5. Experimental 4. Experimental Each experiment was performed with the aim of retrieving the maximum of information regarding each lens. Unsuccessful examinations are still considered a result and valuable to this project. Samples 1 and 2 (See details in Appendix B) were received first, with instructions not to damage or contaminate them. They were contained in BSS and all surfaces they came in contact with were throughly cleaned with Isapropanol solution and then rinsed with distilled water. They were handled with a plastic tweezers, to cause minimal damage, and were cleaned in the same way. Sterile gloves were used during all the experiments. After non-destructive testing these samples were sent back to Italy. Shortly after, samples 3, 4 and 5 (see details in Appendix B) arrived. These came with permission to conduct destructive testing. All non-destructive tests were performed first with the same care as samples 1 and 2. While still in possession of samples 3, 4 and 5, samples 1 and 2 arrived back from italy, this time with permission to conduct destructive testing. 4.1 Non-Destructive Testing The following non-destructive analyses were performed on the samples available at the time of testing. 4.1.1 Microscopy Apparatus: 49 5. Experimental An Olympus optical microscope was used, model: bx60m with a TIC-1280E colous video camera connected to sufficient software on the computer to capture bitmap images. This microscope can be seen in Figure 4.1 below. Samples available: Sample 1, 2, 3 and 5. Sample 4 was a special interest lens due to its nature of retrieval. No unnecessary risks of contamination were being performed at this time. Procedure: - Glass slides and a plastic tweezers were cleaned as mentioned earlier. Each sample was taken out of the BSS and placed on a slide. - Both the anterior and posterior sides of each sample were then viewed using different magnifications. 50 5. Experimental - Any images of interest were captured using the camera and saved as bitmap images for future scrutiny. 4.1.2 Metrology Apparatus: A Shadow graft made by Baty International using Quadra check 2000, shown in Figure 4.2, was used to measure the samples. Quadra check is an automatic measurement instrument which is connected to a light sensitive probe. The sample is moveable in the x and y plane (horizontal and vertical direction) relative to the probe. The probe senses and records the shadow created by the sample and automatically measures the distance until the next light change i.e. the end of the shadow. The machine is calibrated in the z-direction (distance from the sample to the back surface) to give correct measurements. An Auto cad drawing package was also used. 51 5. Experimental Samples available: Sample 1, 2, 3 and 5. Sample 4 was a special interest lens due to its nature of retrieval. No unnecessary risks of contamination were being performed on sample 4 at the time in question. As a result it was not involved in this procedure. Procedure: - Glass slides and a plastic tweezers were cleaned as outlined above. Each sample was taken out of the BSS and placed on a slide. - Each slide with sample was then, in turn, placed in front of the light source on the designated spot to project an enlarged view of the IOL on the back surface. - The length from haptic was measured, then the diameter of the optic using the light prob. The results were read from the LED display on the Quadra check. - The curvature of the IOL was then measured by taking three points along the shadow of the curve and recording their x and y positions. - The three x and y values were then input into a drawing package which determined the best fit curve and subsequently, the radius. 4.1.3 Optical Measurements Focal length is the distance from the lens to the image plane when parallel rays of light impinge upon the lens. Apparatus: The Apparatus used included a metal bar with a small hole at the end of it and an adjustable height, a light source from far away which creates parallel rays of light, a ruler and a sheet of white paper. This set-up is illustrated in Figure 4.3. 52 5. Experimental Samples available: Samples 1, 2, 3 and 5. No unnecessary risks of contamination were being performed on sample 4 at the time in question. Procedure: - The lenses were taken from solution and blotted dry. - They were in turn placed at the end of the bar with the lens optic over the hole. - The bar and lens arrangement was then held under beams of parallel light and the height adjusted until a sharp spot of light could be seen on the sheet of paper. - The height of the lens from the sheet of paper was then recorded, this is the focal length, f. - Using this value and applying it to equations (1) and (2) in section 3.1 of the theoretical analysis, values for D, the dioptric power, were calculated. 53 5. Experimental 4.1.4 Atomic Force Microscopy AFM is a branch of Scanning Probe Microscopy (SPM). It produces images by scanning a surface using a probe mounted on a cantilever that is driven by piezoelectric ceramic materials, and gives information about the outermost surface of the sample. Deflections in the very small spring-cantilever occur due to probe-sample interaction forces as features are encountered on the sample surface during scanning, giving very accurate height measurements of any irregularities or deposits. A signal is then collected and processed to produce a 3D image of the sample surface. AFM uses the forces that occur when two objects are brought within nanometres of each other. In contact mode the probe obtains a repulsive force. AFM can also be conducted in non contact mode, when the probe is a few nanometres away, causing an attractive force. The forces are usually in the nano-Newton range. AFM can be utilised in air, liquid or a vacuum (Russell et al., 2004). One of the methods for measuring the deflection of the cantilever is the use of a laser beam, reflected from the back of the spring cantilever. The motion of the probe is sensed by the displacement of the reflected beam on a photodetector seen in Figure 4.4. Apparatus: The AFM images reported in this study were obtained using a TopoMetrix Explorer Scanning Probe Microscope (ThermoMicroscope-VEECO), seen in Figure 4.5, in contact mode in the BSS solution. The tip used was silicon nitride with a 4micron base, 4micron long and a tip of radius <50nm. This tip was attached to a 200micron long cantilever. The tip and lever can be seen in Figure 4.6. The raw data collected was processed using horizontal and vertical levelling, left shadowing and equalization of the colour. 54 5. Experimental Samples available: Sample 2. Due to funding and financial restrictions only one sample could be analysed. Procedure: 55 5. Experimental - The sample was taken straight from the BSS solution and held between two glass slides under the microscope (There was a glass slide to the left and right of the sample holding it in place and restricting it from moving). - The BSS solution was then poured over the lens so that it covered it completely. - A microscope connected to a video camera showed views of the cantilever and tip, in order to find the correct positioning on the sample. - Four random spots were analysed on the posterior side of the lens in contact mode, meaning the tip is in physical contact with the sample. - The probe moves across the surface in the x-y plane and the movement in the z plane as a response to the probe-sample interaction is stored on the computer system to generate a 3D representation of the surface. 4.1.5 UV Analysis UV light coming from the sun occurs at a wavelength below the visible spectrum (<400nm). These are harmful rays and can cause damage to the inner eye. The natural lens aborbs most of these rays before they reach the retina. This natural protection is removed along with the cataract lens. All IOLs, therefore, contain UV absorption properties. This experiment analyses how well the IOL UV blockers work by graphing their percentage absorbance of UV light. If the harmful rays are absorbed by the IOL they do not reach the inner workings of the eye. Apparatus: The Cary-100 spectrometer was used with a tungsten filament lamp and a deuterium lamp, which operate in different wavelenghts. the IOL sample. A quartz cuvette held Quartz does not absorb UV light, providing an accurate measure of UV absorption of the samples. The Cary-100, seen in Figure 4.7, is a dobule beam, recording spectrophotometer controlled by a computer operating under Windows2000, and running Varian’s WinUV software. 56 5. Experimental Samples Available: Sample 1, 2, 3 and 5. Sample 4 was a special interest lens due to its nature of retrieval. No unnecessary risks of contamination were being performed on sample 4 at the time in question. As a result it was not involved in this procedure. Procedure: - Each sample, in turn, was placed inside the quartz cuvette, positioned so the light beam would strike it at 90° the optic surface. Due to the tacky nature of the hydrated IOLs, they stuck firmly to the side of the cuvette. - The spectrometer was run at wavelengths between 250nm and 600nm for each sample and graphs plotting wavelength versus percentage absorption were recorded on the software. 4.1.6 Swell test 57 5. Experimental A swell test tells how much water the lens absorbs or takes in. As two of the samples used were hydrophobic any water taken on indicated a polymer gap caused by manufacturing and could be the cause of glistenings or voids. Apparatus: Sartorious enclosed electronic weighing scales with accuracy to 5 decimal places seen in Figure 4.8. Samples Available: Sample 3, 4 and 5. Sample 1 and 2 were in Italy for analysis at the time of this procedure. Procedure: - Sample 3 and 5 were firstly weighed in the fully hydrated state. - All samples were left to dry out for a few days. - They were then weighed in the fully dehydrated state. 58 5. Experimental - The samples were placed in distilled water for one hour, taken out using the tweezers, and the excess water was removed by dabbing them with tissue. Again, they were weighed and the weights recorded. - They were placed back in distilled water and left for a further 23 hours. They were again dabbed dry and weighed. These final values were recorded. - Using equation (3) and (4) in section 3.2, theoretical analysis, the Equilibrium water contents(EWCs) were calculated. 4.1.7 Scanning Electron Microscopy (SEM) SEM operates by creating a voltage between a conductive sample and a filament, causing an electron emission from the filament to the sample. The electrons are guided to sample by a series of electromagnetic lenses in the electron column. This occurs in a vacuum environment. The resolution of the image is determined by the beam current and spot size (size of area of the sample being viewed). Electrons interact with the sample and are emitted back from the sample as a signal used for investigations of surface morphology. Once these electrons escape from the sample surface they are detected by a scintillater-photomultiplier detector. The SEM image formed is the result of the intensity of the electron emission from the sample. The sample must be conductive to interact with the electrons (Russell et al., 2004). Apparatus: The Joel 5600 Scanning Electron Miscoscope, seen in Figure 4.9, with compatible image capturing software. 59 5. Experimental Samples available: Sample 1 and 2. Procedure: - Sample 1 was blotted dry and left uncoated to see if any images could be obtained without causing damage to the lens. - It was placed in the SEM chamber on a tungsten disc and viewed with a low accelerated voltage of 3kV. 4.2 Destructive Testing The following destructive analyses were performed on the samples available at the time of testing. 4.2.1 Microscopy The samples may have harboured dirt deposits or dead proteins while examined by microscopy before. By soaking the samples in disinfectant all these unwanted residues were removed. This experiment was considered destructive 60 5. Experimental because it is not known if the fluid removes any deposits which may be of interest. Apparatus: An Olympus optical microscope was used, model: bx60m with a TIC-1280E colour video camera connected to sufficient software on the computer to capture bitmap images. This microscope can be seen in Figure 4.1 above. Samples Available: Sample 3, 4 and 5. Sample 1 and 2 were in Italy for analysis at the time of this procedure. Procedure: - The IOLs were soaked overnight in undiluted Milton sterilising fluid which contains sodium chloride. They were then rinsed with distilled water and viewed as explained earlier in section 4.1.1. 4.2.2 Microhardness Shore hardness is used to determine the relative hardness of soft materials, usually plastic or rubber. It is a measure of the resistance of material to indentation by a 3 spring-loaded indenter. The scale provides an empirical hardness value that does not correlate to other properties or fundamental characteristics. The Shore hardness is measured with an apparatus known as a durometer and as a result is also known as “Durometer hardness” (Matweb, 2005). Apparatus: Zwick Roell Shore D handheld microhardness tester as seen in Figure 4.10. 61 5. Experimental Samples available: Sample 3, 4 and 5. Procedure: - The specimen’s were tested in the hydrated state. The tip of the tester, the surface used and the tweezers were all cleaned as previously described. - The samples were in turn placed on the cleaned, hard, flat surface. - The indenter for the instrument was then pressed into the sample making sure it was parallel to the surface. - The hardness was read within one second of firm contact with the specimen, and each sample was checked to ensure the indenter did not penetrate it completely. 4.2.3 Scanning Electron Microscopy (SEM) SEM is discussed in section 4.1.8. Apparatus: An Edwards Sputter Coater S150B seen in Figure 4.11 and a Jool JSM-840 electron microscope seen in figure 4.12. 62 5. Experimental Samples available: Sample 3, 4 and 5. Sample 1 and 2 were in Italy for analysis at the time of this procedure. Procedure: - After soaking in Milton and subsequently transferred to distilled water the samples were left out of the solution for a number of days to dry. - The posterior side of each of the samples were sputter coated with gold to make them conductive and placed in the SEM chamber. - A very low 3kV accelerated voltage was used to view the samples. 4.2.4 Differential Scanning Calorimetry (DSC) DSC measures the amount of energy absorbed or released by a sample when it is heated or cooled. It provides data on endothermic (heat absorption) and endothermic (heat evolution) processes. Thermal analysis may be defined as the measurement of the physical and chemical properties of materials as a function of temperature and time. The different thermal properties obtained from DSC analysis include the glass transition temperature and solid-state reactions including nucleation temperatures, crystallisation termperatures and endothermic or melting reactions (Anderson Materials Evaluation Inc., 2004). Apparatus: 63 5. Experimental DSC 10 Differential Scanning Calorimetry made by TA instruments, seen in Figure 4.13. Aluminium pans and crimper. Liquid nitrogen and nitrogen gas. Samples Available: All Procedure: - The samples were dry after being left out of solution for some days. A small piece was cut from each. - The sample was placed in an aluminium pan and covered with a lid, Figure 4.14. It was then closed by pressing in a crimper, seen in Figure 4.15. An emptyreference pan was also prepared. 64 5. Experimental - The crimped pans were then placed in the appropriate spots in the furnace shown in Figures 4.16 and 4.17. - The thermal program was then set up on the computer. The material name and temperature information is input. For this experiment the DSC Ran at 10°C per minute from -50°C to 200°C, isothermal for 1 minute. - The calorimeter was set running to burn off any dirt or dust in the chamber. 65 5. Experimental - The sample was then placed in the furnace and Liquid nitrogen was poured on top of it until the temperature reached -50°C. The computer turned on the furnace, making sure that the heating rate stayed the same throughout the experiment and that the two separate pans heated at the same rate as each other. - The furnace heated and the temperature differences between the sample and reference was monitored so that the DSC can keep the temperatures the same. If an exothermic phase transition occurs, the temperature of the sample will tend to surge ahead of the reference. In this case the DSC does not need to provide as much heat to the sample. If an endothermic transition occurs, the DSC must supply more heat to the sample. The electrical power difference between the sample and reference measures the heat flow in the sample. - During the experiment a stream of nitrogen gas was directed inside the furnace to purge it of oxygen. This prevented oxidation of the sample during heating. - The calorimeter stopped automatically on reaching 200°C producing a heat flow versus temperature graph. The temperature is also directly related to the time in minutes. 4.2.5 Calcium Testing and Energy Dispersive X-Ray analysis The calcium stains described in the literature review were not available in the University. Along with these stains researchers such as Yong et al., (2004), used X-Ray analysis on the samples. The X-Ray analyser also proved the presence of Calcium. The analyser can carry out elemental analysis on spot sizes as small as 2nm. Along with electron emission used to form a morphological image of a surface in SEM a number of other signals are emitted as a result of the electron beam impinging on the surface. Each of these signals carries information about the sample which provides clues to its composition. X-Ray signals are frequently used to provide elemental analysis by the addition of an Energy Dispersive Spectrometer to the SEM system. The interaction between the electron beam and the sample results in the expulsion of an inner shell election 66 5. Experimental from the atom creating a vacancy that is filled by an outer shell electron. This jump from outer to inner shell results in a change in energy that produces an XRay. These are then compared against an X-Ray spectrum of the periodic table of elements. Apparatus: The Jool JSM-840 Scanning Electron Microscope seen in Figure 4.12 equipped with an energy dispersive (EDX) analyser. Samples available: Sample 4 and Sample 1. Procedure: - The analysis was run in conjunction with SEM pictures taken as described in section 4.2.3 above. - Sample 4, previously gold coated, was cut in half before being placed in the chamber. A spot on the inner surface of the sample was analysed using the X-Ray. The posterior surface was also examined. Unusual deposits seen previously using SEM were targeted and analysed. - After drying for a number of days sample 1 was sputter coated with gold and analysed using the energy dispersive X-Ray analyser. Sample 1 was taken straight from the BSS before drying and had never been disinfected using Milton. 4.2.6 Fourier Transform Infrared Spectroscopy (FTIR) FTIR is an analytical technique used to identify organic or inorganic materials. It measures the absorption of various infrared light wavelengths of the material of interest, and identifies types of chemical bonds in a molecule. These infrared bands identify specific molecular components and structures. The region between 1500 – 4000 wave numbers is referred to the fingerprint region. Absorption bands in this region are highly specific for each material. This specificity allows a computerised data search to be performed against a 67 5. Experimental reference library to identify the material. FTIR can be applied to identify foreign materials such as particles and residues. Two techniques can be used in this analysis, the transmission technique or the reflection technique. The transmission technique is the most simple of the infrared spectroscopic techniques. Infrared radiation is passed through a sample and the transmitted radiation is measured. It is only applicable to thin samples. The reflection technique requires the sample to be placed on a gold disk. The IR light hits the surface of the metal and reflects off it. Only the molecular entities aligned with the electric vector of the incident light will interact with the light, resulting in an absorption band and providing information about the macromolecules at the surface (Materials Evaluation and Engineering, 2000). Apparatus: A Perkin Elmer spectrum 2000 with a FTIR microscope. Samples available: All. Procedure: - Samples 3, 4 and 5had undergone SEM at this stage so the gold material on them was scratched off the IOL surface. - Each sample was then sectioned and compressed in a dye to obtain a thin cross section for analysis. - The samples were then placed on Potassium Bromide (KBr) solid disk, as this is mostly invisible to the IR method. - The spectrums were obtained using the transmission technique through the lenses with the FTIR microscope. All spectrums obtained are relative to the KBr background disk. 68