AN ABSTRACT OF THE THESIS OF Daniel C. Konnoff for the degree of Master of Science in Radiation Health Physics presented on March 23, 2012. Title: SSPM-Based Optical Fiber Radiation Dosimeter Abstract approved: Abi T. Farsoni Current state-of-the-art environmental, clinical, and in-vivo radiation sensing systems utilizing various inorganic and tissue-equivalent plastic scintillators are not user friendly, suffer from electron-beam-generated noise, and are difficult to deploy successfully for real-time dosimetry. A robust, real-time detection system using different scintillating materials coupled to solid-state detectors by optical fibers is developed. This system enables radiation monitors/clinicians to conduct meaningful real-time measurements using different inorganic scintillators or organic, tissue-equivalent plastic scintillators in harsh clinical and environmental environments. Recent solid state photomultiplier (SSPM) technology has matured, reaching a performance level that is suitable for replacement of the ubiquitous photomultiplier tube in selected applications for environmental radiation monitoring, clinical dosimetry, and medical imaging purposes. The objective of this work is laboratory and clinical testing of the Hamamatsu MPPC (S10362-11-050C), Photonique SSPM (0810G1), and Voxtel SiPM (SQBF-EKAA/SQBF-EIOA) SSPMs coupled to different inorganic scintillator crystals (Prelude 420, BGO), inorganic doped glass scintillator material SiO2 : Cu2+ , and organic BCF-12 plastic scintillating fibers, used as detector elements. Both polymer optical fibers (POFs) and glass optical fibers (GOFs) are used as signal conduits for laboratory and clinical testing. Further, reduction of electron-beam-generated Cerenkov light in optical fibers is facilitated by the inclusion of metalized air-core capillary tubing between the BCF-12 plastic scintillating fiber and the POF. Dose linearity, percent depth dose, and angular measurements for 6 MV/18 MV photon beams and 9 MeV electron beams are compared using the Hamamatsu MPPC with-and without the use of the metalized air-core capillary tubing for BCF-12 plastic scintillating fiber. These same measurements are repeated for SiO2 : Cu2+ scintillator material without air-core capillary tubing. c Copyright by Daniel C. Konnoff March 23, 2012 All Rights Reserved SSPM-Based Optical Fiber Radiation Dosimeter by Daniel C. Konnoff A THESIS submitted to Oregon State University in partial fulfillment of the requirements for the degree of Master of Science Presented March 23, 2012 Commencement June 2012 Master of Science thesis of Daniel C. Konnoff presented on March 23, 2012 APPROVED: Major Professor, representing Radiation Health Physics Head of the Department of Nuclear Engineering & Radiation Health Physics Dean of the Graduate School I understand that my thesis will become part of the permanent collection of Oregon State University libraries. My signature below authorizes release of my thesis to any reader upon request. Daniel C. Konnoff, Author ACKNOWLEDGEMENTS First and foremost, I would like to thank Dr. Thomas Plant from the School of Electrical Engineering and Computer Science for acting as my co-advisor. Dr. Plant has provided me with excellent advice and has worked hard on my behalf, supporting my research efforts and kindly making his own laboratory available to me. Further, Dr. Plant read all of the manuscript and made constructive criticisms. Any remaining mistakes are, of course, my responsibility alone. I would also like to thank Dr. Abbi Farsoni for acting as my advisor. I would like to thank Dr. Kathryn Higley for giving me the opportunity to become an RHP student, and for helping me transition to the on-campus program, and for many thoughtful discussions. I would like to thank Dr. David M. Hamby. I am extremely grateful for the initial research opportunity you extended to me and for making your own laboratory and equipment available. I would like to thank several members of my “unofficial” thesis committee. Dr. Michael Lerner from the Department of Chemistry engaged in many valuable discussions. Manfred Dittrich made all the optical jig-assemblies needed to complete this project in a timely manner. Without Manfred’s excellent machine work none of the laboratory or clinical experiments would have been possible. Elizabeth Shiner at Good Samaritan Hospital in Corvallis Oregon who gave generously of her time and provided LINAC beam time for testing. I would like to thank my friends and those staff and faculty in various Oregon State University departments who assisted me in many ways while I was there. This was not always an easy thing to do. I would also like to thank my two fathers for instilling within me the importance of education at an early age. Moreover, I would like to thank Mr. Roshan Patel from Hamamatsu Corp., San Jose CA., for providing MPPC samples, Dr. David McNally from Photonique SA, Geneva Switzerland. for SSPM samples, Dr. Vinit Dhulla from Voxtel Corp., Beaverton OR. for SiPM samples, Mr. Don Doize from PolyMicro Technologies for samples of their metalized air-core capillary tubing, Dr. Alan Huston from the US Navy Research Laboratory for samples of SiO2 : Cu2+ scintillating fiber material, Dr. David Akselrod from the University of Oklahoma (now Landauer Inc.) for samples of Al2 O3 : C OSL material, and Mr. Michael Mayhugh from Saint-Gobain Crystals for samples of their inorganic/organic scintillator material. This research was sponsored by a seed seed grant ESE151 from the OSU College of Engineering and by employment in the Colleges of Engineering and Business, also by Peet’s Coffee and Tea. TABLE OF CONTENTS Page 1. INTRODUCTION AND PROJECT OBJECTIVES . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2. Goals of this project . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.3. Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 2. OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 2.1. Overview of Photon Interactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 2.2. Overview of Electron Interactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Collisional Stopping Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Radiative Stopping Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 10 2.3. Cerenkov Radiation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 2.4. Specific Ionization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 2.5. Solid State Photomultipliers (SSPMs) as Photomultiplier Tube (PMT) Replacements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 2.5.1 Avalanche Noise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5.2 Spectral Response and Quantum Efficiency . . . . . . . . . . . . . . . . . . . . . . 2.5.3 SSPM Biasing, Equivalent Circuit, and Speed of Response . . . . . . . 2.5.4 Dynamic Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5.5 Dark Count - Causes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Temperature Dependence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Afterpulses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Optical Crosstalk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5.6 SSPM S/N Ratio and Noise Considerations . . . . . . . . . . . . . . . . . . . . . . 2.5.7 Comparison of SSPM vs. Vacuum Tube (PMT) Technology . . . . . 2.6. 14 16 18 21 22 24 25 26 26 29 Important Characteristics of Optical Fibers - Glass, Plastic . . . . . . . . . . . . . 30 2.6.1 Light Propagation in OFs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.6.2 Metalized Air-Core Capillary Tube Cerenkov Light Removal . . . . . Other Cerenkov Removal Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.6.3 Other Optical Fiber Radiation Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 37 38 39 TABLE OF CONTENTS (Continued) Page 2.7. SiO2 : Cu2+ Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 3. EXPERIMENTAL METHODS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 3.1. SSPMs and Device Characterization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 3.1.1 3.1.2 3.1.3 3.1.4 3.1.5 3.1.6 3.2. Photon Counting Electronics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-V and CV Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SSPM Gain Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SSPM DCR Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SSPM PDE Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . GRIN Lens Design for the Voxtel SQBF-EKAA . . . . . . . . . . . . . . . . . 43 46 50 50 52 53 Optical Coupling using Adhesives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 3.2.1 Characteristics of Optical Fibers used in this work . . . . . . . . . . . . . . . 58 3.3. Scintillators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 3.4. Optical Fiber System Efficiency Considerations . . . . . . . . . . . . . . . . . . . . . . . . . 63 3.4.1 Estimated Detected Light from Scintillators . . . . . . . . . . . . . . . . . . . . . 3.4.2 Efficiency: Overall Light Coupling. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4.3 Optical Signal-To-Noise-Ratio (S/N) Considerations . . . . . . . . . . . . . 63 64 67 3.5. SiO2 : Cu2+ Fiber Optic Probe Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70 3.6. BCF-12 Fiber Optic Probe Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 3.6.1 Angular Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 3.7. Dose Linearity Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 3.8. Percent Depth Dose Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79 4. RESULTS AND DISCUSSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 4.1. Photonique SSPM Comparisons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 4.2. Photonique, MPPC, and Voxtel SSPM Comparisons . . . . . . . . . . . . . . . . . . . . 91 TABLE OF CONTENTS (Continued) Page 4.3. Clinical Results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102 4.3.1 Mitsubishi Eska GH4001 POF Cable Clinical Background Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3.2 Angular Dependence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3.3 Dose Linearity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3.4 Depth Dose Measurements: Photon and Electron Beam . . . . . . . . . 102 102 102 107 4.4. Summary of Clinical Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110 4.5. Uncertainty Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114 5. CONCLUSIONS AND RECOMMENDATION FOR FUTURE WORK . . . . . . . 116 5.1. Optical and Electrical Characterization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116 5.2. Recommendations for Future Work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118 BIBLIOGRAPHY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122 APPENDICES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132 A APPENDIX A TEC Cooler Schematic for VOXTEL SQBF-EKAA . . . . 133 B APPENDIX B Fiber Efficiency Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . 135 C APPENDIX C Laboratory Dose Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . 137 D APPENDIX D Decay of Laboratory Sources, Number of Photons Emitted, and Prelude 420 Crystal Activity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138 D1 D2 D3 Decay of Laboratory Sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138 Number of Photons Emitted by Decayed Sources . . . . . . . . . . . . . . . . 138 Prelude 420 (Lu1.8 Y.2 SiO5 : Ce) Crystal Activity . . . . . . . . . . . . . . . . . 139 LIST OF FIGURES Figure Page 1.1 Detector system block diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 2.1 Three principle photon interactions. (source: Knoll) . . . . . . . . . . . . . . . . . . . . . 5 2.2 Compton scattering kinematics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 2.3 Avalanche diagram for SSPM. (source Kasap) . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 2.4 Representative structural diagram for Si SSPMs. (source Photonique) . . . 15 2.5 Optical absorption coefficients for different photodetector materials.(source: Sze) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 2.6 SSPM pixel array and signal shape, bias circuit, SPICE model, and Vov definition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 (a) SSPM array structure and signal shape. . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 (b) SSPM bias circuit, SPICE model, and Vov . . . . . . . . . . . . . . . . . . . . . . . . . 19 2.7 SSPM Dynamic range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 2.8 Photodetection process in SSPM. (source: Sze). . . . . . . . . . . . . . . . . . . . . . . . . . 27 2.9 Optical fiber waveguide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 2.10 Index and light ray profiles for step index optical fiber. . . . . . . . . . . . . . . . . . . 31 2.11 Snell’s Law of reflection and refraction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 2.12 Critical angle for fiber. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 2.13 TIR for fiber when incident angle exceeds the critical angle. . . . . . . . . . . . . . 34 2.14 Important angle relationships for an OF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 2.15 Illustration of fiber coupling issues: light source coupled to fiber. . . . . . . . . 37 2.16 Illustration of metalized air-core glass capillary tube. . . . . . . . . . . . . . . . . . . . . 39 (a) Cross section of metalized air-core capillary tube. (Source: Polymicro Inc.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 LIST OF FIGURES (Continued) Figure (b) Page Interfaces between metalized air-core glass capillary tube and scintillator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 3.1 Block diagram of laboratory and clinical measurement system. . . . . . . . . . . 42 3.2 Photon counter operation. (source: Hamamatsu) . . . . . . . . . . . . . . . . . . . . . . . . 44 3.3 Dark counts showing p.e. levels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 3.4 Typical output from photon counter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 3.5 Circuit diagram for battery operated SSPM power supply. . . . . . . . . . . . . . . . 45 3.6 Circuit diagram for photon counter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 3.7 Power supply circuit diagram for photon counter. . . . . . . . . . . . . . . . . . . . . . . . 48 3.8 ADC output charge frequency distribution. (source SensL) . . . . . . . . . . . . . . 51 3.9 SSPM DCR measurement setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 3.10 SSPM PDE measurement setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 3.11 Voxtel SQBF-EKAA mounted on 3-stage Peltier-cooler showing recessed die. Can diameter is approximately 0.5 inch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 3.12 GRIN lens system showing important dimensions. . . . . . . . . . . . . . . . . . . . . . . . 54 3.13 GRIN lens design parameters for the Voxtel SQBF-EKAA. . . . . . . . . . . . . . . 57 (a) SLW-1.8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 (b) SLW-1.0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 3.14 BCF-12, BC430 scintillator emission spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . 61 3.15 Light collection efficiencies for components in the optical signal chain. . . . 65 3.16 Cerenkov threshold energy vs. material refractive index.. . . . . . . . . . . . . . . . . 67 3.17 Predicted Cerenkov angle vs. incident energy. . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 3.18 Number of Cerenkov photons produced in visible light spectrum. . . . . . . . . 71 LIST OF FIGURES (Continued) Figure Page (a) 1 mm material depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 (b) 1 cm material depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 3.19 Cone of Cerenkov Radiation: electron beam ⊥ to fiber. . . . . . . . . . . . . . . . . . . 72 3.20 Cerenkov cone approaching θcritical for the optical fiber. . . . . . . . . . . . . . . . . . 72 3.21 Schematic of BCF-12 and SiO2 : Cu2+ dosimeter cables. . . . . . . . . . . . . . . . . . 73 3.22 Eska GH4001 POF and BCF-12 scintillator dosimeter cable. . . . . . . . . . . . . 74 3.23 BCF-12 Air-Core Capillary Tube Dosimeter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 3.24 Photo of electron beam angular measurement setup. . . . . . . . . . . . . . . . . . . . . 76 3.25 Photo of dose linearity measurement setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 3.26 Photo of electron beam percent depth dose measurement setup. . . . . . . . . . 80 4.1 SSPM signal shape capacitance SPICE simulation. . . . . . . . . . . . . . . . . . . . . . . 81 4.2 Die capacitance for five Photonique SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 4.3 Dark count rates for Photonique SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84 4.4 4.5 (a) Measured DCR for Photonique SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . . . . 84 (b) Calculated DCR for Photonique SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . . . 84 Gain dependence for two Photonique SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 (a) Gain dependence for 050701GR.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 (b) Gain dependence for 0611B1 and CLNS. . . . . . . . . . . . . . . . . . . . . . . . . . . 85 Gain dependence for two Photonique SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . . . . 86 (a) Gain dependence for 0701BG. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86 (b) Gain dependence for 0810G1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86 4.6 IV-characteristics for MPPC, Photonique, and Voxtel SSPMS. . . . . . . . . . . 91 4.7 Die capacitance for MPPC, Photonique, and Voxtel SSPMs. . . . . . . . . . . . . . 92 LIST OF FIGURES (Continued) Figure 4.8 4.9 Page Dark count rates for Photonique, Voxtel and MPPC SSPMs. . . . . . . . . . . . . 93 (a) Measured DCR for (4) compared SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . . 93 (b) Calculated DCRs for (4) compared SSPMs.. . . . . . . . . . . . . . . . . . . . . . . . 93 Gain dependence for Voxtel and MPPC SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . 94 (a) Gain dependence for Voxtel SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 (b) Gain dependence for MPPC SSPM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 4.10 Optical fiber and plastic scintillator junction ends. . . . . . . . . . . . . . . . . . . . . . . 101 (a) Optical fiber junctions resulting in signal loss. . . . . . . . . . . . . . . . . . . . . . (b) BCF-12 scintillator tip under 10x magnification: tapered end is 485.5 µm diameter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101 101 4.11 GH4001 POF cable background response. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103 (a) GH4001 POF photon background response. . . . . . . . . . . . . . . . . . . . . . . . 103 (b) GH4001 POF electron background response.. . . . . . . . . . . . . . . . . . . . . . . 103 4.12 Angular dependence of standard and capillary tube POF dosimeters: BCF-12 scintillator: measured and theoretical results . . . . . . . . . . . . . . . . . . . 104 4.13 6 MV photon dose linearity: Prelude 420 scintillator. . . . . . . . . . . . . . . . . . . . . 105 4.14 Photon dose linearity: standard and capillary tube dosimeters . . . . . . . . . . . 106 (a) BCF-12 and SiO2 : Cu2+ Dosimeters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 (b) BCF-12 Capillary Tube Dosimeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 4.15 Electron dose linearity: BCF-12 and SiO2 : Cu2+ Dosimeters. . . . . . . . . . . . . 107 4.16 Electron dose linearity: capillary tube dosimeters. . . . . . . . . . . . . . . . . . . . . . . . 108 (a) BCF-12 Capillary Tube Dosimeter: 6 MeV, 9 MeV. . . . . . . . . . . . . . . . 108 (b) BCF-12 Capillary Tube Dosimeter: 12 MeV, 16 MeV, 20 MeV. . . . . 108 4.17 10 cm x 10 cm 6 MV photon depth dose profile in water tank . . . . . . . . . . . 110 LIST OF FIGURES (Continued) Figure Page 4.18 10 cm x 10 cm 18 MV photon depth dose profile in water tank . . . . . . . . . . 111 4.19 10 cm x 10 cm 9 MeV electron depth dose profile in water tank . . . . . . . . . 112 5.1 0.2 Proposed light guide for small area SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121 (a) Light guide schematic drawing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121 (b) Light guide ray trace diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121 Circuit diagram for TEC cooler controller. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134 LIST OF TABLES Table Page 2.1 w for some common radiation detection materials. . . . . . . . . . . . . . . . . . . . . . . 12 2.2 G, PDE, and Bias variations as a function of temperature. . . . . . . . . . . . . . . 26 2.3 Optical photon sensing: vacuum vs. solid-state technology. . . . . . . . . . . . . . . 30 2.4 The three types of optical fiber. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 3.1 c SLW GRIN lenses at λ = 440 nm and Calculated values for SELFOC 580 nm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 3.2 Characteristics of GH4001 cable and SiO2 optical fiber. . . . . . . . . . . . . . . . . . 59 3.3 Summary of published scintillator emission characteristics. . . . . . . . . . . . . . . 62 3.4 Estimated Scintillator Outputs: number of photons and SSPM current I. 64 3.5 v for a few energies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 4.1 Summary of published SSPM parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 4.2 PDE measurements for Photonique SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 4.3 Summary of measured breakdown voltages and device currents for Photonique SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 4.4 Summary of measured Photonique SSPM characteristics. . . . . . . . . . . . . . . . . 89 4.5 Photonique laboratory photon count rates†(cpm) 60 Co: direct scintillatorSSPM attachment and using POF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 4.6 PDE measurements for PH0810G1, Voxtel, and MPPC SSPMs. . . . . . . . . . 95 4.7 Summary of measured breakdown voltages and device currents. . . . . . . . . . 95 4.8 Summary of measured characteristics for Photonique, Hamamatsu, and Voxtel SSPMs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 Laboratory photon count rates†(cpm) 60 Co: direct scintillator-SSPM attachment and using POF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 4.9 4.10 Photon and Electron Dose Linearity Differences (range extrema) from Reference Ion Chamber (in %). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109 LIST OF TABLES (Continued) Table Page 4.11 Photon and Electron Percent Depth Dose Differences (range extrema) from Reference Ion Chamber (in %). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113 0.1 GH4001 plastic fiber cable attenuation characteristics.. . . . . . . . . . . . . . . . . . . 136 DEDICATION This thesis is dedicated to the memory of my two fathers: Morris Nickoliavich Konnoff and Leo Miles Clausen. Together you both made me the man I am today. This thesis is also dedicated to my wife and best friend who endured years of separation while I was thousands of kilometers away at OSU. SSPM-BASED OPTICAL FIBER RADIATION DOSIMETER 1. 1.1. INTRODUCTION AND PROJECT OBJECTIVES Introduction Optical fiber (OF) dosimetry using plastic or glass fibers coupled to a wide variety of scintillators has been reported for clinical, diagnostic radiography, and environmental radiation sensing purposes.[Akselrod et al., 2007, Anderson et al., 2009, Beddar, 1994, Beddar et al., 2001, Hyer et al., 2009] Copper-doped glass (SiO2 : Cu2+ ) may be used as an Optically Stimulated Luminescence (OSL) material or as a real-time Radio Luminescence (RL) scintillator [Huston et al., 2001, Justus et al., 2004] while BCF-12 plastic scintillator may be used in RL mode (photomultiplier tube as photosensor).[Beierholm et al., 2008] Water equivalence and energy independence of plastic scintillators as a detector material are well documented, and considerations of detector size and signal-to-noise (S/N) ratio have been analyzed.[Beddar et al., 2005, Clift et al., 2000, Archambault et al., 2005] BCF-12 plastic scintillator represents a good choice for a dosimeter as variants of this material have produced good results in clinical studies.[Bartesaghi et al., 2007] Recent work by groups in Japan and Russia have advanced solid state photomultiplier (SSPMs, SiPMs, MPPCs) technology [Golovin and Saveliev, 2004, Gomi et al., 2007] making it a viable alternative to photomultiplier tubes (PMTs) for laboratory, environ- 2 mental, clinical dosimetry, medical imaging applications (PET, CT, SPECT)[Heckathorne et al., 2006], dosimetry in diagnostic radiology, and in-vivo applications (catheters and brachytherapy[Anderson et al., 2009, Suchowerska et al., 2007]). Their small size, high gain, low bias voltage, and non-magnetic characteristics are distinct advantages when compared with the PMT.[Saveliev and Golovin, 2000] Previous clinical studies using optical fibers and different inorganic/organic scintillators have used standard photodiodes or PMTs as optical photon detectors. The large signals generated by Linac (linear accelerator) beams coupled with external amplification proved sufficient for analyzable signals.[Beddar, 1994, Lee et al., 2006] However the added advantages of built-in gain, low voltage operation, greater photon detection efficiency (PDE) and ruggedness, together with low noise, make SSPMs more suitable photon detectors than standard photodiodes in a wider variety of applications. 1.2. Goals of this project This project has two goals. The first is to investigate the design and characterization of SSPM-optical fiber-coupled systems (both OSL and RL) as radiation sensors for laboratory, environmental, diagnostic radiology, clinical radiotherapy, and military applications. The second is to characterize and compare the performance of different sampled SSPMs from Hamamatsu, Photonique, and Voxtel when coupled with different scintillator material types and scintillator coupled optical fibers. Note that SensL, currently the other major SSPM manufacturer, refused to supply samples for use in this work.[SensL, 2006] Figure 1.1 shows a block diagram of the detector system. current voltage 3 Ionizing radiation: x−ray, gamma ray, beta particle t Scintillator SSPM t Interface and Readout Electronics Host Computer FIGURE 1.1: Detector system block diagram. 1.3. Applications SSPM-based radiation sensing systems can solve a number of problems that plague PMT-based systems, among them: • 800-1200 V PMT bias voltages and glass envelopes are hazardous and fragile. • High sensitivity to vibration and magnetically unstable. • Large size, not easily portable. • More expensive compared to SSPM-based systems. Advantages of SSPMs are: • Extremely small size, compact. • Low bias voltage, battery operable (< 80 V). • Magnetically insensitive. • High PDE. • High gain. • Fast time resolution. • Not damaged when saturated by ambient light. 4 • Low cost potential. However there are disadvantages, among them: • High room temperature noise rate ( ≥ 100 kHz/mm2 ) for 1 mm2 active areas. • PDE is smaller than QE. • PDE is a function of overvoltage Vov and sensitive to small changes. • Gain and noise are dependent on temperature. • Limited dynamic range. • Optical cross talk issue. • High initial cost currently. This work explores these issues. Many of these disadvantages can be corrected by: using a temperature and voltage stabilized SSPM bias source, and ongoing technology improvements. 5 2. 2.1. OVERVIEW Overview of Photon Interactions There are six primary radiation interactions with matter: photoelectric effect, compton scattering, pair production, Raleigh scattering, photodisintegration, and triplet production. Of these six, photodisintegration and triplet production have low probabilities for the energy range typically encountered in radiation monitoring or oncology treatments (0 to 20 MeV) while Rayleigh scattering is a coherent, elastic process which does not contribute more than a few percent to dose.[Podgorsak, 2005, Martin, 2006] Figure 2.1 shows the three most important photon interaction mechanisms: the photoelectric effect, compton scattering, and pair production.[Turner, 2007, Attix, 1986, Tsoulfanidis, 1995, Knoll, 2000] FIGURE 2.1: Three principle photon interactions. (source: Knoll) 1. The photoelectric effect refers to the interaction of an incident photon with a tightly 6 bound atomic electron. The incident energy of the photon Eγ is absorbed by the bound electron which is then ejected from the atom leaving a vacancy in one of the inner orbitals. For a photon incident on the atom with energy hv, the kinetic energy (KEγ ) of the ejected photoelectron is: KEγ = hv − Ebinding (2.1) where Ebinding is the binding energy of the ejected electron in the atom. The probability of a photoelectric interaction (a function of τ the photoelectric cross section) varies approximately as Z 4−5 /Eγ3.5 (Z is atomic number) at low photon energies and with the atomic number of the target.[Knoll, 2000, Turner, 2007] 2. Scattering by a loosely bound or free electron is known as Compton scattering or the Compton effect; it is the predominant mode of photon interaction especially for low Z absorbing material (see Figure 2.1). Here the incident photon is scattered by an electron with some of the energy transferred to the same electron (know as the recoil electron). Figure 2.2 shows the kinematics of this interaction. recoil electron Ee Eγ φ θ incoming γ −ray E’γ scattered γ−ray FIGURE 2.2: Compton scattering kinematics. The recoil kinetic energy of the Compton electron is: 7 ′ Ekinetic = Ehv − Escattered = Eγ − Eγ ′ (2.2) ′ where (Escattered = Eγ = Ehv ) is the energy of the scattered photon. This energy depends upon the incident photon energy and the scattering angle which is given by ′ ′ Eγ = Ehv = 1+ Eγ Eγ (1 me c2 (2.3) − cosθ) ′ ′ where Eγ = Ehv is the energy of the incoming gamma ray photon, Eγ = Ehv is the energy of the scattered x-ray photon, and me c2 is the rest mass energy of an ′ electron (511 keV). Note that Eγ = Eγ at θ = 0 (scattered photon maximum energy) ′ and at θ = π (scattered photon minimum energy) Eγ = Eγ /1 + 2Eγ 511keV . Compton scattering probability (σ = the Klein-Nishina cross section) generally decreases with increasing incident photon energy, varying approximately as Z/hv and is directly proportional to the number of electrons per gram of material. [Knoll, 2000, Turner, 2007, Tsoulfanidis, 1995, Attix, 1986] 3. When incident photon energy is greater than 1.022 MeV pair production predominates. The photon interacts with the nucleus electromagnetic field, is absorbed, and is replaced by an electron-positron pair. Rest mass energies for both the electron and positron are me c2 (511 keV); thus the minimum incoming photon threshold energy needed is 1.022 MeV (2me c2 ). Excess photon energy is transferred to the electron and positron as kinetic energy: Epp−transf erred = Ehv − 2me c2 (2.4) Transferred KE is equally shared between the electron and positron. Probability of pair production (κ = the pair production cross section) varies approximately as Z 2 ln(Ehv − 2me c2 ).[Knoll, 2000, Turner, 2007, Tsoulfanidis, 1995, Attix, 1986] 8 2.2. Overview of Electron Interactions The primary photon interactions all transfer energy to electrons in the material in the form of KE, setting electrons in motion. Propagation of these electrons causes the downstream deposition of radiation effects or dose. The process of energy transfer between the electrons set in motion in the material can be described four ways:[Turner, 2007, Tsoulfanidis, 1995] 1. Coulumbic interactions with orbital electrons and other nuclei, 2. Emission of electromagnetic radiation (Bremsstrahlung radiation), 3. Nuclear interactions, 4. Emission of Cerenkov radiation. Energy transfer is primarily through Coulombic interactions between the electric field of moving electrons and the electric field of either electrons or atomic nuclei in the material. If the particle is an electron or positron, it may collide with an atomic electron and lose all of its energy in a single collision because the collision involves particles of the same mass, or it may be scattered in a zig-zag pattern with large angles.[Tsoulfanidis, 1995] Contrast this with heavier charged particles (alpha, proton, deuteron, etc.) that lose much smaller amounts of energy with each collision with atomic electrons and experience little deflection. Interactions involving electrons in the material result in ionizations along the initial electron’s trajectory and are called ionizational losses. The rate of energy loss per unit length of material as a result of these interactions is characterized by the linear stopping power dE/dx, where E is the kinetic energy of the initial energetic electron and x is the length traveled by the particle. The total mass ionization stopping power S(E) = − 1ρ (dE/dx) for electrons is given by dividing dE/dx by ρ, the density of the material 9 (minus sign makes S positive). S(E) is separated into two components: the collisional stopping power and the radiative stopping power shown in equation 2.5.[Turner, 2007] 1 S=− ρ dE dx total 1 =− ρ dE dx 1 + − ρ col dE dx [ rad M eV cm2 ] g (2.5) Collisional Stopping Power Collisional stopping power for electrons/positrons is given by equation 2.6. [Turner, 2007, Attix, 1986] 1 − ρ dE dx ± col √ 4πk02 e4 n mc2 τ τ + 2 ± √ = + F (β) ln mc2 β 2 2I (2.6) gives the energy loss per unit thickness of material where: k0 = 8.99 × 109 [N m2 /C 2 ], e is the electronic charge = 1.6 · 10−19 C, n is the number of electrons per unit volume in the material, m is the mass of an electron, β is the electron relativistic speed v/c, τ = E/me c2 is the particle energy divided by electron rest mass energy (511 keV), I is the mean atomic ionization potential [eV] (found in tables, varies with Z)[Turner, 2007, Podgorsak, 2006], F ± is defined differently for electrons and positrons. For electrons F − (β) is: 1 − β2 τ2 F β) = − (2τ + 1)ln2 1+ 2 8 − (2.7) and for positrons F + (β) is: 10 4 14 β2 + + . 23 + F (β) = ln2 − 24 τ + 2 (τ + 2)2 (τ + 2)3 + (2.8) 10 Substituting in the constants and rearranging equation 2.6 gives, − 1 ρ dE dx ± = col with G± (β) given by 5.08 × 10−31 n ± G (β) − ln I β2 √ G± (β) = ln 3.6 × 105 τ τ + 2 + F ± (β). (2.9) (2.10) Radiative Stopping Power Acceleration interactions involving atomic nuclei in a material result in radiative loss of energy through bremsstrahlung processes (when an electron/positron is deflected by a nuclear electric field) that in turn result in the emission of photons. Bremsstrahlung production varies approximately with Z 2 ; losses are greater in high-Z materials. The mass radiative stopping power for electrons is given by equation 2.11. [Turner, 2007, Attix, 1986] 1 − ρ dE dx rad ZE 1 × ≈ 800 ρ dE dx (2.11) col where the individual terms are the same as above. [Podgorsak, 2006] gives complete expressions for radiation stopping power for various ranges of electron kinetic energies. 2.3. Cerenkov Radiation Another process through which energetic electrons can lose energy is through the emission of Cerenkov radiation. Cerenkov emission occurs when a charged particle passes through any medium in which the phase velocity of light is less than the particle velocity (i.e. βn > 1, n being the refractive index of the medium). These conditions occur when high-speed charged particles pass into a transparent dielectric material. While the velocity of the particle is unaltered, the electric field associated with the particle’s charge 11 and the magnetic field associated with the motion of this charge are propagated with a phase-velocity of c/n. As the particle moves ahead from a slower-moving portion of its own electromagnetic field, an electromagnetic wave front is formed. The number of quanta emitted is inversely proportional to the square of the wavelength, thus shorter wavelengths are favored; Cerenkov radiation appears as bluish-white light. The threshold particle energy required to generate Cerenkov radiation is given by equation 2.12. [Jelly, 1958] ECerenkov−threshold = mo c2 r ! 1 −1 1+ 2 n −1 (2.12) Photons are emitted anisotropically with the direction of emission characterized by a cone spreading out at the Cerenkov angle from the direction of the interacting particle, given by equation 2.13. [Jelly, 1958] κ = cos−1 ( 1 ) = Cerenkov angle nβ (2.13) where β is is the ratio of velocity in the material to the speed of light in a vacuum. This is an important noise source in several types of specialized radiation sensing devices which use optical fibers for light signal transmission. The angle at which this light is predicted to be a maximum is given by[Jelly, 1958] θCerenkov−M ax = cos−1 1 (2.14) ncore where ncore is the refractive index of the optical fiber core.[Jelly, 1958] The approximate theoretical intensity curve for the captured Cerenkov radiation in optical fiber is given by Icapture ncore 2 − 1 ≈ρ cos−1 πncore 2 − 1 3 ncore − ∆n − cos γ √ sin γ ncore 2 − 1 (2.15) where ρ is the OF core radius, ∆n is the difference between the OF core and cladding indices of refraction and γ is the angle between the electron beam and the OF axis in the 12 direction of the detector.[Law et al., 2007] 2.4. Specific Ionization An important concept when working with different materials is the number of ionpairs formed per unit distance traveled in a material. The specific ionization is given by equation 2.16. [Tsoulfanidis, 1995] SI = dE/dx [eV /cm] w [eV /ip] (2.16) where w is the energy needed to create an electron-hole (e-h) pair or an ion, which is specific to a material. w for some common radiation detecting materials is shown in Table 2.1.[Turner, 2007, Attix, 1986, Martin, 2006] TABLE 2.1: w for some common radiation detection materials. Material Air Plastic SiO2 Si Ge GaAs (w) e-h pair 34 eV 60 eV 17 eV 3.6 eV 2.8 eV 4.8 eV generation energy 2.5. Solid State Photomultipliers (SSPMs) as Photomultiplier Tube (PMT) Replacements Various commercial manufacturers have recently released SSPM devices that can replace PMTs in selected applications. [Gomi et al., 2007, Swain et al., 2005, Jackson, 2007, Zecotek, 2008] These devices are small, low voltage, non-magnetic (important for active MRI and military uses), rugged and relatively low cost (PMTs of comparable or greater performance are currently more expensive). Their principle drawbacks at present 13 are higher dark current (2-3 or more orders of magnitude greater than low noise PMTs) and, until recently, small active areas. Each potential application must be evaluated individually, as one cannot simply drop SSPMs in as a PMT replacement. SSPMs are similar to PIN diodes in function having a more complex structure (an array of diodes connected in parallel), with the main exception being built-in gain from operating in the geiger-mode avalanche breakdown region of the I-V curve (also known as geiger breakdown).[Sadygov et al., 1996, Saveliev and Golovin, 2000, Golovin and Saveliev, 2004, Sadygov et al., 2006] Avalanche breakdown is caused by impact ionization where an electron or a hole gains sufficient energy from the applied electric field such that the energy gained initiates a transition of an electron from the valance band to the conduction band, thus creating a new electron-hole (e-h) pair.[Sze, 2006] Newly created e-h pairs are also accelerated by the applied electric field, generating further new e-h pairs by process repetition. If the bias voltage Vbias is high (hence a high E-field) an uncontrolled rise in the current is the result. Left unchecked this process leads either to device destruction or current limiting by an external load resistor, Rquenching , for individual SSPM pixels. Figure 2.3 shows the e-h avalanche diagram for an SSPM. FIGURE 2.3: Avalanche diagram for SSPM. (source Kasap) SSPM multiplication gain is approximated by equation 2.17. [Sze, 2006] 14 M= 1 1 = IM /Ip = RL 1 − α(x)Wd 1 − 0 α(x)dx (2.17) where α is the electron or hole multiplication coefficient, L is the electron space charge region width, IM is the average value of the total multiplied output current, Ip is the primary unmultiplied photocurrent, and Wd is the depletion region width. This is true for equal ionization coefficients α = αn = αp .[Sze, 2006]Note that αWd = 1 corresponds to the device breakdown voltage Vbreakdown . α(x)Wd is the number of ionization’s in width Wd ; this is a function of electric field E and thus on reverse bias voltage Vbias , m with C a constant. Here m is an empirical exponent; approximated as α(x)Wd = CVbias a function of the material and its doping levels. Using the above C can be expressed as −m C = Vbreakdown .[Bar-Lev, 1993, Sze, 2006] Substituting into equation 2.17 gives the common form for M M= 1 1 − (Vbias /Vbreakdown )−m (2.18) Guard ring structures are used to fabricate SSPM devices to minimize leakage currents caused by regions of high electric field at the junction edges. These areas of local high E fields, known as micro-plasmas, have low breakdown voltages and the potential for uncontrolled avalanches.[Pellion et al., 2009, Golovin and Saveliev, 2004, Renker and Lorenz, 2009, McNally and Golovin, 2009] Figure 2.4 shows a representative structural diagram for a Si SSPM showing trench based optical isolation and E-field intensity by region. 2.5.1 Avalanche Noise If m is the statistically varying SSPM gain, then < m2 > > < m >2 = M 2 where the (<>) represent an ensemble average, and < m >= M is the average carrier multiplication as defined in equation 2.17.[McIntyre, 1966] This noise results from the avalanche process itself and depends upon < m2 >, the mean square gain, Thus SSPM avalanche noise can 15 FIGURE 2.4: Representative structural diagram for Si SSPMs. (source Photonique) be high.[Fyath and O’Reilly, 1988] It has been found empirically that < m2 > can be approximated as: < m2 >= M 2+x (2.19) where x varies between 0 and 1 and is a function of SSPM crystal material (Si, InGaAs, etc.) and device structure.[Scansen and Kasap, 1992, Kasap, 2001] The excess noise factor, F(M), is a measure of the increase in SSPM noise resulting from the randomness of the multiplication process.[Sze, 2006] F is the ratio of the actual noise generated in an SSPM to the noise that would exist if all carrier pairs were multiplied by M. It is given by F = <m2 > <m>2 = <m2 > .[Kim M2 et al., 1997] From equation 2.19, F(M) is approximated as: F = Mx (2.20) where x = 0.3 for Si and x = 0.7 for InGaAs.[Scansen and Kasap, 1992, Kasap, 2001] 16 2.5.2 Spectral Response and Quantum Efficiency Figure 2.5 shows optical absorption coefficients for different optical materials. From this figure one can see that Si and modified Ga compounds are suitable materials for SSPMs. FIGURE 2.5: Optical absorption coefficients for different photodetector materials.(source: Sze) The value of the long wavelength cutoff frequency λc is given by equation 2.21: hc Eg (2.21) 1.241 [µm] Eg (eV ) (2.22) λc = which is commonly expressed as: λc = Room temperature bandgap energies are 1.124 eV for Si and 1.42 eV for GaAs; the corresponding λc cutoffs for each material are 1.1 µm and 0.87 µm, respectively. 17 Quantum efficiency η(λ) for a SSPM is the number of e-h pairs generated per incident photon, given by equation 2.23: η(λ) = Iphoto q P hλ = number of emitted electrons number of incident photons (2.23) where Iphoto is the photocurrent generated by incident optical power P. Overall responsivity is defined by equation 2.24: R= Iphoto η(λ)q η(λ)λ = M = Ro M = M [A/W ] P hν 1.241 (2.24) (with λ in µm), as the ratio of the photocurrent Iphoto to the incident optical power P, and Ro is the unity gain responsivity.[Kasap, 2001, Palais, 2005] In order to have high η(λ), an SSPM must have enough absorption layer width, which is determined by the light absorption coefficients as shown in Figure 2.5. Most of the incident light on an SSPM will be absorbed when the depletion width Wd is of the same order as 1/α. For semiconductor homojunctions, the depletion region must extend from the material surface to avoid surface absorption. For semiconductor heterojunctions such as modified Ga compounds, η(λ) does not depend on the junction-to-surface distance. The larger bandgap material is used as a window for light transmission. Indeed, this is how wide spectral range optical photon harvesting detectors are constructed. Antireflection coatings must be used to minimize light reflection at the semiconductor-air interface.[McNally and Golovin, 2009, Barton et al., 2009] Overall SSPM efficiency, sometimes referred to as Photon Detection Efficiency (PDE) is given by equation 2.25, ǫ = η(λ) × ǫGeiger × ǫGeometry = PDE (2.25) Here η(λ) is the quantum efficiency as defined in equation 2.23, ǫGeiger is the probability of an avalanche breakdown, and ǫGeometry is the SSPM photosensitive area packing 18 factor.[Buzhan et al., 2006] SSPM pixel capacitance at the breakdown voltage Vbreakdown is used to calculate gain using[Buzhan et al., 2006] Gain = R i(t)dt = Qpixel /q = Vov Cpixel /q q (2.26) where gain is proportional to the overvoltage, Vov = Vbias − Vbreakdown above the SSPM breakdown voltage. SSPM manufacturers typically recommend an operating voltage at a specific gain, known as Vop . For our laboratory and clinical measurements PDE at the peak scintillator wavelength was the best indicator of overall SSPM system performance, followed by SSPMscintillator/SSPM-OF coupling considerations, SSPM dark count rate (DCR), then by SSPM gain. For low level environmental radiation sensing or when used together with OFs, SSPM DCR must remain small to detect signals attenuated by long POFs (10s of meters), where a high system S/N ratio is necessary for detection of low level scintillator signals.[Pavlov et al., 2005, Beddar, 2007] 2.5.3 SSPM Biasing, Equivalent Circuit, and Speed of Response Figure 2.6 shows the SSPM pixel array structure, gross output signal shape, circuit model, SPICE model for a typical SSPM, and the definition of overvoltage Vov on the I-V curve. Figure 2.6(a) illustrates the diode pixel elements connected in an array with a common output together with the gross signal shape from those same elements. The overall magnitude of the SSPM output signal is the sum of each diode pixel. The SSPM spice model is divided into three zones: an Active zone which represents a diode pixel about to avalanche, during avalanche, or recovering from avalanche, a Passive zone that represents the remaining non-firing diode pixels, and a Parasitic zone which represents other stray capacitance’s on the SPPM die.[Corsi et al., 2006, Seifert et al., 2009] External circuit resistance is the sum of Rs + RL , the shunt resistance of any subsequent amplifier stage, 19 (a) SSPM array structure and signal shape. +Vbias Bias Circuit RL A hv SSPM C shape R shunt Definition of Vov reverse I Spice Model Active Passive Parasitic Vbreakdown Vbias Rquench Cquench R ______ quench (N−1) forward Cparasitic hv Ipulse V Cquench (N−1) Cdiode Cdiode (N−1) Overvoltage(Vov) (b) SSPM bias circuit, SPICE model, and Vov . FIGURE 2.6: SSPM pixel array and signal shape, bias circuit, SPICE model, and Vov definition. 20 and the load resistance, while external circuit capacitance is the sum of CD + Cshape , the pixel junction capacitance and other parasitic capacitance’s. For a typical SSPM the CD and RL terms dominate. Note that CD is a strong function of pixel area and reverse bias voltage. Each diode in the pixel array is in series with a quenching resistor, Rquench , that is typically 1 MΩ or more. Reducing the bias on each pixel element below Vbreakdown stops the impact ionization process. Approximately, the RC time constant Rquench Cdiode determines the time for this to occur. Rquench limits the current recharging the SSPM diode pixel. A loop equation for each pixel element gives: Vbias = Vdiode + IRquench + IRL (2.27) where the resulting pixel element signal (ideal) is given by the following steps. • The diode pixel is in the dark and no current flows (I = 0), and Vdiode = Vbias . Assume a photon impinges on the pixel at some time t0 . • Diode pixel current increases. The diode pixel is charged with τr = Cdiode Rquench time constant. The diode pixel bias will now decrease until it reaches breakdown voltage. • At some time t1 , Vdiode = Vbreakdown , and uncontrolled geiger mode avalanche multiplication is stopped (quenching). • At times > t1 , the diode pixel capacitance discharges with a time constant τf = Cdiode (Rquench + Rs ). • After a finite diode pixel recovery time between approximately 4τf to 9τf , current is reduced to zero and the diode pixel element is reset. This pixel recovery time is strongly dependent on SSPM manufacturer and technology used. 21 The rise time and fall time of an SSPM are defined as the times for the signal to rise or fall from 10% to 90% or 90% to 10% of the final value, respectively. This parameter can be also expressed as frequency response, which is the frequency at which the SSPM output decreases by 3 dB, and is given by equation 2.28:[Kasap, 2001, Palais, 2005] trise = 0.35 f3db (2.28) For an SSPM there are four principle factors which determine the speed of response: • The RC quenching time for each individual pixel. This is the product of the individual pixel capacitance and on-chip quenching resistor. • The RC time constant (tRC = 2.2RC) of the pixel (diode) array and external circuit. • The diffusion (charge) collection time of carriers in the undepleted region. • The drift charge collection time of carriers in the depletion region of each pixel. 2.5.4 Dynamic Range The total number of pixels firing in an SSPM determines the dynamic range. The re- lationship between the number of pixels fired (Nf ired ) and the number of incident photons (Nphotons ) is given by equation 2.29; Nf ired = Ntotal 1 − exp −Nphotons Ntotal (2.29) Ntotal is the total number of SSPM pixels in the array.[Renker, 2006] The signal output is proportional to the number of fired cells. When Nphotons reaches the same order of magnitude as Ntotal , the probability that multiple photons hit the same pixel increases.[Renker, 2006] As a result the output signal will saturate independent of the number of photoelectrons. An SSPM signal can saturate at high light levels without any resulting physical 22 damage. Contrast with a PMT that suffers physical damage at high or ambient light levels. Equation 2.29 is plotted in Figure 2.7, showing Nf ired vs. Nphotons for representative Hamamatsu, Photonique, SensL, and Voxtel SSPMs. SSPM Dynamic Range 10000 100 pixels 516 pixels 556 pixels Number of Fired Pixels 400 pixels 1000 1024 pixels 1600 pixels 3900 pixels 8100 pixels 14560 pixels 100 10 1 1 10 100 1000 10000 Number of Photoelectrons FIGURE 2.7: SSPM Dynamic range. 2.5.5 Dark Count - Causes The three contributors to SSPM dark counts are thermal e-h carrier generation, afterpulses, and optical crosstalk. SSPMs, being made of semiconductor material, suffer from thermal generation of eh pairs in addition to those caused by incoming photons.[Quimby, 2006] These thermally generated e-h pairs typically lead to a dark count rate (DCR) from 105 − 107 counts s−1 /mm2 at room temperature with a threshold at half of one photo-electron amplitude (p.e) approximately 0.5 · 106 .[Quimby, 2006, Renker and Lorenz, 2009] The dark count 23 rate decreases with temperature and decreasing Vov . The density of defects in, and volume of the semiconductor material are the primary dependencies. A major noise parameter in an SSPM is the dark current. Even though no light is present some current will still flow between the terminals of the device. The magnitude of this current differs, dependent on device physical design (layout, isolation trenches, pixel size, pitch, etc.). Dark current arises from random electron-hole pair generated thermally or by tunneling. A dark current generated electron will have the same effects as a photo generated one, thereby multiplication. Tunneling is mostly a matter of design and choice of materials. It is therefore not a process the user can alter by some means of external stress, unless compromising vital properties like bias. However, the rate of thermally generated carriers can be influenced by material temperature. The rate of thermally generated carriers in an intrinsic material is given by the generation rate[Bar-Lev, 1993, Sze, 2006] G(T ) = κn2i (T ) [carrier pairs 1/m3 s] (2.30) where κ is a proportionality constant. Note that G(T ) ≈ n2i r(T ) where r(T) is the carrier recombination rate. At low temperatures (77 K) carrier mobilities are greater than at room temperature (300 K); as temperature is reduced Vbreakdown decreases.[Sze, 2006, Bar-Lev, 1993] The function G(T) has a temperature dependency, decreasing with less temperature and increasing with greater temperature. Decreasing the temperature decreases the dark current, thus cooling the semiconductor material increases low light photon detection ability. The dark current in an SSPM is approximated by I = Ie + Ih = q A n2i (T ) C(T ) −qVbias −qVbias exp − 1 = Idark exp −1 kT kT (2.31) where Ie and Ih are the electron and hole currents, A is the SSPM junction area, C(T) is a 24 constant including diffusion lengths, depletion region widths, and material doping levels. ni (T ) is given by ni (T ) = 2( 2πkT 3/2 ) (me mh )3/4 eEg /2kT h2 (2.32) where Eg is the material energy gap and me , mh are the electron/hole mobilities respectively. Thus, the thermal probability of carrier production can be approximated as Thermal Probability ≈ T 3/2 eEg /2kT (2.33) Note I = dQ/dt = nq/∆t where n is the number of electrons, q is the electronic charge, and ∆t is a time interval (typically ns).[Quimby, 2006] The number of electrons passing a cross-section of material in 1 ns is n= I∆t = 6.26 · 10−9 I q (2.34) From equation 2.34 one can see that less than one electron per ns is needed to realize a dark current in the sub-nanoampere range. Temperature Dependence SSPMs are similar to standard diodes with increased photo sensitivity, where the output current contains a term which is dependent on the incident light intensity on its surface in the operating wavelength range. SSPM output current is given by − qVop Isspm = Iphoto + Idark e kT −1 (2.35) Here Isspm is the SSPM output current, Iphoto is the photo current, Idark is the saturation dark current, and Vop is the operating voltage. 25 At a constant bias voltage and wavelength, the SSPM responsivity increases by decreasing the device temperature. Cooling an SSPM will increase its detected photocurrent. SSPM dark current, given by the second term of equation 2.35, is also dependent on the device operating temperature. In general, decreasing the SSPM temperature will decrease the dark current. SSPM signal amplitude temperature stability is a function of three (3) factors as shown in equation 2.36 Signal Amplitude(T ) = Nphotons × G(T ) × P DE(T ) (2.36) Note that gain G and PDE are strong functions of temperature, additionally both PDE and G are functions of Vbias . SSPM breakdown voltage Vbreakdown increases linearly with temperature[Bar-Lev, 1993]. The breakdown voltage can be approximated for pure avalanche processes as Vbreakdown = Vbreakdown0 (1 + β(T − T0 ) (2.37) where Vbreakdown0 is the reverse breakdown voltage at room temperature (T0 ) and β is the linear growth constant, typically greater than 10−3 K −1 .[Petasecca et al., 2008] Signal amplitude increases with Vbias when SSPM temperature is lowered because G and PDE are negatively correlated with temperature.[McNally and Golovin, 2009] Temperature variation dependencies of bias voltage, G, and PDE are given in Table 2.2 (table from [Shushakov et al., 2008]). A temperature controlled isolation chamber is needed to calculate these variations from measured data; none was available for use in this work. Table 2.2 is included to illustrate these variations are important parameters that must be included in SSPM system design. Afterpulses Afterpulsing occurs when carriers are trapped during one avalanche and undergo a delayed release, after which a new avalanche is triggered. Afterpulses with short delay 26 TABLE 2.2: G, PDE, and Bias variations as a function of temperature. Parameter Definition Unit Typical value Temperature coefficient of Vbias dVbias dT mV /◦ C 20 − 30 Temperature coefficient of G dG dT ·G %/◦ C 0.1 − 0.5 Temperature coefficient of PDE d(P DE) dT ·P DE %/◦ C 0.1 − 0.6 contribute little because the cells are not fully recharged but have an effect on the SSPM recovery time.[Renker and Lorenz, 2009, Vinogradov et al., 2009] Long delay afterpulses contribute to DCR. Optical Crosstalk Interpixel crosstalk occurs during avalanche breakdown between adjacent pixels or nearly adjacent ones. Once a pixel fires, its photons can trigger neighborhood pixels adding to the DCR or an artificial increase in signal amplitude. Approximately three photons per 105 carriers are emitted when the incoming photon energy is greater than 1.14 eV.[Lacaita et al., 1993] It is this optical crosstalk problem that prevented the practical development of SSPMs until the last decade of the 20th century. A typical SSPM physical topology uses trenches to isolate each pixel on an island. Filling the trenches in the pixel array with opaque optical material helps eliminate crosstalk. Crosstalk rate is a function of Vov . 2.5.6 SSPM S/N Ratio and Noise Considerations The power SNR at the output of an optical receiver is given by: S/N = photocurrent signal SSP M noise power + amplif ier noise power (2.38) To obtain the greatest S/N the SSPM must have high η(λ) and any subsequent amplification noise should be low. The S/N sensitivity of an SSPM is described in terms 27 of the minimum detectable optical power. This is the optical power necessary to produce a photocurrent of the same magnitude as the total rms noise current (or a SNR of 1).[Palais, 2005, Motchenbacker and Connelly, 1993, Sze, 2006] The photodetection process for a SSPM is shown in Figure 2.8. The SSPM gain multiplies three important currents: the signal current, the background current, and the dark current or dark current noise.[Sze, 2006] Noise currents have mean square rms valSSPM Amplifier Current Input Output Signal Optical signal Photo− electric effect Background Dark Background signal Input− gain− output circuit Avalanche gain Excess noise Signal + noise Thermal noise FIGURE 2.8: Photodetection process in SSPM. (source: Sze) ues donated by the < > symbol and statistical variance σ 2 .[Motchenbacker and Connelly, 1993] When a modulated optical power P(t) impinges on a SSPM, the primary photocurrent is i(t)photo = ηq hν P (t). This current has two components, a dc average photocurrent Ip and a signal power component ip (t). For a SSPM the mean square signal current is < i2s >= σs2 =< i2p (t) > M 2 . M here is the multiplication gain as discussed in Section 2.5.1. As stated above, the primary sources of noise in SSPMs are: quantum noise, dark current noise generated in the bulk semiconductor material without incident light, and multiplication noise.[Palais, 2005, Motchenbacker and Connelly, 1993, Sze, 2006, Kasap, 2001] Dark current is further divided into two components: bulk dark current < i2BDark > and surface dark current < i2SDark >. 1. Quantum noise also known as shot noise stems from the statistics governing production and collection of photoelectrons when light is incident upon a photodetector. 28 Shot noise obeys the axioms of Poisson processes. Shot noise is given as: 2 < i2quantum >= σquantum = 2qIp BW M 2 M x (2.39) where BW is the bandwidth and M x is F(M) as defined in equation 2.20.[Kasap, 2001, Sze, 2006, Moura and Darwazeh, 2005] Note that for PIN diodes, F(M) and M are equal to 1. 2. Bulk dark current noise originates from electrons/holes thermally generated in pn junctions. However in an SSPM any charge carriers are accelerated and multiplied by avalanche gain. The mean-square value of the bulk dark current from thermally generated electrons/holes is given as: < i2BDark >= σBDark = 2qIBD BW M 2 M x (2.40) where IBD is the unmultiplied SSPM bulk dark current.[Bielecki, 1997, Becker and Johnston, 2004, Sze, 2006] 3. Multiplication noise was discussed previously in Section 2.5.1. 4. Surface dark current noise also called surface leakage current more commonly simple leakage current, is a function of material surface defects and surface area. It is also a function of bias voltage. Guard ring structures are used in SSPMs to reduce this current. The mean square value of this current is given as: 2 < i2SDark >= σSDark = 2qISLeak BW (2.41) where ISLeak is the surface leakage current. Note the surface dark current is not multiplied by avalanche gain whereas bulk dark current is.[Bielecki, 1997, Becker and Johnston, 2004, Sze, 2006] All the dark currents and signal currents are uncorrelated, so the mean square SSPM 29 noise current is: 2 < i2sspm−noise > = σsspm−noise =< i2quantum > + < i2BDark > + < i2SDark > 2 2 2 2 + σSDark = σsspm−noise + σquantum + σBDark (2.42) = 2q(Ip + IBDark )M 2 M x BW + 2qISLeak BW The load resistor RL contributes a mean square thermal noise current, also known as Johnson noise current, which is given as: 2 = 4kT < i2thermal >= σthermal BW RL (2.43) Here, T is absolute temperature in K◦ and k is Boltzmann’s constant.[Palais, 2005, Kasap, 2001, Sze, 2006, Motchenbacker and Connelly, 1993, Moura and Darwazeh, 2005] Johnson noise can be reduced by using a load resistor which is large mindful of overall bandwidth requirements. A trade-off is involved. Substituting equations 2.42 and 2.43 into equation 2.38 gives:[Bielecki, 1997, Sze, 2006, Palais, 2005, Motchenbacker and Connelly, 1993] S/N = < i2p > M 2 2q(Ip + IBDark )M 2 M x BW + 2qISLeak + 4kT BW/RL (2.44) Here the thermal noise is of lesser importance and the SSPM noise typically dominates. Note that the signal power is multiplied by M 2 and the quantum noise plus bulk dark current is multiplied by M 2 M x (or F(M)). [Bielecki, 1997, Palais, 2005, Kasap, 2001, Sze, 2006, Motchenbacker and Connelly, 1993] One SSPM tested in this work (Voxtel) is mounted on a 3-stage thermo-electric cooler (TEC) allowing it to be operated at −20◦ C (or lower depending upon the cooling circuit setpoint). This can reduce the dark rate by an order of magnitude (or more) below its room temperature value.[Voxtel, 2008] 2.5.7 Comparison of SSPM vs. Vacuum Tube (PMT) Technology Table 2.3 summarizes the salient differences between vacuum tube and solid-state technology for optical photon detection circa 2009. 30 TABLE 2.3: Optical photon sensing: vacuum vs. solid-state technology. Device Characteristic PMT PIN APD SSPM† Gain (G) 106 − 107 1 ≈ 200 105 − 106 Dark Count (DCR) < 0.5 Mhz < 0.5 Mhz < 1 Mhz < 1 Mhz Sensitivity 1 p.e. 100s p.e. 10 p.e. 1 p.e. Speed 100s of ps 10s of ns < 10 ns 100s of ps Vbias 500 V - 1.2 kV 10 V - 100 V 100 V - 500 V 15 V - 100 V Battery Operation difficult Yes difficult Yes PDE (Red) ≤ 10% ≈ 90 − 100% ≈ 75% ≈ 40% PDE (Green) ≈ 40% ≈ 80 − 90% ≈ 60 − 70% ≈ 40 − 50% PDE (Blue) ≈ 20% ≈ 50% ≈ 40 − 50% ≈ 20 − 30% Temperature Sensitive Yes Yes Yes Yes Magnetic Sensitive Yes No No No Mechanical fragile small small small bulky rugged rugged rugged Rate †Approximate die size 1mm x 1mm. 2.6. Important Characteristics of Optical Fibers - Glass, Plastic Figure 2.9 shows an optical fiber waveguide (hereafter simply optical fibers or OFs). There are different indices of refraction for the core and cladding n1 , ncore or n2 , nclad , respectively, and ncore > nclad always. Refractive index may decrease abruptly or gradually from core to cladding depending upon how a fiber is manufactured. A gradual index change means having a diffuse material interface between core and cladding. This results in rays that do not change direction abruptly upon reflection by the material interface. 31 FIGURE 2.9: Optical fiber waveguide. An abrupt decrease in index corresponds to a stepped interface where a ray changes direction abruptly upon reflection by the core-cladding interface. Fibers with abrupt changes in index are called step index fibers, while fibers with a gradual change in index are called graded index fibers (GRIN). GRIN fibers give sharper output pulses with less pulse distortion in response to an input pulse, when compared with stepped index fibers. Figure 2.10 shows index and light ray profiles for multimode step index fiber used in this work. GRIN optical fibers exhibit a gradual refractive index shift from the core outward to the cladding as opposed to step index optical fibers where the index changes in a discrete step.[Palais, 2005, Kasap, 2001] An optical fiber has different core diameters. Small n2 n2 n2 n1 core θ n n1 cladding n1 no n2 2a FIGURE 2.10: Index and light ray profiles for step index optical fiber. cores (e.g., 5 µm) diameter mean that only rays parallel to the fiber axis travel completely through the fiber; off-axis rays need to be reflected many times during fiber travel and escape. Large core (e.g., 500 µm) diameters mean that both off-axis and on-axis rays travel through the fiber. Fibers with large cores are called multimode while fibers with 32 small cores are called single-mode. Multimode fibers have greater pulse distortion than single-mode fibers. Light intensity traveling through single-mode fibers is less than for multimode fibers. Single mode fibers are stepped index, while multimode fibers may be either stepped index or graded index. Table 2.4 summarizes the characteristics of the three types of optical fiber. TABLE 2.4: The three types of optical fiber. Fiber type single mode multimode multimode stepped index graded index stepped index Light travel on-axis (||) to fiber both on- and off-axis both on- and off-axis Pulse distortion lowest higher highest Size small < 5 µm large > 500 µm large > 500 µm 2.6.1 Light Propagation in OFs When an incident ray traveling in a material of refractive index n1 encounters the interface with a material of refractive index n2 , as shown in Figure 2.11, it is partly reflected and partly transmitted. The reflected ray is symmetric with the incident ray around the normal to the material interface, such that θ1 (angle between the incident ray and the normal) equals the angle between the reflected ray and the normal, which is the case for mirror reflection. The transmitted ray is not in the same direction as the incident ray, if n1 is not equal to n2 . This is known as refraction. The angle of refraction θ2 (angle between the refracted ray and the normal) is not equal to the angle of incidence θ1 . Snell’s Law describes the relationship between the angle of incidence θ1 and the angle of refraction θ2 n1 sin θ1 = n2 sin θ2 (2.45) 33 Refracted Ray Material refractive index n 2 Material n 1 refractive index θ 1 θ2 θ1 Reflected ray Incident ray FIGURE 2.11: Snell’s Law of reflection and refraction. From Snell’s Law one can see that if n1 > n2 , then θ2 > θ1 . A larger refractive index means lower speed in a material, if n1 > n2 , then v2 > v1 . The material with the larger speed is associated with a larger angle between the internal ray and the normal as shown in the figure. When θ2 = 90◦ , the refracted ray travels along the material interface, this condition is satisfied when θ2 in equation 2.45 is 90◦ , see Figure 2.12. This value of θ1 corresponding ο θ 2 = 90 Material refractive index n 2 Material n 1 refractive index θ c Incident ray Refracted Ray θ c Reflected ray FIGURE 2.12: Critical angle for fiber. to θ2 = 90◦ is called the critical angle (θc ). 34 When θ1 > θc , there is no refracted ray at all and the incident ray is completely reflected; this is call total internal reflection (TIR), see Figure 2.13. TIR allows a solid optical fiber to guide light within it such that light rays remain internal even when the fiber is bent. Maximum bending radii are part of the specifications for optical fibers. The reason is that optical fiber has a core with refractive index n1 and a cladding with Material refractive index n 2 Material n 1 refractive index θ 1 θ1 Incident ray Reflected ray FIGURE 2.13: TIR for fiber when incident angle exceeds the critical angle. refractive index n2 , where n1 > n2 ; TIR occurs when θ1 > θc . The critical angle is given by equation 2.46 for TIR.[Palais, 2005] θc = sin−1 ( n2 ) n1 (2.46) Provided the refractive index is lower on the other side of the core/cladding interface, no light can pass through, thus most of the light is reflected back into the core. For light to remain constrained within the fiber core an incident ray must have an angle of incidence less than θc . Figure 2.14 shows the structural cross section of an optical fiber waveguide illustrating the angle of acceptance, θaccept and the critical angle θc for an optical fiber. The result is that incident rays greater than angle θna from the normal axis of the fiber are not constrained by the core. The optical fiber acceptance angle is defined as θaccept = 2 θc Rays at angles within θaccept are completed constrained within the fiber core. (2.47) 35 Acceptance angle n2 θ NA Cladding n1 θc Core n2 Cladding Core Refractive Index (n 1) = 1.49 Cladding Refractive Index (n 2) = 1.39 − 1.42 Numerical Aperture (NA) = sin θ c= 0.3 − 0.6 Acceptance Angle = 2 θ c= 60 degrees FIGURE 2.14: Important angle relationships for an OF. Numerical aperture (NA) is a dimensionless number that describes the range of angles over which an optical fiber can accept or emit light. NA is defined as N A = n1 sin θna (2.48) From figure 2.14 θna = 90◦ − θc , thus N A = n1 sin θna = n1 sin(90 − θc ) = n1 cos θc (2.49) From equation 2.46 and sin2 θ + cos2 θ = 1, substituting θc for θ gives NA in terms of the core/cladding indices of refraction cos2 θ = 1 − sin2 θ → cos θc = 1 − n22 /n21 r √ 2 2 p n1 −n2 n21 −n22 2 2 cos θc = 1 − n2 /n1 = = n1 n21 p thus N A = n1 cos θc = n21 − n22 (2.50) Meridional rays are guided down the fiber only if their external angle of incidence on the fiber end-face is smaller than sin−1 (N A), the acceptance angle from equation 2.47. Equation 2.51 gives θaccept in terms of core and cladding indices respectively. q θaccept = sin−1 (N A) = sin−1 ( n2core − n2cladding ) (2.51) Light rays impinging on a fiber at angles greater than the maximum acceptance angle are refracted into the fiber, but are not guided for a long distance. This is because they 36 are not totally reflected at the core/cladding interface, but instead are partially refracted back into the cladding. The maximum acceptance angle, 2θaccept , is a function of both fiber NA and refractive index of the external material. Note that these equations apply only for meridional rays; skew rays have greater acceptance angles.[Kasap, 2001, Palais, 2005] Light is absorbed as it travels through any material (gas, liquid, solid); the intensity I at a distance x is related to the intensity Io at x = 0 by I = Io exp−αx (2.52) where α is the absorption coefficient, which varies from one material to another. Define attenuation loss(dB) = −10log(I/Io ) (2.53) For example, when I/Io = 0.1 the attenuation loss is 10 dB, when I/Io = 0.01 the attenuation loss is 20 dB, when I/Io = 0.001 the attenuation loss is 30 dB, and so on. Rewriting equation 2.52 after taking logarithms and converting to base 10 gives attenuation loss(dB) = 10αx 2.3 (2.54) Note that in equation 2.54 the attenuation loss is directly proportional to x, the distance light travels in the fiber. The attenuation loss per unit length of an optical fiber varies with wavelength because the absorptivity and light scattering out of a material are functions of wavelength. Absorption losses occur when the frequency of the light is resonant with oscillations in the electronic or molecular structure of the fiber material. Different ions and functional groups have characteristic absorption peaks at well-defined frequencies. Scattering that occurs at inhomogeneities in the fiber is linear and there is no change in frequency. Scattering due to phonon-phonon interactions or Raman scattering is nonlinear and there is a frequency change. Typical losses for glass fibers are approximately 1 - 10 dB/km@650 nm. While 37 polymers are not as robust as glass for use as a fiber material due to their high attenuation loss, for example 190 dB/km@650 nm, they are considerably more rugged. Another source of attenuation loss called coupling loss, typically 10-15 db, occurs due to poor coupling between the light source and an optical fiber or between optical fiber and detector. The loss is the result of light from the source that has rays greater than θaccept for the optical fiber, illustrated by Figure 2.15. For example, when a light emitting diode (LED) or scintillator with exit rays within 80◦ to 100◦ is directly attached to the optical fiber, coupling loss still occurs. Using a laser less coupling loss occurs, because laser light diverges less as it travels than diffuse light. Lost power area 11111111111111111111111111111111111111 00000000000000000000000000000000000000 00000000000000000000000000000000000000 11111111111111111111111111111111111111 Optical power source 11111111111111111111111111111111111111 00000000000000000000000000000000000000 00000000000000000000000000000000000000 (LED, fiber, scintillator) 11111111111111111111111111111111111111 000000000000000000 111111111111111111 00000000000000000000000000000000000000 11111111111111111111111111111111111111 000000000000000000 111111111111111111 000000000000000000 111111111111111111 00000000000000000000000000000000000000 11111111111111111111111111111111111111 000000000000000000 111111111111111111 000000000000000000 111111111111111111 00000000000000000000000000000000000000 11111111111111111111111111111111111111 000000000000000000 111111111111111111 000000000000000000 111111111111111111 00 11 00000000000000000000000000000000000000 11111111111111111111111111111111111111 000000000000000000 111111111111111111 00 Emitting11 000000000000000000 111111111111111111 00 11 Optical power coupled in 00000000000000000000000000000000000000 11111111111111111111111111111111111111 000000000000000000 111111111111111111 00 11 area 000000000000000000 111111111111111111 00 11 00000000000000000000000000000000000000 11111111111111111111111111111111111111 000000000000000000 111111111111111111 00 11 000000000000000000 111111111111111111 00000000000000000000000000000000000000 11111111111111111111111111111111111111 000000000000000000 111111111111111111 000000000000000000 111111111111111111 00000000000000000000000000000000000000 11111111111111111111111111111111111111 000000000000000000 111111111111111111 000000000000000000 111111111111111111 00000000000000000000000000000000000000 11111111111111111111111111111111111111 00000000000000000000000000000000000000 11111111111111111111111111111111111111 00000000000000000000000000000000000000 11111111111111111111111111111111111111 00000000000000000000000000000000000000 11111111111111111111111111111111111111 00000000000000000000000000000000000000 11111111111111111111111111111111111111 Junction Cladding Fiber acceptance angle Core Cladding Source radiation envelope FIGURE 2.15: Illustration of fiber coupling issues: light source coupled to fiber. 2.6.2 Metalized Air-Core Capillary Tube Cerenkov Light Removal Various Cerenkov light reduction schemes proposed by different groups typically use one of several methodologies: a background fiber in parallel with the signal fiber, optical filters to attenuate the Cerenkov light which is centered at a different frequency than the signal of interest, temporal methods, spectral discrimination, or more recently the use of metalized air-core capillary tubing.[Matsuura and Miyagi, 2004, Beddar et al., 2004, 38 Justus et al., 2006, Frelin et al., 2006, Lee et al., 2007a] The generation of Cerenkov light within OFs when using electron beams degrades the S/N ratio significantly, introducing additional noise into the system.[Beddar et al., 1992a, Elsey et al., 2007] This light varies with the speed of the particle in a material and the index of refraction of the material, and is a function of impinging electron beam angle on the OF. [Law et al., 2007] However, it is an important noise source only for electron beams above 200 keV and can be neglected for diagnostic radiation detection purposes and low energy radiation detection. In this work metalized air-core capillary tube was chosen to reduce the Cerenkov light generated within OFs, as shown in Figure 2.16.[Lambert et al., 2008] No Cerenkov light is generated within the air-core. Cerenkov light generated within the glass material of the capillary tube is reflected by the metal film. Within the air-core itself, scintillation light is guided by grazing angle reflections from metal film into the attached optical fiber. Other Cerenkov Removal Methods The time signature of light signals originating within optical fibers when irradiated by high energy electron beams is different, thus providing a way to separate the Cerenkov noise from the radiation dependent signal: Cerenkov - ps, native fluorescence - ns, phosphorescence - 100s of µs. [Justus et al., 2004, 2006] In theory, careful time discrimination allows one to exclude the Cerenkov noise signals when using a high energy pulsed radiation source (as a LINAC). [Beddar, 2007] describes a PMMA fiber coupled to plastic scintillator dosimeter that does not require optical stimulation for reading. This material has one principle advantage in that it is near tissue equivalent.[Beddar et al., 1992a,b] Cerenkov noise signals are removed using two adjacent identical fibers, one with a scintillator and one without, and mathematically subtracting the different signals.[Beddar et al., 2001] 39 (a) Cross section of metalized air-core capillary tube. (Source: Polymicro Inc.) Plastic Optical Fiber 111111 000000 000000 111111 000000 111111 Glass Cladding 000000 111111 000000 111111 000000 111111 111111111111 000000000000 000000 111111 1111111111111111111 0000000000000000000 00000000 11111111 000 111 11111111111111111111111111111111 00000000000000000000000000000000 1010 00000000 11111111 000 111 Air Core Fiber Core 00000000 11111111 000 111 1010 0000000000000000000 1111111111111111111 00000000 11111111 000 111 000000000000 11111111111111111111111111111111 111111111111 00000000000000000000000000000000 0000000000000000000 1111111111111111111 000000000000 111111111111 Scintillator Air−Core Silver Coated Capillary Tubing Metal Film Silver Coating (b) Interfaces between metalized air-core glass capillary tube and scintillator. FIGURE 2.16: Illustration of metalized air-core glass capillary tube. 2.6.3 Other Optical Fiber Radiation Effects An extensive overview of radiation effects on optical fibers in various environments (Space, Nuclear Industry) was completed by NATO.[Berghmans et al., 2007] Radiation Induced Attenuation (RIA) in optical fibers is the result of trapping radiation generated electrons and holes at defect sites creating color centers. Typically, optical fiber color centers absorb photons at specific wavelengths but can be annealed out, decreasing RIA, by thermal and optical means. Many factors influence optical fiber response to ionizing radiation, among them:[Berghmans et al., 2007] • previous irradiation episodes and thermal cycling, 40 • chemical composition, • ambient temperature, • type of radiation exposure, dose-rate, and total dose, • test parameters (optical wavelength and power), • time lag between radiation exposure and optical measurement. In the past, simple power-law and complex nth-order kinetics have been used to model RIA in optical fibers without much success. At this time there is no commonly accepted dose effect model. To date there have been studies investigating standard SiO2 fiber material for use as Thermo-Luminescent-Dosimeters (TLDs) with mixed results.[Espinoza et al., 2006] 2.7. SiO2 : Cu2+ Material Justus and Huston from the U.S. Naval Research Laboratory describe scintillation and OSL from Cu1+ doped fused quartz fibers for use as a dosimeter.[Benevides et al., 2007, Huston et al., 2001, 2002, Justus et al., 2004, 2006] This material has an OSL peak at 500 nm and a stimulation peak at 390 nm. Due to issues with the OSL mode of operation they abandoned this method is favor of a more passive system using the material drawn into a fiber used as a scintillator. The resulting signal is similar to OSL but no optical stimulation is needed when using the fibers as radiation sensors or dosimeters.(This is what other OSL researchers call the Radio Luminescence or (RL) signal.) For this work the SiO2 : Cu2+ fibers themselves will be used as radiation sensors in RL mode; no OSL mode of operation is utilized. 41 3. 3.1. EXPERIMENTAL METHODS SSPMs and Device Characterization The primary SSPMs tested in this work include Hamamatsu MPPC S10362-11050C, Photonique 0810G1, and Voxtel SQBF-EKAA/SQBF-EIOA. All are Si devices utilizing an MRS structure for pixel quenching where a typical broad spectrum device extending from the UV/blue range through red-yellow range is a p+pn+ structure. Sizes tested in this work had 1 x 1 mm2 or 1.1 x1.1 mm2 active areas divided into different array sizes. Individual pixels are separated by grooves filled with an optically opaque material for suppression of optical crosstalk which reduces SSPM noise. Each photosensitive area is protected with 1.58 index of refraction transparent epoxy. Custom electronic charge and transimpedance (MAR-8ASM+ RF) amplifiers were constructed for the three SSPMs tested: MPPC S10362-11-050C, Photonique 0810G1, and Voxtel SQBF-EKAA/SQBF-EIOA.[Hamamatsu, 2007, Voxtel, 2008, Photonique, 2006] Two versions of Voxtel SSPMs were evaluated: one a room temperature device (SQBFEIOA) in a T0-8 package and another using the same semiconductor die mounted to a 3-stage thermoelectric cooler in a modified T0-8 package (SQBF-EKAA) to minimize DCR.[Dhulla et al., 2007] Dedicated circuitry (Maxim MAX1978) was used for the temperature controller of the SQBF-EKAA for laboratory tests. In addition to the Photonique 0810G1, four other Photonique SSPM devices were evaluated: 0701BG, 0611B1, 050701GR, and a Custom Low Noise Sample (CLNS). The amplified SSPM pulse or charge signal was coupled to a custom designed discriminator circuit feeding an ORTEC 996 counter/timer or host computer running analysis software, as shown in Figure 3.1. The photocurrent for SSPM tests was generated using blue and red light from high intensity LEDs using a custom designed pulser circuit with extensive modifications based upon [Kapustinsky et al., 1985]. This pulser outputs 1 ns pulses which gave a few photons 42 Scintillation crystals Reference PMT H5783P Custom fiber and scintillator connector heads LED Pulser 1) GH4001 POF with BCF12 scintillator 2) Glass OF with SiO2:Cu2+ scintillator USB connected PC Running analysis software SSPM Detectors: MPPC Photonique, Voxtel Amplifier Threshold Discriminator Frequency Counter Ortec 996 Counter/Timer Oscilloscope FIGURE 3.1: Block diagram of laboratory and clinical measurement system. 43 from the LEDs after attenuation. A reference photomultiplier tube (Hamamatsu PMT H5783P) was used as a benchmark photon detector coupled to the same photon counting electronics used with the SSPMs. A Hamamatsu S1223 Si PIN photodiode was used as a reference photodetector for calibration purposes due to its known responsivity. A Tiffen P Series ND.6 Grad SE Neutral Density Filter mounted to an optical xyz stage was used to attenuate the photon flux from the LEDs and LEDs coupled to test fibers from the pulser. A Newport Digital Power Meter Model 815 was used to measure the optical power. Special light-tight mating jigs were used to hold in place the different scintillation crystals or POFs to each SSPM device and PMT for laboratory testing purposes. Signal c acquisition was also done using LabVIEWsoftware controlling Tektronix oscilloscopes. 3.1.1 Photon Counting Electronics A brief overview of the photon counter operation is provided with reference to Figure 3.2. The window comparator circuits (U4, U5 in Figure 3.6) allow a lower threshold level to be set that screens out SSPM dark noise (0.5 p.e. black line in Figure 3.2). The amplitude of this threshold is arbitrary, depending upon the application. This same circuit allows an upper threshold level to be set. Both upper and lower threshold levels are used to create a window than can be used for energy discrimination purposes. This feature was not used for this work. Figure 3.3 shows scope photos of MPPC S10362-11-050C dark noise pulses (note the individual p.e. levels), and Figure 3.4 shows scope output pulses from the photon counter. Because each SSPM tested had a different reverse breakdown voltage and for portability reasons (battery operation), the circuit shown in Figure 3.5 was designed and built for laboratory and clinical testing purposes. This is a switching power supply circuit that provides stable, extremely low ripple SSPM power from 18 V to 80 V for biasing purposes. Figure 3.6 shows the single channel analyzer (SCA - photon counter) used in this work. This circuit gives an output pulse when an input voltage from the MAR8+ amplifier 44 FIGURE 3.2: Photon counter operation. (source: Hamamatsu) FIGURE 3.3: Dark counts showing p.e. levels. FIGURE 3.4: Typical output from photon counter. 45 FIGURE 3.5: Circuit diagram for battery operated SSPM power supply. 46 falls between upper and lower threshold voltage levels (set by the user using potentiometers or host computer interface). It is a high speed window comparator design with feedback hysteresis to prevent oscillation when switching from high to low voltage levels and low to high levels. Figure 3.7 shows the reference power supplies for the window comparator and logic circuits. The DC reference level is provided by U12, a variable shunt regulator driven from a current source U11, R11. The maximum DC reference level is 10 V for compatibility reasons with NIM-BIN systems. With this circuit the reference voltage levels for the window comparator circuit may be adjusted to screen out SSPM noise ( photo-electron equivalent, p.e.), thereby setting a lower limit of detection. Measured absolute photon counts (cpm), detector efficiency, and scintillator coupling issues are extremely important for low level environmental radiation sensing. Measured activities based on photon counts are not reported for two reasons: coupling issues (area mismatch between scintillators and SSPM die, OF coupling inefficiencies) and the lack of suitable calibration radiation sources. (The exception was when using the reference PMT together with the BGO scintillator directly attached as the BGO diameter closely matched the PMT diameter.) 3.1.2 I-V and CV Measurements Current-Voltage (I-V) curves for all SSPMs were measured with an Agilent 4156 semiconductor parameter analyzer, pixel capacitance was measured with an HP 4263B LCR meter - both measurements with the Device Under Test (DUT) SSPM in total darkness. These measurements were used to assist with characterizing the DCR and gain. Operating temperature for all SSPM tests was ambient, 22.5 ◦ C. For the SQBF-EKAA, operating DUT temperature was set at −20 ◦ C. Thirty (30) engineering samples of the room temperature Voxtel SQBF-EIOA SSPM were I-V tested to optimally select a device for low dark current. 47 FIGURE 3.6: Circuit diagram for photon counter. 48 FIGURE 3.7: Power supply circuit diagram for photon counter. 49 The multiplication gain of each SSPM was calculated using the measured die capacitance and the photocurrent/dark current measurement results using equation 2.26. The C-V measurement is used to determine the parasitic capacitance of the SSPM at a given reverse bias voltage. It is also used to determine the punch-through voltage of the SSPM, when its value is ambiguous from the I-V results. The SSPM capacitance is extremely sensitive to the reverse bias due to the strong dependence on the depletion layer thickness. For an SSPM device whose punch-through is difficult to determine by I-V measurements, C-V measurement is an effective diagnostic tool. Another parameter one can extract from C-V data is the doping level in the absorption region of the SSPM. How fast the capacitance decreases with the bias depends on the doping level of the absorption region. Capacitance, hence doping level ND is calculated from equation 3.1 where ǫ is the permittivity of the absorbing material. s Csspm = A ǫǫ0 qND [Farads] 2(Vbias − Vovervoltage ) (3.1) Above the breakdown voltage, capacitance cannot easily be measured. It is assumed that the total capacitance after breakdown is the same as was measured before breakdown. From this information the capacitance per pixel is calculated as Cpixel = Ctotal /n, where n is the number of pixels in the SSPM array. The actual SSPM photosensitive area is found by dividing the total area by (1 - fill factor), where “fill factor” is the total area of active pixel elements. For the Photonique, Hamamatsu, and Voxtel devices this is approximately 1 mm2 , thus A = Aπ/(1 − f illf actor). Now the thickness of the avalanche region can be determined: tavalanche = Aǫ0 ǫSi [m] Ctotal (3.2) 50 3.1.3 SSPM Gain Measurement The definition of SSPM gain typically uses a charge amplifier coupled to an Analog- to-Digital Converter (ADC) that produces a frequency distribution of charge vs. channel number; Figure 3.8 shows an example. This is essentially the same functionality used in energy spectroscopy Multi-Channel Analyzers (MCAs). The LeCroy 4300B 16-channel charge ADC is an example with a sensitivity of 0.25 pC/count. SSPM gain is given by: Gain = ADC conversion rate × number of ADC channels between two peaks q Note that this equation is the same as equation 2.26. For example, if the distance between two adjacent peaks is 100 channels (Figure 3.8), then the gain is 100 × 0.25 · 10−12 C/1.6 · 10−19 C = 156250. A charge amplifier ADC and MCA were not available for this work. Instead the method described here was used. In this work the SSPM charge signal was measured using an oscilloscope to determine area. Together with equation 2.26 and a constant 50Ω load, Qsspm was determined by integrating the area on the oscilloscope. Z V (t) dt = Z R · I(t) dt = 50Ω · Qsspm(t) where gain is proportional to the overvoltage Vov . Note that gain (G) can also be calculated using the leakage current Ileak and the DCR as shown in equation 3.3.[Petasecca et al., 2008] G= 3.1.4 Ileak q DCR (3.3) SSPM DCR Measurement DCR was measured using the setup shown in Figure 3.9. Referring to Figure 3.3, threshold was set to 0.5 p.e. for this measurement. Frequency was recorded using a 51 FIGURE 3.8: ADC output charge frequency distribution. (source SensL) 52 Tektronix DC 502 frequency counter. Counting the number of pulses exceeding the 0.5 p.e. threshold gives the number of times one or more photons has been detected. Optical Fiber LED Pulser Optical Fiber Neutral density filter in xyz optical stage SSPM MAR8+ Amplifier Oscilloscope Pulser trigger signal Frequency Counter FIGURE 3.9: SSPM DCR measurement setup. 3.1.5 SSPM PDE Measurement PDE(λ) was measured using the setup shown in Figure 3.10. In this measurement the S1223 photodiode was used first to measure the photon flux, then replaced by the DUT SSPM. PDE is determined by equation 3.4. Because the number of photons detected are calculated from a photocurrent, afterpulsing and optical crosstalk effects are also included.[Gomi et al., 2007, Hamamatsu, 2007] P DE = Optical Fiber LED Pulser P IN Active Area SSP M detected photons SSP M Active Area P IN incident photons Optical Fiber Neutral density filter in xyz optical stage (3.4) Dark Box SSPM or photodiode Ammeter MAR8+ Amplifier Oscilloscope FIGURE 3.10: SSPM PDE measurement setup. Light power, P, is the number of photons impinging on a photodetector (PIN, SSPM, PMT) each second, Nphotons , multiplied by the power of each photon. The power of each photon is h ν. Thus P = Nphotons h ν and Nphotons = P/h ν = P λ/h c. The number 53 of optical photons incident on a photodetector (PIN, SSPM, PMT) can be approximated using an optical power meter; this is given by equation 3.5 Nphotons ≈ P (W ) λ(m) Rate(Hz) hc(Jm) (3.5) where P(W) is the power measured using a Newport Digital Power Meter Model 815, λ is the peak wavelength of the LED, and Rate is the pulse repetition rate for the LED. Alignment was done on an optical rail. 3.1.6 GRIN Lens Design for the Voxtel SQBF-EKAA The photosensitive die of the SQBF-EKAA is recessed 2.73 mm behind the clear glass window in the Peltier-cooled T0-8 package, see Figure 3.11. To optimally couple FIGURE 3.11: Voxtel SQBF-EKAA mounted on 3-stage Peltier-cooler showing recessed die. Can diameter is approximately 0.5 inch. 54 light from a scintillator crystal or OF to the die, some type of lens system is needed. For this work a GRIN lens system was chosen. A bi-convex lens system was evaluated in the optical lab but found too difficult for practical use with the cooled TO-8 package and 1 mm or 400 µm diameter OFs. Figure 3.12 shows the key GRIN lens system parameters used for the SQBF-EKAA. The design information for SELFOC GRIN lenses from NSG America was used to calculate parameters for the SQBF-EKAA SSPM.[NSG, 2008] GRIN optic area Diam(fiber) Diam(sspm) active area θ1 θ2 Grin Lens n1 n2 Fiber core SSPM Z L1 L2 FIGURE 3.12: GRIN lens system showing important dimensions. The on axis refractive index with λ in µm is given in equation 3.6. N0 (λ) = 1.5868 + 8.14 · 10−3 λ2 The index gradient constant for the SLW-1.0 with λ in µm and √ A(λ) = 0.5945 + A(λ) = 0.3238 + √ A in mm−1 : 3.936 · 10−3 5.539 · 10−4 + λ2 λ2 The index gradient constant for the SLW-1.8 with λ in µm and √ (3.6) 5.364 · 10−3 2.626 · 10−4 + λ2 λ2 For pitch P and lens length Z the relationship is: (3.7) √ A in mm−1 : (3.8) 55 √ 2πP = Z A (3.9) Table 3.1 shows calculated values for SLW GRIN lenses. c SLW GRIN lenses at λ = 440 nm and 580 TABLE 3.1: Calculated values for SELFOC nm. SLW-1.0 (440 nm, 580 nm) SLW-1.8 (440 nm, 580 nm) N0 - on axis refractive index √ A in mm−1 1.6288, 1.6101 1.6288, 1.6101 0.6296, 0.6111 0.3585, 0.3420 Pitch length P - 0.23 (mm) 2.2953, 2.2364 4.0301, 4.2247 Pitch length P - 0.25 (mm) 2.4948, 2.5705 4.3814, 4.5920 For the image distance itself L2 : √ √ √ −(n1 n2 / A sin(Z A) − n2 N0 L1 cos(Z A)) √ √ √ L2 = n1 N0 cos(Z A) − N02 L1 A sin(Z A) (3.10) For the transverse magnification MT : MT = n1 √ √ √ n1 cos(Z A) − N0 L1 A sin(Z A) (3.11) For this system the height of the object on the SSPM die Hobj is: Hobj = Diamimage |MT | (3.12) To simplify the optical system construction L2 was set equal to 2.73 mm; the rear face of the GRIN lens is butted directly against the SQBF-EKAA cover glass centered over the photosensitive die area. The image size is also a constant equal to 1.0 mm for SQBF-EKAA (the diameter of the photosensitive area). The object distance L1 can be found by solving equation 3.10 for L1 : 56 √ √ √ −(n1 n2 / A sin(Z A) − n1 N0 L2 cos(Z A)) √ √ √ L1 = n2 N0 cos(Z A) − N02 L2 A sin(Z A) (3.13) Because SELFOC GRIN lenses have an active area on the frontal plane equal to 60% of the lens diameter, the maximum fiber diameter that can be used is Diamf iber = 0.6Diamlens − 2L1 tan(θ1/2 ) (3.14) The NA for the GH4001 POF is 0.50; the half-angle theta1/2 used in equation 3.14 is arcsin(0.5) = 30◦ . Figure 3.13 shows the results for object height Hobj , object distance L1 , and fiber core diameter Diamf iber for both 400 µm and 1.0 mm cores. Looking at the graphs one can make some design decisions: • The height of the projected image object on the SQBF-EKAA photosensitive die when using the SLW-1.8 GRIN is greater than for the SLW-1.0 GRIN. • Where the object distance L1 and object height Hobj lines cross in Figures 3.13(a), and 3.13(b), there are solutions which give lens lengths Z = 2.65 mm and 1.85 mm, respectively. • Overall, there is a poor GRINS lens solution for 1 mm core diameter optical fibers when using off the shelf SELFOC lenses. The end result of the GRIN lens optical analysis is that there is no optimum solution when using off the shelf lenses for 1.0 mm core diameter optical fibers coupled to the 1.0 mm recessed die SQBF-EKAA. To maximize the light collection efficiency, a custom GRIN lens must be designed and manufactured to optimize three parameters: optical fiber core diameter, projected object height on the SQBF-EKAA die, and object distance. Given these constraints the system was laboratory tested using the off the shelf SLW18-S0250130-NCO and SLH10-S0250-130-NCO GRIN lenses coupled directly to the SQBF-EKAA 57 Dependencies for SLW18 Grin Lens 2.5 L1, Hsspm, Dfiber (mm) 2.0 1.5 1.0 0.5 L1 Object Distance 0.0 Hsspm Dfiber -0.5 2 3 4 Z (mm) (a) SLW-1.8 Dependencies for SLW10 Grin Lens 2.0 1.5 L1, Hsspm, Dfiber (mm) L1 Object Distance Hsspm Dfiber 1.0 0.5 0.0 -0.5 1.5 2.0 2.5 3.0 Z (mm) (b) SLW-1.0 FIGURE 3.13: GRIN lens design parameters for the Voxtel SQBF-EKAA. 58 glass cover using special light tight mating jigs. It was prohibitively expensive to do a custom GRIN lens manufacturing run for this project. However, Voxtel has available a custom GRIN lens solution for 65/125 GOFs for use in DNA testing. 3.2. Optical Coupling using Adhesives To ensure durable yet light transmissive joints between scintillators and optical fibers (or capillary tubes) different types of UV cement were evaluated: Norland NOA68, TRI-Con F114, Dymax OP-52/OP-4-20639, and Master Bond EP21-7P-Clear. These optical cements attempt to balance the need for rigid connections with refractive index matching between two different materials. When curing these cements, a UV light source is required. A 337 nm UV laser (Laser Science model VSL-337-NDS) was used for this work with an initial curing time of 4 hours. Setting was completed overnight. After repeated laboratory tests with pieces of GH4001 optical fiber, plastic scintillators, and glass optical fiber; Norland NOA-68 cement was chosen for this work. Note that polyacrylate glue (“Crazy Glue”) degrades OF material if not applied with extreme care. 3.2.1 Characteristics of Optical Fibers used in this work Table 3.2 shows key parameters for the PMMA ESKA GH4001 optical fiber cable and the glass SiO2 optical fiber used in this work. 3.3. Scintillators Two different inorganic scintillator materials and two organic materials were used during SSPM laboratory testing: 1 mm x 1 mm x 5 mm square cylindrical Prelude 420 (Lu1.8 Y.2 SiO5 : Ce), 8 mm x 10 mm long cylindrical BGO (Bi4 Ge3 O12 ) crystals, 1 mm x 3 mm long cylindrical BC430 plastic scintillator, and 1 mm x 3 mm long lengths of cylin- 59 TABLE 3.2: Characteristics of GH4001 cable and SiO2 optical fiber. Specification† GH4001 SiO2 Fiber Diameter 1 mm 400 µm Refractive indices ncore = 1.492 ncore = 1.458 nclad = 1.402 nclad = 1.441 Angle of acceptance (θcritical ) 60◦ 25.4◦ Numerical aperture 0.50 0.22 400 - 1000 nm 200 - 1600 nm 190db/km@650 nm 10db/km@650 nm 25 mm 275 mm -55◦ C - 70◦ C -40◦ C - 85◦ C Black polyethylene Black Nylon Spectral range Attenuation Minimum bend radius Temperature range Jacket †- Data are from Mitsubishi and US Naval Laboratory. 60 drical BCF-12 plastic scintillating fiber (Saint-Gobain Crystals, Hiram OH.). Inorganic scintillators were first painted with BaSO4 reflective paint, wrapped with Teflon tape, and covered with black epoxy before being coupled to each SSPM active area using optical grease (Saint-Gobain Crystals, Hiram OH.). BC430 plastic scintillator was chosen to test the greater SSPM PDE in the visible light ranges; it was tested in addition to BCF-12 scintillating fiber during laboratory tests. BCF-12 and BC430 scintillators were polished (both ends), painted with BaSO4 reflective paint, wrapped with Teflon tape, and covered with black epoxy before coupling to each SSPM active area using optical grease.[Webber and Christ, 2003] The count rate from each SSPM-scintillator combination was measured when exited by low activity radiation sources: (1.842 µCi 60 Co, 0.661 µGy/s and 1 µCi 137 Cs, 0.201 µGy/s) were used to stimulate the BGO and BCF-12 scintillators in the laboratory. The Prelude 420 crystals are self-poisoned with radioactive 176 Lu, providing ≈ 1.3 Hz scintillations that can be detected without using the low activity external radiation sources (see Appendix D3). Figure 3.14 shows the BCF-12 scintillating fiber emission spectrum and the BC430 emission spectrum measured with an Ocean Optics PC1000 spectrometer. The peaks (Hg) in the emission spectrum are from the room lights on during the measurement. Table 3.3 shows the industry published emission characteristics of scintillators used in this work for comparison purposes. The photon count rate (cpm) with 0.5 p.e. threshold was measured (5 samples each averaged over 1 minute) for each SSPM with Prelude 420, BGO, BC430, and BCF-12 scintillators directly attached to the transparent window surface of the package. These same scintillators were then attached to 2.5 m lengths of Mitsubishi Eska GH4001 POF, coupled to each SSPM, and the photon counting measurements repeated. Clinical measurements with the GOF coupled SiO2 : Cu2+ and POF BCF-12 scintillators were done using the MPPC S10362-11-050C SSPM together with a USB-based photon counter and PC software supplied by Hamamatsu. A Varian Clinac 2100CD Lin- 61 BCF-12, BC430 Scintillator Absorption Spectrum 2500 BCF12 BC430 Intensity (counts) 2000 1500 1000 500 0 400 500 600 Wavelength (nm) FIGURE 3.14: BCF-12, BC430 scintillator emission spectrum 62 TABLE 3.3: Summary of published scintillator emission characteristics. Scintillator Material Emission Photons Decay time Density λ (nm) (per keV) (ns) (g/cm3 ) SiO2 : Cu2+ 500 † † 2.634 BCF-12 435 ≈8 3.2 1.05 BC-430 580 ≈9 16.8 1.032 Lu1.8 Y.2 SiO5 : Ce 420 32 41 7.1 480 8-10 300 7.13 (Prelude 420) Bi4 Ge3 O12 (BGO) †- Not available. Sources: http://www.detectors.saint-gobain.com/MaterialsGasTubes.aspx http://scintillator.lbl.gov 63 ear Accelerator was used for percent depth dose, dose linearity, and angular dependence measurements. All clinical measurements were verified with a benchmark calibration using a Wellhofer CC13 ion chamber with 0.13 cm3 detector volume (hereafter called the expected value). Unless otherwise noted 10 cm x 10 cm radiation fields (photon, electron beam) were used during measurements. 3.4. 3.4.1 Optical Fiber System Efficiency Considerations Estimated Detected Light from Scintillators Light output from scintillators when using 60 Co and 137 Cs low activity sources was used to estimate the number of photons detected by each SSPM when directly coupled. The number of detected photons is: Nphotons = Energy × scintillator light yield × P DE(λ) × area f actor (3.15) where the area factor is the ratio of the SSPM die area to the scintillator area, PDE is the manufacturer’s given photon detection efficiency, and scintillator light yields are from Table 3.3. For the S10362-11-050C SSPM directly coupled to the Prelude 420 scintillator, the number of detected photons is: Nphotons = 662 keV × 32 photons/keV × 0.45 × 1 = 9532.3 p.e.. Assuming a conservative Gain M = 2 × 105 for each SSPM, the number of electrons at the SSPM output is; Nphotons = 9532.3 p.e. × 2 × 105 = 1.9 × 109 , multiplying by q gives the charge Q = 1.9 × 109 × 1.6 × 10−19 = 3.05 × 10−10 C. Now using an appropriate time integration window, which is a function of the scintillator decay time τ , one can estimate the SSPM output current I: I(t) = dq −q −t/τ = e = 3.05 × 10−10 /50 × 10−9 = 6.1 mA dt τ (3.16) for this example. These calculations predict that oscilloscope signals in the 10s of mV range should be observed without external amplification when using a 50 Ω load. During laboratory tests the reference PMT scintillator combination exhibited oscilloscope signals 64 in the range that could be viewed without external amplification. This illustrates that light coupling losses between the SSPM die and various scintillators is significant; measured PDE for the SSPMs was sometimes less than indicated by the manufacturer. Table 3.4 summarizes the expected number of detected photons and output current I when using 137 Cs as stimulator for photoelectrons. TABLE 3.4: Estimated Scintillator Outputs: number of photons and SSPM current I. Scintillator Material BCF-12 PH0701BG PH0810G1 S10362-11-050C SBQF-EIOA 79440 52960 238320 66200 254.2 mA 84.7 mA 762.6 mA 42.37 mA 119160 119160 104265 52132 152.5 mA 152.5 mA 133.46 mA 66.73 mA Lu1.8 Y.2 SiO5 : Ce 3177.6 2118.4 9532.8 2648 (Prelude 420) 2.0 mA 1.35 mA 6.1 mA 1.69 mA Bi4 Ge3 O12 215.15 182.05 324.38 165.5 (BGO) 19.7 uA 16.65 uA 29.65 uA 15.13 uA BC-430 3.4.2 Efficiency: Overall Light Coupling Details of the overall light collection efficiency calculations for an OF radiation sensing system are in Appendix B.[Beddar, 2007, Lacroix et al., 2008] Figure 3.15 shows the coupling efficiencies for the components in the optical signal chain. The coupling 65 efficiencies at each material junction in the signal path, when combined together, provide a gross estimate of (fiber-only and metalized air-core capillary tube-fiber) overall optical system losses. Fiber Coupling Efficiencies Contributing to S/N Ratio scintillator optical fiber εtransmit−OF εaccept εscint−OF scintillator εaccept SSPM εPDE−sspm εOF−sspm capillary tube optical fiber εtransmit−CAP εtransmit−OF εscint−cap εcap−OF SSPM εPDE−sspm εOF−sspm FIGURE 3.15: Light collection efficiencies for components in the optical signal chain. For the non-capillary tube system: ǫlight−collection = ǫaccept · ǫscint−OF · ǫtransmit−OF · ǫOF −sspm and for the metalized air-core capillary tube OF system: ǫlight−collection−captube = ǫaccept · ǫscint−cap · ǫtransmit−cap · ǫcap−OF · ·ǫtransmit−OF · ǫOF −sspm where ǫaccept is the fraction of light photons produced by the scintillator traveling towards the optical fiber that fall within the optical fiber core acceptance cone. It is estimated by the solid angle of isotropic light originating from the scintillator center which falls within the acceptance cone of the OF. Here there is no assumption that light produced in the scintillator itself is 100% reflected back by reflector materials (BaSO4 optical paint, reflective tape). Worst case calculation gives ǫaccept = 0.01. ǫscint−OF is the fraction of light coupled from the scintillator into the optical fiber at the interface between the two. Here a value of 0.5 was assumed.[Attix, 1986] ǫtransmit−OF is the transmission efficiency of the optical fiber itself. For 2.5 m of 66 Eska GH4001 fiber ǫtransmit = 0.896 (see Appendix B). ǫtransmit−cap is the transmission efficiency of 30 cm of capillary tube. This value was measured using the reference photodiode (green LED), = 0.6. ǫscint−cap is the fraction of light coupled from the scintillator into the capillary tube distal end. Measurements using the reference photodiode and Prelude 420 crystal as scintillator were inconclusive. A course estimate was obtained using (green LED) = 0.27. ǫcap−OF is the fraction of light coupled from the proximal capillary tube end into the optical fiber. This value was measured using the reference photodiode (green LED), = 0.4. ǫOF −sspm is the fraction of light coupled from the OF to the SSPM photosensitive die. Attempts to measure this value with consistency in the laboratory were not conclusive. A value of 0.5 was assumed. This may be further broken down into a component that includes the OF-FC fiber connector interface. This was not done in this study. However, during laboratory testing the FC fiber connector needed to be positioned nearly flush against the SSPM photosensitive die to achieve a measurable signal with low level light sources. Combining these numbers gives: ǫlight−collection = 0.01 · 0.5 · 0.896 · 0.5 = 0.0022 and for the capillary tube/optical fiber system, ǫlight−collection−captube = 0.01 · 0.27 · 0.6 · 0.4 · 0.896 · 0.5 = 0.00029. It is clear from this analysis that ǫaccept is the limiting term. Robust fiber coupling techniques are needed to optimize this value (polishing, optical concentrators, etc.). Further, the inclusion of the capillary tube reduces that light coupled into the fiber by a factor of 0.6. 67 3.4.3 Optical Signal-To-Noise-Ratio (S/N) Considerations Whether Cerenkov light is produced is a function of material refractive index and incident electron energy given by equation 2.12. Figure 3.16 shows the Cerenkov threshold energy vs. material refractive index for different OFs and scintillating materials used in this work. As the material refractive index increases, the threshold energy decreases. Cerenkov Threshold Energy vs. Material Refractive Index Cerenkov Threshold Energy (MeV) 3.0 2.5 2.0 1.5 Cerenkov Energy 1.0 0.5 0.0 1.0 1.2 1.4 1.6 1.8 2.0 Index of Refraction FIGURE 3.16: Cerenkov threshold energy vs. material refractive index. Using equation 2.13, Figure 3.17 plots the angle where Cerenkov light begins to be an important noise source vs. incident energy for different OFs and scintillating materials used in this work. Note that at diagnostic imaging energies and for low level radiation detection, Cerenkov photons can be ignored. 68 Cerenkov Angle vs. Incident Electron Energy Cerenkov Angle (degrees) 50 45 GH4001 POF, ncore = 1.4902 SiO2:Cu2+ OF, ncore = 1.458 BC-430, ncore = 1.58 BCF-12. ncore = 1.6 40 0 2 4 6 8 10 12 14 16 18 20 Energy (MeV) FIGURE 3.17: Predicted Cerenkov angle vs. incident energy. 69 For light in the visible range (λ2 − λ1 , 650 nm - 400 nm) one can calculate the expected number of light photons emitted by Cerenkov radiation in optical fiber. For the BCF-12 and SiO2 : Cu2+ scintillators with center signal wavelengths between 420 nm - 440 nm equation 3.17 gives,[Jelly, 1958] Nphotons = 2παl( 1 1 1 − )(1 − 2 2 ) λ2 λ1 β ncore (3.17) where α = e2 /hc ≈ 1/137 is the fine-structure constant. l is electron depth in the scintillator material. λ2 is the upper wavelength of interest. λ1 is the lower wavelength of interest. ncore is the refractive index of the fiber core. β is the ratio of velocity in the fiber to the speed of light in a vacuum. For a 2 given electron energy its velocity is calculated using KE = √ mo c2 1−v /c2 − mo c2 . For a 2 MeV electron this gives v = 0.979c. β = vparticle−in−f iber /c, thus 0.979c/c = 0.979 and β 2 = 0.9584. Table 3.5 shows v for a few energies. TABLE 3.5: v for a few energies. Energy (MeV) Particle speed v 0.1 0.5482c 0.25 0.7410c 2 0.9790c 6 0.9969c 9 0.9985c For 6 MeV electrons moving through a depth l of plastic or glass optical fiber (1 70 mm or 400 µm, repsectively) with core refractive indices of 1.492 or 1.458 in the visible light range, the results are: Nplastic ≈ 24.2 light photons Nglass ≈ 9.3 light photons which shows that Cerenkov light is a more significant noise contributor in plastic fibers than in glass fibers. At higher energies more Cerenkov photons are produced, reaching a maximum as β 2 approaches c, which degrades the S/N ratio. Figure 3.18 shows the number of Cerenkov photons produced vs. energy for different material indices of refraction (1 mm x 1 cm long material lengths). Using the refractive index information from Table 3.2, one can calculate the critical angle for the plastic fiber and the maximum Cerenkov angle from equation 2.46 and equation 2.14 respectively. The results are 70◦ and 47.91◦ , respectively. Figure 3.19 shows the situation when the electron beam is perpendicular to the optical fiber, and the cone of Cerenkov radiation is less than the angle θcritical for the optical fiber. Figure 3.20 shows the cone of Cerenkov radiation when it approaches the θcritical angle for the optical fiber as the electron beam gantry angle from the LINAC is changed. A response maximum is observed when the cone of Cerenkov radiation rotates with the electron beam, and the Cerenkov cone angle approaches θcritical for the optical fiber itself. The inclusion of metalized air-core capillary tubing to mitigate Cerenkov produced light photons attenuates further the optical signal as described in the previous section. 3.5. SiO2 : Cu2+ Fiber Optic Probe Design A 400 um diameter x 5 mm long round cylindrical length of SiO2 : Cu2+ scintillator, supplied by the Naval Research Laboratory, Washington DC., was butt coupled using Norland UV cement and cyanoacrylate glue to a 2.5 m length of GOF. An FC fiber 71 Cerenkov Photons Emitted vs. Material Refractive Index Cerenkov Photons between 400nm and 650nm 28 26 24 22 20 GH4001 POF, ncore = 1.4902 18 GH4001 POF, nclad = 1.402 16 SiO2:Cu2+ OF, ncore = 1.458 SiO2:Cu2+ OF, nclad = 1.441 14 BCF-12, ncore = 1.6 12 BCF-12, nclad = 1.49 BC-430, ncore = 1.58 10 8 6 4 2 0 0 2 4 6 8 10 12 14 16 18 20 Energy (MeV) (a) 1 mm material depth Cerenkov Photons Emitted vs. Material Refractive Index Cerenkov Photons between 400nm and 650nm 280 260 240 220 200 GH4001 POF, ncore = 1.4902 GH4001 POF, nclad = 1.402 180 SiO2:Cu2+ OF, ncore = 1.458 SiO2:Cu2+ OF, nclad = 1.441 BCF-12, ncore = 1.6 160 BCF-12, nclad = 1.49 BC-430, ncore = 1.58 140 0 5 10 15 20 Energy (MeV) (b) 1 cm material depth FIGURE 3.18: Number of Cerenkov photons produced in visible light spectrum. 72 Electron beam cladding n2 = 1.492 Optical fiber core θ critical θ Cerenkov−Max = 47.91 = 70 cladding n1 = 1.402 FIGURE 3.19: Cone of Cerenkov Radiation: electron beam ⊥ to fiber. Electron beam cladding n2 = 1.492 Optical fiber core θ critical Cerenkov cone angle changes with beam angle = 70 cladding n1 = 1.402 FIGURE 3.20: Cerenkov cone approaching θcritical for the optical fiber. 73 connector was used to mate this fiber with the photon counting electronics. The scintillator head and fiber junction were coated with black epoxy and covered with Dupont TefzelTM to shield light. Due to the stimulation energy needed to generate an RL signal using this material, it was not laboratory tested with the low activity gamma radiation sources but was tested clinically using the linear accelerator as noted above. Figure 3.21 shows the construction of the SiO2 : Cu2+ dosimeter cable. Clinical testing was done using the linear accelerator as noted above. FC Connector BCF−12 Scintillator BaS04 Reflective Paint Dupont Tefzel SH4001 Plastic Optical Fiber Jacket Cladding 0.98mm Core GH4001 Plastic Optical Fiber 1mm 3mm 2.5m Jacket Opaque Glue Dupont Tefzel FC Connector SiO2Cu2+ Scintillator Cladding 400um 400um Core SiO2 Optical Fiber 5mm 2.5m FIGURE 3.21: Schematic of BCF-12 and SiO2 : Cu2+ dosimeter cables. 74 3.6. BCF-12 Fiber Optic Probe Design A 1 mm diameter x 3 mm long round cylindrical BCF-12 scintillator was butt coupled using Norland UV cement and cyanoacrylate glue to a 2.5 m length of Mitsubishi Eska Premier GH4001 jacketed POF using an FC fiber connector to mate with the photon counting electronics. BCF-98 clear light guide was also evaluated but rejected due to permanent radiation damage and RL considerations within the fiber itself.[Wick and T. Zoufal, 2001, Nowotny, 2007] Each scintillator was covered with BaSO4 reflective paint, wrapped in Teflon tape, and further covered with Dupont TefzelTM to shield light.[Webber and Christ, 2003] POF and BCF-12 scintillator ends were prepared following the general procedures outlined in [Lee et al., 2004, Ayotte et al., 2006]. Figure 3.21 shows the construction of the BCF-12 fiber optic probe and Figure 3.22 shows the completed BCF-12 fiber optic probe with FC connector and scintillator end. FIGURE 3.22: Eska GH4001 POF and BCF-12 scintillator dosimeter cable. 75 Laboratory testing of absolute count rate was measured for each of the MPPC, Photonique, and Voxtel SSPMs with the BCF-12 scintillator stimulated using low activity radiation sources as noted above. Clinical testing was done using the linear accelerator as noted above with appropriate buildup material to approximate Charge Particle Equilibrium (CPE).[Attix, 1986] 3.6.1 Angular Measurements To minimize the effect of Cerenkov Radiation generated in POFs when using electron beams clinically, a 980 nm internal diamater x 30 cm length of cyclindrical silver coated (inside surface coating thickness 100 nm to 200 nm) air-core capillary tubing (TS1000 series, PolyMicro Technologies LLC, Scottsdale AZ.) was inserted between the BCF-12 scintillator and Mitsubishi Eska GH4001 POF (Norland UV cement, Polyacrylate glue).[Lambert et al., 2008] Figure 3.23 shows the completed cable. BaS04 Reflective Paint BCF−12 Scintillator Silver Capillary Tubing GH4001 Plastic Optical Fiber Jacket FC Connector Dupont Tefzel Cladding Air Core 1mm 3mm 0.98mm Core GH4001 Plastic Optical Fiber 0.5mm 10−20cm 2.5m FIGURE 3.23: BCF-12 Air-Core Capillary Tube Dosimeter. Clinical measurement of signal magnitude was recorded as a function of electron beam angle for BCF-12 and BCF-12 capillary tube probes. The measurement setup shown in Figure 3.24 was used. Two 25 cm x 25 cm styrofoam blocks spaced 30 cm apart positioned the dosimeter at the isocenter of the electron beam. The dosimeter was fixed in position between the blocks using Micropore tape and small cylindrical paper tubes to 76 keep scattered radiation to a minimum. A constant 4 cm x 20 cm radiation field size was used throughout the angular dependence measurements. Five readings each at dose rate 400 MU min−1 with 160 MU total dose delivered were integrated and averaged for each type of fiber optic probe, each with an electron beam (9 MeV) using angles from 0◦ to 100◦ in 10◦ steps. FIGURE 3.24: Photo of electron beam angular measurement setup. 77 3.7. Dose Linearity Measurements Dose linearity measures the ability of the OF probes to accurately determine dose between scaled monitor units (MUs) or activity from ionizing radiation sources. The signal ratio to MU from the linear accelerator should be a constant value for each probe tested, but will not be the same value between probes using different scintillators. For example, ideally each raw signal value should be 2x the previous value ±3% (20 MU = 2 x 10 MU) but also (4 x 5 MU), etc. The variation between the signal ratio within the same data set was analyzed using the slope of the linear best fit line. The slope of the best fit line is the average of the signal/MU ratio. The y-intercept is not zero because the accelerator is not linear when delivering dose at the low end of the range. Accelerators typically under-deliver in the low MU (< 4 MU) ranges which is why a lower limit of 5 MUs per segment is used for IMRT radiation fields. A typical clinical specification requires that the linearity results are between ±3%. Dose linearity measurements were done on top of an acrylic 20 cm x 30 cm x 5 cm phantom placed on top of the Clinac 2100CD patient table as shown in Figure 3.25. These were repeated for both types of BCF-12 probe, and the SiO2 : Cu2+ probe. The Prelude 420 crystal was tested at 6 MV. Different electron (6 MeV, 9 MeV, 12 MeV, 16 MeV, 20 MeV) and photon (6 MV, 18 MV) energies were used. Five readings each were integrated and averaged, then compared against the expected value over a dose range of 5 MU to 160 MU (400 MUmin−1 dose rate). 78 FIGURE 3.25: Photo of dose linearity measurement setup. 79 3.8. Percent Depth Dose Measurements Percent depth dose, equation 3.18, measures the dose deposited at a particular depth in water, normalized to the depth of maximum dose. PDD = Dd x100 Ddmax (3.18) The depth dose distribution (or profile) is a function of two parameters: type of radiation used and its energy in a radiotherapy treatment setting. Each of the depth dose figures shows the results for both types of BCF-12 probe and the SiO2 : Cu2+ probe measured in an acrylic water tank placed on top of the Clinac 2100CD patient table as shown in Figure 3.26. Different electron (9 MeV) and photon (6 MV, 18 MV) energies were used at 400 MUmin−1 dose rate, then compared against the expected value. Each percent depth dose data set was normalized to its own dmax depth: 1.5 cm for 6 MV photons, 3.0 cm for 18 MV photons, and 2.0 cm for 9 MeV electrons. Positional accuracy error for percent depth dose was ±2mm from the effective point of measurement in the water bath. A water phantom/tank from Med-Tec, Inc., model MT-DDA with a hand-cranked gear driven depth adjustment (in increments of 0.1 mm) was used. 80 FIGURE 3.26: Photo of electron beam percent depth dose measurement setup. 81 4. RESULTS AND DISCUSSION The overall SSPM signal shape is approximately a decaying exponential characterized by a time constant, τRt Ct , which is technology dependent (Rquenching , Rs , Cj , etc.) and also depends on other factors discussed in section 2.5.3. Signal rise time is fast, typically less than 2 ns for tested devices. Signal rise/fall times vary by manufacturer, number of pixel elements in the array, overall die capacitance, and external circuit elements. SSPM devices with die sizes ranging from 1 mm2 , 3 mm2 , 4 mm2 , 6 mm2 , and 9 mm2 exhibit increasing capacitance from 10 pF to hundreds of pF, with pixel element numbers ranging from 100 to nearly 10000.[Hamamatsu, 2007, Photonique, 2006, Voxtel, 2008, Zecotek, 2008, SensL, 2006] Figure 4.1 shows SPICE simulation results of signal shape vs. changing SSPM die capacitance in parallel with fixed load resistances simulating typical preamplifier or oscilloscope impedance’s. SSPM Signal Shape vs. Die Capacitance 0 2e-09 4e-09 6e-09 8e-09 2e-05 1e-08 2e-05 Ct = 0 pf Ct = 10 pf Ct = 20 pf Ct = 35 pf 1.5e-05 1.5e-05 Ct = 45 pf Ct = 100 pf Current (A) Ct = 300pf 1e-05 1e-05 5e-06 5e-06 0 0 2e-09 4e-09 6e-09 Time(s) 8e-09 0 1e-08 FIGURE 4.1: SSPM signal shape capacitance SPICE simulation. 82 An exponential current signal with fast rise and fall times is used to stimulate the passive network. What a load impedance actually sees is the SSPM die capacitance in parallel with two resistances: one the oscilloscope or preamplifier input impedance (50 Ω used in the simulation) and the second the SSPM external circuit load resistance, RL , which has a unique recommended value for each SSPM by manufacturer (50 Ω used in the simulation, see Figure 3.5). Results show that SSPM die capacitance significantly influences rise time and pulse height. Table 4.1 shows the manufacturer data for the SSPMs tested in this work; FF is the pixel fill factor in %, C is the device capacitance prior to breakdown, and gain (G) is quoted for the recommended operating voltage Vop . TABLE 4.1: Summary of published SSPM parameters. SSPM Area #cells Pixel size FF Idark C(pf) Vop (V) G PDE PH050701GR 1mm2 516 44µ2 60 2µA 35 40 0.8·106 30@600nm PH0701BG 1mm2 556 44µ2 60 10µA 40 20 0.4·106 40@560nm PH0611B1 1mm2 516 44µ2 60 10µA 40 30 0.18·106 20@440nm PH-CLNS 1mm2 516 44µ2 60 5µA 40 30 0.18·106 20@440nm PH0810G1 1.1mm2 556 44µ2 70 630nA 40 34.5 0.5·106 40@520nm MPPC 1mm2 400 50µ2 61.5 50nA 70 0.75·106 50@400nm SQBF-EIOA 1mm2 1024 32µ2 † 100nA 13 43.5 1.3·106 29@500nm SQBF-EKAA 1mm2 1024 32µ2 † 25nA 43.5 1.3·106 29@500nm †- not available. 35 10 83 4.1. Photonique SSPM Comparisons Figure 4.2 shows the measured pixel capacitance up to the breakdown voltage for each Photonique SSPM. Photonique SSPM Die Capacitance 75 70 050701GR 0701BG Die Capacitance (pF) 65 0611B1 CLNS 60 0810G1 55 50 45 40 35 0 5 10 15 20 25 30 35 Voltage (V) FIGURE 4.2: Die capacitance for five Photonique SSPMs. All Photonique DCR measurements were performed at 22.5◦ C. The following figures show DCRs for the Photonique: PH050701GR, PH0701BG, PH0611B1, PH-CLNS, and PH0810G1 Figure 4.3(a) shows measured DCRs while Figure 4.3(b) shows calculated DCRs based upon measured I/V curves and pixel capacitance. All Photonique gain measurements were done at 22.5◦ C. The following figures show the measured gain for the: PH050701GR (Figure 4.4(a)), PH0701BG (Figure 4.5(a)), PH0611B1 and CLNS (Figure 4.4(b)), and PH0810G1 (Figure 4.5(b)). Table 4.2 shows the PDE summary for the Photonique SSPMs. PDE increases with 84 SSPM Dark Count Rate (Log10 Hz) Measured Dark Count Rate: Photonique SSPMs 1E7 1000000 100000 PH050701GR PH0811B1 10000 PH0701BG PH0611B1 PHCLNS 1000 100 0 2 4 6 8 Overvoltage (V) (a) Measured DCR for Photonique SSPMs. Dark Count Rate Per Pixel (Log10 Hz) Photonique DCR vs. Overvoltage 100000 10000 PH0701BG 1000 PH0611B1 PH050701GR PH0810G1 100 0 1 2 3 4 5 6 Overvoltage (V) (b) Calculated DCR for Photonique SSPMs. FIGURE 4.3: Dark count rates for Photonique SSPMs. 85 Photonique 050701GR Gain vs. Bias 1.0 Equation y = a + b*x Weight No Weighting 0.01249 Residual Sum of Squares 0.98165 Adj. R-Square 0.8 Value Intercept Gain (10^6) B Slope Standard Error -7.38365 0.35478 0.1911 0.0087 0.6 0.4 0.2 PH050701GR Linear Fit of PH050701GR 0.0 39 40 41 42 43 Bias (V) (a) Gain dependence for 050701GR. Photonique CLNS, 0611B1: Gain vs. Bias 2.0 PH0611B1 1.8 PHCLNS Linear Fit of PH0611B1 1.6 Linear Fit of PHCLNS Gain (10^5) 1.4 1.2 1.0 0.8 0.6 Equation y = a + b*x Weight No Weighting 0.02685 Residual Sum of 0.04058 Squares 0.98987 Adj. R-Square 0.4 0.98439 Value Intercept 0611B1 Slope 0.2 Intercept CLNS Slope Standard Error -48.0508 1.5675 1.62891 0.05207 -47.32818 1.92725 1.60909 0.06402 0.0 29.6 29.8 30.0 30.2 30.4 30.6 Bias (V) (b) Gain dependence for 0611B1 and CLNS. FIGURE 4.4: Gain dependence for two Photonique SSPMs. 86 Photonique 0701BG Gain vs. Bias 4 3 PH0701BG Gain (10^5) Linear Fit of PH0701BG 2 1 Equation y = a + b*x Weight No Weighting 0.10306 Residual Sum of Squares 0.99083 Adj. R-Square Value Intercept B 0 18 Slope 19 Standard Error -15.8775 0.61225 0.92167 0.03133 20 21 Bias (V) (a) Gain dependence for 0701BG. PH0810G1 Gain vs. Bias 0.7 0.6 PH0810G1 Linear Fit of PH0810G1 Gain (10^6) 0.5 0.4 0.3 0.2 Equation y = a + b*x Weight No Weighting 0.00194 Residual Sum of 0.1 Squares 0.99481 Adj. R-Square Value Intercept 0.0 B 31 32 Slope 33 34 Standard Error -3.79365 0.09387 0.1226 0.0028 35 36 Bias (V) (b) Gain dependence for 0810G1. FIGURE 4.5: Gain dependence for two Photonique SSPMs. 87 bias voltage; here the bias voltage was fixed at manufacturer suggested Vop . TABLE 4.2: PDE measurements for Photonique SSPMs. Photon Detector 475nm (Blue LED) 640nm (Red LED) PH050701GR 13.1% 39.7% PH0701BG 25.4% 33.8% PH0611B1 20.2% 11.4% PH-CLNS 21.3% 12.1% PH0810G1 19.7% 39.0% ∗ - All biased at manufacturer Vop . Breakdown voltages (Vbreakdown ) and dark current (Idark ) ranges for Photonique SSPMs are shown in table 4.3, TABLE 4.3: Summary of measured breakdown voltages and device currents for Photonique SSPMs. Photon Detector Vbreakdown (V) Idark @Vbreakdown (A) Vbias (V) Idark @Vbias (A) PH050701GR 38.2 2.0 · 10−8 44.0 8.1 · 10−2 PH0701BG 16.5 2.2 · 10−9 23.0 3.2 · 10−5 PH0611B1 28.2 2.6 · 10−8 35.0 5.18 · 10−5 PH-CLNS 28.2 2.3 · 10−8 35.0 5.0 · 10−5 PH0810G1 27.7 2.2 · 10−9 40.0 4.32 · 10−5 Table 4.4 summarizes the measured Photonique SSPM characteristics. Gain was first measured then calculated using equation 2.26. τ ·DCR shows the expected number of dark counts added to a detected signal when using a τ = 1 µs sampling time. DCR was 88 first measured then calculated using equation 4.1.[Pavlov et al., 2005] The DCR for the PH050701GR is greater than measured by Pavlov05 but consistent in magnitude. DCR(Vov ) = I(Vov )dark /Vov Cpixel Npixel (4.1) To decrease the DCR contribution to any signal the sampling time τ must be shorter or device DCR must be smaller. When using fast scintillators such as BCF12 or Prelude 420 (table 3.3, 3.2 ns and 41 ns, respectively), τ can be shortened. This can be done primarily by lowering I(Vov )dark holding the other parameters constant. For the SSPMs tested in this study, I(Vov )dark was indeed decreased with each new generation of device. Vop (hence Vov ), Cpixel , and Npixel often change with each device improvement and topology.[McNally and Golovin, 2009] Based upon these test results the Photonique PH0810G1 was chosen for further testing in part due to its wider Vov range and smaller dark count rate. The PH0701BG had greater PDE in the blue/green light range, but a higher DCR and restricted Vov range. Note that the calculated gain based upon the measured pixel capacitance was consistently larger than measured when using the pulser. Table 4.5 shows the measured counts for the reference PMT and each Photonique SSPM type with BGO, Prelude 420, and BCF-12 scintillators directly attached and attached using 2.5 m Mitsubishi Eska GH4001 POF. 89 TABLE 4.4: Summary of measured Photonique SSPM characteristics. SSPM PH050701GR I(Vov )dark (A) Pixel Cap. (fF) Gain (105 ) 1.15 · 10−5 77.9 0.98 DCR τ ·DCR 15.9 kHz/pix 8.2 8.2 MHz† PH0701BG 6.35 · 10−7 78.2 0.89 9.74 kHz/pix 5.41 5.41 MHz† PH0611B1 9.0 · 10−6 75.6 0.95 14.24 kHz/pix 7.35 7.35 MHz† PH-CLNS 9.0 · 10−6 81.4 1.02 13.27 kHz/pix 6.83 6.83 MHz† PH0810G1 7.1 · 10−7 80.0 0.91 4.2 kHz/pixel 2.33 2.335 MHz† †- At Vov : 1.8V, 1.5V, 1.8V, 1.8V, 3.8V, respectively, vertically down DCR column. 90 TABLE 4.5: Photonique laboratory photon count rates†(cpm) 60 Co: direct scintillator- SSPM attachment and using POF. Photon Detector† BGO P450 BCF12 BGO-POF P450-POF BCF12-POF H5783PMT-ref 5713 3672 2378 732 1466 639 PH050701GR 683 1948 746 87 273 ∗ PH0701BG 1015 3572 1489 114 407 310 PH0611B1 830 3069 965 110 351 271 PH-CLNS 919 3021 1006 103 380 284 PH0810G1 927 3109 1127 92 385 302 ∗ - Undetectable †- At Vop : 41.5V, 19.5V, 31V, 31V, 31.5V, respectively, vertically down Photon Detector column. 91 4.2. Photonique, MPPC, and Voxtel SSPM Comparisons Based upon the measured results from the previous section, the Photonique PH0810G1 was compared against the MPPC and Voxtel SSPMs. Figure 4.6 shows the measured I-V characteristics for each SSPM and Figure 4.7 shows the measured pixel capacitance up to the breakdown voltage for each SSPM. I-V Comparison: Four Different SSPMs 1E-4 S10362-11-050C 1E-5 PH0810G1 SBQF-EKOA SBQF-EIOA -20 deg. C Current (A) 1E-6 1E-7 1E-8 1E-9 1E-10 1E-11 10 20 30 40 50 60 70 Bias (V) FIGURE 4.6: IV-characteristics for MPPC, Photonique, and Voxtel SSPMS. All DCR measurements were done at 22.5◦ C. The following figures show DCRs for the Photonique PH0810G1, Voxtel SQBF-EIOA/EKAA, and MPPC S10362-11-050C. Here Figure 4.8(a) shows measured DCRs while Figure 4.8(b) shows calculated DCRs based upon measured I/V curves and pixel capacitance. All gain measurements were done at 22.5◦ C. The following figures show the measured gain for the: Voxtel SQBF-EIOA/EKAA (Figure 4.9(a)), and MPPC (Figure 4.9(b)). 92 SSPM Die Capacitance vs. Voltage 70 60 Capacitance (pF) 50 40 30 SBQF-EIOA SBQF-EKOA -20 deg. C PH0810G1 20 S10362-11-050C 10 0 10 20 30 40 50 60 70 Voltage (V) FIGURE 4.7: Die capacitance for MPPC, Photonique, and Voxtel SSPMs. 93 Measured Dark Count rate: 4 SSPMs SSPM Dark Count Rate (Log10 Hz) 1E7 1000000 100000 10000 1000 PH0810G1 SBQF-EIOA SBQF-EKOA 100 10362-11-050C 10 0 2 4 6 8 10 12 Overvoltage (V) (a) Measured DCR for (4) compared SSPMs. Dark Count Rate vs. Overvoltage Dark Count Rate Per Pixel (Log10 Hz) 1E7 PH0810G1 S01362-11-050C 1000000 SBQF-EIOA SBQF-EKOA, -20 C 100000 10000 1000 100 10 0 2 4 6 8 10 Overvoltage (V) (b) Calculated DCRs for (4) compared SSPMs. FIGURE 4.8: Dark count rates for Photonique, Voxtel and MPPC SSPMs. 94 Voxtel SSPMs: Gain vs. Bias 0.8 SBQF-EIOA SBQF-EKOA 0.7 Linear Fit of EIOA Linear Fit of EKOA 0.6 Gain (10^6) 0.5 Equation y = a + b*x W eight No W eighting 0.00406 Residual Sum of Squares 0.98922 Adj. R-Square Value Intercept 0.4 C Standard Error -2.15541 0.07907 0.05993 0.00189 Slope 0.3 0.2 Equation y = a + b*x W eight No W eighting 0.00132 Residual Sum of Squares 0.1 0.99398 Adj. R-Square Value Intercept B Slope Standard Error -2.30566 0.0662 0.05969 0.00155 0.0 36 38 40 42 44 46 48 Bias (V) (a) Gain dependence for Voxtel SSPMs. MPPC S10362-11-050U: Gain vs. Bias 1.2 MPPC 1.0 Linear Fit of MPPC Gain (10^6) 0.8 0.6 0.4 Equation y = a + b*x Weight No Weighting Residual Sum 0.02531 of Squares Adj. R-Squar 0.2 0.97412 Value Intercept B Slope Standard Err -22.9830 1.35082 0.33733 0.01941 0.0 68.0 68.4 68.8 69.2 69.6 70.0 70.4 70.8 71.2 Bias (V) (b) Gain dependence for MPPC SSPM. FIGURE 4.9: Gain dependence for Voxtel and MPPC SSPMs. 95 Table 4.6 shows the PDE summary for the four SSPMs. PDE increases with bias voltage, here the bias voltage was fixed at manufacturer suggested Vop . TABLE 4.6: PDE measurements for PH0810G1, Voxtel, and MPPC SSPMs. Photon Detector∗ 475nm (Blue LED) 640nm (Red LED) PH0810G1 19.7% 39.0% S10362-11-050C 38.0% 18.4% SQBF-EIOA 18.1% 14.3% SQBF-EKAA 19.0% 15.6% ∗ All biased at manufacturers suggested Vop . Breakdown voltages (Vbreakdown ) and dark current (Idark ) ranges for the PH0810G1, S10362-11-050, SQBF-EIOA, and SQBF-EKAA are shown in table 4.7, TABLE 4.7: Summary of measured breakdown voltages and device currents. Photon Detector Vbreakdown (V) Idark @Vbreakdown (A) Vbias (V) Idark @Vbias (A) PH0810G1 27.7 2.2 · 10−9 40.0 4.32 · 10−5 S10362-11-050C 68.0 6.0 · 10−10 72.5 6.38 · 10−3 SQBF-EIOA 36.5 1.0 · 10−9 49.0 1.29 · 10−5 SQBF-EKAA 33.0 4.2 · 10−11 49.0 1.36 · 10−6 Table 4.8 summarizes the measured characteristics. Gain and DCR were calculated as described in section 4.1.. Table 4.9 shows the measured count rate (cpm) using the reference PMT (H5783) and each SSPM type with Prelude BGO, Prelude 420, and BCF-12 scintillators directly attached and attached using 2.5 m Mitsubishi Eska GH4001 POF. 96 TABLE 4.8: Summary of measured characteristics for Photonique, Hamamatsu, and Voxtel SSPMs. SSPM Npixel I(Vov )dark (A) Pixel Cap.(fF) Gain (105 ) PH0810G1 556 2.5 · 10−7 80.0 DCR 10.0 τ ·DCR 2585.4 Hz/pixel 1.44 1.44 MHz† S10362-11-050C 400 1.0 · 10−7 92.75 8.69 1796.9 Hz/pixel 0.71 718.7 kHz† SQBF-EIOA 1024 1.5 · 10−7 12.3 2.30 3969.8 Hz/pixel 4.06 7.2 MHz† SQBF-EKAA 1024 2.5 · 10−8 8.88 3.88 392.8 Hz/pixel ≈ 0.4 402.2 kHz† †- At Vov : 2V, 1.5V, 3V, 7V, respectively, verticlly down SSPM column. TABLE 4.9: Laboratory photon count rates†(cpm) 60 Co: direct scintillator-SSPM attachment and using POF. Photon Detector BGO P450 BCF12 BGO-POF P450-POF BCF12-POF H5783PMT-ref 5713 3672 2378 732 1466 1039 PH0810G1 927 3109 1127 92 385 302 S10362-11-050C 1082 3281 1682 182 526 413 SQBF-EIOA 358 2613 173 ∗ ∗ ∗ SQBF-EKAA ∗ 138 ∗ ∗ ∗ ∗ ∗ - Undetectable †- At Vop : 30.5V, 70V, 39.5V, 39.5V, respectively, vertically down photon detector column. 97 Laboratory measurements confirm that all three room temperature SSPM types may be used with inorganic scintillators coupled directly to their photosensitive die areas. These results suggest that uses for photon counting, PET, CT, diagnostic radiology, gamma spectroscopy functions, and remote environmental radiation monitoring are possible. The poor count rate measurement performance when using OFs, as with the Photonique SSPMs described earlier, has its roots in a number of areas. Chief among them is the size mismatch between the OF and the active optical area of the SSPM (approximately 1 mm2 for all). The H5783PMT reference PMT has a photocathode effective area that closely matched the BGO crystal, thus optical coupling was good. The Prelude 450 and BCF-12 scintillators, being 1 mm in diameter, did not couple well directly to the large photocathode area reference PMT when using the custom machined mating jigs. When using custom machined FC fiber connectors with the reference PMT the problem persisted as the GH4001 fiber cable has a core diameter of 1 mm. Prelude 450 and BCF-12 scintillators, each being approximately 1 mm in diameter, exhibited significant losses when coupled to the GH4001 fiber cable. To improve the S/N ratio industry standard fiber preparation and scintillator cleaving/polishing techniques must be used. In this work GH4001 fiber ends, BCF-12, and BC430 scintillators were cut with a razor blade beneath a microscope and hand polished. Figure 4.10 illustrates the types of junction defects for optical fibers and scintillators that result in signal loss due non-standard preparation techniques. BGO and Prelude 420 scintillators were supplied professionally polished by Saint-Gobain. Increased count rates due to the professional polishing were immediately apparent when coupling Prelude 420 and BGO scintillators directly to SSPMs without using GH4001 cable. BGO crystals being of larger diameter suffered the most performance loss when coupled to GH4001 cable, as no optical concentrator other than Teflon tape and plastic mating jigs were used (see Figure 5.1). Being larger, the BGO crystals actually absorb the most radiation from the low activity sources. 98 However due to GH4001 fiber coupling losses the count rates are smaller than Prelude 450 and BCF-12 fiber coupled probes. When testing SSPM/scintillator combinations in the laboratory using low activity gamma radiation sources, FC fiber connectors were observed to cause excessive signal loss. It was necessary to place the fiber core in direct contact with the SSPM photosensitive area, eliminating any air gap to maximize the acquired signal. The Voxtel (SQBF-EIOA) and Photonique (0810G1) use clear epoxy windows as a protective material covering their photosensitive areas. This material is hard yet scratchable when butting sharp crystal edges or OF cores against its surface, however shallow scratches may be polished out. The MPPC (S10362-11-050C) uses a clear gelatin-like polymer as a protective material covering its photosensitive area which easily scratches whereby sharp crystal edges or OF cores can penetrate to the photosensitive area, potentially damaging the die or breaking the bonding wires. When using the S10362-11-050C or its many newer variants, it is necessary to design a coupling solution that prevents sharp edges from crystals or OF cores from actually touching the material - adding to system cost. SSPM-scintillator coupling was an issue for the cooled TO-8 packaged Voxtel (SQBFEKAA) with its transparent glass window where the photo sensitive die is set back approximately 2.75 mm from the coupling surface. A custom lens design (GRIN or convex system) is needed to take full advantage of this cooled SSPM when used in large OF core diameter systems. Indeed even when using the SELFOC GRIN lens for 1 mm POF-SSPM coupling, it was not possible to transmit usable amounts of light to the photosensitive die for POF testing purposes during laboratory tests. Voxtel supplied a 62.5/125 GRIN fiber coupling jig for testing purposes. Laboratory tests with the LED pulser confirmed that the cooled SQBF-EKAA can be used successfully with a custom designed optical system, as the 62.5/125 GRIN improves the coupling substantially. However the 62.5/125 fiber coupler is suboptimal for used with 1 mm diameter fibers and large scintillators. A simple package redesign, which places the photosensitive die nearer (< 0.05 mm) 99 the OF core or scintillator crystal surface while eliminating the need for custom lenses (inert gas is a requirement to prevent condensation), will take advantage of the superior dark noise performance of the SQBF-EKAA making it an ideal candidate for low-noise (G,P)OF-SSPM coupled systems or direct scintillator-coupled (diagnostic, CT, PET, environmental) monitoring systems. When using the uncooled TO-8 packaged Voxtel (SQBF-EIOA) with 1 mm core POF centered directly over the photosensitive die (butted against the protective epoxy), the lower PDE of this device when compared with the MPPC (S10362-11-050C) and Photonique (0810G1) during laboratory tests (as shown by lower photon count rates), rendered it unusable for a (G,P)OF-SSPM clinical detection system. However, it can be used successfully with die sized bright scintillators coupled directly to the photosensitive area. Semiconductor device theory predicts[Sze, 2006] that thermally generated free carriers contributing to SSPM dark count noise are reduced by approximately a factor of 2 for every 8 ◦ C drop in temperature.[Renker and Lorenz, 2009] A similar dependency was noted for the SQBF-EKAA (see I/V curve Figure 4.6). This metric alone and others that depend upon temperature (I/V, Gain) are not sufficient by themselves to choose one SSPM over another for a given fiber-based application as noted above. The MPPC (S10362-11-050C) is a lower noise, higher gain device at small overvoltage (Vov ) [with better PDE at the scintillator emission frequencies tested here,] than the Photonique (0810G1) and the Voxtel SQBF-EIOA (Figure 4.6, table 4.8) at room temperature. However, the Photonique (0810G1) has greater dynamic latitude for overvoltage (Vov ) than the MPPC (S1036211-050C) which can be seen from the slope of their respective I/V curves in Figure 4.6. This characteristic is important, for example, when trading off noise against gain (bright scintillator light source, long fiber length, remote monitoring of isotopes) or signal acquisition without a preamplifier where high gain is a requirement. Test results indicated these two devices are closely matched. The room temperature Voxtel SQBF-EIOA, de- 100 spite a large dynamic overvoltage (Vov ) range, exibited lower gain with higher noise that the MPPC and Photonique counterparts, table 4.8). Optical signal coupling with 1 mm POF/scintillator combinations was poor for this device. Based upon these results (measured I/V, DCR, and Gain at Vov ), PDE at scintillator wavelength, and system analysis considerations (fiber coupling issues, and availability of portable electronics), the MPPC (S10362-11-050C) was chosen for the initial clinical portion of this work. 101 (a) Optical fiber junctions resulting in signal loss. (b) BCF-12 scintillator tip under 10x magnification: tapered end is 485.5 µm diameter. FIGURE 4.10: Optical fiber and plastic scintillator junction ends. 102 4.3. 4.3.1 Clinical Results Mitsubishi Eska GH4001 POF Cable Clinical Background Measurements Optical fibers exhibit luminescence while in photon orelectron beams.[Clift et al., 2002, Archambault et al., 2006] This is a noise source which must be subtracted from the measured count rate to obtain the true count rate. To gain information about the background counts generated in GH4001 POF cable, it was tested without an attached scintillator using the Clinac 2100, for both photon and electron beams (dose rate 400 MU/min−1 ). Figure 4.11(a) shows the background photon response measured for the GH4001 POF cable and Figure 4.11(b) shows the background electron response measured for the GH4001 POF cable at 0◦ gantry angle. Measurements confirm the literature reports that POFs generate optical noise when in photon beams as cited above. Electron beam noise is due in part to the generated Cerenkov radiation. 4.3.2 Angular Dependence Figure 4.12 shows the results for the electron beam (9 MeV) angular dependence tests. Using equation 2.14 the Cerenkov generated maximum peak at approximately 47.91◦ is clearly visible for the standard BCF-12 dosimeter, as is a 40% reduction when using the BCF-12 capillary tube POF dosimeter. The theoretical Cerenkov intensity curve is also plotted for comparison using equation 2.15. Earlier published work reports that the spectrum of light generated is independent of the angle between the beam and the fiber axis.[Lambert et al., 2009] Measured background counts were subtracted from the subsequent clinical measurements when using the GH4001 POF cable. 4.3.3 Dose Linearity An initial dose linearity test was done to check the clinical measurement setup using a Prelude 420 scintillation crystal attached to 2.5 m GH4001 POF. The result is shown 103 Eska GH4001 POF: Photon Response 18000 16000 Equation y = a + b*x Weight Equation No Weighting y = a + b*x 26269.84826 No Weighting Residual Sum of Weight Squares 14000 Residual Sum of Adj. R-Square Squares Response (Counts) Intercept 12000 Slope Intercept 6xbg 18xbg 0.99971 Value 0.98072 Adj. R-Square 6xbg 47889.61487 26269.84826 0.98072 Intercept Slope Slope Standard Error 147.53731 46.08477 Value Standard Error 9.76453 147.53731 0.61108 46.08477 -39.57164 9.76453 62.22277 0.61108 107.65089 0.82507 10000 8000 6MV 18MV 6000 Linear Fit 6MV Linear Fit 18MV 4000 2000 0 0 20 40 60 80 100 120 140 160 Dose (MU) (a) GH4001 POF photon background response. Eska GH4001 POF: Electron Response 10000 6 MeV, Linear Fit of 6 MeV 9 MeV, Linear Fit of 9 MeV 12 MeV, Linear Fit of 12 MeV 8000 18 MeV, Linear Fit of 18 MeV Response (Counts) 20 MeV, Linear Fit of 20 MeV 6000 4000 2000 0 0 20 40 60 80 100 120 140 160 Dose (MU) (b) GH4001 POF electron background response. FIGURE 4.11: GH4001 POF cable background response. 104 BCF-12 Dosimeter Response vs. Angle 1.0 Theoretical BCF12 BCF12 Cap Tube Normalized Response 0.8 0.6 0.4 0.2 0.0 0 20 40 60 80 100 Angle between beam axis and POF axis (degrees) FIGURE 4.12: Angular dependence of standard and capillary tube POF dosimeters: BCF12 scintillator: measured and theoretical results 105 in figure 4.13. 2100CD Clinac 6MV Photon Dose Linearity: Prelude 420 Crystal 25000 6MV Linear Fit of 6MV Response (Counts) 20000 15000 10000 Equation y = a + b*x Weight No Weighting 375069.7299 Residual Sum 2 of Squares 5000 0.99893 Adj. R-Square Value 6x 0 0 20 40 60 80 Standard Erro Intercept 14.73134 174.13444 Slope 157.4575 2.309 100 120 140 160 180 Dose (MU) FIGURE 4.13: 6 MV photon dose linearity: Prelude 420 scintillator. The 6 MV ratio and linear fit analysis to data points at doses of (5, 10, 20, 40, 80, & 160 MUs) gave percent errors of (-34.0, -16.2, 2.10057, 5.2, 2.1, -0.9)%, respectively. Note the large percent error at the 5 MU and 10 MU dose ranges, indicating a large under-response (hence nonlinearity) in this range. The linear dose responses of the OF dosimeters in the RL scintillating mode are shown in the following figures. Figure 4.14(a)shows the measured photon dose linearity for the standard POF dosimeter and GOF dosimeter: BCF-12, and SiO2 : Cu2+ scintillators at 6 MV and 18 MV. Figure 4.14(b) shows the measured photon dose linearity for the capillary tube POF dosimeter: BCF-12 scintillator at 6 MV and 18 MV. Figure 4.15 shows the measured electron dose linearity (9 MeV - 20 MeV) for the standard POF dosimeter, GOF dosimeter: BCF-12, and SiO2 : Cu2+ scintillators. 106 2100CD Clinac Photon Dose Linearity 400000 6MV, 350000 Linear Fit of 18MV 6MV SiO2:Cu2+, 300000 Response (Counts) Linear Fit of 6MV 18MV, Linear fit of 6MV 250000 200000 150000 100000 50000 0 0 20 40 60 80 100 120 140 160 Dose (MU) (a) BCF-12 and SiO2 : Cu2+ Dosimeters 2100CD Clinac Photon Dose Linearity 120000 6MV Cap Tube, 18MV Cap Tube, Linear Fit of 6MV Linear Fit of 18MV Response (Counts) 100000 80000 60000 40000 20000 0 0 20 40 60 80 100 120 140 160 Dose (MU) (b) BCF-12 Capillary Tube Dosimeter FIGURE 4.14: Photon dose linearity: standard and capillary tube dosimeters 107 2100CD Clinac Electron Dose Linearity 300000 250000 6MeV, Linear Fit of 6MeV 9MeV, Linear Fit of 9MeV 12MeV, Linear Fit of 12MeV 16MeV, Linear Fit of 16MeV 20MeV, Linear Fit of 20MeV Response (Counts) 9MeV SiO2:Cu2+, Linear Fit of 9MeV 200000 150000 100000 50000 0 0 20 40 60 80 100 120 140 160 Dose (MU) FIGURE 4.15: Electron dose linearity: BCF-12 and SiO2 : Cu2+ Dosimeters. Figure 4.16(a) shows the measured electron dose linearity for the capillary tube POF dosimeter (6 MeV, 9 MeV): BCF-12 scintillator. Figure 4.16(b) shows the measured electron dose linearity for the capillary tube POF dosimeter (12 MeV, 16 MeV, 20 MeV): BCF-12 scintillator. The figures are non-linear in dose ranges of interest for radiation oncology treatments. Table 4.10 compares the measured photon and electron dose linearity for the GOF dosimeter, standard POF dosimeter, and capillary tube dosimeters: BCF-12, and SiO2 : Cu2+ scintillators. 4.3.4 Depth Dose Measurements: Photon and Electron Beam Section 3.8. describes depth dose as measured in this work. Each of the depth dose figures shows the results for the OF dosimeters measured in a water tank, using different electron and photon energies, and comparing against the reference ion chamber. Figure 4.17 shows the result of the 6 MV photon depth dose measurements for the 108 2100CD Clinac Electron Dose Linearity: Capillary Tube 180000 160000 Response (Counts) 140000 6MeV, Linear Fit of 6MeV 9MeV, Linear Fit of 9MeV 120000 100000 80000 60000 40000 20000 0 0 100 200 300 400 500 Dose (MU) (a) BCF-12 Capillary Tube Dosimeter: 6 MeV, 9 MeV. 2100CD Clinac Electron Dose Linearity: Capillary Tube 16000 14000 Response (Counts) 12000 12MeV, Linear Fit of 12MeV 16MeV, Linear Fit of 16MeV 20MeV, Linear Fit of 20MeV 10000 8000 6000 4000 2000 0 0 20 40 60 80 100 120 140 160 Dose (MU) (b) BCF-12 Capillary Tube Dosimeter: 12 MeV, 16 MeV, 20 MeV. FIGURE 4.16: Electron dose linearity: capillary tube dosimeters. 109 TABLE 4.10: Photon and Electron Dose Linearity Differences (range extrema) from Reference Ion Chamber (in %). Dosimeter 6MV SiO2 : Cu2+ -5.5 18MV † 6MeV † +0.5 BCF-12 BCF-12 Cap. Tube 9MeV -4.4 12MeV 16MeV 20MeV † † † +0.2 +9.0 -3.0 -0.8 -16.0 -6.5 -5.1 -6.2 +0.6 +6.4 +7.1 +8.2 +0.8 +9.4 +5.0 -21.1 -19.2 -17.8 -15.7 -20.3 -15.4 -18.2 -0.8 +22.4 +1.6 +7.1 +14.0 +8.0 +6.6 †- Not tested. standard POF dosimeter, using both BCF-12 and SiO2 : Cu2+ scintillators (always used with GOF) as well as the capillary tube dosimeter using a BCF-12 scintillator. Figure 4.18 shows the result of the 18 MV photon depth dose measurements for the standard and capillary tube dosimeters: BCF-12 and SiO2 : Cu2+ scintillators. Figure 4.19 shows the result of the 9 MeV electron depth dose measurements for the standard BCF-12 POF dosimeter, the GOF SiO2 : Cu2+ dosimeter and the BCF-12 capillary tube dosimeter. Overall the figures show the agreement is good for the SiO2 : Cu2+ and standard BCF-12 dosimeters; however, in a clinical setting no more than 3%-5% error is permitted. The capillary tube dosimeter measured a large error, 21%. The errors are too large for actual clinical measurements using the current laboratory manufactured OF dosimeters and further characterization is needed. Table 4.11 compares the measured photon and electron depth doses for the GOF dosimeter, standard POF dosimeter, and capillary tube dosimeters: BCF-12, and SiO2 : Cu2+ scintillators. However, the accuracy is certainly closeenough to use these small, all solid state dosimeters for real-time detection of 110 Relative Response 2100CD Clinac - Reference % Depth Dose: 6 MV Photons 100 IonChamber 90 SiO2:Cu2+ 80 BCF12 70 BCF12 Cap Tube 60 50 40 30 0 5 10 15 20 % Difference from Reference Ion Chamber: 6 MV Photons 0 SiO2:Cu2+ Residuals (%) -2 BCF12 -4 BCF12 Cap Tube -6 -8 -10 -12 0 5 10 15 20 Depth (cm) FIGURE 4.17: 10 cm x 10 cm 6 MV photon depth dose profile in water tank equipment malfunctioning. The equipment could be shutdown before injuring the patient. 4.4. Summary of Clinical Results Angular measurements with the capillary-tube-coupled BCF-12 scintillator in an electron beam show a Cerenkov signal reduction of 40% at the measured peak angle of 50◦ , while not completely eliminating the Cerenkov noise. Due to a scarcity of material to meet the clinical testing schedule, the length of air-core capillary tube used for this measurement was not of sufficient length to completely avoid Cerenkov effects (as described in [Lambert et al., 2008]) from the beam. This greater variation is likely due to three factors: the shorter length of air-core capillary tubing used during the angular measurements, the direction of the beam incident to the POF (refer to Figure 3.24) positive or negative 111 2100CD Clinac - Reference % Depth Dose: 18 MV Photons Relative Response 100 90 80 70 IonChamber 60 SiO2:Cu2+ 50 BCF12 40 BCF12 Cap Tube 30 20 0 5 10 15 20 % Difference from Reference Ion Chamber: 18 MV Photons 15 Residuals (%) SiO2:Cu2+ 10 BCF12 BCF12 Cap Tube 5 0 -5 -10 -15 0 5 10 15 20 Depth (cm) FIGURE 4.18: 10 cm x 10 cm 18 MV photon depth dose profile in water tank gantry angle), and effects from the 2100CD Clinac generated noise. As the 2100CD Clinac rotates to extreme angles more of the POF is in the beam, coupling issues between the POF-air-core capillary tubing-scintillator junctions, and manufacturing differences between the dosimeter cables are significant. There remains a measurable Cerenkov effect when irradiating the BCF-12 capillary tube POF dosimeter. The dose ranges discussed below and shown in the tables are range extremas (extremes of the measured results). When using SSPMs-OFs as a photon detector, variation in dose linearity for the SiO2 : Cu2+ (6 MV only) and standard BCF-12 dosimeters (6 MV, 18 MV) confirms previous studies on the accuracy of OF dosimetry.[Beddar, 1994, Beddar et al., 2001, Huston et al., 2001] Here the variation in dose linearity ranged from −5.5, +0.5% for SiO2 : Cu2+ dosimeters and ranged from +9.0, +0.6%, -3.0, +6.4% for BCF-12 dosimeters respectively, when compared against the expected values from the ion 112 Relative Response 2100CD Clinac - Reference % Depth Dose: 9 MeV Electrons 100 80 60 IonChamber 40 SiO2:Cu2+ BCF12 20 BCF12 Cap Tube 0 0 1 2 3 4 5 % Difference from Reference Ion Chamber: 9 MeV Electrons 10 Residuals (%) 5 0 -5 -10 SiO2:Cu2+ -15 BCF12 -20 BCF12 Cap Tube -25 0 1 2 3 4 5 Depth (cm) FIGURE 4.19: 10 cm x 10 cm 9 MeV electron depth dose profile in water tank chamber. For the BCF-12 air-core capillary tube dosimeter (6 MV, 18 MV), the variation in dose linearity ranged from -21.2, -0.8% and -19.2, +22.4% when compared against the expected values from the ion chamber. When using SSPM-OFs as electron detectors variations in dose linearity for the SiO2 : Cu2+ dosimeter (9 MeV only) ranged from -4.4, -0.2% and for the standard BCF12 dosimeter (6 MeV, 9 MeV, 12 MeV, 16 MeV, 20 MeV) ranged from -0.8, +7.1%, -16.0, +8.2%, -6.5, +0.8%, -5.1, +9.4%, and -6.2, +5.0%, respectively, when compared against the expected values from the ion chamber. For the BCF-12 air-core capillary tube dosimeter, the variation in dose linearity ranged from -17.8, +1.6%, -15.7, +7.1%, -20.3, +14.0%, -15.4, +8.0%, and -18.2, +6.6% when compared against the expected values from the ion chamber. Depth dose range discussed below and shown in the tables are range extrema. When 113 TABLE 4.11: Photon and Electron Percent Depth Dose Differences (range extrema) from Reference Ion Chamber (in %). Dosimeter 6MV 18MV 9MeV SiO2 : Cu2+ -7.4 -12.9 -7.3 -1.2 +1.8 +4.9 -9.1 -1.8 -11.3 -1.3 +7.4 +6.4 -11.0 -2.7 -21.2 -1.4 +15.3 +3.9 BCF-12 BCF-12 Capillary Tube using SSPMs-OFs as a photon detector for depth dose (6 MV, 18 MV), variation in accuracy for SiO2 : Cu2+ , BCF-12, and BCF-12 capillary tube dosimeters ranged from: -7.4, -1.2%, -9.1, -1.3%, -11.0, -1.4% and -12.9, +1.8%, -1.8, +7.4%, -2.7, +15.3% respectively, when compared against the expected values from the ion chamber. When using SSPMs-OFs as an electron detector for depth dose (9 MeV), variation in accuracy for SiO2 : Cu2+ , BCF-12, and BCF-12 capillary tube dosimeters ranged from: -7.3, +4.9%, -11.3, +6.4%, and -21.2, +3.9% respectively, when compared against the expected values from the ion chamber. The SiO2 : Cu2+ dosimeter had the most accurate response in this study when compared against the expected values from the ion chamber. Both BCF-12 dosimeter depth dose measurements were consistently below the reference ion chamber measurements, indicating a S/N ratio, coupling, or scintillator efficiency issue when compared to the SiO2 : Cu2+ dosimeter. 114 4.5. Uncertainty Analysis SiO2 : Cu2+ and BCF-12 material dose linearities have been well characterized [Huston et al., 2001, Justus et al., 2004, 2006, Beierholm et al., 2008] while a capillary tube dosimeter using a PMT optical photon sensor showed much better linearity than measured here in a previous study using a 60 cm capillary tube.[Lambert et al., 2008] In an SSPM-POF coupled radiation sensing system sources of signal degradation are numerous: three optical junctions in a standard cable (scintillator-POF, POF-FC, FC-SSPM), four optical junctions in a capillary tube cable (scintillator-capillary tube, capillary tube-POF, POF-FC, FC-SSPM). Each of these junctions has associated with it a signal coupling attenuation factor, which when multiplied together greatly reduce the number of optical photons reaching the SSPM.[Beddar, 2007] In this work scintillators of varying sizes (1 mm square, 1 mm diameter circular, and 4 mm diameter circular) were used to gain insight into coupling onto 1 mm2 die area SSPMs and 1 mm diameter POFs for remote monitoring applications. Photon counting results indicate that significant light is lost when using a larger diameter scintillator without an optical concentrator system. Using a smaller scintillator (smaller than die area) introduces excess noise into the system due to DCR considerations. ESKA GH4001 plastic fiber cable attenuates light at 0.19 db/m (650 nm) and 0.22 db/m (400 nm). For example, at 15 m this is 2.85 db with a signal transmission factor of 0.518 (2.85db = −10logT , see Appendix B). For remote radiation monitoring applications even this short POF length degrades the signal substantially (fewer optical photons reaching SSPM), underscoring the need for a low DCR SSPM with high gain and high PDE at scintillator emission wavelengths. Similarly, in clinical applications it is desirable to have electronics physically separated from noise sources (removed from Linac vaults, x-ray generator rooms, etc.). At long POF cable lengths needed to achieve this separation, signal attenuation is significant; multiple junctions further degrade the S/N ratio. 115 Multiple junction POF systems necessitate that precision manufacturing techniques be used on each component to avoid optical photon attenuation and loss. Current state of the art SSPMs have greater DCRs than PMTs (200 nA vs. 2 nA).[Hamamatsu, 2008, 2007, Zecotek, 2008, SensL, 2006] Replacing PMTs in radiation monitoring applications using POFs as signal conduits requires that the system S/N ratio budget be carefully considered. For direct coupled, low-level radiation detection systems, the size of the SSPM die and scintillator diameter must be closely matched or an optical concentrator system must be used. Scintillator end faces must be carefully polished to optical industry standards, and index matching gel should be used at the interfaceto the SSPM. Finally, the PDE of the SSPM must be high at the emission wavelength of the scintillator. Once POF is introduced into the system the importance of these constraints is magnified as noted above. During laboratory tests, length extensions of POF beyond 2.5 m with low PDE SSPMs resulted in undetectable signals when using low activity radiation sources. Glass optical fiber which is less lossy but fragile, (400 µm or 500 µm diameter) was not invistigated in this work. 116 5. 5.1. CONCLUSIONS AND RECOMMENDATION FOR FUTURE WORK Optical and Electrical Characterization Use of SSPMs as photon detectors for laboratory, environmental, diagnostic radiation sensing and dosimetry applications have many advantages over PMTs: SSPMs are rugged, small size, have high gain, operate at low bias voltages, are non-magnetically sensitive, have high quantum efficiency, and (potentially) much lower cost. This work has shown that, based upon their performance as radiation sensors and optical fiber dosimeters, SSPMs can be used to replace PMTs for laboratory, clinical, and environmental applications. The MPPC S10362-11-050C and Photonique 0810G1 had the best laboratory and clinical performance of the SSPMs tested. Their gain measurements and photon counts with scintillators directly attached indicate that these two devices are somewhat closely matched in the low Vov range, with the Photonique 0810G1 exhibiting an advantage with a much wider Vov range (Figure 4.6). The MPPC S10362-11-050C has a measurable edge in dark current (6.0 · 10−10 A at Vbreakdown vs. the Photonique 0810G1 (2.2 · 10−9 A), and SQBF-EIOA (1.0 · 10−9 A) respectively (Table 4.7). The two Voxtel devices ( SQBFEIOA/SQBF-EKAA) and the Photonique 0810G1 exhibited much wider usable Vov ranges than the MPPC S10362-11-050C, allowing the user greater flexibility to trade off gain vs. noise at a given Vov . However, the MPPC S10362-11-050C exhibited a higher PDE at the scintillator wavelengths used in this work (Table 4.6). Since the same techniques were used to attach scintillators to optical fibers for use with the SSPMs, the performance degradation relative to each device was observed to be similar. Much time was spent testing the Voxtel SQBF-EIOA (room temperature) and SQBF-EKAA (cooled) devices in the laboratory in an effort to approximately match the 117 performance of the MPPC and Photonique SSPMs. Thirty engineering samples of the SQBF-EIOA were tested for minimal dark current, with the four best chosen for further laboratory work. The SQBF-EKAA (cooled, −20◦ C) SSPM had the lowest dark current of any device tested, 4.2 · 10−11 A at Vbreakdown (Table 4.7). This result is expected; in semiconductors much of the dark current is thermally generated. On the test bench the photon counts for the SQBF-EIOA were lower than MPPC and Photonique SSPMs (Table 4.9). Further discussion with Voxtel disclosed that PDE was not directly measured for the sample devices - PDE was measured by an outside vendor at the initial design and fabrication stage. Although PDE was also not measured through the 380 nm - 650 nm range in the laboratory for MPPC and Photonique SSPMs, photon counting results indicate that the Voxtel device suffered from a significantly lower PDE at the scintillation wavelengths of interest (420 nm - 450 nm) when compared with its Hamamatsu and Photonique counterparts. When used with the 580 nm BC-430 plastic scintillator the SQBF-EIOA demonstrated significantly larger photon counts, indicating that the decreased PDE for this device was a significant factor in laboratory tests. The SQBF-EKAA (cooled) SSPM shows much promise as a device with lowered dark pulses when compared with the MPPC and Photonique SSPMs as shown in Figure 4.6. Due to its current package design, robust optical coupling was not possible with off-theshelf GRIN lenses. Indeed Voxtel had an optical manufacturer fabricate a custom GRIN lens for use with 62.5/125 diameter optical fibers (for DNA testing) coupled to their Peltier-cooled T0-8 package. In this work budgetary and time constraints prohibited the design of a custom GRIN lens for 1mm diameter optical fibers. The introduction of metalized air-core capillary tubing in the optical signal path essentially functions as an attenuator and Cerenkov noise reducer during radiotherapy applications. Under clinical conditions the SSPM-air-core capillary tube-OF dosimeter provided a noisy, fluctuating signal for analysis. Angular measurements confirm the value of the air-core capillary tube for reducing Cerenkov radiation generated within the OFs. 118 Additional lengths of this material (up to 1 m) can completely eliminate measurable Cerenkov noise generated in the OFs, whilst providing a simpler noise reduction methodology when compared with other noise reduction schemes discussed previously.[Lee et al., 2007b, Frelin et al., 2006] Depth dose profiles can be measured by the SiO2 : Cu2+ dosimeter, standard BCF-12 dosimeter, and BCF-12 metalized air-core capillary tube dosimeter for electon and photon beams. The standard SSPM-OF dosimeters exhibit linear response characteristics to dose and energy independence when using both plastic and SiO2 : Cu2+ materials, though not within the commonly accepted +-3% error tolerance needed for clinical measurements. The SSPM-air-core capillary tube-OF dosimeter while suffering from noise and accuracy problems, can be improved with better manufacturing processes to approach the performance of the standard SSPM-OF dosimeters. The S/N ratio of the non-capillary tube SSPM-OF system is adequate for both clinical dosimetry and short length environmental radiation detection. 5.2. Recommendations for Future Work The two apparent limitations of the Voxtel SQBF-EKAA (cooled) device, PDE and optical coupling, can be addressed by device testing/redesign, and an improved compact package (modified T0-8 with inert gas atmosphere to prevent H2 O condensation) that decreases the separation of the photosensitive die area for coupled light sources to distances < 0.1 mm, eliminating the need for a custom GRIN lens. A thin optical coating similar to that used on the MPPC and Photonique SSPMs is adequate for this purpose. Laboratory photon-counting electronics used in this work can be improved by creating a portable USB based system that accommodates SSPMs with different working voltages and gain requirements. The power supply and amplification stages for a portable device were presented in this work. The USB interface and software blocks are currently 119 missing from a realization. Use of a light guide when attaching large scintillators, large diameter optical fibers, or fiber bundles to small SSPMs facilitates light concentration onto the small die area. Larger (2 mm x 2 mm and 3 mm x 3 mm) SSPMs are available. However, at present, these have significantly higher DCRs.[Hamamatsu, 2007, Photonique, 2006, Zecotek, 2008, SensL, 2006] Figure 5.1 shows a proposed light guide schematic drawing and ray trace diagram. Importantly, calibrated low activity radiation source testing is needed to determine accuracy for environmental radiation measurements when using SSPMs and POFs. Testing of the fiber optic dosimeters and direct scintillator coupled SSPMs with low energy diagnostic radiation sources (x-ray generators, mammography, CT scanners), where the number of daily procedures is much greater than clinical radiotherapy, is needed for system validation. Together with the low energy clinical testing regime, an x-ray generator and mCi activity radiation sources are required for laboratory testing purposes. Dose rate tests, both in the laboratory and in the clinic are needed to assess system reproducibility. CT scans in particular are experiencing explosive growth worldwide. The ability to easily record diagnostic radiation dose in patient charts is likely to be become standard practice in the future. The low energies used in clinical diagnostics render the need for Cerenkov noise cancellation moot; for these applications no capillary tube noise reduction is necessary. In addition, monte-carlo simulation studies using various simulators (Slitrani, GEANT4, EGS4, MCNP) can be used to validate the scintillation light generated by different radiation fields and in external light/environment shielding materials. In particular, studies to determine the effect of light shielding materials, light photon propagation, capillary tube signal loss, and fiber junction signal loss characteristics are needed. Though no long-term environmental or clinical study of the SSPM-OF system has yet been performed, the first results presented here are encouraging. For future prototypes the air-core capillary tube-dosimeter-OF coupling must be optimized by cutting/polishing 120 the OFs, scintillators, and capillary tubes with tighter manufacturing tolerances to improve performance and reduce signal variability. Moreover, computer simulation of the optical photon signal path will assist in identifying coupling problems and in improving the S/N ratio. Further laboratory and clinical studies with new OF-SSPM prototypes, using the Photonique (0810G1) and Voxtel (SBQF-EKAA) with modified package (outside the LINAC vault, with and without a preamplifier), will include low energy (< 300 keV) dose and dose rate characterization, long-term measurements, ion chamber OF dosimeter calibration measurements, and improved software to operate the system. 121 (Large diameter scintillator or fiber bundle) Plastic lens material SSPM die cover material o 0.9mm SSPM die approx. 4mm 10 111111 000000 000000 111111 SSPM case (a) Light guide schematic drawing. (b) Light guide ray trace diagram. FIGURE 5.1: Proposed light guide for small area SSPMs. 122 BIBLIOGRAPHY M.S. Akselrod, L. Botter-Jenson, and S.W.S. McKeever. Optically stimulated luminescence and its use in medical dosimetry. Radiation Measurements, 41:S78–S99, 2007. C.E. Anderson, Nielsen S.K., S. Greilich, J. Helt-Hansen, J.C. Lindegaard, and K. Tanderup. Characterization of a fiber-coupled Al2 O3 : C luminescence dosimetry system for online in vivo dose verification during 192 Ir brachytherapy. Med. Phys., 36 (3):708–718, 2009. L. Archambault, J. Arsenault, A.S. Beddar, R. Roy, and L. Beaulieu. Plastic scintillation dosimetry: Optimal selection of scintillating fibers and scintillators. Med. Phys., 32(7): 2271–2278, 2005. L. Archambault, A.S. Beddar, L. Gingras, R. Roy, and L. Beaulieu. Measurement accuracy and cerenkov removal for high performance, high spatial resolution scintillation dosimetry. Med. Phys., 33(1):128–135, 2006. F.H. Attix. Introduction to Radiological Physics and Radiation Dosimetry. Wiley- Interscience, Berlin, 1986. G. Ayotte, L. Archambault, L. Gingras, F. Lacroix, A.S. Beddar, and L. Beaulieu. Surface preparation and coupling in plastic scintillator dosimetry. Med. Phys., 33(9):3519–3525, 2006. A. Bar-Lev. Semiconductors and Electronic Devices, 3rd ed. Prentice Hall, New York, 1993. G. Bartesaghi, V. Conti, D. Bolognini, S. Grigioni, V. Mascagna, M. Prest, S. Scazzi, A. Mozzanica, P. Cappelletti, M. Frigerio, S. Gelosa, A. Monti, A. Ostinelli, G. Giannini, and E. Vallazza. A scintillating fiber dosimeter for radiotherapy. Nuc. Instr. Meth. Phys. Res. A., 581(1-2):80–83, 2007. 123 P. Barton, C. Stapels, E. Johnson, J. Christian, W.W. Moses, M. Janecek, and D. Wehe. Effect of sspm surface coating on light collection efficiency and optical crosstalk for scintillation detection. Nuc. Instr. Meth. Phys. Res. A, 610:393–396, 2009. H.N. Becker and A.L. Johnston. Dark current degradation of near infrared avalanche photodiodes from proton irradiation. IEEE Trans. Nuc. Sci., 51(6-2):3572–3578, 2004. A.S. Beddar. A new scintillator detector system for the quality assurance of 60 Co and high-energy therapy machines. Phys. Med. Biol., 39:253–263, 1994. A.S. Beddar. Plastic scintillation dosimetry and its application to radiotherapy. Radiation Measurements, 41:S124–S133, 2007. A.S. Beddar, T.R. Mackie, and F.H. Attix. Cerenkov light generated in optical fibers and other light pipes irradiated by electron beams. Phys. Med. Biol., 37(4):925–935, 1992a. A.S. Beddar, T.R. Mackie, and F.H. Attix. Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. physical characteristics and theoretical considerations. Phys. Med. Biol., 37(10):1883–1900, 1992b. A.S. Beddar, T.J. Kinsella, A. Ikhlef, and H. Sibata. A minature “scintillator-fiberopticpmt” detector system for the dosimetry of small fields in sterotactic radiosurgery. IEEE Trans. Nuc. Sci., 48(3):924–928, 2001. A.S. Beddar, N. Suchowerska, and S.H. Law. Plastic scintillation dosimetry for radiation therapy: minimizing capture of cerenkov radiation noise. Phys. Med. Biol., 49:783–790, 2004. A.S. Beddar, T.M. Briere, F.A. Mourtada, O.N. Vassiliev, H.H. Liu, and R. Mohan. Monte carlo calculations of the absorbed dose and energy dependence of plastic scintillators. Medical Physics, 32(5):1265–1269, 2005. 124 A.R. Beierholm, C.E. Anderson, L.R. Lindvold, F. Kjaer-Kristoffersen, and J. Medlin. A comparison of BCF-12 organic scintillators and Al2 O3 : C crystals for real-time medical dosimetry. Radiation Measurements, 43:898–903, 2008. L.A. Benevides, A.L Huston, B.L. Justus, P. Falkenstein, L.F. Brateman, and D.E. Hintenlang. Characterization of a fiber-optic radioluminescent detector for application in the mammography energy range. Medical Physics, 34(6):2220–2227, 2007. F. Berghmans, B. Brichard, and S. Girard. An Introduction to Radiation Effects on Optical Components And Fiber Optic Sensors. NATO Science for Peace and Security Series B: Physics and Biophysics, Springer, Netherlands, 2007. Z. Bielecki. Analysis of operation conditions of avalanche photodiode on signal to noise ratio. Opto-Electr. Rev., 5(4):127–165, 1997. P. Buzhan, B. Dolgoshein, L. Filatov, A. Ilyin, V. Kaplin, A. Karakash, S. Klemin, R. Mirzoyan, A.N. Otte, E. Popova, V. Sosnovtsev, and M. Teshima. Large area silicon photomultipliers: Performance and applications. Nuc. Instr. Meth. Phys. Res. A., 567(1): 78–82, 2006. H. Cember. Introduction to Health Physics, 3rd ed. McGraw-Hill Medical, New York, 1998. M.A. Clift, R.A. Sutton, and D.V. Webb. Water equivalence of plastic organic scintillators in magavoltage radiotherapy bremsstrahlung beams. Phys. Med. Biol., 45:1885–1895, 2000. M.A. Clift, P.N. Johnston, and D.V. Webb. A temporal method of avoiding Cerenkov radiation generated in organic scintillator dosimeters by pulsed mega-voltage electron and photon beams. Phys. Med. Biol., 47:1421–1433, 2002. F. Corsi, A. Dragone, C. Marzocca, A. Del Guerra, P. Delizia, N. Dinu, C. Piemonte, 125 M. Boscardin, and G.F. Dalla Betta. Modelling a silicon photomultiplier (SiPM) as a sognal source for optimum front-end design. Nuc. Instr. Meth. Phys. Res. A., 572: 416–418, 2006. V.H. Dhulla, G. Gudkov, D. Gavrilov, A. Stepukhovich, A. Tsupryk, O. Kosobokova, A. Borodin, B. Gorbovitski, and V. Gorfinkel. Single Photon Counting Module Based on Large Area APD and Novel Logic Circuit for Quench and Reset Pulse Generation. IEEE J. S. Topics Quantum Electronics, 13(4):926–933, 2007. J. Elsey, D.R. McKenzie, J. Lambert, N. Suchowerska, S.L. Law, and S.C. Fleming. Optimal coupling of light from a cylindrical scintillator into an optical fiber. Applied Optics, 46(3):397–404, 2007. G. Espinoza, J. I. Golzarri, J. S. Bogard, and J. Garcia-Macedo and. Commercial Optical Fibre As TLD Material. Rad. Prot. Dos., 119(1-4):197–200, 2006. A.-M. Frelin, J.-M. Fontebonne, G. Ban, A. Batalla, J. Colin, A. Isambert, M. Labalme, T. Leroux, and A. Vela. A New Scintillating Fiber Dosimeter Using a Single Optical Fiber and a CCD Camera. IEEE Trans. Nuc. Sci., 53(3):1113–1117, 2006. R. S. Fyath and J. J. O’Reilly. Analysis of nonideal solid-state photomultiplier (sspm) for optical receivers. IEE Proc., 135(6):423–434, 1988. V. Golovin and V. Saveliev. Novel type of a photodetector with geiger mode operation. Nuc. Instr. Meth. Phys. Res. A., 518(1-2):560–564, 2004. S. Gomi, H. Hano, T. Lijima, S. Itoh, K. Kawagoe, S.H. Kim, T. Kubota, T. Maeda, T. Matsumura, Y. Mazuka, K. Miyabayasi, H. Miyata, T. Murakami, T. Nakadaira, T. Nakaya, H. Otono, E. Sano, T. Shinkawa, Y. Sudo, T. Takeshita, M. Taguchi, T. Tsubokawa, S. Uozumi, M. Yamaoka, H. Yamazaki, M. Yokoyama, K. Yoshimura, and T. Yoshioka. Development and study of the multi pixel photon counter. Nuc. Instr. Meth. Phys. Res. A., 581:427–432, 2007. 126 Photonics Hamamatsu. S10362-11-050C, April 2007. URL http://sales. hamamatsu.com/en/products/solid-state-division/si-photodiode-series/ mppc/part-s10362-11-050c.php. Photonics Hamamatsu. H5783-03, June 2008. URL http://sales.hamamatsu.com/ en/products/electron-tube-division/detectors/photomultiplier-modules/ part-h5783-20.php. E. Heckathorne, R. Silverman, F. Daghighian, and M. Dahlbom. Prototype solid state photomultiplier based intra-operative beta camera. Nuclear Science Symposium Conference Record, 6:3571–3575, 2006. A. L. Huston, B. L. Justus, P. L. Falkenstein, R. W. Miller, H. Ning, and R. Altemus. Optically stimulated luminescent glass optical fibre dosimeter. Rad. Prot. Dos., 101 (1-4):23–26, 2002. A.L. Huston, B.L. Justus, P.L. Falkenstein, R.W. Miller, N. Hing, and R. Altemus. Remote optical fiber dosimetry. Nuc. Instr. Meth. Phys. Res. B., 184:55–67, 2001. D.E. Hyer, R.F. Fisher, and D.E. Hintenlang. Characterization of a water-equivalent fiber-optic coupled dosimeter for use in diagnostic radiology. Medical Physics, 36(5): 1711–1716, 2009. C. Jackson. Silicon Photomultiplier Detectors for Low Light Detection. Photonics Spectra, pages 64–66, December 2007. J.V. Jelly. Cerenkov Radiation and its Applications. Pergamon Press, London UK., 1958. B.L. Justus, P. Falkenstein, A.L. Huston, and M.C. Plazas. Gated fiber-optic-coupled detector for in vivo real-time radiation dosimetry. Applied Optics, 43(8):1663–1668, 2004. 127 B.L. Justus, P. Falkenstein, A.L. Huston, and M.C. Plazas. Elimination of cerenkov interference in a fibre-optic-coupled radiation dosemeter. Radiation protection Dosimetry, 120(1-4):20–23, 2006. J. S. Kapustinsky, R. M. DeVries, N. J. DiGiacomo, W. E. Sondheim, and J. W. Sunier. A fast timing light pulser for scintillation detectors. Nuc. Instr. Meth. Phys. Res. A., 241:612–613, 1985. S. O. Kasap. Optoelectronics and Photonics: Principles and Practices, 1st ed. PrenticeHall, New York, 2001. J. Kim, Y. Yamamoto, and H. Hogue. Noise-free avalanche multiplication in Si solid state photomultipliers. Appl. Phys. Lett., 70(21):2852–2854, 1997. G. F. Knoll. Radiation Detection and Measurement, 3rd ed. John Wiley & Sons, New York, 2000. A. L. Lacaita, F. Zappa, S. Bigliardi, and F. Manfredi. On the bremsstrahlung origin of hot-carrier-induced photons in silicon devices. IEEE TED., 40(3):577–582, 1993. F. Lacroix, A. S. Beddar, M. Guillot, L. Beaulieu, and L. Gingras. A design methodology using signal-to-noise ratio for plastic scintillation detectors design and performance optimization. Med. Phys., 36(11):5214–5220, 2008. J. Lambert, Y. Yin, D.R. McKenzie, S. Law, and N. Suchowerska. Cerenkov-free scintillation dosimetry in external beam radiotherapy with an air core light guide. Phys. Med. Biol., 53:3071–3080, 2008. J. Lambert, Y. Yin, D.R. McKenzie, S. Law, and N. Suchowerska. Cerenkov light spectrum in an optical fiber exposed to a photon or electron radiation therapy beam. Applied Optics, 48(18):3362–3367, 2009. 128 S.H. Law, N. Suchowerska, D.R. McKenzie, S.C. Fleming, and T. Lin. Transmission of cerenkov radiation in optical fibers. Optics Letters, 32(10):1205–1207, 2007. B.S. Lee, Y.M. Hwang, H.S. Cho, and S. Cho. Fabrication of fiber-optic radiation sensor tips with inorganic scintillator for remote sensing of x or γ-ray. Nuclear Science Symposium Conference Record, 7:865–868, 2004. B.S. Lee, DH. Cho, KW. Jang, SC. Chung, JW. Lee, S. Kim, and H. Cho. Measurements and characterization of cerenkov light in fiber-optic radiation sensor irradiated by highenergy electron beam. Jap. J. Appl. Phys., 45(10A):7980–7982, 2006. B.S. Lee, K.W. Jang, D.H. Cho, SC. Chung, GR. Tack, GM. Eom, and JH. Yi. Fabrication and characterization of a minature fiber-optic scintillating detector for electron-beam therapy dosimetry. J. Kor. Phys. Soc., 51(5):1645–1648, 2007a. B.S. Lee, KW. Jang, DH. Cho, WJ. Yoo, GR. Tack, SC. Chung, S. Kim, and H. Cho. Measurements and elimination of cherenkov light in fiber-optic scintillating detector for electon beam therapy dosimetry. Nuc. Instr. Meth. Phys. Res. A., 579:344–348, 2007b. J.E. Martin. Physics for Radiation Protection, 2nd ed. Wiley-VCH, Berlin, 2006. Y. Matsuura and M. Miyagi. Hollow Optical Fibers for Ultraviolet and Vacuum Ultraviolet Light. IEEE J. Quan. Elec., 10(6):1430–1434, 2004. R.J. McIntyre. Multiplication Noise in Uniform Avalanche Diodes. IEEE TED., 13(1): 164–168, 1966. D. McNally and V. Golovin. Review of solidstate photomultiplier developments by CPTA and Photonique SA. Nuc. Instr. Meth. Phys. Res. A., 610(1):150–153, 2009. C. D. Motchenbacker and J. A. Connelly. Low Noise Electronic System Design. John Wiley & Sons, Inc., New York, New York, 1993. 129 L. Moura and I. Darwazeh. Introduction to Linear Circuit Analysis and Modeling From DC to RF. Elsevier, Amsterdam, 2005. R. Nowotny. Radioluminescence of some optical fibres. Phys. Med. Biol., 52:N67–N73, 2007. America NSG. GRIN Design Parameters, September 2008. URL http://www. nsgamerica.com/index.php?lang=english&page=selfoc_materials. J. C. Palais. Fiber Optic Communications, 5th ed. Pearson Prentice-Hall, Upper Saddle River, New Jersey, 2005. N. Pavlov, G. Maehlum, and D. Meier. Gamma spectroscopy using a silicon photomultiplier and a scintillator. 2005 IEEE Nuclear Science Symposium Conference Record., pages 173–180, 2005. D. Pellion, K. Jradi, F. Moutier, F. Oms, J.P. Gardou, C. Magenc, T. Camps, D. Esteve, and A.R. Bazer-Bachi. APD photodetectors in the Geiger photon counting mode. Nuc. Instr. Meth. Phys. Res. A, 610:410–414, 2009. M. Petasecca, B. Alpat, G. Ambrosi, P. Azzarello, R. Battinson, M. Ionica, A. Papi, G.U. Pignatel, and S. Haino. Thermal and electrical characterization of silicon photomultiplier. IEEE Trans. Nuc. Sci., 55(3):1681–1690, 2008. SA Photonique. PH0810G1, May 2006. URL http://www.photonique.ch. E.B. Podgorsak. Radiation Oncology Physics: A Handbook for Teachers and Students. International Atomic Energy Agency, Vienna, 2005. E.B. Podgorsak. Radiation Physics for Medical Physicists. Springer, Berlin, 2006. R.S. Quimby. Photonics and Lasers. John Wiley & Sons, Inc., New Jersey, 2006. D. Renker. New trends on photo detectors. Nuc Instr Meth. Phys. Res. A., 571:1–6, 2006. 130 D. Renker and E. Lorenz. Advances in solid state photon detectors. J. Inst., 4:1–47, 2009. Z. Sadygov, A. Olshevski, I. Chirikov, I. Zheleznykh, and A. Novikov. Three advanced designs of micro-pixel photodiodes: Their present status, maximum possibilities and limitations. Nuc. Instr. Meth. Phys. Res. A., 567:70–73, 2006. Z.Y. Sadygov, I.M. Zheleznykh, N.A. Malakhov, V.N. Jejer, and T.A. Kirillova. Avalanche Semiconductor Radiation Detectors. IEEE Trans Nuc. Sci., 43(3):1009–1013, 1996. V. Saveliev and V. Golovin. Silicon avalanche photodiodes on the base of metal-resistorsemiconductor (MRS) structures. Nuc. Instr. Meth. Phys. Res. A., 442:223–229, 2000. D. Scansen and S. O. Kasap. Excess noise, gain, and dark current in ge avalanche photodiodes. Can. J. Phys., 70(10-11):1070–1075, 1992. S. Seifert, D. T. van Dam, J. Huizenga, R. Vinke, P. Dendooven, H. Lohner, and D. R. Schaart. Simulation of silicon photomultiplier signals. IEEE Trans. Nuc. Sci., 56(06): 3726–3733, 2009. Inc. SensL. Silicon Photomultipliers, December 2006. URL http://sensl.com/ products/silicon-photomultipliers/. D. Shushakov, V. Shubin, K. Sitarsky, E. Levin, E. Shelegeda, and S. Vinogradov. New approach to solid state photomultipliers. SORMA West, Berkeley CA., 1094:1–5, 2008. N. Suchowerska, J. Lambert, T. Nakano, S. Law, J. Elsey, and D.R. McKenzie. A fibre optic dosimeter customized for brackytherapy. Radiation Measurements, 42:929–932, 2007. J. Swain, S. Reucroft, and Y. Musienko. Recent Developments in Pixellated Avalanche Photodiodes. IEEE Nuc. Sci. Symp. Conf. Rec., J03-45:2944–2948, 2005. S. M. Sze. Physics of Semiconductor Devices 3rd ed. Wiley-Interscience, New York, 2006. 131 N. Tsoulfanidis. Measurement and Detection of Radiation, 2nd ed. Taylor & Francis, Washington D.C., 1995. J. E. Turner. Atoms, Radiation, and Radiation Protection, 3rd ed. WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim, 2007. S. Vinogradov, T. Vinogradova, V. Shubin, D. Shushakov, and K. Sitarsky. Probabilistic distribution and noise factor of solid state photomultiplier signals, taking into account cross-talk processes. Bulletin of the Lebedev Physics Institute, 36(9):3–13, 2009. Inc. Voxtel. SBQF-EKOA, April 2008. URL http://www.voxtel-inc.com/index.php/ products/apds-product-page/. S. Webber and D. Christ. Comparison of LuYAP, LSO, and BGO as scintillators for high resolution pet detectors. IEEE Trans. Nuc. Sci., 50(3):1370–1372, 2003. K. Wick and T. T. Zoufal. Unexpected behavior of polystyrene-based scintillating fibers during irradiation at low doses and low dose rates. Nuc. Instr. Meth. Phys. Res. B., 185:341–345, 2001. Inc. Zecotek. Micro-pixel Avalanche Photo Diodes (MAPD), April 2008. URL http: //www.zecotek.com/503/1107/. 132 APPENDICES 133 A APPENDIX A TEC Cooler Schematic for VOXTEL SQBF-EKAA 134 FIGURE 0.2: Circuit diagram for TEC cooler controller. 135 B APPENDIX B Fiber Efficiency Calculations • ǫaccept: For butt-joined OFs or OF-scintillator junctions, NA is modified by a term n1 /no where n1 is the index of the OF core (1.492 for GH4001 fiber cable) and no is the plastic scintillator refractive index (1.58). This gives a new NA value of 0.482. Thus the new acceptance cone angle is 2 · sin−1 (newN A) = 57.62◦ . Now ǫaccept is approximately the elementary solid angle that falls within the 57.62◦ R acceptance cone angle. It is (1/4π) sinθdθdφ = 0.01. • ǫtransmit−OF : GH4001 fiber cable has an attenuation of 190 db/km or 0.19 db/m @ 650 nm. This is 0.475 db for the 2.5 m length used in this work. The attenuation factor is: 0.475db = −10logT . Solving for T gives a transmission value of 0.896 or 89.6%. Table 0.1 shows attenuation characteristics for GH4001 plastic fiber cable out to 100 meters. 136 TABLE 0.1: GH4001 plastic fiber cable attenuation characteristics. GH4001 length (m) Attenuation (db) Attenuation factor 0 0 0 2.5 0.475 0.896 5 0.95 0.8035 7.5 1.425 0.7202 10 1.9 0.6456 12.5 2.375 0.5787 15 2.85 0.5188 20 3.8 0.4169 25 4.75 0.335 30 5.7 0.2691 35 6.65 0.2163 40 7.6 0.1738 45 8.55 0.14 50 9.5 0.1122 100 19 0.0126 137 C APPENDIX C Laboratory Dose Calculations Dose rate in Gy/s from a radioactive point source is given by: Ḋ = S µen · · Eγ · 1.602 × 10−10 [Gy/s] 2 4πr ρ (C.1) where the 1.602 × 10−10 term (conversion factor) is for unit conversion to Gy from MeV and g etc., r is the distance to the point source (in cm), S is the isotope energy flux in [d/s or dps] (from appendix D), µen /ρ, is the mass attenuation coefficient for the isotope in (cm2 /g), and Eγ is the energy of the isotope (in MeV). The 60 Co source had 1.1 mCi activity on 06/09/1960 (age 48.6 years). 29284.6d/s (0.0326 4π(0.2 cm)2 • For 137 Cs: • For 60 Co: 68154.16d/s (0.0243 4π(0.2 cm)2 cm2 /g)(0.662 M eV )(1.602 × 10−10 ) = 0.201 µGy/s cm2 /g)(1.25 M eV )(1.602 × 10−10 ) = 0.661 µGy/s These calculations confirm the laboratory observations that the stimulated the scintillators better than 137 Cs 60 Co point source using the same physical geometry. 138 D APPENDIX D Decay of Laboratory Sources, Number of Photons Emitted, and Prelude 420 Crystal Activity D1 Decay of Laboratory Sources Radioactive decay is governed by equation D.1: A = Ao e−λt ln 2 T1/2 where λ = = 0.693 T1/2 , (D.1) t is time, T1/2 is half-life, and activity A is in [Ci] or [Bq], often expressed as disintegrations per second (d/s or dps) or transformations per second (t/s or tps). The µCi 137 Cs 60 Co source had 1.1 mCi activity on 06/09/1960 (age 48.6 years). For the 1 and 1.1 mCi 60 Co sources that were 3 years and 48.6 years old, respectively, when delivered from campus Radiation Safety. Using equation D.1: 137 Cs 60 Co D2 = (1x10−6 Ci)(3.7x1010 dps/Ci)e −0.693x3yrs 30.07yrs = (1.1x10−3 Ci)(3.7x1010 dps/Ci)e = 34528 dps −0.693x48.6yrs 5.27yrs = 68154.16 dps Number of Photons Emitted by Decayed Sources The total number of photons emitted by the decayed sources is given by equa- tion D.2: N = t × A × TB (D.2) where t is the time, A is the activity (equation D.1), and TB is the total branching ratio for the isotope (the probability of specific transitions in the decay chain). TB itself is the product of the branching ratio for that photon energy (BR) and the branching fraction for the mode of decay (BF): T B = BR × BF . N137 Cs = 1s × 34528 dps × 0.85 = 29248.6 662 keV photons in 1 s 139 N60 Co = 1s × 68154.16 dps × 1 = 68154.159 1.773, 1.332 MeV photons in 1 s 1 . D3 Prelude 420 (Lu1.8 Y.2 SiO5 : Ce) Crystal Activity Prelude 420 crystals contain 176 Lu, a natural beta emitter, which emits a 307 keV gamma ray as part of its decay cascade. Here the Specific Activity (SA) of the 1 mm x 1 mm x 5 mm crystals used in the laboratory is estimated using equation D.3. [Cember, 1998] SA = Activity Nλ λNa = = m m Ao (D.3) where Na is Avogadro’s number (6.025E23 atoms/mol), λ is the same as equation D.1, and Ao is the compound mass in g/mol. • Mass of Lu1.8 Y.2 SiO5 : Ce = 7.1 g/cm3 .2 • Volume of 1 mm x 1 mm x 5 mm Prelude 420 crystal is 0.1 x 0.1 x 0.5 = 0.005 cm3 . • Mass of 1 mm x 1 mm x 5 mm Prelude 420 crystal is 7.1 x 0.005 = 0.0355 g. The mass percentage M of Lu in Lu1.8 Y.2 SiO5 : Ce is M ass% = = 1.8 M assLu 1.8 M assLu + 0.2M assY + M assSi + 5 M assO 1.8 × 174.967 = 71.44% 1.8 × 174.967 + 0.2 × 88.905 + 28.085 + 5 × 15.999 The mass of Lu in Lu1.8 Y.2 SiO5 : Ce is 0.714 x 0.0355 = 0.02536 g. There is 2.59% of 176 Lu in natural Lu; the radioactive mass in each Prelude 420 crystal is 0.0259 x 0.02536 g = 6.57E-4 g. 1 Data for branching ratios are from Martin [2006], Chart of the Nuclides, and http://www.detectors. saint-gobain.com 2 http://www.detectors.saint-gobain.com/PreLude420.aspx 140 Using equation D.3 the specific activity of each 1 mm x 1 mm x 5 mm Prelude 420 crystal is SA = ln2 × 6.025E23 atoms/mol = 2002.3 Bq/g 1.192E18 s × 174.967 g/mol thus the specific activity is 2002.3 Bq/g x 6.57E-4 g = 1.315 Bq.