Optical Frequency Domain Imaging of Human Retina and Choroid by Edward Chin Wang Lee B.A.Sc. Engineering Physics University of British Columbia, 2004 SUBMITTED TO THE DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE OF MASTER OF SCIENCE IN ELECTRICAL ENGINEERING AND COMPUTER SCIENCE AT THE MASSACHUSETTS INSTITUTE OF TECHNOLOGY JUNE 2006 0 2006 Massachusetts Institute of Technology / 9 /All/pights reserved Signature of Author: Department of Electrical Engineering and Computer Science May 20, 2006 Certified by: Seok-Hyun Yun Assistant Professor of Dermatology, Harvard Medical School Thesis Supervisor Certified by: Brett E. Bouma Associate Professor of Dermatology, Harvard Medical School Member of the Faculty of the Harvard-MIT Divigion of Health Sciences and Technology Thesis Sunervisor Accepted by: Arthtir C. Smith MASSACHUSETTS NSTITUTE 0 F IECH LOGY NOV 0 2 2006 L IES6 LIBRARIES Professor of Electrical Engineering and Computer Science Chairman, Committee for Graduate Students BARKER Optical Frequency Domain Imaging of Human Retina and Choroid by Edward Chin Wang Lee Submitted to the Department of Electrical Engineering and Computer Science on May 20, 2006 in Partial Fulfillment of the Requirements for the Degree of Master of Science in Electrical Engineering and Computer Science ABSTRACT Optical coherence tomography (OCT) has emerged as a practical noninvasive technology for imaging the microstructure of the human eye in vivo. Using optical interferometry to spatiallyresolve backreflections from within tissue, this high-resolution technique provides cross-sectional images of the anterior and posterior eye segments that had previously only been possible with Current commercially-available OCT systems suffer limitations in speed and histology. sensitivity, preventing them from effective screening of the retina and having a larger impact on the clinical environment. While other technological advances have addressed this problem, they are inadequate for imaging the choroid, which can be useful for evaluating choroidal disorders as well as early stages of retinal diseases. The objective of this thesis was to develop a new ophthalmic imaging method, termed optical frequency domain imaging (OFDI), to overcome these limitations. Preliminary imaging of the posterior segment of human eyes in vivo was performed to evaluate the utility of this instrument for comprehensive ophthalmic examination. The 1050-nm OFDI system developed for this thesis comprised a novel wavelength-swept laser that delivered 2.7 mW of average power at a sweep rate of 18.8 kHz, representing a two-order-ofmagnitude improvement in speed over previously-demonstrated lasers in the 1050-nm range and below. The system, with an optical exposure level of 550 gW, achieved resolution of 10 gm in tissue and sensitivity of >92 dB over a depth range of 2.4 mm. Two healthy volunteers were imaged with the OFDI system, with 200,000 A-lines over 10.6 seconds in each imaging session. In comparison to results from a state-of-the-art spectral-domain OCT system, the OFDI system provided deeper penetration into the choroid. This thesis demonstrates OFDI's capability for comprehensive imaging of the human retina, optic disc, and choroid in vivo. The deep penetration power of the system enabled the first simultaneous visualization of retinal and choroidal vasculature without the exogenous dyes required by angiography. The combined capability for imaging microstructure and vasculature using a single instrument may be a significant factor influencing clinical acceptance of ophthalmic OFDI technology. Thesis Supervisor: Seok-Hyun Yun Title: Assistant Professor of Dermatology, Harvard Medical School Thesis Supervisor: Brett E. Bouma Title: Associate Professor of Dermatology, Harvard Medical School Member of the Faculty of the Harvard-MIT Division of Health Sciences and Technology 3 4 Acknowledgements All I ever wanted to learn about research, I learnt at MIT and Wellman - from tinkering during rotation, to gaining project ownership, to eventually building a state-of-the-art imaging system from scratch. Yet, none of this could have been possible without the tremendous help I received from the incredible people around me. My supervisor Andy Yun has been an amazing inspiration to me, with his enthusiastic attitude towards science and seemingly unlimited energy. And how often do you get to shoot a laser into your boss' eye and get published for it? Brett Bouma, my co-supervisor, is the main reason behind my finishing this thesis and graduating on time. I am most grateful for his guidance and genuine interest in my personal development over the last two years. I would also like to thank Johannes de Boer and Mircea Mujat for their indispensable roles in this marvelous collaboration. Johannes provided his super-stable eye and much- needed expert opinion on retinal imaging, while Mircea remained extremely courteous and helpful in spite of my constant harassment. Aside from Andy and myself, Catherine Bolliet had the honor (misfortune) of spending the most time examining my OFDI images. I cannot overstate the value of her feedback and artwork for this thesis. It is rare in life to work in a truly collegial environment, but that is what I have found at the Wellman Center for Photomedicine. I will definitely miss the marshmallows from Alyx Chau and her expertise in Word and MATLAB. William Oh and Pilhan Kim have my gratitude for their patient reception to stupid questions; Jason Motz and Ben Vakoc have my appreciation for their invaluable technical assistance. I am also indebted to Gary Tearney and Seemantini Nadkarni for their supervision during my rotation through mini-projects. Jason Bressner has been a great buddy, and was gracious enough to agree to be the first person besides my supervisors to read this thesis. And I would like to extend a big thank-you to everyone else at the lab for their impressive work ethics and high tolerance for noise. Like everything else in life, money can be a practical problem for poor graduate students, even those with high ideals. As someone without the noblest ideals, I fortunately have 5 had the blessing of the Canadian government and education system - I benefited from Canada's high-quality secondary education and excellent undergraduate training for free. I would also like to acknowledge the scholarship support for this graduate research from the Natural Sciences and Engineering Research Council of Canada. Last but certainly not least, my family and friends have been an inexhaustible source of love and encouragement. In particular, thank you to Mom and Dad and Susan for your unwavering support. It means a lot to me to know that, no matter what path I take, I will always have you on my side. For one more time, thank you. Artistic Rendering of Retinal Vasculature by CatherineBolliet 6 Table of Contents CHA PTER 1: IN TR OD U CTION .................................................................................. 1.1 OPHTHALMIC IMAGING .............................................................................................. 11 12 1.1.1 A Brief History of Inventions and Advances..................................................................... 13 1.1.2 Optical Coherence Tomographyfor Ophthalmic Imaging............................................... 15 THEORY OF OPTICAL COHERENCE TOMOGRAPHY (OCT)......................................... 17 1.2 .1 Interferom etry........................................................................................................................ 17 1.2.2 From OCDR to OCT ............................................................................................................. 20 1.2 .3 SD -O C T ................................................................................................................................. 22 1.2 OPTICAL FREQUENCY DOMAIN IMAGING (OFDI)..................................................... 24 1.3 .1 B ackground ........................................................................................................................... 24 1.3.2 Sensitivity Advantage ................................................ 25 1.3.3 OFDIvs SD-OCT .................................................................................................................. 29 NEW OPPORTUNITY - IMAGING HUMAN RETINA AND CHOROID WITH OFDI ...... 31 1.3 1.4 CH A PTER 2: LA SER .................................................................................................... 37 2.1 INTRODUCTION ............................................................................................................. 37 2.2 SETUP............................................................................................................................ 37 2.2.1 Polygon-BasedFilter............................................................................................................ 2.2.2 Design and Operation...........................................................................................................39 RESULTS ....................................................................................................................... 2.3 CHA PTER 3: O FD I SY STEM ...................................................................................... 37 40 43 3.1 INTRODUCTION .......................................................................................................... 43 3.2 SETUP............................................................................................................................ 43 3.2.1 Design and Operation....................................................................................................... 3.2.2 DataAcquisition....................................................................................................................45 43 SIGNAL PROCESSING .................................................................................................... 47 3.3.1 BackgroundSubtraction................................................................................................... 48 3.3.2 Windowing and FourierTransform................................................................................... 49 3.3.3 Interpolationto Linear k-Space..........................................................................................49 3.3.4 DispersionMismatch Compensation................................................................................. 3.3.5 Image Construction...............................................................................................................52 3.3 3.4 RESULTS ....................................................................................................................... 51 54 7 CHAPTER 4: IN VIVO IMAGING OF HUMAN RETINA AND CHOROID......... 61 4.1 INTRODUCTION ............................................................................................................. 61 4.2 OFDI IMAGING ............................................................................................................. 61 4.3 COMPARISON TO AN 840-NM SD-OCT SYSTEM....................................................... 63 CHAPTER 5: VISUALIZING RETINAL AND CHOROIDAL VASCULATURE 67 5.1 INTRODUCTION ............................................................................................................. 67 5.2 AUTOMATIC DEPTH-SECTIONING ALGORITHM ......................................................... 68 5.3 FUNDUS-TYPE IMAGES .............................................................................................. 72 CHAPTER 6: SUMMARY AND DISCUSSION ......................................................... 6.1 SUMMARY..................................................................................................................... 77 6.2 DISCUSSION .................................................................................................................. 79 REFERENCES................................................................................................................ 8 77 81 List of Figures MICHELSON INTERFEROMETER AND DETECTION CURRENT. ....................................................... 12 14 16 18 TYPICAL SETUP OF AN OCT SYSTEM. ...................................................................................... 20 CREATION OF CROSS-SECTIONAL OCT IMAGE BY SUCCESSIVE AXIAL SCANS. ........................ 21 TYPICAL SETUP OF A SD-OCT SYSTEM. .................................................................................... 24 FIGURE I -1. SIDE VIEW OF THE EYE. ............................................................................................................. FIGURE FIGURE FIGURE FIGURE FIGURE FIGURE 1-2. 1-3. 1-4. 1-5. 1-6. 1-7. INDOCYANINE GREEN ANGIOGRAMS. ..................................................................................... COMPARISON OF OCT IMAGE TO HISTOLOGY. ............................................................................ 24 FIGURE 1-8. TYPICAL SETUP OF AN OFDI SYSTEM. ...................................................................................... FIGURE 1-9. WATER ABSORPTION'S WAVELENGTH DEPENDANCE AND ITS IMPLICATIONS ON CLINICAL A PPLIC ATIO N S. ......................................................................................................................... FIGURE 2-1. WAVELENGTH-SCANNING FILTER. ............................................................................................. 31 38 FIGURE 2-2. EXPERIMENTAL SETUP OF WAVELENGTH-SWEPT LASER.........................................................40 FIGURE 2-3. MEASURED LASER OUTPUT CHARACTERISTICS. ..................................................................... 41 FIGURE 3-1. EXPERIMENTAL SETUP OF THE OFDI SYSTEM. ......................................................................... 44 FIGURE 3-2. TIMING DIAGRAM FOR THE OFDI SYSTEM. .............................................................................. 46 FIGURE 3-3. LABVIEW USER INTERFACE FOR THE OFDI SYSTEM. ................................................................ 47 FIGURE 3-4. INITIAL PROCESSING OF DETECTED FRINGES. ........................................................................ 48 FIGURE 3-5. NUMERICAL SIMULATION OF ACHIEVABLE RESOLUTION. ....................................................... 50 FIGURE 3-6. NUMERICAL MAPPING TO A UNIFORM K-SPACE BY INTERPOLATION. ..................................... 50 FIGURE 3-7. MEASURED SNR OF A SAMPLE REFLECTOR AS A FUNCTION OF REFERENCE ARM POWER. ........... 55 FIGURE 3-8. GLASS SLIDE FOR LASER TUNING CALIBRATION. .................................................................... 56 FIGURE 3-9. TREMENDOUS IMPROVEMENT TO REFLECTIVITY PROFILE AFTER INTERPOLATION AND DISPERSION COMPENSATION. ................................................................................................ 57 FIGURE 3-10. RESOLUTION PERFORMANCE ACROSS THE ENTIRE DEPTH RANGE. ....................................... 57 FIGURE 3-11. POINT SPREAD FUNCTIONS MEASURED AT VARIOUS DEPTHS. .................................................. 58 FIGURE 3-12. ORIGINAL PLOT OF POINT SPREAD FUNCTIONS BEFORE NOISE FLOOR SUBTRACTION. .............. 59 FIGURE 4-1. REPRESENTATIVE OFDI IMAGE FRAME. .................................................................................. FIGURE 4-2. COMPARISON OF TWO IMAGING SYSTEMS (OFDI AT 1050 NM AND SD-OCT AT 840 NM). 62 ............ 64 FIGURE 5-1. ILLUSTRATION OF AUTOMATIC DEPTH-SECTIONING ALGORITHM. .......................................... 71 FIGURE 5-2. RETINAL AND CHOROIDAL VASCULATURE. ............................................................................ 72 FIGURE 5-3. COMPARISON OF OFDI FUNDUS-TYPE IMAGES TO SLO IMAGE. ................................................ 74 9 ... on eyes ... You cannot depend on your eyes when your imaginationis out of focus. -- Mark Twain 10 Chapter 1: Introduction Humans might have seen the world for ages, but the world has only seen inside the living human eye for a little over 150 years. Until the invention of the ophthalmoscope by Helmholtz in 1851, the fine structures inside the living eye had remained an inaccessible mystery. Since then, there have been many exciting advances in the field of ophthalmic imaging. In particular, optical coherence tomography (OCT) has emerged in the last decade as a practical noninvasive technology that can provide clinically-meaningful images of the human eye in vivo and in real time. With its unique capability in highresolution cross-sectional imaging, OCT offers a compelling advantage over other existing technologies for ophthalmic clinical applications. Current commercially-available OCT systems suffer limitations in speed and sensitivity, preventing them from effective screening of the retina and having a larger impact on the clinical environment. While other technological advances have addressed this problem, they are inadequate for imaging the choroid, which can be useful for evaluating choroidal disorders as well as early stages of retinal diseases. The objective of this thesis was to develop a new ophthalmic imaging method, termed optical frequency domain imaging (OFDI), to overcome these limitations. Preliminary imaging of the posterior segment of human eyes in vivo was performed to evaluate the utility of this instrument for comprehensive ophthalmic examination. The thesis is organized as follows. The rest of this chapter reviews the background for this thesis, and explores the new opportunity in posterior eye segment imaging with OFDI at 1050 nm. Chapter 2 describes the source for the OFDI system - a novel wavelength-swept laser. Chapter 3 focuses on system design and operation. Then Chapter 4 demonstrates the first OFDI imaging of the human posterior eye and Chapter 5 presents the first simultaneous visualization of both retinal and choroidal vasculature without the exogenous dyes required by angiography. Finally, Chapter 6 provides a summary and discussion of this thesis research. 11 1.1 Ophthalmic Imaging The human eye is a complex organ of numerous components (Figure 1-1). The posterior eye segment includes the vitreous, retina, and choroid and is essential to our vision. The ability to image the posterior segment plays a crucial role in the detection, monitoring, and treatment of common blinding eye diseases such as glaucoma, diabetic retinopathy, and macular degeneration. Vitreous Humor Ciliary Muscle Sclera Body Aqueous /Retina Zonules Choroid Cornea-Fovea Lens .... . . .... ii-Visual Axis-- ; ;- Lens Sack Iris ------- -----.... M acula Optic Disk- Canals of Schlemm Optic Conjunctiva Nerve Orbital Muscles Retinal Blood Vessels Figure 1-1. Side view of the eye. (reproduced from Charlie Web's website on vision loss and blindness [1]) Light enters the eye through the cornea and is focused by the lens through the vitreous humor onto the retina, where photoreceptive cells translate optical images into electrical impulses that the brain understands. Directly opposite the lens, the macular region on the retina has a dip in its center called the fovea. Densely packed with photoreceptive cone cells, the fovea provides color vision and enables high acuity. The optic nerve is responsible for transmitting electrical signals to the brain, but the nerve cells of the retina reside inside the multiple layers that absorb excess radiation and supply nutrients. In order for the optic nerve to connect to nerve cells of the retina, the optic nerve pierces the retina at a point near the macula called the optic disc. The optic disc is also known as the blind spot because no photosensitive cells exist there. Beyond the retina lies the choroid, a vascular layer that supplies retinal cells with oxygen and nourishment. It is the reflection of light from the choroidal blood vessels that causes the red eye effect in photography. 12 1.1.1 A Brief History of Inventions and Advances The invention of the ophthalmoscope by Helmholtz in 1851 marked the first milestone towards the goal of imaging the posterior segment. Helmoholtz's design consisted of a partially-reflecting mirror that directed light from a source onto the retina. The reflected light transmitted through the partially-reflecting mirror and was magnified to form an image. With lenses and mirrors, the ophthalmoscope equipped scientists with a tool for examining the retina. In 1886, Jackman and Webster recorded the first in-vivo human retinal photograph, showing the optic disc and larger blood vessels [2]. Such en face retinal photography known as fundus photography was commercialized by Zeiss in 1920. It was at first limited in clinical use due to the slow speed of film and long exposure time with the then carbon-arc illumination system, until the invention of the electronic flash in the 1950s. Shortly after in 1961, the first successful fluorescein angiography was administered in humans. Through intravenous injection of the fluorescent fluorescein dye, fluorescein angiography has become the main diagnostic tool for study of retinal circulation [3]. The choroid, however, is usually not visible in either fundus photography or fluorescein angiography, due to the strong scattering and absorption of the retinal pigment epithelium above it. By the early 1990s, indocyanine green angiography [4] has gained clinical acceptance for the study of choroidal circulation. The indocyanine green dye facilitates penetration into the choroid with its infrared emittance and excitation spectra. Its strong binding to blood proteins also results in slow diffusion out of the fenestrated choriocapillaris in contrast to the rapid leakage of fluorescein dye, which prevents visualization of choroidal vascular details. Webb's invention of the scanning laser ophthalmoscope (SLO) in the early 1980s [5] thrust the field of fundus imaging into a new era. Instead of capturing the image as a whole, the SLO samples the retina point by point in a raster-like fashion with its laser beam. The SLO's high light efficiency allows the laser beam better penetration through the lens and corneal opacities even at a low light level, resulting in improved spatial resolution and contrast. With real-time continuous imaging, the SLO can be used in conjunction with angiography to monitor dye arrival and leakage (Figure 1-2) [4]. The SLO is housed in a human interface that uses a high-power condensing lens to image the 13 retina onto a plane within the instrument, which is in turn imaged by another lens to the eye of the operator or a recording device. b a Figure 1-2. Indocyanine green angiograms for a 48-year-old woman without significant retinal pathology: (a) 72 seconds and (b) 22 minutes 32 seconds after injection. The choroidal vessels discernible in Figure (a) are usually not visible in fluorescein angiograms. With the dye exited from the vasculature in Figure (b), the angiogram is identical to a normal fundus image. ( reproduced from Jozik et al's publication in Retina [4] ) The traditional two-dimensional fundus view from an ophthalmoscope is limited in clinical diagnostic value and often requires additional techniques such as fluorescein angiography and visual field testing that are sensitive to the physiologic consequences of structural abnormalities. One of the greatest modern developments in ophthalmic imaging is the ability to evaluate posterior microanatomy in three dimensions. aforementioned SLO was combined with confocal optics in 1987 [6]. The In addition to obtaining a higher contrast by reducing light scatter from other ocular structures, the confocal SLO is capable of depth-sectioning and enables en face fundus imaging with micron-scale transverse and -300-im axial resolution. Meanwhile, ultrasound has been widely used clinically for quantitative measurements of intraocular distances. With its principle of operation similar to radar detection of aircraft, ultrasound determines distances within the eye from the echo delay of sound from different boundaries within the eye. It has the inconvenient requirement of direct contact of the ultrasound measuring 14 device to the cornea or immersion of the eye in a liquid which facilitates transmission of sound waves into the globe. Standard ultrasound offers axial resolution of 150 pm [7]; although higher-frequency ultrasound can offer higher resolution approaching 20 pm, it has been limited to use for the anterior segment due to its strong attenuation in biological tissues [8]. The resolutions of computed tomography and magnetic resonance imaging are also limited to hundreds of microns [9, 10]. All these current techniques do not have sufficient depth resolution to provide useful cross-sectional images of retinal structure. In comparison, optical coherence tomography has emerged as a promising technology for three-dimensional imaging of the posterior segment, by offering high transverse and axial resolution (<10 pm) in a noninvasive and non-contact manner. 1.1.2 Optical Coherence Tomography for Ophthalmic Imaging Optical coherence tomography (OCT) is analogous to ultrasonography in operation. However, instead of sound waves, OCT measures the echo time delay and intensity of backscattered light from sites within the eye. Because the high speed of light does not permit direct detection of echo signals, OCT uses low-coherence interferometry. light with The retina is virtually transparent with extremely low optical backscattering, but the high sensitivity of OCT enables detection of such weak signals. In contrast to conventional microscopy, OCT decouples the governing mechanisms for the axial and transverse resolution, and thus allows for high resolution in all three dimensions. In short, OCT is a noninvasive, cross-sectional diagnostic imaging modality that is capable of producing a highly-accurate structural representation of the retina. OCT was first demonstrated in 1991 for in vitro imaging of the human retina and atherosclerotic plaque [11]. Then in 1993, the human optic disc and macula were imaged in vivo [12, 13]. Fortunately for the field of OCT, the rise of the telecommunication industry brought numerous technological advances in fiber optics, and the industry's downturn provided sophisticated components at affordable prices. Hence, it is now possible to engineer a compact and robust OCT system at low cost. Carl Zeiss Meditec introduced the first commercial OCT system to the ophthalmic marketplace in 1996. To 15 date, partly due to the ease of optical access to the eye, OCT has made the largest clinical impact in ophthalmology. The primary strength of OCT in ophthalmic imaging lies in its high-resolution, noninvasive imaging of the retina in vivo, as the precise visualization of pathology is critical for the diagnosis and staging of ocular diseases. OCT's axial resolution far exceeds that of ultrasound or confocal SLO, and approaches that of conventional histology. Although the imaging depth is limited by the high optical scattering of biological tissue to a few millimeters, it is on the same scale as histology, sufficient for imaging the entire thickness of the retina. Figure 1-3 shows the remarkable resemblance of OCT images to histology. Since excisional biopsy of the retina is unviable, OCT can serve as an excellent noninvasive tool for diagnosis and monitoring of diseases, as well as evaluating response to therapeutic intervention. In addition, OCT's ability to examine posterior microanatomy in three dimensions facilitates detection of diseases in their earliest stages, when treatment is most effective and irreversible damage can be most easily prevented or delayed. a) b) Remarkable resemblance of OCT image to histology. Figure 1-3. (a) In vivo OCT image of a healthy volunteer near the macular region (obtained with the system built for this thesis). (b) Histology of a different subject. ( Figure 3b reproduced from Uniformed Services University's website [14]) 16 Other advantages of OCT make it a practical tool with significant clinical impact. For example, the new high-speed OCT systems make real-time diagnosis a reality. Also, compared to conventional fundus photography, OCT requires no dilation and causes minimal discomfort to the patient with its low-intensity infrared illumination. Unlike ultrasound, OCT is a non-contact method that is well tolerated by patients. Moreover, objective and reproducible quantitative values can be derived from OCT images. For example, the thickness of a retinal structure is simply the thickness measured from the OCT image divided by the group refractive index of the retina (n = 1.38 [15, 16]). This is important, since thickness maps of retinal structures can be useful for detection of pathology. For instance, a thickness map of the nerve fiber layer is of great diagnostic value for macular diseases and glaucoma [17, 18]. Last but not least, OCT can be extended to functional imaging applications such as Doppler blood flow measurements [15, 19], blood oxygenation quantification with spectroscopy [20], and tissue birefringence measurements with a polarization-sensitive system [21]. With its long list of benefits, optical coherence tomography is no longer a research curiosity. It is gaining acceptance as a clinical diagnostic tool for the three leading causes of blindness [22] - glaucoma [18], diabetic retinopathy [23], and macular degeneration [24]. Clinical studies have also been performed to investigate its feasibility for diagnosis and monitoring of other retinal diseases such as macular edema [17], macular hole [25], central serous chorioretinopathy [26], epiretinal membranes [27], and optic disc pits [28]. 1.2 Theory of Optical Coherence Tomography (OCT) 1.2.1 Interferometry The heart of optical coherence tomography is the employment of interferometry, a method with a long history and numerous applications in diverse areas. The Michelson interferometer commonly used in OCT was invented around 1881. Before its application in OCT, it provided the famous first evidence against the existence of the aether and paved the path to modem techniques in optical precision measurements. 17 Figure 1-4(a) illustrates the free-space configuration of a Michelson interferometer. A collimated light beam is split by a beamsplitter into two arms. Light in the sample arm probes the sample and its backscattered signal is recombined with light from the reference arm at the beamsplitter. Assuming a simplified case where the source is perfectly coherent (monochromatic) with wavenumber k and the sample is a partial reflector of reflectance R, the detector current can be expressed as [29]: idet(t) oc (1.1) 2 PrjPRcos(2kzjt)) where Pr is the optical power reflected from the reference arm at the photodetector, P, is the optical power reflected from the sample arm at the photodetector assuming a perfect mirror sample, and zo is the sample's position relative to the scanning reference mirror (or the path length difference between the two arms). In other words, the detector current varies sinusoidally when the reference mirror is scanned back and forth mechanically (Figure 1-4b). The DC components of the detector current are neglected in Equation (1.1), since the desired sample information is contained in the interferometric term. Reference Mirror a) Linht ;nrc 77 liJ b ) Detector c) Short Coherence Length Envelope Long Coherence Length C Co co t,z t~z Figure 1-4. (a) Free-space configuration of a Michelson interferometer. (b) Sinusoidal detection current with a scanning reference mirror and a perfectly-coherent source. (c) Detection current with a low-coherence source. The coherence length, 6z, is also shown. 18 Now consider the case of a low-coherence source with finite bandwidth AA. Also assume that the reflectors are spectrally uniform and that the sample and reference arms consist of a uniform, linear, and non-dispersive material. The detector current can be expressed as [29]: idet(t) C R -real edJWOATP JS(o - co, )ej(CO)A r d( - wo 2;r )} (1.2) where S(w-wo) is the spectrum of the source with center frequency 0-)o, Azr is the phase delay mismatch, and Arg is the group delay mismatch. This time, as the reference arm is scanned, the detector current still oscillates at the carrier frequency (first exponential term), but is now modulated by an envelope (integral term) that is the inverse Fourier Transform of the source power spectrum. The envelope is the interferometer's detected signal of the mirror sample and thus characterizes the system's point spread function (Figure 1-4c). The width of the envelope (FWHM value), also known as the coherence length or resolution, is given as follows [29]: 0Z= 22 n 7r AA (1.3) assuming a Gaussian source spectral profile of center wavelength AO and bandwidth AA, and a sample of refractive index n. This low-coherence interferometry was first applied in the telecommunication industry in 1987 and was called optical coherence domain reflectometry (OCDR). Since interference is observed only when the lengths of the two arms of the interferometer are matched to within the coherence length, a short coherence length translates to a high system resolution. The interferometric approach also offers an extremely high sensitivity. As in optical heterodyne detection, the weak field from the sample is amplified by the strong field from the reference beam. Furthermore, the detector current is proportional to the field of the sample signal rather than its intensity, giving rise to a high dynamic range and sensitivity. Therefore, by employing low-coherence light and demodulating the interference output, OCDR is a nondestructive method used for high-resolution, highsensitivity measurements of optoelectronic devices [30, 31]. 19 1.2.2 From OCDR to OCT It was not long after the development of OCDR when the potential of low-coherence interferometry for biomedical imaging applications became obvious. Figure 1-5 depicts a typical OCT system, which has required a few modifications from an OCDR system for applications in biomedical imaging. First of all, most clinical OCT systems employ fiber optics for its environmental stability and compactness. Second, the paramount importance of resolution in biomedical imaging demands a large source bandwidth, since the axial resolution is inversely proportion to the bandwidth of the source, as indicated in Equation (1.3). Short-pulse lasers in laboratories are extremely broadband, but compact and cost-effective superluminescent diodes or semiconductor-based light sources are more suitable for building commercial systems. The wavelength of the source also bears important implications for possible clinical applications with OCT, considering the absorption curves of tissue constituents (Section 1.4). Third, the requirements for the mechanical scanning of the reference mirror are different. Higher speed is desired for imaging to minimize motion artifacts, while the depth for imaging is much less than that for OCDR due to tissue scattering and absorption. .. _.-Broadband Source Detector .... . Mirror reference arm (50150) sample arm Sample Figure 1-5. Typical setup of an OCT system. The 50/50 fiberoptic coupler replaces the free-space beam splitter. OCT is capable of providing three-dimensional information, whereas OCDR operates in one dimension. With the reference mirror scanning for information in the axial direction, two orthogonal galvanometer mirrors are used to scan in the transverse directions. The collimated light in the sample arm reflects off the galvanometer mirrors and is focused by a lens onto the sample. As the galvanometers change the angles of the mirrors, the beam focus is scanned across the sample. OCT's trademark cross-sectional 2D image can be 20 created with one galvanometer mirror by successive axial scans at different transverse locations (Figure 1-6). For 3D data, the second galvanometer mirror also scans slowly in the other transverse direction for successive cross-sectional images. The compiled 3D data is often displayed as a movie sequence of cross-sectional images, or used for 3D rendering of the sample. .... Scan Transverse............................... .................................... Reflectivit No Depth Figure 1-6. OCT's trademark cross-sectional 2D image is created by successive axial scans at different transverse locations. On the right of the OCT image is a reflectivity versus depth plot for one axial scan (blue arrow). For retinal imaging, the galvanometer mirrors are housed in a human interface similar to that used for the scanning laser ophthalmoscope. The transverse resolution depends on the imaging optics. Assuming a Gaussian beam at wavelength A, it can be shown [29] that a lens with a focal lengthf and filled aperture D gives a spot size (1/e 8x = Azfcus = 2 width) of: (1.4) rD and a depth of focus of: (1.5) Thus better transverse resolution requires a decrease in the depth of focus, as in conventional microscopy. Unlike conventional microscopy or the SLO, OCT's axial resolution depends only on the temporal coherence properties of the source, and not on the pupil-limited numerical aperture of the eye or ocular aberrations. In addition, recent works have shown that adaptive optics can further improve transverse resolution for retinal imaging [32]. 21 Measurements of axial eye length and corneal thickness were some of the first biomedical applications of low-coherence interferometry [33, 34]. Since then, OCT has been used to investigate numerous clinically-meaningful physical properties that change the amplitude, phase, or polarization of backscattered light. It is important to keep in mind that, despite its accurate description of the retina, OCT reports optical properties of the tissue and does not necessarily reflect the true histopathologic morphology. As well, artifacts and other noises can arise when light is strongly attenuated by media opacities such as corneal edema, significant cataract, and vitreous hemorrhage. Finally, OCT images are constructed based on the time delay of reflected light. If the reflected light is multiply scattered before being collected, it would appear to originate from a site deeper than the actual location of the reflection. Furthermore, multiple scattering can also cause speckle, an inherent noise source of coherent imaging. Fortunately, such multiple scattering is a minor concern. The retina is a relatively non-turbid tissue, the confocal configuration of most OCT systems spatially selects singly-scattered light, and multiplyscattered light tends to lose temporal coherence with successive scattering events preventing detection by OCT. 1.2.3 SD-OCT As OCT progressed from the research laboratory into the clinical setting, scientists and clinicians recognized the need to increase acquisition speed without compromising sensitivity or resolution. This need has been fulfilled by a second-generation technology called spectral-domain optical coherence tomography (SD-OCT), also known as FourierDomain OCT (FD-OCT) [35, 36]. In the original time-domain method (TD-OCT), the reference mirror is mechanically scanned to obtain the interference pattern for each sample depth sequentially in time. In SD-OCT, the entire depth profile is interrogated all at once, while the reference arm pathlength is kept constant. The spectrum of the interference signal is acquired by a spectrometer and then analyzed to yield the desired depth profile. 22 The spectrometer is often custom-built and consists of a grating, a lens and a CCD array. The collimated interference light is dispersed by the grating and focused by the lens onto the CCD array. Assuming a single reflector of reflectance R at depth zo in the sample arm, the interferometric signal measured by the CCD array is [29]: ispec(k) oc 2 PrPRcos(2kzo) (1.6) neglecting the DC terms. This means that a single reflector sample induces in the k-space domain a characteristic sinusoid, whose frequency and amplitude are directly proportional to the depth and reflectance of the sample, respectively. Clearly, the depth and reflectance of the sample can be obtained via the Fourier Transform relation. In the case of a more complex sample, additional surface reflections simply superimpose sinusoids of different frequencies corresponding to their depths and reflectances. Because the Fourier Transform is a linear operation, the complete reflectance profile of the sample can be reconstructed from the Fourier Transform of ispec(k). SD-OCT has enabled video-rate imaging at unprecedented speed and resolution. Because the sensitivity of a SD system does not suffer the same inverse relationship with resolution as a TD system does [37, 38], ultrahigh resolution is now possible [39, 40]. Also, SD systems can afford to operate at very high speed while maintaining sufficient sensitivity, owing to their intrinsic sensitivity advantage (Section 1.3.2). Furthermore, SD-OCT has eliminated inconveniences, such as nonlinearity, associated with the mechanical scanning of the reference mirror. A TD system is limited in speed due to the mechanics of its scanning reference mirror, but a SD system is only restrained by the detection rate of its CCD array. Since affordable broadband sources and spectrometers are readily available in the 800-nm range, SD-OCT is mostly applied at this wavelength for ophthalmic applications. Figure 1-7 depicts a typical SD system. 23 J JLJ Mirror , ._.- !Broadbandi Source reference arm (i 50/50) Spectrometer sample armi Sample -_ -_._ -_._._._._! Figure 1-7. Typical setup of a SD-OCT system. 1.3 Optical Frequency Domain Imaging (OFDI) 1.3.1 Background Optical frequency domain imaging (OFDI), also known as swept-source OCT (SS-OCT), is another second-generation method for OCT [41]. Like OCT itself, OFDI is based on a technology from the telecommunication industry - optical frequency domain reflectometry (OFDR). OFDR is used for characterizing optoelectronic devices [42] as well as fiber-optic cables [43], and OFDI is its biomedical imaging counterpart. An OFDI system (Figure 1-8) uses a wavelength-swept laser source and a single photodetector in place of the broadband source and spectrometer in SD-OCT. Similar to SD-OCT, OFDI has no scanning reference mirror, and offers the same sensitivity advantage [37] that makes simultaneous high resolution, speed, and sensitivity feasible. It enjoys several additional benefits such as reduced susceptibility to motion-induced signal fading [44], a simple polarization-sensitivity or diversity scheme [45], and a long ranging depth [41]. Mirror reference arm i Tunable Source (50/50) sample arm Detector ! !1 . I Sample Figure 1-8. Typical setup of an OFDI system. 24 As its name suggests, a wavelength-swept laser sweeps its output wavelength periodically in time in a monotonic fashion. Assuming a single reflector of reflectance R at depth zo in the sample arm, the detector current without the DC terms is [41]: idet(t) oc 2 PPRcos(2k(t)zo ) (1.7) where k(t) = 27r / A(t) is the wavenumber of the swept laser at time t. Suppose that the tuning of the laser obeys the linear relation k(t) = ko + k1 t. Then the linear tuning of the laser has effectively mapped the k-space into the time domain; the discrete-time detector current recorded by the data acquisition board can be written as i(t,)=i(k). Equation (1.7) is thus similar to Equation (1.6) for SD-OCT, with surface reflections inducing characteristic sinusoids in the detected signal. By analogy, the Fourier Transform will also yield the reflectance profile of the sample. In reality, the laser does not tune linearly in k-space, but this can be corrected numerically [41, 46], as explained in Section 3.3.3. OFDI has been extensively applied in the 1300-nm wavelength range. A main reason for this is the wide availability of commercial fiber-optic components at 1300 nm and the lack of a wide-tuning rapidly-swept light source outside this range [47-49]. Besides this, OFDI can achieve a larger usable imaging depth at 1300 nm as a result of lower tissue scattering at longer wavelengths. Finally, OFDI is typically preferred over SD-OCT at 1300 nm because affordable high-speed spectrometers are currently unavailable. The drawback of using long-wavelength sources is the quadratic dependence of resolution on wavelength (Equation (1.3)): 10-pm resolution requires only 28 nm of bandwidth at 800 nm, but would require 75 nm at 1300 nm. 1.3.2 Sensitivity Advantage The sensitivity of an OCT system is a measure of the minimum detectable reflectivity R2 in the sample arm. Since the measured signal is proportional to the reflectance R, the sensitivity is equal to the ratio of the time-averaged signal power to the time-averaged noise power. 25 1.3.2.1 Noise Sources There are four main sources of noise in an OCT system: thermal or Johnson noise, digitization noise, relative intensity noise, and shot noise. Thermal noise arises from random particle motion in resistors due to their thermal energy. Digitization noise refers to the excess noise generated in the data acquisition board. Relative intensity noise (RIN) describes any noise source with a power spectral density that scales linearly with the mean photocurrent power. Optical source power fluctuation is one example. Shot noise, on the other hand, is a white-noise process that is a consequence of the quantized nature of light and charge. The photodetector emits charge at a mean rate that depends on the detected optical power; however, the time between specific emissions is random. Such current fluctuations are termed shot noise. OCT offers extremely high sensitivity for detection of weak reflections from the retina. Besides utilizing optical heterodyne detection and enjoying the advantage of measuring optical field rather than intensity, OCT can achieve quantum-limited performance. Thermal noise and electrical noise can be minimized with a high-gain electrical amplifier circuit placed before the data acquisition board. RIN can be reduced by appropriately selecting the reference power and employing dual-balanced detection. When shot noise becomes the dominant noise source, the OCT system is said to be shot-noise-limited. 1.3.2.2 Sensitivity of TD-OCT For a shot-noise-limited TD system with detector signal current is and noise current in, the sensitivity can be shown to be [29]: SNRTD - i (t)) ((t) _ -- _____ (1.8) 2 E, NEB where q is the quantum efficiency of the photodetector, E. is the energy of a single photon, NEB is the noise-equivalent-bandwidth of the system, and P, is again the optical power reflected from the sample arm at the photodetector assuming a perfect mirror sample. The bracket < > denotes the time average and Equation (1.8) assumes that 26 PsR2<<Pr,which is generally true for retinal imaging. The noise-equivalent-bandwidth is essentially the electronic detection bandwidth, which is linearly proportional to the system's axial-line (A-line) rate fA and optical bandwidth AAZ. Since the optical bandwidth is inversely proportional to the axial resolution 3z, it follows that: NEB x fAAA oc fA/z and C power -resolution speed SNRTD Equation (1.10) describes the tradeoff in the design of a TD system. (1.9) (1.10) Most retinal imaging applications require a level of sensitivity close to 100 dB and cannot tolerate a reduction in sensitivity to achieve a higher frame rate or better resolution. Although the source is the ultimate limitation on maximum optical power, the maximum power used for imaging is usually constrained by the safety limit for retinal exposure and is not considered a design variable. Resolution is critical to identifying retinal structures and pathologies, while high speed is required to minimize motion artifact and patient discomfort. 1.3.2.3 Sensitivity of OFDI For the case of OFDI, recall that the reflectance profile R(z) can be obtained via Fourier Transform. This derivation [37, 41] assumes a square-profile spectral envelope and 100% tuning duty cycle for the source, i.e., constant output power in time. A Discrete Fourier Transform (DFT) of the detector current i(k,) with M samples gives: M F(z) -j2rlm i(k,)e = (1.11) . m=1 Parseval's theorem, I F 2 = MXi2 , holds for both signal and noise [50]. The sampled noise current can be shown to be mutually uncorrelated in the case of Nyquist sampling [38, 51]. Hence, the white-noise power adds incoherently, yielding: (F, = I F! M = (f) n = M(i). (1.12) 27 Now consider the signal power in the Fourier domain (reflectivity) for a mirror at depth zo. It is zero everywhere except for two peaks at zl=±zo, giving F, (z = +zo 1=1 = 2 - MD, = i2 2K~)(.3 S (1.13) with the coherent addition of signal power. Therefore, SNROFDI - |Fs((zi =+Zo 20 M 2 (F2) SNRTD (1.14) and in the shot-noise limit, SNROFDI t s Ev f (1.15) where the A-line ratefA is the tuning speed of the source for OFDI. It can be shown [37, 41] that Equation (1.15) is valid for a more general case where the source spectral envelope is not square and the tuning duty cycle is less than 100%, if Ps is taken to be the time-averaged value over one tuning cycle. In all cases, while the noise power is distributed across all frequencies, the signal power of a single discrete reflection remains concentrated in two peaks in the Frequency domain (±zo). Compared to Equation (1.8), the noise-equivalent-bandwidth for OFDI is equal to only the A-line rate, rather than the electronic bandwidth which is the A-line rate multiplied by the optical bandwidth. The significance of this result and the meaning of Equation (1.14) may be better understood when one considers that M/2 corresponds to the number of spatially-resolvable points in the ranging depth [41]. For a system of better resolution, M is larger and the sensitivity gain of OFDI is even more profound. That is because TD-OCT, unlike OFDI, suffers from the explicit inverse relationship between sensitivity and resolution (Equation (1.10)). 1.3.2.4 Sensitivity of SD-OCT The derivation of sensitivity for SD-OCT [37] is straightforward following the above analysis. As indicated by Equation (1.6), the SD-OCT signal is also discrete in wavenumber. While OFDI discretizes its signal by sampling the detected light in time, SD-OCT performs discretization by dispersing the detected light onto M discrete 28 detectors of the CCD array. The analysis for SD-OCT can be related to that for OFDI by comparing the integrated signal from one CCD detector, ispec(km), to one sample value of the photodetector current, idet(kd. If the source power is the same for OFDI and SDOCT, the optical power arriving at one CCD detector is reduced by a factor of M. Yet, the CCD detector's integration time is the entire duration of an axial scan, reducing the detection bandwidth also by a factor of M. Therefore, SNRsD = E, (f4 IM) = SNROFDI (1.16) showing that both SD-OCT and OFDI enjoy the same inherent sensitivity advantage in Fourier-domain techniques. The resultant higher sensitivity allows higher image acquisition speed without sacrificing sensitivity. 1.3.3 OFDI vs SD-OCT This section discusses the two second-generation OCT technologies with respect to key system performances. Alike in their Fourier-domain analyses, the two technologies differ in their hardware configurations. OFDI maps k-space to time with a tuning source whilst SD-OCT maps to spatial location with a spectrometer. This difference in hardware contributes to some marked differences in performance. For both technologies, the reflectance profile is obtained via DFT. As a result, the highest detectable frequency of the signal corresponds to a reflection at the maximum depth, which is defined by [41]: Az = 4n9A (1.17) and depends on the sampling wavelength interval 6A=AA/M according to Nyquist Theorem. A reflection from outside the depth range will appear in the image by aliasing. For OFDI, the sampling rate of the data acquisition board limits the sampling wavelength interval and hence the depth range. For SD-OCT, the limiting factor is the spectral resolution of the spectrometer. Since a real signal's Fourier Transform is symmetric, there exists an ambiguity between positive and negative depths. To avoid the 29 superposition of the positive-depth image upon the negative-depth image, the reference arm can be adjusted to position the sample entirely on the positive or negative side. Techniques have also been developed to remove this depth degeneracy [52, 53]. Not the entire depth range is always usable due to sensitivity dropoff. As explained earlier for TD-OCT, interference occurs only when the pathlength difference is within the coherence length of the source. In that case, a large bandwidth is desired for a short coherence length (high resolution). Yet for OFDI, at any instant in time, the laser light is interrogating the entire depth range simultaneously. A narrow instantaneous linewidth of the tuning laser enables interference with reflections from large depths, and corresponds to a slowly-decaying coherence function, a measure of sensitivity dropoff over z. Also note that the sampling interval should be smaller than the instantaneous linewidth of the source; otherwise, the coherence function would decay more quickly. For SD-OCT, the finite width of pixels in the CCD array leads to spatial integration of the interference spectrum, and causes a strong dependence of sensitivity on depth [51, 54]. In short, OFDI enjoys a greater ranging depth and a greater usable depth, due to its higher sampling wavelength interval and narrow linewidth, respectively. Other qualities of the different hardware lead to further differences in characteristics. Availability of affordable equipment dictates the operating wavelength range and the possible clinical applications of each technology accordingly. The A-line rate depends on the repetition rate of the tuning source and the speed of the CCD array for OFDI and SD-OCT, respectively. Because an additional photodetector is much more convenient and economical than an additional spectrometer, polarization diversity, polarization sensitivity, and dual-balanced detection are more easily implemented in OFDI. Lastly, SD-OCT's CCD array integrates its signal during the entire A-line acquisition. Consequently, SD-OCT is more sensitive to phase instabilities [44] induced by motion or blood flow, with fringe washout resulting in a weaker signal and a smaller dynamic range for Doppler flow measurements. 30 OFDI and SD-OCT share almost the same relations in terms of resolution. The axial point-spread function of the system is given by the Fourier Transform of the source power spectrum, with the axial resolution being inversely proportional to the spectral width. For OFDI, the spectrum refers to the time-averaged spectrum of the tuning source. The transverse resolution is determined by the focusing properties of the optics and the eye. A high-NA lens in the human interface is desired for a small spot size on the retina, since a shallow depth of focus can be tolerated with the small retinal thickness. 1.4 New OpportunityImaging Human Retina and Choroid with OFDI Among the numerous applications of optical coherence tomography (OCT), ophthalmic imaging is the most clinically-advanced area to date. Ophthalmic imaging presents unique challenges in comparison to other imaging applications, because the aqueous and yitreous humors are 99% water. For this reason, water absorption plays a pivotal role in any optical imaging system for the posterior eye segment. Figure 1-9 shows water absorption's dependence on optical wavelength [55] for roundtrip propagation in a typical human eye that is modeled as a water volume of 21 -mm length. 4030 - OFDI 200 10 - retinal imaging) SD-OCT - retinal imaging 0 700 800 900 1000 1100 0D OFDI non-retinal tissue imaging 1200 1300 Wavelength (nm) Figure 1-9. Water absorption for roundtrip propagation in a typical human eye that is modeled as a water volume of 21-mm length. The applications of SD-OCT and OFDI are also shown at their respective wavelengths of operation. The potential of OFDI for retinal imaging is investigated in this thesis for the 1 -pm range. 31 The standard spectral range of conventional ophthalmic OCT has been between 700 and 900 nm. Not only does near-infrared light transmit well through the vitreous, it minimizes patient discomfort in comparison to visible light. The availability of broadband sources was an equally-important incentive that invited the development of the first-generation time-domain OCT (TD-OCT) systems in the 800-nm range. Spectraldomain OCT (SD-OCT) was also developed at 800 nm to take advantage of the broadband sources and fast CCD cameras at this wavelength. It has since enabled threedimensional retinal imaging in vivo with superior image acquisition speed and sensitivity compared to TD-OCT. Optical frequency domain imaging (OFDI) delivers the same improvements in imaging speed and sensitivity as SD-OCT and offers several additional advantages. The reduced scattering at its long operating wavelength (1300 nm) affords greater light penetration depth, but water absorption at this wavelength becomes a dominator factor for retinal imaging. Even when assuming a perfectly-reflecting retina, less than 0.5% of incident light can be measured in reflection. For this reason, only the human anterior eye segment has been imaged by 1300-nm systems [56, 57]. Until now, however, a clinically-viable OFDI system has been unavailable outside the 1300-nm range, primarily due to the lack of a wide-tuning rapidly-swept light source. Therefore, retinal imaging has been out of reach for OFDI systems. Recent studies have suggested that the 1-pm region [58-60] could be a viable alternative operating window for retinal imaging. The zero-dispersion point of water occurs around 1 tm so operation in that range can lead to easier dispersion management. The 1-pm region could also benefit from less attenuation from scattering in opaque eye media, that occurs in older patients with cataract lenses and haze in cornea [58]. Most importantly, there exists a local minimum in water absorption (Figure 1-9) and the small absorption loss at this wavelength can be compensated with higher incident optical power. The ANSI (American National Standards Institute) standards govern the maximum limits on optical power incident on human eyes [61]. With the retina being the main safety concern, the ANSI standards are concerned with the actual power impinging the retina. 32 Hence, the ANSI power limitations, having taken water absorption into account, are less restrictive at longer wavelengths. For continuous exposure up to eight hours, the maximum power of light into the pupil is 600 ptW at 800 nm. At 1050 nm where roundtrip water absorption loss is -3 dB, the maximum power is 1.9 mW. Therefore, the absorption loss at 1050 nm can be compensated by imaging at a higher power that is still well below the ANSI safety limit. In contrast, the huge loss of -20 dB at 1300 nm requires power beyond the capability of existing lasers and the ANSI safety limit, which does not rise as fast as water absorption beyond 1 pim. It should be noted that the correct value for maximum power is obtained by multiplying the area of the human pupil by the maximum power density from the ANSI standards. Imaging in the 1-pm region could also potentially offer deeper penetration into the choroid below the retinal pigment epithelium (RPE) [58]. The highly absorbing and scattering nature of the RPE becomes evident in the case of RPE atrophy, in which enhanced penetration and visualization of the choroid is observed [58]. Most of the eye structure is designed to facilitate transmittance of light to the retina, where light is absorbed by photoreceptors. Melanin, a chromophore in the RPE and choroid, absorbs any excess radiation to prevent disruptive reflections within the eye that might otherwise result in the perception of confusing images. In fact, it is the choroid that gives the inner eye a dark color. Due to melanin's strong absorption, typical 800-nm OCT images show weak signals from only superficial layers of the choroid. Melanin does have a decreasing absorption spectrum from 600 nm to 1200 nm and scattering in biological tissue exhibits a similar trend. Therefore, there exists a window of opportunity in the 1 -pm region for deeper penetration into the choroid. The visualization of morphological features in the choroid can offer substantial benefits. Early stages of retinal pathologies such as age-related macular degeneration and proliferative diabetic retinopathy are often accompanied by choroidal neovascularization [62], an extensive growth of new blood vessels in the choroid and retina which irreversibly impairs vision in the affected regions. Therefore, the ability to image the choroid and detect the onset of choroidal neovascularization can provide valuable insight 33 to retinal specialists. Many diseases of the retina also have characteristic findings in choroidal circulation, whose current imaging method requires intravenous injection of indocyanine green dye. The visualization of this choroidal vasculature with a non- contact, non-invasive OCT technology could have a significant clinical impact. This thesis reports the development of a high-performance wavelength-swept laser with a center wavelength at 1050 nm. The laser source was incorporated into an OFDI system and the first OFDI imaging of posterior segments of the human eye in vivo with high image acquisition speed, sensitivity, and penetration depth was demonstrated. With the system's enhanced penetration, depth-sectioned fundus-type reflectivity images of the choroidal capillary and vascular networks were also obtained. 34 35 ... on seeing ... "The trick is to love somebody. Ifyou love one person, you see everybody else diffkrently -- 36 James Baldwin Chapter 2: Laser 2.1 Introduction During the last decade, the development of rapidly scanning, widely tuning laser sources has been driven by diverse applications in optical reflectometry, sensor interrogation, test and measurement applications, and biomedical imaging. A commonly-used technique is to employ an intracavity narrowband wavelength-scanning filter. Although single- frequency operation was demonstrated with sophisticated grating filter design [63], it is not essential for imaging applications and can be compromised to enhance tuning speed. Sufficiently-narrow linewidths and wide sweep ranges were achieved by the use of rapidly-tuning elements such as acousto-optic filters and Fabry-Perot filters [64, 65]. Yet, their speed had been less than 1 kHz, inadequate for video-rate biomedical imaging. In 2003, a novel wavelength-scanning filter based on a polygon scanner and diffraction grating was developed. The filter was incorporated into a 1300-nm extended-cavity laser [66] that achieved a variable repetition rate an order of magnitude faster than previously demonstrated. The rest of this chapter describes the principles, design, and characteristics of the rapidly scanning, widely tuning laser - the key enabling element of the 1050-nm OFDI system for posterior eye imaging. 2.2 Setup 2.2.1 Polygon-Based Filter As shown in Figure 2-1, the wavelength-scanning filter comprises a diffraction grating, a telescope with two lenses in an infinite-conjugate configuration, and a polygon mirror scanner. With the grating at the front focal plane of the first lens and the polygon spin axis at the back focal plane of the second lens, the telescope serves two distinct roles: the conversion of diverging angular dispersion into converging angular dispersion, and the control of the imaged beam size and convergence angle at the polygon. As indicated in 37 Figure 2-1, from all the light converging onto the polygon scanner, only a narrow band that is normal to the front mirror facet is reflected back at any instant in time. As the polygon scanner rotates, the filter selects the narrow spectral band of light that is reflected back through the telescope into the laser cavity, and thus accomplishes wavelength tuning. The actual direction of wavelength tuning depends on the orientation of the beam's incidence angle and rotation direction of the polygon. For example, Figure 2-1 illustrates a sweep in increasing wavelength. fibeir-o Polyon mirror r Figure 2-1. Wavelength-scanning filter. F, and F2 are the focal lengths of Lens 1 and Lens 2, respectively. ( reproduced from Yun et al's publication in Optics Letters [66] ) The following derivation [66] assumes a collimated Gaussian beam incident on the grating. By the grating equation, the center wavelength of the filter's tuning range is: -% p (sin a+ sin/p) = where p is the grating pitch, and a and p (2.1) are the angles of the incident beam and the optical axis of the telescope, respectively, with respect to the grating normal. The instantaneous linewidth of light from the filter's output can be shown to be: F1FWHM where A = = AAGpm)cos(a/W) (2.2) 41n2/;r, m is the diffraction order, and W is the l/e 2 width of the Gaussian beam at the collimator. If the angular range of the spectrum incident upon the polygon is greater than the facet angle (6 = 2n/N), the N-sided polygon mirror can retroreflect more than one spectral component at a given time. The spacing of these spectral components is called the free spectral range: AAFSR 38 = p9(F2 /F)cosp8 (2.3) where F1 and F 2 are the focal lengths of the first and second lens, respectively. Although the tuning range of the filter is fundamentally limited by the finite numerical aperture of the first lens, in practice it is the free spectral range that determines the tuning range of the laser for a homogenously-broadened gain medium. Finally, in order to maintain the duty cycle of the laser sweep at 100%, all the beams within the spectral tuning range should fall within a mirror facet without clipping, or equivalently, (F2 -S)9+W' < 2L (F 2 -S)o -W' > 0 and (2.4) (2.5) where W'= W(cos f/cos aXF2 /FI) is the beam size at the polygon mirror and S is the distance between the second lens and the front of the polygon mirror. The filter of the 1050-nm system in this thesis was designed in accordance with the above equations and in consideration of the bandwidth of the available semiconductor optical amplifier (SOA). A laser source with a large bandwidth is desired for high system resolution, but the limited bandwidth of the SOA presents a tradeoff. If the filter's free spectral range is too small, the full bandwidth of the SOA would be underutilized; if it is too large, the laser will operate with a reduced duty cycle. Optical components were selected with the following optimal parameters: p = 1200 lines/mm, F1 = 100 mm, F2 = 50 mm, N = 40, m = 1, a = 65 deg, and / = 21.5 deg. Corresponding to these design parameters, the theoretical linewidth was ~0. 1 nm and the free spectral range was 61 nm. 2.2.2 Design and Operation Figure 2-2 depicts a schematic of the laser source incorporating the wavelength-scanning filter. The linear-cavity configuration is an attractive alternative to the previous ring cavity design [66] as low-loss, low-cost circulators and isolators are not readily available at 1050 nm. The gain medium was a SOA that was recently introduced to the commercial market (QPhotonics, Inc., QSOA-1050). It was bi-directional and driven at an injection current level of 400 mA. One port of the amplifier was coupled to the filter. The other port was spliced to a Sagnac loop mirror made of a 50/50 coupler. The Sagnac loop also served as an output coupler [67]. Two counterpropagating waves traveled 39 along identical paths in the loop, but were in different polarization states during different parts of the loop, depending on the tuning and location of the polarization controller PC,. Hence the reflectivity and output coupling ratio were complementary and optimized by adjusting PC, to tune the amount of birefringence-induced non-reciprocity in the loop. Sweep repetition rates of up to 36 kHz were possible with 100% duty cycle, representing a significant improvement over previously demonstrated swept lasers that offered tuning rates of a few hundred Hz in the 1050-nm range and below [47-49]. In order to achieve a good depth range given the speed limitation of the available data acquisition board, the laser was operated at a repetition rate of 18.8 kHz in the OFDI system, producing a polarized output with an average output power of 2.7 mW. lO"p minrw PC, 50150 Pc G Lens Lens Polygon scanning fter Figure 2-2. Experimental setup of wavelength-swept laser. 2.3 Results Figure 2-3(a) depicts the output spectrum measured with an optical spectrum analyzer in peak-hold mode (resolution = 0.1 nm). The output spectrum spanned from 1019 to 1081 nm over a range of 62 nm determined by the free spectral range of the filter. The spectral range coincided with a local transparent window of the eye. The roundtrip optical absorption in human vitreous and aqueous humors was estimated to be 3 - 4 dB based on known absorption characteristics of water (Figure 2-3a) [55]. Using a variable-delay Michelson interferometer, the coherence length of the laser output, defined as the roundtrip delay resulting in 50% reduction in interference fringe visibility, was measured to be approximately 4.4 mm in air. This value represented the entire usable depth range, 40 including the positive and negative regions. From this value, the instantaneous linewidth of laser output was calculated to be 0.11 nm. (a) (b) ' 08 6 \ j. 08 s 0.4 4 E 25 2 1020 1040 10 1080 3 0--50 0,111I",.T (nm) 0 Time 50 100 (ps) Figure 2-3. Measured laser output characteristics. (a) Peak-hold output spectrum (blue curve) and optical absorption in water (red curve) for 42-mm propagation distance corresponding to a roundtrip in typical human vitreous. (b) Time-domain output trace. Figure 2-3(b) depicts an oscilloscope trace of laser output showing 100% tuning duty cycle at 18.8 kHz (single shot, 5-MHz detection bandwidth). represents instantaneous optical power. The y-axis of the trace When lasing was suppressed by blocking the intracavity beam in the polygon filter, the total power of amplified spontaneous emission (ASE) in the output was ~0.5 mW. Since ASE is significantly suppressed during lasing, it is expected that the ASE level in the laser output should be negligible. The laser output exhibited significant intensity fluctuation (-8% pp). The fluctuation was a consequence of an etalon effect originating from relatively large facet reflections at the SOA chip, with a thickness equivalent to 2.5 mm in air. In the imaging system, the etalon reflection could cause interference with sample reflections, but due to its low intensity, no ghost image was observed for retinal imaging. 41 ... on life (and science) ... "Ihear and I target. I see and I remember. I do and1 I understand. -- Confucius 42 Chapter 3: OFDI System 3.1 Introduction Optical frequency domain imaging (OFDI) was championed in the 1300-nm region [41], driven by the development of high-speed wavelength-swept sources based on available semiconductor optical amplifiers [66, 68]. In addition to delivering improvements in imaging speed and sensitivity over TD-OCT, OFDI offers several additional advantages, such as reduced susceptibility to motion-induced signal fading [44], simple polarizationsensitivity or diversity scheme [45], and long ranging depth [41]. The rest of this chapter describes the setup, operation, and characteristics of a 1050-nm OFDI system developed for posterior eye imaging. 3.2 Setup 3.2.1 Design and Operation An OFDI system, comprising a fiber-optic interferometer, a human interface for retinal imaging, detection electronics and a computer, was constructed using the 1050-nm wavelength-swept laser as a light source (Figure 3-1). The single-mode fiber-optic implementation of the interferometer had the advantages of layout simplicity, alignment convenience, and portability. It also ensured the mutual spatial coherence of the sample and reference light incident on the detector. Furthermore, single-mode wideband fiberoptic couplers are commercially available with arbitrary splitting ratios, and have proved to be excellent replacements of beamsplitters. The sample arm (30% port) was connected to a human interface. The human interface was originally designed by Cense et al [21] based on a slit lamp. The slip lamp, an instrument commonly used by ophthalmologists for routine corneal and retinal exams, is a bio-microscope on a movable table with a headrest. When adapted with suitable lenses 43 and galvanometer scanners, it became a clinically-applicable human interface for retina imaging. Aberrations in the eye limited the diameter of the collimated beam to 2.5 mm and the focal beam size (transverse resolution) to approximately 10 pm in tissue (index = 1.38). The optical power level at the entrance pupil of the eye was measured to be 550 pW, well below the 1.9-mW maximum exposure level at 1050 nm according to the ANSI laser safety standard for continuous exposure. XYscannr swept laser Eye 30(70 Computer 90110 NO arm Figure 3-1. Experimental setup of the OFDI system. The reference arm (70% port) employed a transmission-type variable delay line. A neutral-density attenuator was used to obtain the optimal reference-arm power. Light returning from the sample was combined with the reference light at a 50/50 coupler and the resulting interference signals were measured with an InGaAs dual-balanced detector (New Focus, Inc., 1811). By using a transmission-type variable delay line instead of the usual reflection-type, the OFDI system achieved dual-balanced detection without any circulators. Not only does balanced detection reduce RIN, it also improves the dynamic range, reduces fixed-pattern noise, and suppresses self-interference noise originating from multiple reflections within the sample and optical components including the laser [51, 69]. The balanced detector achieved a common-noise rejection efficiency of -25 dB in the range between DC and 5 MHz. The detector signal was further amplified (10 dB) to minimize the effects of excess electrical noise. It was then low-pass filtered and digitized at 10 MS/s with a 12-bit data acquisition board (National Instruments, Inc., PCI6115). 44 In the reference arm, a 10% tap coupler was used to generate sampling trigger signals for data acquisition. The tapped laser output was launched into free space from a collimator, dispersed by a grating, and focused by a lens onto a fast InGaAs photodetector. This setup acted as a narrowband fixed-wavelength filter; the detector generated a pulse when the output spectrum of the laser swept through the narrow passband of the filter. The detector pulse was converted to a TTL pulse train with an electronic circuit and the TTL pulses were used to generate gating pulses for signal sampling at the data acquisition board. 3.2.2 Data Acquisition A complete OFDI imaging system is a complex instrument of many optical and electronic components. In order to acquire meaningful image data, the dynamic functions of the system's laser driver, scanning optics, and digitization electronics have to be coordinated. With the advance of digitization and computer technology, the hardware comprising the synchronization and image acquisition electronics can be as simple as a multifunctional data acquisition board (DAQ) residing in a personal computer, with a LabVIEW user interface. Figure 3-2 is the timing diagram for the OFDI system. The speed of the polygon scanner determined the A-line rate and was controlled by a driver that took a 5-Vpp square wave as its input. For this particular polygon scanner, the A-line rate fA was ten times the frequency of the input square wave. A function generator generated the square wave, but the clock output of the DAQ could also be used. As the polygon scanner rotated, a triggering signal (Figure 3-2a) was created for each A-line scan by the narrowband filter and converted into a TTL pulse train by an electronic circuit. Upon reception of each TTL pulse, the DAQ acquired M samples. Recall that M, the number of samples in each A-line, is proportional to the imaging depth range. For the purpose of retinal imaging, M was chosen to be 512 for an imaging depth of 2.44 mm in air (1.77 mm in tissue). This also means that the maximum A-line rate was 19.5 kHz, limited by the DAQ's maximum sampling rate of 10 MS/s. At a lower A-line rate, the system could gain a larger depth 45 range with a higher value of M, but a higher A-line rate could only be achieved if the system sacrificed its depth range or employed a faster DAQ. In the experiment, the system was operated at 18.8 kHz with 512 samples per A-line (Figure 3-2b). In this configuration, the data samples at the very beginning and end of each sweep were excluded and their discontinuous slopes would not distort the depth profile in the Fourier domain. Electrical ramp signals were also used to control the galvanometer mirrors of the human interface for transverse scanning (Figure 3-2c). a) fA = 18.8 kHz b) Trigger f= 18.8 kHz M = 512 samples >*-. 500 A-lines -Signal DAQ Sampling C) Period~b = 500 A-lines/frame x 200 frames/scant. pr .. ...cvi c ra e Periodfast = 500 A-lines/frame ---Fast Transverse Axis - = = Slow Transverse Axis Figure 3-2. Timing diagram for the OFDI system. (a) Triggering signal for each A-line. (b) Sampling with the data acquisition board. (c) Ramp signals for transverse scanning. The user-friendly LabVIEW interface (Figure 3-3) provided convenient control and realtime feedback of the imaging environment. The window at the top left corner displayed the acquired samples for each A-line in real time. Although the triggering signal was not automatically synchronized to the beginning of each sweep, a potentiometer on the TTL 46 triggering circuit could be used for adjustment by monitoring the displayed A-line. The operator could also use the interface to output a sawtooth waveform from the DAQ to drive a galvanometer-mirror for transverse scanning. The period corresponded to the number of A-lines in the cross-sectional frame, which was displayed in real time at the bottom right corner. Similarly, the interface could be used to synchronize transverse scanning to the beginning of each frame acquisition. An external function generator drove the slower axis of transverse scanning, whose period corresponded to the number of frames in a scan across the retina. This could also be replaced by using the second signal output of the DAQ. Finally, the top right window displayed the depth profile, the Fourier Transform of the acquired A-line in the top left window. With a sample reflector, the user could optimize the imaging system's reference arm power, polarization alignment, and dual-balanced detection by monitoring and maximizing the signal in the depth profile. Figure 3-3. LabVIEW user interface for the OFDI system. 3.3 Signal Processing Signal processing involves several steps, including reference subtraction, envelope apodization (windowing), interpolation into linear k-space, and dispersion mismatch compensation. The subtleties of image construction are also discussed below. 47 3.3.1 Background Subtraction At the beginning of every image acquisition, a reference image was obtained by blocking the sample arm. This reference image consisted of a residual background signal from the reference light, as a result of the wavelength-dependent splitting ratio of the 50/50 coupler [41] and imperfect symmetry of the balanced detectors. Subtracting the reference from the interference signal can account for source fluctuations between measurements and eliminate fixed-pattern artifacts at low frequencies (depths). The raw data of a typical A-line after balanced detection is depicted in Figure 3-4(a). The desired interference signal is contained in the fringes that are superimposed on the reference. It should be noted that the reference also fluctuates in optical power. Since this fluctuation is random in phase, it can be removed by averaging all A-lines in the reference image. Figure 3-4(b) shows the fringes after subtraction of the averaged reference. The fringes were then normalized with their envelope, before apodization with a Gaussian window, as illustrated in Figure 3-4(c,d). Raw Data a) -- b) ---- After Reference Subtraction C) -- d) 0 --- 100 200 300 Time (samples) Reference - 400 Normalized Gaussian-Windowed 500 600 Figure 3-4. Initial processing of detected fringes. (a) Raw data - fringes superimposed on the reference. (b) Fringes after subtraction of the averaged reference. (c) Normalized fringes. (d) Apodized fringes. 48 3.3.2 Windowing and Fourier Transform Apodizing interference fringes with a proper window function before Fourier Transform can suppress sideband ripples to avoid image artifacts and realize better image contrast [70, 71]. On the other hand, by the convolution property of Fourier Transform, window functions impose a limit on the achievable width of the point spread function. Consequently, the choice of an appropriate window function represents a tradeoff between ripple suppression and resolution [72]. Consider a numerical simulation with a perfect sinusoid as the time-domain data (Figure 3-5a). The use of no windowing, Gaussian windowing, and Hamming windowing yielded resolutions of 11 ptm, 13 pim, and 17 pim, respectively. It is important to realize that, as opposed to the theoretical value predicted by Equation (1.3), these values were the practical limitations on achievable resolutions, in the sense that they were the best attainable values even when assuming perfect raw data and no error sources. As expected, the best resolution was achieved with no windowing, at the expense of substantial ripples (Figure 3-5b). For this thesis research, the Gaussian window was chosen for its high resolution and reasonable ripple suppression. By inspection, the Gaussian window indeed produced the highestquality retinal images, compared to other window functions. Note that windowing can also impose a penalty on sensitivity if the effective average power is lower after normalization and windowing. For the experiment in this thesis, this corresponded to a sensitivity loss of -2 dB, given the envelope of the time-domain data. 3.3.3 Interpolation to Linear k-Space The Fourier-Transform relationship (Equation (1.11)) between the time-domain fringe data and the Fourier-domain depth profile assumes a tuning source that sweeps linearly in k-space. Nonlinearity in the tuning curve of the laser results in chirping of the signal, and this variation in the characteristic frequency of a given reflection leads to resolution degradation in z-space. A solution to this image blurring problem is to sample the detector signal in nonlinear time intervals to compensate for the frequency chirping of the source [56]. Alternatively, the existing chirped signal can be numerically mapped to a uniform k-space by interpolation prior to Fourier Transform (Figure 3-6) [41, 46]. 49 Although both methods were previously demonstrated to yield a transform-limited axial resolution, the second method is preferred for its ease of implementation in software. a) Perfect Sinusoid -C-- --- Gaussian Envelope Hamming Envelope 'U Time b) Figr 1-. NNo Window wGaussian windowing, -usnw LeTHamming 0.6 - - =0.4 -- w0.2- 0 190 200 z4 210 Depth (pm) 220 230 240 Figure 3-5. Numerical simulation of achievable resolution. (a) Perfect sinusoid shown with Gaussian and Hamming envelopes. (b) Point-spread functions with no windowing, Gaussian windowing, and Hamming windowing. - onLinear Tuning Nonlinear Tuning t Figure 3-6. Numerical mapping to a uniform k-space by interpolation. 50 The actual tuning curve of the laser is unknown, unless one explicitly measures it. Alternatively, one can estimate its polynomial expansion by iteratively searching for parameters that optimize the system's point spread function, in terms of resolution and symmetry. This procedure can be implemented with an automated algorithm, but it was carried out manually for this thesis, because it only had to be done once for a given laser configuration. The use of two parameters was found to be sufficient for this application, and the duration of the entire procedure was about 5 - 10 minutes. In the laboratory setting, the laser tuning curve remained stable over time and did not require recalibration. Sometimes more data points in the z-domain (512 by default) are needed to better visualize the reflectivity profile or compute a more precise value for resolution. Direct interpolation (linear, spline, or FFT) from a reflectivity profile is possible, but it is more accurate to perform an N-point DFT (N>512) of the time-domain data. 3.3.4 Dispersion Mismatch Compensation Chromatic dispersion arises from the wavelength dependence of the speed of light and increases linearly with propagation length in a dispersive medium. In optical communications and ultrafast measurements, dispersion compensation is essential for preserving pulse shape and temporal resolution [73]. Fortunately for OCT technologies, only the dispersion mismatch between the two arms of the interferometer must be compensated for optimal resolution. Compensation in hardware was demonstrated by matching the optical materials and path lengths in the two arms [22, 74], and compensation in software was proved feasible in TD-OCT systems [75, 76]. Recently, a software compensation method for SD-OCT was introduced by Cense et al [39]. This flexible method is well suited for retinal imaging, because dispersion compensation needs to be tailored to individual subjects with different and unknown axial eye lengths. The method is also directly applicable to OFDI, which has the spectral fringe pattern readily available in its detected signal. Dispersion mismatch introduces a phase shift e'o(k) in the detected spectral fringe pattern i(k). By Taylor series expansion, the phase shift can be expressed as: 51 9(k) = 0(ko)+ (k) (kO -k)+ ak I0( 182(k) aa(2 k (k0-k)2)2 +...+- 1 a"o(k k (k,-k)", 2 Ok n! ak ) (3.1) where ko is the wavenumber corresponding to the center wavelength. The third term in the equation represents group-velocity dispersion, which is largely responsible for dispersion mismatch. Higher-order dispersion is represented by the higher-order terms. Multiplying the spectral fringes by a nonlinear phase distorts the depth profile in the Fourier domain, resulting in a loss of axial resolution. Therefore, to correct for the chromatic mismatch, the spectral fringe pattern was multiplied with the inverse phase shift e~'O(k) [39], upon completing interpolation and prior to Fourier Transform. Like the tuning curve of the laser, the amount of dispersion mismatch is an unknown factor. In the case of a single-reflector sample, the phase shift can also be estimated in an iterative manner, searching for parameters that optimize the resolution and symmetry of the computed reflectivity profile. However, in the case of retinal imaging, isolated reflections are often unavailable, so the sharpness of the image has to be evaluated instead. For this thesis, the procedure was carried out manually, once for each subject, with the image sharpness assessed by the operator. A more objective assessment would require a sharpness metric function [40], which could be used in an automated algorithm. It should be noted that this numerical dispersion compensation procedure was done after data acquisition and did not interfere with the imaging session. 3.3.5 Image Construction The last major step of signal processing is visualization of the OFDI data. The 3D data set is sometimes visualized with volume rendering using commercially-available software, but this has not gained wide acceptance. Retinal specialists are used to 2D histological sections and fundus images, so 3D-rendered visualization is difficult for them to interpret. Instead, the 3D OFDI data set is usually visualized as a movie sequence of 2D cross-sectional reflectivity plots. These cross-sectional images can be shown realistically in a 1:1 scale, but often they are expanded vertically to reveal the microanatomy of the layered structures. The choice of mapping used to represent data 52 values can have a profound effect on the appearance and interpretability of the images. Typically, the images display the logarithm of reflectivity in units of decibels (dB), because the dynamic range of OCT images approaches 50 dB. Humans can sense brightness variation of only 3-4 decades and in practice monitors and printers place even more stringent limitations. Puliafito et al advocated for false color mapping [77], but false colors may produce image artifacts and lead to incorrect interpretation of physical structures. With standard grayscale mapping, strong reflections appear white on a black background. The inverse grayscale, with black structures on a white background, was primarily motivated by its relative ease in printing and photocopying. As will become obvious in the next chapter, the two scales lead to significant differences in image perception, as a result of the nonlinear response of human visual systems. The dimensions of OCT images, especially in retinal imaging, are computed rather than directly measured. The transverse dimensions are calculated from the angular deviations of the scanning mirrors using geometric optics. In the axial dimension, the image depth range can be calculated by Equation (1.17). The axial scale can also be experimentally determined by moving a reflector in the sample arm and dividing the distance of translation in air by that in the image. Both methods require a further division by the refractive index of tissue to yield correct intraocular distances. Since the local refractive index usually does not vary over more than a few percent, the error induced by assuming a constant index over the entire retina is negligible [78]. The value of the index used for this thesis was 1.38, based on previous studies in the 800-nm region [15, 16], since the index was not expected to differ much at 1050 nm. Although the refractive index value might not be highly accurate, OCT enjoys a high degree of repeatability in thickness measurements, which is the primary consideration for clinical diagnostics. Koozekanani et al [79] conducted a clinical study with a commercial time-domain system (Humphrey Instruments) to image twenty-six volunteers (15 images over 3 sessions for each person). The average retinal thickness was found to be 274 ± 17 ptm for a 1-mm long region 0.75 mm from the fovea. They obtained a 99% confidence interval that individual scan averages of the retinal thickness would be within 11.2 pm of the true subject value, and 53 that sessional averages would be within 7.0 pm. These results of high repeatability were comparable with those of Hee et al [80] and Baumann et al [81]. Image processing was applied extensively in the past for motion correction. Natural and saccadic patient motion was fast compared to the image acquisition time of the slow TDOCT systems, degrading image resolution and inducing artifacts. The advent of highspeed OFDI technology has in large part alleviated this concern. Even without using any motion-correction algorithm, the cross-sectional images in this thesis do not exhibit motion artifacts, although movie sequences do show a slow drift of the eye. The only image processing tool employed was a median filter for smoothing images and mitigating speckle noise. 3.4 Results Now that the theory, practical issues, and optimization tools have been discussed, this section will describe the characterization of the 1050-nm OFDI system. The test sample was a partial reflector (-73 dB) that comprised a neutral-density attenuator and a goldcoated mirror. The sensitivity of the system was the measured SNR value of the sample plus 73 dB. The operating parameters of the system were first optimized to maximize sensitivity. The performance of dual-balanced detection was originally suboptimal due to the imperfect symmetry of the balanced detector and fiber coupling at the two ports. By loosening the fiber at one port, one could carefully fine-tune the balanced detection. Now consider a sample at a fixed depth. By measuring the SNR of the sample as a function of the reference arm power, the optimal reference arm power for maximum sensitivity was determined to be 2.6 p.W at each detection port of the dual-balanced detector. Figure 3-7 illustrates the relationship between signal and noise. Note that these measurements were obtained for a previous setup that had a slightly lower optimal power. At high power, RIN noise dominated; at low power, signal dropped off faster than noise. Sensitivity was maximized when the shot noise equaled the RIN noise at the optimal power of 2.6 p.W. 54 The relatively low value is attributed to the relatively large intensity noise of the laser that cannot not be completely suppressed in balanced detection. The system was further improved by adjusting the polarization controller to align the polarization states of the two arms, while monitoring the SNR with the LabVIEW interface. a) 30 20 - z &0 100 10 -2 10, 100 10~1 Reference Arm Power (pW) 10 110 2 b) 4030 C, Signal Noise 20 100-10 10. 102 10~ 100 10 102 Reference Arm Power (pW) Figure 3-7. (a) Measured SNR of a sample reflector as a function of reference arm power. (b) Signal and noise power as a function of reference arm power. The system's tuning curve and dispersion mismatch were determined next. As explained previously, both nonlinear tuning and dispersion mismatch give rise to peak broadening. If the partial reflector's reflectivity profile were used for estimating the tuning curve's polynomial expansion, it would yield an incorrect value by including the effect of dispersion mismatch. Instead, the sample arm was blocked and a 1 -mm-thick glass slide was inserted in the reference arm. The air-glass interface reflection generated a delayed version of the reference beam (Figure 3-8). In this experiment's configuration of a transmission-type reference arm and reflection-type sample arm, the delayed reference beam's beating with the original reference beam appeared as a sample reflection at a depth equivalent to the optical thickness of the slide (~1.3 mm). Unlike a real reflection 55 ... . ..... . from the sample arm, this signal did not experience the dispersion mismatch of the two arms, except for negligible dispersion in the glass slide. Therefore, the use of the glass slide led to a more accurate estimate of the tuning curve. In this case, balanced detectin was not used, as balanced detection would have substantially reduced the selfinterference signal from the reference arm. Figure 3-8. Glass slide in the reference arm and its self-interference for laser tuning calibration. Peak broadening that results from nonlinear tuning is strongly dependent on the depth of the sample, while broadening from dispersion is depth-independent. In order to verify this, the resolution (width) of the reflectivity profiles of the partial reflector at different depths was measured by moving the reference mirror. In the case where the resolution versus depth curve was not flat after interpolation, the polynomial coefficients of the tuning curve were adjusted accordingly. The reflectivity profile of the partial reflector, after correction for interpolation, was then used to estimate dispersion mismatch. Given a flat resolution versus depth curve after interpolation, the curve should remain flat after dispersion compensation. Figure 3-9 illustrates the tremendous improvement to the reflectivity profile of the partial reflector after interpolation and dispersion compensation. The theoretical curve was obtained from a numerical simulation based on the source spectrum (Figure 2-3a) and the Gaussian window function. Before applying the Gaussian window, random noise was added to a perfect sinusoid to match the curve to the system's theoretical sensitivity. Figure 3-10 highlights the system's resolution performance. The measured values of axial resolution were 14 - 16 gm in air, slightly increasing with the depth, while the theoretical value was computed to be 13 pm from the curve in Figure 3-5(b). Errors in interpolation and dispersion compensation due to higher order terms probably account for the discrepancy. 56 IIII I II -70 Original - - - - After Interpolation After Dispersion Compensation - Theoretical \ -80 a -85 0 0~ Ij -90 0 -95 -100 I it -105 1.45 1.4 1.35 1.5 Depth (mm) 1.6 1.55 1.65 Figure 3-9. Tremendous improvement to the reflectivity profile of the partial reflector after interpolation and dispersion compensation. The theoretical curve was obtained from a numerical simulation with a perfect sinusoid plus random noise. 30 28262422- ---- After Interpolation & After Dispersion Compensation - --. Theoretical .2 200 18 16141210 -2.5 I -2 -1.5 -1 -0.5 .5 0 Depth (mm) 1 1.5 2 2.5 Figure 3-10. System performance in resolution across the entire depth range. 57 .................. Figure 3-11 depicts the point spread functions measured at various depths. The maximum SNR was 25 dB, corresponding to a maximum sensitivity of 98 dB. The theoretical shot-noise limit of sensitivity, after accounting for Gaussian windowing, was calculated to be 106 dB; the 8-dB deficiency in sensitivity of our system seems reasonable, considering that the residual laser intensity noise and imperfect polarization alignment between the sample and reference light, among many other practical details, contributed to SNR loss. In addition, due to absorption by water in the eye, the actual SNR for the human retina would be 3 - 4 dB lower than the values measured with the mirror sample. 3025- a 20- 10- 5 - 0 -2.5 -2 -1.5 -1 -0.5 0 0.5 Depth (mm) 1 1.5 2 2.5 Figure 3-11. Point spread functions measured at various depths for a sample reflectivity of -73 dB. As indicated in Figure 3-11, sensitivity decreased from 98 dB to 92 dB as the path length increased to a depth of 2.4 mm, due to the finite coherence length of the laser output. Compared to the previous time-domain system using a broadband source at 1040 nm (50nm bandwidth and 10-pm resolution in air) [58], the OFDI system offers a higher sensitivity at a 100-fold faster image acquisition speed and using only one sixth of sample arm power. The high sensitivity and depth range of the OFDI system compare favorably 58 with those of state-of-the-art spectral-domain systems using broadband sources in the 700 - 900 nm spectral range [39, 40]. The above system characterization was performed averaging over 500 A-lines at constant depth. This way random noises were removed and only systematic noises remained. The reflectivity profiles were averaged after undergoing all prior processing steps individually. The noise floor in sensitivity analysis, obtained from reference images, exhibited variations of -5 dB across the depth range due to the frequency response of the electronic low-pass filter. Hence, Figure 3-11 was produced from the original plot of point spread functions (Figure 3-12) after subtraction of the noise floor. I I N sI IFI -.. Noise Floor 30- - 25 20@1 0 *1 15 ci) 10 5 0 -2.5 -2 -1.5 -1 -0.5 0 .5 1 1.5 2 2.5 Depth (mm) Figure 3-12. Original plot of point spread functions before noise floor subtraction. 59 ... on science (and life) ... "Genius? Nothing ... Sticking to it is genius. I've failed my wiay to success. -- Thomas Edison 60 Chapter 4: In Vivo Imaging of Human Retina and Choroid 4.1 Introduction The compelling advantages of optical frequency domain imaging (OFDI) have already been shown for imaging skin, coronary artery, esophagus, and anterior eye segments [41, 45, 56, 82-85]. Until now, however, a clinically-viable OFDI system for imaging posterior eye segments has been unavailable. This chapter presents the results obtained with the OFDI system developed for this thesis. The first OFDI imaging of posterior segments of the human eye in vivo with high image acquisition speed, sensitivity, and penetration depth was demonstrated. The system's deep penetration power into the choroid was confirmed by a comparison to a state-of-the-art spectral-domain system. 4.2 OFDI Imaging OFDI imaging was conducted with two healthy volunteers (A: 36-year-old Asian male, B: 41-year-old Caucasian male). With real-time feedback (images) on the LabVIEW interface, the operator focused light through the human interface onto the retina of each subject and selected the region of interest. The OFDI system acquired 200,000 A-lines over 10.6 seconds in each imaging session as the focused sample beam was scanned over an area of 6 mm (horizontal) by 5.2 mm (vertical) across the macular region. The movie sequence of images recorded from volunteer A at a frame rate of 18.8 Hz was published [86] and can be viewed online (http://www.opticsinfobase.org/abstract.cfm?URI=oe-1410-4403). One-hundred-twenty of a total of two-hundred image frames are shown in the movie. Each frame was constructed from a thousand A-line scans with an inverse logarithmic grayscale table mapping to the reflectivity range. Figure 4-1(a) is a representative image frame from the movie. The OFDI image allows clear visualization of the anatomical layers in the retina and provides deep penetration into the choroid up to the interface with the sclera. 61 a) 2oownL. Figure 4-1. (a) Representative OFDI image frame from a movie sequence of a healthy volunteer (movie downloadable from http://www.opticsinfobase.org/abstract.cfm?URI=oe-14-10-4403). The image allows clear visualization of the anatomical layers in the retina and provides deep penetration into the choroid up to the interface with the sclera. Each image frame consists of 1000 axial lines, spans over 6.0 mm (horizontal) and 1.8 mm (depth) in tissue, and was produced with inverse logarithmic grayscale mapping. (b) Same image produced with standard logarithmic grayscale mapping. As mentioned previously, the choice of mapping in creating images can have a significant impact on the viewer's perception. Figure 4-1(b) uses the same data as in Figure 4-1(a) but applies a standard grayscale. A typical viewer can observe in Figure 4-1(a) more details of high-reflectivity areas (e.g. capillaries and pigment cells appearing as black spots in the choriocapillary layer just beneath the RPE); however in Figure 4-1(b), the low-reflectivity areas (e.g. choroidal vessels appearing as dark ellipsoid regions) are more noticeable to the eye. In addition, the mapping range corresponds directly to the reflectivity dynamic range displayed in the image and thus determines the image contrast. In this thesis, the mapping range was chosen to match the dynamic range of the detected signals (-47 dB). 62 Figure 4-2(A) in the next page depicts a vertically-expanded image of the macular region. The layered structure of the retina visualized in the OFDI image correlates well with previously published OCT images and histological findings [14, 39, 40, 58]. The center of the fovea appears as a highly-reflective spot in the center of the image, at the top of the retina. The thick dark band at the top left and right of the image is the retinal nerve fiber layer (RNFL), which becomes thinner closer to the fovea. The layer below the RNFL is the inner plexiform layer (IPL) that grows very thick near the fovea. The two white bands underneath the IPL are the inner and outer nuclear layers (INL and ONL), respectively, sandwiching a darker band called the outer plexiform layer (OPL). Further down, the dark line that rises directly below the center of the fovea is the interface between the inner and outer segments of the photoreceptor layer (IPRL). The bottom layer is the retinal pigment epithelium (RPE). Finally, the choroid (C) lies below the RPE. 4.3 Comparison to an 840-nm SD-OCT System To assess the penetration power of the OFDI system, three-dimensional imaging was performed on the two volunteers A and B with both the OFDI system and a state-of-theart SD-OCT system previously developed for video-rate retinal imaging [35]. The SD system employed a superluminescent diode with a center wavelength of 840 nm and a 3dB spectral bandwidth of 50 nm, offering an axial resolution of 8 - 9 pm in air. At an Aline rate of 29 kHz and a sample arm power level of 600 p.W, the SD system offered a peak sensitivity of 98 dB at zero delay that decreased to 82 dB at the maximum ranging depth of 2.4 mm in air. In comparison, the OFDI system offered resolution of 14 - 16 pm and sensitivity of 89 - 95 dB over a similar depth range, after accounting for the additional water absorption at 1050 nm. Figure 4-2 represents side-by-side comparison of OFDI and SD-OCT images near the fovea and optic disc. Figures Al - A4 were obtained from a 36-year-old Asian male, and Figures B 1 - B2 were from a 41-year-old Caucasian male. Evidently, the OFDI images exhibit considerably deeper penetration into the choroid compared to the SD-OCT 63 In particular, penetration into the choroid is generally much shorter in the images. macular region than near the optic disc due to the macula's high melanin content [87], a phenomenon observed in the SD-OCT image (Figure A3). In contrast, the OFDI image (Figure Al) shows deep penetration across the entire retina. On the other hand, the higher axial resolution in the SD-OCT images provides better contrast between retinal layers. The apparently superior penetration of the OFDI system to the SD system is attributed to the lower absorption and scattering in RPE at 1050 nm than 840 nm [58] as well as the higher sensitivity of OFDI at large depth. However, verifying this hypothesis would require detailed analysis involving more human subjects and imaging systems. B1 B2 Figure 4-2. Comparison of two imaging systems (OFDI at 1050 nm and SD-OCT at 840 nm). Al and A2: OFDI images at fovea and optic nerve head, respectively, from volunteer A, 36-year-old Asian male. A3 and A4: SD-OCT images from the same volunteer at similar tissue locations. B1 and B2: OFDI and SD-OCT images, respectively, obtained from volunteer B, 41-year-old Caucasian male. OFDI images exhibit considerably deeper penetration in tissue than SD-OCT images in all the data sets. The OFDI image (Al) shows the anatomical layered structure: RNFL; retinal nerve fiber layer, IPL; inner plexiform layer, INL; inner nuclear layer, OPL; outer plexiform layer, ONL; outer nuclear layer, IPRL; interface between the inner and outer segments of the photoreceptor layer, RPE; retinal pigmented epithelium, and C; choriocapillaris and choroid. 64 65 ... on grad school ... "A gradstudent in procrastinationtend to stay in procrastination unless an externalforce is applied to it. -- phdcomics.com 66 Chapter 5: Visualizing Retinal and Choroidal Vasculature 5.1 Introduction A primary obstacle to the widespread acceptance of OCT is the difficulty for an ophthalmologist to follow and interpret video-rate OCT movie sequences of crosssectional images. Ophthalmologists are familiar with transverse fundus images and have access to large fundus image databases for various diseases. Precise spatial registration of OCT structures to fundus landmarks can capitalize on this knowledge to facilitate the analysis of OCT data. For instance, a fundus image can be shown as a still picture beside an OCT movie, with a line across the fundus image indicating the transverse position of the OCT cross-sectional scan along the retinal surface [88, 89]. This technique helps emphasize the relation between the fundus image and the underlying 3D OCT data set. The problem of spatial registration has been tackled in different ways. A basic approach is to separately acquire OCT and fundus images, and perform image registration using prominent features of the fundus image such as the retinal vasculature and the optic disc. Precision with this approach is unsatisfactory due to its inherent differences in image acquisition in terms of time and retina orientation. Another method is to combine OCT with the capability of a scanning laser ophthalmoscope (SLO). A TD-OCT system was demonstrated that acquired images transversely, one depth at a time [90]. This system split reflected sample light into two detection channels to produce transverse OCT and fundus images simultaneously. The system's pixel-to-pixel correspondence enabled precise spatial registration, but required a complicated setup and suffered sensitivity loss due to the splitting of reflected light. A simpler method is to employ a transverse-mode OCT system and construct a fundus-type image by numerical integration of transverse images along the depth axis [91]. Recently, Jiao et al [88] developed a technique with SD-OCT to acquire fundus-type images. Since the fundus-type image was constructed from the same raw spectra used to generate OCT data, the problem of spatial registration was solved, and both fundus-type and cross-sectional OCT images could be displayed in 67 real time. Mujat et al [89] demonstrated similar fundus-type images with SD-OCT by integrating cross-sectional images along the depth profile, which is equivalent to Jiao's method by Parseval's Theorem. Their comparisons to images from SLO [88, 89] and fluorescein angiography [89] indicated that these fundus-type images could be used as a reliable representation of the retinal vasculature. Jiao et al [88] also showed that image contrast of fundus-type images could be further enhanced by integrating only selective regions in cross-sectional images, based on anatomical structures. In this thesis, this technique has been extended to produce depthsectioned fundus images. For the first time, the choroidal vasculature was clearly visualized in a noninvasive manner. 5.2 Automatic Depth-Sectioning Algorithm The study of image segmentation, or partitioning of an image into selective regions, has a long history in the fields of computer vision and image processing. With OCT, segmentation algorithms have been applied to retinal imaging for estimating thicknesses of various retinal layers [92, 93]. For this thesis, an automatic algorithm similar to Mujat's method [89] was developed. Utilizing MATLAB's image processing tools [94], the algorithm can segment retinal layers by boundary detection in cross-sectional frames and construct depth-sectioned fundus-type images from an OCT movie sequence. The algorithm was applied to areas away from the optic disc, because the optic disc is difficult to track and does not provide relevant depth-resolved information of the vasculature. The algorithm operates as follows: Image Segmentation: 1. Convert a cross-sectional frame (Figure 5-la) into a binary frame of edges (Figure 5-1b) with the MA TLAB function edge. This function performs edge detection by gradient calculation and binary conversion with a threshold value. In a typical retinal image, the top boundary of the retinal nerve fiber layer (RNFL) 68 and the interface between the inner and outer segments of the photoreceptor layer (IPRL) appear as two prominent boundaries and serve as a good starting point for segmentation. The two boundaries often contain holes due to image noise or shadow of structures above (e.g. retinal vessels). On the other hand, additional edges may appear elsewhere. The threshold in the edge function represents a tradeoff between holes and additional edges. 2. Remove holes and additional edges (Figure 5-1c) by performing morphological operations on binary images with MATLAB functions bwmorph('bridge') and bwareaopen. The bwmorph('bridge') function removes holes by bridging continuous boundaries with short broken sections. Then the bwareaopen function eliminates additional edges by removing short isolated sections. 3. Trace the IPRL by tracking its detected boundary across the image one pixel at a time. Even after Steps 1 - 2, holes and additional edges are often still present. They are identified with sharp discontinuities in the boundary, since the IRPL is slowly-varying in depth. When a hole or an additional edge is detected, the tracking algorithm corrects for it by referring back to the previous value in the traced boundary or the value from the last traced frame at the same transverse location. 4. Smooth the traced boundary with the MATLAB function sgolayfilt. The presence of holes and additional edges can sometimes distort some part of the traced boundary. The Savitzky-Golay smoothing filter [94, 95] can minimize such distortion effect with its generalized moving average algorithm. Selective Integration: 5. Draw a parallel line to the traced IPRL to mark the RPE (Figure 5-1d). Because the curvatures of retinal layers below the IPRL are similar across the region of interest, the RPE can be approximated by a parallel line to the IPRL. The boundaries used for integration in the choroid region are also created this way. 69 6. Integrate pixel values along the depth in the region defined by the two boundaries (Figure 5-1d). This projects the 2D region into a ID data set (Figure 5-le) for each frame. The integration is performed in the logarithmic scale to yield a smoother fundus-type image [89]. Fundus-Type Image Production: 7. Repeat steps 1 - 6for all frames in the movie sequence to produce afundus-type image (Figure 5-1e). The 1D data set from each frame is displayed as a line in the 2D fundus-type image of the retinal vasculature. The red line corresponds to the result of integration (Figure 5-1d) in the selected region shown in Figure 5-1c. The ditch in Figure 5-1d indeed shows up as a dark segment (vessel) in the fundus-type image. 8. Correct for motion artifact by correlation. Discontinuities in the retinal vasculature, resulting from eye movements between frames, can be readily observed in the fundus-type image (Figure 5-le). The prominent retinal vasculature lends itself to an automatic image realignment procedure. Each transverse line of the fundus-type image is compared to the next, and the offset (if any) can be found from correlation. It should be noted that the offset information can potentially be used to realign the cross-sectional movie sequence. The corrected image (Figure 5-1f) is slightly narrower than Figure 5-le, because transverse realignment results in the need for removal of part of the image. The visible intensity gradient in the vertical direction in Figure 5-le has also been corrected. 9. The depth-sectioned fundus image has been produced (Figure 5-1f). 70 b) a) f d) C) - ',~ -~ IPRL RPE --- Integrated Signal Figure 5-1. Automatic depth-sectioning algorithm. (a) Original cross-sectional image frame. (b) Binary frame of edges (step 1). (c) Binary frame after removal of holes and additional edges (step 2). (d) Original cross-sectional image frame with traced boundaries of IPRL and RPE (steps 3 - 5). The integrated signal from the selected integration region is also shown (step 6). (e) Resultant fundustype image from repeating steps 1 - 6 for all frames in the movie sequence (step 7). (f) Final fundustype image after correction for motion artifact (step 8). The corrected image is slightly narrower, because transverse realignment results in the need for removal of part of the image. 71 5.3 Fundus-Type Images Given three-dimensional tomographic data of the eye's posterior segment, integrating the pixel values along the entire depth axis readily produces a two-dimensional fundus-type reflectivity image [88, 89]. Figure 5-2(A) depicts a fundus-type image generated from the entire OFDI image sequence for volunteer A. The image visualizes the optic nerve head, fovea, retinal vessels, and the faint outline of the deep choroidal vasculature; however, the depth information is completely lost. To overcome this limitation of the conventional method, one can integrate only selective regions based on anatomical structures. Seltivo integradon _E Figure 5-2. The retinal and choroidal vasculature extracted from the three-dimensional OFDI data set of volunteer A. (A) Two-dimensional reflectivity image (5.3 x 5.2 mm 2 ) obtained with the conventional full-range integration method. Higher (lower) reflectivity is represented by white (black) in the grayscale. (B) Illustration of the depth-sectioning integration method, with the different integration regions labeled C, D, E corresponding to the following fundus-type reflectivity images, respectively: (C) retinal reflectivity image showing the shadow of retinal vasculature (3.8 x 5.2 mm 2 ), (D) reflectivity image obtained from the upper part of the choroid, and (E) reflectivity image from the center of the choroid revealing the choroidal vasculature. Shadows of retinal vasculature are also visible in D and E. Scale bars: 0.5 mm. 72 For example, to visualize the retinal vasculature selectively with improved contrast, the aforementioned automatic algorithm was applied to integrate reflectivity in a selected region. A retinal blood vessel produces a large reflection by strong scattering and casts a shadow in the layers below. Integrating over the entire retina including vessels and their shadows often results in a lower contrast in the vasculature. Therefore, integration was performed in the region between IPRL and RPE (marked by red lines and labeled C in Figure 5-2B), where the shadows created by the retinal vessels above appear most distinctly [88]. Figure 5-2(C) depicts the fundus-type image (shadow) of the retinal vessels produced with this method. In this thesis, this method has been extended to produce depth-sectioned fundus-type images of the choriocapillary layer and the choroidal vasculature. The choriocapillary layer contains abundant small blood vessels and pigment cells [62, 96]. It is visualized (Figure 5-2D) using a thin integration region in the upper part of choroid (labeled D in Figure 5-2B). To obtain a fundus-type image of the complete choroid region, the bottom integration region (marked by blue lines and labeled E in Figure 5-2B) was used. In contrast to retinal vessels, choroidal vessels appear as low-signal (white in inverse grayscale) regions compared to their surroundings in the cross-sectional image. The reason for this phenomenon has not been completely understood yet. In the resulting fundus-type image, the choroidal vasculature (dark from integrating over low-signal regions) is clearly visualized (Figure 5-2E). Fundus-type images with similar qualities were obtained for volunteer B. During the last decade, an increasing number of retinal specialists have adopted indocyanine green (ICG) angiography [4] to visualize the choroidal vasculature for disease diagnosis. It is because fundus cameras [2] and scanning laser ophthalmoscopes (SLOs) [6] do not have access to the choroid otherwise, except for subjects with a low level of pigmentation. The above in vivo fundus-type image of human choroidal vasculature (Figure 5-2E) was created from the three-dimensional data set acquired by the 1050-nm OFDI system. In Figure 5-3, the choroid image (with selective integration), along with the image produced from the entire 3D data set (without selective integration), 73 is compared to a fundus image of the same eye obtained with a state-of-the-art commercial SLO (Heidelberg Engineering - HRA 2; laser wavelength = 820 nm; resolution ~ 5 pm; power = 46 p.W; speed = 5 Hz). This shows that the OFDI fundustype images are reliable representation of the retinal vasculature. Furthermore, the choroidal vasculature, which is not visible in the SLO image, is clearly visualized with the OFDI depth-sectioning method. The method does not require dilation or intravenous injection of fluorescent dyes. The entire imaging session took only approximately 10 seconds, while ICG imaging usually required up to 30 minutes, excluding the time for dilation and angiography. In addition to offering clinical convenience, the new non- invasive method would enable clinicians to image the choroidal vasculature of a larger patient population, some of whom might be allergic to dyes. The high-speed OFDI machine, now capable of high-resolution cross-sectional imaging as well as fundus imaging of retinal and choroidal vasculature, makes an ideal candidate for a one-stop clinical system for a wide range of ophthalmic applications. OFDI SLO Figure 5-3. A comparison of OFDI fundus-type images (A,B) to a SLO image (C). Figure A is a reliable representation of the retinal vasculature and Figure B clearly visualizes the choroidal vasculature, which is not visible in the SLO image. Figures A and B are Figure 5-2A and Figure 5-2E rescaled to match Figure C. 74 75 ... on life after grad school "WIELL ... we'll see." -- 76 Edward Chin Wang Lee Chapter 6: Summary and Discussion This thesis introduces a new ophthalmic technology that allows for comprehensive imaging of human retina, optic disc, and choroid in vivo. The technology's ability to produce high-resolution cross-sectional images and visualize retinal and choroidal vasculatures without angiography has been demonstrated. This final chapter presents a summary of the thesis and a discussion about the technology's future. 6.1 Summary From the first retinal photograph with the ophthalmoscope to the invention of the scanning laser ophthalmoscope (SLO), the field of retinal imaging has seen many exciting advances. Analogous to ultrasound in operation but based on interferometry for signal detection, optical coherence tomography (OCT) has emerged as a practical noninvasive technology that can produce a highly-accurate structural representation of the human retina in vivo with its high-resolution cross-sectional imaging. Optical frequency domain imaging (OFDI) is a second-generation method for OCT. It employs a wavelength-swept laser source to encode depth information into characteristic frequencies in the detected signal. Like spectral-domain (SD) OCT, OFDI has made significant improvements in imaging speed and sensitivity. Furthermore, OFDI offers additional benefits such as a long usable ranging depth and reduced sensitivity to motioninduced signal fading. However, while retinal imaging has been demonstrated for SDOCT with affordable spectrometers in the 800-nm region, this application has been out of reach for OFDI to date, due to the high water absorption in the 1300-nm region and the lack of a wide-tuning rapidly-swept light source in a low water-absorption window. Since the 1-pm region was shown to be a viable alternative window for retinal imaging and could even potentially offer deeper penetration into the choroid, the recent introduction of a commercial semiconductor optical amplifier in the 1 -pm region has opened a window of opportunities for OFDI. 77 Chapter 2 describes in detail the novel wavelength-swept laser developed for this thesis. The laser adopted a linear-cavity configuration and comprised a custom-built polygonbased intracavity scanning filter. With its output spectrum spanning 62 nm, the laser delivered 2.7 mW of average power at a sweep rate of 18.8 kHz. Chapter 3 explains the design and operation of the OFDI system constructed with this light source. The resultant system, with a sample arm power of 550 pW, achieved resolution of 10 pm in tissue and a peak sensitivity of 98 dB that dropped to 92 dB at the maximum depth of 2.4 mm. Compared to the previous time-domain system using a broadband source at 1040 nm [58], this OFDI system offers a higher sensitivity at a 100-fold faster image acquisition speed and using only one sixth of sample arm power. In Chapter 4, comprehensive in vivo imaging of human retina and choroid was demonstrated for two healthy volunteers. The movie sequence of images was recorded at a frame rate of 18.8 Hz for 10.6 seconds, with a thousand A-lines in each frame and a reflectivity dynamic range of 47 dB. The important steps in data processing included background subtraction, windowing, interpolation into linear k-space, and dispersion mismatch compensation. The resultant images clearly visualize the anatomical layers in the retina and correlate well with previously-published OCT images and histological findings. For the first time, posterior eye segment imaging has been realized with OFDI. Also for the first time, video-rate imaging has been demonstrated in the 1-pm region. The OFDI system further provides deeper choroid penetration in comparison to a stateof-the-art SD-OCT system at 840 nm. The improved penetration may be clinically useful for evaluating early stages of retinal pathologies, such as age-related macular degeneration, that are accompanied by choroidal neovascularization. The production of fundus-type images was examined next in Chapter 5. Previous studies showed that fundus-type images could be produced from a 3D OCT data set by integrating along the depth. These fundus-type images are helpful to retinal specialists, who can use them for precise spatial registration of OCT structures. It had also been shown that the prominent retinal vasculature in a fundus-type image could be enhanced in contrast by integrating only selective regions. This thesis has extended the technique to 78 produce depth-sectioned fundus-type images, with an automatic algorithm that was developed for selective integration based on anatomical structures. In combination with the system's enhanced penetration, the choroidal vasculature was, for the first time, clearly visualized in a noninvasive manner. This approach does not employ angiography and thus eliminates angiography's associated disadvantages such as a long measuring time, required dilation, and possible allergic consequences. 6.2 Discussion The unique characteristics of the 1050-nm OFDI system present challenges as well as opportunities in the future. Increasing the saturation power of the semiconductor optical amplifier (SOA) in the source and decreasing its relative intensity noise can drive the system sensitivity higher. Similarly, increasing the bandwidth of the SOA can result in higher resolution. However, in the 1-pm region, the increased water absorption away from the local minimum presents an obstacle to be overcome. Even in ultrahigh- resolution SD-OCT in the 800-nm window, the effective bandwidth in the eye can be limited due to increased absorption above 920 nm in the vitreous [39]. Yet in a positive light, with this thesis' demonstration of high-speed, high-sensitivity imaging in the 1050nm window and Lim et al's development of another high-speed swept source in the 800nm window [97], it is now possible to conceive a system that has a bandwidth spanning from the 800-nm window to the 1050-nm window. In such a Fourier-domain system (OFDI or SD-OCT), the time-domain fringes would be enveloped by the product of the source spectral shape and the water absorption spectral profile. Even though the fringes corresponding to the high-absorption region between the two windows would be smaller and noisier, the fringes could be normalized before Fourier Transform. The overall achievable resolution would still be higher than even that of the current ultrahighresolution systems. Nevertheless, all of this is contingent on technological advances in commercial SOAs. In the immediate future, source development would be the deciding factor between operation at 1050 nm (deep penetration and demonstrated high-quality images) and 800 nm (proven wavelength for clinical ophthalmic applications). 79 The powerful features of OFDI have profound implications for applications in posterior segment imaging. For example, the system can be readily adapted for polarization diversity and polarization-sensitive measurements. Polarization-sensitive measurements can quantify loss of retinal birefringence for early detection of glaucoma, but require two detection channels for the two different polarization states [21]. Thus, OFDI's use of simple photodetectors is appealing compared to SD-OCT's spectrometers. Besides this, OFDI's capability for phase-resolved measurements can enable the imaging of blood flow dynamics. Its reduced susceptibility to motion-induced fringe washout also allows measurements of flow rates much higher than possible with SD-OCT [19]. One can expect that a future phase-resolved OFDI system at 1050 nm can enhance the contrast of retinal and choroidal vessels in both cross-sectional images and fundus-type images. Furthermore, while fundus-type images cannot show a dye's filling of or leakage from vessels as in angiography, a phase-resolved system can produce depth-sectioned fundustype images that illustrate the retinal blood flow dynamics. With several thousand units of commercial TD-OCT systems in operation, the OFDI system developed for this thesis offers a glimpse into the future of second-generation OCT systems that might reside in the majority of ophthalmic clinics worldwide. The 1050-nm OFDI machine, with its high peak sensitivity and slow dropoff in addition to a deep penetration power, could potentially become the de facto platform for routine eye examinations. Compared to the system's current level of retinal exposure, the ANSI maximum permissible exposure allows for an increase by a factor of three, which would translate to a three-fold speed improvement at the same high sensitivity. This means that a complete three-dimensional data set of a patient's retina that includes information on retinal and choroidal vasculature could be obtained in 3 - 4 seconds without eye dilation or angiography. OFDI's multifunctional capability such as Doppler blood flow imaging and polarization-sensitive glaucoma diagnostics could further provide a universal platform for diverse imaging applications. With its demonstrated performance and impressive potential, OFDI is a promising technology for comprehensive ophthalmic examination. 80 References [1] C. Web, "The eye: information about vision loss and blindness," (2005), http://www.99main.com/~charlief/Blindness.htm. [2] P. J. Saine, and M. E. Tyler, Ophthalmic Photography:Retinal Photography,Angiography, and Electronic Imaging (Butterworth-Heinemann, Woburn, MA, 2002). [3] S. Wolf, F. Jung, H. Kiesewetter, N. Korber, and M. Reim, "Video Fluorescein Angiography Method and Clinical-Application," Graefes Archive for Clinical and Experimental Ophthalmology 227, 145-151 (1989). [4] J. J. Jorzik, A. Bindewald, S. Dithmar, and F. G. Holz, "Digital simultaneous fluorescein and indocyanine green angiography, autofluorescence, and red-free imaging with a solid-state laserbased confocal scanning laser ophthalmoscope," Retina-the Journal of Retinal and Vitreous Diseases 25, 405-416 (2005). [5] R. H. Webb, and G. W. Hughes, "Scanning Laser Ophthalmoscope," Ieee Transactions on Biomedical Engineering 28, 488-492 (1981). [6] R. H. Webb, G. W. Hughes, and F. C. Delori, "Confocal Scanning Laser Ophthalmoscope," Applied Optics 26, 1492-1499 (1987). [7] T. Olsen, "The Accuracy of Ultrasonic Determination of Axial Length in Pseudophakic Eyes," Acta Ophthalmologica 67, 141-144 (1989). [8] C. J. Pavlin, K. Harasiewicz, M. D. Sherar, and E. S. Foster, "Clinical Use of Ultrasound Biomicroscopy," Ophthalmology 98, 287-295 (1991). [9] D. F. Chang, General Ophthalmology (Appleton & Lange, Norwalk, Conn, 1992). [10] T. Seiler, and T. Bende, Noninvasive Diagnostic Techniques in Ophthalmology(Springer-Verlag NY Inc, New York, NY, 1990). [11] D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, "Optical Coherence Tomography," Science 254, 1178-1181 (1991). [12] A. F. Fercher, C. K. Hitzenberger, W. Drexler, G. Kamp, and H. Sattmann, "In-Vivo Optical Coherence Tomography," American Journal of Ophthalmology 116, 113-115 (1993). [13] E. A. Swanson, J. A. Izatt, M. R. Hee, D. Huang, C. P. Lin, J. S. Schuman, C. A. Puliafito, and J. G. Fujimoto, "In-Vivo Retinal Imaging by Optical Coherence Tomography," Optics Letters 18, 1864-1866 (1993). [14] Uniformed Services University of the Health Sciences, "HistoWeb," (2000), http://bicmra.usuhs.mi/M RA/Histology/Eye/HTMLdocs/eye05.html. [15] S. Yazdanfar, A. M. Rollins, and J. A. Izatt, "Imaging and velocimetry of the human retinal circulation with color Doppler optical coherence tomography," Optics Letters 25, 1448-1450 (2000). 81 [16] J. F. de Boer, B. Cense, N. Nassif, B. R. White, B. H. Park, M. C. Pierce, G. J. Tearney, B. E. Bouma, and T. C. Chen, "Ultra-high speed and ultra-high resolution optical coherence tomography and optical Doppler tomography," Investigative Ophthalmology & Visual Science 46 (2005). [17] M. R. Hee, C. A. Puliafito, C. Wong, J. S. Duker, E. Reichel, B. Rutledge, J. S. Schuman, E. A. Swanson, and J. G. Fujimoto, "Quantitative Assessment of Macular Edema with Optical Coherence Tomography," Archives of Ophthalmology 113, 1019-1029 (1995). [18] J. S. Schuman, M. R. Hee, C. A. Puliafito, C. Wong, T. Pedutkloizman, C. P. Lin, E. Hertzmark, J. A. Izatt, E. A. Swanson, and J. G. Fujimoto, "Quantification of Nerve-Fiber Layer Thickness in Normal and Glaucomatous Eyes Using Optical Coherence Tomography - a Pilot-Study," Archives of Ophthalmology 113, 586-596 (1995). [19] B. J. Vakoc, S. H. Yun, J. F. de Boer, G. J. Tearney, and B. E. Bouma, "Phase-resolved optical frequency domain imaging," Optics Express 13, 5483-5493 (2005), http://www.opticsexpress.org/abstract.cfm?id=84914. [20] D. J. Faber, E. G. Mik, M. C. G. Aalders, and T. G. van Leeuwen, "Light absorption of (oxy)hemoglobin assessed by spectroscopic optical coherence tomography," Optics Letters 28, 14361438 (2003). [21] B. Cense, H. C. Chen, B. H. Park, M. C. Pierce, and J. F. de Boer, "In vivo birefringence and thickness measurements of the human retinal nerve fiber layer using polarization-sensitive optical coherence tomography," Journal of Biomedical Optics 9, 121-125 (2004). [22] W. Drexler, U. Morgner, R. K. Ghanta, F. X. Kartner, J. S. Schuman, and J. G. Fujimoto, "Ultrahigh-resolution ophthalmic optical coherence tomography," Nature Medicine 7, 502-507 (2001). [23] W. Goebel, and T. Kretzchmar-Gross, "Retinal thickness in diabetic retinopathy - A study using optical coherence tomography (OCT)," Retina-the Journal of Retinal and Vitreous Diseases 22, 759-767 (2002). [24] M. R. Hee, C. R. Baumal, C. A. Puliafito, J. S. Duker, E. Reichel, J. R. Wilkins, J. G. Coker, J. S. Schuman, E. A. Swanson, and J. G. Fujimoto, "Optical coherence tomography of age-related macular degeneration and choroidal neovascularization," Ophthalmology 103, 1260-1270 (1996). [25] M. R. Hee, C. A. Puliafito, C. Wong, J. S. Duker, E. Reichel, J. S. Schuman, E. A. Swanson, and J. G. Fujimoto, "Optical Coherence Tomography of Macular Holes," Ophthalmology 102, 748-756 (1995). [26] M. R. Hee, C. A. Puliafito, C. Wong, E. Reichel, J. S. Duker, J. S. Schuman, E. A. Swanson, and J. G. Fujimoto, "Optical Coherence Tomography of Central Serous Chorioretinopathy," American Journal of Ophthalmology 120, 65-74 (1995). [27] J. R. Wilkins, C. A. Puliafito, M. R. Hee, J. S. Duker, E. Reichel, J. G. Coker, J. S. Schuman, E. A. Swanson, and J. G. Fujimoto, "Characterization of epiretinal membranes using optical coherence tomography," Ophthalmology 103, 2142-2151 (1996). [28] B. K. Rutledge, C. A. Puliafito, J. S. Duker, M. R. Hee, and M. S. Cox, "Optical coherence tomography of macular lesions associated with optic nerve head pits," Ophthalmology 103, 10471053 (1996). [29] B. E. Bouma, and G. J. Tearney, Handbook of Optical Coherence Tomography (Dekker, New York, NY, 2001). 82 [30] K. Takada, I. Yokohama, K. Chida, and J. Noda, "New Measurement System for Fault Location in Optical Wave-Guide Devices Based on an Interferometric-Technique," Applied Optics 26, 16031606 (1987). [31] R. C. Youngquist, S. Carr, and D. E. N. Davies, "Optical Coherence-Domain Reflectometry - a New Optical Evaluation Technique," Optics Letters 12, 158-160 (1987). [32] B. Hermann, E. J. Fernandez, A. Unterhuber, H. Sattmann, A. F. Fercher, W. Drexler, P. M. Prieto, and P. Artal, "Adaptive-optics ultrahigh-resolution optical coherence tomography," Optics Letters 29, 2142-2144 (2004). [33] A. F. Fercher, K. Mengedoht, and W. Werner, "Eye-Length Measurement by Interferometry with Partially Coherent-Light," Optics Letters 13, 186-188 (1988). [34] D. Huang, J. P. Wang, C. P. Lin, C. A. Puliafito, and J. G. Fujimoto, "Micron-Resolution Ranging of Cornea Anterior-Chamber by Optical Reflectometry," Lasers in Surgery and Medicine 11, 419425 (1991). [35] N. A. Nassif, B. Cense, B. H. Park, M. C. Pierce, S. H. Yun, B. E. Bouma, G. J. Tearney, T. C. Chen, and J. F. de Boer, "In vivo high-resolution video-rate spectral-domain optical coherence tomography of the human retina and optic nerve," Optics Express 12, 367-376 (2004), http://www.opticsexpress.org/abstract.cfi?id=78765. [36] M. Wojtkowski, R. Leitgeb, A. Kowalczyk, T. Bajraszewski, and A. F. Fercher, "In vivo human retinal imaging by Fourier domain optical coherence tomography," Journal of Biomedical Optics 7, 457-463 (2002). [37] M. A. Choma, M. V. Sarunic, C. H. Yang, and J. A. Izatt, "Sensitivity advantage of swept source and Fourier domain optical coherence tomography," Optics Express 11, 2183-2189 (2003), http://www.opticsexpress.orgabstract.cfm?id=78787. [38] J. F. de Boer, B. Cense, B. H. Park, M. C. Pierce, G. J. Tearney, and B. E. Bouma, "Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography," Optics Letters 28, 2067-2069 (2003). [39] B. Cense, N. A. Nassif, T. C. Chen, M. C. Pierce, S. H. Yun, B. H. Park, B. E. Bouma, G. J. Tearney, and J. F. de Boer, "Ultrahigh-resolution high-speed retinal imaging using spectraldomain optical coherence tomography," Optics Express 12, 2435-2447 (2004), http://www.opticsexpress.org/abstract.cfm?id=80 150. [40] M. Wojtkowski, V. J. Srinivasan, T. H. Ko, J. G. Fujimoto, A. Kowalczyk, and J. S. Duker, "Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation," Optics Express 12, 2404-2422 (2004), http://www.opticsexpress.or/abstract.cfm?id=80147. [41] S. H. Yun, G. J. Teamey, J. F. de Boer, N. Iftimia, and B. E. Bouma, "High-speed optical frequency-domain imaging," Optics Express 11, 2953-2963 (2003), http://www.opticsexpress.org/abstract.cfm?id=77825. [42] H. Barfuss, and E. Brinkmeyer, "Modified Optical Frequency-Domain Reflectometry with High Spatial-Resolution for Components of Integrated Optic Systems," Journal of Lightwave Technology 7, 3-10 (1989). [43] W. Eickhoff, and R. Ulrich, "Optical Frequency-Domain Reflectometry in Single-Mode Fiber," Applied Physics Letters 39, 693-695 (1981). 83 [44] S. H. Yun, G. J. Tearney, J. F. de Boer, and B. E. Bouma, "Motion artifacts in optical coherence tomography with frequency-domain ranging," Optics Express 12, 2977-2998 (2004), http://www.opticsexpress.org/abstract.cfmn?id=80320. [45] S. H. Yun, G. J. Tearney, J. F. de Boer, M. Shishkov, W. Y. Oh, and B. E. Bouma, "Catheterbased optical frequency domain imaging at 36 frames per second," in Coherence Domain Optical Methods and Optical Coherence Tomography in Biomedicine IX, 5690-16(San Jose, CA, 2005). [46] E. Brinkmeyer, and R. Ulrich, "High-Resolution Ocdr in Dispersive Wave-Guides," Electronics Letters 26, 413-414 (1990). [47] S. R. Chinn, E. A. Swanson, and J. G. Fujimoto, "Optical coherence tomography using a frequency-tunable optical source," Optics Letters 22, 340-342 (1997). [48] F. D. Nielsen, L. Thrane, J. Black, K. Hsu, A. Bjarklev, and P. E. Andersen, "Swept-Wavelength Source for Optical Coherence Tomography in the 1 pm Range," in European Congress on Biomedical Optics (ECBO)(Munich,Germany, 12-16 June 2005). [49] M. V. Shramenko, E. V. Andreeva, D. S. Mamedov, V. R. Shidlovski, and S. D. Yakubovich, "NIR semiconductor laser with fast broadband tuning," in Coherence Domain OpticalMethods and Optical Coherence Tomography in Biomedicine X, 6079-42(2006). [50] A. V. Oppenheim, R. W. Schafer, and J. R. Buck, Discrete-Time Signal Processing(Prentice Hall, New Jersey, 1999). [51] R. Leitgeb, C. K. Hitzenberger, and A. F. Fercher, "Performance of fourier domain vs. time domain optical coherence tomography," Optics Express 11, 889-894 (2003), http://www.opticsexpress.org/abstract.cfm?id=71990. [52] M. Wojtkowski, A. Kowalczyk, R. Leitgeb, and A. F. Fercher, "Full range complex spectral optical coherence tomography technique in eye imaging," Optics Letters 27, 1415-1417 (2002). [53] S. H. Yun, G. J. Tearney, J. F. de Boer, and B. E. Bouma, "Removing the depth-degeneracy in optical frequency domain imaging with frequency shifting," Optics Express 12, 4822-4828 (2004), http://www.opticsexpress.org/abstract.cfin?id=81308. [54] C. Dorrer, N. Belabas, J. P. Likforman, and M. Joffre, "Spectral resolution and sampling issues in Fourier-transform spectral interferometry," Journal of the Optical Society of America B-Optical Physics 17, 1795-1802 (2000). [55] D. J. Segelstein, The complex refractive index of water (University of Missouri-Kansas City, 1981). [56] M. A. Choma, K. Hsu, and J. A. Izatt, "Swept source optical coherence tomography using an allfiber 1300-nm ring laser source," Journal of Biomedical Optics 10 (2005). [57] Y. Yasuno, V. D. Madjarova, S. Makita, M. Akiba, A. Morosawa, C. Chong, T. Sakai, K. P. Chan, M. Itoh, and T. Yatagai, "Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments," Optics Express 13, 1065210664 (2005), http://www.opticsexpress.org/abstract.cfm?id=86669. [58] A. Unterhuber, B. Povazay, B. Hermann, H. Sattmann, A. Chavez-Pirson, and W. Drexler, "In vivo retinal optical coherence tomography at 1040 nm-enhanced penetration into the choroid," Optics Express 13, 3252-3258 (2005), http://www.opticsexpress.org/abstract.cfm?id=83715. 84 [59] S. Bourquin, A. D. Aguirre, I. Hart, P. Hsiung, T. H. Ko, J. G. Fujimoto, T. A. Birks, W. J. Wadsworth, U. Bunting, and D. Kopf, "Ultrahigh resolution real time OCT imaging using a compact femtosecond Nd : Glass laser and nonlinear fiber," Optics Express 11, 3290-3297 (2003), http://www.opticsexpress.org/abstract.cfm?id=7809 1. [60] H. Lim, Y. Jiang, Y. M. Wang, Y. C. Huang, Z. P. Chen, and F. W. Wise, "Ultrahigh-resolution optical coherence tomography with a fiber laser source at 1 pm," Optics Letters 30, 1171-1173 (2005). [61] American National Standards Institute, American nationalstandardforsafe use of lasersz136.1 (American National Standards Institute, Orlando, FL, 2000). [62] D. V. Alfaro, P. E. Liggett, W. F. Mieler, H. Quiroz-Mercado, R. D. Jager, and Y. Tano, Agerelatedmacular degneration:a comprehensive textbook (Lippincott Williams & Wilkins, Philadelphia, PA, 2006). [63] K. C. Harvey, and C. J. Myatt, "External-Cavity Diode-Laser Using a Grazing-Incidence Diffraction Grating," Optics Letters 16, 910-912 (1991). [64] B. Golubovic, B. E. Bouma, G. J. Tearney, and J. G. Fujimoto, "Optical frequency-domain reflectometry using rapid wavelength tuning of a Cr4+:forsterite laser," Optics Letters 22, 17041706 (1997). [65] S. H. Yun, D. J. Richardson, and B. Y. Kim, "Interrogation of fiber grating sensor arrays with a wavelength-swept fiber laser," Optics Letters 23, 843-845 (1998). [66] S. H. Yun, C. Boudoux, G. J. Tearney, and B. E. Bouma, "High-speed wavelength-swept semiconductor laser with a polygon-scanner-based wavelength filter," Optics Letters 28, 19811983 (2003). [67] B. Culshaw, "The optical fibre Sagnac interferometer: an overview of its principles and applications," Measurement Science & Technology 17, Rl-R16 (2006). [68] W. Y. Oh, S. H. Yun, B. J. Vakoc, G. J. Tearney, and B. E. Bouma, "Ultrahigh-speed optical frequency domain imaging and application to laser ablation monitoring," Applied Physics Letters 88, 103902 (2006). [69] K. Takada, K. Yukimatsu, M. Kobayashi, and J. Noda, "Rayleigh Backscattering Measurement of Single-Mode Fibers by Low Coherence Optical Time-Domain Reflectometer with 14-Mu-M Spatial-Resolution," Applied Physics Letters 59, 143-145 (1991). [70] J. A. C. Lee, and D. C. Munson, "Spatially variant apodization for image reconstruction from partial Fourier data," Ieee Transactions on Image Processing 9, 1914-1925 (2000). [71] F. J. Harris, "On the Use of Windows for Harmonic Analysis with the Discrete Fourier Transform," Proceedings of the IEEE 66, 51-84. [72] A. H. Nuttall, "Some Windows with Very Good Sidelobe Behavior," Ieee Transactions on Acoustics Speech and Signal Processing 29, 84-91 (1981). [73] B. E. Lemoff, and C. P. J. Barty, "Cubic-Phase-Free Dispersion Compensation in Solid-State Ultrashort-Pulse Lasers," Optics Letters 18, 57-59 (1993). [74] B. Bouma, G. J. Tearney, S. A. Boppart, M. R. Hee, M. E. Brezinski, and J. G. Fujimoto, "HighResolution Optical Coherence Tomographic Imaging Using a Mode-Locked Ti-Al2o3 Laser Source," Optics Letters 20, 1486-1488 (1995). 85 [75] J. F. de Boer, C. E. Saxer, and J. S. Nelson, "Stable carrier generation and phase-resolved digital data processing in optical coherence tomography," Applied Optics 40, 5787-5790 (2001). [76] A. F. Fercher, C. K. Hitzenberger, M. Sticker, R. Zawadzki, B. Karamata, and T. Lasser, "Dispersion compensation for optical coherence tomography depth-scan signals by a numerical technique," Optics Communications 204, 67-74 (2002). [77] C. A. Puliafito, M. R. Hee, J. S. Schuman, and J. G. Fujimoto, Optical coherence tomography of ocular diseases (SLACK Incorporated, Thorofare, NJ, 1996). [78] M. R. Hee, J. A. Izatt, E. A. Swanson, D. Huang, J. S. Schuman, C. P. Lin, C. A. Puliafito, and J. G. Fujimoto, "Optical Coherence Tomography of the Human Retina," Archives of Ophthalmology 113, 325-332 (1995). [79] D. Koozekanani, C. Roberts, S. E. Katz, and E. E. Herderick, "Intersession repeatability of macular thickness measurements with the Humphrey 2000 OCT," Investigative Ophthalmology & Visual Science 41, 1486-1491 (2000). [80] M. R. Hee, C. A. Puliafito, J. S. Duker, E. Reichel, J. G. Coker, J. R. Wilkins, J. S. Schuman, E. A. Swanson, and J. G. Fujimoto, "Topography of diabetic macular edema with optical coherence tomography," Ophthalmology 105, 360-370 (1998). [81] M. Baumann, R. C. Gentile, J. M. Liebmann, and R. Ritch, "Reproducibility of retinal thickness measurements in normal eyes using optical coherence tomography," Ophthalmic Surgery and Lasers 29, 280-285 (1998). [82] R. Huber, M. Wojtkowski, J. G. Fujimoto, J. Y. Jiang, and A. E. Cable, "Three-dimensional and C-mode OCT imaging with a compact, frequency swept laser source at 1300 nm," Optics Express 13, 10523-10538 (2005), http://www.opticsexpress.org/abstract.cfm?id=86656. [83] R. Huber, M. Wojtkowski, K. Taira, J. G. Fujimoto, and K. Hsu, "Amplified, frequency swept lasers for frequency domain reflectometry and OCT imaging: design and scaling principles," Optics Express 13, 3513-3528 (2005), http://www.opticsexpress.orgy/abstract.cfm?id=83745. [84] B. J. Vakoc, S. H. Yun, M. S. Shishkov, W. Oh, J. A. Evans, N. S. Nishioka, B. E. Bouma, and G. J. Teamey, "Comprehensive microscopy of the esophagus using optical frequency domain imaging," in Coherence Domain OpticalMethods and Optical Coherence Tomography in BiomedicineX, 6082-13(San Jose, CA, 2006). [85] J. Zhang, and Z. P. Chen, "In vivo blood flow imaging by a swept laser source based Fourier domain optical Doppler tomography," Optics Express 13, 7449-7457 (2005), http://www.opticsexpress.org/abstract.cfm?id=85477. [86] E. C. Lee, J. F. de Boer, M. Mujat, H. Lim, and S. H. Yun, "In vivo optical frequency domain imaging of human retina and choroid," Optics Express 14, 4403-4411 (2006), http://www.opticsinfobase.org/abstract.cfn?URI=oe- 14-10-4403. [87] R. Seth, and P. Gouras, "Assessing macular pigment from SLO images," Documenta Ophthalmologica 108, 197-202 (2004). [88] S. L. Jiao, R. Knighton, X. R. Huang, G. Gregori, and C. A. Puliafito, "Simultaneous acquisition of sectional and fundus ophthalmic images with spectral-domain optical coherence tomography," Optics Express 13, 444-452 (2005), http://www.opticsexpress.org/abstract.cfm?id=82381. 86 [89] M. Mujat, R. C. Chan, B. Cense, B. H. Park, C. Joo, T. Akkin, T. C. Chen, and J. F. de Boer, "Retinal nerve fiber layer thickness map determined from optical coherence tomography images," Optics Express 13, 9480-9491 (2005), http://www.opticsexpress.org/abstract.cfm?id=86235. [90] A. G. Podoleanu, G. M. Dobre, R. G. Cucu, R. Rosen, P. Garcia, J. Nieto, D. Will, R. Gentile, T. Muldoon, J. Walsh, L. A. Yannuzzi, Y. Fisher, D. Orlock, R. Weitz, J. A. Rogers, S. Dunne, and A. Boxer, "Combined multiplanar optical coherence tomography and confocal scanning ophthalmoscopy," Journal of Biomedical Optics 9, 86-93 (2004). [91] C. K. Hitzenberger, P. Trost, P. W. Lo, and Q. Y. Zhou, "Three-dimensional imaging of the human retina by high-speed optical coherence tomography," Optics Express 11, 2753-2761 (2003), http://www.opticsexpress.org/abstract.cfm?id=77668. [92] H. Ishikawa, D. M. Stein, G. Wollstein, S. Beaton, J. G. Fujimoto, and J. S. Schuman, "Macular segmentation with optical coherence tomography," Investigative Ophthalmology & Visual Science 46, 2012-2017 (2005). [93] D. Koozekanani, K. Boyer, and C. Roberts, "Retinal thickness measurements from optical coherence tomography using a Markov boundary model," Ieee Transactions on Medical Imaging 20, 900-916 (2001). [94] MATLAB, "MATLAB Documentation," (2006), http://www.mathworks.com/access/helpdesk/help/techdoc/matlab.html. [95] S. J. Orfanidis, Introduction to Signal Processing(Prentice-Hall, Englewood Cliffs, NJ, 1996). [96] V. Poukens, B. J. Glasgow, and J. L. Demer, "Nonvascular contractile cells in sclera and choroid of humans and monkeys," Investigative Ophthalmology & Visual Science 39, 1765-1774 (1998). [97] H. Lim, J. F. de Boer, B. H. Park, E. C. Lee, R. Yelin, and S. H. Yun, "Optical frequency domain imaging with a rapidly swept laser in the 815-870 nm range," Optics Express (submitted 2006). 87 Artistic Rendering of Choroidal Vasculature by CatherineBolliet