18 generator column. The effect of tem­ perature on the sampling rates amounted to a factor of two to three. Differences between the uptake kinetics of organic contaminants by LDPE tubing using three forced flow techniques amounted to a factor of two. Sampling rates of SPMDs and LDPE membranes were the highest reported so far in the literature. Performance Reference Compounds (PRCs) were spiked into LDPE membranes and silicone tubing by in­ cubation in aqueous-methanolic PRC solutions. During the spiking procedure all PRCs reached equilibrium within a few hours. Best results were obtained with a methanol/water volume ratio of 80/20. The precision of the PRC spiking was 1-6% depending on the com­ pound. In February and October 1999 field deployments of SPMDs and LDPE membranes were made under ambi­ ent flow conditions in the Scheldt estuary (two stations) and in the Southern North Sea (three stations) along with mussel exposures by the National Institute for Coastal and Marine Management /RIKZ. Inter-site dif­ ferences in absorbed contaminant amounts were more pronounced in SPMDs and LDPE membranes than in mussels. The dissipation rates of PRCs from SPMDs and LDPE membranes indicated that the sampling rates were constant within a factor of 1.3 for all stations. wide, with a wall thickness of 50 to 90 urn. A wide range of non-polar to moderately polar contaminants has been sampled successfully with SPMDs: PAHs (Lebo et al., 1992; Huckins et al., 1993), organochlorine pesti­ cides (Prest et al., 1992; Ellis et al., 1995; Herve et al., 1995; Petty et al., 1995; Prest et al., 1995b; Prest et al., 1995c), PCBs (Prest et al., 1992; Petty et al., 1993; Prest et al., 1995b; Prest et al., 1995c), dioxins/dibenzofurans (Prest et al., 1992) and chlorophenols/anisols/veratols (Ellis et al., 1995; Herve et al., 1995). When SPMDs are exposed to the water phase organic contaminants initially are absorbed at a constant rate (linear uptake phase, Fig. 1). N = C w Rs t (1) with N the absorbed amount, Cw the aqeuous concen­ tration, and Rs the sampling rate. For low Kow com­ pounds water-SPMD sorption equilibrium may be ap­ proached, causing the uptake curve to level off at the equilibrium amount Neq (Fig. 1, equilibrium phase). Neq= V s KSwCw (2) with Vs the SPMD volume and KSw the SPMD-water partition coefficient. Equations 1 and 2 indicate that contaminant amounts absorbed by SPMDs are linearly proportional to their concentration in the water phase, regardless of the degree of equilibrium that has been attained between SPMD and water. In principle this means that for each individual compound a higher ab­ sorbed amount indicates a higher aqueous concentra­ tion. Comparison of absorbed amounts for different compounds is more complicated, because both the SPMDs and LDPE membranes were exposed under ambient and forced flow conditions in the Marsdiep (Western Wadden Sea). In addition, silicone tubing was tested as a passive sampler of organic contaminants. The silicone tubing sampler was much more efficient in sampling low Kow compounds when compared to sam­ pling by SPMDs, due to a larger phase volume for the silicone sampler. This study was carried out within the framework of proj­ ect TNW*MILLAB of the National Institute for Coastal and Marine Management /RIKZ, under contract RKZ 583. INTRODUCTION Passive sampling of dissolved organic contaminants by Semi-Permeable Membrane Devices (SPMDs) has re­ ceived great deal of attention during the previous decade (Huckins et a l, 1990; Prest et al., 1995a). SPMDs consist of a low-density polyethylene (LDPE) lay-flat tubing, filled with either natural lipids or the mod­ el lipid triolein (1,2,3-tri[cis-9-octadecenoyl]glycerol). Typically, SPMDs are 10 to 230 cm long and 2.5 to 5 cm Fig. 1. Uptake curves for three com pounds that differ in hydrophobicity: low Kow (thick), m edium Kow (thin) and high Kow (dashed). A t sampling tim e 1 all com pounds are in the linear uptake phase. A t sampling tim e 2 the low Kow com pound has reached equilibrium , the high Kow com pound still is in the lin­ ear uptake phase, and the m edium Kow com pound is in an in­ term ediate phase. sampling rates (Rs in eq. 1) and the SPMD-water parti­ tion coefficients (KSw in eq. 2) are dependent on the compound properties. A complicating factor in the inter­ site comparison of contaminant amounts is that the sampling rates increase with increasing water flow rates (Booij et al., 1998). The flow effect can in principle be accounted for by measuring the dissipation rates of Performance Reference Compounds (PRCs) that are spiked into the triolein phase prior to exposure (Prest et al., 1992; Huckins et al., 1993; Booij et al., 1998). When PRCs are chosen that do not occur in the envi­ ronment their dissipation can be described by N=N0 exp(-ke t) (3) where N0 is the PRC amount at t=0 and ke is the ex­ change rate coefficient, or the dissipation rate constant. Comparison of ke values obtained at different exposure sites therefore allows for estimating the effect of inter­ site differences in flow on the absorbed amounts. Suitable PRCs are isotopically labelled compounds and a number of polychlorinated biphenyls or other com­ pounds that do not occur in the environment in appre­ ciably amounts (e.g. CB29, CB155). Another complicating factor in the field application of SPMDs is that both the sampling rates and the SPMDwater partition coefficients may be temperature de­ pendent. This may cause problems in the interpretation of SPMD results obtained at different temperatures. In a previous laboratory study we showed that the effect of differences in flow velocity on the sampling rates could be accounted for by changes in the dissipation rate constants of the PRCs (Booij et al., 1998). It also appeared that for high Kow (>106) compounds the LDPE membranes were just as efficient in sampling or­ ganic contaminants as were (triolein filled) SPMDs. In the present study we aim to evaluate the effect of tem­ perature on the sampling rates and the PRC dissipation rates for SPMDs, LDPE membranes and silicone tub­ ing, using laboratory experiments under controlled con­ ditions. For a number of field experiments in the Dutch coastal waters we assessed the merits of these passive samplers compared to mussel exposures for monitoring programmes. In addition we investigated whether or not the application of a forced flow regime during the sam­ pler exposures would facilitate the interpretation of the results. We also developed a method for spiking PRCs into LDPE membranes and silicone tubing. MATERIAL Internal standards and Performance Reference Compounds (PRCs). Prior to the exposure of LDPE membranes and SPMDs, the samplers were spiked with a number of PRCs: perdeuterated PAHs (acenaphtene-D10, phenanthrene-D10, chrysene-D12) and PCBs (CB 4, CB 29, CB 155 and CB 204). For SPMDs the PRCs were spiked into the triolein. For LDPE mem­ branes and silicone tubing, the PRCs were loaded by equilibrating the polymers in PRC spiked methanol/wa­ ter mixtures (80/20 v/v) overnight. Details of the poly­ mer spiking are discussed below. Analytical internal standards were added to the samples prior to extrac­ tion. For the lab experiments CB112, pyrene-D10 and perylene-D12 were used. For the field samples a total of six internal standards were used: CB53, CB112, CB198, pyrene-D10, 13C6-chrysene, and peryleneD12. LDPE membranes and SPMDs. Low-density polyeth­ ylene tubing (2.5 cm wide, wall thickness 70 urn) that contained no additives was supplied by Brentwood Plastics Inc. (Brentwood, Missouri, USA). Prior to use the membranes were extracted twice by soaking in pentane for 24 hours. After exposure the membranes were extracted 24 hours with 100 mL pentane. SPMDs were prepared following the method described earlier (Booij et al., 1998) After exposure SPMDs were extracted twice with 100mL pentane for 24 hours. Both LDPE membranes and SPMDs had a length of about 30 cm. SPMDs contained 300 uL triolein, corresponding to the standard design value of 1 mL m"1. Silicone tubing. Silicone tubing was purchased from Fischer Scientific ('s Hertogenbosch, The Netherlands; 6 mm id, 9 mm o.d.), Cole-Palmer (Niles, IL, USA; Masterflex 96410-18, 8 mm id, 11 mm o.d.), Rubber BV (Hilversum, The Netherlands, 2 mm id, 6 mm o.d.). The method development for silicone tubing extraction and clean-up is described in the Results and Discussion I section. The Fisher Scientific tubing was selected for the exposure experiments. MeOH/Acetonitrile/water mixtures. Contaminant con­ centrations in mixtures of water and methanol or acetonitrile were determined by bringing the water content to > 80% using hexane extracted bidistilled water, and extracting twice with pentane after addition of the inter­ nal standard solutions and about 1 mL 2,2,4-trimethylpentane (TMPe). The extracts were transferred to a round bottom flask equipped with a mi­ cro Snyder column, and were concentrated to about 5 mL. After transferring the concentrates to a 20 mL grad­ uated test tube the samples were further concentrated to about 0.5 mL. The final sample volume was 1 mL in TMPe prior to injection. Water samples. Water volumes from the exposure sys­ tem (4.8 L) were siphoned off through a stainless steel tubing into a 5 L volumetric flask. The water was ex­ tracted with about 100 mL pentane for 30 min using a magnetic stirrer and a stirrer bar from which the Teflon coating had been stripped off. After a clean-up step us­ ing 2 g silica (6% water, elution with 40 mL pentane) the sample volumes were concentrated to 100 uL prior to injection. Sample clean-up. The raw concentrated extracts were passed over a 6 mm id LC column that contained 2 g silica (6 % water, elution with 40 mL pentane). For the lab experiments the eluate was concentrated to 0.1 or 1 mL followed by analysis for PCBs/chlorobenzenes and PAHs. After analysis for PAHs the field samples were passed again over a silica column (elution with 10 mL pentane) prior to analysis for PCBs/chlorobenzenes. GC-analysis. Samples were analysed for PCBs and chlorobenzenes on a Carlo-Erba 5160 equipped with an A200S autosampler, a CP-Sil8 capillary column (50m, 0.25 mm id, film thickness 0.25 um, carrier gas H2) and an ECD-80 model electron capture detector (constant-current mode, 340 °C), injection mode splitless, temperature program 90 °C (hold 2 min), increase with 10 °C min"1 to 215 °C (hold 10 min) increase with 8 °C min'1 to 275 °C (hold 17 min). For the lab experi­ ment samples a faster temperature program was used: 90 °C (4 min), increase with 15 °C min"1 to 300 °C (15 min). Samples were quantified using logarithmic inter­ polation (internal standard method). PAHs were analysed on a HP5890 GC equipped with a HP7673 autosampler and a CP-Sil 8 capillary column (50 m, 0.25 mm id, film thickness 0.25 urn, carrier gas He), in­ jection mode splitless, temperature program 100 °C (0 min), increase with 15 °C min' 1 to 200 °C , increase with 6 °C min"1 to 320 °C (0 min). Analytes were detect­ ed with a Fisons Auto-Spec Ultima mass spectrometer, ionisation mode El (70 eV, source temperature 250 °C), detection mode Selected Ion Recording of the molecu­ lar ions (dwell time 80 ms, equilibration 10 ms). Samples were quantified using linear interpolation of the responses (internal standard method). Preparation of the generator column. About one kg of silica bonded C18 (Polygoprep 100-30 C18, Machery-Nagel, Düren, Germany) was loaded in a 1 L pressure filtration unit, using a 25 urn nylon gauze as a filter. The column was pre-flushed three times by circu­ lating about one litre of acetonitrile through the column for 0.5 hour. Between the successive flushings the ex­ cess acetonitrile was removed as far as possible. Then 600 mL of a spike solution containing three chlorobenzenes, seven PCBs and ten PAHs in acetonitrile (100 ug - 5 mg) was circulated through the column for one hour. The water content ofthe solution was increased to 90% in steps of 10-20% by adding bidistilled water, fol­ lowed by equilibration for at least one hour. After reach­ ing a water content of 90% the column was flushed four times by circulating 1L of Milli-Q water. Initial experi­ ments showed that the volume rate of flow through the column was insufficient. Therefore, the C18 material was removed, homogenised and distributed over two identical column systems. Forced flow exposure apparatus. Three techniques for exposing LDPE membranes under forced flow con­ ditions were tested for mechanical strength and ease of operation (Fig. 2). One system consisted of an LDPE disk (14 cm diameter, thickness 70 urn) rotating at a speed of 180 min"1 . A first try with an unsupported disk (Fig. 2A) showed that the membrane was wrapped up around the main shaft. Therefore the disk was posi­ tioned between two support disks (gap width 5 mm) with 102 holes with a diameter of 8 mm (Fig. 2B). An as­ sembly with four supported disks was used in the expo­ sure experiments. The second system was a two­ toothed fork rotating at 60 min'1 (Fig. 2C). The distance between the teeth was 30 cm. On each ofthe teeth a to­ tal of five LDPE membranes or SPMDs could be clamped. In the third system water was pumped through a i m lay-flat tubing at a rate of 25 L h"1 (Fig. 2D). A metal rod was used to prevent the empty tubing to be wrapped up in the simultaneously exposed rotat­ ing disk assembly. The temperature and flow depen­ dence of the sampling rate for LDPE membranes and SPMDs was studied using the fork sampler (Fig. 2C). The exchange kinetics between silicone tubing and wa­ ter was studied at 13 °C by pumping water through a 1 support disk Fig. 2. Forced flow sampling set-ups used. m long silicone tubing. For the exposures at 2 and 30 °C two 20 cm long silicone tubings were mounted on a 14 cm diameter disk that rotated at 60 min"1 , with one end opposed to the flow direction, and the other end ex­ tending 5 mm through the disk as a chimney. Exposure set-up. A 200 L round bottom flask was placed in a 750 L insulated temperature controlled wa­ ter bath (Fig.3). A volume of 200 L Milli-Q water was cir­ culated through the C18 columns at a rate of 900 L d‘ 1. Six kg sodium chloride was added to bring the salinity at about 30. Sodium azide (400 g) was added in order to inhibit bacterial growth and subsequent DOC produc­ tion during the exposures. The exposure experiments started after an equilibration period of 20 h. Method development for spiking PRCs on empty membranes and silicone tubing. Two series of exper­ iments were carried out to optimise the spiking of or­ ganic contaminants in LPDE. In each experiment 100 mL of a methanol-water mixture was prepared in a 250 mL erlenmeyer flask by mixing the appropriate volumes of water, methanol and 1 mL of a spike solution in methanol. Five pieces of LDPE membrane (5x2.5 cm, single layer) were added, and the flask was placed on a table shaker operated at 100 min"1. At regular time in­ tervals one of the membranes was removed. The solu­ tion sticking to the membranes was removed by wiping the membranes with a paper tissue. Consequently the membranes were extracted 24 h with 100 mL pentane in a round bottom flask. One series of experiments was carried out with technical PCB mixtures (Clophen-A30 and Clophen-A60). In the second experiment a number of PRCs was used. The spiking of silicone tubing with PRCs was studied by incubating five 1 cm long tubings in a MeOH/water mixture (80/20 v/v) on a table shaker. After incubation the tubings were extracted by soaking in pentane for 24 h. Field exposure. LDPE membranes and SPMDs (2 each) were exposed in the Western Scheldt estuary (2 stations) and at the southern North Sea (3 stations) for 6-8 weeks in January-March 1999 (Fig. 4). A second field survey was made in October-November 1999. Due to bad weather conditions during the second survey on­ ly the stations in the Scheldt estuary and the location west of Schouwen could be sampled. The Katwijk and Noordwijk stations could only be visited in January 2000, but then it was observed that the membranes and SPMDs were lost. From October 1999 until February 2000 a number of field exposures was made in the Marsdiep. The purpose of this field study was to assess the merits of forced flow versus passive exposures and to test the merits of passive sampling using silicone tub­ ing in the field. A total of five simultaneous exposures were started for the three methods. Due to repeated leakage of the engine house of the forced flow appara­ tus two exposures were unsuccessful and one was on­ ly partly successful. Two parallel exposures under am­ bient and forced flow conditions could be finalised without technical difficulties. Analysis of the samples from the second successful parallel exposure could not be finalised before completion of this report. RESULTS AND DISCUSSION I : LAB EXPERIMENTS Testing three forced flow techniques for LDPE membranes. A typical example of the uptake kinetics for the three samplers (Fig. 2) is shown in Fig. 5 for CB153. The differences between the three methods are limited to a factor of 2. We decided to use the rotating fork sampler (Fig. 2C) for studying the temperature ef­ fect on the uptake rates because this sampler could support both SPMDs and empty LDPE tubing. I 6° Fig. 3. Exposure set-up used in the laboratory experiments Fig. 4. Pyrene amounts absorbed by SPMDs during the January-March 1999 field exposure. 1: Hansweert, 2: Vlissingen, 3: Katwijk6 , 4: Schouwen25, 5: Noordwijk60. The bar for station 1 represents an amount of 1802 ng pyrene. Silicone tubing pre-treatment. Synthetic polymers commonly contain large amounts of additives and lowmolecular weight polymer units that may cause signifi­ cant problems during sample clean-up and GC analy­ sis. A number of experiments were carried out to eliminate these co-extracts. As a first step samples of the three brands of silicone tubing (length 1-3 cm, mass 1-2 g) were extracted overnight using pentane, toluene and DCM as extrac­ tion fluids. Volumetric polymer swell factors and extractable mass were evaluated. The largest swelling was observed for pentane (~ factor of 3.5). Dichloromethane caused the smallest swelling (~ a factor of 2.0). Differences between the various tubing brands were insignificant (-10%). Approximately the same extractable mass was found for the three solvents used. The highest extractable mass was found for the Masterflex tubing (25 mg g"1), followed by the Fischer Scientific tubing (22 mg g '1) and the Rubber BV tubing (18 mg g~1). The extracted amounts had an oily appear­ ance, and probably mainly consisted of silicone oil. These quantities can be expected to give large prob­ lems in sample clean-up and GC analysis. Amounts of 10 mg and 34 mg extracted residue of the Masterflex tubing were passed over an LC column filled with 2 g silica that was deactivated with 6% water. The samples were eluted with 40 mL pentane, followed by elution with 10 mL DCM. About 90% of the residue was re­ tained by the column during 40 mL elution with pentane, both for the 10 mg and for the 34 mg loading The pre-extracted tubings were extracted a second time with pentane. The concentrated extracts were passed over silica using the same procedure. The eluate was split in two parts. The first part was used for de­ termining GC blanks, the second part was used for de­ termining extractable mass residues in the LC column eluate. For the Masterflex tubing a small residue was found in the eluate (0.16 mg g ‘1 tubing). For the other tubings and for a pentane control no residue was found (detection limit 0.02 mg g"1. Therefore we concluded that the Masterflex tubing was not suitable for passive sampler construction. In addition, we concluded that for the other two tubing brands a double pre-extraction with pentane is sufficient as a tubing pre-treatment, and that a single silica clean-up removes remaining co-extracts that may be hazardous for the GC. Polymer PRC loading. Figure 6 shows the uptake curves for PRC spiking of LDPE and silicone respec­ tively. Both sampling phases can be rapidly spiked with PRCs by incubation in 80/20 (v/v) methanol-water mix­ tures. Equilibrium is generally attained within a few hours. The within-batch coefficient of variation amount­ ed to 2-6 % for PRC spiking of LDPE, as determined for five series of four control membranes prepared over a time period of about a year. For SPMDs (spiking via ad­ dition to the triolein) the variability amounted to 6-10 %. For silicone tubing not enough series have been pre­ pared so far to assess the within batch variability, but the results of the first series are promising (1-4 %) t(d ) t(d ) Fig. 5. CB153 uptake by LDPE membranes for three forced flow exposure methods: rotating disk (circles), rotating fork (squares) and pumping though empty tubing (triangles). Absorbed amounts were normalised on the sampler mass (1 , 0.5 and 0.4 g respectively) Fig 6. Uptake kinetics of Performance Reference Compounds by LDPE membranes (upper panel) and silicone tubing (lower panel) from 80/20 (v/v) methanol-water mixtures. CB4 (closed circles), CB29 (closed squares), CB155 (closed triangles), CB204 (asterisks), acenaphtene-D10 (open circles), phenanthrene-D10 (open squares) and chrysene-D12 (open trian­ gles). day # Fig. 7. Time evolution of aqueous concentrations during the laboratory experiments for pentachlorobenzene (circles), CB153 (squares), and CB194 (asterisks). All concentrations were normalised on the average concentration during the laboratory ex­ periments. Stability of the aqueous concentrations. The evolu­ tion of aqueous concentrations during the lab incuba­ tions is shown for pentachlorobenzene, CB153 and CB194 in Fig. 7. The triangles near the x-axis indicate the temperature changes from 2-30, 30-13, 13-2 and 230 °C respectively. The first temperature change form 2 to 30 °C caused a slight increase in concentrations over a one month period, followed by a large increase during the first experiment at 30 °C. Following the temperature change from 30 to 13 °C the concentrations first dropped rapidly, then increased steadily, and finally sta­ bilised. The temperature drop from 13 to 2 °C induced a small decrease in concentrations to a level that was much higher than at the start of the experiments. The temperature rise from 13 to 30 °C caused a large and rapid increase. One reason for the unstable concentra­ tions may be that the uptake rate of the samplers was of the same order as the delivery rate of the generator col­ umn. The volume rate of flow from the generator col­ umn was about 900 L d '1. The sampling rate of the SPMDs and the LDPE membranes during the 30 °C ex­ periment was about 60 L d "1 for each device. During the first day a total of 10 devices was present in the set-up, which means that the total sampling rate was about 600 L d '1. During the course of the experiments the number of samplers in the set-up decreased from 10 to 2 , there­ by enabling the generator column to better keep up with the contaminant removal by the samplers. This conclu­ sion is supported by the observation that the apparent sampling rates for the first day of the 30 °C experiment were much higher that for the later time periods. The sampling rate for day 1 was based on the concentration measured at this moment, but the average concentra­ tion during day one most likely was much higher. The decrease in concentrations following the temperature drop form 30 to 13 °C is caused in the first place by low­ er exit concentration levels from the generator column. The concentration of pentachlorobenzene showed a steady increase, probably caused by a decrease in sampling rate from about 300 to 40 L d '1. The initially steady decrease in concentration for CB153 and CB194 is probably caused by a longer equilibration time for these (more hydrophobic) compounds in the gener­ ator column. For the experiments with silicone tubing (from day 110 onwards) the total sampling rates were smaller than 40 L d '1, which allowed the generator col­ umn to efficiently stabilise the aqueous concentrations. The overall conclusion should be that the generator col­ umn method as such is suitable for stabilising the aque­ ous concentrations, but that the total volume rate of flow from the columns should have been much higher. A second conclusion is that column equilibration periods following temperature changes should be at least three weeks. Sampling rates. A summary of observed sampling rates is shown in figures 8 to 11. Sampling rates of LDPE membranes and SPMDs at a flow of 90 cm s'1 do not depend on logKow (Fig. 8, Table 1). For LDPE mem­ branes a temperature increase from 2 to 13 °C has no effect on the sampling rate (18.9 ± 3.9 L d'1 versus 18.6 ± 4.6 L d '1). A further increase to 30 °C causes an in­ crease in sampling rate by a factor of two (36.4 + 8.1 L d~1). Sampling rates for SPMDs continuously increase with temperature (18.5 ± 5.8, 25.7 ± 8.0, and 56.2 ± log K ow logK ow Fig 8 . Sampling rates (R s) of LDPE membranes (left panel) and SPMDs (right panel) at 2 °C (circles), 13°C (squares) and 30°C (asterisks). 6.5 L d "1 at 2, 13 and 30 °C respectively), but the effect of a temperature increase from 2 to 13 °C is quite small (~ factor of 1.4). For PAHs sampling rates of about 3.8 L d' 1 for 100 cm long SPMDs have been reported at wa­ ter flow velocities of 1 cm s "1 (Huckins et al., 1999). The present SPMD sampling rates at 13 °C are higher by a factor of 23 when converted to a 100 cm SPMD length (multiply by 100/30). The present values are also con­ siderably higher than PAH/PCB sampling rates of about 7 L d' 1 at 13 °C at a flow rate of 30 cm s for 30 cm long SPMDs (Booij et al., 1998). These observations once again stress the importance of water flow velocities for the observed sampling rates. The observed sampling rates at higher flows (225 cm s"1) are shown in Fig. 9. Increased sampling rates of about 40 L d "1 are observed for the logKow range 5 to 5.8. At higher logKow values sampling rates decrease to about 6 L d '1 at logKow= 7. During the high flow experi­ ments the LDPE membranes were ripped to several parallel strips that often extended up to the mounting loop seal. No damage was observed for the SPMDs. The only explanation for the decreasing sampling rates at higher Kow we can think of is that the very high shear stresses induce a change in the molecular structure of the membrane, which causes a sharp decrease in membrane permeability for higher molecular weight compounds. Sampling rates for silicone tubing are slightly de­ pendent on contaminant hydrophobicity (Figs. 10 and 11). During the 13 °C silicone experiment water was pumped through the tubing at a rate of about 25 L d '1. Sampling rates decreased from 3 L d"1 at logKOJ(l,=4 to 1.6 L d at logKovv=7. For the 2 °C and the 30 °C sili­ cone experiment the tubing was mounted on a rotating disk that rotated at 60 RPM, with one end directed to­ wards the flow and the other end extending perpendicu­ larly through the disk, as a chimney. For these experi­ ments the observed sampling rates were virtually independent of logKow, with average values of 4.1 ±0.6 and 9.3±1.3 L d"1 at 2 °C and 30 °C respectively. The difference in sampling rates between the silicone tubing and the SPMDs and LDPE membranes is largely caused by differences in surface area and flow rates. The smaller surface area (70 versus 150 cm2) accounts for a factor of two difference in sampling rates. The tub­ ing rotated at a radius of 14 cm whereas the SPMDs and LDPE membranes rotated at a 30 cm radius (~ fac­ tor 4 difference in flow rate). Although there is no simple relationship between ambient flow rates and sampling rates the effect of a decrease in ambient flow by a factor of four would be substantial. The higher mechanical strength of silicone tubing compared to LDPE most like­ ly allows for exposing these samplers under much high­ er forced flow regimes. PRC dissipation rates. Dissipation rate constants (Table 2) are shown as a function of Kow in Fig. 12. Values for LDPE membranes are higher than for SPMDs by a factor of 9.5 */+ 1.5. Each 11 to 17 °C tem­ perature increase causes an increase in dissipation rates by a factor of 2.3 */+ 1.7, both for SPMDs and LDPE membranes, with no apparent relation with T able 1. Summary o f sampling rates (R s) for SPMDs and LDPE membranes (30x2.5 cm), and silicone tubing (15cm, 6 mm id, 9 mm id) _________________ obtained during the lab experiments__________________ 2 °C 13 °C 30 °C R. SPMD LDPE silicone tubing (L d-') 19 ± 6 19 ± 4 4.3 + 1.0 Rs (L d '1) 26 ± 8 19 ± 5 Rs (L d-1) 56 + 6 36 ± 8 9.9 ± 1.2 log K ow Fig. 9. Sampling rates of LDPE membranes (circles) and SPMDs (squares) at 13°C at conditions of increased flow (225 cm s'1) log K ow Fig. 10. Sampling rates of silicone tubing at 13°C. W aterw as sucked through the tubing at a rate of 25 L d"1. \ogKow logK olv Fig. 11. Sampling rates of silicone tubing at 2 °C (circles) and 30 C (squares). The tubings were mounted on a 15 cm diam­ eter disk that rotated at 60 RPM. logKow. In a previous study we showed that the effect of flow on sampling rates and on PRC dissipation rate constants is the same (Booij et al., 1998). The value of PRC dissipation for accounting for differences in sam­ pling rates as a result of temperature differences is lim­ ited, however. For LDPE membranes no difference in sampling rates is found between 2 and 13 °C, whereas PRC dissipation rate constants increase by a factor of 2. For the 13 to 30 °C temperature range the increase in Rs by a factor of 2 is indeed matched by an increase in PRC dissipation rate constants by a factor of about 2. For SPMDs increased sampling rates by a factor of 1.4 and 2.1 are reasonably matched by increased dissipa­ tion rates by a factor of two. It does not seem wise, Fig. 12. Dissipation rate constants (ke) of the Performance Reference Compounds for LDPE membranes (filled symbols) and SPMDs (open symbols) at 2 °C (circles), 13 °C (squares) and 30 °C (triangles) however, to rely on PRC dissipation to account for the effect of temperature on the sampling rates, since the exposure temperature generally is known with reason­ able accuracy. The temperature correction in the dissi­ pation rate constants is therefore easily made, which al­ lows for estimating the pure flow effect on the sampling rates from the PRC dissipation rate constants. RESULTS AND DISCUSSION II : FIELD EXPER­ IMENTS Field survey Scheldt estuary and Southern North Sea. The amounts absorbed by SPMDs and LDPE T able 2. Summary of dissipation rate constants (ke) of the Performance Reference Compounds obtained during the lab experiments _________for SPMDs and LDPE membranes (30x2.5 cm).__________ 30 °C 13 °C 2°C ke ke SPMD ACE-D10 PHEN-D10 CB004 CHR-D12 CB029 LDPE ACE-D10 PHEN-D10 CB004 CHR-D12 CB029 ( d 1) 0.124 0.048 0.034 (d-1) 0.151 0.108 0.317 0.231 1.35 0.41 0.66 0.017 0.092 membranes during the February/March survey are shown in Fig. 13. These graphs reflect the intuitively ex­ pected geographical trend in the contamination of the water phase: Hansweert> Vlissingen- Katwijk 6 > Schouwen25 > Noordwijk60. A typical example of the geographical distribution of the contaminant amounts absorbed by SPMDs is shown for pyrene in Fig. 4. PRC dissipation data are shown in table 3. Amounts of CB155 and CB204 retained during the exposures aver­ aged 97% and 100% of the amounts in the controls, with standard deviations of 8 %. Reliable dissipation rate constants could be estimated for CB4, CB29 and chrysene-D12, both for SPMDs and for LDPE mem­ branes. We did not consider the dissipation rate con­ stants for phenanthrene-D10 to be very reliable, be­ cause of a phenanthrene-D10 contamination in the pyrene-D10 batch that was used to prepare the PAH analytical internal standard solution. The chemical noise in the acenaphtene-D10 trace on the MS, com­ bined with the short dissipation time scale (~ 6 days) prevented the calculation of a reliable dissipation rate constant for this compound. Dissipation rate constants were virtually the same for all locations, with coeffi­ K (d-1) 0.87 0.91 0.028 0.076 3.63 2.45 2.30 0.168 cients of variation of 7 to 20 %, with the exception of CB4 for the SPMDs which had a CV value of 42% due to an exceptionally high value for the Schouwen25 loca­ tion (Table 3). This high value is most likely due to stor­ age of the Schouwen25 sample at room temperature for several months instead of storage at -20 °C. The high similarity of dissipation rate constant for the differ­ ent locations means that the uptake rate constants also were the same for all locations within a small error mar­ gin. The half life times for the attainment of equilibrium can be estimated as 11 days for CB4 (logKovv=5.0), 58 days for CB29 (logKow=5.9) and 102 days for chrysene-D10 (logK 0kV=5.8). Results for LDPE membranes and for SPMDs are quite similar. The SPMD/LDPE amount ratio is shown as a function of logKow in Fig. 14. For HCB, tri- and tetrachlorobiphenyls, and three- to four ring PAHs the amounts absorbed by SPMDs are much higher than for LDPE membranes. For larger PCBs and five-ring PAHs this difference is negligible, which indicates that the sorption capacity of both samplers is sufficient to keep the sampling in the linear uptake phase for these com­ pounds. Table 3. Dissipation rate constants of the Performance Reference Compounds for the February-March _______________________ field survey (average and coefficient of variation).________________________________ ____________ log/Çm Hansweert_______ Vlissingen________ Katwijk________ Schouwen_______ Noordwijk SPMDs CV CV CV ke.SPM D ke.SPM D CV k e,SPMD CV ke.SPM D ke.SPM D CB004 CHR-D12 CB029 4.97 5.81 5.90 (d-‘) (%) 0.0495 13 0.0070 6 0.0109 9 LDPE membranes k-e. LDPE CV _____________________(d-') (%) CB004 CHR-D12 CB029 4.97 5.81 5.90 0.164 2 0.025 6 0.066 0 ( d ‘) (%) 0.0529 2 0.0064 4 0.0120 0 ( d 1) 0.0510 0.0052 0.0104 (%) - CV k e,W P E CV (d-‘) (%) (d~‘) (%) k e,LDPE 0.174 8 0.021 6 0.094 4 0.182 0.019 0.081 - (d-‘) (0.1131) 0.0066 0.0127 (%) - (d-1) (%) 0.0579 2 0.0088 8 0.0133 1 CV k eJ D P E CV (d-1) (%) (d-1) (%) k e LDPE - - 0.196 7 0.017 24 0.084 12 700 i Hansweert 600 _500 O) r 400 c I 300 200 100 - JL. jn . o T— o T— X X X 00 00 T— CNJ O CD O X o CÛ O o T— CÛ O o T— o T— o T— O T— o T— o T— o T— o T— o T— X X X X X X X X X CNJ LO O T— LU CD 00 CO 00 o 00 N- CO o o CO o Q Û Cl Q_ CÛ O T— in CÛ O CÛ O CO CO o CÛ o o LU LU X CL o Z> £ û- < m a: X o ,Q 0Û CL CO sCD CÛ O Fig. 13. Amounts absorbed by SPMDs (hatched) and LDPE membranes (white) during 38 - 43 days exposures in FebruaryMarch 1999 in the Scheldt estuary and the Southern North Sea at locations Hansweert (top), Vlissingen (middle), Katwijk6 (bot­ tom) (continued on next page) 700 n S ch o u w en 25 600 _500 O) “ 400 C o 300 H ro 200 100 0 a X CD 0 1 o T— X CO T— o CD O n n r— I ,n n n ■ «n ■ 1 1 1 «n 1 « 1 1 O o o o O o o o o O LU T— T— T— T— T— T— T— T— T— T— o X X X X X X X X X X < T— LU CM 00 <J) 00 CO 00 o o T— in Csl in o CO 00 Q T — T — O o Q CD CD CÛ CD CD CD CD CD Q. CD O O O O O O o o CL O XL LU X O < Z) CÛ CL r-d-- r-=- a: ll CL CL X o CD CD CD QZ LL^ CL CL 700 Noordwijk60 600 _500 O) r4 0 0 H C o 300 H ” 200 100 0 O T— T— T— T— o T— X X X X X X CD O X 00 00 CNJ to o CD t— o o o T— o CD O CNJ o CD o o o O T— o o T— X X T— o CD o LU Q Q Q- Q. CD o o T— o T— T— T— X X X X 00 T— CO LO 00 o 00 o CO CD CD CD CD CD O LU Ik UJ I CL fh O Z> < JD CD X o & CD CD CD T— CD O O G> o o Fig. 13 (continued). Amounts absorbed by SPMDs (hatched) and LDPE membranes (white) during 38 - 43 days exposures in February-March 1999 in the Scheldt estuary and the Southern North Sea at locations Schouwen25 (top), and Noordwijk60 (bot­ tom). LU Q. O CL to o c D o Fig. 14. Ratio of amounts absorbed by SPMDs and LDPE membranes during the February-March 1999 field survey as a function of logKow for the locations Hansweert (circles), Vlissingen (squares), Katwijkô (triangles) and Noordwijk60 (asterisks). logK c 1 0 0 0 -I Hansweert 800 ( ß /6 u ) 0 600 ■ 400 200 ■ ., o dij u. h ~i o T— o T— o T— o T— o T— o T— o T— o T— o T— X X X X X X X X X X CO 00 T— o CO 00 CNJ o m CNI in o CO X T— LU Û CT> o CO LO 00 CO o CO o h- CO Û o o o o X o Q. n-. I Q. o T— CÛ CO CÛ cd CÛ O o o o o O o o LU % z LU X CL o Z) _1 LL £ a. < CO h r-n cr X o CD ca x & rfl CL CL s CO CD CL CL 1000 Vlissingen 800 (ß /ß u ) 600 - 0 400 200 n_ o ,-Œi o 'T" X T— X CD 00 o X T— o CD o 1000 ! o rm,d T— X o 00 T— X CNJ o o CNJ CD o o T— X T— in CÛ X— X o LU Q CD Û o O ,-E L o T~ X O) 7— X CO CO CO IT) CO CO Cl Cl T— X o O o o T— X T— X o o CO CO 00 T— CO o o LU LU I O Z) co o O. m ÜL m O Katwijk6 800 (ß /ß u ) 600 - O 400 200 - m r -n rJl XL Jb- o T— o T~ o T— o T— o T— O T— o T— o T— o T— o T— X X X X X X X X X X X 00 T— o CO CNJ O CNJ in o T— o T— LU Q CT> O CO in ^— CO CO T— o 00 7— o NT— CO X CO CO CO CO O O O o Q Q- CL T— CO CO CO CO CO o o o o o LU X LU X CL 1 1,m ,m < ÜL co . n r~n .=n (Z x o CO ÛL CL CÛ CÛ Fig. 15. Contaminant concentrations in SPMDs (hatched) and mussels (white) during 38 - 43 days exposures in FebruaryMarch 1999 in the Scheldt estuary and the Southern North Sea at locations Hansweert (top), Vlissingen (middle), and Katwijk6 (bottom). 1000 Schouwen25 800 600 ■ c O 400 200 - n 1— n 1 1-1 1 1 r l 1 -Q , o T— o 1— o T--- o T— o T— X X X X X X 00 T— CO CNJ CsJ LO O CD O T— LU CD O o CD o o 00 o o— CD O O Q T pC o T— o T— X X X X X O) CO LO CO o 00 o CO T— CD Q. Q. T— LJ-I o o T— T— o T— o CD CD CÛ o o o o T o . 1— 1 -1 . r ~l~ ' 2 < LU . 1-1— 1 . CL LL^ CL CL o CÛ 00 CD X X CD Q. h- o CD O A !, LU o o 1000 1 Noordwijk60 800 • 600 " c O 400 ■ 200 - .IL _n oT o X X O CO T— o CÜ — CD X o T--- o T--- o7 X X CNJ o CD o CD T— o o CN — T— 7— T— X X X X X HI 0> CO in — X T— 00 oT o LO Q Û CD o o Q. CO CD CD Q. 00 o CÛ o — XU LU X o CO CD LU X ~1 , n - , , O D CL CL cc X o CD m co CD o O Fig. 15 (continued). Contaminant concentrations in SPMDs (hatched) and mussels (white) during 38 - 43 days exposures in February-March 1999 in the Scheldt estuary and the Southern North Sea at locations Schouwen25 (top), and Noordwijk60 (bot­ tom). 300 n 250 - 250 ■ 200 ^200 - O) I 150 o E 100 § 150 o E 100 50 ■ 50 ■ .rh .rh . r h LU O < Z LU I CL DC < > Q. 2 , CO cr X o LL ■Q CD rh CL Q_ <© CD .n ~ i ,l~h ,r h , d - i JÜ . CÛ LU î£ z LU X CL O => _i LL CO Fig. 16. Amount absorbed by SPMDs (left panel) and LDPE membranes (right panel) exposed under forced flow (hatched) and ambient flow conditions (white) during the 07/10/99 to 28/10/99 Marsdiep experiment. A comparison between contaminant absorption by mussels and by SPMDs is shown in Fig. 15. All ab­ sorbed amounts were normalised on dry weight in order to express data for mussels and SPMDs on the same scale. Geographical trends in contaminant concentra­ tions are more pronounced in SPMDs than in mussels. The concentrations of CB153 in mussels, for example, are lower at Schouwen25 than at Hansweert by a factor of two. SPMDs indicate a difference in exposure con­ centrations by a factor of seven. CB153 concentration differences between Noordwijk60 and Hansweert are a factor of three for mussels and a factor of 48 for SPMDs. Marsdiep field experiments. The first purpose of the Marsdiep experiments was to assess the merits of ex­ posing SPMDs and LDPE membranes under a forced flow regime, as opposed to exposures under ambient flow conditions. The amounts absorbed for the 07/10/1999 to 28/10/1999 period are shown in Fig. 16. For the three and four-ring PAHs the amounts absorbed by SPMDs under forced flow conditions is higher than for LPDE membranes, indicating that equilibrium has been attained for these compounds. For the five-ring PAHs no difference was observed. The amounts ab­ sorbed by SPMDs under forced flow conditions were higher than for the ambient flow exposures, with the ex­ ception of acenaphtene and phenanthrene. The differ­ ence increases with the hydrophobicity of the com­ pound and levels off to a factor of 1.8 for the compounds with the highest Kow. For LDPE mem­ branes no difference in absorbed amounts is observed up to pyrene and fluoranthene, which indicates that these compounds have attained equilibrium both for the ambient flow and the forced flow conditions. The differ­ ence between ambient and forced flow exposures is al­ so reflected in the dissipation rate constants (ke) of the PRCs (table 4). For SPMDs and LDPE membranes the ke estimates are higher for the forced flow conditions by a factor of 2.5 */+ 1.2 and 1.2 */+ 1.1 respectively. It should be borne in mind, however, that the purpose of the forced flow exposures was to gain better control over the sampling rates, rather than to increase the sampling rates. The second aim of the Marsdiep experiments was to test passive sampling by silicone tubing in the field. Figure 17 shows a large difference between the amounts of low Kow compounds absorbed by the sili­ cone sampler compared to the LDPE forced flow sam­ pler. This is caused by the higher sorption capacity of the silicone sampler. The silicone-LDPE amount ratio decreases with compound hydrophobicity, and levels off to a value of 0.5 in the high Kow range (Fig. 18). On the basis of the laboratory experiments a somewhat lower ratio was expected for the high Kow compounds. With a temperature of about 8 °C sampling rates of about 6 L d'1 are expected for the silicone tubing, com­ pared to 19 L d~1 for the LDPE membranes, which would result in an ratio of absorbed amounts of 6/19 » 0.3. The higher ratios observed in the field may indicate that the silicone sampler was insufficiently shielded off from the ambient flow. This also indicates that much higher sampling rates for the silicone sampler can be realised. CONCLUSIONS AND OUTLOOK • • • • • • Sampling rates attained by SPMDs and LDPE membranes are dramatically increased at a flow rates of about 1 m s '1, compared to literature values obtained at 1 and 30 cm s . CB4, CB29 and chrysene-D12 (logKow= 5.0, 5.9, 5.8) are suitable Performance Reference Compounds for six week exposures of SPMDs and LDPE membranes in the Dutch coastal waters. Perdeuterated fluoranthene (logKow=5.2) may be used in the future to supplement this set of PRCs. PRC dissipation rate constants for these com­ pounds in the field could be estimated with a preci­ sion of 6-25%. The temperature effect on the PRC dissipation rate constants amounts to a factor of about 2 for each 10°C temperature change. The effect of a temperature increase from 2 to 13°C on the sampling rates is limited to a factor of 1.0 and 1.3 for LDPE membranes and SPMDs respectively. For the 13 to 30 °C temperature increase the change in sampling rates amounts to a factor of 2 for both samplers. Passive sampling with silicone tubing has a great potential for monitoring organic contaminants in the Dutch coastal waters, particularly in the low Kow range (< 104 ). Passive sampling by SPMDs and LDPE mem­ branes is quite suitable for monitoring PCBs, chlorobenzenes and PAHs in the Dutch coastal wa­ ters. The differences in dissipation rate constants for the field stations in the Scheldt estuary and the southern North Sea were limited to a factor of 1.3, indicating that the sampling rates attained for ambi­ ent flow exposures are sufficiently similar. Therefore the application of forced flow samplers at these lo­ cations is not worth the effort. RECOMMENDATIONS It is useful to initiate a follow-up study on monitoring PCBs and PAHs in the Dutch coastal waters using pas­ sive samplers. We recommend that this study be struc­ tured along the following main lines. • • • The programme covers a period of five years in which the stations Hansweert, Vlissingen, Schouwen25, Katwijk 6 and Noordwijk60 are sam­ pled twice each year. Sampling is done using silicone tubing and SPMDs. The use of silicone tubing facilitates the detection of low Kow compounds, and also allows for time inte­ grative sampling of these compounds. It seems wise to also include SPMDs in the programme, in order to link the results for silicone tubing to the re­ sults obtained by the more generally accepted SPMD method. Performance Reference Compounds are used to assess differences in sampling rates. Suitable PRCs for SPMDs that are exposed for six weeks are CB4, CB29, phenanthrene-D10, fluoranthene- 300 250 .200 D) C c 150 D o E 100 50 LU Q < 2 LU X Q. O a: < d l ce ll a 3 ^ (ü X ;0 eu r-J co O cû m q. aj m Fig. 17. Amounts absorbed by silicone tubing sampler (hatched) and LDPE membranes (white) under forced flow conditions during the 07/10/99 to 28/10/99 Marsdiep experi­ ment. • • 10 and chrysene-D12. We suggest that also CB155 and/or CB204 be included as PRCs. The recoveries of these compounds can be used as control param­ eters for extraction efficiency and analytical stability. For exposures using silicone tubing we suggest that less hydrophobic PRCs are included, such as naphtalene-D 8 (logK0kV=3.35) At the beginning of the programme an operating procedure is formulated that describes in detail the sampler construction, storage, exposure, extraction, clean-up, and analysis. Special attention should be given to minimise sample handling outside the labo­ ratory, particularly when use is made of vessels that lack clean analytical facilities. Internal Reference Material (IRM) samples are pre­ pared that are analysed at regular time intervals in order to assess the overall stability of the method. REFERENCES Booij.K., Sleiderink.H.M., and Smedes,F., 1998. Calibrating the uptake kinetics of semi-permeable membrane devices using exposure standards. Environ.Toxicol.Chem. 17: 1236-1245. Ellis,G.S., Huckins.J.N., Rostad.C.E., Schmitt,C.J., Petty,J.D., and MacCarthy.P, 1995. Evaluation of logKolv Fig. 18. Ratio of amounts absorbed by silicone tubing sampler and LDPE membranes exposed under forced flow conditions during the 07/10/99 to 28/10/99 Marsdiep experiment. lipid-containing semipermeable membrane devices for monitoring organochlorine contaminants in the Upper Mississippi River. Environ.Toxicol.Chem. 14: 1875-1884. Herve.S., Prest, H. F., Heinonen.P, Hyötyläinen.T., Koistinen.J., and Paasivirta.J., 1995. Lipid-filled semipermeable membrane devices and mussels as samplers of organochlorine compounds in lake wa­ ter. Environ.Sei.Pollut.Res.Int. 2: 24-29. Huckins.J.N., Manuweera.G.K., Petty,J.D., Mackay.D., and Lebo.J.A., 1993. Lipid-containing semipermeable membrane devices for monitoring organic con­ taminants in water. Environ.Sei.Technol. 27: 24892496. Huckins.J.N., Petty,J.D., Orazio.C.E., Lebo,J.A., Clark, R. C., Gibson,V.L., Gala.W.R., and Echols,K.R., 1999. Determination of uptake kinetics (sampling rates) by lipid-containing semipermeable membrane devices (SPMDs) for polycyclic aromatic hydrocarbons (PAHs) in water. Environ.Sei.Technol. 33: 3918-3923. Huckins.J.N., Tubergen.M.W., and Manuweera.G.K., 1990. Semipermeable membrane devices contain­ ing model lipid: a new approach to monitoring the bioavailability of lipophylic contaminants and esti­ mating their bioconcentration potential. Chemosphere 20: 533-552. Table 4. Dissipation rate constants of the Performance Reference Compounds for the 07/10/99 to 28/10/99 Marsdiep exposure(average and coefficient of variation). ambient flow SPMDs forced flow ambient flow forced flow SPMDs LDPE membranes LDPE membranes iogA:„„. ke CV ke CV CV CV ke K (cf1) (d-1) (d-1) ( d 1) (%) (%) (%) (%) CB004 4.56 0.062 10 0.135 12 0.285 0 0.336 25 PHEN-D10 4.97 0.093 7 0.202 9 1 0.275 0.261 3 CHR-D12 5.81 0.006 16 0.017 30 0.041 0.032 3 4 Lebo.J.A., Zajicek.J.L., Huckins.J.N., Petty,J.D., and Peterman,P.H., 1992. Use of semipermeable mem­ brane devices for in situ monitoring of polycyclic aromatic hydrocarbons in aquatic environments. Chemosphere 25: 697-718. Petty,J.D., Huckins.J.N., Martin,D.B., and Adornato.T.G., 1995. Use of semipermeable mem­ brane devices (SPMDs) to determine bioavailable organochlorine pesticide residues in streams receiv­ ing irrigation drainwater. Chemosphere 30: 18911904. Petty,J.D., Huckins.J.N., and Zajicek.J.L., 1993. Application of semipermeable membrane devices (SPMDs) as passive air samplers. Chemosphere 27: 1609-1624. Prest,H.F., Huckins.J.N., Petty,J.D., Herve.S., Paasivirta.J., and Heinonen.P., 1995a. A survey of recent results in passive sampling of water and air by Semipermeable Membrane Devices. Mar.Pollut.Bull. 31: 306-312. Prest,H.F., Richardson,B.J., Jacobson,L.A., Vedder.J., and Martin,M., 1995b. Monitoring organochlorines with Semi-Permeable Membrane Devices (SPMDs) and mussels (Mytilus edulis) in Corio Bay, Victoria, Australia. Mar.Pollut.Bull. 30: 543-554. Prest,H.F., Jacobson,L.A., and Huckins.J.N., 1995c. Passive sampling of water and coastal air via semipermeable membrane devices. Chemosphere 30: 1351-1362. Prest,H.F., Jarman,W.M., Weismuller,T., Martin,M., and Huckins.J.N., 1992. Passive water sampling via semipermeable membrane devices (SPDMs) in concert with bivalves in the Sacramento/San Joaquin river delta. Chemosphere 25: 1811-1823. CONTENTS 1 1 C N Tt CO Summary ............................................................................................................................................................................ Introduction ........................................................................................................................................................................ Material ......................................................... Results and discussion I : Lab experiments Results and discussion II : Field experiments Conclusions and outlook ..................................................................................... ............................................................. Recommendations .............................................................................................................................................................. References ........................................................................................................................................................................ 14 14 15