1 Pressure Measurements 2011 Pressure Measurements Lecture Notes Systems & Biomedical Engineering Department Faculty of Engineering, Cairo University Prof. Bassel Tawfik Biomedical Measurements 1/1/2011 Pressure Measurements 2 Lecture Outline 1. 2. 3. 4. 5. 6. Definition of pressure Terminology Strain gages Resonant wire Piezo-electric transducers Optical 1. Pressure 1.1 Definitions Types of Pressure Measurements There are three types (or definitions) of pressure: absolute, differential, and gauge. Absolute pressure is measured relative to a perfect vacuum. An example is atmospheric pressure. A common unit of measure is pounds per square inch absolute (psia). Differential pressure is the difference in pressure between two points of measurement. This is commonly measured in units of pounds per square inch differential (psid). DP transducers are often used in flow measurement where they can measure the pressure differential across a venturi, orifice, or any other type of flow resistances. The detected pressure differential is related to fluid velocity and therefore to volumetric flow. Last, gage pressure is measured relative to ambient pressure. Blood pressure is one example. Common measurement units are pressure per square inch gauge (psig). Intake manifold vacuum in an automobile engine is an example of a vacuum gauge measurement (vacuum is negative gauge pressure). Figure 7. The same sensor can be used for all three types of pressure measurement; only the references differ. The three types of measurements are shown in Figure 7. Note that the same sensor may be used for all three types; only the reference is different. Differential pressures may be measured anywhere in the range—above, below, and around atmospheric pressure. Biomedical Measurements | Bassel Tawfik 3 Pressure Measurements Pressure Units As previously noted, pressure is force per unit area and historically a great variety of units have been used, depending on their suitability for the application. For example, blood pressure is usually measured in mmHg because mercury manometers were used originally. Atmospheric pressure is usually expressed in in.Hg for the same reason. Other units used for atmospheric pressure are bar and atm. The following conversion factors should help in dealing with the various units: 1 psi = 51.714 mmHg = 2.0359 in.Hg = 27.680 1 in.H2O = 6.8946 kPa 1 bar = 14.504 psi 1 atm = 14.696 psi Example: Convert 200 mmHg to psi: 200 mmHg • 1 psi/51.714 mmHg = 3.867 psi Terminology Often, the terms pressure gauge, sensor, transducer, and transmitter are used interchangeably. The term pressure gage usually refers to a self-contained indicator that converts the detected process pressure into a mechanical motion of a pointer. A pressure transducer might combine the sensor element of a gage with a mechanical-toelectrical or mechanical-to-pneumatic converter and a power supply. The system may also have a signal conditioner that converts the transducer signal into a standardized output. When transducers are directly interfaced with digital data acquisition systems and are located at some distance from the data acquisition hardware, it is preferable that the output voltage signal is high enough in order to be immune from electromagnetic interference (EMI) during its travelling path to the receiver. Accuracy: Transducer accuracy refers to the degree of conformity of the measured value to an accepted standard. It is usually expressed as a percentage of either the full scale or of the actual reading of the instrument. In case of percent-full-scale devices, error increases as the absolute value of the measurement drops. Repeatability refers to the closeness of agreement among a number of consecutive measurements of the same variable. Linearity is a measure of how well the transducer output increases linearly with increasing pressure. Biomedical Measurements | Bassel Tawfik 4 Pressure Measurements Hysteresis error describes the phenomenon whereby the same process pressure results in different output signals depending upon whether the current pressure is approached (from a lower or higher pressure). Practical Considerations Strain gage transducers are available for pressure ranges as low as 3 inches of water to as high as 200,000 psig (1400 MPa). Inaccuracy ranges from 0.1% of span to 0.25% of full scale. Additional error sources can be a 0.25% of full-scale drift over six months and a 0.25% full scale temperature effect per 1000 F. The way differential pressure transducers look like in reality Strain Gages When a strain gage, as described in detail in the previous lecture, is used to measure the deflection of an elastic diaphragm or Bourdon tube, it becomes a component in a pressure transducer. Strain-gage transducers are used for narrow-span pressure and for differential pressure measurements. Essentially, the strain gage is used to measure the Figure 3-4: Strain-Gage Based Pressure Cell displacement of an elastic diaphragm due to a difference in pressure across the diaphragm. These devices can detect gage pressure if the low pressure port is left open to the atmosphere. If the low pressure side is a sealed vacuum reference, the transmitter will act as an absolute pressure transmitter. Biomedical Measurements | Bassel Tawfik 5 Pressure Measurements Capacitance Type Capacitance pressure transducers were originally developed for use in low vacuum research. This capacitance change results from the movement of a diaphragm element (Figure 3-5). The diaphragm is usually metal or metalcoated quartz and is exposed to the process pressure on one side and to the reference pressure on the other. Depending on the type of pressure, the capacitive transducer can be either an absolute, gauge, or differential pressure transducer. Figure 3-5: Capacitance-Based Pressure Cell In a capacitance-type pressure sensor, a high-frequency, high-voltage oscillator is used to charge the sensing electrode elements. In a two-plate capacitor sensor design, the movement of the diaphragm between the plates is detected as an indication of the changes in process pressure. As shown in Figure 3-5, the deflection of the diaphragm causes a change in capacitance that is detected by a bridge circuit. This circuit can be operated in either a balanced or unbalanced mode. In balanced mode, the output voltage is fed to a null detector and the capacitor arms are varied to maintain the bridge at null. Therefore, in the balanced mode, the null setting itself is a measure of process pressure. When operated in unbalanced mode, the process pressure measurement is related to the ratio between the output voltage and the excitation voltage. Capacitance pressure transducers are widespread in part because of their wide rangeability, from high vacuums in the micron range to 10,000 psig (70 MPa). Differential pressures as low as 0.01 inches of water can readily be measured. And, compared with strain gage transducers, they do not drift much. Better designs are Biomedical Measurements | Bassel Tawfik 6 Pressure Measurements available that are accurate to within 0.1% of reading or 0.01% of full scale. A typical temperature effect is 0.25% of full scale per 1000 F. Capacitance-type sensors are often used as secondary standards, especially in lowdifferential and low-absolute pressure applications. They also are quite responsive, because the distance the diaphragm must physically travel is only a few microns. Newer capacitance pressure transducers are more resistant to corrosion and are less sensitive to stray capacitance and vibration effects that used to cause "reading jitters" in older designs. Potentiometric The potentiometric pressure sensor provides a simple method for obtaining an electronic output from a mechanical pressure gauge. The device consists of a precision potentiometer, whose wiper arm is mechanically linked to a Bourdon or bellows element. The movement of the wiper arm across the potentiometer converts the mechanically detected sensor deflection into a resistance measurement, using a Wheatstone bridge circuit (Figure 3-6). Figure 3-6: Potentiometric Pressure Transducer The mechanical nature of the linkages connecting the wiper arm to the Bourdon tube, bellows, or diaphragm element introduces unavoidable errors into this type of measurement. Temperature effects cause additional errors because of the differences in thermal expansion coefficients of the metallic components of the system. Errors also will develop due to mechanical wear of the components and of the contacts. Potentiometric transducers can be made extremely small and installed in very tight quarters, such as inside the housing of a 4.5-in. dial pressure gauge. They also provide a strong output that can be read without additional amplification. This permits them to be used in low power applications. They are also inexpensive. Potentiometric transducers can detect pressures between 5 and 10,000 psig (35 KPa to 70 MPa). Their accuracy is between 0.5% and 1% of full scale, not including drift and the effects of temperature. Biomedical Measurements | Bassel Tawfik 7 Pressure Measurements Resonant Wire The resonant-wire pressure transducer was introduced in the late 1970s. In this design (Figure 3-7), a wire is gripped by a static member at one end, and by the sensing diaphragm at the other. An oscillator circuit causes the wire to oscillate at its resonant frequency. A change in process pressure changes the wire tension, which in turn changes the resonant frequency of the wire. A digital counter circuit detects the shift. Because this change in frequency can be detected quite precisely, this type of transducer can be used for low differential pressure applications as well as to detect absolute and gauge pressures. Figure 3-7: Resonant-Wire Pressure Transducer The most significant advantage of the resonant wire pressure transducer is that it generates an inherently digital signal, and therefore can be sent directly to a stable crystal clock in a microprocessor. Limitations include sensitivity to temperature variation, a nonlinear output signal, and some sensitivity to shock and vibration. These limitations typically are minimized by using a microprocessor to compensate for nonlinearities as well as ambient and process temperature variations. Resonant wire transducers can detect absolute pressures from 10 mm Hg, differential pressures up to 750 in. water, and gauge pressures up to 6,000 psig (42 MPa). Typical accuracy is 0.1% of calibrated span, with six-month drift of 0.1% and a temperature effect of 0.2% per 1000 F. Biomedical Measurements | Bassel Tawfik 8 Pressure Measurements Piezoelectric When pressure, force or acceleration is applied to a quartz crystal, a charge is developed across the crystal that is proportional to the force applied (Figure 3-8). The fundamental difference between these crystal sensors and static-force devices such as strain gages is that the electric signal generated by the crystal decays rapidly. This characteristic makes these Figure 3-8: Typical Piezoelectric Pressure Sensor sensors unsuitable for the measurement of static forces or pressures but useful for dynamic measurements. Piezoelectric devices can further be classified according to the measured variable, being either the crystal's electrostatic charge, its resistivity, or its resonant frequency. Depending on which phenomenon is used, the crystal sensor can be called electrostatic, piezo-resistive, or resonant. When pressure is applied to a crystal, it is elastically deformed. This deformation results in a flow of electric charge (which lasts for a period of a few seconds). The resulting electric signal can be measured as an indication of the pressure which was applied to the crystal. These sensors cannot detect static pressures, but are used to measure rapidly changing pressures resulting from blasts, explosions, pressure pulsations (in rocket motors, engines, compressors) or other sources of shock or vibration. Some of these rugged sensors can detect pressure events having "rise times" on the order of a millionth of a second. The output of such dynamic pressure sensors is often expressed in "relative" pressure units (such as psir instead of psig), thereby referencing the measurement to the initial condition of the crystal. The maximum range of such sensors is 5,000 or 10,000 psir. The desirable features of piezoelectric sensors include their rugged construction, small size, high speed, and self-generated signal. On the other hand, they are sensitive to temperature variations and require special cabling and amplification. They also require special care during installation: One such consideration is that their mounting torque should duplicate the torque at which they were calibrated (usually 30 Biomedical Measurements | Bassel Tawfik 9 Pressure Measurements in.-lbs). Another factor that can harm their performance by slowing response speed is the depth of the empty cavity below the cavity. The larger the cavity, the slower the response. Therefore, it is recommended that the depth of the cavity be minimized and not be deeper than the diameter of the probe (usually about 0.25-in.). Electrostatic pressure transducers are small and rugged. Force to the crystal can be applied longitudinally or in the transverse direction, and in either case will cause a high voltage output proportional to the force applied. The crystal's selfgenerated voltage signal is useful where providing power to the sensor is impractical or impossible. Figure 3-9: Acceleration-Compensated Piezoelectric Sensor These sensors also provide high speed responses (30 kHz with peaks to 100 kHz), which makes them ideal for measuring transient phenomena. Figure 3-9 illustrates an acceleration-compensated pressure sensor. In this design, the compensation is provided by the addition of a seismic mass and a separate "compensation crystal" of reverse polarity. These components are scaled to exactly cancel the inertial effect of the masses (the end piece and diaphragm) which act upon the pressure-sensing crystal stack when accelerated. Because quartz is a common and naturally occurring mineral, these transducers are generally inexpensive. Tourmaline, a naturally occurring semi-precious form of quartz, has sub-microsecond responsiveness and is useful in the measurement of very rapid transients. By selecting the crystal properly, the designer can ensure both good linearity and reduced temperature sensitivity. Although piezoelectric transducers are not capable of measuring static pressures, they are widely used to evaluate dynamic pressure phenomena associated with explosions, pulsations, or dynamic pressure conditions in motors, rocket engines, compressors, and other pressurized devices that experience rapid changes. They can detect pressures between 0.1 and 10,000 psig (0.7 KPa to 70 MPa). Typical accuracy is 1% full scale with an additional 1% full scale per 1000¡ temperature effect. Biomedical Measurements | Bassel Tawfik 10 Pressure Measurements Piezoresistive pressure sensors operate based on the resistivity dependence of silicon under stress. Similar to a strain gage, a piezoresistive sensor consists of a diaphragm onto which four pairs of silicon resistors are bonded. Unlike the construction of a strain gage sensor, here the diaphragm itself is made of silicon and the resistors are diffused into the silicon during the manufacturing process. The diaphragm is completed by bonding the diaphragm to an unprocessed wafer of silicon. If the sensor is to be used to measure absolute pressure, the bonding process is performed under vacuum. If the sensor is to be referenced, the cavity behind the diaphragm is ported either to the atmosphere or to the reference pressure source. When used in a process sensor, the silicon diaphragm is shielded from direct contact with the process materials by a fluid-filled protective diaphragm made of stainless steel or some other alloy that meets the corrosion requirements of the service. Piezoresistive pressure sensors are sensitive to changes in temperature and must be temperature compensated. Piezoresistive pressure sensors can be used from about 3 psi to a maximum of about 14,000 psi (21 KPa to 100 MPa). Resonant piezoelectric pressure sensors measure the variation in resonant frequency of quartz crystals under an applied force. The sensor can consist of a suspended beam that oscillates while isolated from all other forces. The beam is maintained in oscillation at its resonant frequency. Changes in the applied force result in resonant frequency changes. The relationship between the applied pressure P and the oscillation frequency is: Where TO is the period of oscillation when the applied pressure is zero, T is the period of oscillation when the applied pressure is P, and A and B are calibration constants for the transducer. These transducers can be used for absolute pressure measurements with spans from 015 psia to 0-900 psia (0-100 kPa to 0-6 MPa) or for differential pressure measurements with spans from 0-6 psid to 0-40 psid (0-40 kPa to 0-275 kPa). Biomedical Measurements | Bassel Tawfik 11 Pressure Measurements Inductive/Reluctive A number of early pressure transducer designs were based on magnetic phenomena. These included the use of inductance, reluctance, and eddy currents. Inductance is that property of an electric circuit that expresses the amount of electromotive force (emf) induced by a given rate of change of current flow in the circuit. Reluctance is resistance to magnetic flow, the opposition offered by a magnetic substance to magnetic flux. In these sensors, a change in pressure produces a movement, which in turn changes the inductance or reluctance of an electric circuit. Flush-mount pressure sensor fits 1/4-in. NPT threads. Biomedical Measurements | Bassel Tawfik 12 Pressure Measurements Figure 3-10A illustrates the use of a linear variable differential transformer (LVDT) as the working element of a pressure transmitter. The LVDT operates on the inductance ratio principle. In this design, three coils are wired onto an insulating tube containing an iron core, which is positioned within the tube by the pressure sensor. Figure 3-10: Magnetic Pressure Transducer Designs Alternating current is applied to the primary coil in the center, and if the core also is centered, equal voltages will be induced in the secondary coils (#1 and #2). Because the coils are wired in series, this condition will result in a zero output. As the process pressure changes and the core moves, the differential in the voltages induced in the secondary coils is proportional to the pressure causing the movement. LVDT-type pressure transducers are available with 0.5% full scale accuracy and with ranges from 0-30 psig (0-210 kPa) to 0-10,000 psig (0-70 MPa). They can detect absolute, gauge, or differential pressures. Their main limitations are susceptibility to mechanical wear and sensitivity to vibration and magnetic interference. Reluctance is the equivalent of resistance in a magnetic circuit. If a change in pressure changes the gaps in the magnetic flux paths of the two cores, the ratio of inductances L1/L2 will be related to the change in process pressure (Figure 3-10B). Reluctance-based pressure transducers have a very high output signal (on the order of 40 mV/volt of excitation), but must be excited by ac voltage. They are susceptible to stray magnetic fields and to temperature effects of about 2% per 1000¡ F. Because of their very high output signals, they are often used in applications where high resolution over a relatively small range is desired. They can cover pressure ranges from 1 in. water to 10,000 psig (250 Pa to 70 MPa). Typical accuracy is 0.5% full scale. Biomedical Measurements | Bassel Tawfik 13 Pressure Measurements Optical Optical pressure transducers detect the effects of minute motions due to changes in process pressure and generate a corresponding electronic output signal (Figure 3-11). A light emitting diode (LED) is used as the light source, and a vane blocks some of the light as it is moved by the diaphragm. As the process pressure moves the vane between the source diode and the measuring diode, the amount of infrared light received changes. The optical transducer must compensate for aging of the LED light Figure 3-11: Optical Pressure Transducer source by means of a reference diode, which is never blocked by the vane. This reference diode also compensates the signal for build-up of dirt or other coating materials on the optical surfaces. The optical pressure transducer is immune to temperature effects, because the source, measurement and reference diodes are affected equally by changes in temperature. Moreover, because the amount of movement required to make the measurement is very small (under 0.5 mm), hysteresis and repeatability errors are nearly zero. Optical pressure transducers do not require much maintenance. They have excellent stability and are designed for long-duration measurements. They are available with ranges from 5 psig to 60,000 psig (35 kPa to 413 MPa) and with 0.1% full scale accuracy. Practical Considerations In industrial applications, good repeatability often is more important then absolute accuracy. If process pressures vary over a wide range, transducers with good linearity and low hysteresis are the preferred choice. Ambient and process temperature variations also cause errors in pressure measurements, particularly in detecting low pressures and small differential pressures. In such applications, temperature compensators must be used. Biomedical Measurements | Bassel Tawfik 14 Pressure Measurements Thick-film silicon pressure sensor is available in ranges from 10 to 30,000 psia. Power supply variations also lower the performance of pressure transducers. The sensitivity (S) of a transducer determines the amount of change that occurs in the output voltage (VO) when the supply voltage (VS) changes, with the measured pressure (Pm) and the rated pressure of the transducer (Pr) remaining constant: In a pressure measurement system, the total error can be calculated using the rootsum-square method: the total error is equal to the square root of the sums of all the individual errors squared. Selection Criteria Pressure transducers usually generate output signals in the mV range (spans of 100 mV to 250 mV). When used in transmitters, these are often amplified to the voltage level (1 to 5 V) and converted to current loops, usually 4-20 mA dc. The transducer housing should be selected to meet both the electrical area classification and the corrosion requirements of the particular installation. Corrosion protection must take into account both splashing of corrosive liquids or exposure to corrosive gases on the outside of the housing, as well as exposure of the sensing element to corrosive process materials. The corrosion requirements of the installation are met by selecting Biomedical Measurements | Bassel Tawfik 15 Pressure Measurements corrosion-resistant materials, coatings, and by the use of chemical seals, which are discussed later in this chapter. If the installation is in an area where explosive vapors may be present, the transducer or transmitter and its power supply must be suitable for these environments. This is usually achieved either by placing them inside purged or explosion-proof housings, or by using intrinsically safe designs. Probably the single most important decision in selecting a pressure transducer is the range. One must keep in mind two conflicting considerations: the instrument's accuracy and its protection from overpressure. From an accuracy point of view, the range of a transmitter should be low (normal operating pressure at around the middle of the range), so that error, usually a percentage of full scale, is minimized. On the other hand, one must always consider the consequences of overpressure damage due to operating errors, faulty design (waterhammer), or failure to isolate the instrument during pressure-testing and start-up. Therefore, it is important to specify not only the required range, but also the amount of overpressure protection needed. Most pressure instruments are provided with overpressure protection of 50% to 200% of range (Figure 3-12). These protectors satisfy the majority of applications. Where higher overpressures are expected and their nature is temporary (pressure spikes of short duration--seconds or less), snubbers can be installed. These Figure 3-12: Bourdon Tube Overpressure Protection filter out spikes, but cause the measurement to be less responsive. If excessive overpressure is expected to be of longer duration, one can protect the sensor by installing a pressure relief valve. However, this will result in a loss of measurement when the relief valve is open. If the transmitter is to operate under high ambient temperatures, the housing can be cooled electrically (Peltier effect) or by water, or it can be relocated in an airconditioned area. When freezing temperatures are expected, resistance heating or steam tracing should be used in combination with thermal insulation. Biomedical Measurements | Bassel Tawfik 16 Pressure Measurements When high process temperatures are present, one can consider the use of various methods of isolating the pressure instrument from the process. These include loop seals, siphons, chemical seals with capillary tubing for remote mounting, and purging. Maintenance Without exception, pressure sensors require scheduled, periodic maintenance and/or recalibration. It is necessary to periodically remove the transducer from the process and to make sure that this procedure does not require shutting down the process and does not cause injury or damage. Because the process fluid may be toxic, corrosive, or otherwise noxious to personnel or the environment, it is necessary to protect against the release of such fluids during maintenance. Figure 3-13: Three-Valve Manifold for Instrument Isolation A three-way manifold (Figure 3-13) can provide such protection. In the illustration, valve P is used to isolate the process and valve D serves to discharge the trapped process fluid from the instrument into some safe containment. The purpose of valve T is to allow the application of a known calibration or test pressure to the instrument. As all the components of the manifold are pre-assembled into a compact package, space and field assembly time are saved and chances for leaks are reduced. Calibration Pressure transducers can be recalibrated on-line or in a calibration laboratory. Laboratory recalibration typically is preferred, but often is not possible or necessary. In the laboratory, there usually are two types of calibration devices: deadweight testers that provide primary, base-line standards, and "laboratory" or "field" standard calibration devices that are periodically recalibrated against the primary. Of course, these secondary standards are less accurate than the primary, but they provide a more convenient means of testing other instruments. A deadweight tester consists of a pumping piston with a screw that presses it into the reservoir, a primary piston that carries the dead weight, and the gauge or transducer to be tested (Figure 3-14). It works by loading the primary piston (of cross sectional area A), with the amount of weight (W) that corresponds to the desired calibration pressure Biomedical Measurements | Bassel Tawfik 17 Pressure Measurements (P = W/A). The pumping piston then pressurizes the whole system by pressing more fluid into the reservoir cylinder, until the dead weight lifts off its support. Today's deadweight testers are more accurate and more complex than the instrument in Figure 3-14, but the essential operating principles are the same. Sophisticated features include temperature compensation and the means to rotate the piston in its cylinder to negate the effects of friction. Figure 3-14: Dead-Weight Tester Schematic In the United States, the National Institute of Standards & Technology (NIST) provides certified weights and calibrates laboratory piston gauges by measuring the diameter of the piston. Deadweight testers can be used to calibrate at pressure levels as low as 5 psig (35 kPa) and as high as 100,000 psig (690 MPa). Tilting type, air-lubricated designs can detect pressures in the mm Hg range. NIST calibrated deadweight testers can be accurate to 5 parts in 100,000 at pressures below 40,000 psig (280 MPa). For an industrial quality deadweight tester, error is typically 0.1% of span. A typical secondary standard used for calibrating industrial pressure transducers contains a precision power supply, an accurate digital readout, and a high-accuracy resonant (quartz) pressure sensor. It is precise enough to be used to calibrate most industrial pressure transducers, but must be NIST-traceable to be used as an official calibration standard. The best accuracy claimed by the manufacturers is typically 0.05% full scale. Installation & Accessories When possible, pressure instrumentation should be installed in visible, readily accessible locations. Readouts should be located at eye elevation. Headroom should be provided for instrument removal, as well as any space for tools and test equipment that might be needed. Biomedical Measurements | Bassel Tawfik 18 Pressure Measurements In some applications, it is desirable to prevent the process fluid from coming in contact with the sensing element. The process may be noxious, poisonous, corrosive, abrasive, have the tendency to gel, freeze or decompose at ambient temperatures, or be hotter or colder than the sensor can tolerate. Other reasons for inserting accessories between the process and the pressure instrument are to filter out potentially plugging solids or to remove potentially damaging pressure spikes or vibrations. Snubbers & Pulsation Dampers An unprotected pressure sensor on the discharge of a positive displacement pump or compressor would never come to rest, and its pointer would cycle continuously. To filter out pressure spikes, or to average out pressure pulses, snubbers and Figure 3-15: Pulsation Damper & Snubber Designs pulsation dampers are installed between the process and the instrument (Figure 3-15). The first design shown in the illustration uses a corrosion-resistant porous metal filter to delay the pressure reading by about 10 seconds. Other designs provide shorter delays via fixed or variable pistons or restrictions. The advantage of an adjustable restriction is that if, for example, a pressure gauge is placed on the discharge of a compressor, one can see when the pointer cycling has stopped. Naturally, when one is interested in the measurement of fast, transient pressures (such as to initiate safety interlocks on rising pressures), snubbers must not be used, as they delay the response of the safety system. Figure 3-16: Chemical Seal Alternatives Chemical Seals The chemical seal is also known as a "diaphragm protector." Its main components (the upper and lower body and the clean-out ring) are shown in Figure 3-16A. The pressure instrument is screwed into the upper body, which can be made of standard materials because it contacts only the non-corrosive filling fluid, usually a silicone oil. The top section with the filled diaphragm capsule can be removed with the pressure instrument while the operator cleans out the material accumulated in the bottom housing. This lower body is made of "pipe specification" (process compatible) materials and can be continuously or periodically cleaned by purging. Biomedical Measurements | Bassel Tawfik 19 Pressure Measurements The seal shown in Figure 3-16A is an off-line design; an in-line design is shown in Figure 3-16B. In-line devices are less likely to plug, but the process has to be shut down if maintenance is required. The ultimate in selfcleaning designs is shown in Figure 316C, in which all sharp edges and dead-ended cavities (where solids could accumulate) have been eliminated. The flexible cylinder can be made of a variety of plastics, including PFA, and is available in spool and wafer configurations. Miniature pressure sensor fits in tight spots. As the process pressure changes, the amount of liquid displaced by the sealing diaphragm is small, and is sometimes insufficient to fill and operate bellows-type sensors. In that case, larger displacement "rolling" diaphragms are used. Volumetric seal elements (Figure 3-17) also can eliminate cavities and sharp edges where material might accumulate. They also are well suited for high pressure and high viscosity applications such as extruders. Biomedical Measurements | Bassel Tawfik 20 Pressure Measurements Figure 3-17: Volumetric Seal Element Designs Adding seals to a press measurement device can cause the following problems: Long or large bore capillaries increase the volume of the filling fluid, increasing the temperature error. Smaller diameter diaphragms are stiff and increase error, particularly at low temperatures. Filling fluid viscosity, acceptable at normal ambient temperatures, may be unacceptably high at low temperatures. Long capillary lengths or smaller bores can cause slow response. Uneven heating/cooling of seals and capillaries can cause errors. Some fill fluids expand excessively with temperature and damage the instrument by overextending the diaphragm. High temperature and/or high vacuum may vaporize the fill fluid and damage the instrument. Fluid may contract excessively at low temperatures, bottoming the diaphragm and preventing operation. Frozen fill fluid also will prevent operation. For a successful seal installation, the following must also be considered: Process and ambient temperature range. Relative elevation of the seals and the instrument and the hydrostatic head of the fill fluid. Instrument should be rezeroed after installation to correct for elevation. Temperature, pressure, and physical damage potentials during cleaning and emptying. Possible consequences of diaphragm rupture in terms of hazard and contamination. Identical seals and capillary lengths for both sides of a differential pressure device. Seal and instrument performance at maximum temperature/minimum pressure and minimum pressure/temperature combinations. Biomedical Measurements | Bassel Tawfik 21 Pressure Measurements Wet Legs & Seal Pots When one or both impulse lines to a differential pressure device are filled with a stable, process compatible fluid, the installation is called a "wet-leg" installation. The net effect of the legs' height above the instrument and specific gravity of the fluid must be considered in the calibration. Wet leg design must also allow for the filling and draining of the leg(s). Seal pots are used with wet legs when the instrument displaces a large volume of liquid as the measurement changes. A seal pot is a small pressure vessel about one quart in volume that is mounted at the top of the wet leg line. If two wet legs are used in a differential application, the pots must be mounted at the same elevation. Each pot acts as a reservoir in the impulse line where large volume changes will result in minimal elevation change so that seal liquid is not dumped into the process line and elevation shifts of the wet leg liquid do not cause measurement errors. Biomedical Measurements | Bassel Tawfik 22 Pressure Measurements A check that the answer is correct is to verify that it is in the desired units and that mmHg cancels. Pressure Sensing Pressure is sensed by mechanical elements such as plates, shells, and tubes that are designed and constructed to deflect when pressure is applied. This is Figure 8. The typical pressure sensor has three functional blocks. the basic mechanism converting pressure to physical movement. Next, this movement must be transduced to obtain an electrical or other output. Finally, signal conditioning may be needed, depending on the type of sensor and the application. Figure 8 illustrates the three functional blocks. Sensing Elements The main types of sensing elements are Bourdon tubes, diaphragms, capsules, and bellows (see Figure 9). The Bourdon tube is a sealed tube that deflects in response to applied pressure. All except diaphragms provide a fairly large displacement that is useful in mechanical gauges and for electrical sensors that require a significant movement. Mechanical Pressure Gauges In mechanical gauges, the motion created by the sensing element is read directly by a dial or pointer. These devices are typically seen in Figure 9. The basic pressure sensing element can be low-performance applications, including configured as a C-shaped Bourdon tube (A); a helical Bourdon tube (B); flat diaphragm (C); a convoluted diaphragm (D); a blood pressure measurement and automotive capsule (E); or a set of bellows (F). pressure gauges. The mechanical approach used to couple the sensing element to the readout can introduce repeatability Biomedical Measurements | Bassel Tawfik 23 Pressure Measurements Figure 10. Temperature has an effect on the offset of a pressure sensor. errors, which will be discussed later. The mechanical mass of the gauges also limits the frequency response and makes these sensors suitable only for slowly changing measurements. Electromechanical Pressure Sensors Electromechanical pressure sensors convert the applied pressure to an electrical signal. A wide variety of materials and technologies has been used in these devices, resulting in performance vs. cost tradeoffs and suitability for applications. The electrical output signal also provides a variety of choices for various applications. Sensor Effects A pressure sensor may be modeled as: VOUT = kO + k1P (13) Where: kO = offset k1 = pressure sensitivity in V/pressure unit Figure 11. The sensitivity of a pressure sensor is also affected by temperature. Biomedical Measurements | Bassel Tawfik 24 Pressure Measurements A sensor will typically exhibit temperature coefficients of offset (also called null shift) and sensitivity (see Figures 10 and 11). Linearity, on the other hand, refers to deviations from the ideal straight line described by Equation (13). One way to measure linearity is to use the least squares method, which gives a best fit straight line (see Figure 12). Figure 12. The least squares method can be used to measure a pressure sensor's linearity. Repeatability refers to the sensor's ability to produce the same output with consecutive applications of the same pressure (see Figure 13). Hysteresis refers to the ability of the sensor to give the same output when the same increasing and then decreasing pressures are applied consecutively (see Figure 14). Figure 14. Hysteresis is a sensor's ability to give the same output at a given temperature before and after a temperature cycle. Figure 13. Repeatability refers to the ability of a pressure sensor to provide the same output with successive applications of the same pressure. Temperature hysteresis refers to the ability of the sensor to give the same output at a given temperature before and after a temperature cycle. Repeatability and hysteresis effects are not easily compensated and are indicators of the basic stability of the device. Biomedical Measurements | Bassel Tawfik 25 Pressure Measurements Gauge factor is a measure of the sensitivity of a sensor. It is defined as the ratio of the change in an electrical transduction parameter over the full range of pressure to the value of that parameter at zero pressure. Thus, the gauge factor of a resistive sensor is: R/R (14) where: R = base resistance R = resistance change with full-scale pressure, for example Pressure Sensor Technologies: Potentiometric Pressure Sensors Potentiometric pressure sensors use a Bourdon tube, capsule, or bellows to drive a wiper arm on a resistive element. For reliable operation the wiper must bear on the element with some force, which leads to repeatability and hysteresis errors. These devices are very low cost, however, and are used in lowperformance applications such as dashboard oil pressure gauges. An example is shown in Figure 15. Figure 15. Potentiometric pressure sensors use a Bourdon tube, capsule, or bellows to drive a wiper arm on a resistive element. Such sensors tend to be inexpensive, but subject to repeatability and hysteresis errors. Biomedical Measurements | Bassel Tawfik 26 Pressure Measurements Inductive Pressure Sensors Several configurations based on varying inductance or inductive coupling are used in pressure sensors. They all require AC excitation of the coil(s) and, if a DC output is desired, subsequent demodulation and filtering. The linear variable differential transformer (LVDT) types have a fairly low frequency response due to the necessity of driving the moving core of the differential transformer (see Figure 16). The LVDT uses the moving core to vary the inductive coupling between the transformer primary and secondary. Figure 16. An LVDT pressure sensor, one configuration of inductive devices, drives a moving core that varies the inductive coupling between the transformer primary and secondary. Capacitive Pressure Sensors Capacitive pressure sensors typically use a thin diaphragm as one plate of a capacitor. Applied pressure causes the diaphragm to deflect and the capacitance to change. This change may or may not be linear and is typically on the order of several picofarads out of a total capacitance of 50-100 pF. The change in capacitance may be used to control the frequency of an oscillator or to vary the coupling of an AC signal through a network. The electronics for signal conditioning should be located close to the sensing element to prevent errors due to stray capacitance. Figure 17. The basic capacitive pressure sensor consists of two plates with a vacuum between them. The capacitance of two parallel plates is given by: C = µA/d (15) where: Biomedical Measurements | Bassel Tawfik 27 Pressure Measurements µ = dielectric constant of the material between the plates A = area of the plates D = spacing between the plates If the dielectric constant of the material between the plates isn't kept constant, errors may result. Capacitive absolute pressure sensors with a vacuum between the plates are ideal in this respect. Because the capacitance of this sensor depends only Figure 18. A more complex capacitive pressure on physical parameters, sensors with good sensor can be built to detect differential pressure performance can be constructed using materials with low coefficients of thermal expansion. Since the device has to be fairly large to obtain a usable signal, frequency response may be a problem in some applications. Also, low-pressure capacitive sensors exhibit acceleration and vibration sensitivity due to the necessity for a large, thin diaphragm. A basic capacitive sensor is shown in Figure 17 and a more complex differential pressure capsule is shown in Figure 18. Piezoelectric Pressure Sensors Piezoelectric elements are bi-directional transducers capable of converting stress into an electric potential and vice versa. They consist of metallized quartz or ceramic materials. One important factor to remember is that this is a dynamic effect, providing an output only when the input is changing. This means that these sensors can be used only for varying pressures. The piezoelectric element has a high-impedance output and care must be taken to avoid loading the output by the interface electronics. Some piezoelectric pressure sensors include an internal amplifier to provide an easy electrical interface (see Figure 19). Strain Gauge Pressure Sensors Strain gauge sensors originally used a metal diaphragm with strain gauges bonded to it. A strain gauge measures the strain in a material subjected to applied stress. Consider a strip of metallic material (see Figure 20) with electrical resistance given by: RO = L/WT (16) Biomedical Measurements | Bassel Tawfik 28 Pressure Measurements where: = resistivity L, W, T = length, width, thickness Figure 19. Piezoelectric sensors convert stress into an electric potential and vice versa. Sensors based on this technology are used to measure varying pressures. Figure 20. Strain gauge pressure sensors measure the strain in a material that is subjected to applied stress. Metallic strain gauges depend only on dimensional changes to produce a change in resistance. A stress applied to the strip causes it to become slightly longer, narrower, and thinner, resulting in a resistance of: Biomedical Measurements | Bassel Tawfik 29 Pressure Measurements R = (L + L) / (W R RO(1 + 3 ) W)(T - T), or (17) As might be expected, the signal due to deformation of the material is small, on the order of 0.1% of the base resistance. Figure 21. A silicon bar can be bonded to a diaphragm to yield a strain gauge sensor with a relatively high output. Semiconductor strain gauges are widely used, both bonded and integrated into a silicon diaphragm, because the response to applied stress is an order of magnitude larger than for a metallic strain gauge. When the crystal lattice structure of silicon is deformed by applied stress, the resistance changes. This is called the piezoresistive effect. Following are some of the types of strain gauges used in pressure sensors. Photo 1. IC processing is used to form the piezoresistors on the surface of a silicon wafer to fabricate an integrated piezoresistive pressure sensor. Deposited strain gauge. Metallic strain gauges can be formed on a diaphragm by means of thin film deposition. This construction minimizes the effects of repeatability and hysteresis that bonded strain gauges exhibit. These sensors exhibit the relatively low output of metallic strain gauges. Biomedical Measurements | Bassel Tawfik 30 Pressure Measurements Bonded semiconductor strain gauge A silicon bar may be bonded to a diaphragm to form a sensor with relatively high output. Making the diaphragm from a chemically inert material allows this sensor to interface with a wide variety of media (see Figure 21). Piezoresistive Integrated Semiconductor. IC processing is used to form the piezoresistors on the surface of a silicon wafer (see Photo 1). There are four piezoresistors within the diaphragm area on the sensor. Two are subjected to tangential stress and two to radial stress when the diaphragm is deflected. They are connected in a four-element bridge configuration (see Figure 22) and provide the following output: VOUT/VCC = R/R (18) where: VCC = supply voltage R = base resistance of the piezoresistor R = change with applied pressure and is typically ~2.5% of the full R Figure 22. Piezoresistive integrated semiconductor pressure sensors incorporate four piezoresistors in the diaphragm. When the diaphragm is deflected, two resistors are subjected to tangential stress and two to radial stress. The four are connected to a four-element bridge. Etching of the back of the wafer is used to form the diaphragm (see Figure 23). The high output of the bonded strain gauge is combined with the low hysteresis of the deposited strain gauge in this design, due to the integrated construction and the nearly perfect elasticity of single-crystal silicon. The Figure 23. The back of a wafer is etched out to form the cost of the sensing element is low since diaphragm of a piezoresistive pressure sensor. a large number of devices fit on a silicon wafer. Typical die size is 0.1 in. square with a 50 mil square Biomedical Measurements | Bassel Tawfik 31 Pressure Measurements diaphragm. The circuitry needed for amplification, temperature compensation, and calibration may also be included on the same IC. Various pressure ranges are accommodated by varying the diaphragm thickness and, for very low pressures, by varying the diaphragm diameter. This device can be used to construct absolute, differential, and gauge pressure sensors, depending on the reference (see Figure 24). Because the sensing element is so small (see Photo 2), the package can have a great deal Photo 2: This totally integrated silicon of mounting and pressure sensor measures 0.52 in. long port interface by 0.44 in. wide by 0.75 in. high, including the port flexibility. Also, the small size means that it has a wide frequency response and may be used for dynamic pressure measurements without concern about errors. Mechanical vibration and acceleration have a negligible effect. Figure 24. Piezoresistive pressure sensors can be configured to provide absolute, differential, or gauge pressure readings, depending on the reference. The diaphragm is shown here as it deflects under applied differential pressures. Using a Flow Sensor to Measure Pressure A recently developed micromachined flow sensor (see Photo 3) represents a new approach to low-pressure measurement in applications where a small flow across the sensing element may be tolerated. The sensor consists of a thin film thermally isolated bridge structure suspended over a cavity in the silicon IC. Photo 3. The recently developed microbridge mass airflow sensor consists of a thin thermally isolated bridge structure suspended over a cavity in the silicon IC. The chip is 67 mil square Because of its small size and excellent thermal resistance, only a few mW of power are required to achieve high air flow sensitivity. By calibrating the device with a flow restriction for a specified pressure drop, a sensor capable of measuring pressures of a few inches of water results. This is ideal for airflow measurement in HVAC applications, for example. Biomedical Measurements | Bassel Tawfik 32 Pressure Measurements The sensor operates on the principle of heat transfer due to mass airflow directed across the surface of the sensing element (see Figure 25). Output voltage varies in proportion to the mass air (or other gas flow) through the inlet and outlet ports of the package (see Photo 4). Figure 25. A silicon bar can be bonded to a diaphragm to yield a strain gauge The sensor has a unique sensor with a relatively high output. silicon chip based on recent advances in microstructure technology. It consists of a thin film, thermally isolated, bridge structure containing a heater and temperature-sensing elements. The bridge structure provides a sensitive and fast response to the flow of air over the chip. Dual sensing elements flanking a central heating element indicate direction as well as rate of flow. A 4. The output voltage of the microbridge specially designed housing precisely directs and Photo mass air flow sensor varies in proportion to the controls the airflow across the sensing mass flow of air or other gas through the inlet and ports of the package. The packaged sensor is microstructure. Highly effective thermal outlet 1.20 in. wide by 1.24 in. long by 0.61 in. high isolation for the heater and sensing resistors is attained by the etched cavity air space beneath the flow sensor bridge. The small size and thermal isolation of the microbridge mass airflow sensor are responsible for the remarkably fast response and high sensitivity to flows. Pressure Switches Pressure switches, combining a diaphragm or other pressure measuring means with a precision snap switch, can provide precise single-point pressure sensing. Alternatively, simple electronic switches may be combined with electrical sensors to construct a pressure switch with an adjustable set point and hysteresis. Biomedical Measurements | Bassel Tawfik 33 Pressure Measurements Electrical Interfacing Care must be taken with the output from a pressure sensor to avoid corrupting the signal by noise or 60 Hz AC pickup. If the signal must be run some distance to the interface circuitry, twisted and/or shielded wire should be considered. A decoupling capacitor located at the sensor and connected from the supply to ground will also filter noise, as will a capacitor from output to ground. For long runs, a current output sensor should be considered. These devices have a 2wire interface and modulate the supply current in response to applied pressure. Obviously, wire resistance has no effect and noise must change the loop current, not simply impress a voltage on the signal. The industry standard interface is: PL = 4 mA (19) PH = 20 mA (20) Where: PL = low pressure range limit PH = high pressure range limit Calibration Manufacturers of pressure sensors have elaborate calibration facilities to verify the accuracy of their production test equipment. These include dead weight testers and temperaturecontrolled servo-rebalance testers using Bourdon tubes made from high-stability quartz. Dead-Weight Tester A dead-weight tester uses calibrated weights that Figure 26. Dead-weight testers, used to calibrate pressure sensors, incorporate calibrated weights that exert force on a piston which in turn acts on a fluid to produce a test pressure. Oil-type testers like the one shown here are commonly used for high pressures; pneumatic air bearing devices are the usual choice for lower pressures. Biomedical Measurements | Bassel Tawfik 34 Pressure Measurements exert force on a piston which then acts on a fluid to produce a test pressure. For high pressures (>500 psi), oil is typically used (see Figure 26); for lower pressures, pneumatic air bearing testers are available and are much more convenient as well as less messy to use. Manometer A mercury manometer is a simple pressure standard and may be used for gauge, differential, and absolute measurements with a suitable reference. It is useful mainly for lower pressure work because the height of the column of mercury will otherwise become unwieldy. The difference in column heights gives the pressure reading (see Figure 27). Figure 27. The mercury manometer, another calibration option for pressure sensors, can be used on gauge, differential, and absolute sensors with a suitable reference. The difference between column heights gives the pressure reading. Manometers are used mainly to calibrate sensors designed to measure in the lower pressure ranges Low-Cost Calibration Many of the higher performance commercially available pressure sensors are furnished with individual test data. A sensor with excellent repeatability and hysteresis makes an excellent low-cost in-house pressure calibration reference when combined with a pneumatic pressure regulator and a source of air pressure (see Figure 28). Figure 26. Dead-weight testers, used to calibrate pressure sensors, incorporate calibrated weights that exert force on a piston which in turn acts on a fluid to produce a test pressure. Oil-type testers like the one shown here are commonly used for high pressures; pneumatic air Biomedical Measurements | Bassel Tawfik bearing devices are the usual choice for lower pressures. 35 Pressure Measurements Selection Considerations Selection of a pressure sensor involves consideration of the medium for compatibility with the materials used in the sensor, the type (gauge, absolute, differential) of measurement, the range, the type of electrical output, and the accuracy required. Manufacturer's specifications usually apply to Figure 28. A sensor with excellent repeatability and hysteresis can be combined with a pneumatic pressure regulator and a source of air pressure to yield an inexpensive in-house pressure calibration reference. a particular temperature range. If the range of operation in a given application is smaller, for example, the errors should ratio down. Total error can be computed by adding the individual errors (worst-case) or by computing the geometric sum or root sum of the squares (RSS). The latter is more realistic since it treats them as independent errors that typically vary randomly. Following is a comparison of the two methods. Given the following error terms: Linearity = 1% F.S. Null calibration = 1% F.S. Sensitivity calibration = 1% F.S. Temperature errors are sometimes given as coefficients per ºC referenced to 25ºC. Simply multiply the coefficient by the temperature range of the application to obtain the total error. Temperature error = 0.5% F.S. Repeatability and hysteresis = 0.1% F.S. (21) Worst case error is equal to the sum of all the maximum errors: Worst case error = 1 + 1 + 1 + 0.5 + 0.1 = 3.6% (22) Biomedical Measurements | Bassel Tawfik 36 Pressure Measurements Applications Industrial. Fluid level in a tank: A gauge pressure sensor located to measure the pressure at the bottom of a tank can be used for a remote indication of fluid level using the relation: h = P/ g (23) Fluid flow: An orifice plate placed in a pipe section creates a pressure drop. This approach is widely used to measure flow because the pressure drop may be kept small in comparison to some other types of flowmeters and because it is impervious to clogging, which may otherwise be a problem when measuring flow of a viscous medium or one containing particulate matter. The relation is: (24) In some cases, differential pressures of only a few inches of water are measured in the presence of common-mode pressures of thousands of pounds per square inch. These pressure sensors are built with elaborate mechanisms to prevent damage due to the high common-mode pressures and also frequently have remotely controllable pressure ranging. Automotive. A wide variety of pressure applications exist in the modern electronically controlled auto. Among the most important are: Manifold absolute pressure (MAP). Many engine control systems use the speed-density approach to intake air mass flow rate measurement. The mass flow rate must be known so that the optimum amount of fuel can be injected. MAP is used in conjunction with intake air temperature to compute the air density. This requires a 15 psia range or higher (for supercharged or turbocharged engines). It is also desirable to include an altitude correction in the control system, and this requires measurement of barometric absolute pressure (BAP). Some systems use a separate sensor, but it is more common for the MAP sensor to do double duty since it reads atmospheric pressure for two conditions. One, before the engine begins cranking and two, whenever the throttle is wide open. Engine oil pressure. Engine lubrication requires pressures of 10-15 psig. The oil pump is sized to achieve this pressure at idle and the pressure increases with engine speed. A Biomedical Measurements | Bassel Tawfik 37 Pressure Measurements potentiometric gauge or pressure switch is used for this function since precision isn't required. Evaporative purge system leak detection. To reduce emissions, modern fuel systems are not vented to the atmosphere. This means that fumes resulting from temperatureinduced pressure changes in the fuel tank are captured in a carbon canister and later recycled through the engine. Government regulations require that leaks in this system be detected by the onboard diagnostics system. One approach is to pressurize the system and measure pressure decay over a fixed time interval. A 1 psig sensor is used for this function. Tire pressure. Recent development of the "run-flat" tire has prompted development of a remote tire pressure measurement system. The reason is that a flat tire of this type is difficult to detect visually and the distance over which it can be used without any pressure is limited. Biomedical Measurements | Bassel Tawfik Pressure Measurements 38 Appendix (A) Basic Physics of Pressure Basic Physics of Pressure Sensing1 This tutorial addresses the basic physics of pressure, three types of pressure measurement, seven pressure sensor technologies, and three calibration techniques. Static Pressure Pressure, P, is defined as force, F, per unit area, A: P = F/A (1) The measurement of pressure is generally associated with fluids, either liquids or gases. A container filled with a liquid (see Figure 1) has a pressure (due to the weight of the liquid) at a point in the liquid of: P= F/ A = hw (2) Where: h = distance from the surface to the point w = weight of the liquid (most liquids are nearly incompressible) 1 Source: Sensor Magazine (1998)Nov 1, 1998 - By: Robert E. Bicking Biomedical Measurements | Bassel Tawfik 39 Pressure Measurements Figure 1. The pressure at any given point in a confined liquid is determined by the weight of the liquid and the distance from the point to the surface. The weight per unit volume, V, is given by: w = mg/V (3) Where: m = mass g = gravitational acceleration Note that this relation can be used to determine the height of the column of liquid in a tank by measuring the pressure. The density, = m/V , is given by: (4) Thus, the density of a liquid determines the pressure, P, exerted for a given height. Mercury is 13.6 3 denser than water, so would exert a pressure 13.6 3 that of water for a column of the same height. It should be noted that the pressure due to the height of a column of liquid is in addition to the atmospheric pressure acting on the surface of the liquid. The height of a column of liquid is: h = P/ g (5) Archimedes's principle states that "A body wholly or partially submerged in a liquid is buoyed up by a force equal to the weight of the liquid displaced." Given a block of material submerged in a container of liquid (see Figure 2) with area A and length L, the downward pressure exerted on the top face is: PD = h g (6) Biomedical Measurements | Bassel Tawfik 40 Pressure Measurements Figure 2. Archimedes' principle states that an object submerged in a confined liquid will be buoyed up by the weight of the liquid it displaces. and the upward pressure exerted on the bottom face is: PU = (h + L) g (7) giving a resultant pressure of: PU - PD = L g (8) and the force is equal to the volume of the block multiplied by the weight of the liquid displaced. Figure 3. The pressure transmitted to a confined liquid is normal to the liquid's surface, as can be observed by breaching one of the walls of the container. A liquid exhibits nearly no shear stress, which leads to some interesting results. Pressure is transmitted to the inside of a container normal to the surfaces, a fact that can be most easily proved by punching a hole in a container of water and observing the stream as it exits the hole (see Figure 3). This is important in the construction of dams, given that they must resist the force of water. This pressure is called the static pressure. Biomedical Measurements | Bassel Tawfik 41 Pressure Measurements Pascal's law states that an increase in pressure at any point in a liquid results in a like increase at every other point in the liquid. This principle is used in hydraulic systems such as jacks and automobile brakes and is the fluidic equivalent of the principle of the lever, which allows large forces to be generated easily by trading large movement of a small piston for small movement of a large piston (see Figure 4). Figure 4. Pascal's law states that an increase in pressure at any point in a liquid causes a corresponding increase at every other point in the liquid. Pressure in Moving Fluids The pressure in a moving fluid exerted parallel to the flow direction is called the impact pressure, PI. This is due to the kinetic energy of the fluid: PI = VO²/2 (9) where: = fluid density VO = fluid velocity Bernoulli's theorem states that for horizontal flow the following relation holds: PS = PO + PI (10) where: PS = stagnation (or total) pressure PO = static pressure This relationship is useful for determining the flow velocity in a variety of applications. Rearranging the equation gives: (11) Biomedical Measurements | Bassel Tawfik 42 Pressure Measurements Fluid flow velocity may be measured with the Pitot tube assembly shown in Figure 5. The tube facing the flow measures total pressure and the tube normal to the flow measures static pressure. This approach is used in HVAC applications and in aircraft to measure flow velocity. Figure 5. A Pitot tube assembly can be used to measure the velocity of a moving fluid. Gases Gases differ from liquids in two respects: they are very compressible, and they completely fill any closed vessel in which they are placed. The nonlinear air pressure variation with altitude shown in Figure 6 is an example of the effect of the compressibility of gases. Figure 6. The compressibility of gases is illustrated by nonlinear atmosphere pressure as a function of altitude Dynamic Effects Static pressure is measured under steady-state or equilibrium conditions, but most reallife applications deal with dynamic or changing pressure. For example, the measurement of blood pressure usually gives the two steady-state values of systolic and diastolic pressure. There is much additional information in the shape of the blood pressure signal, however, which is the reason for the monitors used in critical-care situations. To measure changing pressures, the frequency response of the sensor must be considered. As a rough approximation, the sensor frequency response should be 5-10 × the highest frequency component in the pressure signal. The frequency response is defined as the highest frequency that the sensor will measure without distortion or attenuation. Sometimes the response time is given instead. For a first-order system, they are related as follows: Biomedical Measurements | Bassel Tawfik 43 Pressure Measurements FB = 1/2 (12) where: FB = frequency where the response is reduced by 50% = time constant where the output rises to 63% of its final value following a step input change Another issue is the remote measurement of pressure where a liquid coupling medium is used. Care must be taken to purge all air because its compressibility will corrupt the waveform. Biomedical Measurements | Bassel Tawfik