AARST MODEL QUALITY ASSURANCE PLAN Notes to AARST Users: This model Quality Assurance Plan (QAP) was developed by AARST for its membership to provide the beginning of a QAP for companies and individuals involved in making radon and radon decay product measurements. This document is however not complete. Each company must expand this QAP with specifics for their company and device. A standard operating procedure (SOP) for that device can be written as a separate document that is referenced in the QAP or it can be incorporated directly into the QAP. If a company has multiple devices then separate SOPs are recommended. The Quality Assurance Plan (QAP) is the written document which presents, in specific terms, the objectives, policies, organization, functional activities, and specific Quality Control (QC) measurements such as duplicates and spikes. The QAP provides management with a document that can be used to assess whether the planned QC activities are being implemented. Quality Assurance (QA) is an umbrella term that includes many activities that provide the evidence needed to establish confidence that the data obtained are of the required precision and accuracy. Quality Control (QC) is the actual measurements that are made for the purpose of assessing and monitoring data quality, such as blank, spike and duplicate measurements. Standard Operating Procedures (SOP) is the written instructions for performing QA/QC work, particularly the analytical and testing functions. The SOP describes how the QA officer monitors compliance with the SOP and recommend revisions when appropriate. The Quality Assurance Officer (QAO) is a vital link in the success of a QA program. The QA Officer is generally responsible for * writing and revising the QA procedures so they are consistent with current state and federal guidelines or requirements as well as assuring that they are clear, thorough and practical. * ensuring the implementation of the QA Plan, including making written and oral recommendations to management regarding data quality; * developing the procedures for chain-of-custody, statistical analyses, data verification, training in QA procedures, and whatever else is necessary to properly implement the QA Plan; page 1 of 2 The objective of quality assurance is to maintain a quality measurement program and to document that quality. In addition, a QA program adds greatly to an operators understanding of the methods they use and provides early detection of problems so that they can be rectified quickly. A testing company that has inadequate QA records to document the validity of measurements, especially those made during real estate transactions, may have serious liability potential. 1. The text that appears in italics (as is this note) are notes to you, and should be deleted prior to final printing. Some of the italics may precede sections which are only applicable to some measurement methods and both the italics text and some of the text that follows should be deleted. 2 This model QA plan is for testing companies and individuals and not for device manufacturers or charcoal or alpha track laboratories. 3. The text enclosed in brackets < > is replaced by the appropriate information about your organization and services. 4. Use as specific information as possible, including names of people and model names of equipment. The QA Plan is a "living document" that is frequently revised and referred to. Even handwritten changes can be used on a temporary basis, if they are initialled, dated, and the responsible officials initial and date their signatures on the signature page. 5. This QA Plan can contain all the detail in a Standard Operating Procedure (SOP) if the QAP refers to only one type of device. For multiple devices a single QAP can refer to separate SOPs written for each measurement method. The SOP describes all the detector operating procedures in sufficient detail to serve as a reference for training new employees. References to a SOP in the QA Plan should list the detector the SOP is written for and the latest revision date. (The EPA provides a brief description of what a SOP should contain in the EPA Homes Protocols [EPA 1993], in Section 4.5, page 4-8). 6. Attached to the QA Plan, as an Appendix, are copies of blank forms used routinely, such as log entry forms, customer instruction sheets or reports. 7. This plan is a guide to be molded to your needs; much more information needs to be added; for some organizations some information may be more appropriate in different sections, and in other cases some of these sections may not be appropriate. This model QAP needs to be modified in order to include more than one detector type. This model Quality Assurance Plan was written by: Melinda Ronca-Battista 14642 South 25th Place Phoenix, AR 85044 and Bill Brodhead 2844 Slifer Valley Rd. Riegelsville, PA 18077 page 2 of 2 (602) 759-1544 (610) 346-8004 page 3 of 2 <Detector Name> QUALITY ASSURANCE PLAN for <Company Organization (if applicable)> <Company Name> <Address> <Phone #> Prepared by <Preparer's Name> <Title> <Date> Revision Date: <Date of most recent revision of any page herein> Signatures and Dates: ______________________ Quality Assurance Officer < Name > ____________________________ President or Principal Investigator < Name > ___________________________________ Individual Operator or Laboratory Director < Name > Note: Use these or applicable titles, complete names, signatures, and date the signatures. These signatures represent the awareness of, approval of, and responsibility for this plan of all persons who have significant responsibility for ensuring that the provisions of this plan are implemented. CONTENTS 1.0 INTRODUCTION & DISTRIBUTION LIST ........................... Rev 1 - 6/10/94 1.1 Introduction ...................................................................................... 1-1 1.2 Distribution List ................................................................................... 1-1 2.0 DESCRIPTION OF OPERATIONS ....................................... Rev 1 - 6/10/94 2.1 Description of Equipment ................................................................. 2-1 2.1 Operation Description ....................................................................... 2-1 3.0 ORGANIZATION & RESPONSIBILITIES.............................. Rev 1 - 6/10/94 4.0 MEASUREMENT PROCEDURES ........................................ Rev 1 - 6/10/94 4.1 Measurement Definitions ................................................................. 4-1 4.2 Measurement Types ........................................................................ 4-2 4.3 Guidelines, State Regulations, & Federal Protocols ........................ 4-2 4.4 Radon Survey by Test Technician ................................................... 4-3 4.5 Measurement Location ..................................................................... 4-3 4.6 Number of Measurements ................................................................ 4-4 4.7 Measurement Exposure Time .......................................................... 4-4 4.8 Detector Non-Interference ................................................................ 4-5 4.9 Closed House Conditions ................................................................. 4-5 4.10 Informing Client about Test Conditions ............................................ 4-6 4.11 Safety ............................................................................................... 4-6 4.12 Minimum Test Condition Verification ................................................ 4-6 4.13 New Construction Test Conditions ................................................... 4-7 Section No.i Rev No. 1 Date: 6/10/94 Page ii of iii 5.0 CHAIN of CUSTODY & DOCUMENTATION ........................ Rev 1 - 6/10/94 6.0 ANALYTICAL PROCEDURES .............................................. Rev 1 - 6/10/94 7.0 CALIBRATION PROCEDURES ............................................ Rev 1 - 6/10/94 8.0 DATA CALCULATION, VALIDATION, & REPORTING ........ Rev 1 - 6/10/94 8.1 Data Calculation ................................................................................ 8-1 8.2 Data Validation .................................................................................. 8-1 8.3 Data Reporting .................................................................................. 8-1 8.4 Maintaining Records ......................................................................... 8-2 9.0 INTERNAL QUALITY CONTROL.......................................... Rev 1 - 6/10/94 9.1 Routine Instrument Checks & Preventive Maintenance ................... 9-1 9.2 Routine Spiked Measurements made to Assess.............................. 9-1 9.2 Cross Checks made to Assess Bias ................................................ 9-2 9.3 Duplicate Measurements made to Assess Precision ....................... 9-3 9.4 Exposure of Field Blanks for Passive Detectors .............................. 9-4 10.0 QUALITY ASSURANCE OBJECTIVES & PROCEDURES for ASSESSING PRECISION & BIAS ......... Rev 1 - 6/10/94 10.1 Monitoring Bias ............................................................................ 10-1 10.2 Monitoring Precision ..................................................................... 10-2 11.0 QUALITY ASSURANCE AUDITS & MANAGEMENT REPORTS .................................................. Rev 1 - 6/10/94 12.0 CORRECTIVE ACTION ........................................................ Rev 1 - 6/10/94 13.0 QUALITY ASSURANCE TRAINING ..................................... Rev 1 - 6/10/94 Section No.i Rev No. 1 Date: 6/10/94 Page iii of iii 14.0 REFERENCES ...................................................................... Rev 1 - 6/10/94 15.0 EXHIBITS ............................................................................ Rev 1 - 6/10/94 APPENDIXES 1.0 INTRODUCTION AND DISTRIBUTION LIST 1.1 Introduction This Quality Assurance Plan QAP is consistent with the guidance issued by the U.S. Environmental Protection Agency (EPA Interim Guidelines for Preparing QA Plans [EPA 1980]; EPA Requirements for QA Project Plans [EPA 1992a]). This Plan is formatted in a way that allows <organization name> staff to reference easily pertinent portions of this document. The 16 elements recommended by the EPA (EPA Interim Guidelines for Performing QA Plans [EPA 1980]) are all included although some have been combined. For example, corrective actions are included with each section for which it is appropriate, instead of discussing all corrective actions in a separate section. The nomenclature used in this QAP is appropriate for the operations of <organization name>, and every effort has been made to maintain consistency with the latest EPA guidance (QA Guidance for RMP Participants [EPA 1993 draft]). 1.2 Distribution List This report reflects current operations, and therefore is often updated and revised. The QA Officer has responsibility for incorporating changes, and ensuring that the changes are reviewed and approved by the management, as indicated by their dated signatures on the signature page. After significant revisions, revised copies of this QAP are distributed to the following key personnel: <QA officer> <Lab director> <Co. manager or Co. president> <Names of other Co. employees> 2.0 DESCRIPTION OF OPERATIONS & EQUIPMENT 2.1 Operation Description <Organization name> performs measurements of <radon and/or radon decay product> concentrations for <residences, commercial buildings, and schools>. In general, the number of measurements performed ranges from < > to < > per month. The staff consists of <staff names and titles>, and <QA officer name>, the Quality Assurance Officer. The <laboratory or office> is located at <address>. 2.2 Description of Equipment Used <This organization> measures <radon and/or RDP> using <detector name>. Note: This section includes specific model numbers, serial numbers, if appropriate, of all the equipment used. Include a brief description of the detectors and how they function. If a laboratory is used for analysis, or if analyses are conducted in different locations, this is briefly described here. 3.0 ORGANIZATION & RESPONSIBILITIES The following <company name> employees have key responsibilities in the production of <radon and/or decay product> measurement operations. Note: Describe the organization, with names, titles, and functions. The responsibilities of each person or department is briefly described. A chart may be useful. Two important components are the quality control procedures and how the Quality Assurance Officer fits into the organization. If the QA Officer makes recommendations, to whom are they made? Who has responsibility for implementing the changes? Who is responsible for documenting changes in the Standard Operating Procedures (SOP) and/or in the QA Plan? Who has responsibility for training a new employee or retraining employees in the case of changes in procedures? 4.0 MEASUREMENT PROCEDURES This section covers the procedures that are followed to ensure that the measurements are made in appropriate locations and test conditions consistent with EPA recommendations and clients' needs. 4.1 Measurement Definitions Client - Person, persons, or businesses who have contracted with <company name> to perform a radon survey in a dwelling involved in a real estate transaction. Long-Term Measurement - Any radon or radon decay product measurement that is acknowledged as appropriate and acceptable in the EPA Measurement Protocols and has a duration of more than ninety days. Long term measurements are normally made using <list long term device that your company uses> exposed continuously over the exposure period. A one year follow-up measurement can also be made by averaging the results of a series of 3 or more equally spaced short-term (2 to 7 day) measurements taken throughout the year under normal living house conditions. Long term measurements for dwellings involved in a real estate transfer are duplicate measurements in the lowest tested area. Occupant - A person living in a dwelling who may or may not be the owner of the dwelling and is responsible for the dwelling. Primary Measurement - A radon or radon decay product measurement that provides an averaged concentration over the exposure period. The detector is be located as specified in the EPA Measurement Protocols. The detector is be operated in accordance with the recommendations of the detector manufacturer or supplier. The detector exposure time is not less than the recommended time as specified in the EPA Measurement Protocols, the Citizen's Guide or any EPA real estate testing protocols. The detector is not exposed for fewer than 48 continuous hours. Radon Survey - The process of a test company following the guidelines of this QAP and making one or more primary measurements to sample and analyze the air in a dwelling, either passively or actively, so as to measure the radon or radon decay product concentration during the test period in the areas being sampled. Responsible Individual - This refers to the person or persons responsible for assuring that the test conditions required by this QAP are being followed during a radon survey. This responsible individual does not necessarily have to be the owner of the dwelling. Short-Term Testing - Any radon or radon decay product measurement that is a primary measurement and has a duration of from two to ninety days. Section No.4 Rev No. 1 Date: 6/10/94 Page 2 of 7 Test Technician - The person responsible for placing and retrieving the <radon or radon decay product> detector. This person may be either the owner, employee or subcontractor of the test company. The test technician abides by all the requirements of the state in which the test is being conducted. The test technician is under the supervision of <company name>. The test technician has, as a minimum, attended a state or federally approved radon testing course that fulfills any necessary educational requirements for state certification in the state in which the test is being performed, or has been continually employed for one year as a test technician under the supervision of a state certified company. The test technician has obtained RMP listed individual participant status or equivalent status with a certifying state agency. 4.2 Measurement Types The EPA defines three types of radon gas measurements; initial measurement, follow-up measurement and post-mitigation measurement. An initial measurement is of a relatively short-term duration (usually 2 to 7 days) to detect a radon problem in a home. Short term tests must be carried out under closedhouse conditions and in the lowest lived-in area of a house or lowest area suitable for occupancy if the house is involved in a real estate transaction. If the result of the initial measurement done under closed house conditions is below 4 pCi/L, a follow-up test is not necessary. A follow-up measurement is a measurement made to verify and characterize a radon problem as indicated by a previously made short term measurement value at or above 4 pCi/L. In general, the higher the initial measurement, the sooner the follow-up measurement is made and the shorter the exposure period in order to minimize the health risk to the residents; viz., (1) if the initial test result was between 4 and 10 pCi/L, a long term measurement lasting up to a full 12 months is recommended, (2) if the initial measurement was greater than 10 pCi/L, a short term measurement is recommended. A post-mitigation measurement is a short term measurement made in the lowest lived-in area of the dwelling or the lowest level suitable for occupancy if the dwelling is involved in a real estate transaction. The test must not be started sooner than 24 hours after the installation of the radon system. The test must have 24 hours of closed house conditions before the start of the test and closed house conditions during the test. 4.3 Guidelines, State Regulations, and Federal Protocols Federal or state regulations or certification requirements in the state in which the radon survey is being performed is followed for measurements performed in that state. All applicable existing laws, including but not limited to statutes, ordinances and Section No.4 Rev No. 1 Date: 6/10/94 Page 3 of 7 regulations, are complied with by <company name> and <company name> test technicians in all locations in which the survey is being conducted. The test company does not offer or perform any act or service contrary to law. 4.4 Radon Survey by Test Technician During a radon survey of a dwelling involved in a real estate transfer, primary measurement placement and retrieval of the detector(s) is only performed by a test technician as defined in 4.1. The test technician ensures that the client has obtained the test technician's RMP identification number <and/or state certification number if applicable>. 4.5 Measurement Location Measurement locations are chosen according to the following criteria. o The locations are consistent with the guidelines set forth by the EPA (EPA Device Protocols [EPA 1992], Section 1.2.3, page 1-3). These guidelines are paraphrased in the instructions to clients. o The purpose of the measurement and type of building dictates the measurement strategy and location; specific EPA guidance is used for measurements in homes (EPA Homes Protocols [EPA 1993]), for real estate transactions (EPA Buyer's and Seller's Guide to Radon [EPA 1993a]), and in schools (EPA Radon Measurements in Schools [EPA 1993b]). o Specific needs or requests of clients are considered. The following general criteria are to be used in selecting the location of a detector within a room. o A location must be selected where the detector is not be disturbed during the measurement period. o The detector must not be placed near drafts caused by HVAC vents, windows, and doors. Avoid locations near excessive heat, such as fireplaces or in direct sunlight. o The detector can be suspended from the ceiling if it is within the breathing zone (6' to 8' high) or placed on a shelf or table at least 50 centimeters (20 inches) above floor level and with its top face at least 10 centimeters (4 inches) from other objects. Section No.4 Rev No. 1 Date: 6/10/94 Page 4 of 7 o o o The detector must not be placed within 30 centimeters (12") of the outside walls of the house or within 90 centimeters (36") of any opening in the outside walls such as windows. Nothing must impede air flow around the detector. In general, kitchens, bathrooms, laundry rooms, root cellars, garages, crawl spaces, or sumps are not suitable measurement locations. 4.6 Number of Measurements In order to determine comprehensively a dwelling's potential to have elevated radon levels, a radon survey has a minimum of one primary measurement in each lowest structural area suitable for occupancy. An example of the required test areas would be a split level style of dwelling with a finished basement, a slab on grade room, and a room over a crawl space. Each of these three separate lowest structural areas would require one primary measurement. The measurement placement conforms to the current EPA measurement protocols for radon measurements. To ensure that the radon survey is a valid representation of the radon levels, <either an active detector or a duplicate (simultaneous) measurement made with two detectors, side by side or a measurement followed by a second measurement (sequential) is made>. If sequential measurements are made, then the results of the first measurement are not revealed until the completion of the second measurement. The requirement to use <an active detector, or simultaneous measurement or sequential measurement>, only pertains to the measurement taken in the lowest level of the dwelling. 4.7 Measurement Exposure Time The measurement exposure time is a minimum of 48 hours. Short-term measurement exposure time should be in increments of 24 hours plus or minus 1 hour for each day of exposure in order to minimize the effect of diurnal variations. This means that a three-day test should be exposed from 69 to 75 hours. The exposure time is not be less than the manufacturer's or supplier's recommendations. If the radon survey is to be a long-term measurement, closed-house conditions need not be maintained. The test company should, however, recommend to the owner or occupant of the dwelling that at least half the test period should be during the season that the dwelling is most likely be operated with closed-house conditions and that reasonable closed-house conditions should be maintained during the test period so that the results of the test are more accurate indicators of the yearly average. Long term measurements for dwellings involved in a real estate transfer are duplicate measurements in the lowest tested area. Section No.4 Rev No. 1 Date: 6/10/94 Page 5 of 7 4.8 Detector Non-Interference Test Technicians should not initiate a measurement if the occupant is moving or is planning remodeling, changes in the heating, ventilating and air conditioning (HVAC) system, or other modifications that may influence the radon concentration during the measurement period. <Detector name> should not be deployed if the occupant's schedule precludes termination of the measurement at the appropriate time. The testing device is not be moved, covered or have its performance altered during the radon survey by anyone. Examples of performance alteration would be covering the detector, moving it to another location during the test period, or opening windows during a short-term test. The responsible individual is requested to sign a non-interference agreement that indicates a knowledge of the testing conditions of this QAP and a willingness to cooperate in maintaining the required test conditions. If such an agreement cannot or will not be signed by the responsible individual, <company name> indicates in the Report why the signature was not obtained. Appendix A is an example of Non-Interference Agreement. <List other tamper methods used by your company.> 4.9 Closed-house Conditions The weather and building ventilation conditions prior to and during the measurement are in accordance with the guidelines set forth by the EPA (Device Protocols [EPA 1992], Section 1.2.2, page 1-2). These guidelines are paraphrased in the instructions to clients. Closed-house conditions require that all the windows are kept closed. External and any basement to first floor doors are closed except for normal momentary entering and exiting. Structural openings due to disrepair or structural defects that allow a significant amount of ventilation is repaired to correct their condition before initiation of closedhouse conditions. All exterior windows and doors are inspected by the test technician at the placement and retrieval of the detector. The test technician walks around the outside of the house to inspect for closed-house condition. Heating, air conditioning, and heat recovery ventilators can be operated normally. Operation of dryers, range hoods, and bathroom fans should be kept to a minimum. The responsible individual, however, should be informed that overuse of an appliance that exhausts air may influence the final readings. Section No.4 Rev No. 1 Date: 6/10/94 Page 6 of 7 Whole house fans are not operated. Portable window fans are removed from the window or sealed in place. Window air conditioning units are only operated in a recirculating mode. If the dwelling contains an air handling system, the air handling system is not set for continuous operation unless the air handling equipment is specifically used for radon control and labeled as such. Fireplaces or combustion appliances are not be operated unless they are the primary sources of heat for the dwelling. Ceiling fans, portable dehumidifiers, portable humidifiers, portable air filters and window air conditioners are not operated within twenty feet of the detector. Closed-house conditions are maintained for twelve hours prior to the start of the radon survey, if the test period is four days or fewer in duration and recommended for tests up to seven days long. <Company name> inquires with the responsible individual to determine if closed-house conditions have been maintained for the twelve hours prior to the start of the test. The test technician takes due and proper care in obtaining closed-house conditions during short-term testing in order to increase measurement reproducibility. If at the initiation of the test, the test technician discovers or visually observes that closed-house conditions were not maintained, then the radon survey is not initiated until twelve hours of prior closed-house conditions have been maintained or the radon test period is extended to four days or more with an appropriate detector after closed-house conditions are initiated. 4.10 Informing Client about Test Conditions <Company name> makes reasonable efforts to determine who is the responsible individual for the dwelling during the test period. Before any primary measurements are begun, the responsible individual is informed or attempted to be informed about the requirements of and the need for closed-house conditions as well as all other conditions for which they are responsible. A copy of the EPA document, "Home Buyers and Sellers Guide to Radon" should be given to each client. In order to inform occupants of the dwelling about the test and the conditions of the test, a "Radon Survey in Progress" notification form with the conditions of the test stated on the notification, is posted upon initiation of a short- term test in a conspicuous location of the dwelling. Appendix B is an example of a Radon Survey in Progress notification form used by <company name>. 4.11 Safety The test technician should not enter any area or perform any test that would damage property or risk the test technician's own or another's safety. If it is known that closed-house conditions are detrimental to the health of the occupants, then the radon survey using a short-term test is not done. Section No.4 Rev No. 1 Date: 6/10/94 Page 7 of 7 4.12 Minimum Test Condition Verification <Company name>'s minimum requirements for verifying test conditions is fulfilled by the following: informing the responsible individual of the test conditions, obtaining or attempting to obtain a signed non-interference agreement, conducting a visual inspection of the dwelling upon placement and retrieval of the detector, and posting a Radon Survey in Progress notification form. This QAP does not require the test technician to be responsible for inspecting for closed-house conditions 12 hours before the start of the test or between placement and retrieval of the detectors. 4.13 New Construction Test Conditions Newly constructed dwellings are tested in accordance with this QAP. The following items, if such items are part of the completed dwelling, are installed and completed before the radon survey is initiated: all insulation, all exterior doors, all windows, all fireplaces and fireplace dampers, all heating appliances, all ceiling coverings, all interior trim and coverings for the exterior walls, all exterior siding, weatherproofing and caulking. If <company name> knows construction work, which will likely affect the test results, is to be done inside the dwelling during the test period, then the <company name> schedules the test during a time when such interference is less likely to take place. 5.0 CHAIN OF CUSTODY AND DOCUMENTATION Note: This section includes the record keeping procedures, with a list of the information to be gathered with each measurement; copies or descriptions of the logbooks and forms used (reference copies in appendix); the record keeping procedures for maintaining records of QA practices and QC measurements, including calibration results, background measurement results and any changes in operators, materials, or procedures. Logbooks should be bound, and records made in pen. Every entry includes the name of the person making the entry and the date. Chain of custody for each <detector name> includes documenting the exposure location, times and dates in/out, and persons responsible for the detectors prior to, during and after their exposure, up to their <analysis or shipment to the analysis laboratory>. <Name and/or previously listed title> is responsible for ensuring that this information is properly recorded for each measurement. It is important that enough information about each measurement be recorded on the proper forms by the Field Technician or homeowner who deploys the detector (see "Operating Instructions" form in Appendix <list detector instruction sheet appendix number> and by the Laboratory Director in the permanent "Detector Placement Log Book." This data is important in case further interpretations and measurements are required at a later date. The information called for on the data form included with the detector (Appendix <list appendix number>) must be accurately completed by the <homeowner or field technician> when the detector is placed and/or retrieved. The detector serial number is recorded by <lab director> on a separate log sheet and the <homeowner or Field Technician> records a description of the location in the house where it was placed. The detector is not to be relocated during the measurement period; however, if it is necessary to do so (e.g., during a long-term test), the Field Technician should note the new location on all relevant log sheets. The information to be logged must include: o Address of measurement client and building measured. o The start and stop, date and times of each measurement. o The detector type and identification numbers. o Whether the homeowner has agreed in writing to abide by the closed house conditions (if applicable) throughout the test period and twelve hours before the test. o Exact locations of the detector. Section No.5 Rev No. 1 Date: 6/10/94 Page 2 of 2 o Other easily gathered information that may be useful, such as the type of house, type of heating system, condition of crawl space vents. Note: This next section is appropriate only for users of those methods that involve analysis (e.g., primary organizations). 6.0 ANALYTICAL PROCEDURES The measurement procedures recommended by EPA (Device Protocols [EPA 1992]) and <detector manufacturer > are followed. These include <list basic steps used to obtain a measurement>. <If separate SOP is written then it can be a brief description here with reference to the specific SOP.> Note: Include specific information on corrective action, which is what to do and who is responsible when something in the analysis goes wrong or is not as expected. (e.g., background of active monitor is approaching the activity of the air to be sampled.) Note: This section is not relevant for organizations that use passive detectors that have analysis equipment that is calibrated by the manufacturer or another company. 7.0 CALIBRATION PROCEDURES Calibration of the <primary detector name(s)> is conducted prior to initiating field measurements and at least every 12 months in accordance with the following procedures. Note: This section briefly describes the calibration process, including specific information, such as the pre-calibration performance checks (background, high voltage adjustment, flowrate, etc.) and adjustments, as well as the range of radon/RDP exposure the detector is exposed to. Factors which may not be always achieved can be stated as goals. If instruments or systems are calibrated by the manufacturer or other laboratory, their procedures for calibration can be referenced (by title and date). The documentation of the calibration facility, where exposures are performed, should be attached to the QAP or referenced. This documentation indicates the successful participation in federal intercomparison exercises or other traceability to a federal laboratory standard. 8.0 DATA CALCULATION, VALIDATION, AND REPORTING Note: This section includes a description of the methods by which the readings produced by the instrument or detector are recorded, transformed to radon concentrations, and reported to the client. 8.1 Data Calculation Note: This section is not relevant for secondary organizations that receive results from an independent laboratory. Primary organizations that are responsible for producing the result reported to clients list here the formulas and procedures used to obtain the <radon or RDP> result. 8.2 Data Validation Note: This section describes the procedures used to avoid transcription errors and to spot-check reports to clients for accuracy. <In this section describe the procedures for hand-calculating entries for errors in calculation, transcription, or in the computer database. The procedures should state how often, how many and by whom entries are checked; the frequency should be at least semi-annually, and after any change in procedure, personnel, or equipment.> A transcription error rate greater than < 5 > significant errors in 100 fields is cause for notification of the QA Officer, who investigates and recommends re-entry of data if necessary. 8.3 Data Reporting Final results are reported in units of <pCi/L or WL>, in the format provided by the <Company name> test report (see sample report in Appendix <reference appendix #>). The information furnished to clients in the test report follows the recommendations made by EPA (Homes Protocols [EPA 1993]). Values are reported to <one figure after the decimal for radon concentrations, and to three figures after the decimal for radon decay product concentrations>. <Name or listed title> is responsible for checking the accuracy and completeness of the test reports before they are sent to the client. The Report is in writing and either mailed first class, faxed or hand delivered to the client within five business days after the results are obtained by the test company. The report contains the following. o Address of measurement client and building measured. o o The start and stop, date and times of each measurement. The detector type and identification numbers. Section No.8 Rev No. 1 Date: 6/10/94 Page 2 of 3 o Whether the homeowner has agreed in writing to abide by the closed house conditions (if applicable) throughout the test period and twelve hours before the test. o Exact location of the detector. o Test Technicians EPA RMP or equivalent state certification numbers . o Any required statement by the state, EPA or this QAP. o Observed status of mitigation system if applicable. o Any observed structural openings, such as crawl space vents. o Any observed deviations from required test conditions. The test report includes an interpretation of the measurement results that uses the EPA "A Citizens Guide to Radon (Second Addition)" and/or the "Home Buyers and Sellers Guide to Radon" as its source. The measurements are reported in units that are appropriate to the measurement method. Any test report that converts measurement results to the unit of another product discloses, as a minimum, the limitations and the possibility for variations of such conversion as well as the equilibrium ratio used to make that conversion. The Report should describe the general limitations of the test such as the following statement: There is an uncertainty with any measurement result due to statistical variations and other factors such as daily and seasonal variations in radon concentrations due to changes in the weather and operation of the dwelling as well as possible interference with the necessary test conditions that may or may not influence the results. All test results include a statement which recommends that the dwelling be retested in each of the following cases whether or not the dwelling has been mitigated: o o o o o o Occupancy by a new owner Two years since the previous test A new addition is added An alteration is made that could change the ventilation pattern Major cracks or penetrations occur in the foundation walls or slab Significant nearby construction blasting or earthquakes occur Section No.8 Rev No. 1 Date: 6/10/94 Page 3 of 3 o o Changes are made or happen to an installed mitigation system Occupation of a ground contact area that was not previously tested 8.4 Maintaining Records Radon and radon decay product measurements are a radiation exposure measurement. Because the effects of radiation exposure may not become apparent until many years have passed, measurement data should be maintained permanently. 9.0 INTERNAL QUALITY CONTROL (QC) This section describes specific procedures for four types of internal quality control (QC) checks, measurements and procedures: a) Routine instrument performance checks b) Analysis of measurements made to assess relative bias c) Analysis of duplicate measurements made to assess precision. d) Laboratory analysis of blank field detectors 9.1 Routine Instrument Performance Checks and Preventive Maintenance Specific procedures relevant to quality assurance include: <Here describe those procedures conducted to check the quality of the measured result. For some methods this may be minimal, for other detectors more extensive. For example, routine active instrument performance checks may consist of checking the battery, pump flow rates, and filters, etc.> For each type of check, the following procedures are followed. <Name> is responsible for conducting the following checks. The performance checks are conducted <every measurement, once a week>. The results are documented in <where>, after each performance check. The numerical limits are that <any measured parameter (e.g., max/min flow rate, maximum background) and the procedures for corrective action if the results of the checks are not within these limits>. This documentation is <periodically> reviewed by <QA officer> and <Name or listed title> is responsible for any additional investigation and/or corrective action. Note: Include the passive or active method used below in 9.2 and delete the other method. Passive Method 9.2 Spiked Measurements Made to Assess Bias Any consistent bias of the <detector name> method is evaluated by exposing the detectors to a known concentration (spiked). Spiked measurements made to assess bias are made at a frequency of at least 3 per 100 measurements (e.g., if approximately 300 measurements are estimated for the next 6 months, then at least nine spiked Section No.9 Rev No. 1 Date: 6/10/94 Page 2 of 4 measurements are made at a radon calibration chamber.) The passive detector is exposed in a radon chamber that intercompares its results with an EPA or DOE radon chamber at least every six months or the passive detectors are carefully exposed by an independent company in a known volume chamber that contains a NIST radium standard and NIST approved procedures for such an exposure are used. The QA officer is responsible for ensuring that <chamber facility name> used by <Company name> provides documentation of its own frequent and regular successful participation in EPA and/or U.S. DOE intercomparison exercises or traceability to a NIST standard. <Chamber facility name> makes available its procedures for stringent internal QA including documentation of "traceability" to the EPA, NIST or U.S. DOE laboratory standards. A copy of the most recent federal intercomparison records of <chamber facility name> is on file. Note: It is recommended that chamber blanks and blanks to determine any exposure in transport (if relevant) be considered part of the spike procedure. The <detector name>s to be exposed in the calibration chamber are selected from the stock of detectors by the QA Officer. Specifications for spiking the exposure of the detectors is made by the QA Officer, in written form, directly to the calibration facility. Note: If relevant the radon chamber average gamma background, relative humidity or temperature is requested. Results of the analysis of spiked samples are evaluated by the QA Officer following the procedures described in Section 10.1 and recorded on a control chart as shown in Exhibit 1 to determine any consistent bias in the results and to identify the need to contact the detector manufacturer about making corrections. Active Monitor 9.2 Crosschecks Made to Assess Bias Any consistent bias shown by the instrument(s) is evaluated by performing regular crosschecks with a recently calibrated (see Section 7.0) instrument. These crosschecks are performed during the 12-month interval between calibration, and approximately six months after calibration, so that no more than about six months elapses between either a calibration or a crosscheck. Refer to section 7.0 for calibration procedures. Crosschecks are conducted with the following procedures. a) <list the detector manufacturer's instructions regarding checking the unit for background and conducting instrument performance checks (use instructions from the detector user manual)> Section No.9 Rev No. 1 Date: 6/10/94 Page 3 of 4 b) The monitor is exposed in an environment that averages, if possible, greater than 4 pCi/L for a minimum of 48 hours directly adjacent to a recently calibrated active monitor (not necessarily the same type or model, but one that produces results in the same units and has been calibrated within the last 30 days). The environment where the monitors are compared is, if possible, chosen for its <radon or RDP> stability, and in an area where radon levels are considerably greater than each unit's lower limit of detection. Both units measure over the same time interval, although the recently calibrated monitor can have the capability to measure in shorter intervals. The initial four hours of data is not used in the calculation. The results for each monitor are averaged and analyzed in terms of percent difference (%D), as described in Section 10.1. The results of crosschecks are analyzed following the procedures described in Section 10.1.1 and recorded on a control chart as shown in Exhibit 2. The QA Officer is responsible for checking the results of these crosschecks and determining the maximum amount of disagreement (upper bounds of agreement) between devices. <The upper bound may be based upon a sSTD as calculated from the control chart or may be a standard level specified as a percent difference (%D). Refer to exhibit 2 and 4. For example, corrective action might be required if the devices do not agree to within 15 %D.> 9.3 Duplicate Measurements Made to Assess Precision The precision of the measurement method is evaluated on an ongoing basis through the use of duplicate measurements. Note: Include the passive and/or active method used below and delete what is not relevant. Passive Duplicate: Passive duplicates are defined as identical detectors placed approximately four inches apart. Every tenth measurement receives two detectors instead of one, until at least 25 pairs (50 total detectors) are placed each month. Duplicate results are logged in the duplicate file as Result A and Result B. To ensure that some radon concentrations greater than 4 pCi/L are measured by duplicates, it may be necessary to place duplicates occasionally in locations of elevated radon concentrations (e.g., staff person's crawl space). If duplicate pairs are to be exposed by the homeowner, test instructions clearly explain how this is to be done. Note: Occasional use of nonidentical detectors for comparison may help identify a bias in either measurement system under field conditions and may be very useful, but these measurements are not considered true duplicates. Section No.9 Rev No. 1 Date: 6/10/94 Page 4 of 4 Active Duplicate: Active duplicates are defined as measurements made with the same type of instrument over identical time periods with air intakes adjacent. When two monitors are available at the same location, side-by-side measurements in approximately 10 percent of the total number of measurements, or up to 25 each month is made. If only one similar active monitor is available, one out of every 10 measurements is checked with a second device (not necessarily a "recently-calibrated" active instrument but any passive or active device that is RMP listed) placed adjacent to the active device for the same time period. Organizations with only one available monitor are not able to assess precision. Precision of a single instrument can only be roughly estimated from the analysis error provided by the detector manufacturer, (the manufacturer can measure precision because they have multiple units available, although they sometimes use counting statistics which can be misleading). Precision error for organizations with only one monitor is therefore referenced from information provided by the manufacturer. The results of duplicates are analyzed following the procedures described in Section 10.2.1 and recorded on a control chart as shown in Exhibit 3. The QA Officer is responsible for checking the results of these duplicates and determining upper bounds for agreement between devices. <The upper bound may be based upon a sigma level as calculated from 20 in control duplicate pairs, (reference exhibit 3), or may be a standard level specified as a relative percent difference.> Note: Section 9.4 is for passive methods: 9.4 Exposure of Field Blanks for Passive Detectors The EPA recommends that field blanks for passive detectors number approximately five percent of the detectors deployed. At least <5> per <100 measurements> are set aside as blanks and not exposed. The location and length of storage of blanks is determined by the QA Officer. Blanks are, if possible, shipped and handled as other exposed detectors. The results of these blanks are assessed by the QA Officer. If any blank result is significantly above the LLD, the QA Officer assesses the cause for the elevated reading and/or contacts the analysis laboratory. If appropriate, as determined by the QA Officer in consultation with the analysis laboratory, the results of the blanks are subtracted from the results of the exposed detectors. Note: This QA Plan is not designed to encompass the range of procedures (including laboratory blanks) faced by an analysis laboratory for ATD or charcoal devices; routine assessment of several types of laboratory background is important for such laboratories but is too detailed and laboratory-specific to be covered here. 10.0 QUALITY ASSURANCE OBJECTIVES AND PROCEDURES FOR ASSESSING PRECISION AND BIAS The objectives for precision and bias are consistent with U.S. Environmental Protection Agency precedents (Device Protocols [EPA 1992]), and are sufficient to meet the needs of clients. 10.1 Monitoring Bias The objective for relative bias for the <detector name> is a percent difference (%D) of + or - 10% . This relative bias is calculated as the difference between the measured and the reference value divided by the reference value. Note: Different values may be set at the QA Officer's discretion. % D= Measured Value - Reference Value Reference Value x 100 Note that the definition of percent difference is similar to the definition of Individual Relative Error (IRE), as defined in the RMP Program Handbook (RMP Handbook [EPA 1991]), except that the numerator of the IRE is the absolute value of the difference while %D can have positive or negative values. To assess the results of <spiked or cross checked measurements> over time, the results are plotted on a means control chart (see Exhibit <1 or 2>). 10.1.1 Means Control Chart to Evaluate Relative Bias From the Results of <Spikes (for passive) or Crosschecks (for active)> The results of <spikes or crosschecks> are plotted on a means control chart (see Exhibits <1 and/or 2> and the associated examples in Exhibits <1A and/or 2A>). The mean line is set at zero, and the warning and control limits are determined by calculating the sample standard deviation (sSTD) of the resultant %D values from at least 20 in-control <spikes or crosschecks>. (refer to Exhibit 4 for an explanation of sSTD) The QA Officer plots the results from the <spikes or crosschecks> on the appropriate control chart as results are available, and checks the results as soon as they are plotted. 10.1.2 Corrective Action Based on the Determination of High Bias The following guidelines, taken from Taylor (Taylor 1987) are used to determine whether the measurement system exhibits a bias high enough to warrant corrective action. Section No.10 Rev No. 1 Date: 6/10/94 Page 2 of 2 As the data is plotted, indicators that the measurement system may be "out-ofcontrol" include: o Two successive points outside the warning limits (2 X sSTD) o Four successive points outside the one-sigma limits (1 X sSTD) o A systematic trend high or low A systematic trend includes a series of points in the same direction or successive points all on the same side of the mean, even if all are within the control limits. If the data exhibit any of these indicators, the measurement system should be checked and additional QC measurements made as per the QA officer's recommendations. 10.2 Monitoring Precision The objective for precision of the <detector name> method is an average relative percent difference (RPD) between duplicates of 14 percent or less as calculated from duplicates with an average exposure of 4 pCi/L or greater. Relative percent difference is defined as the absolute difference between duplicates divided by their mean times 100, as shown below. RPD = Larger Result - Smaller Result X 100 Average of Both Results The precision objective for RPD values can be set at the discretion of the QA Officer using the methods described in Exhibit 3. 10.2.1 The Analysis of Duplicate Measurements To assess the range of differences between duplicates over time, the results of duplicates are plotted on a control chart (see Exhibit 3). The control lines for the chart in exhibit 3 are taken directly from EPA guidance (Homes Protocols [EPA 1993], Appendix B, page B6) for measurement systems with an "in control" precision of about 14% RPD. The determination of whether the measurement system is "in control" is made following the guidelines described in Section 10.1.2, as taken from Taylor (Taylor 1987). The results from duplicates are plotted on the duplicate control chart as they are available, and <name and/or title> checks the results as soon as they are plotted. 11.0 QUALITY ASSURANCE AUDITS AND REPORTS TO MANAGEMENT The QA Officer conducts an audit of QA operations at least once every <six months>. Specific focus of the audit is on: a) recordkeeping, including results of routine instrument checks b) an adequate number and type of QC measurements c) maintenance and revisions of SOP d) adequate training and retraining of staff The QA Officer submits a written audit report to <management or company name> within 30 days following the audit. The audit report includes an assessment of bias and precision, and includes recommendations as appropriate. Note: Other information on how management receives periodic status reports on QA/QC (via memos, or orally during staff meetings) should be described here. 12.0 CORRECTIVE ACTION Failure of quality control measurements to be within the defined limits of this QA plan results in immediate action to identify, correct, and document the problem. The QA officer is responsible for ensuring timely solutions to identified problems. 13.0 QUALITY ASSURANCE TRAINING The QA Officer has responsibility for reviewing the training plans for new staff and the plans for retraining when procedures change. Adequate training is given high priority, since the implementation of this QA plan is dependent upon the staff's understanding of its requirements. 13.1 Personnel Training All <company name> personnel are responsible for knowing everything in this QA Plan which falls within their particular area of responsibility as defined under Section 3.1 of this plan and/or their particular company Job Description. This QA Plan is the principle source document for the QA procedures and protocols which must be known and practiced by responsible company personnel. The QA Officer provides each employee with a copy of this QA Plan in which the specific QA activities and responsibilities for that particular employee are clearly marked and indexed. At the end of the first month of employment and at least annually thereafter, the QA Officer checks each involved employee's knowledge and understanding of their QA duties and responsibilities as defined in this Plan. If, in the judgement of the QA Officer, an employee does not adequately understand his/her responsibilities, follow-up instructions and checks are carried out until adequate understanding is demonstrated. The QA Officer notifies the employee's supervisor of each check result and these results are given prominent consideration in compensation and job advancement reviews. 14.0 REFERENCES EPA 1980 U.S. Environmental Protection Agency, 1980, "Interim Guidelines and Specifications for Preparing Quality Assurance Project Plans," QAMS-005/80, Office of Monitoring Systems and Quality Assurance, Office of Radiation Programs, Washington, D.C. 20460. EPA 1991 U.S. Environmental Protection Agency, 1991, "Radon Measurement Proficiency Program Handbook," EPA 520/1-91-006, Office of Radiation and Indoor Air, Washington, DC 20460. EPA 1992 U.S. Environmental Protection Agency, 1992, "Indoor Radon and Radon Decay Product Measurement Device Protocols," EPA 402R-92-004, Office of Radiation and Indoor Air, Washington, DC 20460. EPA 1992a U.S. Environmental Protection Agency, 1992, "EPA Requirements for Quality Assurance Project Plans," EPA QA/R-5, Quality Assurance Management Staff, Washington, DC 20460. EPA 1993 U.S. Environmental Protection Agency, 1993, "Protocols for Radon and Radon Decay Product Measurements in Homes," EPA 402-R92-003, Office of Radiation and Indoor Air, Washington, DC 20460. EPA 1993a U.S. Environmental Protection Agency, 1993, "Home Buyer's and Seller's Guide to Radon," EPA 402-R-93-003, Office of Radiation and Indoor Air, Washington, DC 20460. EPA 1993b U.S. Environmental Protection Agency, 1993, "Radon Measurement in Schools -- Revised Edition," EPA 402-R-92-014, Office of Radiation and Indoor Air, Washington, DC 20460. EPA 1993 draft U.S. Environmental Protection Agency, June 1993 draft, "Radon Measurement Proficiency Program: Guidance on Quality Assurance," Office of Radiation and Indoor Air, Washington, DC 20460. NRC 1986 U.S. Nuclear Regulatory Commission, 1986, "Accuracy and Detection Limits for Bio Assay Measurements in Radiation Protection", NUREG-1156, Office of Nuclear Regulatory Research, Washington, DC Taylor 1987 Taylor, J.K. Quality Assurance of Chemical Measurements, Lewis Publishers; 1987. Exhibit 1 Control Chart of Bias of Spiked <passive detector Name> Measurements % D= Measured value - Reference Value Reference Value x 100 Measured Value is the spiked detector Reference Value is the chamber or true value %D 30% 20% 0% -20% -30% ¦ ¦ ¦ control limit ¦-------------------------------------------------------¦ ¦ warning level ¦-------------------------------------------------------¦ ¦ ¦ ¦ ¦ ¦-------------------------------------------------------¦ ¦ ¦ ¦ ¦ warning level ¦-------------------------------------------------------¦ ¦ control limit ¦-------------------------------------------------------¦ ¦ +-------------------------------------------------------Run number or spike date Note: Control limits are 0 + - (3 X sSTD), and the warning levels are 0 + - (2 X sSTD). The value of sSTD may first be assumed to be 10%, and then after gathering at least 20 spiked results (if possible, within the same range of radon concentrations) the sSTD of the %D values can be calculated. If it is significantly different from 10%, the above limits can be changed by the QA Officer. See Exhibit 4 for description of how to calculate sSTD. Note: The value for sSTD may be different at different radon levels. If possible, assess %D at different radon concentrations. If appropriate, keep control charts for ranges of radon levels (e.g., 4-20 pCi/L, another chart for 20-100 pCi/L, etc.). The need for separate charts is determined by the QA Officer. Section No.15 Rev No. 1 Date: 6/10/94 Page 2 of 6 Exhibit 1A Example Example Control Chart of Bias of Spiked <passive detector Name> Measurements % D= Measured Value - Reference Value Reference Value x 100 Measured Value is the spiked detector Reference Value is the chamber or true value %D 30% 20% 0% -20% -30% ¦ ¦ ¦ control limit ¦-------------------------------------------------------¦ ¦ warning level ¦-------------------------------------------*-----------¦ * ¦ ¦ * ¦ ¦ ¦-------------------------------------------------------¦ ¦ ¦ * * ¦ ¦ * warning level ¦-------------------------------------------------------¦ ¦ control limit ¦-------------------------------------------------------¦ ¦ +-------------------------------------------------------12/93 12/93 12/93 4/94 4/94 4/94 ... Data: Measured Value: Date: Reference Value: %D: 4.4 pCi/L 5.9 4.1 8.9 11.0 5.0 pCi/L 5.0 5.0 10.1 10.1 -12% +18% -18% -12% + 9% Spike 12/93 12/93 12/93 4/94 4/94 Section No.15 Rev No. 1 Date: 6/10/94 Page 3 of 6 12.2 10.1 Note: The sSTD of these six %D results is equal to about 17%. +21% 4/94 Section No.15 Rev No. 1 Date: 6/10/94 Page 4 of 6 Exhibit 2 Control Chart to determine <active detector Name and serial #> bias from crosschecked measurements % D= Measured Value - Reference Value Reference Value x 100 Measured Value is the cross-checked detector measurements Reference Value is the recently calibrated detector measurements %D 30% 20% 0% -20% -30% ¦ ¦ ¦ control limit ¦-------------------------------------------------------¦ ¦ warning level ¦-------------------------------------------------------¦ ¦ ¦ ¦ ¦ ¦-------------------------------------------------------¦ ¦ ¦ ¦ ¦ warning level ¦-------------------------------------------------------¦ ¦ control limit ¦-------------------------------------------------------¦ ¦ +-------------------------------------------------------Run number or cross check date Note: Control limits are 0 + - (3 X sSTD), and the warning levels are 0 + - (2 X sSTD). The value of sSTD may first be assumed to be 10%, and then after gathering at least 20 in-control crosschecked results (if possible, within the same range of radon concentrations) the sSTD of the %D values can be calculated. If it is significantly different from 10%, the above limits can be changed. See Exhibit 4 to describe how to calculate the sSTD. Note: The value for sSTD may be different at different radon levels. If possible, assess %D values at different radon concentrations. If appropriate, keep control charts for ranges of radon Section No.15 Rev No. 1 Date: 6/10/94 Page 5 of 6 levels (e.g., one chart for 4-20 pCi/L, another chart for 20-100 pCi/L, etc.). The need for separate charts is determined by the QA Officer. Section No.15 Rev No. 1 Date: 6/10/94 Page 6 of 6 Exhibit 2A Example Example Control Chart for a single active monitor to determine bias from crosschecked measurements % D= Measured Value - Reference Value Reference Value x 100 Measured Value is the cross-checked detector measurements Reference Value is the recently calibrated detector measurements %D 30% 20% 0% -20% -30% -- ¦ ¦ ¦ control limit ¦--------------------------------------------------------¦ ¦ warning level ¦----------------------------------------*------------¦ * ¦ ¦ * ¦ ¦ ------------------------------------------------------¦ ¦ ¦ * * ¦ ¦ * warning level ¦--------------------------------------------------------¦ ¦ ¦ control limit ¦------------------------------------------------------¦ ¦ +--------------------------------------------------------8/93 10/93 12/93 2/94 4/94 6/94 ... Data: Measured Value: Date: 6.3 pCi/L 5.9 6.2 Reference Value: 6.0 pCi/L 5.0 7.5 %D: -12% 18% -18% Cross Check 8/93 10/93 12/93 Section No.15 Rev No. 1 Date: 6/10/94 Page 7 of 6 8.9 8.8 19.6 10.1 8.1 16.2 -12% 9% 21% 2/94 4/94 6/94 Note: The sSTD of these six %D results from an individual active monitor is equal to about 17%. Section No.15 Rev No. 1 Date: 6/10/94 Page 8 of 6 Exhibit 3 Duplicate Control Chart for determination of <detector Name> precision This chart only uses duplicates from similar detectors where the average of the two is > 4 pCi/L RPD = RPD 36% 28% 14% Larger Result - Smaller Result X 100 Average of Both Results ¦ ¦ ¦ ¦ ¦ control limit Note: expect to see results above ¦----------------------------------------------------¦ this level only 1% of the time ¦ ¦ warning level ¦----------------------------------------------------¦ ¦ ¦ ¦ ¦ in-control level Note: results are routinely ¦----------------------------------------------------¦ around this level of precision ¦ ¦ ¦ ¦ ¦ +------------------------------------------------------ Note: Control limits are 0 + - (2.6 X sSTD), and the warning levels are 0 + - (2 X sSTD). The value of sSTD may first be assumed to be 14%, and then after gathering at least 20 in-control duplicate results (if possible, within the same range of radon concentrations) the sSTD of the RPD values can be calculated. If it is significantly different from 14%, the above limits can be changed. See Exhibit 4 for description of how to calculate the sSTD. Note: This chart's control limit is approximately 2.6 times the sSTD (EPA 1993), and represents a level which is expected to be exceeded by chance about 1% of the time. (If a 3 X sSTD level or 42% RPD were used it would be expected to be exceeded by chance less than 0.5% of the time). The sSTD value of 14% can be used until a sufficient number (at least 20 pairs) of duplicates that are in-control and measure greater than 4 pCi/L have been made. If the sSTD of the RPD of Section No.15 Rev No. 1 Date: 6/10/94 Page 9 of 6 these pairs is significantly different than 14%, the draft EPA guidance on QA for RMP participants (EPA 1993 draft) or the EPA Homes Protocols (EPA 1993) guidance on how to set the proper control limits is consulted to determine how to set different limits. Section No.15 Rev No. 1 Date: 6/10/94 Page 10 of 6 Exhibit 4 Calculation of Sample Standard Deviation The sample standard deviation (sSTD) can be obtained using a spreadsheet computer program or a scientific calculator. For sSTD calculations in this QA plan the following formula can be used. n ( x - x i sSTD = 2 avg ) i=1 n -1 = square root of the value n = number of values n = the sum of whatever follows i=1 x i = each value x avg = avg of the values ( x i - x avg )2 = square of each value difference from the avg Note: In this QA plan only sSTD (sample or population standard deviation) is used. The difference between standard deviation (STD or sigma) and sSTD is that sSTD is obtained by multiplying STD by ( n/n-1 ). Both sSTD and STD are usually obtained by using a scientific hand calculator or computer spreadsheet program. To obtain the correct control chart levels, the sSTD of at least 20 %D or RPD values, which were judged to be in-control, are computed. The control chart warning level is set at 2 times the sSTD. The out of control level is set at 3 times the sSTD for cross checks and spikes and 2.6 times the sSTD for duplicates. Duplicates have a more stringent out of control level because they are only evaluating precision and not system bias. APPENDIX A NON-INTERFERENCE RADON SURVEY AGREEMENT REQUIRED TEST CONDITIONS Radon and radon decay product concentrations in a dwelling fluctuate from hour to hour, day to day and season to season. The following test recommendations were developed by the EPA to provide standardized conditions under which a short-term radon survey is to be performed in order to reduce the variation in radon levels in a dwelling. These conditions tend to maximize the radon measurement in order to determine if a dwelling has the "potential" to have an elevated radon level. If the result is elevated, the EPA recommends further testing to determine better the yearly average concentration. If the test conditions below are not adhered to, the test results may be deemed invalid. The following conditions must be read, understood and followed: All exterior windows must be kept closed. All exterior and the basement to first floor doors must be kept closed except for normal, momentary entering and exiting. The radon detector cannot be moved, covered or altered in any way. Heating, air conditioning, dryers, range hoods, bathroom fans and attic ventilators can be operated normally. If any heating, air conditioning or ventilating equipment has a built in outdoor air supply that is manually controlled, it is turned off or the inlet closed. Fireplaces or wood stoves are not operated, unless they are a primary heat source. Whole house fans are not operated. Window fans are removed or sealed shut. These test conditions are maintained for 12 hours prior to the start of the radon detector being exposed, unless the test is longer than four days in duration. If there are any questions, or test conditions are not met, please contact the test company at (Co. Ph #) I/We the responsible individual or building custodian understand and will inform all occupants of this dwelling of the above conditions of the test. I/we agree to maintain these conditions during the test period. Property Address: Detector Type & Locations Test Technician: 1st: Responsible Individual: 2nd: Real Estate Agent: 3rd: Date Signed: Installed Date/Time: Comments: Retrieval Date/Time: APPENDIX B RADON SURVEY IN PROGRESS DO NOT REMOVE THIS NOTIFICATION The following conditions must be maintained: Do not open any windows. Do not open any external or basement to first floor doors except for normal momentary entering and exiting. Do not touch, cover, move or alter the performance of any radon detectors or non-interference controls. Do not operate any whole house fan(s). Do not use any fireplace(s) or wood stove(s), unless they are the primary heat source. Operate heating and air conditioning normally. Turn off and keep off any equipment that supplies fresh air to the dwelling unless it is make up air to a combustion appliance. NOTE: The dryer, range hood, bathroom fan or attic ventilating fan can be operated. This equipment is operated normally because any exhaust fan or any combustion appliance may increase the negative pressure in the dwelling, which can raise or lower the radon concentration. Windows must be kept closed because they can create negative pressure in the lower portions of the dwelling due to the warm air escaping or the direction of the wind, which can raise or lower the radon levels. Please contact (Testing Company Name & Phone Number) if there are any questions or concerns. Test Period From: Responsible Individual: Date: To: APPENDIX C EXAMPLE RADON TEST REPORT Test Company Street Address City, State Zip Phone & Fax Number Radon levels tend to vary. We accept no liability for any or all loss due to the results of this or these tests Customer Name Street Address City, State Zip Date: Test id#: Test Type: Cus H Ph: Cus W Ph: Test Location: <Street, City, State Zip> Note: Customer signed agreement to maintain closed-house conditions Detector Placed by: <Technician Name & RMP ID #> Comments: None ---------------------------------------------------------------------------------------------------------------------------------------The following measurements were made with: <Detector Type> Detr Id # ------5475 Bas 5085 Bas Test Start Start Rm Date ----------------02/21/94 09:40 02/21/94 09:40 Stop Stop Time Date -------- --------02/23/94 10:30 02/23/94 10:30 | Radon | | pCi/L --------| 5.6 | | 5.2 | ------Bas Average = 5.4 Time -------- +/| pCi/L -------0.3 0.3 pCi/L 6027 1Fl 02/21/94 09:45 02/23/94 10:25 | 1.8 | 0.2 ------------------------------------------------------------------------------------------------------------------------------------------THE AVERAGE RESULTS WERE ABOVE EPA GUIDELINES The EPA recommends that any lived-in areas of a home that measure greater than 4 pCi/L (or any areas that are suitable for occupancy without renovation, if the home is involved in a Real Estate Transaction) be mitigated to below 4 pCi/L. If the measurement is below 4 pCi/L then additional mitigation to reduce the levels is not necessary. It is the client's obligation to inform the dwelling occupants of the results of the measurement. Note: Homeowner is responsible for additional follow-up testing. Retesting should take place with each change of ownership, structural alterations to the dwelling that changes the ventilation pattern, major cracks occurring in the foundation or nearby blasting or earthquakes, and after every two years. There is uncertainty with any measurement result due to: statistical variations, daily and seasonal radon variations due to changes in the weather and operation of the dwelling as well as possible interference with the necessary test conditions that may or may not influence the results.