Document 10552169

advertisement
Mini Kleenpak™ Syringe Filters with Supor® EAV Membrane
Description
Pall Supor® EAV filters are designed for effective bioburden and particle control. The Pall-patented Supor
machV membrane incorporated in these filters ensures high throughputs and flow rates when utilized for the
protection of buffers and biological process fluids.
Supor EAV filters allow for reduced sizing of filter systems with improved process efficiencies when use of
validated sterilizing grade filters is not essential, but reliable bioburden control is. Supor EAV filters are also
effective prefilters for protection and extended life of 0.2 µm sterilizing grade and finer membrane filters where
required.
The Mini Kleenpak capsule format is designed for use in the laboratory. As part of the Pall UpScale Programsm
of scalable filter products, these filter capsules feature the lowest surface area for a pleated filter unit and a
very small hold up volume, making it ideal for filtering small quantities of fluids (up to 100 mL) with minimal
losses.
Key Features and Benefits
Titer reduction claim in excess of 6 log guarantees excellent microbial filtration efficiency in processes
with variable bioburden
Broad pH compatibility PES membrane means Supor EAV filters are suitable for processing a wide
range of fluids
Polyethersulfone membrane has low adsorption profile for maximum transmission of proteins
Supplied pre-sterilised for immediate use
Quality Standards
Meet USP biological reactivity tests in-vivo, in accordance with USP Class VI 121 °C for biological
safety
Endotoxin Level < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Cleanliness per USP particulates in injectables
Non fiber releasing per CFR 211.72 and CFR 210.3(b)(6)
Specifications
Materials of Construction
Filter Membrane
Hydrophilic PES
Housing, Vent Plug and Support Material
Polypropylene
Sealing Technology
Insert molding
Operating Parameters and Temperatures1
5.4 bar (80 psi) @ 20 °C
2.1 bar (30 psi) @ 60 °C
1 In compatible fluids which do not soften, swell or adversely affect the filter or its materials of construction.
Typical Hold-up Volume
< 2.5 mL
Sterilization2
Available pre-sterilized, subject to a minimum of 25 kGy of Gamma-irradiation
2 This product may not be sterilized by in-situ steaming.
Warning: Pre-sterilized Mini Kleenpak filters should not be re-irradiated or autoclaved.
Nominal Dimensions
Capsule Length
21 mm (0.8 in.)
Capsule Diameter
29 mm (1.2 in.)
Nominal Effective Filter Area (EFA)
2.8 cm2 (0.43 in2)
Typical Flow Rates
Ordering Information
To order, quote Pall part number:
KM2EAVS
Supplied in boxes of 50 filters
Contact Information