DSC

advertisement
2005 Training School
DSC
DSC: Terminology
Amorphous Phase - The portion of material whose molecules are randomly oriented in
space. Liquids and glassy or rubbery solids. Thermosets and some thermoplastics.
zCrystalline Phase - The portion of material whose molecules are regularly arranged
into well defined structures consisting of repeat units. Very few polymers are 100%
crystalline.
zSemi-crystalline Polymers - Polymers whose solid phases are partially amorphous
and partially crystalline. Most common thermoplastics are semi-crystalline.
zMelting - The endothermic transition upon heating from a crystalline solid to the liquid
state. This process is also called fusion. The melt is another term for the polymer liquid
phase.
zCrystallization - The exothermic transition upon cooling from liquid to crystalline solid.
Crystallization is a function of time and temperature.
zCold Crystallization - The exothermic transition upon heating from the amorphous
rubbery state to the crystalline state. This only occurs in semi-crystalline polymers that
have been quenched (very rapidly cooled from the melt) into a highly amorphous state.
zEnthalpy of Melting/Crystallization - The heat energy required for melting or released
upon crystallization. This is calculated by integrating the area of the DSC peak on a time
basis.
z
TA Instruments DSC’s
DSC 2010
DSC 2910
DSC 2920
Q系列™ DSC
Q1000 是顶级的带有多种可选配置的研究级DSC
Q100 是具有扩展功能的研究级DSC
Q10 是基本型 DSC
DSC Q10P & DSC Q1000 with Pressure Cell
DSC 光量热附件
SDT Q600
•SDT Q600 同步 DSC-TGA: 同时测量热流
和重量变化信号
DSC: 典型 DSC 转变
热流 -> 放热
氧化
或分解
熔化
玻璃化转
变
交联
结晶
温度
(固化)
Understanding DSC Signals
Heat Flow
• Relative Heat Flow: measured by all DSCs
except TA Q1000. The absolute value of the
signal is not relevant, only absolute changes
are used.
• Absolute Heat Flow: used by Q1000.
Dividing the signal by the measured heating
rate converts the heat flow signal into a heat
capacity signal
From desktop, double-click on Q
Series Explorer Icon
Q Series Explorer shows all the instruments
attached to your system
Signal Display
Method Segments
Real-Time Plot
Summary,
Procedures, &
Notes
Runs in the
Sequence
Calibration
• Q100 & Q1000
¾Tzero™ Calibration
¾Cell Constant and Temperature Calibration
• Q10
¾Baseline Calibration
¾Cell Constant and Temperature Calibration
Calibration
• Q100 & Q1000
¾Tzero™ Calibration
¾Cell Constant and Temperature Calibration
• Q10
¾Baseline Calibration
¾Cell Constant and Temperature Calibration
To begin calibration start
DSC Calibration Wizard
Select Heat flow signal
& type of cooler
Q1000 = T4P
Q100 = T4
Q10
= T1
Select which
calibration to perform
Tzero Calibration
Enter parameters for
first run (empty cell)
Start experiment
Enter weight of sapphire samples
When run is completed, capacitance &
resistance are plotted and saved
Always run Indium for Cell Constant
Enter parameters for Indium sample
Enter temperatures for Indium run
Start experiment
Data is analyzed
automatically
Calibration
• Q100 & Q1000
¾Tzero™ Calibration
¾Cell Constant and Temperature Calibration
• Q10
¾Baseline Calibration
¾Cell Constant and Temperature Calibration
To begin calibration start
DSC Calibration Wizard
Select type of
calibration to run
Enter parameters
Review summary
Enter sample
information
Finish entering
sample information
Review checklist
Baseline calibration running
Start calibration analysis
File is opened
automatically
Select limits then click on
Limits Ok button
Click on Accept to
save calibration
Once the file is analyzed and the
results are saved, a checkmark
appears next to the filename
To begin calibration start
DSC Calibration Wizard
Select type of
calibration to run
Enter parameters
Review summary
Enter sample
information
Finish entering
sample information
Review checklist
Start calibration analysis
File is opened and analyzed
automatically. Click analyze
to change limits, or accept
Tzero™ Heat Flow Equation
Heat Flow
Sensor Model
qr
qs
Cs
Cr
Tr
Ts
Besides the three
temperatures (Ts, Tr, T0);
what other values do we
need to calculate Heat
Flow?
Rr
Rs
How do we calculate these?
T0
⎛ 1
dTs
d∆ T
∆T
1 ⎞
q=−
+ ∆T0 ⎜⎜ − ⎟⎟ + (Cr − Cs )
− Cr
Rr
dτ
dτ
⎝ Rs Rr ⎠
Measuring the C’s & R’s
• Tzero™ Calibration calculates the C’s & R’s
• Calibration is a misnomer, THIS IS NOT A
•
CALIBRATION, but rather a measurement of
the Capacitance (C) and Resistance (R) of
each DSC cell
After determination of these values, they can be
used in the Four Term Heat Flow Equation
showed previously
Measuring the C’s & R’s
• Preformed using Tzero™ Calibration Wizard
1. Run Empty Cell
2. Run Sapphire on both Sample & Reference
side
A few words about the Cs and Rs
• The curves should be smooth and continuous,
•
•
•
without evidence of noise or artifacts
Capacitance values should increase with
temperature (with a decreasing slope)
Resistance values should decrease with
temperature (also with a decreasing slope)
It is not unusual for there to be a difference
between the two sides, although often they are
very close to identical
Good Tzero™ Calibration Run
Enable and Select Diagnostic Signals
Check this box!
Enable and Select Diagnostic Signals
Select 1-8 for an RCS
or all of them for an LNCS
More on Diagnostic Signals later
Advanced Tzero™ Technology
• During transitions and MDSC® experiments, the heating
•
•
•
rates of the sample pan, sample calorimeter, reference
pan and reference calorimeter may be very different.
Sample pans have thermal resistance and heat capacity
and sample and reference pans rarely have the same
mass.
Advanced Tzero includes the heat capacity of the pans
and the heating rate differences between the sample and
reference calorimeters and pans.
Peaks are taller and sharper, hence both resolution and
sensitivity are dramatically improved.
Advanced Tzero™ Model
Advanced Tzero is a further refinement of the Tzero model and takes the
measurement up to the sample pan,
one step closer to the actual sample
q sam
Advanced Tzero
model includes
the Pans
Q1000
m pr c pan
m ps c pan
T pr
T ps
Rp
qs
Q100
Tzero models
the Calorimeters
Cpan
Rp
Rs
Rp
Ts
qr
Tr
Cr
Cs
Rs
Rr
T0
Advanced Tzero™
• The pan type MUST match the pan that you are
using
• Be sure and enter pan weights
Coolers
• Refrigerated Cooling Accessory (RCS)
• Liquid Nitrogen Cooling Accessory (LNCS)
• Finned Air Cooling Accessory (FACS)
• Quench Cooling Accessory (QCA)
Coolers
• Cooling device sits on
flange
• Nickel cooling rods
transfer cooling to
furnace/cell
Used to select
type of cooler, and
to specify status
between runs
RCS
• Temperature Range
¾-90°C to 550°C
RCS
• Temperature Range
•
•
¾-90°C to 550°C
Cell purge and base purge required when RCS
is running
Secondary purge is optional, but recommended
when using in highly humid environments
¾Dry N2 to cooling gas port on back of DSC
RCS
• Temperature Range
•
•
•
¾-90°C to 550°C
Cell purge and base purge required when RCS
is running
Secondary purge is optional, but recommended
when using in highly humid environments
¾Dry N2 to cooling gas port on back of DSC
Ensure that cell remains above ambient using
unload range and standby temperature
Unload range is the
temperature that the
cell will go to at the
end of a run
If unload temperature range
is not active, cell can be
controlled by using standby
temperature
LNCS
• Temperature Range
¾-180°C to 550°C
LNCS
• Temperature Range
•
•
•
¾-180°C to 550°C
Cell purge and base purge required when
LNCS is running
Secondary purge is optional, but recommended
when using in highly humid environments
¾Dry N2 to cooling gas port on back of DSC
Ensure that cell remains above ambient using
unload range and standby temperature
FACS
• Temperature Range
¾Ambient to 725°C
FACS
• Temperature Range
¾Ambient to 725°C
• Dry air to cooling gas port
• Quench cooler available to cool down quicker
between runs (not for sub-ambient use)
QCA
• Temperature Range
•
¾-180°C to 550°C
Cell purge and base purge required when using
a QCA
Topic
•
•
•
•
Keeping your DSC cell clean
Calibration
Sample Preparation
Thermal Method
Keeping the DSC Cell Clean
• One of the first steps to ensuring good data is
•
to keep the DSC cell clean
How do DSC cells get dirty?
¾Decomposing samples during DSC runs
¾Samples spilling out of the pan
¾Transfer from bottom of pan to sensor
How do we keep DSC cells clean?
• DO NOT DECOMPOSE SAMPLES IN
THE DSC CELL!!!
• Run TGA to determine the decomposition
•
•
•
temperature
¾Stay below that temperature!
Make sure bottom of pans stay clean
Use lids
Use hermetic pans if necessary
TGA Gives Decomposition Temperature
Cleaning Cell
• If the cell gets dirty
¾Clean w/ brush
ƒ Brush gently both sensors and cell if
necessary
ƒ Be careful with the Tzero™ thermocouple
ƒ
Blow out any remaining particles
Brushing the Sample Sensor
Heat Flow Calibration (Cell Constant)
• Heat Flow Calibration of Differential Scanning
•
•
•
Calorimeters – ASTM E-968
Enthalpy Calibration
Performed using Calibration Wizard
One Run
¾Indium metal
ƒ
ƒ
ƒ
ƒ
Sample Weight 1-5mg
Pre-melt sample the first time you run it
Heating rate of 10°C/min
Dependent upon purge gas
Cell Constant
• The cell constant is calculated as the ratio of
the theoretical heat of fusion of a standard
material, to the measured heat of fusion
Cell Constant = Hf (Theoretical) / Hf (Measured)
• Cell Constant should be 0.95-1.20 in N2
Temperature Calibration
• Temperature Calibration of Differential
•
•
•
Scanning Calorimeters – ASTM E-967
Performed using Calibration Wizard
Indium Cell constant run also performs
temperature calibration
Can do up to 5 standards
¾Pure metals typically used - In, Sn, Zn, Pb
¾We’ve found that on the Q series DSC’s one
temperature calibration point is all that is
usually needed
Sample Pans
• Type of pan depends on¾Sample form
¾Volatilization
¾Temperature range
• Use lightest, flattest pan possible
• Always use reference pan of the same type as
sample pan
Standard DSC Pans (Crimped)
• Pan & lid weighs ~23mg, bottom of pan is flat
• Used for solid non-volatile samples
• Always use lid (see exceptions)
•
¾Lid improves thermal contact
¾Keeps sample from moving
Exceptions to using a lid
¾Running oxidative experiment
¾Running PCA experiment
Standard DSC Pans (Crimped)
• Crimped pans are available in:
•
¾Aluminum - up to 600°C
¾Copper
- up to 725°C (in N2)
¾Gold
- up to 725°C
Standard Pans without lids
¾Graphite
- up to 725°C (in N2)
¾Platinum
- up to 725°C
Hermetic Pans (Sealed)
• Pan & Lid weigh ~55mg, bottom of pan is not
•
•
•
as flat as std pans
Used for liquid samples and samples with
volatiles
Always use lid (same exceptions as before)
After sealing pans, should form dome
Hermetic Pans (Sealed)
• Hermetic Pans are available in:
¾Aluminum – <600°C; <3 atm (300 kPa gage)
¾Alodined Aluminum - <600°C; <3 atm (300 kPa gage)
ƒ
•
(For aqueous samples)
¾Gold – <725°C, <6 atm (600 kPa gage)
Specialized Sealed Pans
¾High Volume - 100µL; <250°C; 600 psig(4.1 MPa)
¾High Pressure - 35µL; <300°C; 1450 psig(10 MPa)
• Note: 3 atm is approximately 44 psig
It Does Matter What Pan you use
Monohydrate
Pharmaceutical
sample
Sample Shape
• Keep sample thin
• Cover as much as the bottom of pan as
possible
Sample Shape
• Cut sample to make thin, don’t crush
• If pellet, cut cross section
Sample Shape
• Cut sample to make thin, don’t crush
• If pellet, cut cross section
• If powder, spread evenly over the bottom of the
pan
Using Sample Press
• When using crimped pans, don’t over crimp
• Bottom of pan should remain flat after crimping
Crimped Pans
Hermetic Pans
Good
Not
Sealed Sealed
Bad
• When using Hermetic pans, a little more
•
pressure is needed
Hermetic pans are sealed by forming a cold
wield on the Aluminum pans
Sample Size
• Larger samples will increase sensitivity
•
but…………….
Larger samples will decrease resolution
• Goal is to have heat flow of 0.1-10mW going
through a transition
Sample Size
• Sample size depends on what you are
measuring
¾If running an extremely reactive sample (like
an explosive) run very small samples (<1mg)
¾Pure organic materials, pharmaceuticals
(1-5mg)
¾Polymers - ~10mg
¾Composites – 15-20mg
Purge Gas
• Purge gas should always be used during DSC
•
experiments
¾Provides dry,inert atmosphere
¾Ensures even heating
¾Helps sweep away any off gases that might
be released
Nitrogen
¾Most common
¾Increases Sensitivity
¾Typical flow rate of 50ml/min
Purge Gas
• Helium
•
¾Must be used with LNCS
¾High Thermo-conductivity
¾Increases Resolution
¾Upper temp limited to 350°C
¾Typical flow rate of 25ml/min
Air or Oxygen
¾Used to view oxidative effects
¾Typical flow rate of 50ml/min
Sample Temperature Range
• Rule of Thumb
¾Have 2-3 minutes of baseline before and
after transitions of interest - if possible
ƒ DO
NOT DECOMPOSE SAMPLES
IN DSC CELL
¾Temperature range can affect choice of pans
¾Just because the instrument has a
temperature range of –90°C to 550°C
(with RCS) doesn’t mean you need to heat
every sample to 550°!
Effect of Heating Rate
PMMA
10.04mg
Thermal History
• The thermal history of a sample can and will
•
affect the results
The cooling rate that the sample undergoes can
affect :
¾ Crystallinity of semi-crystalline materials
¾ Enthalpic recovery at the glass transition
• Run Heat-Cool Heat experiments to see effect
of & eliminate thermal history
¾ Heat at 10°C/min
¾ Cool at 10°C/min
¾ Heat at 10°C/min
Heat-Cool-Heat of PET
1.5
1.0
Cool
Heat Flow (W/g)
0.5
Second Heat
0.0
First Heat
-0.5
-1.0
-1.5
20
60
100
140
180
Temperature (°C)
220
260
Help
• Instrument Error messages
• Instrument Log
• On-line Help
• Getting Started Guide
To see help for error, click
on error description, then
click on error help
HELP!!
Download