Management Cover

advertisement
Management Cover
for Field Action Notes
Att:
All
915-248
Product: ABL700/ABL800
December 6, 2006
_________________________________________________________________________
Subject:
Update on FAN 915-248
Background: We have found two issues with the current service software, which have initiated new
software versions:
1.
After introduction of a new type cuvette in the hemolyzer, the Complete Oximetri test in
the ABL7XX and ABL8XX could fail the Cavitation absorbance test with a perfectly good cuvette.
2.
Limits in the Inlet service program for ABL8XX were tighter than the limits in the normal
inlet program. This could result in a failed calibration in the service program even the ABL has performed
correctly in normal use. Limits mentioned in the service manual will be updated.
Action: We recommend that software versions 3.834 and 5.26 are replaced with versions 3.835 and 5.27
respectively for upcoming upgrades.
For analyzers already upgraded we recommend that they are upgraded at the next service visit.
If the analyzer needs to have the Hemolyzer replaced the software MUST be upgraded.
The software will be available under the following part numbers shortly:
- ABL700 version 3.835 part number 914-848
- ABL800 version 5.27 part number 914-849
Instructions on how to download the software are included.
Documentation:
This does not change the originally set deadlines for the FAN.
___________________________________________________________________________
Prepared by:
Christian Knudsen
Service Specialist
Technical Support
ABL700 Software
Version 3.835, 914-848
(ONLY for ABL700 analyzers running Windows 95)
Download the software files:
Click on the link below to download the software 3.835 or type it in your browser. http://83.90.92.170/ABL700SW3835-914-848.EXE
This file is 250MB. The download time will depend of you Internet connection; see the list below to estimate the
total download time:
Internet connection type
Time to download a 1MB file
Modem 28.8
Modem 33.6
Modem 56K (46K connect)
64 Kbit ISDN
128 Kbit ISDN
256 Kbit ADSL
512 Kbit ADSL
1024 Kbit/1 Mbit ADSL
2048 Kbit/2 Mbit GS-HDSL
5 Minutes
4 Minutes, 30 seconds
3 Minutes, 10 seconds
2 Minutes, 8 seconds
1 Minute, 4 seconds
32 seconds
16 seconds
8 seconds
4 seconds
Download
ratio
3.3 Kb/sec
3.9 Kb/sec
6 Kb/sec
7.4 Kb/sec
14 Kb/sec
27 Kb/sec
55 Kb/sec
115 Kb/sec
230 Kb/sec
ABL700 application software:
- The file “ABL700-SW3835-914-848.EXE” is a self-extracting file. So run the file and select the destination for
unpacking the contained files. (Must be an empty folder)
- Burn the unpacked files on a CD. Make sure that the file structure of the CD looks EXACTLY as below.
ABL800 FLEX Software**
ABL700 Series software*
Version 5.27, code number 914-849.
(*The software is ONLY valid ABL700 analyzers running on Windows XP)
(**For ABL800 FLEX with Crea module installed use software version 5.42)
Download the software files:
Click on the link below to download the software 5.26 or type it in your browser.
http://83.90.92.170/ABL800FLEX-SW527-914-849.EXE
This file is 185MB. The download time will depend of you Internet connection; see the list below to estimate the
total download time:
Internet connection type
Time to download a 1MB file
Modem 28.8
Modem 33.6
Modem 56K (46K connect)
64 Kbit ISDN
128 Kbit ISDN
256 Kbit ADSL
512 Kbit ADSL
1024 Kbit/1 Mbit ADSL
2048 Kbit/2 Mbit GS-HDSL
5 Minutes
4 Minutes, 30 seconds
3 Minutes, 10 seconds
2 Minutes, 8 seconds
1 Minute, 4 seconds
32 seconds
16 seconds
8 seconds
4 seconds
Download
ratio
3.3 Kb/sec
3.9 Kb/sec
6 Kb/sec
7.4 Kb/sec
14 Kb/sec
27 Kb/sec
55 Kb/sec
115 Kb/sec
230 Kb/sec
ABL800 FLEX application software:
- The file “ABL800FLEX-SW527-914-849.EXE” is a self-extracting file. So run the file and select the destination
for unpacking the contained files. (Must be an empty folder)
- Burn the unpacked files on a CD. Make sure that the file structure of the CD looks EXACTLY as below.