OneWorld Community Health Centers Policy and Procedure INFECTION CONTROL TITLE: Instrument Sterilization Policy APPLICABLE STANDARDS: Joint Commission on Accreditation of Healthcare Organizations (JCAHO’s) Comprehensive Accreditation Manual for Ambulatory Care (CAMAC) IC.02.02.01 OSHA Standard 29 CFR 1910.1030 REFERENCES: CDC MMR Volume 42/No RR-8 OBJECTIVE: To remove, transport and sterilize used biomedical instruments properly in order to minimize the risks of contaminating other instruments or transmitting an infection to a patient or staff member. PROCEDURE(S): 1. SEPARATION: Once the procedure requiring the instruments has been completed, the patient will be escorted out of the room, as appropriate. 2. COLLECTION: Nursing Personnel, wearing gloves, will collect all instruments and place them in a covered, puncture resistant, plastic instrument container with the appropriate bio-hazardous label affixed. The empty containers should be kept under the sink in each exam/procedure room. 3. TRANSPORTATION: The instrument container, with cover firmly affixed, shall be transported by appropriately trained and protected personnel to the instrument cleaning area. Metal reusable instruments will be removed taken directly to the sterile processing room in a covered transport container after the patient encounter. Dirty speculums will be removed from the exam rooms at the end of the clinic day and transported directly to the sterile processing room in a covered transport container. 4. LOCATION: The soiled instrument container will be placed at a designated location of the cleaning cycle area. All speculums and instruments will be kept soaking (completely covered) until they can be put through the ultrasonic cleaner. 5. PROTECTION: Appropriately trained staff shall wear scrubs, lab coat or plastic apron, rubber utility gloves when transferring instruments from the transport container to the Ultrasonic cleaner. Protective eyewear, goggles and/or mask may be during this phase of the cleaning process if the likelihood of splashing occurs. 6. EVALUATION: Appropriately trained staff shall evaluate each instrument to determine if any of the instruments are visibly soiled. If they are visibly soiled, then the staff member should try rinsing the instruments off under running water first, then scrubbing with an appropriate brush, if necessary. Instruments to be sterilized must be clean, dry and free from any residual matter such as tissue and/or blood. 7. IMMERSION: The container will then be filled with Empower Metrex for one hour. The liquid should completely submerge all instruments. An Empower Metrex is recommended for this step as it further assists in the cleaning and disinfection process. Page 1 of 7 D:\533571433.doc Reviewed 2/10 OneWorld Community Health Centers Policy and Procedure INFECTION CONTROL 8. DISINFECTION: The instruments will be allowed to soak according to the manufacturer’s guidelines for destruction of HIV virus, TB, and other infectious agents. However, in the absence of such guidelines, a minimum soak time of at least 1 hour shall be required. 9. EVACUATION: Once the instruments have been through the ultrasonic cleaning process using Metriclean 2 Metrex, appropriately trained staff shall don new gloves, a lab coat or scrubs. The instruments shall then be removed one at a time from the ultrasonic machine rinsed under a steady stream of tap water. The objective in this stage is to evacuate the instruments and remove any remaining debris or visual signs of contamination. If visual contamination is present, the instrument may be scrubbed with risk of body fluid exposure. 10. EVAPORATION: Once the instruments have been ultrasonically cleaned and rinsed, they shall be placed on a towel in a relatively isolated area and allowed to air dry completely. 11. ISOLATION: Appropriately trained staff will then remove their protective equipment used in steps five (5) though nine (9), wash their hands and don new gloves. The instruments will then be packed as follows: a. Speculums are placed directly on the tray along with an indicator strip. b. Other individual instruments will be placed in individual autoclave bags (indicator strip is on the outside of each bag). Scissors and forceps will be placed in the pouch with pointed end facing up (the open end). c. Autoclave bags should then be sealed and placed on the autoclave instrument trays. 12. STERILIZATION: Once the instruments have been properly packed, they should be placed directly into the autoclave and run through a complete cycle, as follows: a. Packages with sharp instruments in them should not be allowed to touch each other. Use sterile gauze or cotton between the instruments to protect each other from puncture during the sterilization process. b. Any items that might hold water should be placed in a manner which facilitates drainage c. DO NOT overload sterilizer tray. d. Operating the Sterilizer (note: staff operating the sterilizer should also be familiar with the manufacturer’s instructions for safely running the equipment): I. Make sure unit is plugged into electrical outlet II. Check the distilled water level in the sterilizer, the water line should be in the GREEN zone. If the water line is not in the GREEN zone, add more distilled water until the water level raises into the green zone. III. Load autoclave and place an indicator strip in with load to be sterilized. DO NOT PUT PLASTIC INSTRUMENTS IN THE AUTOCLAVE. IV. Shut the door (pull up on the handle to close tight). V. Press the “On/Standby” button (bottom right). VI. Select either the “unwrapped” button or the “pouches” button (bottom left) VII. Push the green “start” button (bottom right) VIII. When the cycle is done, the door will automatically open between the sterilizing and dry cycles. IX. When the door opens the autoclave is cooling down the load. Page 2 of 7 D:\533571433.doc Reviewed 2/10 OneWorld Community Health Centers Policy and Procedure INFECTION CONTROL X. When the instruments are cool, remove the load. XI. Check sterilization indicator strips If the indicator strip has changed from pale yellow to black then the item is now sterilized. If the indicator strip has not changed from pale yellow to black (that is, it is still yellow), resterilze the instruments using a new package. DO NOT REUSE THE SAME PACKAGE. XII. Autoclave cleaning to be done monthly using 1 oz of Speed-Clean Midmark per gallon of water. 13. INDICATOR TESTING: A sterilizer indicator test will be run with each load. 14. RESTORATION: The newly sterilized packs should be removed carefully from the autoclave and distributed to the appropriate storage area. (See attachment A) PLEASE NOTE: Sterilized Autoclave packs are assumed to be sterile indefinitely unless the packaging has been damaged in some way (i.e. package tear, water stains, etc.). This “Event Related Sterilization” process means that autoclave packets do not need to be dated. Additionally, nursing staff should monthly visually review the instruments to determine if they must be sterilized again. ** Instrument sterilization packs will be dated, but this date will be used for batch inventory purposes only. 15. SPORE TESTING OF STERILIZER: An internal spore test indicator will be performed weekly. All spore-testing results will be maintained for reference and tracking purposes. All results must be logged in the white “Biological Indicator Monitoring Log Book.” To effectively monitor the performance of the organization’s sterilizers Spore Testing is performed to ensure that patient assessment and care equipment are safe for use on other patients. The CDC MMR Volume 42/No RR-8 (1993) and subsequent revisions dating up to the current year (2000), states the following: “Proper functioning of sterilization cycles should be verified by the periodic use (at least weekly) of biologic indicators (i.e., spore tests) (3, 9)-. Heat-sensitive chemical indicators (e.g., those that change color after exposure to heat) alone do not ensure adequacy of a sterilization cycle but may be used on the outside of each pack to identify packs that have been processed through the heating cycle.” Therefore, OneWorld staff will perform biological (spore) testing of every sterilizer in the organization at least weekly following this established procedure: A. Exposure: 1. Place one or more ProSpore2 vials in a horizontal position in the most difficult to sterilize locations per manufacturer’s instruction along with a normal instrument load. Run a normal sterilization cycle. DO NOT PROCESS CONTROL AMPOULE. Page 3 of 7 D:\533571433.doc Reviewed 2/10 OneWorld Community Health Centers Policy and Procedure INFECTION CONTROL 2. Sterilize according to the normal procedure. 3. After the sterilization cycle is completed, remove the test ampoule(s) from the autoclave after cooled 1015 minutes. B. Activation: Items needed: Processed vial(s) that have been cooled but not longer than 15 minutes. One unprocessed (control) vial. 1. Seal the cap by pressing down firmly until flush with the tube. 2. Crush the media ampoule with the tool provided, or by squeezing the sides of the plastic tubes. 3. Inspect the unit to make sure that the spore disc is immersed in the growth media. If the spore disc is not immersed, hold the vial in a vertical position tapping the bottom on a hard surface until the disc becomes saturated. C. Incubation: Place the processed vial(s) labeled with the date, time and autoclave being tested and one unprocessed (control) vial labeled with the date, time and “control” in a vertical position in an incubator at 55-66 degrees Celsius for a period of 24 hours. D. Monitoring: Examine the vials daily during incubation recording observations. All positive vials will be recorded and disposed of immediately. E. Interpretation: Control: The control vial should exhibit a color change and/or turbidity. If the control ampoule does not show signs of growth, consider the test invalid. Test: A failed sterilization cycle is indicated by turbidity and/ or change in color to or toward yellow. A test ampoule that retains its purple color indicates an adequate sterilization cycle. 1. Test results are logged by the person who performed the test. 2. If the culture is positive, which means it has failed will require that the autoclave be tested again, repaired if necessary, and a “pass” spore test result must be obtained prior to the autoclave’s reuse. Page 4 of 7 D:\533571433.doc Reviewed 2/10 OneWorld Community Health Centers Policy and Procedure INFECTION CONTROL 3. Any failed results will require that ALL instruments that were autoclaved since the last spore test to be reprocessed. If these instruments cannot be identified separately then ALL instruments will be resterilized. 4. A log of all spore testing results must be turned in to the review by the Risk Management Committee on a quarterly basis (see policy, Risk Management Sub-Committee). 5. An Incident report will be filled out and given to the COO or PI nurse. 16. In addition to the weekly internal biological (spore) testing of every sterilizer in the organization, OneWorld will perform and send a biological spore testing strip monthly to an outside vendor/testing to serve as an additional infection control measure following this established procedure: 1. Place one test strip (Maxi Test) per manufacturer’s instruction in the center of the sterilizer, and the other biological test strip in the most difficult to sterilize locations along with a normal instrument load. Run a normal sterilization cycle. DO NOT PROCESS CONTROL STRIP. 2. Sterilize according to the normal procedure. 3. After the sterilization cycle is completed, remove the test strips from the autoclave. 4. Complete the Sterilization Log and testing paperwork per manufacturer’s guidelines. 5. Place testing packages per manufacturer’s guidelines into the pre-addressed envelope that already contains the viability control piece and mail the envelope at the earliest possible convenience. Note: (a) The two test pieces plus the control piece must contain the same lot control number. (b) Test data must be mailed in the envelopes provided. 6. Test results will be accessed via Internet. 7. Test results are logged by the person who performed the test using the “Monthly Sterilization and Spore Testing Log” (Attachment A). 8. If the culture is positive (failed), the vendor/testing lab will notify the clinic by phone and then fax the results. 9. A failed result will require that the autoclave be tested again, repaired if necessary, and a “pass” spore test result must be obtained prior to the autoclave’s reuse. Page 5 of 7 D:\533571433.doc Reviewed 2/10 OneWorld Community Health Centers Policy and Procedure INFECTION CONTROL 10. Any failed results will require that ALL instruments that were autoclaved since the last spore test to be reprocessed. If these instruments cannot be identified separately then ALL instruments will be resterilized. 11. A log of all spore testing results must be turned in to the review by the Risk Management Committee on a quarterly basis (see policy, Risk Management Sub-Committee). 12. An Incident report will be filled out and given to the COO or PI nurse. ____________________________________________________ SIGNATURE: Dr. Kristine McVea, MD MPH CMO, OneWorld CHC __________________ DATE Page 6 of 7 D:\533571433.doc Reviewed 2/10 OneWorld Community Health Centers Policy and Procedure Attachment A MONTHLY STERILIZATION & SPORE TESTING LOG DATE CYCLE TYPE STERILIZATION OR SPORE TEST PACK/SPORE TEST ID # CYCLE COMPLETED YES OR NO SPORE TEST RESULTS CONTROL TEST RESULTS INITIALS 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Page 7 of 7