G16.4427 Practical MRI 1 Pulse Design for Parallel MR Transmission G16.4427 Practical MRI 1 – 12th March 2015 Outline • K-space analysis of small tip-angle excitation • RF shimming and Parallel MR Transmission • B1 mapping • Accelerated MR excitations • Homogeneity and SAR minimization G16.4427 Practical MRI 1 – 12th March 2015 K-Space Interpretation of Small Tip Angle Excitation • New point of view for analyzing selective excitation • Similar approach as that of data acquisition and reconstruction • Strictly valid only for small tip angles excitations but holds also for flip angles ~ 90° æ 0 Gix - B1, y ö æ M x ö æ Mx ö ç ÷ç ç ÷ ÷ 0 B1,x ÷ ç M y ÷ ç M y ÷ = g ç -Gix ç ÷ç ç ÷ ÷ çè M ÷ø çè B ÷ø çè M ÷ø B 0 1, y 1,x z z G16.4427 Practical MRI 1 – 12th March 2015 Bloch equation in the rotating frame (ignoring T1 and T2) Excitation K-Space M z » M0 M xy = -ig GixM xy + ig B1 M 0 M xy » M x + iM y T M xy (x) = ig M 0 ò B1 (t)e B1 » B1,x + iB1, y 0 T k(t) = -g ò G(s) ds t (parametrically describes a path through the spatial frequency space) S(k) = ò T 0 { d (k(t) - k) k(t) }dt 3 Parametric description of the unit-weight trajectory ( sampling structure) T òt - ig xi G(s) ds dt T M xy (x) = ig M 0 ò B1 (t)e- ig xik (t ) dt 0 M xy (x) = ig M 0 ò W (k)S(k)eik ix dk K W (k(t)) = B1 (t) g G(t) Spatial-frequency weighting of the k-space trajectory G16.4427 Practical MRI 1 – 12th March 2015 Application to Slice-Selective Excitation Conventional slice-selective excitation pulse sequence k-space interpretation • K-space is scanned linearly as the RF field is applied • The location in k-space at time t is the integral of the remaining gradient waveform – Origin reached when the remaining waveform integrates to zero – RF weighting is symmetric with respect to the origin – Slice profile (Fourier transform of RF weighting) is in phase • The role of the refocusing lobe is to shift the k-space origin back in the middle of the RF excitation G16.4427 Practical MRI 1 – 12th March 2015 Multiple Coil Excitations RF shimming a2, φ2 • Distinct but time-constant amplitudes and phases for each element a1, φ1 RF a3, φ3 a4, φ4 a8, φ8 a7, φ7 • Common gradient and RF waveform a5, φ5 a6, φ6 RF Parallel Transmission • Distinct and time-varying amplitudes and phases for each element • Common gradient waveform but distinct RF waveform RF RF RF a2(t), φ2(t) a6(t), φ6(t) a3(t), φ3(t) a7(t), φ7(t) RF G16.4427 Practical MRI 1 – 12th March 2015 RF RF RF Parallel RF Transmission • Parallel transmission may be used to correct RF inhomogeneities, control SAR, tailor excitations • Requires calibration of coil array excitation patterns, and operates in close analogy to parallel reception G16.4427 Practical MRI 1 – 12th March 2015 Small Flip Angle Excitation • Homogeneous volume coil excitation k-space sampling trajectory (controlled by the switching gradients) (Mxy is obtained by multiplying the profile by iγM0) m (x) = b(x) ò W (k)S(k)eik ix dk K B1 spatial weighting Question: what is the B1 spatial weighting? spatial-frequency weighting (proportional to the coil driving current) N • Transmit coil array b̂l (x) = å cn,l bn (x) n=1 L m (x) = å b̂l (x) ò Wl (k)S(k)e l=1 K ik ix dk G16.4427 Practical MRI 1 – 12th March 2015 Effective spatial weighting, to account for coupling-induced intercoil correlations Illustration of Parallel Transmission L (x) bˆl (x) Wl (k )S (k )ei 2 k x dk l 1 k f l ( x) 1 RF pulse + the gradient pulse G16.4427 Practical MRI 1 – 12th March 2015 Illustration of Parallel Transmission L (x) bˆl (x) Wl (k )S (k )ei 2 k x dk l 1 k f l ( x) 1 RF pulse + the gradient pulse + B1 weighting G16.4427 Practical MRI 1 – 12th March 2015 Illustration of Parallel Transmission L (x) bˆl (x) Wl (k )S (k )ei 2 k x dk l 1 k f l ( x) L (coils) RF pulses + the gradient pulse + B1 weighting G16.4427 Practical MRI 1 – 12th March 2015 Example: 2D Selective Excitation G16.4427 Practical MRI 1 – 12th March 2015 Example: 3D Selective Excitation G16.4427 Practical MRI 1 – 12th March 2015 Outline • K-space analysis of small tip-angle excitation • RF shimming and Parallel MR Transmission • B1 mapping • Accelerated MR excitations • Homogeneity and SAR minimization G16.4427 Practical MRI 1 – 12th March 2015 B1 Mapping • Accurate transmit RF field (B1+) or flip angle maps are needed for many MR applications. Examples? – Correct the results of quantitative methods – Validate theoretical models for EM calculations – Testing MR compatibility of implanted objects – Compensate for B1 inhomogeneities • Image-based RF field measurements are needed for in-vivo applications • Several B1 mapping techniques exists, but further improvements (time efficiency, anatomical coverage, accuracy) are needed to use them in the routine practice and for parallel transmission G16.4427 Practical MRI 1 – 12th March 2015 Multi-Point Intensity Method • Non-selective RF pulse (long TR) and FID – Signal is largely independent from T1 and T2 – S ∝sin(α) • Step through transmit voltage until the first signal maximum is found (i.e. α = 90°) • Other pulse amplitudes would then be set relative to this calibration pulse – For GRE: S Smax = B1+ + B1,max sin(g B1+t ) G16.4427 Practical MRI 1 – 12th March 2015 Double Angle Method (DAM) • Collect two scans, one of which uses twice the RF amplitude of the other. sin(2 j ) image value at unknown actual “Double angle formula” 2cos( ) j pixel j flip angle sin( j ) S j1 = m j sin(a j ) + e1 j S j 2 = m j sin(2a j ) + e 2 j object magnetization at voxel j error G16.4427 Practical MRI 1 – 12th March 2015 æ 1 S j2 ö â j = arccos ç ÷ è 2 S j1 ø Double Angle Method (DAM) • Collect two scans, one of which uses twice the RF amplitude of the other. sin(2 j ) image value at unknown actual “Double angle formula” 2cos( ) j pixel j flip angle sin( j ) S j1 = m j sin(a j ) + e1 j S j 2 = m j sin(2a j ) + e 2 j object magnetization at voxel j æ 1 S j2 ö â j = arccos ç ÷ è 2 S j1 ø error • Inefficient method (TR ≥ 5T1 required) • Performs poorly in regions of low signal • 2 ambiguities if j is too large, sensitive to noise if j is too small G16.4427 Practical MRI 1 – 12th March 2015 Phase-Based Method • Exploits the fact that rotations do not commute • Final Mxy differs by a phase that depends on the magnitude of the flip angle α αx αy α-x α-y αx αy αy αx α-y α-x αy αx G16.4427 Practical MRI 1 – 12th March 2015 3 Df @ a 2 2 Question: What are pros and cons of the phasebased method? G16.4427 Practical MRI 1 – 12th March 2015 Phase-Based Method • Exploits the fact that rotations do not commute • Final Mxy differs by a phase that depends on the magnitude of the flip angle α αx αy α-x α-y αx αy αy αx α-y α-x αy αx 3 Df @ a 2 2 • Works better for larger α and shorter pulses ( SAR limitation) • Only for 3D and sensitive to motion/flow G16.4427 Practical MRI 1 – 12th March 2015 Actual Flip Angle Imaging (AFI) • Two identical RF pulses followed by two delays of different duration (TR1 < TR2 < T1) Assumption: at the end of both TR1 and TR2 the transverse magnetization is completely spoiled (need RF spoiling with dummy repetition to reach steady state) Before each excitation pulse: M z1 = M 0 M z 2 = M0 E1,2 = e - 1- E2 + (1- E1 )E2 cos a 1- E1 E2 cos 2 a The observed signals are: S1,2 = M z1,2 e - TE T2* sin a 1- E1 + (1- E2 )E1 cos a TR1,2 T1 1- E1 E2 cos 2 a Their ratio is: S2 1- E1 + (1- E2 )E1 cos a = S1 1- E2 + (1- E1 )E2 cos a » æ ç a » arccos ç ç çè ö S2 TR2 - 1÷ S1 TR1 ÷ TR2 S2 ÷ TR1 S1 ÷ø G16.4427 Practical MRI 1 – 12th March 2015 1+ (TR2 / TR1 )cos a (TR2 / TR1 ) + cos a Any questions? G16.4427 Practical MRI 1 – 12th March 2015 Outline • K-space analysis of small tip-angle excitation • RF shimming and Parallel MR Transmission • B1 mapping • Accelerated MR excitations • Homogeneity and SAR minimization G16.4427 Practical MRI 1 – 12th March 2015 Parallel Transmit For 2D EPI Excitation L m (x) = å b̂l (x) ò Wl (k)S(k)eik ix dk K l=1 Complex-valued excitation profile In the case of a 2D EPI excitation trajectory, let’s define: fl (x, y) = òò Wl (kx ,k y )e i( k x x+k y y ) dk x dk y k x ,k y L +¥ l=1 m=-¥ Periodic excitation pattern associated with the RF pulse of the lth transmit coil m (x, y) = å b̂l (x, y) å fl (x - mD x , y) If the sampling interval in excitation k-space is sufficiently small, then Δx = 1/Δkx is big enough that all the aliasing lobes are outside the FOV: L m (x, y) = å b̂l (x, y) f l (x, y) l=1 If we undersample (i.e. use a larger sampling interval) in excitation k-space, then M lobes will alias inside the FOV: L M -1 l=1 m=0 m (x, y) = å b̂l (x, y) å f l (x - mD x , y) G16.4427 Practical MRI 1 – 12th March 2015 Accelerated Parallel MR Excitations To design our parallel transmit pulse design we need to find the periodic excitation patterns for each transmit coil such that: ( æ f (x , y ) ö 1 n n ç ÷ ... ç ÷ b̂1 (xn , yn ) ... b̂l (xn , yn ) ... b̂L (xn , yn ) ç f l (xn , yn ) ÷ = m (xn , yn ) ç ÷ ... ç ÷ ç f (x , y ) ÷ è L n n ø ) (in the case of the EPI excitation trajectory we can treat each position separately) Question: how would the equation above change for an accelerated RF excitation? G16.4427 Practical MRI 1 – 12th March 2015 Accelerated Parallel MR Excitations To design our parallel transmit pulse design we need to find the periodic excitation patterns for each transmit coil such that: ( æ f (x , y ) ö 1 n n ç ÷ ... ç ÷ b̂1 (xn , yn ) ... b̂l (xn , yn ) ... b̂L (xn , yn ) ç f l (xn , yn ) ÷ = m (xn , yn ) ç ÷ ... ç ÷ ç f (x , y ) ÷ è L n n ø ) (in the case of the EPI excitation trajectory we can treat each position separately) We can exploit the extra degrees of freedom to under sample the excitation by a factor M: æ b̂1 (xn , yn ) ç ... ç ç b̂ (x + mD , y ) x n ç 1 n ç ... ç çè b̂1 (xn + M D x , yn ) ... b̂l (xn , yn ) ... ... ... ... ... b̂l (xn + mD x , yn ) ... ... ... ... ... b̂l (xn + M D x , yn ) ... ö æ f (x , y ) ö æ m (xn , yn ) ÷ç 1 n n ÷ ç ... ÷ç ... ... ÷ ç b̂L (xn + mD x , yn ) ÷ ç fl (xn , yn ) ÷ = ç m (xn + mD x , yn ) ÷ç ÷ ç ÷ ... ... ç ÷ ç ... ÷ç ÷ ç b̂L (xn + M D x , yn ) ÷ø è f L (xn , yn ) ø è m (xn + M D x , yn ) b̂L (xn , yn ) G16.4427 Practical MRI 1 – 12th March 2015 ö ÷ ÷ ÷ ÷ ÷ ÷ ø Outline • K-space analysis of small tip-angle excitation • RF shimming and Parallel MR Transmission • B1 mapping • Accelerated MR excitations • Homogeneity and SAR minimization G16.4427 Practical MRI 1 – 12th March 2015 SAR and RF Homogeneity • SAR management and RF homogeneity are critical issues at high magnetic field strengths – SAR is a potentially elevated safety concern – B1 focusing compromises the underlying SNR increase G16.4427 Practical MRI 1 – 12th March 2015 RF Power Deposition in Multiple Coil Excitations unit current electric field L Net Electric Field E(r, pDt) = å I l ( pDt)e l (r) l=1 RF energy dissipated in Dt xp = 2 1 H s (r) E (r, pDt) dtdv = I F Ip p òò 2 V Dt electric noise fieldcovariance covariancematrix matrix 1 P H x = åI p F I p P p=1 Global SAR EPI excitation trajectory Small tip angle { } I l ( pDt) µF -1 fl (rn ) RF excitation patterns Image-domain global SAR 1 N H x = å fn F fn N n=1 G16.4427 Practical MRI 1 – 12th March 2015 f n « f (rn ) Pulse Design for SAR Reduction target excitation profile at rn Homogeneous excitation with minimum SAR weighting Optimal excitation patterns for Parallel Transmission Minimum global SAR Time-independent RF shimming ij I l ( pDt) = A l e l I( pDt) = a l I p optimal modulation Optimal modulation coefficients shared excitation profile G16.4427 Practical MRI 1 – 12th March 2015 Minimum SAR with Parallel Tx 8 coils 12 coils 20 coils SAR = 7.9 SAR = 5.5 SAR = 3.3 Parallel Transmission - 7 Tesla - No Acceleration G16.4427 Practical MRI 1 – 12th March 2015 Ultimate Basis Set SAR = 1 SAR vs. Profile Homogeneity Bo = 7T G16.4427 Practical MRI 1 – 12th March 2015 Calibrating the Phi Matrix 1 2 3 x1global é F11 F12 êF F = [ 10 0 0 ] ê 21 22 F F ê 31 32 êë F41 F 42 F13 F14 F 23 F24 F33 F34 F 43 F44 ù úé 1 ù ú = F11 úê0 0 ú êë 0 úû úû x2global é F11 F12 êF F = [ 0 10 0 ] ê 21 22 F F ê 31 32 êë F41 F 42 F13 F14 F 23 F24 F33 F34 F 43 F44 ù úé0 ù ú ê 01 ú = F 22 ú êë 0 úû úû 4 1 2 3 4 G16.4427 Practical MRI 1 – 20th March 2012 Calibrating the Phi Matrix 1 2 3 x3global é F11 F12 êF F = [ 0 0 10 ] ê 21 22 F F ê 31 32 êë F41 F 42 F13 F14 F 23 F24 F33 F34 F 43 F44 ù úé0 ù ú ê 01 ú = F33 ú êë 0 úû úû x4global é F11 F12 êF F = [ 0 0 0 1 ] ê 21 22 F F ê 31 32 êë F41 F 42 F13 F14 F 23 F24 F33 F34 F 43 F44 ù úé0 ù ú = F 44 úê0 0 ú êë 1 úû úû 4 1 2 3 4 G16.4427 Practical MRI 1 – 20th March 2012 Calibrating the Phi Matrix 1 2 x5global 3 é F11 F12 êF F = [ 110 0 ] ê 21 22 F F ê 31 32 êë F 41 F 42 F13 F14 F 23 F 24 F33 F34 F 43 F 44 ù úé 1 ù ú ê 01 ú ú êë 0 úû úû = F11 + F12 + F 21 + F 22 4 F12 = (F21 )* 1 2 3 4 x5global é F11 F12 êF F = [ 1 i 0 0 ] ê 21 22 F F ê 31 32 êë F 41 F 42 F13 F14 F 23 F 24 F33 F34 F 43 F 44 ù úé 1 ù ú ê 0i ú ú êë 0 úû úû = F11 + iF12 + iF 21 - F 22 iF12 = (iF21 )* G16.4427 Practical MRI 1 – 20th March 2012 How Many Measurements? Channel Experiment 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 1 2 3 4 é F11 F12 F13 F14 êF F F F ê F21 F22 F23 F24 ê 31 32 33 34 ë F 41 F 42 F 43 F 44 4 real diagonal elements 6 complex off-diagonal elements ù Question: ú Why 16 measurements? ú ú û 16 variables to estimate 16 (# of channels × # of channels) measurements required G16.4427 Practical MRI 1 – 20th March 2012 Simulation Results Simulation set up Measured and predicted power Measured and Predicted Power 0.3 1 0.25 Power(Watts) 2 3 4 0.2 0.15 0.1 0.05 1 2 3 4 5 Experiment Number 6 7 8 1 2 3 4 5 Experiment Number 6 7 8 % Error 3 2 1 0 Φ calibrated using simulated E fields Φ calibrated using power measurements Identical ! G16.4427 Practical MRI 1 – 20th March 2012 Experimental Set Up (7 Tesla) Directional couplers Knee setup, 8-channel parallel Tx stripline coil Power meter Rhodes & Schwarz NRP-Z11 Computer that automates the measurement RF switch National Instrument Dual 16x1 MUX G16.4427 Practical MRI 1 – 20th March 2012 In-Vivo Results Φ calibrated at 60V 40-miliseconds - measured power • predicted power Φ calibrated at 120V 40-miliseconds G16.4427 Practical MRI 1 – 20th March 2012 Uses of Phi Matrix Calibration • Prediction and real-time monitoring of global SAR • Prediction and real-time monitoring of individual channel FWD and RFL power • Real-time detection of Tx chain hardware failures • Optimization of RF pulse design for RF shimming and parallel transmission G16.4427 Practical MRI 1 – 20th March 2012 Maximum Efficiency RF Shimming Array transmit efficiency metric: Average B1+ strength squared: Total power deposition w H w H w w B1 strength squared dissipated power B 1 B1 M H 1 w H M 1C H Cw w H w 2 1 wwHl w bl,x ibl ,y Cw r E (r) Bdv 1 # of spatial l 2 V locations Maximum Efficiency RF Shimming Find w that maximize η It can be treated as a generalized eigenvalue problem: • Largest eigenvalue = maximum transmit efficiency ηmax • Corresponding eigenvector = wmax for maximum efficiency RF shimming G16.4427 Practical MRI 1 – 20th March 2012 Experiment: Hip Imaging at 7 T No Shimming Excitation: 4 ch Tx/Rx loop coils Receive: 10 ch Tx/Rx (5 loop/stripline modules) Conservative parallel transmit SAR limits were used Φ matrix computed from forward and reflected power measures Flip Angle Maps 90 Γ - matrix Φ - matrix 0 π -π RF Shim weights G16.4427 Practical MRI 1 – 20th March 2012 RF Shimming Weights Calculation B1+ map acquisition and field extraction: ~1 min Maximum efficiency weights calculation: < 1 s Φ matrix calibration: < 5 seconds G16.4427 Practical MRI 1 – 20th March 2012 Results: RF Shimming at 7 T No RF Shimming Maximum Efficiency RF Shimming Transmit Efficiency (η) 117 286 Measured Total Average Energy Deposition (Watts) 155 58.8 28.8° ± 10.7° 25.1° ± 10.9° Mean Flip angle in ROI G16.4427 Practical MRI 1 – 20th March 2012 Any questions? G16.4427 Practical MRI 1 – 12th March 2015 See you after Spring Break! G16.4427 Practical MRI 1 – 12th March 2015