Management Cover for Field Action Notes Att: All 915-248 Product: ABL700/ABL800 December 6, 2006 _________________________________________________________________________ Subject: Update on FAN 915-248 Background: We have found two issues with the current service software, which have initiated new software versions: 1. After introduction of a new type cuvette in the hemolyzer, the Complete Oximetri test in the ABL7XX and ABL8XX could fail the Cavitation absorbance test with a perfectly good cuvette. 2. Limits in the Inlet service program for ABL8XX were tighter than the limits in the normal inlet program. This could result in a failed calibration in the service program even the ABL has performed correctly in normal use. Limits mentioned in the service manual will be updated. Action: We recommend that software versions 3.834 and 5.26 are replaced with versions 3.835 and 5.27 respectively for upcoming upgrades. For analyzers already upgraded we recommend that they are upgraded at the next service visit. If the analyzer needs to have the Hemolyzer replaced the software MUST be upgraded. The software will be available under the following part numbers shortly: - ABL700 version 3.835 part number 914-848 - ABL800 version 5.27 part number 914-849 Instructions on how to download the software are included. Documentation: This does not change the originally set deadlines for the FAN. ___________________________________________________________________________ Prepared by: Christian Knudsen Service Specialist Technical Support ABL700 Software Version 3.835, 914-848 (ONLY for ABL700 analyzers running Windows 95) Download the software files: Click on the link below to download the software 3.835 or type it in your browser. http://83.90.92.170/ABL700SW3835-914-848.EXE This file is 250MB. The download time will depend of you Internet connection; see the list below to estimate the total download time: Internet connection type Time to download a 1MB file Modem 28.8 Modem 33.6 Modem 56K (46K connect) 64 Kbit ISDN 128 Kbit ISDN 256 Kbit ADSL 512 Kbit ADSL 1024 Kbit/1 Mbit ADSL 2048 Kbit/2 Mbit GS-HDSL 5 Minutes 4 Minutes, 30 seconds 3 Minutes, 10 seconds 2 Minutes, 8 seconds 1 Minute, 4 seconds 32 seconds 16 seconds 8 seconds 4 seconds Download ratio 3.3 Kb/sec 3.9 Kb/sec 6 Kb/sec 7.4 Kb/sec 14 Kb/sec 27 Kb/sec 55 Kb/sec 115 Kb/sec 230 Kb/sec ABL700 application software: - The file “ABL700-SW3835-914-848.EXE” is a self-extracting file. So run the file and select the destination for unpacking the contained files. (Must be an empty folder) - Burn the unpacked files on a CD. Make sure that the file structure of the CD looks EXACTLY as below. ABL800 FLEX Software** ABL700 Series software* Version 5.27, code number 914-849. (*The software is ONLY valid ABL700 analyzers running on Windows XP) (**For ABL800 FLEX with Crea module installed use software version 5.42) Download the software files: Click on the link below to download the software 5.26 or type it in your browser. http://83.90.92.170/ABL800FLEX-SW527-914-849.EXE This file is 185MB. The download time will depend of you Internet connection; see the list below to estimate the total download time: Internet connection type Time to download a 1MB file Modem 28.8 Modem 33.6 Modem 56K (46K connect) 64 Kbit ISDN 128 Kbit ISDN 256 Kbit ADSL 512 Kbit ADSL 1024 Kbit/1 Mbit ADSL 2048 Kbit/2 Mbit GS-HDSL 5 Minutes 4 Minutes, 30 seconds 3 Minutes, 10 seconds 2 Minutes, 8 seconds 1 Minute, 4 seconds 32 seconds 16 seconds 8 seconds 4 seconds Download ratio 3.3 Kb/sec 3.9 Kb/sec 6 Kb/sec 7.4 Kb/sec 14 Kb/sec 27 Kb/sec 55 Kb/sec 115 Kb/sec 230 Kb/sec ABL800 FLEX application software: - The file “ABL800FLEX-SW527-914-849.EXE” is a self-extracting file. So run the file and select the destination for unpacking the contained files. (Must be an empty folder) - Burn the unpacked files on a CD. Make sure that the file structure of the CD looks EXACTLY as below.