Unit 5A - Workforce Solutions

advertisement
UNIT 5A
Introduction to
Radiographic Equipment
General Radiographic, Fluoroscopic
Dedicated, Multipurpose &
Mobile
Conventional vs. Digital
General Radiographic Equipment
• Multi-purpose / general radiographic
equipment - a wide range of procedures
can be performed using this type of
equipment
• It is used for general and/or routine
radiography.
• Examples: orthopedic radiography, chest,
podiatric, abdomen, spine and skull
radiography
General Radiographic Equipment
• Multi-purpose
general
radiographic
equipment
– Stationary
• Conventional
(cassette-based
film)
• Digital (CR –
cassette-based)
DR (Direct Digital) Radiographic
Equipment
• Multipurpose
– stationary
• IR integrated into
the equipment
(cassetteless)
Terminology
• Conventional - Uses film in cassettes (film
holders)
• Digital - Filmless; image is viewed on
monitor
– CR – computed radiography – uses
conventional equipment with cassettes
– DR – Direct digital radiography – uses special
equipment; cassetteless
Image Receptors
• Defined as the device that receives the
energy of the x-ray beam and forms the
image of the body part
• Four types
– Cassette with film (conventional)
– Image plate (IP – used in CR)
– Direct radiography (detector)
– Fluoroscopic screen
The X-ray Tube
• X-rays are produced in a cathode ray
tube
• They are produced from a series of
energy conversions
Requirements for the production of
x-rays include:
• A source of
electrons (the
filament or cathode)
• A means to
accelerate the
electrons (kVp)
• A way to stop the
electrons (the
target or anode)
The x-ray tube
• Is an evacuated glass envelope
• Has a positive (anode) and negative
(cathode) electrode.
• The cathode (filament) serves as the source of
electrons
• High voltage is applied (kVp) to accelerate the
electrons across the tube
• The anode (target) stops the electrons suddenly
which results in the production of x-rays
X-ray production
• The energy conversion involves the
conversion of
– electrical energy to..
– heat energy to..
– kinetic energy to..
– x-ray energy and heat energy.
• >99% of the energy conversion results in
heat
• <1% of the energy conversion results in xrays
The x-ray tube
• Tube housing protects the
patient and operator from
radiation emanating in all
directions.
• The collimator decreases or
increases the size of the xray field (inc or dec pt. &
operator exposure.)
– PBL-positive beam
limitation-automatic
collimation – detects size
of image receptor and
automatically
reduces/increases beam
size to correspond.
Tube Stand
• The device that supports and permits
the x-ray tube to be moved in different
directions.
• Floor-mounted – runs along a track on
the floor.
• Floor-to-ceiling or Floor-to-wall - has
add’l tracks on wall or ceiling
• Ceiling-mounted – provides most
flexibility
X-ray tube
controls
*SID is the distance
between the source (anode)
and the image receptor (IR)
• Longitudinal lock locks tube into position
along the length of the
table
• Transverse locklocks tube into position
across the width of the
table
• Vertical lock-locks
tube vertically to set
the SID*
• Collimator controls
X-ray tube
controls
• Detent lock-locks the
tube into the center of
the bucky tray (where IR
is) transversely.
• Tube angulation lockallows angulation of the
tube cephalad (towards
the head) and caudad
(towards the feet)
• Tube angle indicatorindicates the degrees of
tube angulation
Radiographic
Tables
• The table is
designed to support
the patient in a
position that
enhances the
radiographic
examination
• It is not designed
for comfort!
Radiographic
Tables
• The table must be
uniformly
radiolucent (allows
x-rays to easily pass
through)
• It must be easily
cleaned.
• It must be hard to
scratch.
• Some tables are
stationary (ie.
They don’t move)
Radiographic Tables
• Tables must have a space for a tray (Bucky
tray) to hold image receptors (IR) and a
radiographic grid
IR
Bucky Tray
Image Receptors
• Image receptors
(IR) can either be
film cassettes or
• CR (computerized
radiography)
imaging plates
(IP).
IR Orientation
Radiographic Table Accessories
• Grid –is installed over the tray & consists of Pb strips and
is used to absorb scattered radiation which causes fog
that degrades the visibility of image detail.
– Some grids consist of a mechanism that automatically
moves the grid during exposure (reciprocating grid).
– Others use a stationary grid (one that does not move).
Radiographic Tables
• Tables can be “fixed” or “tilting”
• Fixed (stationary) tables do not
permit tilting of patient’s head or feet
• Tilting tables are used mostly for
radiographic/fluoroscopic applications
• Table height is typically 30-40” from the
floor.
– Some fixed tables have adjustable heights
to assist patients on and off the table
Radiographic Tables
• Tabletops are usually movable
– Some are motor-driven along their
length
– Some are “floating tops” which move
along length and width simultaneously
– Brakes can be controlled by hand, knee or
foot.
– Movable tables are an advantage in
moving/positioning large patients!
Radiographic Table Accessories
• Footboards are an accessory used to
support the patient in a standing position
when tilting tables are employed.
•
should be exercised to
assure the footboard is
attached correctly to prevent injury to the
patient!
Wall-Mounted Bucky System
and Cassette Holders
• Allows standing radiographs
• Cassette holder – adjustable;
holds various size cassettes.
• Wall-Mounted Bucky –
Cassette holder (tray) with grid.
• DR (Digital Radiography)
does not use a bucky tray;
image receptor does not have to
be changed between exposures
Generator or
Control panel
• KvP Selector
• mA Selector
• Exposure time
Selector
• Table bucky vs
upright Bucky
switch
• On/Off switch
• Exposure switch
kVp Selector
• Selects the kilovoltage applied to the x-ray tube
to produce x-rays
• Diagnostic Range 45kVp-150kVp
• Major/Minor (increments of 10 kV/1kV)
• KvP controls the penetrating power of the xray beam (and therefore the # of x-rays
reaching the film)
• KvP controls the energy of the beam
• KvP affects density (darkness/brightness) and
contrast of the image (more on this in a later
lesson)
mA selector
• Selects the milliamperage (mA;x-ray tube current)
• Heats the filament; boils off electrons
(thermionic emission)
• Controls the # of electrons flowing from
cathode to anode (ie. Tube current)
• Controls the amount of x-rays produced
• Controls the amount of exposure delivered to
the patient.
• Controls the density (darkness) or brightness
of the image
• Range 25mA-1200mA (varies with equipment)
Exposure time selector
• Selects the length of the exposure
• Range from milliseconds to 6 seconds
(varies with equipment)
• Controls the time electrons flow from
cathode to anode
• Controls the amount of exposure
• Controls the density or brightness of
the image
mAs
• mAs is the product of mA X exposure time
• It represents the quantity of radiation
exposure delivered.
• Ex: 100 mA @ 1/20 second = 5 mAs
• Ex: 500 mA @ .01 second = 5 mAs
Table bucky vs
upright Bucky
switch
• Available when room
has a table bucky and
an upright bucky
• Activates the
reciprocating bucky
in table or upright.
Other controls
• On/Off switch - turns power to controls on
or off
• Exposure switch - remote control switch
which controls exposure
– prep stage - heats cathode and rotates anode*
– exposure stage - makes exposure(applies kV)
– *Holding the “prep” or “rotor” down for an
extended time decreases the life of the xray tube!
Other controls
• AEC - Automatic Exposure Control Automatically terminates exposure when
predetermined amount of exposure is
reached.
• Determines the amount of exposure needed
to produce a diagnostic image.
• Actually controls the exposure time.
• Operator selects kVp (and sometimes the
mA) and appropriate cells (detectors)
AEC (Automatic Exposure Control)
Provides automatic
control of x-ray
exposure
exposure automatically
terminates when the film
has received desired
amount of film
AEC
Radiographic/Fluoroscopic Equipment
• Radiographic equipment produces static
images only.
• Fluoroscopic equipment produces static
AND dynamic images.
Radiographic/Fluoroscopic Equipment
• In addition to the
components already
discussed,
fluoroscopic
equipment has a
movable
fluoroscopic
carriage which
includes:
– Image intensifier
– An additional x-ray
tube under the table
Radiographic/Fluoroscopic Equipment
• Spot film device – permits the
radiologist to obtain static images
during the fluoroscopic exam.
– Uses various cassettes
– Allows one or more exposures/image
receptor.
• Digital fluoroscopy allows digital
acquisition of spot films (cassetteless)
– Stores static images on a computer.
Mobile Radiographic and Fluoroscopic
Equipment
• Used for radiography:
– At the patient’s bedside
– In the surgical suite
– In the emergency department
• Mobile equipment has similar controls
to stationary equipment but no table.
Mobile Radiographic and Fluoroscopic
Equipment
• A mobile
fluoroscopic unit is
sometimes called a
“c-arm”
Introduction to the
Darkroom
Cassettes
Radiographic Film
Storage and Handling
Essentials of the Darkroom
Automatic Processor
Cassettes
• Cassettes are a lighttight film holder.
• They have a
radiolucent (allows xrays to pass through)
front (tube side)
• The back side usually
has Pb (lead) or Al
(aluminum) to absorb
back scatter.
• Cassettes come in a
variety of standard
sizes.
Standard Cassette Sizes
English
• 14” X 17”
• 14” X 14”
• 11” X 14”
• 10” X 12”
• 9.5” X 9.5”
• 8” X 10”
• 7” X 17”
Metric
• 35 X
• 35 X
• 28 X
• 24 X
• 24 X
• 20 X
• 18 X
43cm
35cm
35cm
30cm
24cm
25cm
43cm
Cassettes
• Most cassettes have intensifying screens
mounted on the front and back inside
surface.
• An Intensifying screen is a fluorescent
material that converts x-ray energy to light
energy which exposes the film.
• The purpose of intensifying screens is to
reduce patient exposure.
• Some cassettes have a single intensifying
screen; some have two intensifying
screens.
Intensifying Screens
• An intensifying screen is composed of
fluorescent phosphors which give off light
when struck by x-rays.
• Different phosphors give off different colors
(spectra) of light
• In radiography, most emit blue, blue/violet
or green light.
• Radiography film is most sensitive to one of
these spectrum.
Luminescence: Fluorescence vs.
Phosphorescence
• Luminescence - emission of light
• Fluorescence - when a phosphor emits light
upon being struck by x-rays, and light
emission ceases when the x-ray source
ceases.
• Phosphorescence - (aka “screen lag”) when
a phosphor emits light upon being struck by
x-rays, but the light continues to glow after
the x-ray source ceases.
Radiographic Film
• Radiographic film is
the primary medium
of conventional
radiography.
• It is composed of a
polyester base
which is tinted blue,
• And a coating of
emulsion (the light &
x-ray sensitive
component of film)
which is composed
of AgBr crystals
suspended in gelatin
• Film can either have
Radiographic Film emulsion coated on
one side (single
emulsion)
• or can be coated on
both sides (duplitized).
• Single emulsion film is
used with cassettes
with single screens.
• Duplitized film is used
with dual-screen
cassettes.
Radiographic Film Classifications:
• Speed - reflects the sensitivity (response)
of the emulsion to x-ray and light
exposure.
– ex: 100, 200, 400, 800 speed; the higher the
number, the more sensitive the emulsion
(requires less exposure)
• Contrast - how the film exhibits the
relationships between blacks and whites.
• Resolution (detail) - how the film
reproduces details of the subject.
Film Storage and Handling
• Ideal storage temperature - 68o
• Ideal humidity - 30-60%
• Stored film must be protected from:
– radiation
– light
– age
– pressure
– high temperatures
– too high/low humidity
Artifacts-foreign marks on the film
• Fog – gray/black appearance of film
– Caused by: Exposure to radiation, light or age
– Fog destroys visibility of details.
• Static-black “lightening”-like artifacts
– Caused by: Static discharge in the darkroom
due to low humidity & electrification by friction.
• Crescent (“crinkle”) marks- black moonshaped artifact.
– Caused by: Bent film (mishandling)
Darkroom
• Should be a safe environment for handling
radiographic film.
• Safelight-allows visibility in the darkroom
without fogging the film!
– Maximum intensity of bulb- 15 watts
– Safelight filters
• GBX - red light (for green or blue sensitive films
• Wratten 6B amber light for blue sensitive only (not
used today)
– Safelight can be direct or indirect .
Darkroom - film storage
• Film bins provide various sizes of film a lighttight & radiation proof storage .
• Films are stored largest film to smallest film
(front to back)
• Boxes of film are stored on end to prevent
pressure artifacts.
• The FIFO method (First in; first out) of film
usage prevents film from being used after it’s
expiration date.
• Expiration dates are stamped on each box
Darkroom –
The Automatic
Processor
• There are four racks
in an automatic
processor:
Developer Fixer,
Wash, Dryer
• Processing
temperature is 9095o F.
• Most processors
process films in 90
seconds.
Automatic Processor
Identify the following in the Darkroom:
• Main switch ( on wall)
• Power on/off (on processor)
• Film tray
• Water valve
• Replenishment tanks
– One for developer; one for fixer
– hoses
Sponsored by:
•
This workforce solution was funded by a grant awarded under the President’s CommunityBased Job Training Grants as implemented by the U.S. Department of Labor’s Employment
and Training Administration. The solution was created by the grantee and does not
necessarily reflect the official position of the U.S. Department of Labor. The Department of
Labor makes no guarantees, warranties, or assurances of any kind, express or implied, with
respect to such information, including any information on linked sites and including, but not
limited to, accuracy of the information or it’s completeness, timeliness, usefulness,
adequacy, continued availability, or ownership. This solution is copyrighted by the
institution that created it. Internal use by an organization and/or personal use by an
individual for non-commercial purposes is permissible. All other uses require the prior
authorization of the copyright owner.
Download