AN ABSTRACT OF THE THESIS OF Pajongwit Pongsivapai for the degree of Master of Science in Chemical Engineering presented on January 14, 1994. Title: Residence Time Distribution of Solids in a Multi-Compartment Fluidized Bed System. Abstract approved:_ Redacted for Privacy VV -7 01 CUL jvv ant., v it.. In a conventional well-mixed fluidized bed the solids are almost completely mixed and the residence time for individual particles may vary between zero and infinity. For materials especially sensitive to the processing time, a wide distribution of residence time is extremely undesirable. A multicompartment fluidized bed was proposed to minimize this problem. The twocompartment fluidized bed system was designed and experimentally investigated with positive results. Residence time distributions were measured with glass bead particles of 379 p.m mean diameter size. Tracer particles were colored glass beads which had exactly the same properties as the particles used for bed material. Two theoretical models were proposed to predict the flow behavior of solids through the two-compartment fluidized bed vessel. The results show the solids flow pattern can be described by the axial dispersion plug flow model and the tanks-in-series model. The parameters of both models can be determined by minimizing the sum of the squared differences between experimental data and model predictions curves. The influence of fluidization velocity, diameter of the orifice connecting the two adjacent compartments, and the height of the overflow exit orifice were investigated. The two-compartment fluidized bed can improve the residence time distribution of solids from a single fluidized bed. The residence time distribution of solids was also improved with decreasing fluidizing gas velocity. Orifice diameter and height of the overflow exit orifice did not influence significantly on particles flow in the two-compartment fluidized bed system. However, a solids mass flow through the partition orifice is predicted reasonably well, within moderate range of variance, by equation developed by de Jong (1965) [12]. Residence Time Distribution of Solids in a Multi-Compartment Fluidized Bed System by Pajongwit Pongsivapai A THESIS submitted to Oregon State University in partial fulfillment of the requirement for the degree of Master of Science Completed January 14, 1994 Commencement June 1994 APPROVED: Z) Redacted for Privacy Professor op enemiczngineering in cnarge of major _ in Redacted for Privacy Head of Dtefiartnient of Chemical Engineering Redacted for Privacy Dean of th. .... . .0..6- 0 Date thesis is presented January 14,1994 Typed by Pajongwit Pongsivapai for Pajongwit Pongsivapai TABLE OF CONTENTS CHAPTER 1 : INTRODUCTION 1.1 Objectives 1 CHAPTER 2 : LITERATURE REVIEW 8 CHAPTER 3 : THEORETICAL APPROACHES Tracer techniques 3.2 Residence time distribution (RTD) analysis 3.3 Model of flow of solids in a non-ideal system 3.3.1 Axial dispersion plug flow model 3.3.2 Tanks-in-series model 3.4 Model of the fluidized solids flow through the orifice 3.1 CHAPTER 4 : EXPERIMENTAL EQUIPMENT AND PROCEDURES 4.1 Experimental apparatus 4.1.1 Fluidized bed 4.1.2 Fluidized bed material 4.1.3 Air supply system 4.1.4 Pressure measuring system 4.1.5 Feed system 4.1.6 Tracers 4.2 4.3 4.4 Experimental mechanism Experimental procedures Experimental plan CHAPTER 5 : RESULTS AND DISCUSSION 6 15 15 16 21 21 26 29 34 34 34 35 36 36 37 37 37 39 41 43 5.1 5.2 5.3 5.4 5.5 5.6 The measurement of minimum fluidizing velocity Tracer response curves Axial dispersion plug flow modeling Tanks-in-series modeling The effect of the fluidized bed comparmentation The effect of the orifice size in the partition 5.7 The effect of the height of exit orifice 43 44 53 63 73 76 78 5.8 The effect of fluidizing gas velocity 5.9 The modeling of solid flow through the orifice 81 5.10 Tracer recovery CHAPTER 6 : CONCLUSIONS AND RECOMMENDATIONS Conclusions 6.2 Recommendations for future work 6.1 BIBLIOGRAPHY 84 85 86 86 87 88 APPENDICES Appendix 1 : Dispersion program code Appendix 2 : The paired t-analysis Appendix 3 : Calculation of minimum fluidizing velocity Appendix 4 : Percentiles of t-distributions Appendix 5 : Gamma function table Appendix 6 : Apo and W 91 98 103 104 105 106 LIST OF FIGURES Page Figure 1-1 Flow regime in fluidized bed vessel with increasing number of compartments 3 1-2 Solids flow from dense bed to lean bed 4 3-1 E curve for solids flowing through a vessel, RTD curve 17 3-2 F curve, the cumulative distribution curve 17 3-3 E curves in closed-closed vessels for various extents of dispersion 26 3-4 E curves for various extents of number of tanks in series 29 3-5 The illustration of solid flow through the orifice 30 4-1 Schematic of apparatus for the two-compartment fluidized bed system 38 4-2 Concentration of tracer versus absorbance calibration curve 41 5-1 Pressure drop across a fluidized bed versus gas fluidizing velocity of glass beads (379 gm) 43 5-2 C curve of experiment #1 45 5-3 C curve of experiment #2 45 5-4 C curve of experiment #3 46 5-5 C curve of experiment #4 46 5-6 C curve of experiment #5 47 5-7 C curve of experiment #6 47 5-8 C curve of experiment #7 48 5-9 C curve of experiment #8 48 5-10 C curve of experiment #9 49 Page Figure 5-11 C curve of experiment #10 49 5-12 C curve of experiment #11 50 5-13 C curve of experiment #12 50 5-14 C curve of experiment #13 51 5-15 C curve of experiment #14 51 5-16 C curve of experiment #15 52 5-17 C curve of experiment #16 52 5-18 C curve of experiment #17 53 5-19 Optimization approach on experiment #9 54 5-20 Dispersion modeling for experiment #1 55 5-21 Dispersion modeling for experiment #2 55 5-22 Dispersion modeling for experiment #3 56 5-23 Dispersion modeling for experiment #4 56 5-24 Dispersion modeling for experiment #5 57 5-25 Dispersion modeling for experiment #6 57 5-26 Dispersion modeling for experiment #7 58 5-27 Dispersion modeling for experiment #8 58 5-28 Dispersion modeling for experiment #9 59 5-29 Dispersion modeling for experiment #10 59 5-30 Dispersion modeling for experiment #11 60 5-31 Dispersion modeling for experiment #12 60 5-32 Dispersion modeling for experiment #13 61 5-33 Dispersion modeling for experiment #14 61 5-34 Dispersion modeling for experiment #15 62 Page Figure 5-35 Dispersion modeling for experiment #16 62 5-36 Dispersion modeling for experiment #17 63 5-37 Tanks-in-series modeling for experiment #1 64 5-38 Tanks-in-series modeling for experiment #2 64 5-39 Tanks-in-series modeling for experiment #3 65 5-40 Tanks-in-series modeling for experiment #4 65 5-41 Tanks-in-series modeling for experiment #5 66 5-42 Tanks-in-series modeling for experiment #6 66 5-43 Tanks-in-series modeling for experiment #7 67 5-44 Tanks-in-series modeling for experiment #8 67 5-45 Tanks-in-series modeling for experiment #9 68 5-46 Tanks-in-series modeling for experiment #10 68 5-47 Tanks-in-series modeling for experiment #11 69 5-48 Tanks-in-series modeling for experiment #12 69 5-49 Tanks-in-series modeling for experiment #13 70 5-50 Tanks-in-series modeling for experiment #14 70 5-51 Tanks-in-series modeling for experiment #15 71 5-52 Tanks-in-series modeling for experiment #16 71 5-53 Tanks-in-series modeling for experiment #17 72 5-54 Comparison of single- and two-compartment RTD curves 73 5-55 Effect of compartmentation by axial dispersion plug flow model 74 5-56 Effect of compartmentation by tanks-in-series model 75 Figure Page 5-57 Effect of orifice size by tanks-in-series model 76 5-58 Effect of orifice size by axial dispersion plug flow model 77 5-59 The E curves conducted with different height of exit orifice 78 5-60 Effect of exit orifice height by axial dispersion plug flow model 79 5-61 Effect of exit orifice height by tanks-in-series model 80 5-62 Effect of fluidizing gas velocity by tanks-in-series model 81 5-63 Effect of fluidizing gas velocity by axial dispersion plug flow model 82 5-64 Measured versus predicted mass flow rate of solids through the orifice 84 5-65 The percent recovery of tracer 85 A2-1 t-distribution curve 99 A6-1 Pressure drop across the orifice for 16 experiments 106 A6-2 Weight of solid hold-up after finishing 16 experiments 107 LIST OF TABLES Em Table 2-1 Tracers used in a liquid-carrier system 2-2 Tracers used in a gas-carrier system 11 2-3 Tracers used in a two-phase system 13 2-4 Tracers used in a solid system 14 3-1 Dispersion models, Bischoff and Levenspiel 4-1 The glass beads material size distribution 36 4-2 The physical characteristics of bed material 36 4-3 Experimental plan 42 A2-1 Effect of exit orifice height when d.= 5.1 mm 100 A2-2 Effect of exit orifice height when d.= 3.5 mm 100 A2-3 Effect of orifice size when h = 10 cm 100 A2-4 Effect of orifice size when h = cm 100 A2-5 Effect of exit orifice height when d.= 5.1 mm 101 A2-6 Effect of exit orifice height when d.= 3.5 mm 101 A2-7 Effect of orifice size when h = 10 cm 101 A2-8 Effect of orifice size when h = 20 cm 101 20 8 (1962) [14] 22 NOMENCLATURES A Surface area of orifice, m2 Ab Crossectional area of bed, m2 C Tracer concentration, kg of tracer/kg of sample CD Orifice discharge coefficient, s.(m/kg)" CD Modified orifice discharge coefficient, s.(m/kg)" dp Particle diameter, m do Orifice diameter, m Tube diameter, m D Dispersion coefficient, m2/s Di, Axial dispersion coefficient, dispersion plug flow model, m2/s D'L Axial dispersion coefficient, axial dispersion plug flow model, m2/s Db. Mean value of DL(R), m2/s D'Ln, Axial dispersion coefficient, uniform dispersion model, m2/s DL(R) Axial dispersion coefficient at radial position R, general dispersion in cylindrical co-ordinates, m2/s DR Radial dispersion coefficient, dispersion plug flow model, m2/s DR, Mean value of DR(R), m2/s D 'IN, Radial dispersion coefficient, uniform dispersion model, m2/ s DR(R) Radial dispersion coefficient at radial position R, general dispersion in cylindrical co-ordinates, m2/s E(t) Exit age distribution function, s"' E(0) Exit age distribution function, dimensionless F, Flow rate of solids, kg/s g Gravitational acceleration, m/s2 h Height of exit orifice (measure from distributer plate), m hd Height of dense bed, m hi Height of lean bed, m L Distance between measurement points for dispersion experiment M Amount of tracer introduced into the vessel, kg N Number of tanks in series Apb Pressure drop cross the bed of solids, Pa Apo Pressure drop cross the orifice, Pa R Radial position, m r, Rate of chemical reaction, mole/time-volume S Source term t Time, s Mean residense of solids in vessel, s Average interstitial velocity of fluid in axial direction, m/s Fluidizing gas velocity, m/s u-und Excess gas velocity, m/s U-U Relative excess gas velocity, m/s mf uL Dispersion number 2D V Volume of tank, m3 W Average mass of soids in the vessel, kg x Axial position measured from start of test section, m Greek symbols 0 .t e Porosity or fraction voids in a fluidized bed, m3 gas/m3 bed A Solids density, kg/m3 Os Flow rate of solids through the orifice, kg/s t , reduced time, dimensionless Subscripts b Bed d Dense bed f At bubbling fluidizing condition / Lean bed mf At minimum fluidizing condition RESIDENCE TIME DISTRIBUTION OF SOLIDS IN A MULTI-COMPARTMENT FLUIDIZED BED SYSTEM CHAPTER 1 INTRODUCTION Fluidization is a subject which engineers and scientists have been studying since its impact on industrial application of a catalytic cracking process in 1942. Kunii and Levenspiel (1991) [27] define the term gas fluidization as "the operation by which solid particles are transformed into a fluidlike state through suspension in a gas". This mechanism of gas-solid contacting has some unusual characteristics which depicts several advantages of fluidization. Fluidized bed reactors have been widely acclaimed in the process industry for their highly advantageous characteristics of good mixing and contacting, high rates of heat and mass transfer, mechanical robustness, and capability of continuous operation. However, fluidized bed also has undesirable characteristics. The rapid mixing of solids in the bed gives nonuniform residence times of solids in continuous gas-solid flow vessels which is a consequence of individual solid particles having different lengths of stay in the bed. For continuous treatment of solids this leads to a nonuniform product and lower performance especially at high conversion levels. For example, in particle drying processes, the wide distribution of residence times yields a variation in moisture concentration in dried-particle products. The effect may be uneven and too high final moisture content. 2 This is a disadvantage especially when the time to achieve the desired level of moisture content is longer than the mean residence time. Solids residing too long in a dryer may deteriorate if the material of solids is sensitive to heat. Furthermore, in catalytic reactions, the movement of porous catalyst partides which continually capture and release reactant gas molecules, contributes to the backmixing of gaseous reactant, thereby reducing yield and performance, Kunii and Levenspiel (1991) [27]. The fundamental idea of this research is the improvement of residence time distribution of solid particles flowing through a fluidized bed. From the viewpoint of the residence time distribution (RTD) of solids the flow of solids through a single-compartment fluidized bed vessel can be considered as flow through a single ideal mixed vessel, Kunii and Levenspiel (1991) [27]. The best RTD of solids that we can expect in a single-compartment fluidized bed is the RTD obtained in an ideal mixed vessel. Very often, due to the short circuiting of solid flow, between inlet and outlet, much less favorable RTD's are obtained. However, our goal is to obtain RTD which are much more favorable than RTD for ideal mixing. To create conditions where all particles may have the same residence time, a basic idea of plug flow is considered. Plug flow is defined as the flow where all solids have the same residence time in the vessel. There is no diffusion relative to the solid bulk flow and no backmixing of solids of different ages. For most gas-solid reactions where high conversion of solids is expected the idea conditions of plug flow of solids is desirable. One way to approach the plug flow of solids is to connect several fluidized beds in series. Then the RTD is narrowed and moved towards that 3 (a) (b) (c) (d) Figure 1-1 Flow regime in fluidized bed vessel with increasing number of compartments 4 have been reported in literature (Guanglin et a/.(1985)[18] and Kato et ai.(1985)[24]) design to overcome the unfavorable backmixing. This idea is illustrated in Figure 1-1. Figure 1-la shows a possibility of bypassing the flow of solids which can be prevented by partitioning a single fluidized bed vessel as shown in Figure 1-1b, c and d. This approach can reduce a gross backmixing and improve the flow pattern close to plug flow. In this study the design of a two-compartment fluidized bed vessel is proposed to improve the RTD of solids in the fluidized bed vessel. The schematic of the two-compartment fluidized bed vessel is given in Figure 1-2. Solids in Solids out .A) 4 Gas Alif i ..."' Gas Figure 1-2 Solids flow from dense bed to lean bed. 5 Two fluidized beds are separated from each other by a partition. The driving force that establishes the flow of solids through the system is generated from the pressure difference across the orifice surface between two compartments. From Figure 1-2, bed 1 (left bed) and bed 2 (right bed) will be refered to as a dense and lean bed respectively. The fresh solids are fed from the top of the dense bed and discharged through the orifice to the lean bed then overflow out of the vessel at the exit orifice. RTD measurements have become an essential analytical tool in the study of solid flow through fluidized bed vessel. To study the flow pattern in a vessel it is necessary to have a good method of measuring the residence time distribution. Tracer technique which has been used by chemical engineers for almost 50 years is also employed in this work. It is a powerful tool for determining the parameters of any particular flow model representing the system. Monitoring and interpreting the actual flow pattern of individual particles is usually impractical. Therefore, the approach taken is to postulate a flow model which reasonably approximates real flow, and then use this flow model for predictive purposes. If a flow model closely reflects a real situation, its predicted response curves will closely match the tracer- response curve of the real system. This is only one of the requirements in selecting a satisfactory model. The extent of the deviation from ideal flow model will depend mainly on the geometric characteristics of the bed, the mean residence time of particles, the fluidizing gas velocity, and the size and density of the particles. Individual particles of the same size have different lengths of stay within the fluidized bed. All these effects must be taken into account if we wish to 6 control and predict the behavior of solids that pass through the system. Besides the main future of the design of the two-compartment fluidized bed vessel, this research is concerned about the influence of fluidizing gas velocity, orifice diameter in partition and mean residence time of solids. 1.1 Objectives The study presented in this thesis focuses on the development of the design of the multi-compartment fluidized bed reactor and the parameters influencing its. We are primarily interested in RTD curves of flowing solids inside the multi-compartment fluidized bed vessel. The RTD curves are considered to be a measure of a potential performance of the fluidized bed multiphase (gas-solid) chemical reactor. The specific objectives of the study are: 1. to improve RTD of solids in a single-compartment fluidized bed vessel, 2. to investigate the effects of the orifice size connecting adjacent compartments, 3. to investigate the effects of the bed height, or mean residence time of solids, 4. to investigate the effects of the fluidizing gas velocity on the RTD of solids in two-compartment fluidized bed, 5. to develop models to predict the residence time of solids in the muticompartment fluidized bed vessel, 6. to develop an equation expressing the flow rate of solids passing through the orifice. 7 There are 5 chapters in this thesis. Chapter 2 summarizes a literature review on the relevant work about tracer techniques. Chapter 3 describes the theoretical approaches of tracer techniques and the RTD analysis. A detailed description of experimental apparatus, experimental mechanism, experimental procedures and experimental plan are discussed in chapter 4. The results from experiments are discussed in chapter 5. Two models describing the flow pattern are also presented in this chapter. Finally, chapter 6 contains the conclusions of this study and recommendations for future work. 8 CHAPTER 2 LITERATURE REVIEW The idea of applying the distribution of residence time in the analysis of chemical reactor performance was apparently first systematically attempted by Mac Mullin, et al .(1935) [33]. It was the case of a number of identical completely mixed tanks in series. However, this concept was not to be used until 1952, when Danckwerts (1952) [6] explained how residence-time distribution functions can be defined and measured for real system under the assumption that the flow through out the system was steady. The use of the distribution-functions is illustrated by incorporating them in the calculation of the performance of reactors and blenders, Danckwerts (1952) [6]. A literature survey on tracers used in different studies is summarized in Table 2-1, 2-2, 2-3 and 2-4. Method of tracer detection Conduc- tivity salts Carrier fluid Tracer H2O NaNO3 Instruments Thermal or electrical conductivity cells, recorder or potentiometer Thermal or electrical NaC1 or conductivity cells, KC1 or Cl- + NaCl recorder or potentiometer or HC1 Type of reactors References packed bed [40] fluidized bed tubes tanks packed bed fluidized [40] bed Table 2-1 Tracers used in a liquid-carrier system [2] [35] [40] [40] 9 Method of tracer detection Conduc- tivity salts References Carrier fluid Tracer Instruments Type of reactors Ethyl Acetate NaC1 tubes [40] Na2S203 KC1 Thermal or electrical conductivity cells, recorder or potentiometer Thermal or electrical conductivity cells, recorder or potentiometer colorimeter tanks [40] tubes [40] + H2O black dye Color and H2O + 40% of light sensitive sugar solution dyes 1120 Color and H2O light sensitive packed bed packed red dye bed KMnO4 tubes spectrophotometer fluorescein phototube mlliamper fluidized bed recorder packed Alizarin phototube bed Saphirol [40] fluidized [31] blue dye photoelectric colorimeter colorimeter SES [40] [40] [40] [31] bed dyes sulphite spectrophotometer iodate, or photograph tubes [40] starch Color and H2O sensitive dyes Table 2-1 (continued) light sensitive solution [40] colorimeter tubes 10 Method carrier fluid of tracer detection Color and 1120 sensitive dyes Radioactive tracers Titration and others Tracer Instruments type of reactors sodium per sulphate, potassium iodide, sodium thiosulphate, starch fi-naphthol spectrophotometer tubes [40] packed bed fluidized bed tubes [40] or photograph spectrophotometer petrole- Ba' u rn sb124 F120 radioactive Geiger counter F120 NaCl or Geiger counter titration with AgNO3 HC1 Benzoric acid H2O Teepol Xylene cyanol FF solution benzene ethyl n-butyrate Table 2-1 (continued) References titration with NaOH absorbtometer refractive index measurements [40] [40] [40] tanks fluidized bed packed bed fluidized bed packed bed [40] packed bed [40] [40] [40] [40] [40] 11 Carrier fluid Air Type of Referenreactors ces CO2 thermal conductivity cells packed bed [40] packed bed anemometer [40] fluidized bed [40] gas chromatograph tubes gas analyzer [40] packed bed [40] thermal conductivity cells packed bed He [40] fluidized bed [40] fluidized bed (40] spectrometer thermal conductivity cells gas-liquid [40] 142 contractor multistage (113] gas chromatograph fluidized bed tubes modified Orsat analyzer [40] N2 fluidized bed [40] gas analyzer packed bed Ammo- gas analyzer [30] fluidized bed [30] ni a fluidized bed [40] 0-scintillation detector Kr85 packed bed Hg vapor U.V. absortion [40] Tracer SO2 butane sulfur hexafluoride N2 CO2 H2 KC1 Instruments measurement autometer flame ionization detector not reported thermal conductivity measurement gas chromatograph gas chromatograph potentiometer titration Table 2-2 Tracers used in a gas-carrier system packed bed tubes fluidized bed [40] packed bed fluidized bed tubes tubes packed bed fluidized bed [40] [40] [26] [40] [16] [16] [1] [1] 12 Carrier fluid Tracer N2 KMnO4 Instruments spectrophotometer FeC13 spectrophotometer C2 H4 katharometer SF6 katharometer H2 N2 analytical cell (ionizes the gases passing through cell) , electrical conductivity cell SF6 katharometer He 1,3 U.V. sensitive photobutadiene electric tube, spectrophotometer A gas chromatograph gas chromatograph H2 NH3 gas chromatograph N2 gas chromatograph ammonia KC1 or potentimetric titration synthesis NaC1 1,3 butadi- 1 butyne He ene A Octadeane 04 CO2 H2 N2 A oil gas H2 Sc46, C'33 Celt 3 0 Table 2-2 (continued) U.V. sensitive photoelectric tube spectrophotometer gas analyzer gas chromatograph gas chromatograph katharometer scintillation counter Type of reactors packed bed packed bed tube tube packed bed References [1] [1] [13] [13] [40] tube tube [13] tube tube tube tube packed bed fluidized bed [16] tube tube tube tube tube tube tube regenerator [40] [40] [16] [16] [16]_. [1] [1] [40] [40] [40] [16] [16] [13] [40] 13 Carrier fluid benzene - H2O -H2O Tracer Instruments HC1 for H2O , electric conductivity oil red dye for photoelectric cells benzene sargent osillometer He for air, MgSO4 or KC1 (measures the dielectric strength titration for or HCl for H2O phase kerosene - H2O NaNO3 for H2O, oil blue Air - H2O HC1 for H2O isobutanol - NaNO3 for H2O H2O, oil blue dye for Refer- reactors packed bed ences packed bed [40] solvent [40] [40] H2O phase electric conductivity photoelectric cells extrac- tion column dye for kerosene He for Air, Type of electric conductivity, thermal conductivity cells, millivolt recorder electric conductivity, special electric conductivity cells isobutanol Table 2-3 Tracers used in a two-phase system packed bed [40] packed bed [40] 14 Carrier fluid Tracer Instruments Type of reactors Air colored Si02 not reported gas-solid [18] multistage fluidized bed fluidized bed [25] polycarbonate not reported References resin Air - H2O colored glass not reported spheres Kerosene - H2O oleophilic dye not reported 1120 Benson-Lehner y- rays 'Oscar' film darkened glass particles analyzer Table 2-4 Tracer used in a solid system gas-liquid-solid multistage fluidized bed gas-liquid-liquid multistage fluidized bed liquid-solid fluidized bed [24] [24] [37] 15 CHAPTER 3 THEORETICAL APPROACHES 3.1 Tracer technique The residence time distribution (RTD) for a flowing fluid can be determined easily and directly by a widely used method of inquiry, the so- called "stimulus-response" technique. With this technique, the RTD is determined experimentally by injecting an inert chemical, particle, molecule or atom, called a tracer, into the inlet stream at some time t = 0 and then measuring the corresponding response, the tracer concentration, at the effluent stream of the reactor as a function of time. Then, a suitable flow model can then be selected by matching the experimental residence time curve with that obtained from the mathematical model. To successfully measure a RTD the tracers must have certain properties. Therefore, the type of tracer and techniques used in experimental work are of interest. The basic requirements for a satisfactory tracer experiment can be outlined as follows: 1. The tracer compound should be miscible and have similar physical properties as the main fluid/solid stream when its mixing behavior is under investigation. 2. The tracer compound should be accurately detectable in small concentrations so that the introduction of tracer does not effect the flow pattern of the main stream. 3. Generally, the tracer should be nonreactive, so the analysis of the RTD curve can be kept simple. 4. During an experiment in a multiphase system the tracer should not be transferred from one phase to another phase. 16 5. The tracer itself, its detection device and other auxiliary equipment should be relatively inexpensive. 3.2 Residence time distribution (RTD) analysis To predict the performance of a system such as a chemical reactor, we must know the pattern of fluid/solids motion through the reactor. This pattern can be determined by finding how long the individual molecule of flowing fluid/solids stay in the system; however, the complexity of the flow pattern make this method impractical and often impossible. A better approach to overcome this problem about the flow behavior in a process is to find the age distribution of the elements of the fluid/solids in the exit stream or the residence time distribution (RTD). The RTD describes the distribution of lengths of time spent by elements of the fluid inside the vessel before reaching the exit. This is termed as an exit age distribution function (E curve). The integral form of RTD of solids at the vessel exit can be expressed as a cumulative distribution function (F curve), Danckwerts (1952) [6]. An exit age distribution function describes the fraction of solids, E(t)dt, in the exit stream which has resided a time between t and (t + dt) in the vessel, Figure 3-1. It is the most used of the distribution functions connected with reactor analysis because it characterizes the lengths of time various solids spends within the reactor. The other expression has been defined as the fraction of solids which has spent a time t or less than t , Figure 3-2. 17 Fraction of solids leaving the /vessel spends time between t and t+dt in the vessel. E(t) Total area under E curve =1 t+dt t Figure 3-1 E curve for solids flowing through a vessel, RTD curve. 1.0 The fraction (F(t) ) of solids spends time t or less in the vessel. F(t) 0.0 t t Figure 3-2 F curve, the cumulative distribution curve. 18 Thus, the relation between E and F function is F(t) = 5 E(t) dt (3-1) o Besides the E and F functions, measured at the exit stream, it is possible to define, on the other hand, a distribution of residence times for which the solids currently in the vessel. It is termed as the internal age distribution function, I(t). It is a function such that I(t)dt is the fraction of solids inside the vessel that has been inside the vessel for a period of time between t and t+dt . The function E(t) is viewed outside the vessel and I(t) is viewed inside the vessel. It follows from the definitions of E(t), I(t) and F(t) that 5 E(t) dt = 1 (3-2) o 1 I(t) dt = 1 (3-3) o F(t) = 1 at t .... and (3-4) As stated previously in the tracer technique section, the RTD curve can be established by the stimulus-response technique. The stimulus can be of different forms but the most commonly used methods are pulse input and step input. If the form of stimulus is a pulse input, then the observed response at the vessel exit is called C curve, the tracer concentration versus time curve, which will be modified to E curve. The exit response of a step input yields the F curve. The selection of these two tracer-injection methods is dependent upon the applications. In this study the pulse technique is applied because it is 19 probably the most convenient for conducting this experiment. To satisfied requirements of the pules input method the injection must take place over a short period. Therefore, one of the basic assumptions in conducting this experiment is a negligible amount of dispersion between the point of injection and the entrance to the system. An amount of tracer M is injected into the solids entering the system. This experiment analyzes the injection of a tracer pulse for a single-input and single-output system in which only the flow of solids carries the tracer material across system boundaries, and where the tracer is not deminished from its flowing phase. The other basic assumption on this study is to restrict a steady-state flow with one entering stream and one leaving stream of a single fluid of constant density, Danckwerts(1952) [6]. We focus only on the unchanging size and density of particle throughout the continuous two-compartment fluidized bed system. There is no reaction taking place inside the vessel. From material balance equations, Levenspiel (1989) [28]: M Area under the C curve = F w (3-5) S t =T (3-6) where M is the amount of tracer introduced into the solids entering the vessel, W is the mass of solids in the vessel, and F, is mass flow rate of solids. Principally, the area under the C curve and the mean residence time can be calculated by equation(3-5) and (3-6) but the reactor characteristics may then be the reason for making the theoretical values deviate from experimental values. Therefore, we apply the mathematical tools which are widely used in 20 tracer work so we can calculate the numerical values from the experimental data. 00 Area under the C curve = IC dt (3-7) 0 and 00 itC dt i. o____. (3-8) fC dt 0 Since the distribution curve is known at a number of discrete time values ti then Area under the C curve = I Cl At i (3-9) all i and I, ti cieti all i t (3-10) I, CAti all i It is more convenient to normalize the C curve which will lead to the other type of RTD curve, as stated previously as E curve, in such a way that the area under the curve is equal to one, equation(3-1). Since the flow rate of solids throughout the system is assumed to be steady, we can define the RTD function, E(t), from the tracer concentration C(t) as: C(t) (3-11) E(t) = IC dt 0 21 The integral in the denominator is the area under the C curve, equation(3-7). Frequently, E(t) function is normalized respect to the dimensionless time scale,9 = t t . Therefore, a dimensionless function E(8) can be defined as E(8) = i E(t) (3-12) 3.3 Model of flow of solids in a non-ideal system There are several types of models that have found useful applications, but two are most common, at least when found to be adequate to represent the physical situation of the real system. The first is usually mentioned the "axial dispersion" or "axial dispersion plug flow" model. The second visualize various flow regions connected in series by perfectly mixed tanks, and is called tanks-in-series model. 3.3.1 Axial dispersion plug flow model The fundamental ideas of the axial dispersion plug flow model are derived on the basis of homogeneous quality of flow conditions throughout the vessel. In this study the quality of flow conditions of solids are interrupted in the orifice connecting the two compartments. Therefore, the dispersion model can not exactly explain the flow characteristics of solids in the two-compartment fluidized bed vessel. The influences of an orifice on the quality of solids flow may be eventually incorporated in flow characteristics if enough orifices are present in the vessel. This suggests that with increase in the number of compartments the dispersion model will do a better justice to the experimental data. The interruptive effect of orifices will Name of model Simplifying assumptions or restrictions in addition to those for the model above General dispersion : includes chemical Constant density reaction and sourse term Bulk flow in axial Genaral dispersion direction only. in cylindrical co- ordinates Radial symmetry Uniform dispersion Dispersion coefficients independent of position hence constant Dispersion plug flow Fluid flowing at mean velocity, hence plug flow Parameters of model D, i4 Defining differential equation dC _ Tt + u DC = V . (D. VC) + S + rc dC dC a D (R) D (R) dC R ' L ' at + ii (R) Tx = ax DL(R) Tx ii (R) D' Rm, D'L.,, + dC _ at + u dC (R) ax 1d dC R dR RD R(R) -dR + S + 7. , (T3-2) d2C No variation in properties in the radial direction D'L' 14 Table 3-1 Dispersion models, Bischoff and Levenspiel (1962) [4]. a +S+/.c ii (R) DR, DL, ll D DR = D Tx 7 + --'"L" R dR d2C X ax = 131.--ap- + at + u.. X . PA d _ X 2C R dC dR (T3-3) 43C aR R A +S+rc. Axial dispersion plug flow (T3-1) 5 T. + u Tx = DLTx-2- +S+l-c at (T3-4) (T3-5) 23 be evenly spread throughout the vessel and eventually accounted as homogeneous property of the flow quality. The most widely accepted approach for representing the mixing in flowing systems has been based on models that use diffusion equations with modified diffusion coefficients. These are termed dispersion models, and the coefficients are termed dispersion coefficients. Table 3-1 lists the mathematical equations corresponding to the various dispersion models from the most general and unwieldy to the most restricted but most easily handled. Equation(T3-4) and (T3-5) are the most commonly used because of their simplicity. Naturally, the axial dispersion model, as just stated, are useful mainly to represent flow in tubes and packed beds, however, this model can also be successfully applied to describe a variety of other types of vessels, Bischoff (1990) [15]. An attempt will be made in this thesis to make use of the axial dispersion plug flow model to predict the RTD of solids in the two-compartment fluidized bed vessel. The phenomenon of longitudinal or axial mixing is assumed to be described by the equation(T3-5). If there is no reaction in the system, the aC _ dC crC -ai, + u Ti = rY,.-7 + S + rc (T3-5) reaction rate (re) term is equal to zero. Typically, there is no source term within the system boundary, the tracer injection point is upstream from the section over which the equations are used; thus the source term (S) is equal to zero. The differential equation for the axial dispersion plug flow model is then reduced to: 24 X gC _ dC Ti. =Ddx2 -u ax (3-13) where t is the time from the commencement of the displacement, x is the distance from the point of introduction of displacement, C = C(x,t) is the solid tracer concentration, D is the axial or longitudinal dispersion coefficient which uniquely characterizes the process, it is the characteristic velocity (very often interstitial velocity) of the carrier phase, and L is the characteristic length of the system. Equation(3-13) can be put in dimensionless form by introducing z= (u t+ x) L t tu and 0 = = r the basic differential equation representing this t axial dispersion plug flow model becomes. dC _( D )d2C .dC de itL)W- az where the dimensionless group 1_3' (3-14) is the parameter which reflects the uL extent of axial dispersion. To simplify subsequent formulas, U 14 L 2D is introduced into equation(3-14) and termed the dispersion number. Therefore, the axial dispersion plug flow differential equation can be written as : X do 1 gC X 21/ di dz (3-15) 25 The two-compartment fluidized bed system in this study is considered as "closed-dosed vessel". We assume that solids enter and leave the system in plug flow manner. The entrance and exit point are designed in such a way that once the solids enter the vessel they do not diffuse back out of the system and once they leave the vessel they do not move upstream back to the system. Based on the aforementioned boundary conditions, called "closed-closed vessel", Yagi, et al .(1953) [41] presented the solution of equation(3-15) for initial condition given in the form of "S (Dirac's) function" and x=L as: E(8) = 2 I pn (Usin(pn) + pn cos(pn)) (112+ 211 +,412) pn2 ) + x exp [ U -((1.122U ] (346) with pn positive roots of cot(pn) LI pn 11-1 9 2 = (3-17) U is the parameter which measures the extent of axial dispersion. Thus it LI= L 2D --) e° : negligible dispersion (D=0), hence plug flow : 0 # 1, then E(0) = 0 : 0 = 1, then E(0) = ... U= ic L 2D >0 : large dispersion(D =oo), hence mixed flow : E(0) = exp (-0) Figure 3-3 shows the E(8) versus 8 curves of the model for various extents of back-mixing as predicted by the axial dispersion plug flow model. 26 1.5 i 4 I 1 i i 1.1=0.00 I 1.2 I U = 0.20 , II !; , t 0.9 1. t , 1 1 0 . , 1 1 11=0.50 % I 1 . f, ; " 1 t E(0) I st . 0.6 I 8 I 1 =1.00 1 1 U = 5.00 I . . , 11 i % kAt, , 11=0.69 I 1 ,. t.1,, U = 20.00 d1 % % % % i: 0.3 % i ; II 1 1 i i i '% % 1ili : I la % i I I : % I/ i . . \,%, 0.0 ' 0.0 20 10 .. . 30 40 5.0 0 Figure 3-3 E curves in closed-closed vessels for various extents of dispersion. 3.3.2 Tanks-in-series model The tanks-in-series model is another one-parameter model frequently used to simulate behavior of non-ideal flow. In this case the volume of the real system is represented by a series of identical ideal stirred tanks, and the only parameter of this model is the number of tanks (N) in series. The total volume of these equal-sized tanks is the same as that of the real system. V real = N VCSTR (3-18) 27 It happens that the tanks-in-series model gives tracer response curves that are somewhat similar in shape to those found from the axial dispersion plug flow model. Thus, this model can be applied whenever the axial dispersion model is applied. For not too large a deviation from plug flow, both models give identical results, Levenspiel (1989)[28]. The simplicity is the advantage of the tanks-in-series model. Starting from first principles one can easily develop the function representing the response of the system to Dirac's function type stimulus, Levenspiel (1972) [29]. Ea?) N(NO)(N_i),N-1 ex p (-NO) ;N =1, 2, 3, ... (3-19) The model suffers from one shortcoming, it allows only integer values of N . The usefulness of the model can be improved if N is allowed to assume non-integer values. The idea of using non-integer values of N in tanks-in-series model is reported by Hill (1977) [19]. The most obvious advantage of using non-integer values of N is to match an experimental RTD curve with the theoretical one that lies between those of adjacent N values of the tanks-in-series model. To resolve this problem the gamma function is introduced . The basic relations are: rag + 1). Nr(N) (3-20) and RN). (N-1)! then, the equation(3-19) becomes (3-21) 28 E(0) - N(N 0)14-1 1-(N) exp ( -N0) ;N 0 (3-22) The values of gamma function with non-integer values of N can be found in tables, see appendix 5. Figure 3-4 shows the E(0) versus 0 curves correspond to tank-in-series model, equation(3-22). N -4 1 : single well-mixed tank, hence ideal mixed flow N -4 oo : ideal plug flow : 0 * 1, then E(0) = 0 : 0 = 1, then E(0) = co 29 2.0 I i i 1 I I iP I II i I! , f 'II 1.6 N= ------ I '.I I ) il il !I il 9 1 N=2 N=5 ii N = 10 1.2 i 1 1 I !I I I, MI , . t N = 50 s. ft , 0.8 ------- . I I.. , a , 0.4 i 4, 1 i I I 1 tN. e 0.0 0.0 , 4. t I ; I ; ) i N = 1.5 1 t. .'. r\ , ---------- N = 20 t .1 E(0) .% i -... % .... %. ... 10 20 3.0 40 5.0 0 Figure 3-4 E curves for various extents of number of tanks in series. 3.4 Model of the fluidized solids flow through the orifice Several investigators, J. A. H. de Jong (1965) [11], Davidson (1985) [8] and Korbee et al. [20], studied the discharge of fluidized solids through the orifice. They all agreed that the discharge flow rate of solids depends on: 1. the pressure drop across the orifice, 2. the density of the fluidized solids, 3. the orifice diameter, 4. the particle size and shape. 30 Gas Gas The circle area is enlarged below r . : a,. .11 64, .. .4. ., i. ,-i ..,.4 ., --...,. :. -. 4. ... Partition . Lean bed , .1%, i A % I' I 11 I? . 1 I I 1.1 ,,, ' '4 t Dense bed i.,- " i ..,.. ,.. a. 4, Y- ' '.... . -, 4 s :l':1,.%3- , ....... 4 -1 '.1. * .: ....... 4`; T v I .r;.t4,* P.:;:.' :. ..:, d aPd ,:,. . .../ .,,e.,' e 'v:4 : " ... ,'. or, ', 4,, ,... .... ' ' y: .; ' : e :-, ., I ft -. .. .:.,...- )., ...,. .. -..-' I ,. I I " A 4 % , 44 .. Pt g 04, 4 i ,7 'I I... s Solid flow direction /. -.4 I. 1 9 1. ..p.: ... i ,, . 4 4,.... .4 . A, : :4 ..-.. I. * 1.4 44% . r 'i ,1.4 di .0 %. ... ` I . e I %%4 :?.- .8 !I. Dense bed I .4 c.:4* , I 4. t It Lean bed %. a, t ,..r b t Partition .. Figure 3-5 The illustration of solid flew through the orifice. The driving force for this flow is the pressure drop across the orifice (Apo) between the dense and lean bed, Figure 3-5. On the basis of the mechanical energy balance, a semi-empirical equation of solid mass flow has 31 been derived. In order to describe a solid mass flow through the orifice by a simple model it is assumed that particles in the vicinity of the orifice are in the state of minimum fluidization. The flow of the solids through the orifice can be viewed as a flow of liquid with a density of the fluidized bed at minimum fluidization. By assuming the uniform velocity through the crosssectional area of the orifice, the mechanical energy balance may be written in the form: 0s2 CD2 A: (3-23) where psnl = density of the fluidized solids at minimum fluidization, kg/m3 = ps (1- Cm? Ao 1/ = ivc102) = area of the orifice, m2 J. A. H. de Jong (1969) [11] concludes from his experimental data that values for discharge coefficients (CD) in equation(3-23) range from 0.45 to 0.51. In the two-compartment fluidized bed situation, the solids flowing through the orifice from the dense bed to the lean bed are assumed to have the same composition as the dense bed itself, Korbee, et al [10]. Therefore, equation(323) can be applied to derive a relation for the solids discharge rate: Co A02 2ps (1- Ed) Apo (3-24) Inside the orifice there is an area where particles can not pass or where the concentration of particles is low, Figure 3-5. Therefore, in order to correct the solids mass flow rate, Davidson (1985) [8] suggests the use of an effective 32 orifice diameter given by (do - kd,). Equation(3-24) can then be expressed as, de Jong (1965) [12] : = C,132 A.,2 2ps (1.. Ed) Apo (3-25) (1-k L do )2 (3-26) with A'. = A ci and de Jong (1969) [11] reports that k = 2.9 for sand = 1.4 for spherical particle leading to 2/ Os = CD Ao(1 -kd; k2p, (1- Ed)" (Apo)" (3-27) Then, all experiments give the constant orifice discharge coefficient (C'D ) value of 0.53 ± 0.01, de Jong (1969) [11]. In this work to determine the solid flow rate(Os) from equation(3-27), the pressure drop across the orifice (AO and the voidage of dense bed (Ed) are considered. The pressure drop across the orifice is measured directly by using the u-tube manometer while performing an experiment. The voidage of dense bed can then be calculated by using the following equation. E d = 1- APa hd(ps-Pg) g (3-28) 33 where Apd dense bed. is the pressure drop across the dense bed and hd is the height of the 34 CHAPTER 4 EXPERIMENTAL EQUIPMENT AND PROCEDURES 4.1 Experimental apparatus 4.1.1 Fluidized bed Experimental measurements of solid flow behavior were conducted in a Plexiglass cylinder with a 100.0 mm internal diameter and 6.3 mm wall thickness. The length of the cylinder measured from above the distributor plate to the top is 590.0 mm and below the distributor plate to the bottom where the air entrance holes were located is 190.0 mm. The fluidized bed cylinder was divided into two compartments by a 6.3 mm-thick Plexiglass partition; both the upper and the lower part of the vessel were partitioned. To facilitate the flow of the bed material through the system an orifice is provided at the bottom of the partition near the distributor plate. The size of the orifice was one of the investigated system parameters. The center of the orifice was located vertically 10.0 mm above the distributor plate and near the center of the partition. The fluidizing air was introduced at bottom of the fluidized bed. The air flow rate, i.e. the fluidizing velocity, was controlled independently for each compartment by a control valve and measured by a calibrated rotameter. A perforated "sandwich" design plate was used as a gas distributor. The distributor was a 3.6 mm-thick Plexiglass plate covered with a metal screen to prevent solids from raining through the orifices when the gas flow is stopped. The size and number of openings depend upon the diameter and the density 35 of particles and the height of the bed. The plate has 86 of 1.1 mm-diameter holes positioned in a triangular pitch. 4.1.2 Fluidized bed material Glass beads were used as a fluidized bed material. The minimum fluidizing velocity (um? was determined by measuring the pressure drop across the bed as a function of a fluidizing gas velocity. Data are shown in Figure 5-1. The experimental value for minimum fluidizing velocity was compared to the calculated unif from theoretical equation in appendix 3. Since the solids had a distribution of sizes, it was necessary to calculate the mean diameter for a mixture of particles of different sizes. The mean surface particle diameter was determined from the sieve size analysis. The following formula is applied to calculate the surface average diameter for mixture particles, Kuni and Levenspiel (1991) [27]. (IP 1 = (4-1) x1 all i dpi where x. = the mass fraction of particles of size interval i dpi = the average diameter of particles in interval i 1 The glass beads particles were seived and the size distribution of the batch of glass beads particles is shown in table 4-1. 36 Sieve Aperture (p.m) Weight Percent (%) 125-250 8 297-417 32 417-500 60 Table 4-1 The glass beads material size distribution. By using equation(4-1) the mean surface partide diameter of glass beads is calculated to be 379 gm. The density of glass beads was determined from volumetric and weight measurements in water. Table 4-2 summarizes the physical characteristics of the glass beads. Bed Materials Glass beads d range (I P P. u mf (exp) u mf (theo) (gm) (gm) (kg/m3) (cm /s) (cm/s) 125-500 379 2415.9 16.1 15.0 P Table 4-2 The physical characteristics of bed materials. 4.1.3 Air supply system Air was used as fluidization gas. It was supplied from the compressor available in the chemical engineering building. 4.1.4 Pressure measuring system The pressure measuring system consists of pressure taps and U-tube manometers. The pressure taps used to determine the pressure difference are positioned at both side of the vessel along the vertical axis. Pressure taps were installed right above and below the distributor plate and along the wall 37 of the fluidizing column, figure 4-1. Fine screen was added at each tap to keep the solids from plugging the lines. 4.1.5 Feed system The bed material is introduced into the fluidized bed by a gravity feed through the conical hopper and 300.0 mm long glass standpipe which are located 400.0 mm above the dense side of the two-compartment fluidized bed. 4.1.6 Tracers The basic requirements for a satisfactory tracer, are described in chapter 3. The tracer particles were prepared from glass beads and food color. These particles are identical to the particles used in the fluidized bed. The particles are coated with a solution of the Durkee-French Foods Company green food coloring and propanol. After soaking particles long enough to retain color solution on their surface the mixture is poured into a tray, spread evenly, and allowed to dry at room temperature. 4.2 Experimental Mechanism The experimental equipment is installed as shown in Figure 4-1. The fluidizing air is supplied from the building pipeline and flows through a pressure regulator which allows a maximum pressure of 24.13 kPa to reach the rest of the equipment. The air is divided into 2 lines and passed through 38 Ink 1. Two-compartment fluidized bed vessel 2. Partition 3. Partide hopper 4. Rotameter 5. Pressure probe 6. U-tube manometer 7. Pressure guage 8. Sampling cup 9. Exit orifice 10. Pressure regulator valve Figure 4-1 Schematic of apparatus for the two-compartment fluidized bed system. 39 two rotameters. Two independent lines are connected to each side of the fluidized bed. The particles are fed continuously from the hopper located above the system. Small orifice placed at the bottom of the partition, allows the particles to flow from one compartment to the other, Figure 1-1b. The pressure drop across the orifice is the driving force for this flow. The partides leave the system via an adjustable overflow exit orifice, Figure 1-2. 4.3 Experimental Procedures To begin the procedure for each experiment the preliminary experimental set-up procedures are required. 1. Determine the minimum fluidizing velocity (IQ by using the conventional fluidized bed vessel. The pressure drop across the bed (Ap b) is measured at fluidizing gas velocities uo being increased (aeration) from 0.0 cm/s to 28.0 cm/s and then at U. being decreased (deaeration) back to 0.0 cm/s. The results are shown in chapter 5, Figure 5-1. 2. Assemble and inspect the two-compartment fluidized bed vessel as shown in Figure 4-1. 3. Set the pressure regulator to 24.13 kPa. 4. Adjust the control valves of both rotameters to allow the fluidizing gas to fluidize the partides in each compartment. 5. Select the orifice size in the partition wall by changing the partition, and 6. Adjust the height of the exit orifice. After all preliminary set-up procedures are complete, we proceed as follows: 40 7. Fill the hopper with particles. 8. Allow the flow of partides reaches the steady state, 9. At time equal to zero (t = 0), instantaneously introduce the tracer (10 grams of colored partides) into the system through the top of the vessel together with the solids entering the system from the hopper. At the same time, start to collect the all particles at the exit orifice at prescribed intervals of time. 10. Measure the pressure drop across the dense (Apd) and lean (Apt) bed by the u-tube manometers at both sides of the vessel. 11. When almost all of tracer particles are out of the vessel (typically after 60 minutes of run), turn off the fluidizing gas on both sides of the bed. 12. Measure the height of dense (hd) and lean (hi) bed, and weigh the amount of solids remaining in the vessel. 13. Measure the weight of all collected sample and analyze the content of each sample for tracer particles. To find the amount of tracer in each sample, use the absorption ability of samples. Each sample is mixed with 250 cm3 of distilled water, stired and left overnight to insure that the color on the surface of the tracer particles is dissolved. Absorbance of the colored solution is then measured by using the " spectrophotometer " (Beckman, model UD-62). This instrument operates from 200 to 900 nm. It uses stable beam technology to take reading in either absorbance or transmittance. Using a visible lamp, the wavelenght of light is set to 420 nm. To learn how much of tracer was in each sample cup, the absorbance reading is converted to the concentration units by the calibration curve given in Figure 4-2. 41 7.0 0' 1 6.0 y = 3.960x - 0.000 r2 =1.000 , x o 1 5.0 a 4.0 W w uets 3.0 1... .,.. ,.... o .... 0 o 5 <4 2.0 1.0 0.0 0 02 04 06 08 1 12 14 1.6 Absorbance ( at wave length = 420 nm) Figure 4-2 Concentration of tracer versus absorbance calibration curve. 4.4 Experimental plan To satisfy all aforementioned objectives of this study, the experimental plan is created and listed as follows. 1. To investigate the improvement of the RTD in the twocompartment fluidized bed, the RTD experiment in the single compartment fluidized bed was conducted for comparison with two-compartment fluidized bed. Both experiments were set-up 42 with identical 20.8 cm/s fluidizing gas velocity (u0) and 100.0 mm exit orifice height (h). 2. To investigate the effect of the fluidizing gas velocity (u0), two sizes of the partition orifice (do = 3.5 and 5.1 mm), and two overflow exit orifice heights (h = 100.0 mm and 200.0 mm) are experimented. 3. To investigate the effect of the size of partition orifice (d.) , four levels of the gas velocity (uo = 16.2, 20.8, 25.6 and 30.5 cm/s) in both compartment, and two overflow exit orifice heights (h = 100.0 mm and 200.0 mm) are experimented. 4. To investigate the overflow exit orifice height (h), four levels of the gas velocity (uo = 16.2, 20.8, 25.6 and 30.5 cm/s), and two sizes of partition orifice (do = 3.5 and 5.1 mm) are experimented. Table 4-3 summarize the experimental plan. Each run is coded as an experimental number. Gas velocity h = 10.0 cm h = 20.0 cm u 0 (cm/s) d0 = 5.1 mm d0 = 3.5 mm d0 = 5.1 mm d0 = 3.5 mm 16.2 Exp. # 1 Exp. # 2 Exp. # 3 Exp. # 4 20.8 Exp. # 5 Exp. # 6 Exp. # 7 Exp. # 8 25.6 Exp. # 9 Exp. # 10 Exp. # 11 Exp. # 12 30.5 Exp. # 13 Exp. # 14 Exp. # 15 Exp. # 16 Table 4-3 Experimental plan. Experiment #17 was conducted in a single compartment fluidized bed vessel with 20.8 cm/s of fluidizing gas velocity and 10.0 cm of the height of exit orifice. 43 CHAPTER 5 RESULTS AND DISCUSSION 5.1 The measurement of minimum fluidizing velocity The minimum fluidizing velocity (um? for glass beads can be determined from the pressure drop-versus-velocity diagram. Figure 5-1 is the pressure drop-versus-velocity diagram of glass beads with the particle size 379 gm. 1.4 1.2 0 -. - -, -7-'Ci 1.0 i 0.8 - I -0- 0.6 -a- Areation Deareation 0.4 0.2 . 0.0 114Inf 50 10.0 15.0' = 16.1 cm/s 20.0 25.0 30.0 U. (cm/s) Figure 5-1 Pressure drop across a fluidized bed versus fluidizing gas velocity of glass beads (379 gm) 44 By increasing and decreasing the fluidizing gas velocity (u0), it has been observed that, for the relatively low flow rates in a fixed bed, the pressure drop is approximately proportional to gas velocity and then remains practically unchanged when the bed is fluidized. It is obviously seen from the diagram that the effect of inter-particle force does not make a big difference between the plot of increasing and decreasing in gas velocity. Therefore, we can neglect the effect of inter-particle force for this particles. The minimum fluidizing velocity Wm? is taken as the intersection of the pressure drop- versus-u 0 line for the fixed bed with the horizontal line. From the aforementioned procedures the minimum fluidizing velocity for glass beads can simply be determined to be 16.1 cm/s. 5.2 The tracer response curve A total of 17 RTD experiments were carried out. Investigated parameters were the fluidizing gas velocity (u), the size of partition orifice (do) and the height of overflow exit orifice (h), i.e. mean residence time. One of 17 experiments was conducted in a single compartment fluidized bed vessel. There are 4 experiments which were performed with the fluidizing gas velocity equal to 16.2 cm/s, close to the unit. No significant quantities of solids were elutriated. The original RTD curves are obtained directly from the tracer concentration, C curve. Figures 5-2 to 5-18 show the C curves obtained for different experimental conditions. 45 0.007 a 0.006 as 0.005 0.004 = 16.2 cm/s h =10.0 cm do = 5.1 mm F's = 2.37 g/s Apo = 0.4 an of H2O Uo A A A A A 0.003 A 0.002 .6A AAA 0.001 as DD 0.000 0.0 10.0 20.0 AA 30.0 40.0 50.0 t (min) Figure 5-2 C curve of experiment #1 0.007 A A A A 0.006 0.005 0 to uo =16.2 cm/s A =10.0 cm = 3.5 mm Fs = 2.32 g/s h A 0.004 do A 0.003 bo CI Apo = 1.4 cm of H20 A 0.002 0.001 A 0.000 0.0 10.0 20.0 30.0 t (min) Figure 5-3 C curve of experiment #2 40.0 50.0 46 0.0 0.0 uo = 16.2 )3 , do Ps = A 0.0 )2 =20.0 cm = 5.1 mm h ,n?6, 2.54 Apo = 0.3 ds - cm/s g/s cm of H2O A A 0.0 )1 AAA /- AA AAAA, A 0.0 )0 A 0.0 AAAAAA AA . . 20.0 10.0 30.0 40.0 50.0 60.0 70 .0 t (min) Figure 5-4 C curve of experiment #3 0.004 0.003 - A A :A 0.002 - A A A A A A uo h = 16.2 cm/s =20.0 cm do = 3.5 mm Fs = 2.37 g/s 4), Apo = 1.3 cm of H20 A 0.001 -s ?N 0.000 0.0 10.0 20.0 30.0 40.0 50.0 t (min) Figure 5-5 C curve of experiment #4 60.0 70.0 47 0.007 0.006 uo = 20.8 cm/s 0.005 =10.0 cm h 0.004 do =5.1 mm A Fs = 2.26 g/s AA 0.003 Apo = 0.3 cm of H20 A 6. 00 0.002 A A 0.001 AA *6AAA,n, A tAAAAAAAL 0.000 30.0 20.0 10.0 0.0 40.0 50.0 t (min) Figure 5-6 C curve of experiment #5 0.007 0.006 Uo A 0.004 0.003 Apo = 1.5 CM Of 0 H2O A A A A 0.002 ,..... .1.. tl _ _ = 20.8 cm/s h =10.0 cm do = 3.5 mm Fs = 2.38 g/s 0.005 AA 0.001 '4AA i 0.000 LAAAAAAAAA i 0.0 10.0 20.0 30.0 t (min) Figure 5-7 C curve experiment #6 40.0 50.0 48 0.004 ,e4A6a . A 0.003 4, A A A A = 20.8 h =20.0 cm = 5.1 nun do Fs = 2.24 A A 0.002 Apo A A A tiA A 0.001 cm/s zio g/s ___ = 0.5 cm of H2O A AAA.° n 0.000 4 AAAA'AAA -AAAAL AA . 10 . t t 20.0 10.0 30.0 40.0 50.0 60.0 70 0 t (min) Figure 5-8 C curve of experiment #7 0.004 AA Fs = 1.92 bn 0.002 _ Uo = 20.8 cm/s =20.0 cm h do = 3.5 mm A A A 0.003 Apo g/s = to cm of H20 0.001 A AA AAA AAAAAAA 0.000 . . /0 10.0 20.0 30.0 40.0 50.0 t (min) Figure 5-9 C curve of experiment #8 60.0 70 .0 49 0.007 0.006 . AAp 0.005 A = 25.6 cm/s =10.0 cm h do = 5.1 mm Fs = 2.31 g/s Apo= 0.4 cm of H2O uo A A aA 0.004 . 0.003 A A 0.002 A A A A °°° 0.001 66,Ap 0.000 . 0.0 i - 30.0 20.0 10.0 40.0 50.0 t (min) Figure 5-10 C curve of experiment #9 0.007 0.006 A 0.005 A . 5 = 3.5 mm Fs = 2.33 g/s Apo = 1.5 cm of H2O do A A 0.003 0.002 h A 0.004 A A =25.6 cm/s =10.0 cm uo °° A A a A 46, 0.001 0.000 AAA . Y 0.0 10.0 AAA A 20.0 A666,66A6,6, ise\mA . 30.0 40.0 t (min) Figure 5-11 C curve of experiment #10 50.0 50 0.005 aE (1) AA 0.004 A A u) '15 00 li0 = 25.6 LA A 0.003 $.4 A V ett I:: cm/s = 20.0 cm do = 5.1 mm Fs = 1.70 g/s Apo = 0.3 cm of H2O h _ A 0.002 '15 b0 AAA z.----% 0.001 A A AAAA A A (..) 0.000 0.0 AAA LAAA AAAA _ - - A AAA, .AAAA 20.0 10.0 30.0 40.0 50.0 60.0 70.0 t (min) Figure 5-12 C curve of experiment #11 0.004 A% A 0.003 =25.6 cm/s =20.0 cm h do = 3.5 mm Fs = 2.05 g/s Apo = 1.1 cm of H2O 6/1 U0 A 6A A 41 0.002 AA 0.001 "AAAAAAAA, --A-AAAAAA-AA AAA 0.000 0.0 10.0 20.0 30.0 40.0 50.0 t (min) Figure 5-13 C curve of experiment #12 60.0 70.0 51 0.007 A 0.006 far A 0.005 bo A A 0.004 AA A 0.003 0 bO C) uo = 30.5 cm/s =10.0 cm h do = 5.1 mm Fs = 2.33 g/s Apo = 0.4 cm of H2O 6.° A 0.002 A A 0.001 0.000 TA6° 46.6ATA6A/A61\6616' 0.0 30.0 20.0 10.0 40.0 50.0 t (min) Figure 5-14 C curve of experiment #13 0.007 ai 0.006 A (3 uo ° 44 bel = 30.5 cm/s =10.0 cm h do = 3.5 mm Fs = 2.32 g/s 0.005 0.004 A c.9 br 0.003 Apo _ = 1.4 cm Of H2O A bo 0.002 tl 0.001 A °° °° AMA6'6466A6'66AAXP 0.000 0.0 10.0 20.0 30.0 40.0 t (min) Figure 5-15 C curve of experiment #14 50.0 52 0.005 w a 4, 0.004 A LA E A cs V3 ,..... o top A A 0.003 038 A 0.002 h =20.0 cm = 5.1 mm _ Ps = 2.23 g/s A A A i..4 --- 30.5 do _ cm/s Uo Apo = 0.3 cm of H2O 66, 4 sr-t 0 66, bo AAA 0.001 .A A AAA AAAAAA t 0.000 AhhA"AAAAA 20.0 10.0 0.0 60.0 50.0 40.0 30.0 70.0 t (min) Figure 5-16 C curve of experiment #15 0.004 uo = 30.5 cm/s 0.003 , , 0.002 do Fs = 1.96 g/s A, A A =20.0 cm = 3.5 mm h AAA Apo = 1.9 cm of H2O YA 4n, 0.001 AAAA AAA , LA A I AAAA AALAiiii AA 0.000 -l ).0 . 1 10.0 20.0 30.0 40.0 50.0 t (min) Figure 5-17 C curve of experiment #16 60.0 70 53 0.009 0.008 a 0.007 u0 =20.8 cm/s A =10.0 cm Single fluidized bed h A 0.006 A A 0.005 A 00 0.004 A 0.003 66,6 A6:68 0.002 0.001 666,6,66.664 0.000 0.0 10.0 20.0 30.0 .40.0 50.0 t (min) Figure 5-18 C curve of experiment #17 5.3 Axial dispersion plug flow modeling To evaluate the model parameter U which represents the best fit to the experimental data, numerical method is applied to resolve equation(3-16) and (3-17). A FORTRAN program is developed, using half-interval approach, to calculate a series of pn . The details of this numerical program are described in appendix 1. The least-square method is used to determine model parameters from experimental data. However, to obtain the best-fit an optimization procedure must be applied. The sum of square differences are calculated for several discreate values of dispersion number (U). Several values of the sum of square differences versus the dispersion number (U), experiment #9, are shown in Figure 5-19. The quadratic equation while represents this set of data is obtained by using curve-fit function in Cricket Graph III Software (version 1.1). 54 0.148 0.147 0.146 0.145 0.144 0.143 0.142 0.76 0.80 0.88 0.84 Ti 0.92 0.96 = 0.86 Dispersion number, U = D/2uL Figure 5-19 Optimization approach on experiment #9. By setting the first derivative of this equation to zero, the dispersion number (U) which yields the best-fit between the measured and the simulated RTD curve is determined. Once the best value for dispersion number is found, the simulated RTD curve can be calculated. Figures 5-20 to 5-36 illustrate the results for the best-fit dispersion number (U) in all 17 experiments. 55 1.0 0.8 uo =16.2 h =10.0 cm do = 5.1 mm . 0.6 cm/s 11 =0.87 E(0) 0.4 0.2 . 0.0 0.0 y 'ilk . ionloW 05 10 15 20 25 30 35 4.0 45 5.0 8 Figure 5-20 Dispersion modeling for experiment #1 1.0 0.8uo = 16.2 cm/s h = 10.0 cm do = 3.5 mm 0.6- U = 0.89 E(0) 0.4- 0.2 'A 0.0 . 0.0 AA . A 05 10 15 20 25 30 35 A 4.0 45 8 Figure 5-21 Dispersion modeling for experiment #2 5.0 56 1.0 0.8 uo = 16.2 cm/s h =20.0 cm 0.6 do = 5.1 mm U=0.71 E(8) 0.4 0.2 eAAAA A AAAAAA 0.0 . 0.0 w Aaa Akik . AA /A w 05 10 15 20 25 30 3.5 4.0 45 5.0 e Figure 5-22 Dispersion modeling for experiment #3 1.0 0.8 uo = 16.2 cm/s h =20.0 cm 0.6 do = 3.5 mm u =0.84 E(0) 0.4 0.2 0.0 0.0 05 10 15 20 25 3.0 3.5 40 45 5.0 0 Figure 5-23 Dispersion modeling for experiment #4 57 1.0 0.8 uo = 20.8 cm/s h =10.0 cm 0.6 do = 5.1 mm E(6) 11 = 0.53 0.4 0.2 . 0.0 0.0 05 10 15 A 20 2.5 AA,..AAA A. m 30 35 4.0 45 5.0 8 Figure 5-24 Dispersion modeling for experiment #5 1.0 0.8 - uo = 20.8 cm/s h =10.0 cm do = 3.5 mm U = 0.75 E(0) 0.0 05 10 15 20 25 30 35 40 45 0 Figure 5-25 Dispersion modeling for experiment #6 5.0 58 1.0 0.8 AA o i U0 = 20.8 h =20.0 cm 0.6 E(8) A 0.4 cm/s do = 5.1 min U = 0.71 p A - 0.2 AA, A 1hiAaa ' 20 25 30 35 40 45 AA, ahALAaA 0.0 A L 0.0 05 10 15 5.0 9 Figure 5-26 Dispersion modeling for experiment #7 1.0 0.8 - uo = 20.8 cm/s h =20.0 cm 0.6 do = 3.5 mm E(0) U = 0.76 0.4 0.2 Atha AhAia 0.0 /ALA, Av 0.0 05 10 15 20 25 3.0 35 40 45 9 Figure 5-27 Dispersion modeling for experiment #8 5.0 59 1.0 0.8 uo = 25.6 cm/s h =10.0 cm 0.6 do = 5.1 mm E(0) 11 = 0.86 0.4 0.2 A 0.0 05 10 15 20 0.0 . Av . , i 30 35 40 45 2.5 5.0 9 Figure 5-28 Dispersion modeling for experiment #9 1.0 0.8 64 A A uo = 25.6 cm/s h =10.0 cm 0.6 do = 3.5 mm U = 0.75 E(0) 0.4 0.2 A 0.0 0.0 ., 05 10 15 20 25 30 35 --4.0 4.5 e Figure 5-29 Dispersion modeling for experiment #10 5.0 60 1.0 0.8 uo = 25.6 cm/s h =20.0 cm 0.6 do = 5.1 mm U = 0.69 E(0) 0.4 0.2 0.0 0.0 10 15 20 25 30 35 40 45 05 5.0 Figure 5-30 Dispersion modeling for experiment #11 1.0 0.8 IJAhk AA UO = 25.6 cm/s h =20.0 cm ._ 0.6 ,A E(8) do = 3.5 mm U = 0.73 A, 0.4 va, .. A A 0.2 -AAA 1 AA,A AA AAAAAA 0.0 ialakAiAL4M/A ! 0.0 0.5 10 15 20 25 3.0 35 40 45 0 Figure 5-31 Dispersion modeling for experiment #12 5.0 61 1.0 0.8 A uo = 30.5 cm/s h =10.0 cm 0.6 do = 51 mm E(8) 1.1 = 0.78 0.4 A 6 0.2 A, AvA.A. 0.0 0.0 05 10 15 20 25 30 35 -- -A 4.0 .\1611416 45 5.0 e Figure 5-32 Dispersion modeling for experiment #13 1.2 A 1.0 Uo = 30.5 0.8 cm/s h =10.0 cm do = 3.5 mm E(0) Li = 0.01 0.6 0.4 0.2 . 0.0 A . 0.0 0.5 . . 10 15 20 25 30 3.5 40 45 9 Figure 5-33 Dispersion modeling for experiment #14 5.0 62 1.0 0.8 uo = 30.5 cm/s h =20.0 cm 0.6 do = 5.1 mm lI = 0.63 E( 0) 0.4 0.2 " AAAAAAA, AIAAA 0.0 05 0.0 10 30 25 20 15 35 4.0 4.5 5.0 0 Figure 5-34 Dispersion modeling for experiment #15 1.0 0.8 0.6 o uo = 30.5 cm/s a,.A h =20.0 cm 3.5 mm U =0.36 A dO = A E(0) IA 0.4 Ill, 0.2 Ilkl *A IL ImA1141,A i hith 0.0 1 0.0 . 1 nr 05 10 15 20 25 30 35 40 45 I . 0 Figure 5-35 Dispersion modeling for experiment #16 5.0 63 1.0 0.8 uo = 20.8 . =10.0 cm Single fluidized bed h 0.6 u = 0.00001 . E(9) 0.4 cm/s AA . i 0.2 ... a AA 0.0 :-.' 0.0 1 05 10 15 . 20 25 30 3.5 40 45 5.0 Figure 5-36 Dispersion modeling for experiment #17 5.4 Tanks-in-series modeling The number of tanks (N) which is a parameter in equation (3-22) is calculated on computer spread sheet by Microsoft Excel (version 3.0). Using the sum of the squared differences technique to compare the experimental points with the calculated points on the RTD curve. The identical optimization approach in the dispersion modeling is also applied in order to determine the value of the best-fit N value. The N values obtained base on this optimization approach were round off to a single decimal number. These outcomes are shown in Figures 5-37 to 5-53. 64 1.0 0.8 A A Uo = 16.2 0.6 cm/s h =10.0 cm A do = 5.1 ram N =2.0 E(8) 0.4 : 0.2 A 0.0 . . . A A &AAP 05 10 15 20 25 30 35 40 45 0.0 5.0 9 Figure 5-37 Tanks-in-series modeling for experiment #1 1.0 A A 0.8 A A uo =16.2 cm/s h =10.0 cm do = 3.5 mm 0.6 E(0) N =2.1 A 0.4 A A 0.2 A A*AA 0.0 AA_AAA 05 10 15 20 25 30 35 40 45 5.0 e Figure 5-38 Tanks-in-series modeling for experiment #2 65 1.0 0.8 uo = 16.2 cm/s h =20.0 cm 0.6 do = 5.1 mm N = 1.9 E(9) 0.4 0.2 0.0 05 10 15 20 0.0 2.5 30 3.5 40 45 5.0 Figure 5-39 Tanks-in-series modeling for experiment #3 1.0 0.8 IAA. uo = 16.2 0.6 cm/s h =20.0 cm A dO = 3.5 min N . 2.0 E(6) 0.4 r 0.2 4A, 'A AA AAAA4A 'AAAA/AAAAAA - AAA 0.0 . 0.0 05 10 15 20 25 30 3.5 APA 40 45 5.0 0 Figure 5-40 Tanks-in-series modeling for experiment #4 66 1.0 0.8 uo = 20.8 0.6 h =10.0 cm A E(0) cm/s do = 5.1 mm N =1.90 AAA ° 0.4 A 0.2 L A AA 4p A,A6AL, AAAAA AA 0.0 05 10 0.0 30 35 40 45 20 2.5 15 A 5.0 9 Figure 5-41 Tanks-in-series modeling for experiment #5 1.0 0.8 uo = 20.8 cm/s h = 10.0 cm 0.6 do = 3.5 mm N = 2.0 E(0) 0.4 0.2 0.0.4 0.0 . . 05 10 15 . . 20 2.5 3.0 35 40 4.5 5.0 9 Figure 5-42 Tanks-in-series modeling for experiment #6 67 1.0 0.8 AAIA Aw =20.8 cm/s h =20.0 cm do = 5.1 mm N =1.9 UP A A A 0.6 A, E(0) p A A 0.4 At 0.2 AAA ,AA I/AAA/AAA AAAPAAA i 0.0 . .0 05 10 15 20 2.5 3.0 35 40 45 5.0 0 Figure 5-43 Tanks-in-series modeling for experiment #7 1.0 0.8 A AA uo =20.8 cm/s A h =20.0 cm 0.6 . 3.5 min N =1.9 dp E(0) 0.4 AAA 0.2 PA Aik A, AkAAA AkiAlAi hApA AA 0.0 AA f ).0 f 05 10 15 1 20 25 i 3_0 3.5 4n ac ci 0 Figure 5-44 Tanks-in-series modeling for experiment #8 68 1.0 0.8 A uo = 25.6 cm/s h =10.0 cm 0.6 do = 5.1 mm N . 1.9 E(0) 0.4 0.2 A A 0.0 A A 0.0 05 10 15 25 30 35 2.0 4.0 45 5.0 Figure 5-45 Tanks-in-series modeling for experiment #9 1.0 0.8 a uo = 25.6 A cm/s h =10.0 cm 0.6 do . 3.5 mm N = 1.9 E(9) 0.4 0.2 AA A 0.0 1 0.0 05 1.0 1 15 20 AA . 2.5 AA A 30 35 40 45 5.0 9 Figure 5-46 Tanks-in-series modeling for experiment #10 69 1.0 0.8 o uo = 25.6 cm/s h =20.0 cm 0.6 do = 5.1 nun N . 1.8 E(9) 0.4 A AA AA A A 1AAA 0.2 A L IIAeAA LAir /"Li AWAA 0.0 1 0.0 05 10 15 2.0 25 30 35 40 45 5.0 9 Figure 5-47 Tanks-in-series modeling for experiment #11 1.0 0.8 AA. AAA U0 = 25.6 0.6 cm/s h = 20.0 cm A do = 3.5 mm A N = 1.8 E(0) 0.4 A IA. 0.2 AA r 1,A AAiA1 A/AAAA11 -AA AAALAA 0.0 I 0.0 I A 05 10 15 20 25 30 35 40 45 5.0 Figure 5-48 Tanks-in-series modeling for experiment #12 70 1.0 A 0.8 A uo = 30.5 cm/s h =10.0 cm 0.6 do = 5.1 mm N = 1.7 E(0) 0.4 0.2 A 0.0 . 0.0 1 . 05 10 15 20 2.5 A 30 35 40 45 5.0 0 Figure 5-49 Tanks-in-series modeling for experiment #13 12 1.0 uo =30.5 cmis 0.8 we) h =10.0 cm do = 3.5 mm N = 1.1 0.6 0.4 A A 0.2 AVA 0.0 05 10 15 , AA 20 2.5 30 35 40 45 A 5.0 e Figure 5-50 Tanks-in-series modeling for experiment #14 71 1.0 0.8 uo = 30.5 cm/s h =20.0 cm 0.6 do = 5.1 mm N =1.8 E(0) 0.4 A 0.2 0.0 05 10 0.0 15 2.0 25 3.0 35 4.0 45 5.0 8 Figure 5-51 Tanks-in-series modeling for experiment #15 1.0 0.8 uo = 30.5 cm/s h =20.0 cm 0.6 do = 3.5 mm N = 1.6 E(0) 0.4 0.2 /AA -A A A. A A'Aii,#,AA, A 0.0 . 0.0 1 I 05 10 15 2.0 25 30 35 40 45 5.0 8 Figure 5-52 Tanks-in-series modeling for experiment #16 72 1.2 1.0 A 0.8 cm/s h =10.0 cm Uo A A E(8) Single fluidized bed N = 1.1 o o 0.6 = 20.8 A A ' 0.4 A A A A 0.2 . A A AAA 0.0 .Z1 0.0 . I 1 05 1.0 . 15 A A A 4.4 -A A 20 25 30 35 40 45 1 5.0 0 Figure 5-53 Tanks-in-series modeling for experiment #17 73 5.5 The effect of fluidized bed compartmentation The tracer experiment in a single compartment fluidized bed vessel is conducted (experiement #17). The obtained RTD curve is compared to the RTD curve obtained from the two-compartment fluidized bed vessel under exactly the same experimental conditions (experiment #6). Figure 5-54 shows that the RTD curve of solids improves substantially to the direction of plug flow when fluidized bed vessel is partitioned into two compartments. 1.2 a 1.0 0.8 o o oath A E(8) 0.6 A A A 0 0 0 A Single-compartment fluidized bed ,,,'.1. .0 5, . 0.0 0.0 0 A 00a 0.2 Two-compartment fluidized bed A 00 A 0 A oo AA o 0 A 0.4 A 05 10 15 20 25 30 A t 35 Yv.. 40 45 5.0 0 Figure 5-54 Comparison of single- and two-compartment RTD curves. 74 To decribe the improvement of RTD curves in Figure 5-54, the axial dispersion plug flow and the tanks-in-series models are used to simulate the curves. Figure 5-55 shows the best-fit curves and model parameter values on experiment #6 and #17 by the axial dispersion plug flow model. The dispersion number (II) is improved from 0.00001 (single compartment) to 0.75 (two compartments). By tanks-in-series model, Figure 5-56 also illustrates the same trend of improvement from the number of tanks (N) 1.1 to 2.0. 1.0 I 0.8 Exp. #6 A %'s t ' I I 4. . U = 0.75 r4 0 44 0 0.6 Exp. #17 u = 0.00001 . E(0) _ \A.6, ..t, 0.4 *. , 0.2 . 0.0 i 0.0 . -,,, sr .... ire ti72,A'VAFA7A7AT I I 05 10 15 20 25 3.0 3.5 V' te TA1 40 45 5.0 9 Figure 5-55 Effect of compartmentation by axial dispersion plug flow model. 75 1.0 0 0.8 As ,& ; 4 ti E( , 0.4 0.2 ; Exp. #6 N = 2.0 o A 0.6 " A o. .. A Exp. #17 _ _ N = 1.1 .A. 4 fA tA4p, S A<Ie A i-AvAt,,,,,,.. Af 0.0 0.0 05 10 15 2.0 25 3.0 I 7 -4° ° 35 40 4.5 5.0 8 Figure 5-56 Effect of compartmentation by tanks-in-series model. 76 5.6 The effect of the orifice size in the partition An increase in gas velocity leads to a decrease in model parameter values. The model parameters-versus-relative excess gas velocity diagrams are shown in Figures 5-57 and 5-58. They demonstrate the effect of orifice size for the same overflow height. Both axial dispersion plug flow and tanks-inseries models give the same tendency of the results. In each figure the slope of both lines are negative, however, the slope for the small orifice diameter (do = 3.5mm) is steeper. 2.4 2.2 2.0 .... 6- -----1.8 - ------ - -------- --- --6-- 1.6 - 1.4 - 0 ---&-1.2 h= 200.0 mm, do = 5.1 mm, do/dp = 14 h = 200.0 mm, do = 3.5 mm, do/dp = 9 1.0 0.0 0.2 0.4 0.6 0.8 1.0 (U-Umf)/ Umf Figure 5-57 Effect of orifice size by tanks-in-series model. 77 1.2 1.0 0.8 O --2 ........ A 0.6 0.4 0.2 _ . 0.0 0 0 ---16.-- h = 200.0 mm, do = 5.1 mm, do/ dp = 14 _ h = 200.0 mm, do = 3.5 mm, do/dp = 9 i 0.0 0.2 0.4 1 1 0.6 0.8 1.0 (U-Umf)/Umf Figure 5-58 Effect of orifice size by axial dispersion plug flow model. Therefore, model parameters are more sensitive to variations of the fluidizing gas velocity for the 3.5 mm diameter orifice size. However, the influence of orifice size can not be supported by a statistical analysis of data. The paired t-analysis (appendix 3) is applied to test the effect of orifice sizes. It is demonstrated in appendix 3 that there is no difference between the orifice size (d) 3.5 mm and 5.1 mm. Thus, the variation of orifice size in this study can not show the influences on model parameters. 78 5.7 The effect of the height of exit orifice The primary effect of the height of the exit orifice can be seen from E(t)- versus-t curve in Figure 5-59. Naturally, the mean residence time of solids is 0.12 Exp # 10 Fs = 2.33 gis h =100.0 nun 0.10 1:P 0.08 - w Exp # 12 Fs = 2.05 gis Mean residence time = 8.85 min (h =100.0 mm) - h = 200.0 nun Mean residence time = 16.18 min (h = 200.0 mm) oA 1:1646AA 0.03 OA .A EI:13 bA66666 A 0.00 ,0 0.0 466A&AA ChICPCP . 10.0 20.0 1666606 PEISITMEITLED 40.0 30.0 50.0 time (min) Figure 5-59 The E curves conducted with different height of exit orifice. longer for the bed with longer solids hold-up and roughly the same solids flow rate; equation(3-6). However, the statistical results show no difference between the exit orifice heights 100.00 mm and 200.00 mm (appendix 3). The paired t-analysis is employed to test the hypothesis that the bed with longer mean residence time (longer exit orifice height, h = 200.0 mm) and the bed with shorter mean 79 residence time (h = 100.0 mm) give identical model parameter values. The results of the statistical analysis show no difference in model parameters due to different bed heights, different mean residence times (appendix 3). Figures 5-60 and 5-61 support this results and illustrate no significant influences of the exit orifice height for orifice size (do) 5.1 mm and 3.5 mm respectively. 1.2 1.0 -.... ..... o 0.8 -.. -.... e 0.6 -... 0.4 -... .. -., . 0.2 . 0.0 --IN-- h = 100.0 mm, do = 3.5 mm 0-- h = 200.0 mm, do = 3.5 mm A 1 0.0 -%.. 0.2 . 1 0.4 - 1 0.6 . t 08 1.0 (14-Umf)/limf Figure 5-60 Effect of exit orifice height by axial dispersion plug flow model. 80 2.4 2.2 2.0 0 1.8 1.6 1.4 - A - -- h= 100.0 mm, do = 5.1 mm 1.2 .0 h = 200.0 mm, do = 5.1 mm 1.0 0.0 0.2 0.4 0.6 08 1.0 (14-Unst)/ Unit Figure 5-61 Effect of exit orifice height by tanks-in-series model. 81 5.8 The effect of fluidizing gas velocity Our visual qualitative observations in all experiments support the statement that the quality of fluidization depends on fluidization velocity. 2.4 2.2 2.0 .... ... ft. .......... .. ..11....... . ..... ..ACA 1.8 - .. ...., ... .. ... ....4..., . . ..... ... ..., A 1.6 h = 100.0 nun , do = 5.1 mm 1.4 h = 200.0 mm, do = 5.1 mm 0-- h= 100.0 mm , do = 3.5 mm 1.2 -0- 1.0 h = 200.0 nun , do = 3.5 mm 1 0.0 0.2 . 1 0.4 . 1 0.6 . 08 1.0 (U-Umf)/ Umf Figure 5-62 Effect of fluidizing gas velocity by tanks-in-series model. The particle mixing in the bed is very sensitive on the excess gas velocity. The mixing quality of the beds are obviously varied by changing the level of fluidizing gas velocity (u0). Therefore, one of the major parameters influencing the solids mixing rate in the bed is the fluidizing gas velocity (u0). 82 When operating at gas velocities close to the minimum fluidization velocity (um ?, it was found that the solids could not be readily fluidized. Often the dense side of the bed would defluidize. However, if the fluidization velocity is increased beyond um! ,the solids would fluidize smoothly. 1.2 1.0 0.8 CY --------- ----- ---......... 0.6 ................... h = 100.0 mm , do = 5.1 mm 0.4 h = 200.0 mm , do = 5.1 mm 0 h= 100.0 mm , do = 3.5 mm 0.2 h = 200.0 mm , do = 3.5 mm 0.0 1 0.0 0.2 0.4 0.6 08 1.0 (U-Umf)/ Umf Figure 5-63 Effect of fluidizing gas velocity by axial dispersion plug flow model. The influences of fluidizing gas velocity on the model parameters, both the number of tanks in series and the dispersion number, are shown in Figures 5-60 and 5-61 respectively. An increasing gas velocity results in a 83 decreasing values of the model parameters; the dispersion number(U) and the number of tanks (N) in series. On the other hand, a decrease in gas velocity leads to higher values for model parameters, thus, improving the flow pattern of solids. 84 5.9 The modeling of solid flow through the orifice The measured mass flow rate of solid through the orifice is plotted against the predicted value in Figure 5-64. The predicted value is calculated from equation(3-27) in chapter 3. The predictions are shown to be in very satisfactory agreement with the experimental results. Raw data of pressure drop across the orifice are plotted in Figure A6-1 (appendix 6). 4.2 3.8 3.4 0 3.0 O/e ie , 1.8 1.4 1.0 ,, 1.0 , , ,, ,- 14 ,, , cy,o o 0 ,-'o , 18 /e , / / / 0 2.6 2.2 / // / // /e / ,/ / // ,, , ,, o 11:6411". o o 22 26 30 34 38 Measured mass flow rate (g/s) Figure 5-64 Measured versus predicted mass flow rate of solids through the orifice. 4.2 85 5.10 Tracer recovery The material balances on the amount of tracers are calculated for 16 experiments. In each experiment the amount of tracers are recovered from sample in each sample cup and the bed material remained in the vessel after finishing that experiment. The percent of recovered tracers are plotted in Figure 5-65. The average of tracer recovery is approximate around 92%, therefore, the tracer technique used in this study is trustworthy. 130 E3 Recovery from sample 120 - Recovery from remained bed 80 70 60 50 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 Experiment # Figure 5-65 The percent recovery of tracer. 86 CHAPTER 6 CONCLUSIONS AND RECOMMENDATIONS 6.1 Conclusions Both the axial dispersion plug flow and the tanks-in-series models were proposed to describe the residence time distribution of solids flowing through the two-compartment fluidized bed vessel. Experiments for investigating the influences of the fluidizing gas velocity, the size of the partition orifice, and the height of the overflow exit orifice were carried out to test the models and evaluate the dispersion number (U) and the number of tanks (N). The results of this study can be summarized as follows. 1. The axial dispersion plug flow and the tanks-in-series models satisfactorily describe the solids flow in the two-compartment fluidized bed vessel. 2. The dispersion number (U) and the number of tanks in series (N) decrease with the increase in gas velocity. The lower the excess gas velocity (uo-umf), the more favorable RTD can be obtained. 3. Within the range of experiments performed in this study, the size of the orifice diameter in the partition and the height of the overflow exit orifice do not effect the model parameters,U andN. 4. The solid mass flow rate through the orifice between two adjacent compartments can be predicted reasonable well by equation(3-27). 87 6.2 Recommendations for future work This study is an initial investigation of multi-compartment fluidized bed reactor. Future work will necessarily incorporate longer number of compartments. We believe that the increase in number of compartments can further improve the residence time distribution of solids. Recommendations for future experimentation are given as follows: 1. To investigate the influence of the orifice size and the exit orifice height. Longer variations of orifice diameter and different exit orifice location are recommend. 2. The location of the orifice in the partition does not necessarily need to be close to the distributor plate. The effect of the position of the orifice may also be of interest. 3. The design of an orifice which would allow solids to move only one direction (one way valve) from one compartment to the other is of extreme importance. 88 BIBLIOGRAPHY 1. Argo, W. B. and Cova, D. R. , Ind. Eng. Chem., Process Design and Development, Vol. 4, pp. 352 (1965). 2. Allen, C. M. and Taylor, E. A. , Trans. Amer. Soc. Mech. Engrs, Vol. 45, pp. 285 (1923). 3. Brenner, H. , "The diffusion model of longitudinal mixing in beds of finite length (numerical value)", Chemical Engineering Science, Vol. 17 pp. 229-243 (1962). 4. Bischoff, K. B. and Levenspiel, 0. , "Fluid dispersion-generalization and comparison of mathematical models_I (Generalization of Models)", Chemical Engineering Science, Vol. 17, pp. 245-255 (1962). 5. Butt, J. B. , Reaction Kinetics and Reactor Design, Prentice-Hall, New Jersey (1980). 6. Danckwerts, P. V. , "Continuous Flow Systems Distribution of Residence Times", Chemical Engineering Science, 1(2):1-13 (1952). 7. Danckwerts, P. V. , Jenkins, J. W. and Place, G. , "The distribution of Residence-Times in an industrial fluidized reactor", Chemical Engineering Science, (3):26-35 (1954). 8. Davidson, J. F. , Clift, R. and Harrison, D. , Fluidization Second Edition , Academic Press, London (1985). 9. Denbigh, K. G. , Turner, J. C. R. , Chemical Reactor Theory Third Edition, Cambridge University Press, London (1984). 10. Dudukovic, M. P. , Mills, P. L. , Chemical and Catalytic Reactor Modeling, American Chemical Society, Washington D.C. (1984). 11. de Jong, J. A. H., "Vertical Air-Controlled Particle Flow from a Bungker Through Circular Orifices", Powder Technol. , 3:297-286 (1969/70). 12. de Jong, J. A. H.,"Aerated Solids Flow Through a Vertical Standpipe Below a Pneumatically Discharged Bunker", Powder Technol. , 12:197-200 (1975). 13. Evan, E. V. and Kenney, C. N. , Proc. Roy. Soc. (London), Vol. A284, pp. 540 (1965). 14. Fogler, H. S., Elements of Chemical Reaction Engineering Second Edition , Prentice-Hall, New Jersey (1992). 89 15. Froment, G.F. , Bischoff, K. B. , Chemical Reactor Analysis and Design second Edition, John Wiley & Sons, New York (1990). 16. Giddings, J. C. and Seager, S. L. , Ind. Eng. Chem. Fundamentals, Vol. 1, pp. 277 (1962). 17. Gilliland, E. R. , Mason, E. A. , Ind. Eng. Chem., 41:1191 (1949). 18. Guanglin, S. , Lin, N. , Xiaoyu, C. and Gantang, C. , "Flow and Performance Study in a Gas-Lift Multistage Fluidized Bed", Fluidization'85 Science and Technology, Science Press, Beijing (1985). 19. Hill, C. G.,Jr. , An Introduction to Chemical Engineering Kinetics & Reactor Design , John Wiley & Son, New York (1977). 20. Korbee, R. , Schouten, J. C. and Van den Bleek, C.M.,"Modelling Interconnected Fluidized Bed Systems", AIChE Symposium Series No. 281, Vol. 87. 22. Klinkenberg, A. , "Residence Time Distributions and Axial Spreading in Flow Systems (With their Application in Chemical Engineering and Other Fields)", Trans. Instn Chem. Engrs., Vol. 43, pp. T141-T156 (1965). 23. Kreyszig, E. , Advanced Engineering Mathematics seventh edition, John Wiley & sons, New York(1993). 24. Kato, Y. , Morooka, S. and Nishiwaki, A. ,"Behavior of Dispersed- and Continuous-Phase in Multi-stage Fluidized Beds for Gas-Liquid-Solid and Gas-Liquid-Liquid systems" ,Fluidization'85 Science and Technology, Science Press, Beijing, pp.217-227 (1985). 25. Kececioglu, I. , Wen, C. Y. and Keairns, D. L. , "Fate of Solids Fed Pneumatically through a Jet into a Fluidized Bed", AIChE Journal, Vol. 30, No. 1, pp. 99-110 (1984). 26. Krambeck, F. J. , Avidan, A. A. , Lee, C. K. and Lo, M. N. , "Predicting Fluid-Bed Reactor Efficientcy Using Absorbing Gas Tracers", AIChE Journal, Vol. 33, No. 10, 1727 (1987). 27. Kunii, D. , Levenspiel, 0. , Fluidization Engineering Second Edition, Butterworth-Heinemann, Stoneham (1991). 28. Levenspiel, 0. , The Chemical Reactor Omnibook, OSU Book Stores, Corvallis (1989). 29. Levenspiel, 0. , Chemical Reaction Engineering Second Edition, John Wiley & Sons, New York (1972). 90 30. Muchi, J. , Mamuro, T. and Sasaki, K. , Kagakukogaku, Chemical Engineering (Japan), Vol. 25, pp. 747 (1961). 31. Murphree, E. V. , Alexis, V., Jr. and Mayer, F. X. , Ind. Eng. Chem., Process Design and Development, Vol. 3, pp. 381 (1964). 32. Molodtsof, Y. , Ould-Dris, A. and Bognar, G.,"Particle Flow Through Orifices and Solids Discharge Rate from Hoppers", Fluidization VII , 363-370. 33. MacMullin, R. B. and Weber, M. , Chem. Met. Eng., 42:254 (1935). 34. Naor, P. and Shinnar, R. , "Representation and Evaluation of Residence Time Distributions", I &EC Fundamentals, 278-286. 35. Parent, J. R. and Roy, P. H. , Symposium on New Data in Fluid Mixing, AIChE 57th Meeting, Boston, Dec. 6-10 (1964). 36. Resnick, W. , Heled, Y. , Klein, A. and Palm, E. , "Effect of differential Pressure on Flow of Granular Solids Through Orfices", I & EC Fundamentals , 5(8):392-396 (1966). 37. Richardso, J. F. and Carlos, C. R. , "Partide Speed Distribution in a Fluidized system", Chemical Engineering Science, Vol. 22, pp. 705-706 (1967). 38. Smith, J. M. , Chemical Engineering Kinetics Third Edition, McGraw Hill, New York (1981). 39. Shah, Y. T. , Gas-Liquid-Solid Reactor Design, McGraw-Hill, New York (1979). 40. Wen, C. Y. , Fan, L. T., Models for Flow Systems and Chemical Reactors , Marcel Decker, New York (1975). 41. Yagi, S. and Miyauchi, T. , "On the Residence Time Curves of the Continuous Reactors", Kagakukogaku, Chemical Engineering (Japan), Vol. 17, No. 10, pp. 382-386 (1953). 42. Zhang, J. Y. and Rudolph, V., 'Packed Solids Downflow in Standpipes", Fluidization VII , 371-379. 43. Mendenhall, W. and Sincich, T. ,Statistics for the Engineering and Computer Sciences Second Edition, Dellen, San Francisco (1988). APPENDICES 91 APPENDIX 1 DISPERSION PROGRAM CODE Variable Listing Program Symbol Definition E E(8) values which are calcylated from equation(11) ETHETA E(0) values read from input file for each experiment IMAX Maximum number of iterations mu p N Number of roots, p, from equation(12) PI it XL Left boundary of current interval XLO Original left boundary of search interval XN Midpoint of current interval XR Right boundary of current interval XRO Original right boundary of search interval THETA 0 values read from input file for each experiment TOL Tolerance criterion U tiL =D , Dispersion number (U) FORTRAN Code for the dispersion program PROGRAM DISPERSION * * The purpose of this program is to determine the dispersion number * (U) which is the parameter in the axial dispersion plug flow model, 92 * equation(11). The result of this analysis is the parameter U that * is the sum of the squared differences between theoretical E(0) values by * equation(11) and experimental E(0) values. * * Main program * REAL XL, XR, XN(1000), TOL, TERM(1000), THETA(100), U REAL mu(1000), E(100), PI, SUM(100), ETHETA(100) INTEGER IMAX, M, N CALL INPUT (N, U, THETA, ETHETA, PI) CALL HALF (N, U, mu, PI) CALL RTD (mu, N, E, U, THETA, ETHETA) CALL LEAST (THETA, ETHETA, E, SUM) END * * The data input subroutine which read the value of 0 and E(0) from * each experimental data. Both values are located in EXPERIMENT file. * SUBROUTINE INPUT (N, U, THETA, ETHETA, PI) REAL U, THETA(100), ETHETA(100), PI INTEGER N PI = 22.0/7.0 N = 30 OPEN (UNIT=111, FILE.'EXPERIMENT.', STATUS='OLD') DO 10 I = 1,100 READ (111,20) THETA(I), ETHETA(I) 10 CONTINUE 93 20 FORMAT (F6.4,1X,F6.4) PRINT *,' ENTER THE VALUE OF "U" ?' READ *, U END The half-interval method is applied as a root-solving technique. The basic idea is initialized with an interval containing a root and reduces that interval by half. SUBROUTINE HALF (N, U, mu, PI) REAL XL, XR, XN(1000), TOL, mu(1000), U, PI INTEGER I, IMAX, M, N TOL =1E -9 IMAX =1000 XL0 = PI/2. XRO = PI XL = XLO XR = XRO INDEX =1 DO 30 M=1,N IF ((INDEX .EQ. 1) .AND. (XL .GT. U)) THEN XRO = XLO XR = XRO XLO = XRO-PI/2. XL = XL0 INDEX = 2 94 ENDIF DO 40 I=1,IMAX XN(M)=(XL+XR)/2.0 FL =FUN (XL, U) FR=FUN (XR, U) FN=FUN (XN(M), U) IF ((FN *FL) .LT. 0.0) THEN XR=XN(M) FR=FN ELSE XL=XN(M) FL=FN ENDIF IF (ABS(XR -XL) .LT. TOL) GOTO 50 40 CONTINUE 50 XL0 = XL0 + PI XR0 = xizo + PI XL = XLO XR = XRO mu(M)=XN(M) OPEN(1,FILE='OUTPUT1',STATUS=1UNKNOWN') WRITE(1,60) M, mu(M), INDEX 30 CONTINUE 60 FORMAT (9X,I3,',1,4X,F15.4,',',1X,I2,',$) END * * The theoretical E(0) values corresponding to 8 values which read from 95 input file are evaluated by using equation(11). * SUBROUTINE RTD (mu, N, E, U, THETA, ETHETA) REAL TERM(1000), THETA(100), E(100), mu(1000) REAL ETHETA(100), U INTEGER I, M, N, Q DO 90 Q=1,100 E(Q) = 0.0 DO 70 M=1,N TERM(M)=2.0 1 *(mu(M)*(U*SIN(mu(M))+mu(M)*COS(mu(M)))) 1 /(U **2.+ 2.*U + mu(M)**2.) 1 *EXP(U-((U**2.+mu(M)**2.)/(2.*U))*THETA(Q)) E(Q) = E(Q) + TERM(M) 70 CONTINUE DO 80 N=1,100 IF (THETA(N) .EQ. 0.0) THEN ETHETA(N) = 0.0 E(N) = 0.0 ENDIF 80 CONTINUE OPEN(1,FILE=tOUTPUT',STATUS='UNKNOWN') WRITE(1,100) THETA(Q), ETHETA(Q), E(Q) 90 100 CONTINUE FORMAT (8X,F9.5,',',7X,E12.7,',',7X,E12.7,1;) END 96 * The experimental data set points is compared to the theoretical one * which calculated from the model. The sum of the squared * differences of all data points is to be minimized. * SUBROUTINE LEAST (THETA, ETHETA, E, SUM) REAL THETA(100), E(100), ETHETA(100) REAL DIFF(100), SUM(100) INTEGER N, I DO 110 N=1,100 IF (THETA(N) .EQ. 0.0) THEN ETHETA(N) = 0.0 E(N) = 0.0 ENDIF 110 CONTINUE SUM(0) = 0.0 DO 120 I=1,100 IF (E(I) .LT. 0.0) THEN E(I) = 0.0 ENDIF DIFF(I) = (ETHETA(I)-E(I))*42. SUM(I) = SUM(I-1) + DIFF(I) 120 CONTINUE PRINT 130,SUM(I-1) 130 FORMAT ('The sum of the squared differences = ',F12.5) END * * This is the function to determine the vulues of "p", equation(12) 97 REAL FUNCTION FUN( X, U ) REAL X, U IF (X .EQ. 0) THEN X =1E -06 ENDIF FUN = 1./TAN(X) - (X/U-U/X)/2. RETURN END 98 APPENDIX 2 THE PAIRED tANALYSIS To perform a paired t-analysis, Mendenhall (1988) [43], we simply calculate the treament (group) difference separately for each pair. The set of differences constitutes a random sample from a single population of such differences. When there is no difference in the original groups, the population of differences will be centered at zero. Therefore the analysis estimates the mean difference with particular attention paid to the possibility that the mean might be zero. In the situation we are concerning, the potency of the population of differences can also be less or greater than the mean difference. So a two-tailed test of a paired t-analysis is applied. Let introduce X and X2 as the data from population 1 and 2 1 respectively, and n is the number of pairs. The population of difference (denoted by D) can be calculated from: Di = X11 X2 1 i = 1, n The null and alternative hypotheses can be presented as follows: Ho : µD =0 H : AD* 0 The null hypothesis (Ho) ) states that the mean of the population 1(/./.1) is equal to the mean of the population 2 (µ2). Therefore, the mean of population of differences (lip ) is equal to zero. On the other hand, the alternative hypothesis (Ha) states that the mean of the population 1(iti) is not 99 equal to the mean of the population 2 (222). Thus, the mean of population of differences (up ) is not equal to zero. For a small-sample with two-tailed test, a test statistic can be decribed by: b t-statistic = SD/(n , 05) (b is the average of the population of differences = n with v = n -1 , pc, is the hypothesized mean difference which is equal to zero in this case, and SD is the standard deviation of the population of differences. In this analysis the critical value (a) which is commonly accepted to be 0.05, indicate the level of significance of t-analysis (95%). The region of rejection can be shown as: f(t) a/2 = 0.025 t 0.0 Rejection re:;11 Figure A2-1 t-distribution curve t 0.975 t ection region 100 The following is a calculation of t-statisic values. The hypothesized mean difference (pp ) is equal to zero and the degrees of freedom are equal to 3. Data from axial dispersion plug flow model h=10 cm h=20 cm difference h=10 cm h=20 cm difference 0.87 0.71 0.16 0.89 0.84 0.05 0.53 0.71 -0.18 0.75 0.76 -0.01 0.86 0.69 0.17 0.75 0.73 0.02 0.78 0.63 0.15 0.10 0.36 -0.26 D 0.0750 D -0.0500 SD 0.1702 SD 0.1421 t-statistic 0.8813 t-statistic -0.7036 Table A2-1 Effect of exit orifice height when do = 5.1 mm 5 d 0=5.1mm d =3.mm difference Table A2-2 Effect of exit orifice height when do = 3.5 mm 1 d0 =5.mm d 0=3.5 rrun difference 0.87 0.89 -0.02 0.71 0.84 -0.13 0.53 0.75 -0.22 0.71 0.76 -0.05 0.86 0.75 0.11 0.69 0.73 -0.04 0.78 0.10 0.68 0.63 0.36 0.27 D 0.1375 D 0.0125 SD 0.3863 SD 0.1763 t-statistic 0.7119 t-statistic 0.1418 Table A2-3 Effect of orifice size when h = 10 cm Table A2-4 Effect of orifice size when h = 20 cm 101 Data from tanks-in-series model h=10 cm h=20 cm 0.1 2.1 2.0 0.1 1.9 0.0 2.0 1.9 0.1 1.9 1.8 0.1 1.9 1.8 0.1 1.7 1.8 -0.1 1.1 1.6 -0.5 h=10 cm h=20 cm 2.0 1.9 1.9 difference difference D 0.0250 D -0.0500 SD 0.0957 SD 0.3000 t-statistic 0.5222 t-statistic -0.3333 Table A2-5 Effect of exit orifice height when do = 5.1 mm d0=5.1mm d 0=3.5mm difference Table A2-6 Effect of exit orifice height when do = 3.5 mm d0=5.1min d 0=3.5mm difference 2.0 2.1 -0.1 1.9 2.0 -0.1 1.9 2.0 -0.1 1.9 1.9 0.0 1.9 1.9 0.0 1.8 1.8 0.0 1.7 1.1 0.6 1.8 1.6 0.2 D 0.1000 D 0.0250 SD 0.33665 SD 0.1258 t-statistic 0.5941 t-statistic 0.39736 Table A2-7 Effect of orifice size when h = 10 cm Table A2-8 Effect of orifice size when h = 20 cm Because the degrees of freedom are equal to 3 and the critical value (a) is equal to 0.05, the tabulated values (appendix 5) for the regions of rejection 102 are t-statistic < -3.182 andt-statistic > 3.182. Therefore, all above tests show that there are no sufficient reasons to belive that the variations of orifice size (3.5 and 5.1 mm) and the height of exit orifice (100.0 mm and 200.0 mm) are influence to the dispersion number (U) and the number of tanks (N). 103 APPENDIX 3 Calculation of Minimum Fluidizing Velocity The minimum fluidizing velocity (u..? can be determined by using the equations from Kunii and Levenspiel (1991) [27] which are recommended by Chitester et al.. dup p mi d3 p g [ (28.7)2 + 0.0494( v g(p9 It where d P = mean diameter of particle, m Pg = density of fluidizing gas, kg/m3 It = viscosity of fluidizing gas, kg/m.s p, = density of solids partide, kg/m3 g = acceleration of gravity, 9.8 m/s' -p)g ' Hun - 28.7 104 APPENDIX 4 Percentiles of t-distributions t (A; v)} = A Entry is t (A; v) where P t (A; v) A 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 .60 .70 .80 .85 .90 .95 .975 0.325 0.289 0.277 0.271 0.267 0.727 0.617 0.584 0.569 0.559 1.376 3.078 1.886 1.638 0.920 1.533 1.476 6.314 2.920 2.353 2.132 2.015 12.706 0.978 1.963 1.386 1.250 1.190 1.156 0.265 0.263 0.262 0.261 0.260 0.553 0.549 0.546 0.543 0.542 0.906 0.896 0.889 0.883 0.879 1.134 1.119 1.108 1.100 1.093 1.440 1.415 1.397 1.383 1.372 1.943 1.895 1.860 1.833 1.812 2.447 2.365 2.306 2.262 2.228 0.260 0.259 0.259 0.258 0.258 0.540 0.539 0.537 0.537 0.536 0.876 0.873 0.870 0.868 0.866 1.088 1.083 1.079 1.076 1.074 1.363 1.356 1.350 1.345 1.796 1.782 2.201 0.258 0.257 0.257 0.257 0.257 0.535 0.534 0.534 0.533 0.533 0.1165 1.071 0.863 0.862 1.069 1.067 1.066 1.064 1.337 11.333 0.257 0.256 0.532 0.532 0.532 0.531 0.531 0.859 0.858 0.858 0.857 0.856 1.063 0.531 0.531 0.256 0.256 0.256 0.256 0.256 0.256 0.256 0.256 40 60 0.255 120 0.254 0.253 40 0.25.4 1.061 0.941 1341 1.771 1.761 1.753 4.303 3.182 2.776 2.571 2.179 2.160 2.145 2.131 1.746 1.740 1.734 1.729 1.725 2.120 2.110 1.721 1.060 1.059 1.058 1.323 1.321 1.319 1.318 1.316 2.080 2.074 2.069 2.064 2.060 1.058 1.057 1.056 1.055 1.055 1.315 1.314 1.313 1.311 1.310 1.706 1.703 0.530 0.530 0.530 0.856 0.855 0.855 0.854 0.854 1.699 1.697 2.056 2.052 2.048 2.045 2.042 0.529 0.527 0.526 0.524 0.851 0.848 0.845 0.842 1.050 1.045 1.303 1.296 1.289 1.282 1.684 2.021 1.671 2.000 1.980 1.960 0.861 0.860 1.061 1.041 1.036 1.330 1.328 1.325 1.717 1.714 1.711 1.708 1.701 1.658 1.645 2.101 2.093 2.086 105 APPENDIX $ Gamma function table x T(x) x 1(x) 1.00 1.01 1.02 1.03 1.04 1.05 1.06 1.07 1.08 1.09 1.00000 .99433 .98884 .98355 .97844 .97350 .96874 .96415 .95973 .95546 1.50 1.51 1.52 1.53 1.54 1.55 1.56 1.57 1.58 1.59 .88623 .88659 .88704 .88757 .88818 .88887 .88964 .89049 .89142 .89243 1.10 1.13 1.14 1.16 1.16 1.17 1.18 1.19 .95135 .94740 .94359 .93993 .93642 .93304 .92980 .92670 .92373 .92089 1.60 1.61 1.62 1.63 1.64 1.65 1.66 1.67 1.68 1.69 .89352 .89468 .89592 .89724 .89864 .90012 .90167 .90330 .90500 .90678 1.20 1.21 1.22 1.23 1.24 1.25 1.26 1.27 1.28 1.29 .91817 .91558 .91311 .91075 .90852 .90640 .90440 .90250 .90072 .89904 1.70 1.71 1.72 1.73 1.74 1.75 1.76 1.77 1.78 1.79 .90864 .91057 .91258 .91467 .91683 .91906 .92137 .92376 .92623 .92877 1.11 1.12 1.30 1.31 1.32 1.33 1.34 1.35 1.36 1.37 1.38 1.39 .89747 .89600 .89464 .89338 .89222 .89115 .89018 .88931 .88854 .88785 1.80 1.81 1.82 1.83 1.84 1.85 1.86 1.87 1.88 1.89 .93138 .93408 .93685 .93969 .94261 .94561 .94869 .95184 .95507 .95838 1.40 1.45 1.46 1.47 1.48 1.49 .88726 .88676 .88636 .88604 .88581 .88566 .88560 .88563 .88575 .88595 1.90 1.91 1.92 1.93 1.94 1.95 1.96 1.97 1.98 1.99 .96177 .96523 .96877 .97240 .97610 .97988 .98374 .98768 .99171 .99581 1.50 .88623 2.00 1.00000 1.41 1.42 1.43 1.44 106 APPENDIX 6 Apo and W Pressure drop across the orifice (Apo) ) and the amount of solid hold-up (W) measured after turning of the fluidization gas for all 16 experiments are shown in the curves below. 2.5 h = 100.0 mm 2.0 - h = 200.0 mm O 0 1.5 - Q 0 0 1.0: o 0.5 GI 0.0 I 0 0 CI 1 o I I o I I 2345 o I I i I I I El iii o I I 6 7 8 9 10 11 12 13 14 15 16 17 Exp # Figure A6-1 Pressure drop across the orifice for 16 experiments. 107 2400.0 o 2000.0- 0 0 0 In 0 Q 0 1600.0- A 1200.0- 4 A A 800.0- 400.0 0 A A 1 2 A h= 100.0 mm 0 h = 200.0 mm 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 Exp # Figure A6-2 Weight of solid hold-up after finishing 16 experiments.