STIS Design Charles Proffitt

advertisement
STIS Design
Charles Proffitt
Outline of Topics







Introduction
Basic Structure of STIS
Detectors
MSM optical elements & observing modes
Slits and Apertures
Target Acquistions
Lamps and Wavecals
STIS Presentation
2
Introduction to STIS
Space Telescope Imaging Spectrograph
 Highly versatile spectrograph
 3 detectors (can use only one at a time)





Long slit 1st order spectra
High dispersion UV echelle spectra
•
•
FUV MAMA ~ 1150 - 1700 Å, 1024 x 1024, ~0.025” pixels
NUV MAMA ~ 1600 - 3200 Å , 1024 x 1024, ~0.025” pixels
CCD ~ 1650 - 11,000 Å , 1024 x 1024, ~ 0.05” pixels
Resolution  up to ~ 100,000
Also slit-less and imaging modes
STIS Presentation
3
STIS Operational History


Replaced GHRS in Axial Bay 1 on Feb 14, 1997 during SM2
STIS Side 1 failed on May 16, 2001




4.25 years and ~ 42,000 hours of operation
Probable short in tantalum capacitor - completely disabled side 1
Failure inaccessible without removing STIS
STIS Side 2 failed on August 3, 2004





3.25 years and ~ 27,000 hours of operation
Failure in Interpoint converter that supplies power to move
mechanisms
STIS other wise appears to be healthy, but no way to move the
mechanisms or get light to the detector
STIS now turned-off (except for heaters) to avoid applying power
to bad component
Repair planned for SM4; will replace LVPS2 board
STIS Presentation
4
STIS Optical Bench
STIS Presentation
5
STIS Presentation
6
STIS Mechanisms

Corrector/Focus mechanism




Echelle Blocker




Keeps scattered light from unused detector’s echelles
Mode Isolation Mechanism


Separately adjustable in tip, tilt, and focus
Last used during SMOV2 in 1997
Will adjust during SMOV4 only if necessary
Blocks direct light from MSM to MAMA unless desired
Calibration Insert Mechanism (CIM)
Slit Wheel
Mode Select Mechanism (MSM)
STIS Presentation
7
STIS Slit Wheel
• Slit wheel is in a image plane
• 77 standard slit wheel positions defined in
aperture table
• 45 distinct clear slits and apertures
• 7 ND slits
• 13 imaging filters
• 12 alternate rotations for barred apers
• Only a subset available for GOs
• For STIS the APERTURE keyword value refers to a
unique slit wheel pos
• Some physical slits have multiple slit wheel
APERTURE positions defined (e.g., barred and regular
long-slit pos)
• Alternate pointings of HST at the same slit wheel
position are captured in PROPAPER keyword (e.g., E1
aperture positions vs. regular positions).
• Sometimes called pseudo-aperture positions
• Usage differs from other instruments.
• In STIS slits & filters are not “optical elements” as far
as most ops data bases are concerned.
STIS Presentation
8
Mode Select Mechanism
•
Tilted cylinders allow MSM to rotate and
“wobble” to allow tip/tilt of elements
•
•
•
•
•
•
21 optical elements in MSM
•
•
16 gratings & 1 PRISM
4 imaging mirrors
10 elements have order sorter filters
For the 4 echelle modes, the crossdisperser is the element in the MSM
Each optical element is intended for use
with only 1 optical path & 1 detector
•
STIS Presentation
Rotation of large wheels more precise than tip-tilt of
small actuators
Repeatability still not perfect
Exceptions for echelle cross dispersers, but
exceptions not used on-orbit
9
STIS Detectors
MAMA Detectors

MAMA = Multi-Anode Microchannel Array




Photocathode produces electron when hit by individual photon
Microchannel plate turns electron into charge cloud (4 x 105 e-)
1K x 1K anode array detects and centroids charge cloud
1024 x 1024 pixels ~ 0.0245” pixel size - 25”x25” FOV


Can be subsampled to 2048 x 2048
Advantages of MAMA detectors


Good FUV and NUV sensitivity
Photon counting





Less sensitive to cosmic rays (just one more dark current count at worst)
Low dark current
More resistant to radiation damage than CCDs



no read-noise
Time-tag mode possible
(but maybe not completely immune to rad. damage)
No charge transfers (no CTI losses or tails)
High spatial resolution
STIS Presentation
11
MAMA Detectors

Dis-advantages of MAMA detectors




Subject to damage if over-illuminated
Cannot operate during SAA (high count rate)
Difficult to manufacture
STIS MAMA peculiarities

STIS MAMAs have optical isolators that scintillate from cosmic rays
 This forces STIS MAMA low voltage to be turned off during SAA
 Prevents STIS MAMAs from observing in any SAA impacted orbit
 HV only on for one ~ 5 - 6 orbit block per day

STIS NUV MAMA has high dark current due to long
phosphorescent window glow excited by charged particle impacts
STIS Presentation
12
FUV MAMA Detector
STIS Presentation
13
MAMA Anode Array



Pulse location positions are
centroided using anode grid
Data routinely sub-sampled
to 2048 x 2048 grid, but flat
fielding issues prevent extra
resolution from being useful
Amount of charge, number
of “folds”, and location used
to choose “valid” events.
STIS Presentation
14
MAMA Detectors

Micro channel plate consists of bundles of curved
glass tubes
1.
2.
3.
4.
5.
STIS Presentation
Photon hits photocathode, ejects electrons
Voltage across plate accelerates electrons down
tubes
Electrons collide with walls, eject more electrons
Average gain of 4 x 105 (electrons out per photon
event)
Size and location of charge cloud used to distinguish
valid events.
15
Differences Between MAMAs

FUV MAMA




CsI photocathode on Micro-channel Plate
Strongest response 1150-2000 Å
Field electrode & repeller wire between window and photocathode
NUV MAMA

CsTe2 photocathode on inside of detector window



Slight defocus from lateral drift of electrons
No repeller wire or field electrode
Strongest response 1700-3200 Å


STIS Presentation
Significant sensitivity down to 1150 Å
Intended as backup for FUV MAMA
• Unused backup modes to replicate FUV abilities
• much lower FUV throughput
16
MAMA Detectors
MAMAs detect individual photons

One event recorded per photon
Invalid events are discarded

ACCUM and TIME-TAG differ mostly in what gets saved


In ACCUM mode, each event increments memory location for
that pixel. Only final accumulated image saved.



Pixel locations shifted on-orbit for orbital doppler correction
In TIME-TAG mode, position and time of each valid event are
saved. Doppler correction done later.
If HV is on, MAMA tubes continue to operate even when
exposures are not in progress - events just aren’t recorded
STIS Presentation
17
MAMA Bright Object Limits

MAMAs can be damaged by excessive illumination



Extracting too much charge too quickly, could limit future charge
extraction, cause localized gain sag, decrease effective sensitivity
Very large count rates can produce gas in tube, perhaps leading to
catastrophic short circuit or gas venting to aft-shroud
CARD (Constraints and Restrictions Document) Limits



Detectors become non-linear at > 300,000 counts/s


Global > 1.5e6 counts for 1 second
Local > 500 counts/lo-res-pixel/s over 4x4 area for > 30 s
Science calibration difficult
Lower screening limits set for operations




STIS Presentation
Global limit 200,000 counts/sec for most modes
30,000 counts/sec for 1st order point sources
Local limit 75 (spec) or 100 (imaging) counts/lo-res-pixel/s
• Brightest pixel
Lower global limits (80,000/12,000) for irregularly variable objects
18
Automatic BOP Mechanisms

Bright Scene Detector (BSD)







Software Global Monitor (SGM)






Every 32nd anode wire monitored by special circuit
Unaffected by high count rates that may saturate normal counting electronics
Detector safed if this monitor is triggered
Trigger equivalent to uniform ~ 2x106 cnt/s global rate
Sparse coverage - bright source could fall between monitored rows
Spectrum at right location could safe detector with only ~ 120,000 counts/s
Uses event counters giving global rate to detect overlight
Monitors all counts (valid or invalid) above set threshold voltage
Triggers at 1x106 c/s (equivalent to ~ 580,000 valid c/s)
Affected by non-linearity at high counts rates (> 4x106 c/s)
Can shut down detector within 0.1 s
Local Rate Check Image (LRC)



300 ms image taken before each MAMA observation
Rebinned 2x2 and 4x4 and brightest pixels compared to limits
Failure of LRC causes detector to be shuttered & and any lamps to be turned off


STIS Presentation
Only shuttered image lost
Check only done at start of observation
19
STIS MAMA Detectors
STIS MAMA Detector Performance Characteristics
Characteristic
FUV-MAMA Performance
NUV-MAMA Performance
----------------------------------------------------------------------------------------------------------Photocathode
CsI
Cs2Te
Wavelength range
1150-1700 Å
1600-3100 Å
Pixel format
1024 x 1024
1024 x 1024
Pixel size
25 x 25 µm
25 x 25 µm
Image mode pixel plate scale ~0.0245” x 0.0247”
~0.0245” x 0.0248”
Field of view
25.1 x 25.3”
25.1 x 25.4”
Quantum efficiency
25% @ 1216 Å
10% @ 2537 Å
-6
-4
Dark count
5 x 10 to 1 x 10 c/s/pixel
8 x 10-4 to 1.7 x 10-3 c/s/pixel
Global count-rate linearity limit1
285,000 counts/s
285,000 counts/s
1
Local count-rate linearity limit
~220 counts/s/pixel
~340 counts/s/pixel
----------------------------------------------------------------------------------------------------------1Rate
at which counting shows 10% deviation from linearity.
These count rates are well above the bright-object limit.
STIS Presentation
20
STIS CCD




STIS CCD: 1024 x 1024 thinned backside illuminated SITe CCD
Thermal electric cooler (TEC) to allow CCD to operate at -83 C.
Includes overscan region to ease bias removal
CCD accumulates charge in each pixel and then is readout by
transferring charge row by row to readout register and then pixel by
pixel to amplifier
STIS Presentation
21
CCD Characteristics



ACCUM mode only - image read out after exposure
Can read out subarrays
Cosmic ray hits can affect numerous pixels


Some charge can lag during readout.




Often need to CR-SPLIT images
Charge Transfer Inefficiency (CTI)
Gets worse as CCD accumulates radiation damage on-orbit
Affects fluxes and causes tails in images
Thinned CCD has good UV sensitivity, but too
transparent in red - photons scatter in chip


Interference fringing in IR
Red light halo
STIS Presentation
22
STIS CCD
STIS CCD Detector Performance Characteristics
Characteristic
CCD Performance
-----------------------------------------------------------------------------------------------Architecture
Thinned, backside illuminated
Wavelength range
~1600-11,000 Å
Pixel format
1024 x 1024 illuminated pixels
Field of view
52 x 52 arcseconds
Pixel size
21 x 21 µm
Pixel plate scale
0.05071 arcseconds
Quantum efficiency
~ 20% @ 3000 Å
~ 67% @ 6000 Å
~ 29% @ 9000 Å
Dark count at -83° C
0.007 e-/s/pixel (but varies with detector T)
Read noise (effective values) 5.4 e- rms at GAIN=1 (1 e- of which is pattern noise)
7.6 e- rms at GAIN=4 (0.2 e- of which is pattern noise)
Full well
144,000 e- over the inner portion of the detector
120,000 e- over the outer portion of the detector
Saturation limit
33,000 e- at GAIN=1 (16 bit A-to-D limit)
144,000 e- at GAIN=4
------------------------------------------------------------------------------------------------
STIS Presentation
23
MSM Optical Elements
and Observing Modes
STIS Observing Modes

Echelle Modes



1st Order and PRISM Modes




E140M, E140H
E230M, E230H
G140L, G230L, G230LB, G430L, G750L
G140M, G230M, G230LM, G430M, G750M
NUV Prism
Imaging Modes

Imaging mirror for each detector


2 for FUV - clear and filtered
Acquistion modes


ACQ mode uses image to align target with
standard slit
ACQ/Peak modes center target in small slit
STIS Presentation
25
STIS Presentation
26
STIS Presentation
27
STIS Presentation
28
STIS Presentation
29
1st Order vs Echelle

1st order gratings (m=1)




Low blaze angle, fine ruling
Large free spectral range
Some use blocking filters to remove higher order
Echelle gratings (orders m = 66 - 747)






High blaze angle, coarser ruling
Use higher order spectra for much higher dispersion
Smaller free spectral range per order~ /m
Use cross-disperser to separate orders
Can image many orders on detector at once
Flux calibration more difficult
STIS Presentation
30
Scanning of gratings

For STIS L-modes, whole spectral range fits onto detector





G140L ~ 1150 - 1736 Å
G230L ~ 1570 - 3180 Å
G230LB ~ 1680 - 3065 Å
G430L ~ 2900 - 5700 Å
G750L ~ 5240 - 10,200 Å (larger  contaminated by 2nd order light)


Medium resolution 1st order
gratings, need to be scanned in
dispersion direction to cover
full spectral range
Echelle gratings, need to be
scanned in cross-dispersion
direction to cover all
wavelengths
STIS Presentation
31
Dispersion,  coverage, & throughput
STIS Presentation
32
Dispersion,  coverage, & throughput
STIS Presentation
33
G140M & G230M Central  Settings

Only pre-defined grating tilts allowed


Prime settings will scan whole range (~10% overlap)
Secondary settings for special purposes
STIS Presentation
34
G230MB, G430M, G750M Cenwaves

CCD M gratings, lower dispersion, but larger
coverage than MAMA M gratings
STIS Presentation
35
Use of 1st order mode with long slit

G750M observation of M84 (Radio Galaxy) nucleus




Bower et al. (1998), ApJL, 492, L111
Long slit (52X0.2) at 6581 CENWAVE across nucleus
Shows N II and S II emission lines from disk
Gives rotation curve with high resolution ~ 0.05”/pixel
STIS Presentation
36
Echelle mode CENWAVE settings

Many secondary settings to allow flexibility.


Used frequently by GOs
E140M covers ~ 1123 to 1710 Å with one setting
STIS Presentation
37
STIS Presentation
38
STIS Echelle Spectra
E140M LINE lamp spectrum
E140M Stellar Spectrum
STIS Presentation
39
Apertures and Filters
STIS Slit Wheel



Very large number of apertures
Slit name convention, e.g, 52X2
 “length in spatial direction” x “length along dispersion direction”
 F preceeds name of filtered apertures
Full list of slit wheel position names
0.05X29
0.09X29
0.2X29
0.05X31NDA
0.05X31NDB
F28X50LP
F28X50OII
F28X50OIII
52X0.1B0.5
52X0.1B1.0
52X0.1B3.0

F25CIII
F25CN182
F25CN270
F25LYA
F25MGII
F25ND3
F25ND5
F25NDQ
F25QTZ
F25SRF2
0.1X0.03
0.1X0.06
0.1X0.09
0.1X0.2
0.2X0.05ND
0.2X0.06
0.2X0.09
0.2X0.2
0.2X0.5
0.3X0.05ND
0.3X0.06
0.3X0.09
0.3X0.2
0.5X0.5
0.2X0.06FPA
0.2X0.06FPB
0.2X0.06FPC
0.2X0.06FPD
0.2X0.06FPE
0.2X0.2FPA
0.2X0.2FPB
0.2X0.2FPC
0.2X0.2FPD
0.2X0.2FPE
1X0.06
1X0.2
2X2
6X0.06
6X0.2
6X0.5
6X6
31X0.05NDA
31X0.05NDB
31X0.05NDC
36X0.05N45
36X0.05P45
36X0.6N45
36X0.6P45
50CCD
50CORON
52X0.05
52X0.05F1
52X0.05F2
52X0.1
52X0.1F1
52X0.1F2
52X0.2
52X0.2F1
52X0.2F2
52X0.5
52X0.5F1
52X0.5F2
52X2
52X2F1
52X2F2
Some wheel position “APERTURES” have multiple target positions defined for same slit
wheel setting. These are not included in this list.
STIS Presentation
41
Slits for 1st order spectroscopy

First order observations usually use 52” long
slits, 52X2, 52X0.5, 52X0.2, 52X0.1, 52X0.05



52X2 for best absolute photometry
Smaller slits for cleaner LSF or extended
targets
The 52” slits all have a pair of fiducial bars

Alternate slit rotations (rotate aperture wheel)
are defined to bring either fiducial bar closer to
center, and target behind bar
STIS Presentation
42
Barred aperture positions



Append F1 or F2 to aperture name (e.g. 52x0.2F1)
Rotation of aperture wheel gives different slit angle
Only 52X0.2F1 “supported”, but other bars “available”
STIS Presentation
43
Alternate Aperture Positions

For some apertures, multiple positions defined, but using same
aperture wheel rotation




For STIS these are often referred to as pseudo-apertures. In
image header in PROPAPER keyword
APERTURE keyword still set to name of physical aperture.
E1 apertures for lower CTI
D1 apertures for lower FUV dark current
STIS Presentation
44
STIS Aperture Selection, continued

Echelle observations usually use short slits


0.2X0.2 for best throughput & photometry
Smaller slits matched to each grating for better LSF




0.2X0.06 for E140M & E230M; 0.2x0.09 for E140H & E230H
Smallest slit 0.1X0.03 to maximize resolution
Small ND slits 0.2X0.05ND (100X) and 0.3X0.05ND (1000X)
FP-SPLIT slits (0.2X0.2FPA-E & 0.2X0.06FPA-E) to dither target
along dispersion direction - solve for fixed-pattern noise.
Dispersion
STIS Presentation
45
STIS Aperture Selection - cont

Most apertures can be used with most gratings



NUV-PRISM and 1st order gratings often used slitless
Can use long slits with echelle for spatially resolved
observations of emission line sources
Filters are also in aperture wheel and can be crossed
with gratings, (e.g., use long-pass filter to block geocoronal lyman-alpha in slitless G140L images)
STIS Presentation
46
Wide slits and extended sources



With wide slit, cannot separate spatial offsets in dispersion direction
from wavelength shifts
Wide slit observations of extended objects degrade spectral resolution
For emission line sources can take advantage of this to do emission
line images.
STIS G750M 6581 52 × 2 Spectral
Image of SN1987A. This shows
the images of the inner circumstellar ring
in [OI], Hα, [NII], and [SII]. Diffuse Hα
emission from the LMC fills the 52 × 2
slit, and broad Hα emission from the SN
is also visible. The continua of stars
produce the horizontal bands. The image
shown is a 950 × 450 subsection of the
1024 × 1024 image. (Figure courtesy of
Jason Pun and George Sonneborn, see
also Sonneborn et al. 1998, ApJ, 492,
L139).
STIS Presentation
47
MAMA Imaging

FUV Imaging



FUV imaging capabilities similar to ACS SBC
Solar blind, but red-leak may be worse than specs
NUV Imaging


Some sensitivity at FUV wavelengths
At long , overlaps with ACS/HRC and WFC3/UVIS abilities
STIS Presentation
48
STIS CCD Imaging Modes

CCD Imaging




Primarily needed for target ACQs, but also for science
Between SM2 and SM3B (ACS install), the unfiltered STIS CCD
was the most sensitive imaging instrument on HST
Only limited filters (OIII has significant red leak)
Also ND filters
STIS Presentation
49
Imaging Coronagraphy

Coronagraphic Mask in filter wheel for use with CCD imaging



Unfiltered coronagraphy only, cannot cross with other filters.
Predefined aperture locations, WEDGEA1.0, etc.
Imaging mode mirror has Lyot stop, but secondary and spider not
aphodized.
STIS Presentation
50
Download