THERMOMETRIC CALIBRATION OF THE HEATING EFFECTS BY 27.12 MHz METTLER DIATHERMY SYSTEM FOR USE IN HYPERTHERMIA SYSTEM FOR TREATMENT OF CANCER Abdul Muqeet Syed B.E., Osmania University, India, 2006 Atif Ahmed Syed B.E., Osmania University, India, 2006 PROJECT Submitted in partial satisfaction of the requirements for the degree of MASTER OF SCIENCE in ELECTRICAL AND ELECTRONIC ENGINEERING at CALIFORNIA STATE UNIVERSITY, SACRAMENTO FALL 2009 THERMOMETRIC CALIBRATION OF THE HEATING EFFECTS BY 27.12 MHz METTLER DIATHERMY SYSTEM FOR USE IN HYPERTHERMIA SYSTEM FOR TREATMENT OF CANCER A Project by Abdul Muqeet Syed Atif Ahmed Syed Approved by: , Committee Chair Dr. Preetham B. Kumar, Ph.D. , Second Reader Prof. Russ Tatro, MS EE. Date ii Name of the Students: Abdul Muqeet Syed Atif Ahmed Syed I certify that these students have met the requirements for format contained in the University format manual, and that this Project is suitable for shelving in the Library and credit is to be awarded to this Project. ____________________________, Graduate Coordinator Dr. Preetham B. Kumar, Ph.D. Department of Electrical and Electronic Engineering iii ____________________ Date Abstract of THERMOMETRIC CALIBRATION OF THE HEATING EFFECTS BY 27.12 MHz METTLER DIATHERMY SYSTEM FOR USE IN HYPERTHERMIA SYSTEM FOR TREATMENT OF CANCER by Abdul Muqeet Syed Atif Ahmed Syed This project will focus on the thermometric calibration of the heating effects by the 27.12 MHz Mettler Diathermy system on non-hardening clay with digital thermometer of plastic probe for use in Hyperthermia system for the treatment of cancer. The non-hardening clay is chosen so that the electrical properties are identical to the biological human tissue. The experimental study on the Mettler diathermy system will include the adjustments of the head of the applicator and different sizes of clay medium. Our goal would be to achieve the temperature of the treatment within the desired time and to improve the conditions of the experiment and to be more efficient and accurate with the temperature measurements with respect to time. , Committee Chair Dr. Preetham B. Kumar, Ph.D. _______________________ Date iv ACKNOWLEDGEMENTS We would like to take this opportunity to convey our sincere regards to Dr. Preetham Kumar, faculty Member, EEE department and Graduate Advisor, for valuable guidance, giving us the opportunity to take on this project and being very helpful throughout this project. It was his support and encouragement that helped the project to start and end at the right time. We thank him again for his constructive feedback throughout the course of our fieldwork. We would also like to acknowledge and thank Professor Russ Tatro, Faculty Member, EEE department for being part of the review committee and extending his guidance for better formulation of our project. We also thank him for his review and comments on the project report. Our sincere appreciation goes to all our family members and friends for their love and support during the entire duration of our coursework and this project. v TABLE OF CONTENTS Page Acknowledgements……...………………………………………………………………...v List of Tables…………………………………………………………… ……………….ix List of Figures…………………………….……………….……………………………...x Chapter 1. INTRODUCTION……..…………...…………….…………………………...…………1 2. BACKGROUND RESEARCH ON MICROWAVE HYPERTHERMIA IN CANCER TREATMENT…………………………………3 2.0 Description on Hyperthermia Radiation ……………………………………………3 2.1 Types of Hyperthermia ……………………………………………………………..3 2.1.1 Local Hyperthermia ………………………………………………..……….4 2.1.2 Regional Hyperthermia ……………………………………………………..5 2.1.3 Whole Body Hyperthermia …………………………………………….…...7 2.2 Effects and Working Methodology of Hyperthermia ….…………………………..8 2.3 Benefits of Hyperthermia ………………….……….......………………………....10 2.4 Risks Involved in Hyperthermia..…………………………………………………12 3. DIATHERMY SYSTEMS FOR HYPERTHERMIA APPLICATIONS ………..……14 3.1 Hyperthermia Application Systems………………………………………………...14 3.1.1 Hot Packs ………………………………………………………………….14 3.1.2 Electrical Heating Pads ………………..….……………………………….15 vi 3.1.3 Heat Lamps …………………………………………………………………..15 3.1.4 Paraffin baths…………………………………………………………………15 3.1.5 Ultrasound…………………………………………………………………….16 3.1.6 Infrared Heating Pads……………………………….……...………………...16 3.2 Diathermy…………………………………………………………………………16 3.2.1 Shortwave Diathermy………………………………………………………17 3.2.1.1 Dielectric Diathermy………………………………………..……….17 3.2.1.2 Inductive Diathermy……………...…...…………………………….17 3.2.2 Microwave Diathermy ……………………………………………………..18 3.3 Absorption of Microwaves………………………………………………………..19 3.3.1 Effect of Hyperthermia in the Flow of Blood …………...………………...21 3.3.2 Features of Diathermy Devices…..…..………………………………….…22 4. METTLER AUTOTHERM EQUIPMENT DESCRIPTION………………………...25 4.1 Mettler Autotherm Equipment Details ………..…………………………………26 4.2 Mettler in Heat Therapy ……………..…………..………………………………27 4.3 Technical Specifications of Mettler Autotherm Equipment ……………….……28 4.4 Advantages of Mettler Shortwave Diathermy Unit …..…………………………29 4.5 Precautions While Using Autotherm ..…..………………………………………29 4.6 Digital Thermometer with Plastic Probe ….…………………………………….32 5. EXPERIMENTAL RESULTS USING DIATHERMY ON CLAY MEDIUM…...….34 5.1 Measurements with the Probe on the Top of the Clay….……..………………...35 vii 5.2 Measurements with Probe Inserted 1 Inch From the Surface…….…..…….……37 5.3 Measurement with the Sliced Clay……….……..….……………………………38 5.4 Measurements Using a Reflector……………….…...…………………………...40 5.5 Measurements with the Applicator Touching Clay………….…….…..……..….41 5.6 Measurement with the Probe Inserted one Inch From the Surface…….…...……43 5.7 Measurements with the Sliced Clay Toucing the Applicator Head….………….44 5.8 Measurements of the Clay with Water………….…….…..….………………….47 6.CONCLUSION………………………………………………………………………..50 References ……………………………………………………………………………….52 viii LIST OF TABLES Page Table 3.1 Penetration Depths of Tissues in cm with High and Low Water Content Tissues………………………………………….19 Table 4.1 Electromagnetic Heating Comparison .………….…………...………………..30 Table 5.1 Variation of Temperature Rise in Thick Clay with Time (Probe on Top of Thick Clay)………………………………….36 Table 5.2 Variation of Temperature Rise in Clay with Time (Probe Inch below Clay Surface)………………………………37 Table 5.3 Variation of Temperature Rise in Sliced Clay with Time (Probe on Top of Clay)…………………………………………39 Table 5.4 Variation of Temperature Rise in Clay with Time Using Reflector (Probe on Top of Clay)…………………………………………40 Table 5.5 Variation of Temperature Rise in Clay with the Applicator Touching (Probe on Top of Clay)…………………………………………42 Table 5.6 Variation of Temperature Rise in Clay with the Applicator Touching the Surface…………………………………………..43 Table 5.7 Variation of Temperature Rise in Sliced Clay with the Applicator Touching the Surface…………………………………………..45 Table 5.8 Variation of Temperature Rise in Clay at Different Depths with the Applicator Touching the Surface……………………………….46 Table 5.9 Variation of Temperature Rise in Clay with Time (One Inch below Clay Surface with Water)………………….…48 ix LIST OF FIGURES Page Figure 2.1 Scheme for Local Hyperthermia ……….……………………………………...4 Figure 2.2 Applicator Types for Local Hyperthermia, such as (a) Waveguide Applicator; (b) Spiral Applicator; and (c) Current Sheet Applicator……………..………5 Figure 2.3 Sigma-60 Applicator……………………………………………..……….…….6 Figure 2.4 Schematic for Aquatherm System ……………………………….…………….8 Figure 2.5 Density Measurements From Stained Tissues ….…..………….....…...……...9 Figure 3.1 Rate of Heat Penetration in Fat, Muscle and Bone Tissues……....….…….….20 Figure 3.2 Range of Intensities of Stray Magnetic Fields around the Diathermy Cables………………………………….….23 Figure 3.3 Range of Intensities of Stray Magnetic Fields around the Diathermy Cables……………………………………..24 Figure 4.1 Mettler Autotherm Diathermy Unit………………………………….………..25 Figure 4.2 Control Knobs on Mettler Autotherm…….……….…………………………..26 Figure 4.3 Patient Input Meter for Autotherm………….….……………………………...27 Figure 4.4 Relative Absorption of RF Power Generated by the Autotherm Equipment…………………………………….28 Figure 4.5 Plastic Probe Digital Thermometer…………....………………………………32 Figure 5.1 Graphical presentation of Temperature Vs Time (Probe on Top of Thick Clay)……………………………..36 Figure 5.2 Graphical Presentation of Temperature Vs Time (Probe Inch below Clay Surface)………………………….38 Figure 5.3 Graphical Presentation of Temperature Vs Time (Probe on Top of Clay)……………………………………39 x Figure 5.4 Graphical Presentation of Temperature Vs Time (Probe on Top of Clay)……………………………………41 Figure 5.5 Graphical Presentation of Temperature Vs Time…….….……………………42 Figure 5.6 Graphical Presentation of Temperature Vs Time (Probe One Inch below the Surface)……………..………..44 Figure 5.7 Graphical Presentation of Temperature Vs time (Probe on Top of the Clay)……………………………….45 Figure 5.8 Graphical Presentation of Temperature Vs time (for all three Depths)……………………………….……47 Figure 5.9 Graphical Presentation of Temperature Vs Time (One Inch below Clay Surface)………………………….48 xi 1 Chapter 1 INTRODUCTION In this phase of experiment our aim is to focus mainly on several aspects to improve the accuracy and the understanding of hyperthermia treatment using Mettler Autotherm. However, we want to improve the experimental procedure in such a way that we can get consistent measurements. Our goal is to get accurate and efficient measurements so that these can become the basis for further study and research in treatment of hyperthermia. There are certain changes that could be done to achieve desired temperature at different depths so that it comes closer to the real time treatment of hyperthermia. We are using the Mettler Autotherm 300 to generate microwaves for heating the clay. Clay is chosen as it closely mimics the properties of human tissue or skin. The modern hyperthermia equipments used these days are highly complicated, bulky and are very expensive. This makes the treatment confined and accessible to very small number of centers. On the other hand the Autotherm 300 is portable, flexible and inexpensive. This is the reason even small treatment centers can take advantage of it. In this experiment we will be replacing the metal probe thermometer with plastic probe digital thermometer. In the first phase of measurements [1] the temperature readings taken by metal probe thermometer were inaccurate as the calibrations were affected by interference between the incident wave energy and the metal probe. As stated earlier, our aim is to obtain higher temperature similar to the temperature used in real 2 time treatment. We will vary the medium of transmission for some measurements. We will also be using the reflector in different positions in order to increase the heating effect. The process of consistent heating will be done to observe the differential heating effect, so that we can understand the variations with respect to time in a better manner. More number of variations in the temperature calibrations is expected throughout the experiment. According to Food and Drug Administration (FDA) the patient should not be exposed to heat more than 20 minutes. This regulation is set to avoid the pain threshold in human during the treatment. The experiment will also involve the ways to achieve this temperature rise (more than 100° F) within 20 minutes of the treatment time. This variation can be achieved if we succeed in reducing the heat loss during the experiment. Our main goal in this experiment hence is to improve the experimental conditions from the first phase and also to understand more efficient and accurate temperature variations with respect to time. The first chapter of this report is an introduction to cancer and the treatments used. Chapter II gives background knowledge and research on hyperthermia for the treatment of cancer. We will also be discussing the benefits and applications of microwave hypothermia. Chapter III explains the dielectric properties and the types of diathermy. Chapter IV looks at the details of Mettler Diathermy equipment and Plastic probe thermometer. Finally Chapter V details the experimental results that were obtained using the Mettler Autotherm unit, followed by the conclusions and References. 3 Chapter 2 BACKGROUND RESEARCH ON MICROWAVE HYPERTHERMIA IN CANCER TREATMENT 2.0 Description on Hyperthermia radiation Hyperthermia is a type of treatment in which body tissues under treatment are exposed to high temperatures. Research has shown that high temperatures can damage and kill cancer cells, usually causing minimum injury to the other normal tissues. It is proposed that by killing cancer cells and damaging proteins and structures within the cells, hyperthermia may shrink tumors [2]. The heat may destroy two types of tumor cells, the first being those that are making deoxyribonucleic acid (DNA) in preparation for division and the second type of cells being those that are acidic and poorly oxygenated. These cell types tend to be resistant to radiation [3]. It is suggested by proponents that heating makes cells more sensitive to radiation and prevents radiationdamaged cells from repairing themselves [4]. 2.1 Types of Hyperthermia There are different kinds of hyperthermia that are under study 1. Local Hyperthermia 2. Regional Hyperthermia 3. Whole Body Hyperthermia 4 2.1.1 Local Hyperthermia In local hyperthermia the cells affected with cancer are heated to at very high temperatures 106° F by using heating elements such as microwave, antennas, heating rods, ultra sound. In this treatment the heat is applied to small areas on the body and when the cancerous cells are exposed to high temperatures continuously they get destroyed. Two factors play a major role in this treatment: the temperature of the cell and the volume of area exposed. The temperature of 106º F for a period of 60 min is used here [5]. There are several applicators are used for the treatment of local hyperthermia such as waveguide, horn, spiral, current sheet and compact applicators. The temperature during the treatment can be controlled either by positioning the applicator or using a power generator. We can see the components used in the hyperthermia system in the Figure 2.1, 2.2. Figure 2.1 Scheme for Local Hyperthermia [6] 5 Figure 2.2 Applicator Types for Local Hyperthermia, such as (a) Waveguide Applicator; (b) Spiral Applicator; and (c) Current Sheet Applicator [6] 2.1.2 Regional Hyperthermia Regional Hyperthermia is used for specific body organ such as hands, legs. It is very popular for the treatment of cancers like sarcomas and melanomas. Deep tissue, regional perfusion and Continuous Hyperthermic Peritonial Perfusion (CHPP) are the different ways that are followed in regional hyperthermia. Deep tissue cancer treatment is done mainly for lung, liver and ovarian carcinoma. The tumor is treated with devices that can produce high energy waves directed to specific areas [7]. In Regional perfusion affected body part is isolated from the rest of the body. Pump oxygenator is used to pump the blood from the isolated region 6 into the heating device. Once the blood is heated to certain temperature the blood is pumped back. Continuous Hyperthermic Peritonial Perfusion (CHPP) is type of regional hyperthermia which is used to treat abdominal cavity cancers such as intestines and digestive organs. The drugs are injected into the cancerous cavity during this treatment and the temperature is maintained up to 108º F. Regional hyperthermia treatment uses the Sigma-60 applicator which is shown below in Figure 2.3 Figure 2.3 Sigma-60 Applicator [6] 7 2.1.3 Whole Body Hyperthermia Whole body hyperthermia (WBH) is utilized when the cancer has spread throughout the body (metastatic cancer) [5]. It is achieved with either radiant heat or extracorporeal technologies, and elevates the temperature of the entire body to at least 105.8° F. In radiant WBH, heat is externally applied to the whole body using hot water blankets, hot wax, inductive coils, or thermal chambers. The patient is sedated throughout the WBH procedure, which lasts approximately four hours. The patient reaches target temperature within approximately 1.3 hours, is maintained at 107.2° F for one hour, and one-hour cooling phase. During treatment, the esophageal, rectal, skin and ambient air temperatures are monitored at 10-minute intervals. Small probes may be inserted into the tumor under a local anesthetic to monitor the temperature of the affected tissue and surrounding tissue. Heart rate, respiratory rate, and cardiac rhythm are continuously monitored. It is always recommended to use temperatures below 111° F otherwise the normal tissues may also get affected [5]. The Aquatherm system is used for the whole body hyperthermia. It is an isolated moisture-saturated chamber which ha steamed tubes (122 – 140 °F) on its inner sides in which the patient is positioned. The whole body is exposed to long-wavelength infrared wave emission. The cabins contain hot water tubes where the patient is kept and the temperature of (140° F) is maintained inside. After a systemic temperature of 107.2° F has been achieved, the patient is thermally isolated with blankets. 8 Figure 2.4 Schematic for Aquatherm System [6] 2.2 Effects and Working Methodology of Hyperthermia Hyperthermia has its own beneficial effects in several ways on the body. Several studies have shown that this treatment increases apoptosis in the body in response to heat. Hyperthermia damages the membranes, cytoskeleton, and nucleus functions of malignant cells and thus kills the tumor cells. There is irreversible damage to cellular perspiration of the cells. Heat above 105.8° F also pushes cancer cells toward acidosis that is decreased cellular pH in body where there is decrease in the cells’ viability and transplantion ability. Hyperthermia when used activates the immune system of the body tissues. From one of the source we get to know that "Heat has a well known stimulatory effect on the immune system causing both increased production of interferon alpha, and increased immune surveillance." [8] Another source mentions the release of lysosomes [7]. 9 Tumors have an effect on the vessels carrying blood to them. The blood vessels grow in such a way that these vessels are unable to dilate and dissipate heat as normal vessels respond to heat. So tumors take longer to heat, but concentration of heat on to them has immense effect on the cells. Tumor blood flow is increased by hyperthermia. The blood flow increases as normal vessels are incorporated into the growing tumor mass and are able to dilate in response to heat, and to channel more blood into the tumorous cells. The figure 2.5 shows the density measurements of blood vessel for different processes. Figure 2.5 Density Measurements From Stained Tissues [9] Tumor masses contain oxygen deprived cells within the inner part of the tumor. These oxygen deprived cells provide resistance to radiation and are very sensitive to heat. This is the reason hyperthermia is considered to be ideal companion to radiation. While radiation kills the oxygenated outer cells, hyperthermia destroys the low oxygenated 10 cells. The hyperthermia oxygenates these cells and makes them more susceptible to radiation damage. It is also thought that hyperthermia’s induced accumulation of proteins stops the malignant cells from repairing the damage done by radiation. We get to know from a source that "It can be hypothesized that hypoxic cells in the center of a tumor are relatively radio resistant but thermo sensitive, whereas wellvascularized peripheral portions of the tumor are more sensitive to irradiation. This supports the use of combined radiation and heat; hyperthermia is especially effective against centrally located hypoxic cells, and irradiation eliminates the tumor cells in the periphery of the tumor, where heat would be less effective." [7] [8]. Hyperthermia when used with radiation is very effective. Studies [8] reveal that combination of hyperthermia and radiation therapy results effective and better treatment, the response rate for radiation without heat was 35% and the response rate of hyperthermia combined with radiation was 75% [8]”. As the research gains momentum, more reasons for the use of hyperthermia are continuously being identified. 2.3 Benefits of Hyperthermia Hyperthermia, when used alone, results in impressive shrinkage sometimes has complete eradication (10-15%) of tumors. These results usually don’t last, and the tumors grow again. Hyperthermia raises the body temperature above normal of 98.6° F as this reduces the foreign organism presence and the impurities in the body. When compared to the body tissues, the invading foreign organisms cannot survive in the high temperatures. 11 This is a certain benefit as the temperature can be increased above certain level which kill the unwanted organisms, bacteria and virus, and thereby killing the cancer cells. Increase in body temperature above normal levels consistently for a period of time helps in drawing out toxins out of the body, cleaning clogged pores, killing of harmful bacteria and viruses, increasing circulation and enhancement of immune system. It is also used in treatments such as healing the muscle aches, pains and injuries which is treatment after accidents [10]. There are several applications of the Hyperthermia in the field of medical sciences. Many diseases can be treated by using this. Hyperthermia and radiation combined together in the treatment has been reported to yield higher and durable responses than radiation alone. In deep seated tumors, the effect of this combined treatment is under research for deep seated tumors. Though it is difficult to increase human tumor temperature, recent clinical trials has shown that radiation with hyperthermia is far more successful in controlling many human tumors radiation alone. Hyperthermia is used in the treatment of upper and lower respiratory tract infections, bladder problems, bronchitis, pneumonia, urinary tract infections such as cystitis, sinusitis and other conditions of the lungs and body cavities. It is also helpful in controlling the bleeding; improve conditions in prostatic hypertrophy and psoriasis [7]. Hyperthermia may also provide additional advantage in drug delivery, being specific hyperthermia is good choice for delivery of drugs with relatively large carriers. Several studies have shown that the delivery of monoclonal antibodies is enhanced by 12 using hyperthermia, particularly to the tumors with resultant improvement in anti tumor effects. The spread of carried chemo drugs into the tissues of liposome is increased considerably with higher temperature when compared to that under normal temperature. Much of the information and research has emerged from hyperthermia studies in several other treatments that may become valuable in the future. Hyperthermia improves the therapeutic index of TBI (total body irradiation), not only by increased neoplastic cell kill, but also by inhibiting the expression of radiation induced damage to the normal cell population. 2.4 Risks Involved in Hyperthermia The side effects of hyperthermia are external methods includes pain, unpleasant sensations and burns in a small percentage of patients. The situation is more complicated in the case of the internal pyrogens. Sometimes there are fatal reactions in humans, as bacterial toxins released during treatment can induce serious imbalances to the tissues, depending on dosage. Ultrasound hyperthermia will cause bone pain in areas where the tumor is over a bone. Whole body hyperthermia can result in neuropathies. Extracorporeal systemic hyperthermia is another mode, where the blood is routed from the body as in dialysis and is heat is applied before returning to the body. It has few benefits such as higher possible temperatures, and more homogeneous heating. The side effects are considerable such as frequent persistent peripheral neuropathies, abnormal or lethal blood coagulation, damages to liver tissues and kidneys, and brain hemorrhaging and some seizures. We get the reason from a source which says that the role of perfusion- 13 induced hyperthermia may be doubtful. There are few risks involved with whole body hyperthermia and these risks include problems such as cardiac and vascular disorders, diarrhea, nausea and vomiting. Extracorporeal systemic hyperthermia can give side effects such as frequent persistent neurophites, damages to liver or kidneys, abnormal blood coagulation, and brain hemorrhaging and few seizures. [11] [12] All these risks and side effects involved with hyperthermia are very rare, negligible and can be avoided by being careful and precise during treatment. The reactions can be avoided by controlled clinical procedures. 14 Chapter 3 DIATHERMY SYSTEMS FOR HYPERTHERMIA APPLICATIONS 3.1 Hyperthermia Application Systems Hyperthermia is any process that uses heat to cure a disease or in order to kill tumors in the human or animal body. This process helps in relaxation of the muscle tissues and also helps in the metabolic process and also in the induction of reflex vasodilatation. Hyperthermia could be the process to kill on the skin or inside the skin. [13] There are different processes that have been used as a process of Hyperthermia, some of which are mentioned below: 3.1.1 Hot Packs In this process of hyperthermia, canvas bags filled with silicon dioxide are used that absorbs more times its own weight. These canvas bags are immersed in hot water for good amount of time and are taken out whenever required. These canvas bags are covered with seven to eight layers of towels before applying to the surface and they are even made sure that the excessive water is taken out. These hot packs stay warm for 30 minutes and have advantage over heat lamps as the heat dries the skin and leaves the skin of the patient moist. 15 3.1.2 Electrical Heating Pads These are used as an alternate to the hot packs, and as these are electrically heated pads, they remain warm for a longer time than the hot packs as the water keeps circulating in this process. As the pads do not cool down by themselves, they should be set on a timer that stops the process within 20 minutes or it would burn the skin. 3.1.3 Heat Lamps The process of heat lamp works on the process of conversion of radiant energy into heat. This process uses 250W incandescent bulb as these bulbs produce large amounts of infrared energy which is good for the process. The heat of this process is adjusted by varying the distance between the bulb and the body. Usually the ideal distance between the bulb and the body is 40 to 50 cm. These are much easier to use than the hot packs for the patient as well as the therapist. For the patient as there is no pressure to be handled as in the case of hot packs. 3.1.4 Paraffin Baths This process is done by using a mixture of mineral oil and paraffin in a container which is heated from 122-129° F. The paraffin soluble can also be heated in small containers or home stoves. It is easier in usage too and can be used by the patient either by immersing the portion to be treated continuously for 30 minutes or even in intervals. 16 3.1.5 Ultrasound This is process in which conversion of the energy helps in producing heat unlike the process of Heat Lamps. In Ultrasound the sound energy is required to process the heat, and this process is usually beneficent with the deep tumors in the skin. This can be even placed on the skin or the application part can be immersed in the ultrasound treated liquid. As this process requires constant check by a therapist, it is not recommended for homes. 3.1.6 Infrared Heating Pads These heating pads use infrared waves in order to generate heat as the infrared rays contains the mechanisms such as arsenide and tungsten. In spite of the greater cost involved with the process, the hot packs and electric heating pads remain much better than the infrared heating pads. [13] 3.2 Diathermy In this process of heating, high frequency electric currents increase the blood flow in the body resulting in acceleration of the repair of the tissue. The heat generated also helps in reducing the nerve fiber sensitivity and also in increasing the threshold of the pain. This process is even used to treat arthritis, painful joints, pelvic infections and sinusitis. [14] There are two types of diathermy which are being used today: Shortwave diathermy and Microwave diathermy. 17 3.2.1 Shortwave Diathermy This diathermy is also known as radio frequency diathermy or high frequency diathermy. This process works on the operating frequency of 27.12 MHz which is assigned by the Federal Communications Commission. The part to be treated in this process is mainly the one that contains thick tissues like the ones at the hip and breast. The part is placed between two capacitor plates and high frequency wave is passed between the plates that treat the part [15]. There are two methods of applying shortwave diathermy, one is dielectric diathermy and the other is inductive diathermy. 3.2.1.1 Dielectric Diathermy The electrodes that are used for the dielectric diathermy produce alternating electric field which is kept either on one side or on both ends of the part to be treated. The electrodes are usually small plates or wire mesh which are covered with cushion or inserted inside the pillow. The electric charge that develops between the plates causes the movement of the molecules, the friction and collision among the molecules generates heat in the tissues. 3.2.1.2 Inductive Diathermy This type of diathermy uses coil for the treatment of the tissue to generate circular electric and magnetic fields into the tissues causing heat in them. These are usually used at the lower RF regions. The coil is normally wound inside the applicator so that the coil 18 could produce reversing magnetic fields. The applicators are usually designed in such a way that it could be easily adjusted and moved to be placed over the area or near to the area to be treated. Shortwave Diathermy has its own contraindications such as malignancy, sensory loss, metallic implants, pregnancy, phlebitis, metal containing intrauterine contraceptive devices, etc. 3.2.2 Microwave Diathermy As the name denotes it, this diathermy treatment uses microwaves to heat the body tissues. The applicator is usually a horn shaped or a direct beam gun that could be utilized to direct the heat to the concerned area. The variation in the heat to be applied is varied by the positioning the space between the applicator and the treated area. The shape of the applicator is usually designed considering the area where it would be used for treatment [15]. This diathermy deals with the frequency from 300 MHz to 300 GHz and this spectrum is after the radio frequency spectrum. These high frequency waves are usually used to cure the tumor on the tissues deep inside the human body. These microwave frequencies help in the repair of the tissues and reduces the stiffness in the joints and muscles. The microwave diathermy unit works on the frequency spectrum of 2.45 MHz and it is not just the frequency which is taken into consideration for the treatment but also 19 the other specifications such as time and power induced for the treatment. The table below shows the penetration depths of microwaves in the human tissues with respect to the water content present in it. Table 3.1 Penetration Depths of Tissues in cm with High and Low Water Content Tissues [6] The penetration of the microwaves depends on the dielectric constant as these both properties are directly proportional to each other. From the above table 3.1 we can see that there is more energy absorbed with high content of water. 3.3 Absorption of Microwaves The percentage of the radiation of the microwaves varies from the range of 50-75 percent depending on the thickness of the skin. All other lower frequencies than the microwaves reach up to the level of 60 percent of penetration. The figure below shows the amount of heat penetration in fat, muscle and bone tissues. 20 Figure 3.1 Rate of Heat Penetration in Fat, Muscle and Bone Tissues The absorption of the microwaves usually works on the Grotthus law which states that “When any radiation meets the surface of a different medium it may either be reflected or penetrate. Those radiations that do penetrate will only have an effect if they are absorbed; thus they will be ineffective if they pass right through”. The heating of the tissues depends on the absorption of the microwaves and rate of heat transfer between the tissues. So, this process of heating is mostly used for heating the tissues of the muscle to increase the flow in blood. [16] 21 3.3.1 Effect of Hyperthermia in the Flow of Blood The process of Hyperthermia is mainly used to kill cancer cells by heating the part to the treated keeping in consideration the pain threshold of the patient. For most treatment in the cure of cancer, hyperthermia only does not help in destroying the tumor and killing the cancer cells but it is usually used in addition with other process like chemotherapy and radiotherapy. [17] The use of heat for the process of curing the tumor of cancer is for the reason that cancer cells react more to heat than the normal cells in the human body. With the tumor in the tissues of the human body, it causes severe abnormalities in the structure of the tissue. As the tumor tissues react more to the increase in the heat, there is more structural damage to the cells which causes the increase in body temperature to about 104 degrees Fahrenheit. As there is structural damage to the tumor tissues, it gets acidic and gets deprived of nutrients. As the acidic conditions increase in the tumor area, the radiations are more effective on these tumors and it causes more effect if it is given with the drugs used for chemotherapy. Therefore, hyperthermia is usually combined with other processes in the treatment of cancer. [17] In both kinds of diathermy, shortwave and microwave, the heating of the tissues is done by the electromagnetic energy radiated from the head of the applicator. Usually in shortwave diathermy devices, the RF field exists around the cable that carries the electrical waves from the generator to the head of the applicator whereas in microwave 22 diathermy there is very little radiation as it is well shielded between the generator and the head. As it absorbs less RF energy from the surroundings, there is very little heat around the cables of the applicator. There are a few factors that are being considered including the generator setting and even the area where the energy is concentrated. Heating either with shortwave or microwave has never caused any harm as the maximum damage that can be done with that heat is that it causes smoke and melting of some materials. [18] 3.3.2 Features of Diathermy Devices The diathermy device uses the radio frequency spectrum to cure the tumor of the human body with heat produced. This process uses an applicator, a generator and a console to check the rise in temperature of the treated part. The power is transmitted from the generator to the head of the applicator and is concentrated on the part where the treatment is to be done. Proper care and supervision should be done so that the temperature should not increase the pain threshold or should not increase to the extent to cause burns on the human skin. The temperature should be elevated to 113° F and should be confined to not more than 30 minutes. There are two kinds of applicators used for the process of diathermy, one is capacitor type electrode and the other is inductor type electrode. In the capacitive type electrode the heating of the tissues is done by the RF electric field. In the inductive type electrode, the heating is done by the eddy currents induced by the magnetic field. In the 23 inductive type of electrode, there is high intensity magnetic field around the cables. The figure below shows the range of intensities around the diathermy cables in electric field and magnetic field. [19] Figure 3.2 Range of Intensities of Stray Magnetic Fields around the Diathermy Cables [19] 24 The figure below shows the range of electric field intensities around the cables of shortwave diathermy device for different types of electrodes and for different power settings. The intensity depends on the distance from the electrode. Figure 3.3 Range of Intensities of Stray Magnetic Fields around the Diathermy Cables [19] 25 Chapter 4 METTLER AUTOTHERM EQUIPMENT DESCRIPTION In this project, all the measurements of the experiment were conducted with the Mettler Autotherm equipment shown in the Figure 4.1. The Autotherm is a equipment which possesses unique induction field circuitry which produces a short wave frequency of 27.12 MHz. This wave can penetrate into muscle tissue with negligible heating in the fatty layer or bone [18]. Figure 4.1 Mettler Autotherm Diathermy Unit 26 4.1 Mettler Autotherm Equipment Details The Autotherm 300 is a continuous shortwave diathermy unit which is designed to be economical and light weight. It is capable of automatic tuning which ensures proper frequency response of the equipment. It is portable with a roller coaster base and is flexible. The arm is made adjustable, so that it can reach different parts of the body and can heat accordingly. This equipment is mainly used where deep heat is required such as the low back, shoulder, neck and hip. This equipment is economical and is easy to operate since it only has two controls, a timer knob and an intensity control knob as shown in figure 4.2 respectively Figure 4.2 Control Knobs on Mettler Autotherm 27 The Autotherm has a timer which is designed to the variation of 0 to 30 minutes. The timer helps to know the timing of treatment. The power meter, as shown in Figure 4.3, displays the energy levels absorbed by body surface. It also monitors the current from the power supply and displays the energy absorbed. Figure 4.3 Patient Input Meter for Autotherm 4.2 Mettler in Heat Therapy Mettler Autotherm is used when deep heat therapy has to be done for any part of the body. It operates with short wave diathermy and is a safe for subcutaneous body tissues. The electro-magnetic field is generated between the equipment and the body. The heat penetrates deep into muscle tissue and eases the tensions and brings relief. This therapy is mainly used for back pain, chronic arthritis, bursitis and other musculo-skeletal conditions [20]. The heat therapy modality is shown below in figure 4.4. 28 Figure 4.4 Relative Absorption of RF Power Generated by the Autotherm Equipment [19] 4.3 Technical Specifications of Mettler Autotherm Equipment Input: 100–240 VAC, 50-60Hz Frequency: 27.12 MHz (Wavelength λ= 11.06 meters) RF output: Continuous mode 100 W Average Power, Pulsed mode 200 W Peak Power Continuous mode: 100 Watts Average Power Pulsed Mode: 200 Watts Peak Power Pulse frequency: 10 Hz, 20 Hz, 50 Hz, 100 Hz, 400 Hz Pulse duration: 65 μs, 100 μs, 200 μs, 300 μs and 400 μs Treatment time: 1–30 minutes [20] 29 Weight: Unit: 30 pounds Dimensions: 40 in (H) x 18 in (W) x 18 in (D), (100 cm (H) x 46 cm (W) x 46cm (D)) 4.4 Advantages of Mettler Shortwave Diathermy Unit 1. Simple to use. 2. Economical 3. Flexible and gives easy access to all body areas 4. Timer switch for treatment lengths 5. Maximum portability 6. No damages are done to the adjacent tissues and bones 7. Light weight 8. A dosage intensity control regulates the power demand 4.5 Precautions While Using Autotherm The use of Autotherm for shortwave and microwave diathermy treatment, tissues are heated by use of electromagnetic energy. This induces or radiates from the head of diathermy unit. This is absorbed in the electrically conductive tissues of the human body and is converted to heat [11]. In most of the shortwave diathermy units, a radio-frequency field exists around the cables in the equipment and carries electrical energy from the source generator to the equipment head. Microwave diathermy units have well-shielded leads within the equipment between the generator and head of the applicator. The 30 generator and applicator head are integrated, and there is normally very little radiation from other source except that of the applicator. There is very little or no heating of the air surrounding the cables since it absorbs little energy from the radio-frequency fields. Table 4.1 shows the variation pattern of electric and magnetic fields in the diathermy unit as a function of distance within the treatment volume. Table 4.1 Electromagnetic Heating Comparison [11] Heating occurs only when conductive or partially conductive material is located within the electric or magnetic field produced by the cables or applicator. The Canadian Bureau of Medical Devices mentions in its data that certain plastics and synthetics such as nylon, polyvinyl chloride, and polyethylene terephthalate and some fabric blends which are usually regarded as good insulators, can also be heated certain temperatures by shortwave and microwave diathermy units [7]. However, two other synthetics widely used in medical practice for several applications, silicone and polytetrafluoroethylene usually called teflon are relatively very less affected by electromagnetic fields. The amplitude of heating depends on a number of factors, including the output setting of the generator and the degree to which the electromagnetic energy waves are 31 concentrated in a small area of the body. A very high density field exists because of three reasons: When a cable is located near a grounded or conductive object, when the cables are positioned at a very less distance. There are some recommendations which will minimize the risk involved in equipment damage and injury and can also reduce the risk of fire associate with equipment [18]. When generator is activated leaning on the equipment or holding the cables may cause unwanted heating or radiation to the body. If done so the cable might break and expose the user to high voltages or radiation. Cables should be spaced apart for each other. We should not keep anything in between these cables to avoid the damage of cables. We should keep all line cords away; there should be no contact between diathermy unit cables. The operating diathermy unit should be kept away from coil line cords. The cables should be kept at least several inches away from any objects or material.There should be no contact between cables and metal or grounded objects. There may be some synthetic or plastic objects which may be nonconductive but may be heated by the diathermy unit. 32 4.6 Digital Thermometer with Plastic Probe Figure 4.5 Plastic Probe Digital Thermometer [23] In the first phase of this project the thermometer with metal probe was used. In this phase of the project we have used the digital thermometer with plastic probe as shown in above figure 4.5. This digital thermometer has easy to read LCD which displays temperature in Fahrenheit or Celsius and its plastic probe accurately senses the temperature. This thermometer gives the readings in 0.1° F and in range of -10° F to 140° F with accuracy of ±1.5° F. [23] This thermometer includes suction cups, replaceable battery, and 39" long temperature probe cord. We can mount this digital thermometer for this experiment by following instructions as follows. Attach suction cups and place the probe in desired location. Press power button to activate the digital thermometer. Slide Fahrenheit/Celsius selector switch to Fahrenheit. Metal probe thermometer was acting as antenna as metal is conductor of electromagnetic energy. The temperature reading shown by this thermometer was fluctuating as it was effected by the interference caused by the microwaves. On the other hand the 33 thermometer with plastic probe is not affected by any interference as the plastic is not a conductor of any electromagnetic waves. The readings shown by this plastic probe are more accurate and are not fluctuating. 34 Chapter 5 EXPERIMENTAL RESULTS USING DIATHERMY ON CLAY MEDIUM As we have discussed about the Mettler diathermy system in the earlier chapter, in this chapter we outline the experiments that were taken by considering many specifications. In the first phase of experiments, [1] clays of two different types were used, one with more water content and one with less water content in it. In all our experiments, we would be using clay with more water content in it as we noticed better results with the non-hardening and that matches more close to the properties of the human body. The clay with more water content resembles to the human skin in many properties such as plasticity, absorbency and cohesion. As with the part of absorbency, oxygen, nitrogen and carbon dioxide can diffuse into the epidermis in small amount constituting to high content of water in the skin. [22] The similar is the case with the clay that it contains 40% of natural water that fills up all the minute pores between all the mineral grains. [21] The plasticity of the clay and the human skin depends on the proportion of water content in them which makes them exhibit similar properties. Another variation from the first phase of experiments [1] is that in this project all the readings have been taken with the plastic probe thermometer. In the set of experiments done in the first phase where the metal probe was used, there were fluctuations in the thermometer display as the probe experienced significant interference from the incident RF radiation from the Autotherm unit. We tried to cover the part of the 35 metal probe that was exposed to the microwaves but that did not help to stop the fluctuations with the thermometer. So, it was decided to use a probe which should not act as an antenna to the thermometer. We researched and got a digital thermometer with the plastic probe which should not fluctuate with the microwaves from the Autotherm unit. The readings were taken by making a closed circuit so that the microwaves remain in the finite area. The system was covered on all sides with absorbers. All the readings were taken with utmost accuracy taking in consideration the time and the distance of the applicator’s head from the clay. All the readings were taking with high intensity on the Autotherm unit. 5.1 Measurements with the Probe on the Top of the Clay This set of measurement was taken by keeping the probe on the top of the clay of about 5 inches thickness and the head of the applicator a one-half inch away from the surface of the clay. The intensity of the Autotherm was kept on maximum throughout the experiment. The readings were taken every 15 minutes and were taken for 2 hours. All the readings were taken in Fahrenheit. Table 5.1 shows the reading taken with the probe on the top of clay medium. 36 Time in Minutes Temperature in Fahrenheit 0 68 15 70.5 30 72.6 45 73.4 60 74.8 75 75.5 90 76.5 105 76.9 120 76.9 Table 5.1 Variation of Temperature Rise in Thick Clay with Time (Probe on Top of Thick Clay) Fig 5.1 shows the graphical presentation of Table 5.1 Temperature in Fahrenheit Probe on top of thick clay 78 76 74 72 70 68 66 64 62 0 15 30 45 60 75 90 105 120 Temperature in 68 70.5 72.6 73.4 74.8 75.5 76.5 76.9 76.9 Fahrenheit Figure 5.1 Graphical presentation of Temperature Vs Time (Probe on Top of Thick Clay) 37 The graph shows that the temperature rises quite rapidly initially over the first 60 minutes, and then tapers off to a peak value.. As we can see there was a gradual change in the temperature with time, but after 90 minutes of the heating, not much change in the temperature could be seen. 5.2 Measurements with Probe Inserted 1 Inch From the Surface This set of readings were taken with probe inserted an inch below the surface of the five inch thick clay. With this set of readings, the head of the applicator was placed half inch above the surface of the clay. All the readings were taken with interval of 15 minutes for 2 hours. Table 5.2 shows the readings taken with the probe an inch under the surface. Time in Minutes Temperature in Fahrenheit 0 72.9 15 83.4 30 85.1 45 86 60 86.4 75 86.5 90 88.1 105 88.3 120 88.3 Table 5.2 Variation of Temperature Rise in Clay with Time (Probe Inch below Clay Surface) 38 Fig 5.2 shows the graph of the readings from the Table 5.2 Temperature in Fahrenheit Probe inserted 1 inch under the surface of the clay 100 90 80 70 60 50 40 30 20 10 0 0 15 30 Temperature in 72.9 83.4 85.1 Fahrenheit 45 86 60 75 90 105 120 86.4 86.5 88.1 88.3 88.3 Figure 5.2 Graphical Presentation of Temperature Vs Time (Probe Inch below Clay Surface) As we can see from the graph that the change in tempertaure was only seen in the first 15 minutes of the experiment. After 30 minutes of heating the clay medium, there was a difference of only 3° F for the next 90 minutes. 5.3 Measurement with the Sliced Clay These set of readings were taken with the sliced clay that was just an inch in thickness. This step was taken as there was no much increase in the temperature after 30 minutes of heating. These readings were taken by keeping the probe on the surface of the clay and the head of the applicator not more than an inch away from the surface. These readings were taken with an interval of 15 minutes for 150 minutes. Table 5.3 shows the readings taken with the sliced clay. 39 Time in Minutes Temperature in Fahrenheit 0 69 15 69.5 30 71.2 45 72.5 60 73.2 75 73.9 90 74.9 105 75.7 120 75.7 135 76.3 150 76.3 Table 5.3 Variation of Temperature Rise in Sliced Clay with Time (Probe on Top of Clay) Temerature in Fahrenheit Readings with sliced clay 78 76 74 72 70 68 66 64 0 15 30 45 60 75 90 105 120 135 150 Temperature in 69 69.5 71.2 72.5 73.2 73.9 74.9 75.7 75.7 76.3 76.3 Fahrenheit Figure 5.3 Graphical Presentation of Temperature Vs Time (Probe on Top of Clay) 40 From the above graph, we can see that there was gradual change throughout the process till 135 minutes then there was no temperature increase. 5.4 Measurements Using a Reflector We even tried taking the readings having a reflector so that we could have the incident wave as well as the reflected so that the time required to reach the desired temperature could be reduced. In this set of readings, we have used the thick clay of 5 inches and the probe was placed on the surface of the clay. All the readings were taken by keeping the head of the applicator an inch above the surface of the clay. Table 5.4 shows the readings taken with the reflector. Time in Minutes Temperature in Fahrenheit 0 69 15 69.5 30 71.2 45 72.5 60 73.2 75 73.9 90 74.9 105 75.7 120 75.7 Table 5.4 Variation of Temperature Rise in Clay with Time Using Reflector (Probe on Top of Clay) 41 Temerature in Fahrenheit Reading with the reflector 77 76 75 74 73 72 71 70 69 0 15 30 45 60 75 90 105 120 Temperature in 71.5 72.9 74 75.5 75.7 75.7 75.7 75.7 75.7 Fahrenheit Figure 5.4 Graphical Presentation of Temperature Vs Time (Probe on Top of Clay) There was no much difference seen with the reflector. There was increase for 45 minutes then it stopped and there was no difference seen for the next 75 minutes. 5.5 Measurements with the Applicator Touching the Clay These set of readings were taken with the head of the applicator touching the clay medium. In real time scenario, the autotherm unit is used in such a way that the head of the applicator touches the human body so that there is no heat loss. Rather than having a vacuum medium, a cloth or a towel is used so that the heat could be transferred to the area to be treated without causing skin burns. 42 Time in Minutes Temperature in Fahrenheit 0 69.3 3 91.6 15 98.7 30 103.9 45 110.2 60 115.2 75 119.4 90 122.4 105 125 120 128.3 Table 5.5 Variation of Temperature Rise in Clay with the Applicator Touching (Probe on Top of Clay) Readings with applicator touching the clay Temperature in Fahrenheit 140 120 100 80 60 40 20 0 0 3 15 30 45 60 75 90 105 120 Temperature in 69.3 91.6 98.7 103.9 110.2 115.2 119.4 122.4 125 128.3 Fahrenheit Figure 5.5 Graphical Presentation of Temperature Vs Time 43 As we can see that there was a large increase of temperature as it reached form 69.3° F to 91.6° F in just 3 minutes. And we have even seen that, the temperature reached 100° F in just 20 minutes which is the requirement of the FDA ( Food and Drug Administration ) that it should not increase the pain threshold of the patient. 5.6 Measurement with the Probe Inserted One Inch From the Surface These set of readings were taken to check the depth of the heat with the applicator head touching the surface of the clay. Table 5.6 shows the readings taken by inserting the probe one inch under the surface of the clay. Time in Minutes Temperature in Fahrenheit 0 71.3 15 72.6 30 75 45 76.9 60 77.8 75 78.5 90 79.8 105 80.4 120 80.9 Table 5.6 Variation of Temperature Rise in Clay with the Applicator Touching the Surface 44 Temperature in Fahrenheit Readings with probe inserted an inch from the surface 82 80 78 76 74 72 70 68 66 0 15 Temperature in 71.3 72.6 Fahrenheit 30 75 45 60 75 90 105 120 76.9 77.8 78.5 79.8 80.4 80.9 Figure 5.6 Graphical Presentation of Temperature Vs Time (Probe One Inch below the Surface) From the above graph we can see that there is not much significant increase in temperature at a depth of one inch from the surface of the clay even after having the applicator head touching the surface of the clay. 5.7 Measurements with the Sliced Clay Toucing the Applicator Head These set of readings were taken with the sliced clay instead of a thick clay to observe the difference in the temperature rise with the thickness of the medium. Table 5.7 shows the readings taken by inserting the probe under the surface of the clay and having the applicator head touching the surface of it. 45 Time in Minutes Temperature in Fahrenheit 0 74.2 15 75.6 30 78.3 45 82.7 60 96.5 75 98.5 90 103.4 105 122.3 120 125.6 Table 5.7 Variation of Temperature Rise in Sliced Clay with the Applicator Touching the Surface Temperature in Fahrenheit Readings of the sliced clay with applicator head touching the clay 140 120 100 80 60 40 20 0 0 15 30 45 60 75 90 105 120 Temperature in 74.2 75.6 78.3 82.7 96.5 98.5 103.4 122.3 125.6 Fahrenheit Figure 5.7 Graphical Presentation of Temperature Vs time (Probe on Top of the Clay) 46 From the above graph we can see that we have achieved the desired temperature but it took 80 minutes to reach the mark of 100° F. As we have achived the desired temperature with the applicator touching the surface of the clay, we will analyze it further by plotting the readings taken at different depths of the probe in the clay. Time in Minutes Temperature in Fahrenheit Probe on clay Temperature in Fahrenheit Probe an inch Temperature in Fahrenheit Sliced clay 0 69.3 71.3 74.2 15 98.7 72.6 75.6 30 103.9 75 78.3 45 110.2 76.9 82.7 60 115.2 77.8 96.5 75 119.4 78.5 98.5 90 122.4 79.8 103.4 105 125 80.4 122.3 120 128.3 80.9 125.6 Table 5.8 Variation of Temperature Rise in Clay at Different Depths with the Applicator Touching the Surface 47 140 120 100 Temperature in Fahrenheit (Probe touching the clay) 80 60 40 20 Temperature in Fahrenheit (probe 1 inch inside the clay) 0 Temperature in Fahrenheit (applicator on sliced clay) Figure 5.8 Graphical Presentation of Temperature Vs time (for all three Depths) From the above graph we can see that the temperature rise in the thick clay and the sliced clay when the probe is inserted under the surface of the clay reaches the same mark of 125° F. But the probe when inserted an inch under the surface of the clay reaches not more than 90° F in 120 minutes. 5.8 Measurements of the Clay with Water These set of experiments were taken by adding more water to the clay to check the difference in rise of temperature having water as a medium. Table 5.8 shows the readings taken by having the applicator head one inch away from the surface of the clay and the probe inserted under the surface. 48 Time in Minutes Temperature in Fahrenheit 0 70.5 15 71.3 30 73.9 45 75.4 60 76.9 75 78.5 90 79.2 105 79.8 120 80.9 Table 5.9 Variation of Temperature Rise in Clay with Time (One Inch below Clay Surface with Water) Fig 5.8 shows the graph of the readings from the table 5.8 Temperature in Fahrenheit Readings with water on the clay 82 80 78 76 74 72 70 68 66 64 0 15 30 45 60 75 90 105 120 Temperature in 70.5 71.3 73.9 75.4 76.9 78.5 79.2 79.8 80.9 Fahrenheit Figure 5.9 Graphical Presentation of Temperature Vs Time (One Inch below Clay Surface) 49 We can see from the above graph that there was no significant increase in temperature by having more water content in the medium. Though there was a gradual change in temperature with the increase in time, the desired temperature was not achieved. From the readings that were taken in this phase of experiment, we have noticed that we achieved the desired temperature twice during the process. 50 Chapter 6 CONCLUSION The hyperthermia treatment system and its effects were thoroughly studied and analyzed. The aspects and the ways hyperthermia is used in the treatment of tumor were reviewed. The applications and benefits of hyperthermia alone and along with radiation in cancer treatment is studied. All simulations were performed using 27.12 MHz Mettler Diathermy Equipment. The duration and intensity of the system were constantly monitored during the experiments. Measurements were successfully taken at intervals of 15 minutes without allowing the clay to cool down so that to avoid the heat loss. The conditions close to hyperthermia treatment in real world were created. The replacement of hardening clay with non-hardening clay was better idea, as it made us understand the effect of the heat on the object which resembled more to the human body. The digital thermometer of the plastic probe gave us more accurate reading than the previously used thermometer of the metal probe. Initially we faced difficulties as the reading on the thermometer was fluctuating because of the interference caused by the microwaves with the metal probe of the thermometer. Though we took intense care in collecting the data the interference had a major effect on the measurements. This problem was addressed as we came to know that the treatment is done in real time by using cloth as medium between the applicator and the patient’s body and not the air. We took various measurements by varying the medium of transmission such as air, water and by variation of length of the clay. We also used 51 reflector in different positions. These measurements we took gave us varied responses as we were not able to generate the high temperature required for hyperthermia. The major success in this project came from two changes: Firstly, we changed the thermometer with the metal probe to another thermometer with a non-metallic probe, and secondly when we heated the clay directly without any medium. The temperature variation was very high as the measurements reached 128.3° F. This also satisfied the regulation set by the Food and Drug Administration (FDA) as the temperature reached 100° F in 23 minutes. Though we were not be able to replicate exactly how the human body responds to heat, we concluded that we made a considerable progress in understanding the way the heat can be effectively transmitted using the autotherm. Further research on this topic can include a different medium between the applicator and the clay so that it can transfer most of the heat. This was all part of the research to get an insight in this treatment. As we work towards the goal of cure for cancer, this effort will help understand the use of Mettler Autotherm in the treatment of hyperthermia. 52 REFERENCES 1. Beena Roshini John William, “Experimental Study of Heating effects of 27.12 MHz Diathermy unit on clay models for Hyperthermia applications in cancer treatment”, Summer 2009; 37-45. 2. http://www.cancer.gov/clinical_trials/ 3. Hyperthermia” http://www.geocities.com/hotsprings/villa/5443/alts/hytherm.html 4. National Cancer Institute (NCI) Resources, ‘Hyperthermia in Cancer Treatment: Questions and Answers’, 2004. 5. http://www.cancercenter.com/glossary/local-hyperthermia.cfm 6. Wust P et al., ‘Hyperthermia in combined treatment of cancer’, The Lancet Oncology, 2002; 3: 487–497 7. http://www.cancer.org/docroot/ETO/content/ETO_1_2x_Hyperthermia.asp 8. http://www.redblufftumor.com/hyperthermia.htm\ 9. http://spie.org/x31557.xml?ArticleID=x31557 10. http://www.steamsaun.com/hyprthr.html 11. http://bmb.oxfordjournals.org/cgi/content/full/83/1/379\\ 12. Hildebrandt B, Wust P, Ahlers O, et al., ‘The cellular and molecular basis of hyperthermia’, Critical Reviews in Oncology/Hematology, 2002; 43:33–56. 13. http://www.aetna.com/cpb/medical/data/500_599/0540.html 14. http://www.encyclopedia.com/doc/1G2-3435100264.html 15. High - Frequency Electrical Equipment in Hospitals; 1970, NFPA No. 76CM, Part III, Section 31, National Fire Protection Association, Boston, Massachusetts 53 16. http://www.scribd.com/doc/6130660/Microwave-Diathermy 17. Gian Franco Baronzio, E Dieter Hager .,’Hyperthermia in cancer treatment a primer’, Georgetown, Tex. : Landes Bioscience, 2006: 79-83 18. http://www.hc-sc.gc.ca/ewh-semt/pubs/radiation/83ehd-dhm98/fields-rf-champseng.php 19. http://www.mdsr.ecri.org/summary/detail.aspx?doc_id=8276 precautions Health and Welfare Canada. Health Protection Branch. Guidelines for microwave diathermy devices. Information Letter, No. 585; Sep 19, 1980. Fire hazard from shortwave and microwave diathermy units. Alert-Medical Devices 1979, No. 24; Aug 24. 20. http://www.mettlerelectronics.com/Brochures/300brochure.pdf 21. http://www.es.ucl.ac.uk/schools/UCL/clays.htm 22. Connor, Steven: The book of skin, Cornell University Press, 2003, pg. 176 23. http://www.drsfostersmith.com/product/prod_display.cfm?pcatid=12089