In vivo dosimetry Eirik Malinen Eva Stabell Bergstrand Dag Rune Olsen In vivo dosimetry error Probalility • In vivo: In the living • Dosimetry: Estimates of radiation dose by theory and measurement • Verification of delivered dose to individual patients • Radiotherapy requires accurate dose delivery Prescribed dose Errors in patient dose • • • • • • • Patient contour / planning basis (CT images) Patient motion Organ motion Dose calculations (inhomogeneities, scatter) Patient positioning Transfer of treatment data from simulator to linac Linac settings (energy, monitor units, field size) and calibration • Beam modifiers (blocks, wedges) Dose characteristics Dose measurements Point detector beam 2D detector array wedge Entrance dose: Output, SSD Patient curvature Wedge, curvature Exit dose: Thickness, density Desired in vivo dosimeter characteristics • Accurate and precise High accuracy Low precision • Multiple readouts • Reusability • No cables • Non-destructive readout Low accuracy High precision In vivo dosimetry principles • Point detector: – Semiconductors (diodes) – Thermoluminescent crystals – EPR (electron paramagnetic resonance) sensitive materials – …. • 2D detector, (electronic) portal imaging device; EPID: – Film – Arrays (ion chambers, semiconductors) Dosimeter reading → absorbed dose • Absorbed dose, D: D RN D Ci i R: dosimeter reading ND: calibration factor Ci: correction factor Calibration • Under reference conditions: beam dosimeter Rcal dmax ion chamber Dcal water phantom Dcal ND Rcal Example – diodes spherical droplet Buildup cap Correction factors • Dosimeter reading may depend on: – Temperature – (Accumulated) Dose – Dose rate – Beam energy – Field size – ... • Accuracy may be reduced if dependence is not corrected Temperature and sensitivity, diodes Detector temperature after placing on patient Sensitivity dependence Accumulated dose and sensitivity, diodes • Regular calibration must be performed Field size and sensitivity, diodes 8 or 18 MV photons Entrance (in) or exit (out) Supralinearity, TLD Energy dependence, TLD Comparison EPR/ alanine TLD Dose rate 1 1 <1 Linearity 1 <1 1 Beam inclination 1 1> 1 Temperature ≈1 1 <1 Energy ≈1 ≈1 ≈1 Stability ≈1 ≈1 Immediate readout 2% Correction factor for Total uncertainty (following corrections) 3-4 % (~1 Gy) 2-3 % Diode Action level • Relative dose difference: Dmeasured r 1 Dprescribed • At what dose difference level should the treatment be revised? 1% ? 2.5 % ? 5 %? • Depends on: – dosimetric accuracy and precision – non-systematic errors –… Clinical example Methods Portal image profile Measured dose / prescribed dose Action level: 2.5% measured dose dose after correction r 1.008 1.2% Frequency distribution of relative dose 2D dose maps Treatment planning algorithm Portal image Collapsed cone algorithm Location of normalization point Novel methods – ”dose guided radiotherapy” prescribed isodose target →OK Backprojection of filtered dose image into patient image →correction dose image Novel methods – ”dose guided radiotherapy” bladder rectum Corrections