Chemistry 472 Fall 2000 UV-Vis Spectrophotometry Introduction This experiment is an introduction to spectrophotometry and consists of four parts. In the first part, the components of two, disassembled spectrophotometers are examined. One is a single beam instrument, a Bausch and Lomb Spectronic 20; the other is a Beckman DB: a double-beam-in-time spectrophotometer. Next, the spectral response curve of a Spec 20 is determined and the Spec 20 is used to collect the absorption spectrum of chromium(III) nitrate. For comparison, this spectrum is also collected with a double-beam-in-time UV-Vis spectrophotometer. In the third part, a calibration curve is made and the concentration of Cr(III) in an unknown is determined using Beer’s Law. Finally, the effect of diffraction grating blaze density and slit width on the resolution of a spectrum is explored using Oriel optical components and a spectral calibration lamp. Background information for this experiment is contained in Chapters 7, 13, and 14 in Principles of Instrumental Analysis 5th Edition by Skoog, Holler, and Neiman. NOTE: A separate station has been set up for each part of the experiment. Parts 1, 2, and 4 may be done in any order. Part 2 must completed before starting Part 3. Part 1. Instrumentation In molecular absorption spectroscopy, a solution is irradiated with light. If the light is of the appropriate wavelength, it can be absorbed by analyte molecules causing them to undergo a transition to a higher energy state. The amount of radiation absorbed by the sample is determined by measuring the intensity of the beam before (Po) and after (P) it passes through the sample. The transmittance T of the solution is defined as T = P/Po. In spectrophotometry, transmittance is actually a measurement of the ratio of a sample beam intensity to a reference beam intensity. The absorbance A is related to the transmittance by A = -log T. At a minimum, the instrumentation for absorption spectrometry consists of a radiation source (a tungsten filament lamp for the visible or a deuterium lamp for the ultraviolet), a monochromator (a prism or grating system to isolate one wavelength region), a sample cell, a photodetector, and a readout device, as shown in Figure 13-12(a) in Skoog, Holler & Neiman. The monochromator consists of an entrance slit, some optics, a prism or grating, and an exit slit. With this single-beam configuration, the percent transmittance of a blank (reference sample) is adjusted to 100%T. The reference cell in then manually replaced with the sample cell and the %T is indicated on the readout device. Anytime the wavelength is changed, this procedure must be repeated. Most modern spectrophotometers are double-beam instruments as shown in Figure 13-12(b & c). These instruments are configured to allow the automatic and simultaneous (or near simultaneous) measurement of the sample and reference beam intensities. This can be accomplished using a beam splitter to divide the beam into the reference beam and sample beam. In a double-beam-in-time spectrophotometer (Figure 13-12(c)), the beams pass through a modulator which allows the detector to see either the reference beam or the sample beam. Since the source beam is directed through a reference cell part of the time and through the sample cell the rest of the time, the term "double-beam-in-time" is used. Examination of the Spec 20 A diagram of the Spec 20 instrument used in this experiment is shown in Figure 13-17(b) in Skoog, Holler, and Neiman. For this part of the experiment, the Spec 20 has been disassembled. Make sure that it is unplugged, lay it on its back, and open the trap door on the bottom. Opening the door exposes the source, filter mount, phototube, and bottom of the sample chamber. Observe the lamp but do not remove it. It is a common 6-volt tungsten filament lamp very similar in design to an ordinary light bulb. 1 Examine a phototube, noting the anode and cathode. The standard Spec 20 phototube is a type S-4 cesium-antimony tube, which is most sensitive to light in the blue region. When working with the Spec 20 in the red region, above 625 nanometers, a red sensitive phototube is used instead. Examine the curved arm of metal at the bottom of the sample chamber. This is the occluder, which, mounted on a pivot, opens and closes the shutter gate which allows light into the sample. Move the occluder by hand and note its action. It is automatically pushed aside by the cuvette when a reading is taken. When your investigation of this area is complete, close and secure the trap door, and set the instrument upright again. The instrument is divided into two sections. The rear contains the electrical components. These include the power supply, the amplifier which strengthens signals coming from the phototube, and the readout device which in this case is a voltmeter calibrated in both absorbance and percent transmittance units. Continuous light over the entire visible spectrum emanates from the tungsten lamp. This light is collected by the field lens and collimated (made parallel) by objective lens. Identify these components and trace out the light path. The white light next falls upon the diffraction grating, which disperses it horizontally into the familiar sequence of spectral colors. To perform this task, the grating is ruled at 600 lines per millimeter. The dispersed light next passes through the light control, the occluder and the exit slit. Only when a cuvette is inserted into the sample chamber is the occluder moved out of the way. The width of the exit slit and the spread or dispersion of the light from the diffraction grating together determine the spectral bandwidth, which in a Spec 20 is 20 nm. Thus, for example, if the instrument is set at 600 nm, light having wavelengths in the range 590 to 610 nm will be passed to the sample and phototube. Only that portion of the light dispersed by the grating which falls on the exit slit is passed to the sample. Selection of the desired portion of the spectrum to be passed is accomplished by adjusting the angle of incidence between the source rays and the diffraction grating. The control on top of the instrument performs this task by physically turning the grating, and is known as the wavelength control knob. Attached to this control is a dial calibrated in nanometers which indicates the setting. In quantitative spectrophotometry one is interested not in the absolute intensity of light passing through a sample, but rather in the relative intensity of such light with respect to the intensity of light passing through a reference or blank solution. It is thus necessary to set the 0 and 100% transmittance limits between which the transmittance of samples will be measured. The 0% T is set with no light reaching the phototube, that is, with no cuvette in the sample chamber and all light blocked by the occluder. This setting is performed with the amplifier control knob on the left front of the instrument. This control adjusts the gain, or sensitivity of the amplifier, and the offset. This determines how much needle deflection is caused by a specific intensity of light striking the phototube. The 100% T is set with the right hand or light control knob, usually with a cuvette of water or another blank solution in the sample chamber. Turning the light control knob causes a V-shaped slit behind the knob to move into or out of the light beam. It is thus a purely mechanical control which simply physically blocks out more or less of the diffracted light. With due respect for the exposed electrical wiring, plug the instrument in and turn it on. Use a sheet of paper or an index card to observe the light spectrum and the effect of the wavelength and light control knobs upon the location and intensity of the dispersed light. Following this examination, turn the instrument off and unplug it. Examination of the Beckman DB The Hitachi U-4001, a "double-beam-in-time" spectrophotometer, will be used in part 2 of this experiment. You will be able to examine the structural details of an older design instrument, the Beckman DB. With the aid of the diagram next to the instrument, trace the light path. Notice the way the instrument switches between the tungsten and deuterium sources. Also, note that instead of a 2 diffraction grating, a mirrored prism is used to disperse the light in this instrument. Finally, observe the vibrating mirrors beneath the sample holder compartment. As these mirrors move back and forth together, the beam alternately passes through the sample and the reference compartments. This is another way of constructing a double-beam-in-time instrument. Part 2. Spectral Response Curve and Chromium Nitrate Spectrum Operation of the Spec 20 In a few brief procedures, the spectral characteristics of the Spec 20 will be examined, and the visible spectrum of a simple substance will be recorded. The handling of cuvettes is very important. Any variation in the cuvettes, such as stains, scratches, or changes in the curvature of the glass, will cause varying results. Thus it is essential to follow several rules: • Do not handle the lower portion of the cuvette through which the light passes. • Always rinse a cuvette with several portions of the solution to be measured before taking the measurement. • Wipe off any liquid drops or smudges with a clean tissue before placing the cuvette in the Spec 20. • When inserting a cuvette, always do so with the index mark facing the front of the instrument, and after the cuvette is seated, line up the index marks exactly. Determination of the Response Curve NOTE: The Spec 20’s take several hours to warm up. They should be on when you come to class – please do NOT turn them off. 1. Using the Zero Control knob (left front) adjust the 0% T with the empty sample chamber. Be sure the chamber is firmly in place and that its cover is shut during this and all future readings to prevent room light from leaking into the phototube. 2. Turn the Light Control knob (right front) fully counter-clockwise, and insert a cuvette filled about one-half full with distilled water. Adjust the wavelength control to 410 nm and rotate the Light Control knob clockwise until the meter reads about 90% T. 3. Rotate the cuvette in the sample chamber and note how readings can change 1% T or so if the cuvette is not consistently positioned. Change the wavelength setting back and forth and note how the response changes with wavelength. 4. Determine the wavelength of light to which the instrument is most sensitive. Depending on the model of Spec 20 you are using, this wavelength may be around 410 nm or 595 nm. Adjust the light control to give 100% T at this wavelength. Then without readjusting either control knob further, record %T readings for the following wavelengths: 350, 375, 400, 425, 450, 475, 500, 512, 550, 575, 600, 612, and 625 nm. The resulting data give the relative response of the instrument vs. wavelength. This is known as a spectral response curve. 5. Place a dry cuvette containing a reflecting chalk into the chamber and rotate the cuvette until the beam reflects upward to view. Observe and record the color of the beam in 50 nm intervals from 350 to 650 nm. Adjust the light control (100%T) if necessary to see the beam, but do not allow the meter to read off scale. 6. Adjust the wavelength to 900 nm and turn the light control fully clockwise. Note the apparent blue-green color of "900" nm light which is actually the second order spectrum of 450 nm light. Turn the light control back down again. Explain the need to use a red filter when working in this region of the spectrum. 3 Collection of an Absorption Spectrum In UV-Vis spectroscopy, the characteristic manner in which an analyte absorbs different wavelengths of radiation depends on the energies and structure of the electronic states for that species. A plot of the absorbance vs. wavelength is called an absorption spectrum. The absorption spectrum is unique to a particular molecule and for this reason, can be used for qualitative identification of a compound. In this experiment, the absorption spectrum of a substance in the visible range, 375 to 625 nm, will be recorded using the Spec 20 and the Hitachi U-4001 spectrometer. The particular species examined is the chromium (Cr+3) cation in an aqueous solution. Spec 20 Instructions 1. Set the Spec 20 to a wavelength of 375 nm and adjust the 0% T as before. Adjust the 100% T with distilled water in the cuvette. Be sure to use the same cuvette filled with distilled water for all future 100% T readings. 2. Fill a second cuvette about one half full with 0.04 M chromium nitrate solution and take a % T reading of this solution at 375 nm. 3. Change the wavelength dial to 400 nm. Set the 0 and 100%T as done previously (no cuvette, cuvette with distilled water). Measure the %T of the 0.04 M chromium nitrate solution.. 4. Continue this procedure at wavelengths of 405, 415, 425, 440, 455, 470, 490, 500, 520, 530, 540, 550, 575, 580, 600, and 625 nm. 5. Empty solutions into the appropriate waste container. Rinse the cuvettes well with distilled water. Hitachi U-4001 Instructions 1. Set the scanning range from 400 – 625 nm (Scan Parameters menu), the slit width to 2 nm (Instrument Parameters menu), and scan a background (Correct User 1) with distilled water in both the sample and reference cuvettes. Disposable cuvettes can be used since the analysis is conducted at wavelengths greater than 350 nm. 2. Fill the sample cuvette with the 0.04 M chromium nitrate solution and collect a spectrum. 3. Print your spectrum and then convert the file to ASCII format. This ASCII file can be opened with EXCEL. Part 3. Determination of Cr(III) Concentration Quantitative Spectrophotometry In quantitative applications, the absorbance A at a given wavelength is related to the analyte concentration c by Beer's law: A = abc, where a is absorptivity of the analyte and b is the sample pathlength. This is a linear equation (y = mx + b) with a slope equal to ab and a y-intercept of 0. A calibration curve is constructed by plotting A versus c for a series of solutions of known concentration. The equation of the best fit line is then used to determine the concentration of the analyte in unknown solutions from the measured absorbance. It should be noted that Beer’s law will be valid and the calibration curve linear only if the solutions are dilute. At high concentrations the analyte molecules interact with each other causing deviations from Beer’s law. Additionally, the calibration curve is only valid within the concentration range used to construct it. It may be necessary to dilute unknowns so that they fall within this range. The wavelength chosen for quantitative measurements must be appreciably absorbed by the substance being determined, otherwise, the measurement will not be sensitive to low concentrations. On the other 4 hand if the substance has too high an absorbance the transmitted light is too weak for accurate measurements. It can be shown mathematically that for spectrophotometers with simple meter readouts such as a Spec 20, the best compromise between too little and too much absorption comes at about 36.8% T (A = 0.43). This is the region where the error in reading the transmittance is least in comparison with the value of the transmittance. Consider a plot of the analyte absorptivity a versus wavelength (absorptivity is determined from the solution absorbance by a rearrangement of Beer’s Law: A/cb = a). An appropriate wavelength for analysis would lie in a region where the slope of the curve approaches zero. Since spectrophotometers isolate a range of wavelengths, and not a single wavelength, it is important that all wavelengths within this instrumental bandwidth exhibit the same molar absorptivity for a given sample concentration. Under these conditions, the sample absorbance is proportional to concentration as if the incident radiation were truly monochromatic. See Figure 13-5 in Skoog, Holler & Neiman for an illustration of some of these points. The basic method by which absorption spectrophotometry is applied quantitatively is to prepare a calibration curve or Beer's law plot. This plot should illustrate the linear relationship between absorbance A and concentration c as given by Beer's law. Preparing a calibration curve allows the analyst to demonstrate that the Beer's law relationship is obeyed for the conditions employed (concentration range and wavelength used). Since the calibration curve makes use of several standards, it will be more accurate than an analysis based on a single standard. In this part of the experiment, a calibration curve will be prepared for a series of Cr3+ standards and used to determine the concentration of Cr3+ in a sample of unknown concentration. 1. Using the 0.200 M chromium nitrate stock solution, 100-mL volumetric flasks, and pipets provided, prepare 6 standards with concentrations from 0.0100 M to 0.0600 M. NOTE: Calculate the volume of the stock solution that will be pipetted for each standard BEFORE you come to lab. 2. Using the spectrum of the chromium nitrate solution collected on the Spec 20, determine the wavelength of maximum absorbance for Cr3+ (λmax). This is the wavelength that will be used to prepare the calibration curve. 3. As done previously, set the 0 %T and 100 %T at λmax. 4. Obtain and record triplicate absorbance readings for each of the standards at this wavelength. 5. Make three measurements of the absorbance for the unknown chromium nitrate solution. Part 4. Slit Width and Blaze Density In this part of the experiment, the effect of slit width and grating on spectral resolution will be examined. The experimental setup consists of (1) spectral calibration lamp as a light source, (2) a monochromator with exchangable diffraction gratings and variable inlet and outlet slits, (3) a PMT detector, and (4) a stripchart recorder as the readout device. The ability of a monochromator to separate different wavelengths of light depends upon its dispersion. The dispersion is detemined by the focal length F of the monochromator and the blaze (or line) density of the diffraction grating. Reciprocal linear dispersion D-1 is often used to describe the ability of the monochromator to disperse light and it can be calculated using the following equation: D −1 = d nF where d is the distance between the lines on the diffraction grating and n is the diffraction order. In essence D-1 describes how “spreadout” the wavelengths of light are at the exit slit of the monochromator 5 (dλ/dy). The dimensions are reported as nm/mm or Å/mm. The lower the value of D-1, the better the monochromator disperses the light. The dimensions of the entrance and exit slits of a monochromator play an important part in the resolving power of the monochromator. Peaks in a spectrum are said to be fully resolved if they are separated from adjacent peaks at the baseline. The effective bandwidth ∆λeff is the span of wavelengths that exit the monochromator for a given wavelength setting and it is related to the reciprocal linear dispersion by the following: ∆λeff = wD −1 where w is the slit width. Theoretically, complete resolution of two lines in a spectrum is achieved if the slit width is adjusted so that the effective bandwidth is equal to one half the wavelength difference between the two lines. Most monochromators are equipped with variable slit widths so that the effective bandwidth can be changed. Narrow slit width are used when the spectral details of narrow absorption or emission bands need to be resolved. Unfortunately, narrowing the slit results in a decrease in signal intensity. Large slit widths are used for quantitative analysis where signal intensity is an advantage and spectral detail is unimportant. Procedure In order to judge the resolving power of the monochromator, a region of the lamp spectrum that contains closely spaced lines must be examined. The source that you will use is an argon spectral calibration lamp and closely spaced lines occur in the region between 417 nm – 421 nm. Choose the smallest step size possible. For two of the gratings (1200 L/mm and either: 1800 L/mm, and 400 L/mm), determine the maximum slit width that can be used and still fully resolve 2 peaks which are about 1 nm apart. The slit widths should always be the same for both the entrance and exit slits, so change them together. Collect data with at least 3 different slit widths for each grating. Operating Instructions: Oriel Components IMPORTANT: Unlike the commercial spectrophotometers that you will use in the other experiments in this class, these components are easily damaged if you are careless about the way that you handle them. In addition, the safety features built into commercial instruments are not present. Please be aware of the following and act accordingly. Failure to follow these guidelines will result in a grade of zero for this experiment. 1. Always wear safety goggles and NEVER look directly at a light source. Many of the light sources emit UV radiation which can damage your eyes. 2. Never put your fingers on the surface of the diffraction grating, handle it by the handle and edges only. Gratings are very expensive and will be permanently damaged by scratches and fingerprints. 3. Do not make connections to a power supply unless it is OFF. This includes the lamps and the PMT, otherwise they will be destroyed. 4. Always check with a TA before powering up a source or turning on the PMT power supply. 5. Be gentle with all components: Carefully turn the micrometer when adjusting the slit width -- do not overcrank it. The monochromator has a manual crank but do not attempt to operate the manual crank while the stepper motor is in operation. 6 Grating Installation The first time that you do this, you must have TA supervision. Make sure that the shutter to the PMT is closed. Open the monochromator compartment and turn the bar above the grating so that is is parallel to the grating. Remember to be extremely careful with the gratings and never allow anything to come in contact with their surfaces!!! The grating should be loose in the mount (if not, DO NOT proceed – see your TA). Carefully lift out the grating with its handles and put its cover on securely. Install grating by carefully placing it in the grating holder while the bar is in the parallel orientation. Then rotate the bar back to the perpendicular direction (point should be directed toward the grating) to hold the grating securely. Close the cover of the monochromator. Monochromator Drive Software Boot the drive program “Mono”. You can control the monochromator from this window. The stepper interface must be turned on (the toggle is on the left back corner of the interface box; when the outer side of the switch is pressed down, the interface is on). Under the “Mono” menu, select Configure Gratings. The available gratings should be listed there. Make sure that the grating currently installed has been selected. Close this window and check the wavelength calibration. The value in the box at the top of the software window is the wavelength that the software thinks the monochromator is set at. The wavelength value displayed on the monochromator is for a 1200 L/mm grating. If you are using a different grating, the actual wavelength is: 1200 L / mm × wavelength displayed = actual wavelength groove spacing ( L / mm ) of grating in use If the actual wavelength is different than the one displayed by the software, double click on the calibration wavelength box and enter the actual wavelength. The monochromator drive has been set to step in 0.01 nm intervals with a 1200 L/mm grating. This is considered by the software to correspond to a 10:1 drive ratio. Be sure that this value and not 1:1 is showing in the lower right hand corner of the software window. Scanning range and intervals are set by entering values in the boxes below “Scan” on the left side of the screen. “Pause” can be set at 0.00 and the software will scan at its maximum rate or you can set a value (in seconds) and the software will scan more slowly. It is suggested that you set the start wavelength in the “Goto” box before doing a scan. That way you can easily activate the chart recorder when the scan begins. Signal Transduction and Readout The voltage to the PMT should not be greater than 500 V. Turn the dial on the PMT power supply fully counterclockwise before turning it on. Power it up and then turn the dial until the reading is at –400 V. Turn on the pre-amp and the chart recorder. Initially, set the pre-amp to a sensitivity of 10-6 amp/V. With the PMT shutter closed, adjust the offset so that 0.000 is displayed. Set the chart recorder to have full scale corresponding to 2 V. Once you start collecting spectra, adjustments to these setting will have to be made. Your TA will show you how to operate the strip chart recorder. You should mark start and stop wavelength for each scan on the chart paper. Adjust the full scale value so that all the peaks are on scale but as large as possible. Always write down all the parameters used for each scan on the chart paper and/or your notebook. PLEASE put the cap back on the pen at the end of the lab period. 7 At this point, you should be ready to collect data. Turn on the light source (MAKE SURE THAT YOUR GOGGLES ARE ON), set the scan parameters, slit widths, open the shutter, and go! Remember to close the shutter to the PMT when you are changing grating and when you are done with the analyses. Data Analysis Using a computer, plot the relative overall response of your Spec 20 vs. wavelength, using the data obtained in the response curve section of this experiment. On the same graph plot a second curve representing the relative response of the phototube vs. wavelength, using the data supplied to you at the end of this section. Along the top of the plot indicate the colors observed for various wavelengths. The relative emission intensity of the lamp (RLI) can be calculated from the two curves plotted as follows. For each wavelength studied, divide the value for the overall instrumental response (from your data) by that for the relative phototube response (from Table I). This gives a series of numbers, the largest of which, will be about three. To convert these numbers to a scale of 100, multiply each of them by the factor 100/3. Thus: Instrument response 100 RLI λ = × Phototube response 3 Finally, plot a third curve on the same graph, representing the relative intensity of the lamp vs. wavelength. Now note how the net effect of the combination of lamp and phototube responses leads to the overall instrument response. This change in response with wavelength is the reason why 0 and 100 % T must be adjusted each time the wavelength of the Spec 20 is changed. Double beam spectrophotometers like the Hitachi are capable of automatic compensation for this change in response with wavelength. The Relative Response of the Phototube The phototube consists of a light sensitive cathode (cesium-antimony type photoemissive cell) and an anode to collect the photoelectrons. The relative response of the phototube to a beam of monochromatic light of constant intensity is given in Table I. Thus it is seen that the phototube is much more sensitive to light of wavelength 400 nm than to light of wavelength 600 nm. This means that the phototube will require a greater intensity of 600 nm monochromatic light, in order for the same % T reading to be registered upon the colorimeter dial than for 400 nm light. Table I Relative Response of a Spec 20 Phototube Wavelength (nm) 350 375 400 425 450 475 500 Relative Response (%) 90 98 100 98 91 81 68 Wavelength (nm) 512 525 550 575 600 612 625 Relative Response (%) 61 53 37 21 10 7 5 Determine which wavelength of light the Spec 20 most responsive to, which wavelength of light is emitted most strongly by the tungsten lamp, and which wavelength of light the phototube is most responsive to. Plot the absorption spectrum of the 0.04 M chromium nitrate solution using the data collected using the Spec 20. On the same plot, plot the spectrum obtained using the Hitachi spectrometer. Locate and record the wavelengths at which absorption maxima occur in both spectra. 8 Using the spectrophotometric data obtained with the Spec 20 for the series of chromium nitrate standards, prepare a plot of absorbance versus concentration. The plot should include error bars indicating the uncertainty in your data points. Report the mean and standard deviation of the unknown concentration. For each diffraction grating you used, calculate the theoretical slit width necessary to fully resolve the two peaks in your argon spectra. Questions Comment on the similarities and differences between the spectra of chromium nitrate collected on the Spec 20 and the Hitachi instruments. What might be the origin(s) for these differences? Discuss this in terms of what you know about the different configurations that these two instruments. Discuss the results of the determination of Cr(III) concentration in your unknown. What are possible sources of error in this measurement? Discuss advantages and disadvantages of using a Spec 20 or a Hitachi U-4001 for collecting an adsorption spectrum. Also discuss advantages and disadvantages of these instruments for measuring the concentration of an unknown. (NOTE: the Hitachi, used in photometry mode, measures absorbance at a selected wavelength). Typically when a double-beam instrument is used to collect a spectrum, a background is collected first. Why is this necessary? Would a background need to be collected with the Beckman DB? Why or why not? How does the slit width and grating blaze density affect peak intensity and the resolution? For each diffraction grating, compare the theoretical slit width necessary to fully resolve two adjacent peaks in your spectrum with the slit width determined experimentally for each. Discuss any differences. 9