Atomic Absorption Spectroscopy Prof Mark A. Buntine Dr Vicky Barnett School of Chemistry University Senior College 1 Atomic Absorption Spectroscopy “This material has been developed as a part of the Australian School Innovation in Science, Technology and Mathematics Project funded by the Australian Government Department of Education, Science and Training as a part of the Boosting Innovation in Science, Technology and Mathematics Teaching (BISTMT) Programme.” 2 Professor Mark A. Buntine Badger Room 232 mark.buntine@adelaide.edu.au Atomic Absorption Spectroscopy • AAS is commonly used for metal analysis • A solution of a metal compound is sprayed into a flame and vaporises • The metal atoms absorb light of a specific frequency, and the amount of light absorbed is a direct measure of the number of atoms of the metal in the solution 4 Atomic Absorption Spectroscopy: An Aussie Invention • Developed by Alan Walsh (below) of the CSIRO in early 1950s. 5 Electromagnetic Radiation Sinusoidally oscillating electric (E) and magnetic (M) fields. Electric & magnetic fields are orthogonal to each other. Electronic spectroscopy concerns interaction of the electric field (E) with matter. 6 The Electromagnetic Spectrum • Names of the regions are historical. • There is no abrupt or fundamental change in going from one region to the next. • Visible light represents only a very small fraction of the electromagnetic spectrum. Frequency (Hz) 1020 1018 -rays X-rays 10-11 Wavelength (m) 1016 1014 1012 UV 10-8 Microwave IR 10-6 108 10-3 Visible 7 400 500 600 700 800 nm The Visible Spectrum l < 400 nm, UV 400 nm < l < 700 nm, VIS l > 700 nm, IR 8 The Electromagnetic Spectrum • Remember that we are dealing with light. • It is convenient to think of light as particles (photons). • Relationship between energy and frequency is: E h 9 Energy & Frequency • Note that energy and frequency are directly proportional. • Consequence: higher frequency radiation is more energetic. E.g. X-ray radiation ( = 1018 Hz): 4.0 x 106 kJ/mol IR radiation ( = 1013 Hz): 39.9 kJ/mol (h = 6.626 x 10-34 J.s) 10 Energy & Wavelength • Given that frequency and wavelength are related: l=c/ • Energy and wavelength are inversely proportional • Consequence: longer wavelength radiation is less energetic eg. -ray radiation (l = 10-11 m): 1.2 x 107 kJ/mol Orange light (l = 600 nm): 199.4 kJ/mol (h = 6.626 x 10-34 J.s c = 2.998 x 108 m/s) 11 Absorption of Light • When a molecule absorbs a photon, the energy of the molecule increases. Ground state photon Excited state • Microwave radiation stimulates rotations • Infrared radiation stimulates vibrations • UV/VIS radiation stimulates electronic transitions • X-rays break chemical bonds and ionize molecules 12 Absorption of Light • When light is absorbed by a sample, the radiant power P (energy per unit time per unit area) of the beam of light decreases. • The energy absorbed may stimulate rotation, vibration or electronic transition depending on the wavelength of the incident light. 13 Atomic Absorption Spectroscopy • Uses absorption of light to measure the concentration of gas-phase atoms. • Since samples are usually liquids or solids, the analyte atoms must be vapourised in a flame (or graphite furnace). 14 Absorption and Emission Excited States Ground State Absorption Emission Multiple Transitions 15 Absorption and Emission Excited States Ground State Absorption Emission 16 Atomic Absorption • When atoms absorb light, the incoming energy excites an electron to a higher energy level. • Electronic transitions are usually observed in the visible or ultraviolet regions of the electromagnetic spectrum. 17 Atomic Absorption Spectrum • An “absorption spectrum” is the absorption of light as a function of wavelength. • The spectrum of an atom depends on its energy level structure. • Absorption spectra are useful for identifying species. 18 Atomic Absorption/Emission/ Fluorescence Spectroscopy 19 Atomic Absorption Spectroscopy • The analyte concentration is determined from the amount of absorption. 20 Atomic Absorption Spectroscopy • The analyte concentration is determined from the amount of absorption. 21 Atomic Absorption Spectroscopy • Emission lamp produces light frequencies unique to the element under investigation • When focussed through the flame these frequencies are readily absorbed by the test element • The ‘excited’ atoms are unstable- energy is emitted in all directions – hence the intensity of the focussed beam that hits the detector plate is diminished • The degree of absorbance indicates the amount of element present 22 Atomic Absorption Spectroscopy • It is possible to measure the concentration of an absorbing species in a sample by applying the Beer-Lambert Law: I Abs log Io = extinction coefficient Abs cb 23 Atomic Absorption Spectroscopy • But what if is unknown? • Concentration measurements can be made from a working curve after calibrating the instrument with standards of known concentration. 24 AAS - Calibration Curve • The instrument is calibrated before use by testing the absorbance with solutions of known concentration. • Consider that you wanted to test the sodium content of bottled water. • The following data was collected using solutions of sodium chloride of known concentration Concentration (ppm) Absorbance 2 4 6 8 0.18 0.38 0.52 0.7625 Calibration Curve for Sodium A b s o r b a n c e 1.0 0.8 0.6 0.4 0.2 2 4 6 Concentration (ppm) 8 26 Use of Calibration curve to determine sodium concentration {sample absorbance = 0.65} A b s o r b a n c e 1.0 0.8 0.6 0.4 Concentration Na+ = 7.3ppm 0.2 2 4 6 Concentration (ppm) 8 27 Atomic Absorption Spectroscopy • Instrumentation • Light Sources • Atomisation • Detection Methods 28 Light Sources • Hollow-Cathode Lamps (most common). • Lasers (more specialised). • Hollow-cathode lamps can be used to detect one or several atomic species simultaneously. Lasers, while more sensitive, have the disadvantage that they can detect only one element at a time. 29 Hollow-Cathode Lamps • Hollow-cathode lamps are a type of discharge lamp that produce narrow emission from atomic species. • They get their name from the cup-shaped cathode, which is made from the element(s) of interest. 30 Hollow-Cathode Lamps • The electric discharge ionises rare gas (Ne or Ar usually) atoms, which in turn, are accelerated into the cathode and sputter metal atoms into the gas phase. 31 Hollow-Cathode Lamps 32 Hollow-Cathode Lamps • The gas-phase metal atoms collide with other atoms (or electrons) and are excited to higher energy levels. The excited atoms decay by emitting light. • The emitted wavelengths are characteristic for each atom. 33 Hollow-Cathode Lamps M* collision-induced excitation M+e M* M + Ar* M* M M* spontaneous emission M M* M + h 34 Hollow-Cathode Spectrum Harris Fig. 21-3: Steel hollowcathode 35 Atomisation • Atomic Absorption Spectroscopy (AAS) requires that the analyte atoms be in the gas phase. • Vapourisation is usually performed by: – Flames – Furnaces – Plasmas 36 Flame Atomisation • Flame AAS can only analyse solutions. • A slot-type burner is used to increase the absorption path length (recall BeerLambert Law). • Solutions are aspirated with the gas flow into a nebulising/mixing chamber to form small droplets prior to entering the flame. 37 Flame Atomisation Harris Fig 21-4(a) 38 Flame Atomisation • Degree of atomisation is temperature dependent. • Vary flame temperature by fuel/oxidant mixture. Fuel Acetylene Acetylene Acetylene Hydrogen Hydrogen Cyanogen Oxidant Temperature (K) Air 2,400 - 2,700 Nitrous Oxide 2,900 - 3,100 Oxygen 3,300 - 3,400 Air 2,300 - 2,400 Oxygen 2,800 - 3,000 Oxygen 4,800 39 Furnaces • Improved sensitivity over flame sources. • (Hence) less sample is required. • Generally, the same temp range as flames. • More difficult to use, but with operator skill at the atomisation step, more precise measurements can be obtained. 40 Furnaces 41 Furnaces 42 Inductively Coupled Plasmas • Enables much higher temperatures to be achieved. Uses Argon gas to generate the plasma. • Temps ~ 6,000-10,000 K. • Used for emission expts rather than absorption expts due to the higher sensitivity and elevated temperatures. • Atoms are generated in excited states and spontaneously emit light. 43 Inductively Coupled Plasmas • Steps Involved: – RF induction coil wrapped around a gas jacket. • Spark ionises the Ar gas. • RF field traps & accelerates the free electrons, which collide with other atoms and initiate a chain reaction of ionisation. 44 Detection • Photomultiplier Tube (PMT). pp 472-473 (Ch. 20) Harris 45 Photomultiplier Tubes • Useful in low intensity applications. • Few photons strike the photocathode. • Electrons emitted and amplified by dynode chain. • Many electrons strike the anode. 46