furnacemar345 Handbook total outer length of fibres: 4-5mm distance crystalalumina-tube: 3mm distance crystal - brass-tube: min: 20.5mm max: 21.5mm alumina-tube: fl 0.5mm, length: 26mm brass-tube: fl 3mm, length: 8mm furnacemar345 Handbook for the high temperature furnace attachment to the mar345 imaging plate detector system Version 1.6 November 2000 Copyright © November 2000 by Michael Estermann, Roman Gubser, Monika Krichel, Katja Lemster, Hans Reifler, Stefan Scheidegger, Walter Steurer. Laboratory of Crystallography Eidgenössische Technische Hochschule ETH ETH Zentrum CH-8092 Zürich Switzerland http://www.kristall.ethz.ch/LFK Contact Michael Estermann Laboratory of Crystallography Eidgenössische Technische Hochschule ETH ETH Zentrum CH-8092 Zürich Switzerland Phone + 41 (0)1 632 64 04 Fax + 41 (0)1 632 11 33 Email michael.estermann@kristall.erdw.ethz.ch furnacemar345 Contents INTRODUCTION ............................................................... 5 Synopsis .................................................................................................... 5 Documentation .......................................................................................... 5 FURNACE ......................................................................... 7 Components .............................................................................................. 7 Furnace housing ..............................................................................................................7 Heating core .....................................................................................................................7 Transformer ..................................................................................................................... 7 Power controller ...............................................................................................................7 Temperature controller .....................................................................................................7 Thermocouple ..................................................................................................................7 Crystal mounting ..............................................................................................................7 Computer control (Silicon Graphics) ................................................................................7 Preparing the mar345 base ...................................................................... 8 Helium beam path (optional) .................................................................. 15 Attach the furnace base ..........................................................................16 Place furnace onto furnace base ............................................................18 Connect furnace to supplies .................................................................. 21 Sample change .........................................................................................21 Changing the heating rods ......................................................................24 CRYSTAL MOUNTING ................................................... 25 High Temperature Crystal Holding Devices ......................................... 25 Building the crystal holding device with alumina fibres ..................................................25 Alumina fibre crystal holder, mounting the crystal ......................................................... 26 Building the crystal holding device with two quartz glass capillaries ............................. 27 Components .............................................................................................29 Alumina tubes ............................................................................................................... 29 Alumina fibres ............................................................................................................... 29 furnacemar345 High temperature cement .............................................................................................. 29 Goniometer head .......................................................................................................... 29 Quartz glass capillaries ................................................................................................. 29 TEMPERATURE CONTROL .......................................... 30 Programmable PID controller, Eurotherm EPC900 ...............................30 Back panel, connections ............................................................................................... Front panel, fuses ......................................................................................................... Front panel keys, EPC900 Controller ............................................................................ Graphic display, EPC900 Controller .............................................................................. Setting limits, EPC900 Controller .................................................................................. Activate Setpoint rate limit SRL, EPC900 Controller ..................................................... Auto Tune (AT), EPC900 Controller .............................................................................. Initiate Auto Tune .......................................................................................................... Instrum. Conf.-Level, EPC900 Controller ...................................................................... 30 30 31 31 31 32 33 33 34 Temperature calibration.......................................................................... 34 Calibration reference thermocouple RTC ...................................................................... 34 Calibration furnace ........................................................................................................ 34 COMPUTER CONTROL ................................................. 35 Hardware at the Swiss-Norwegian Beamline at ESRF ..........................35 EPC900 Controller, Silicon Graphics OCTANE ............................................................. 35 Silicon Graphics OCTANE ............................................................................................ 35 Wiring EPC900 - Silicon Graphics OCTANE ................................................................. 35 EPC900 supervisory program .................................................................36 In situ diffraction at 900 ºC ......................................................................36 furnacemar345 INTRODUCTION Synopsis This handbook describes the installation and operation of the high-temperature furnace attachment for the mar345 imaging plate detector system (X-ray Research GmbH, Norderstedt, Germany). The helium beam path for the mar345 system is also described. Both the high-temperature attachment and the helium beam path were developed at the Laboratory of Crystallography, ETH Zürich, Switzerland. The heating core was developed in collaboration with the Professur für Nichtmetallische Werkstoffe, ETH Zürich, Switzerland. Documentation • The marresearch Imaging Plate System Owner’s Guide, X-ray Research GmbH http://www.marresearch.com • High-temperature furnace for an imaging-plate data-acquisition system. J. Schreuer, A. Baumgarte, M. A. Estermann, W. Steurer and H. Reifler. Journal of Applied Crystallography (1996) 29, 365-370. • A helium beam path for an imaging plate detector system. M. A. Estermann, S. Scheidegger, H. Reifler and W. Steurer. Journal of Applied Crystallography (1998) 30, 1165-1166. • Diffuse scattering data acquisition techniques. M. A. Estermann and W. Steurer. Phase Transitions (1988) 67, 165-196. • A high-temperature furnace for X-ray diffraction with directly machined α-Al2O3 ceramic parts. M. Estermann, H. Reifler, W. Steurer, F. Filser, P. Kocher and L. J. Gauckler. Journal of Applied Crystallography (1999) 32, 833-836. furnacemar345 • THERMOCOAX http://www.thermocoax.fr • 900 EPC Series Enhanced Programmer Controllers Handbook 900 EPC Handbook Supplement 900 Series Digital Communications Handbook Eurotherm Controls http://www.eurotherm.com • Helium flow meter, GEC Marconi 40-500, Elliot Process Instruments 1100 series WISAG Zürich • Helium reduction valve, DL230-0.1, SL Gas, 0-0.1 bar Sauerstoff Werke, Lenzburg AG WISAG Zürich furnacemar345 FURNACE Components Furnace housing Integrated water cooling Entry and exit window (Kapton) for X-ray beam In-situ observation of crystal specimen by CCD-camera Heating core The heating core of the furnace consists of a solid block of sintered aluminium oxide ceramic in which four heating rods are placed. The heating rods have an embedded 0.5mm Kanthal wire. Inside the solid ceramic block is a cavity which hosts the crystal specimen. Access to the central cavity is provided by a number of holes. This includes access for incident and diffracted X-ray beam, optical observation of the crystal specimen with a CCD-camera system, and thermocouple. Transformer Lapp-Textima, 20 volt terminal Power controller 425S Solid state relay with optional Partial Load failure, Eurotherm Controls Temperature controller 4900 EPC Series Enhanced Programmer Controllers, Eurotherm Controls Thermocouple Thermocouple in furnace: THERMOCOAX mantle thermocouple, chromel-alumel, type K, thermocouple SKI10/50, compensation cable 2AB35NN, connector MF7M Thermocouple at position of crystal specimen: THERMOCOAX mantle thermocouple, chromel-alumel, type K, thermocouple SKI05/25, compensation cable 2AB35D, connector MF7M Crystal mounting Crystal specimen is clamped between alumina fibres of 0.01 mm thickness Computer control (Silicon Graphics) Supervisory program for EPC900 controller via RS232 serial interface furnacemar345 Preparing the mar345 base Fig. 2: Attach two magnetic stands onto the base to keep the mar345 in place during installation, put a piece of paper in between stand and marbase. Fig.3 : Remove the connectors from the back of the scanner. Fig. 4: Unscrew the scanner from the scanner base plate, remove the screw completely from the scanner (Allen key, size 2.5mm). Fig. 5: Lift scanner away from the screw, do not touch the Kapton window of the scanner. furnacemar345 Fig. 6: Detector base without scanner. Fig. 7: Remove scanner base plate (Allen key, size 3.0 mm). Fig. 8: Unscrew the white cover at the back of the detector base. Fig. 9: Remove cover, cables remain connected. furnacemar345 Fig. 10: Unscrew the black cover. Fig. 11: First, move the spindle axis to the end-position in direction of the collimation block. Second, remove black cover vertically. Fig. 12: Unscrew white cover at the front of the detector base. Fig. 13: Remove cover. furnacemar345 Fig. 14: Re-insert all unused screws for safe keeping. Fig. 15: Unscrew white side cover. Fig. 16: Remove white side cover. Fig. 17: Unscrew side cover on the opposite side. furnacemar345 Fig. 18: Remove cover. Fig. 19: Unscrew white front cover. Fig. 20: Remove white front cover. Fig. 21: Move white cover at the back, back on. furnacemar345 Fig. 22: Move back scanner base plate (black). Fig. 23: Attach supporting elements (three are of the same type, one is different), leave 5 mm freedom between support element and the table. Fig. 24: Turn brass screw until it contacts with base plate, then tighten. Fig. 25: Attach supporting element at the front. furnacemar345 Fig. 26: Tighten screws. Fig. 27: First, remove screw which fixes the scanner (if not already done). Second, move scanner back onto detector base, and fix with the single screw (Allen key, size 2.5 mm). Fig. 28: Place a sheet of paper onto the detector base (in case a screw falls down), remove beam stop. Fig. 29: Unscrew spindle axis. furnacemar345 Fig. 30: Remove bolt which fixes spindle axis, remove spindle axis. Helium beam path (optional) Fig. 31: Attach clamp. Fig. 32: Attach helium beam path at the clamp, insert special collimator. furnacemar345 Fig. 33: Attach the beam stop. Fig. 34: Move special spindle axis (brass) into position, insert positioning bolt. Attach the furnace base 1 2 Fig. 35: Position the clamps before attaching the furnace base, (1) loose, (2) tightened. Fig. 36: (a) Attach the furnace base by inserting it in a slightly tilted position into the detector base, insert first clamp 1, then clamp 2. (b) Tighten the clamps which hold the furnace base on the detector base. furnacemar345 Fig. 37: Lower clamp. Fig. 38: Tighten support bolt. Fig. 39: Remove the four bolts on the translation table of the furnace base. furnacemar345 Place furnace onto furnace base Fig. 40: Move the translation table as far back as possible (service position), place furnace onto translation table (positioning bolts), tighten the four other bolts. Fig. 41: Remove flange clamp as shown in picture, remove small furnace housing from the rest of the housing and place fully on a even surface. 2 1 Fig. 42: Unscrew the heat shield, remove heat shield, lower screw 1, upper screw 2. Fig. 43: Fully open all collimation slits. furnacemar345 Fig. 44: Reconnect all the cables to the mar345 system. (a) Open beam shutter (hidden switch underneath the detector base). (b) Insert brass needle into collimation system. (c) Close collimation system gently until contact. (d) Open collimation slits slightly so that needle can be moved. (e) Move needle as close as possible (without touching) to the heating core. (f) Close collimation system gently until contact. (g) Centre the heating core with translation tables (vertical, horizontal). (h) Centre the CCD-camera observation hole on the TV-screen by translating the spindle axis (and herewith the CCD camera). Fig. 45: (a) Lock the horizontal and vertical translation tables by the black screws. (b) Remove needle by extracting it through the collimation system. furnacemar345 Fig. 46: (a) Set the zero-position on ring of the horizontal translation table. (b) Move furnace into service position. (c) Screw on heat shield gently. Fig. 47: (a) Move back small furnace case. (b) Align Kapton exit window vertically. Fig. 48: (a) Apply vacuum grease on customised spindle axis (brass). (b) Place flange over spindle axis. (c) Attach spindle axis, insert positioning bolt, screw on spindle axis (Allen key, size 3.0 mm). continue on next page furnacemar345 Fig. 48: Continued from previous page. (d) Translate spindle axis into position. (e) Place goniometer head with needle onto spindle axis. (f) Centre needle. (g) Move furnace forward into measurement position. Connect furnace to supplies • • • • • • cooling-water supply thermocouple plug with the compensation cable electrical heating cable with the terminal clamp helium gas supply vacuum tube to pump vacuum meter Sample change • • • • • • • • • • • • • • • • • • • • • • • Set temperature to room temperature (i.e. epc900> SL 25.0) Turn off thyristor on the temperature controller (for quick cooling) Close He-valve Remove clamp between spindle and furnace Release locking arm and translate furnace back to be able to access sample Remove sample and put on a new one If a capillary powder sample is used , orient capillary while checking on monitor (adjust tilt also !) Translate back furnace to correct position (check position on monitor) If necessary, put vacuum grease on spindle axis and on furnace side of flange Start vacuum pump Open slowly vacuum valve and check that flange between spindle axis and furnace is pulled in Close vacuum valve when Kapton-window has moved inwards Insert clamp between spindle and furnace Check on vacuum meter that vacuum is OK (approx. 1.5 mbar) Close vacuum valve and check that needle moves only slowly clockwise Turn off vacuum meter, close vacuum valve Open slowly He-valve until Kapton window bulges out (He-flow 400 ml/min) Close He-valve, open slowly vacuum valve and check vacuum Repeat the above three points a couple of times to wash out furnace (3-4 times) Finally close vacuum valve, turn off vacuum pump and vacuum meter Open He-valve and adjust flow to approx. 40 ml/min Collect an image at room temperature to check that everything is OK Start heating to desired temperature furnacemar345 Fig. 49: (a) Place goniometer head with crystal onto spindle axis. (b) Move furnace forward into measurement position. (c) Open vacuum valve and pump. (d) Move flange on spindle axis towards the flange of the furnace housing. (e) Check vacuum meter (leaks in vacuum). Fig. 50: Attach the special flange clamp, tighten with the brass screw. Fig. 51: (a) Screw on beam stop. (b) Setup the helium atmosphere by a sequence of pumping and filling (with helium) of the furnace (helium reduction valve 0.02 bar). furnacemar345 Fig. 52: Cooling-water connections. Fig. 53: Electrical heating wires, thermocouple connections. Fig. 54: Helium supply and flow meter. Fig. 55: Overview. furnacemar345 Changing the heating rods Follow the instructions from Fig. 40 to 42. Remove the heat shield, exchange the heating rods, replace the heat shield, and assemble the furnace housing. Fig. 56 & 57: The heating core with embedded heating elements. Loosen screws and replace the heating rods. furnacemar345 CRYSTAL MOUNTING High Temperature Crystal Holding Devices Building the crystal holding device with alumina fibres Alumina tubes of 0.5 mm diameter are cut with a diamond saw to their final length of 26 mm. Five to eight alumina fibres of 0.01 mm thickness are placed into one tube and fixed with high temperature cement based on ceramic compounds and water (Polytec 903HP). The free ends of the fibres are then glued together with cement. The next step is to heat the cement to harden it by loss of water and organic compounds. The alumina tubes with the fibres are placed into the furnace, standing in little alumina crucibles. The heat treatment (A) is as follows: Room temperature — heating rate 120ºC/h — 120ºC for 2h — heating rate 220 or 340ºC/h — 340ºC for 4h — cooling to RT. Another step (B) is a temperature treatment for better stability of the fibres to avoid - or at least minimize distortion and excessive movements once the crystal is clamped in. The alumina tubes with the fibres are heated up to 1000ºC and are held at this temperature for about an hour. It is useful to carry out this step directly after the cement treatment to save time and energy. total outer length of fibres: 4-5mm distance crystalalumina-tube: 3mm distance crystal - brass-tube: min: 20.5mm max: 21.5mm alumina-tube: Ø 0.5mm, length: 26mm brass-tube: Ø 3mm, length: 8mm Fig. 58: High temperature crystal holding device - dimensions. furnacemar345 The heat-treated tube then is cemented into the brass tube. It is necessary to straighten the tubes extremely well, otherwise there could be problems with the centring of the crystal in the high temperature furnace later. The freshly cemented tubes have to dry for about an hour and then are treated again as mentioned above (A). Fig. 59: The fibres holding a crystal Alumina fibre crystal holder, mounting the crystal To mount the crystal, it is necessary to fix a “Crystal Mounting Device” to the microscope (Nikon SMZ-2T). This mounting device consists of two goniometer head holders (Enraf-Nonius). Fig. 60: Crystal attached to a small piece of bee wax. Fig. 61: Crystal clamped between fibres, detached from the bee wax. One holder is fixed to the microscope and the other one to the desk. They are perpendicular to each other. The Huber goniometer head with the prepared crystal holding device is placed onto the first holder, a furnacemar345 goniometer head with a preparation needle is placed onto the other. The difficulty now is to insert the tip of the preparation needle between the fibres of the crystal holding device. This needs several attempts, because of the height adjustment between the two goniometer heads. Once the tip of the preparation needle holds the fibres apart, actual mounting of the crystal can begin. The operator takes a second preparation needle, fixes a very small piece of bees wax on it and picks up the crystal. Now you need a steady hand to put the crystal between the fibres. If the position of the crystal is right, the second goniometer head with the first preparation needle has to be moved back, so that the crystal will be clamped between the fibres. It is useful to check the strength of hold with several light hits against the goniometer head holder which carries the Huber goniometer head with the clamped crystal. If the crystal does not move any more, mounting has been successful. Building the crystal holding device with two quartz glass capillaries Alternatively, a single crystal may be placed onto the end of a small inner capillary, which in turn is placed into the larger outer capillary. Both capillaries are placed into the brass-tube. The measures from Fig. 59a apply. The capillary enters the heating core via a hole with a diameter of 2.5 mm, preferably one should use capillaries of 1.5 mm or less in diameter. We used the Huber, Eucentric Goniometer Head 1005, for all the applications. Do not touch the quartz glass capillary without gloves, the capillary could crystalise at high temperatures. 1 Mount the crystal onto the inner glass capillary with cement Break capillary at desired length (> 20 mm) so that it still can be handled (scratch capillary at breaking point with sandpaper to ease breaking). In order to close the capillary opening, insert cement into the capillary opening, let it dry a bit. Use cement with a lot of water, and place a drop of cement onto the dried cement on the capillary. Place the crystal onto a waxed needle. Place crystal immediately onto the cement drop. Leave it to dry. crystal 4 mm ± 2 mm cement inner capillary (Ø 0.2-0.3 mm) cement glue 24 mm ± 2 mm approx. 36 mm outer capillary filled with argon (Ø 0.5-1.0 mm) brass tube 8 mm Ø 1-2 mm Fig. 59a: Schematic drawing of the mount of a single crystal with double capillaries (prevents recristallisation at high temperatures). The measures also apply for a single capillary mount. 2 Temper the cement to remove water from cement Place capillary into a alumina crucible, and place it into a furnace. Heat at 120°C for one hour furnacemar345 3 Place outer capillary over inner capillary Break outer capillary at desired length (20 mm). If both ends are open, flame-seal one opening (use gas/ oxygen mixture because cigarette lighter flame is not hot enough). Carefully insert inner into outer capillary, crystal should not touch the end of the outer capillary (danger of recrystallisation), 2mm distance. Fix the inner to the outer capillary with glue (Cellit) for further handling. 4 Fill capillaries with argon gas Place the capillaries, open end upside, into the alumina crucible and place into the glove box pass-through. Evacuate the pass-through three times and carefully fill with argon. 5 Flame-seal second capillary opening Remove capillaries from the glove box pass-through, keep capillaries upright, and flame-seal capillaries. 6 Place capillaries into brass tube Place some glue (e.g. Araldit rapid) into the brass tube. Depending on the experiment, the glue should withstand heat. Place capillaries into the brass tube, and align them along the axis of the brass tube. Apply cement at the upper end of the brass tube to seal off brass tube and glue from the high temperatures at the experiment. Fig. 59b: Single-crystal mounted with double capillary technique. furnacemar345 Components Alumina tubes Degussit AL23 insulating tube D 0.5/0.2mm. http://www.friatec.de Email info@friatec.de Alumina fibres Goodfellow AL605725, filament diameter 0.01mm. http://www.goodfellow.com High temperature cement Cement 903 HP, Polytec. Polytec GmbH, Polytec-Platz 1-7, 76337 Waldbronn, Germany. http://www.polytec.com Email info-vi@polytec.de Goniometer head Huber, Eucentric Goniometer Head 1005, 49mmm. HUBER Diffraktionstechnik GmbH, Sommerstraße 4, D-83253 Rimsting, Germany. http://ourworld.compuserve.com/homepages/xhuber Email xhuber@compuserve.com Quartz glass capillaries Quartzglass capillaries (Mark-Röhrchen). inner capillaries 0.2 - 0.3 mm diameter, outer capillaries 0.7 - 1.0 mm diameter. Glass, Technik und Konstruktion, Wolfgang Müller, Glasinstrumentemachermeister, Germanenweg 13c, D-14621 Schönwalde bei Berlin, Germany. furnacemar345 TEMPERATURE CONTROL Programmable PID controller, Eurotherm EPC900 Back panel, connections (1) Fig. 62: (2) (3) (1) RS232 serial communications interface. (2) Connector thermocouple. (3) Exit power controller. (4) Mains 230V. Front panel, fuses (1) Fig. 63: (2) (1) Fuse 250V/500mA. (2) Fuse 600V/20A. furnacemar345 (4) Front panel keys, EPC900 Controller Fig. 64: Front panel EPC900. The instrument can be operated by means of the front panel keys or via the supervisory communications port (RS232). The functioning of the instrument may be accessed by operation of the six front panel keys. The main functions of the instrument are arranged in menu format, which are accessed by the PAGE key, parameters in the menu can be scrolled by the SCROLL key and entered by the VIEW key. selects next menu or returns to the previous menu display SCROLL key, scroll to next parameter or moves highlighted cursor from one entry to another accepts highlighted parameter decrements/increments highlighted numerical parameter Graphic display, EPC900 Controller Fig. 65: graphic display Setting limits, EPC900 Controller Go to access level 2 ( press until access levels reached) Press until SET POINTS reached Select with Proceed to SPLIMITS with and select with Enter the following values SPH = Set Point High= 1000.0 SPL = Set Point low = 0.0 SPR = Set Point Rate / min= 30.0 Move back with to access level 2 and proceed in an analog way with OUTPUTS and proceed to OPLIMITS Enter the following values HO1 = output power high limit = 70 [%] LO1 = output power low limit= 0.0 OPR = output power rate limit dP/dt = 50.0 [%/s] (OP = current output power) furnacemar345 Fig. 66: Limits Activate Setpoint rate limit SRL, EPC900 Controller This function allows the working setpoint to be ramped from the current process variable value to the defined target setpoint. The setpoint rate is in units per minute (unless changed in user configuration menu). The setpoint rate has to be activated after an Auto Tuning or turning the mains off and on again. The setpoint rate limit, SRL has to be selected from the parameter scroll list. Press the SCROLL key until SRL-OFF is displayed and press an up or down key to obtain SRL-ON. To set the setpoint rate SPR with the front panel keys, press the SCROLL key until SPR is displayed. Press the up and down keys to set the required value. The setpoint rate SPR can also be set via the supervisory program located on the Silicon Graphics Octane computer. furnacemar345 Auto Tune (AT), EPC900 Controller The autotune AT is a one-shot algorithm which can be manually initiated at any time to allow the user to retune the instrument control parameters to suit a new setpoint or experimental condition (water pressure, different heating element). Use the autotune in combination with adaptive tune ADT. Initiate Auto Tune & . Turn off controller, turn on again and press both Proceed to Auto Tune page and follow the instructions below. Fig. 67: Initiate Autotune. Adaptive tune will be suspended if manual operation is selected but is automatically reinstated when the instrument is switched back to automatic operation. Ensure that automatic mode is in operation. Activate setpoint rate limit SPR. furnacemar345 Instrum. Conf.-Level, EPC900 Controller 2 1 Fig. 68: Instrument configuration level no changes are required for normal use turn off controller, turn on again and press both & Temperature calibration The temperature is calibrated with a reference thermocouple (RTC) placed at the position of the crystal specimen. The RTC is mounted on a flange which fits onto the corresponding flange of the furnace housing. The temperature is changed in steps of 50º and the thermovoltage from the RTC is measured with a suitable instrument. The RTE, the compensation cable and the voltmeter are a calibrated system. If any of these components are replaced, the complete system has to be re-calibrated. Calibration reference thermocouple RTC 13-JUN-1997, temperature RTC versus thermovoltage temperature [ºC] = 32.201747 + thermovoltage RTC [mV]*25.117049 Calibration furnace 17-JUN-1997, ESRF, temperature furnace versus temperature display EPC900 temperature EPC900 [ºC] = - 0.47647 + temperature furnace [ºC] * 0.95201 14-MAY-1998, ESRF, temperature furnace versus temperature display EPC900 temperature EPC900 [ºC] = - 19.46447 + temperature furnace [ºC] * 1.00279 03-JUN-1998, ESRF, temperature furnace versus temperature display EPC900 temperature EPC900 [ºC] = - 20.58268 + temperature furnace [ºC] * 0.96146 12-JUN-1998, ESRF, temperature furnace versus temperature display EPC900 temperature EPC900 [ºC] = - 11.85021 + temperature furnace [ºC] * 0.936624 29-OCT-1998, ESRF, temperature furnace versus temperature display EPC900 temperature EPC900 [ºC] = - 20.2998 + temperature furnace [ºC] * 0.97259 furnacemar345 COMPUTER CONTROL Hardware at the Swiss-Norwegian Beamline at ESRF EPC900 Controller, Silicon Graphics OCTANE The 900EPC series of controllers and programmer/controllers are high stability, high accuracy, communicating units designed to cope with a wide range of process needs and environments. The 16-bit 68070 microprocessor delivers the power of a PC; a number of ASICs significantly increase the I/O capability. 900 EPC Engineers Handbook 900 Series Digital Communications Handbook Silicon Graphics OCTANE General info OCTANE Technical publications and library Serial communications ports General terminal interfaces Wiring EPC900 - Silicon Graphics OCTANE • • • • • EPC900 controller, serial communications port wired to a standard RS232 connector Silicon Graphics computer OCTANE serial communications port with DB-9 male serial port connector connect the back panel of the EPC900 with a standard RS232 cable (no null-modem cable) connect the OCTANE with a converter from DB-9 to standard RS232 use serial communications port 2 of the OCTANE Eurotherm controller to female connector at the back panel EPC900 RS232 female connector • X1 TD pin 3 • X3 RD pin 2 • X5 GND pin 7 converter cable RS232 to DB-9 RS232 male connector • pin 2 • pin 3 • pin 7 - DB-9 female connector pin 3 TD pin 2 RD pin 5 GND furnacemar345 EPC900 supervisory program A supervisory program (author M. A. Estermann) is installed on the SGI Octane (IRIX 6.4). The program utilises the Serial Communications Port 2, i.e. the DB-9 connector has to plugged into Port 2. The program is self-explanatory and has a help section. The most common commands, such as for setting the temperature, limiting setpoint rate, output power, and setpoint high are always offered. For safety reasons, Auto Tune and changes of the configuration levels can only be performed from the EPC900 console. Start the supervisory program from any unix-shell by typing epc900. The program epc900 displays a list of the most used commands. Fig. 69 & 70: Screen snap shots from a Silicon Graphics OCTANE In situ diffraction at 900 ºC The image was recorded with the mar345 imaging plate detector system and the furnacemar345 at 900ºC. Swiss-Norwegian Beamline at the European Synchrotron Radiation Facility (ESRF), Grenoble, France, Experiment 01-02-156, November 1998. furnacemar345 ETH - Laboratory of Crystallography > Introduction > Access Statistics > Annual Reports > Equipment > Links > Location > Open positions > Research > Seminars > Software > Staff > Teaching Laboratory of Crystallography Swiss Federal Institute of Technology Zürich ETH Zentrum, Sonneggstr. 5, CH-8092 Zürich, Switzerland Phone: +41-1-6323769, FAX: +41-1-6321133, Email: lfketh@kristall.erdw.ethz.ch Head: Prof. Dr. Walter Steurer Radiation safety Quick Links > CWW > DGK > ETH > IUCr > IZA > SGK > QDB http://www.kristall.ethz.ch/LFK/ (1 of 2) [3.11.2000 08:47:49 Uhr] Scientific Index (Bild der Wissenschaft 9/99): The Top Ten in Crystal Research in Austria, Germany and Switzerland. ETH - Laboratory of Crystallography © Laboratory of Crystallography (last updated: 30/Aug/99) http://www.kristall.ethz.ch/LFK/ (2 of 2) [3.11.2000 08:47:49 Uhr] Laboratory of Crystallography - Teaching Teaching The institute is responsible for teaching compulsory introductory courses in crystallography for students of earth and materials sciences (ca. 50-60 students/course), and various specialised courses at higher levels for chemistry, physics, earth and materials science students. The lectures aim to impart the crystallographic-structural way of thinking as basis for professional activities in the field of materials science. The focus is on real structure of crystals, structure-property correlations, and methods for the investigation of single-crystalline and polycrystalline materials. List of courses 06-009 06-101 Röntgenographische Pulvermethoden Kahr, G., Stahel, A. Kristallographie II Steurer, W., Estermann, M. 06-102 Kristallographisches Grundpraktikum Schobinger, P, Fehlmann, M. 06-104 Röntgenographische Einkristallmethoden Gramlich, V., Fehlmann, M. 06-110 Quasikristalle Steurer, W. 06-113 Mineralogische Modelle des Erdmantels und des Erdkerns. Kunz, M. 06-130 Kristallographisches Seminar Steurer, W. 06-131 Kristallographisches Seminar Steurer, W. 06-134 06-801 09-062 Quasikristall-Seminar Steurer, W. Kristallographie I für Physiker Grimmer, H. Mineralogie-Petrographie I Kunz, M., Schmid, R., Schreuer, J. 39-226 Kristallographie I Steurer, W., Haibach, T. 40-009 Chemische Kristallographie II Dobler, M., McCusker, L.B. 40-029 Chemische Kristallographie III Dobler, M., Gramlich, V. 40-039 Chemische Kristallographie I Dobler, M., Gramlich, V. http://www.kristall.ethz.ch/LFK/teaching/ (1 of 2) [3.11.2000 08:47:58 Uhr] Laboratory of Crystallography - Teaching © Laboratory of Crystallography (last updated: 11-Apr-99) http://www.kristall.ethz.ch/LFK/teaching/ (2 of 2) [3.11.2000 08:47:58 Uhr] Laboratory of Crystallography - Teaching: Course 06-023 Teaching: Kristallographie II (W. Steurer, M. Estermann) Ziel: Vermittlung kristallographischer Denkweisen und Untersuchungsmethoden als Grundlage praxisbezogener (erd)materialwissenschaftlicher Tätigkeit. Inhalt: Realstruktur von Kristallen, Beziehungen zwischen Struktur und Eigenschaften, Methoden zur Untersuchung von kristallinem und polykristallinem Material. Skript: Vorhanden. Besonderes: Die 3-stündige Veranstaltung, die auch von Materialwissenschaftlern gehört wird, beinhaltet Vorlesungen und Übungen. Erdwissenschaftler nehmen zusätzlich an 2-stündigen erweiterten Übungen teil, in denen an Hand von praxisbezogenen Fallbeispielen aus dem Erdmaterialbereich kristallographische Untersuchungsmethoden geübt werden. Voraussetzungen: Mineralogie-Petrographie I oder Kristallographie I. Umfang: 3 (+2) Stunden im WS. Download: ● Vorlesungsskript (PDF-Format) ● Übungsaufgaben (PDF-Format) Serie [1] ● Musterlösungen der Übungsaufgaben (PDF-Format) Serie [1] © Laboratorium für Kristallographie (last updated: 2. November 2000, M. Estermann) http://www.kristall.ethz.ch/LFK/teaching/course-06-023.html [3.11.2000 08:48:01 Uhr] Übungsaufgaben zur Kristallographie II WS 2000/2001 Serie 1, Donnerstag 2. November 2000 Seite 1/4 Lösung 1) Frenkel-Defekte, allgemein N Anzahl Gitterplätze NZ Anzahl möglicher Zwischengitterplätze EF Energie um ein Atom von einem Gitterplatz auf einen Zwischengitterplatz zu bringen (Frenkel-Defekt) nF Frenkel-Defekte Innere EnergieU = nF EF Anzahl Konfigurationenen nF identische Atome aus N Gitterplätzen herauszunehmen ❶ ❷ ❸ ❹ ❺ ❻ ❼ ❽ ❾ ❿ …. 1. Atom N Möglichkeiten 2. Atom N-1 Möglichkeiten 3. Atom N-2 Möglichkeiten nF. Atom N-nF+1 Möglichkeiten total M1 = N (N-1) (N-2) (N-3) … (N-nF+1) Möglichkeiten Betrachten wir die Atome als identisch, dann wird eine Sequenz (1, 2, 3, …, nF) nicht von der Sequenz (2, 3, …, nF, 1) zu unterscheiden sein. Beispiel: N=10, nF=4, (❶, ❷, ❸, ❹) ist nicht von (❷, ❸, ❹, ❶) unterscheidbar, letzlich betrachten wir also nur die Menge {(❶, ❷, ❸, ❹) } und nicht mehr die einzelnen Permutationen. Michael Estermann, Laboratorium für Kristallographie, ETH-Zentrum Übungsassistenten: Antonio Cervellino (NOG59.1, Tel 2 3727), Thilo Scholpp (NOG58, Tel. 2 6130) Mirsolav Kobas (NOG63, Tel. 2 3774), Stefano Ortelli (NOG63, Tel. 2 3774) Lösungen: http://www.kristall.ethz.ch/LFK/teaching/ Übungsaufgaben zur Kristallographie II WS 2000/2001 Serie 1, Donnerstag 2. November 2000 Seite 2/4 Lösung Da es insgesamt nF! = nF (nF-1) (nF-2) … 1 Permutationen gibt, reduziert sich die Anzahl der Konfigurationen um den Faktor nF! auf W1 = N ( N - 1)( N - 2)( N - nF + 1) N! = nF ! ( N - nF )!nF ! analog gilt W2 = nF = N NZ ! ( NZ - nF )!nF ! Ê E ˆ NZ expÁ - F ˜ N Ë 2 kBT ¯ S = kB log W1 + kB log W2 , Näherung von Stirling für grosse N: log N! @ N log N log W1 = log N! @ N log N - ( N - nF ) log( N - nF ) - nF log nF ( N - nF )!nF ! log W2 = log NZ ! @ N log N Z - ( N Z - nF ) log( N Z - nF ) - nF log nF ( NZ - nF )!nF ! Z freie Energie F = U - TS soll als Funktion von nF minimal sein, d.h. Ê ∂F ˆ Ê ∂U ˆ Ê ∂S ˆ Á ˜ =Á ˜ - TÁ ˜ =0 Ë ∂nF ¯ T Ë ∂nF ¯ T Ë ∂nF ¯ T È Ê N - nF ˆ Ê N Z - nF ˆ ˘ Ê ∂S ˆ Á ˜ = kB ÍlogÁ ˜ + logÁ ˜˙ Ë ∂nF ¯ Ë nF ¯ ˚ Î Ë nF ¯ Ê ∂U ˆ Á ˜ = EF , Ë ∂nF ¯ Ê ( N - nF )( N Z - nF ) ˆ Ê NN Z ˆ EF = kBT logÁ ˜ @ kBT logÁ 2 ˜ ; nF << N , N Z 2 nF Ë nF ¯ Ë ¯ Ê E ˆ nF = NN Z expÁ - F ˜ Ë 2 kBT ¯ Michael Estermann, Laboratorium für Kristallographie, ETH-Zentrum Übungsassistenten: Antonio Cervellino (NOG59.1, Tel 2 3727), Thilo Scholpp (NOG58, Tel. 2 6130) Mirsolav Kobas (NOG63, Tel. 2 3774), Stefano Ortelli (NOG63, Tel. 2 3774) Lösungen: http://www.kristall.ethz.ch/LFK/teaching/ Übungsaufgaben zur Kristallographie II WS 2000/2001 Serie 1, Donnerstag 2. November 2000 Seite 3/4 Lösung durch dividieren mit N kann dieser Audruck auf Konzentrationen umgerechnet werden nF = N Ê E ˆ NZ expÁ - F ˜ . N Ë 2 kBT ¯ Siehe die Abbildung 1.1-5 im Skript Seite 7, Einheit der x-Achse 1 / T , Einheit der yÊn ˆ Achse lnÁ F ˜ ergibt dann einen linearen Verlauf mit ËN¯ Ê NZ ˆ E Ên ˆ - F lnÁ F ˜ = lnÁ ˜ ËN¯ Ë N ¯ 2 kB T . y = a + bx Beim Abkühlen werden die Punktdefekte ab einer gewissen Temperatur eingefroren, d.h. es kann keine Diffusion mehr stattfinden, und damit auch kein equilibrieren der Defekte mehr. 2) Frenkel-Defekte, Silberchlorid AgCl 4 Ag+ Ionen pro Einheitszelle 4 Cl- Ionen pro Einheitszelle 8 mögliche Zwischengitterplätze (in den Zentren der Achtelswürfel) pro Einheitszelle Für die Berechnung der Konzentrationen können wir uns auf die Einheitszelle beschränken, da die Konzentration für eine Einheitszelle die gleiche wie für den ganzen Kristall ist (Translationssymmetrie). kb = 1.38 E - 23 Joule / K , 1eV = 1.6 E - 19 Joule nF = 8 E - 10 N Michael Estermann, Laboratorium für Kristallographie, ETH-Zentrum Übungsassistenten: Antonio Cervellino (NOG59.1, Tel 2 3727), Thilo Scholpp (NOG58, Tel. 2 6130) Mirsolav Kobas (NOG63, Tel. 2 3774), Stefano Ortelli (NOG63, Tel. 2 3774) Lösungen: http://www.kristall.ethz.ch/LFK/teaching/ Übungsaufgaben zur Kristallographie II WS 2000/2001 Serie 1, Donnerstag 2. November 2000 Seite 4/4 Lösung AgCl, F 4/m -3 2/m, a=5.549Å, mit Zwischengitterplätzen Michael Estermann, Laboratorium für Kristallographie, ETH-Zentrum Übungsassistenten: Antonio Cervellino (NOG59.1, Tel 2 3727), Thilo Scholpp (NOG58, Tel. 2 6130) Mirsolav Kobas (NOG63, Tel. 2 3774), Stefano Ortelli (NOG63, Tel. 2 3774) Lösungen: http://www.kristall.ethz.ch/LFK/teaching/ marresearch: Home Page Welcome to Our Home Page Protein and small molecule crystallographers around the globe know one word for unsurpassed data quality and reliability of X-ray equipment: mar. Please come in and get more information about our products: ❍ mar345 Image Plate Detector ❍ marCCD CCD Detector ❍ marCAM X-ray Beam Monitor ❍ Osmic Confocal Max-Flux® Optics ❍ mar Focussing Mirrors Check out the latest news. Aug 24, 2000: marHKL and marXDS data processing packages now available! marFLM updated (version 2.0)! Our staff members will be glad to help you with more detailed information by email, phone or FAX. © 2000 by X-ray Research GmbH http://www.marresearch.com/ [3.11.2000 08:48:26 Uhr] 1st_page Welcome on our Web site Your expert in thermal devices and temperature measurement English - Deutsch This site is optimized for a 600 x 800 resolution. Sales International THERMOCOAX 40 Bld Henri Sellier 92156 SURESNES Cedex - France Tel: +33 1 41 38 80 50 - Fax: +33 1 41 38 80 58 e-mail : info@thermocoax.fr http://www.thermocoax.fr/ [3.11.2000 08:48:54 Uhr] Navigation : best with NETSCAPE or EXPLORER