RAD 254 Chapter 14 Control of Scatter Break down into: Those that reduce patient dose and those that are geometrical in nature and those not 3 factors affecting scatter (primary) • Increased kVp • Increased field size • Increased patient thickness Spatial Resolution & Contrast Resolution • Spatial resolution may be thought of as geometric in nature (focal spot size, emission spectrum, etc. dealing with geometric image formation • Contrast resolution – driven by scatter and other sources of “noise” Scatter • Increased field sizes = more scatter – collimation is the most readily available and easiest thing to lower the amount of scatter • Patient thickness also increases scatter – compression may be used to help avoid this (IVP’s and mammo are examples) Beam restricting devices limit the radiation to the patient • Aperature diaphram (size and resultant field size are a direct proportion – draw the damn picture and figure the problems) • Cones and cylinders – great for absorbing scatter, but are circular shaped = great for improving contrast and removing scatter BUT requires much more mAs as a result Variable aperture diaphram • Mandated in 1974 by the US Dept of Food and Drug Administration (mandate removed later) – Positive Beam Limitation Devices (PBL’s) • Automatically collimate to the size of cassette/receptor in the bucky and CANNOT be a bigger size than the film/receptor Filtration • Filtration also will decrease the low energy rays and limit patient dose and some scatter The GRID Only “FORWARD” scatter is of any benefit to the radiographic image – all other scatter degrades the image! Scatter = LOWER contrast • Using a grid (alternating strips of fine leaded strips with alternating radiolucent interspace material) can effectively reduce the amount of ANGELED scatter from reaching the film/recepter Grid terms • Grid ratio = height of the lead lines divided by the interspace WIDTH • Grid frequency/lines per inch = the more lines per inch, the more clean up • Grid clean up = scatter w/o a grid vs scatter reaching film with a grid AKA “Contrast Improvement Factor” • Grid function = improve image contrast Bucky Factor • Refers to the AMOUNT of radiation to the patient with a grid vs W/O a grid. – Higher the grid ratio the higher the “bucky factor” – The higher the kVp, the higher the “bucky factor” • Grid weight refers to how heavy it is – duh – the more lead, the heavier it is Grid Types • Parallel • Crossed (cross-hatch) • Focused – Focused - crossed Grid Problems • Grid cut-off = short SID’s result in the vertical, parallel strips absorbing the “diverging” beam at the outer margins of the grid/film/rescepter MOST pronounced at short SID’s • Most grid problems are “positioning” related – – – – Uneven grid/off level grid Off centered (lateral decentering) Off focus grid Upside down focused grid Focused Grid Misalignment • Off level = grid cutoff across image; underexposed image (light) • Off center = ditto • Off focus = CR centered to one side or the other of a focused grid • Upside down grid = severe grid cut-off (no density) at both sides of the image Grid Ratio Selection • 8:1 grid is the most widely used • Grid ratio is kVp driven – Higher kVp’s warrant higher grid ratios – Higher grid ratios = higher patient dose (more radiation needed to produce an image) – As kVp increases past MAXIMUM OPTIMUM kVp, patient dose INCREASES mAs – Grid considerations As grid ratio increases, so much mAs • • • • 5:1 grid = 2 X mAs 8:1 grid = 4 X mAs 12:1 grid = 5 X mAs 16:1 grid = 6 X mAs Air gap technique • By allowing the scatter radiation to “diffuse” in the atmosphere after the patient but BEFORE the film results in a higher contrast image as the scatter diffuses and does NOT reach the film – C-spine lateral is a good example of this