J Am Acad Audiol 10 : 445-457 (1999) Comparison of Performance with Wide Dynamic Range Compression and Linear Amplification Anna C . S . Kam*' Lena L . N . Wong* Abstract This study compared subject performance and preference using a compression-limiting hearing aid set to linear amplification (program 1) and wide dynamic range compression (WDRC, program 2) . The frequency responses of the hearing aid were matched to a 65 dB SPL signal and maximum output to a 90 dB SPL signal . Twenty subjects with moderate to moderately severe sensorineural hearing loss were tested . Speech recognition scores and speech reception thresholds were obtained both in quiet and in noise . Subjective preference for WDRC or linear amplification was measured via a paired-comparison procedure on "loudness appropriateness," "clarity," and "pleasantness" to continuous discourse presented in quiet and in noise . Results suggested that WDRC yielded better speech intelligibility in quiet for low-level signals and no difference in speech intelligibility in noise compared to linear amplification . Subjects preferred WDRC for loudness to both high- and low-level signals and for pleasantness to high-level signals . Key Words: Linear amplification, speech intelligibility, subjective preference, wide dynamic range compression Abbreviations : ANOVA= analysis of variance, AVC =automatic volume control, CL= compression limiting, CVR = consontant-vowel-ratio, HINT = Hearing In Noise Test, P = program, REIG = real-ear insertion gain, REIR = real-ear insertion response, SNR = signal-to-noise ratio, SRS = speech recognition score, SRT = speech reception threshold, WDRC = wide dynamic range compression onlinear amplification or compression systems were originally incorporated Nin hearing aid design to limit the maximum output of a hearing instrument to below the wearer's loudness discomfort level. Before the introduction of compression, peak clipping was the most frequently used method of output limiting (Hawkins and Naidoo, 1993). However, this method causes a considerable amount of distortion that is unpleasant to the listener and may even reduce speech intelligibility (Dreschler, 1988a). To eliminate or minimize these undesirable effects, the use of compression limiting was introduced as an alternative to peak clip*Department of Speech and Hearing Sciences, The University of Hong Kong, Hong Kong ; tcurrently affiliated with Phonak Hearing Centre, Hong Kong Limited Reprint requests : Lena L. N. Wong, Department of Speech and Hearing Sciences, Prince Philip Dental Hospital, 5/F, 34 Hospital Road, Hong Kong ping. Except for people with very profound hearing loss, compression limiting is superior to peak clipping both subjectively and objectively (Gioannini and Franzen, 1978 ; Dreschler, 1988a ; Hawkins and Naidoo, 1993 ; Dillon, 1996) . The use of compression to limit hearing aid output is just one of its applications in hearing aid design . Based upon its function and characteristics, compression used in modern hearing aids can be categorized into three main types: compression limiting (CL), slow-acting automatic volume control (AVC), and wide dynamic range compression (WDRC) (Dillon, 1988 ; Hickson, 1994). The characteristics of the three types of compression systems are summarized in Table 1 . CL is characterized by high compression ratios, high compression thresholds, and short time constants (attack time and release time). This type of compression is designed to limit the output below the listeners' tolerance level, minimize temporal and spectral distortion of 445 Journal of the American Academy of Audiology/Volume 10, Number 8, September 1999 Table 1 Compression System CL AVC WDRC Characteristics of Three Common Types of Compression System (after Fortune, 1996) Compression Threshold (dB SPL) >80 <65 <60 Compression Ratio >5 >5 <5 Attack Release (msec) (msec) <5 10-50 <5 50-100 150-2000 10-100 Time Time CL = compression limiting, AVC = automatic volume control, WDRC = wide dynamic range compression . the acoustic input at low and intermediate levels by providing linear amplification to sounds below the compression threshold, and minimize distortion that may be generated by peak clipping (Dreschler et al, 1984 ; Boothroyd et al, 1988 ; Dillon, 1988 ; Preves, 1991). Hearing aids with AVC typically have intermediate to high compression ratios, low compression thresholds, short attack time, and a very long release time . Because of the long release time, the output of the signal remains relatively constant in the presence of input fluctuations. As a consequence, the need to adjust the volume control of the aid is reduced (Dillon, 1988 ; Hickson, 1994 ; Kuk, 1996). WDRC is characterized by low compression ratios, low compression thresholds, and short time constants. Kuk (1996) and Dillon (1988) limited the definition of WDRC to a system with short release time . This syllabic compression system allows the compression aid to "follow" the envelope fluctuation among syllables seen in speech (Kuk, 1996). The rationale behind WDRC is to match or "squeeze" the normal speech range (from soft to loud speech) to the reduced dynamic range of hearing-impaired people (Steinberg and Gardner, 1937 ; Dillon, 1988). People with sensorineural hearing loss typically have reduced auditory dynamic range (the difference in dB between detection threshold to threshold of discomfort) . In the unaided condition, low-intensity sounds will be inaudible while high-intensity sounds remain loud . Unlike linear amplification, which provides constant gains regardless of the input sound levels, WDRC gives more gain to low-intensity sounds and less gain (gain reduction) to high-intensity sounds . As a result, soft sounds should become audible while listening comfort is ensured even with loud sounds . During the past 2 decades, numerous studies have been carried out to investigate the performance of different amplification systems . Lines of research include comparison of performance between CL and peak clipping (e .g ., Gioannini and Franzen, 1978 ; Dreschler, 1988a; 446 Hawkins and Naidoo, 1993), linear amplification and AVC (e .g., Moore et al, 1991 ; Neuman et al, 1994), linear amplification and WDRC (e .g ., Nabelek, 1983 ; Dreschler, 1988b ; Peterson et al, 1990), and evaluation of multiband compression (e .g ., Plomp, 1988 ; Moore et al, 1992 ; Hohmann and Kollmeier, 1995 ; Yund and Buckles, 1995) . However, the findings in most areas are inconclusive . Theoretical advantages of WDRC over linear amplification are well documented (e .g ., Hickson, 1994 ; Dillon, 1996 ; Kuk, 1996). However, due to problems and differences in the design of various empirical studies, results documenting the advantages are conflicting or inconclusive . Verschuure et al (1993) investigated the effect of syllabic compression (i .e ., WDRC) on speech intelligibility in 19 hearingimpaired listeners. Results obtained via a nonsense consonant-vowel-consonant syllable test in quiet revealed better speech intelligibility with WDRC . The study used experimental hearing aids that were specifically designed so that many parameters of the instrument could be manipulated. Most of the parameters studied could not be altered in commercial hearing aids, making comparisons to other studies very difficult, if not impossible . Accordingly, generalization of the benefits of compression from experimental hearing aids is not readily applicable to commercially available instruments. There are some tentative experimental findings for speech intelligibility improvement in quiet with the use of WDRC instead of linear amplification. For example, Dreschler (1988a) compared the performance using a syllabic compressor with compression limiting and a linear amplifier with peak clipping. Sixteen hearingimpaired subjects with hearing loss ranging from mild to moderately severe participated in a phoneme perception task carried out in quiet. It was found that compression yielded significantly better phoneme identification scores than linear amplification . Nabelek (1983) showed that WDRC was superior to linear amplification Comparison of WDRC and Linear Amplification/Kam and Wong in quiet for 13 subjects with mild-to-severe sensorineural hearing loss . When speech-spectrumshaped noise was introduced, performance with linear amplification was better than with WDRC . An insignificant or negative effect of compression was reported in some studies . Dreschler et al (1984) compared the performance among five hearing aids (a linear aid, two input compressors, two output compressors) . Twelve hearing-impaired subjects were recruited . The speech reception thresholds (SRTs) for sentence material were obtained in quiet and in noise . No significant difference in performance was observed among these hearing aids . Using 16 subjects with mild-to-severe sensorineural hearing loss, Tyler and Kuk (1989) failed to find any significant improvement in consonant identification in babble noise using a single-channel syllabic compressor and its linear version . Hickson et al (1995) evaluated the consonant perception of 15 subjects with mild-to-moderate sensorineural hearing loss using linear amplification and compression amplification with two different compression ratios (1 .3 and 1 .8). No significant difference was found in the scores obtained in a nonsense syllable test in quiet. In the background of babble noise, consonant perception was significantly better with linear amplification than with either form of compression . In this study, speech material was first processed by the hearing instruments and then recorded and played back via headphones during testing. As the hearing aid was not individually fitted, the hearing-impaired subjects' aided hearing ability was not optimized. This also hindered the generalization of the research findings to real-life situations . Different experimental tasks and conditions were employed in the studies, making it difficult to compare findings . For example, some studies have used phonemes (e .g ., Dreschler, 1988b), nonsense syllables (e .g ., Vershuure et al, 1993), and sentences (e .g., Dreschler et al, 1984) as the experimental speech material . When speech intelligibility was measured in noise, some experiments employed multitalker babble (e .g., Hickson et al, 1995), and some used speechspectrum-shaped noise (e .g ., Nabelek, 1983). These differences in experimental tasks and conditions also contributed to the inconsistency of the experimental findings . Experimental Rationale Although the efficacy of WDRC has not been fully validated, such a signal-processing method has already been used in many commercially available hearing aids for some time . Without any well-validated evidence or rationale, what can hearing aid dispensers do when they must select the most appropriate hearing aid or cir- cuit for their clients? Byrne (1996) suggested that the hearing aid dispenser is losing control of the fitting process . The most basic question, "Is compression beneficial to hearing-impaired people?" needs to be answered . An interesting follow-up question to the above is "Is compression beneficial to Cantonesespeaking hearing-impaired people?" In comparison with English, which is an intonation language (i .e ., language uses pitch variations over phrases and sentences to distinguish meaning differences), Cantonese is a tone language, which uses the pitch of individual syllables to contrast meanings (Fromkin and Rodman,1988) . According to Ladefoged (1993), English is a stress-timed language, whereas Cantonese is a syllable-timed language in that syllables tend to recur at regular intervals of time and stress is less important than other prosodic features such as tone . These differences in suprasegmental feature contribute to the difference in spectral and temporal cues of the two languages . Accordingly, it is worthwhile to investigate the effect of WDRC (i .e ., syllabic compression) on Cantonese perception. Previously, performance of two-channel WDRC and one-channel linear amplification on Cantonese perception had been compared in a study done by Wong et al (1996) . In the study, better speech intelligibility using the two-channel aid was found in noise and in quiet. To date, no information is available about the performance of single-band WDRC in Cantonese perception . The main aim of the present study was to investigate the difference in performance with single-channel WDRC and linear amplification in (a) Cantonese speech intelligibility in quiet, (b) Cantonese speech intelligibility in noise, and (c) subjective preference . It was hoped that, with the objective and subjective data, a better comparison of WDRC and linear amplification performance can be obtained . METHOD Subjects Twenty subjects (9 male and 11 female) ranging in age from 16 to 70 (mean = 44 .6, SD = 17 .4) were selected from among experienced hearing aid users who were visiting the 447 Journal of the American Academy of Audiology/Volume 10, Number 8, September 1999 Table 2 Pure-Tone Hearing Thresholds of the Subject's Better Ears (Test Ears) Subject Number 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 Mean SD Hearing Level (dB HL) (re : ANSI S3.6-1989) 250 Hz 500 Hz 65 .00 65 .00 60 .00 65 .00 45 .00 55 .00 55 .00 60 .00 55 .00 45 .00 40 .00 65 .00 45 .00 55 .00 65 .00 60 .00 50 .00 45 .00 40 .00 65 .00 55 .00 65 .00 65 .00 70 .00 80 .00 55 .00 60 .00 70 .00 70 .00 80 .00 70 .00 60 .00 65 .00 60 .00 75 .00 65 .00 55 .00 55 .00 60 .00 70 .00 45 .00 55 .00 60 .00 60 .00 70 .00 50 .00 55 .00 50.00 55 .00 70 .00 55 .00 60 .00 50.00 50 .00 65 .00 60 .00 65 .00 65 .00 65 .00 75 .00 55 .00 55 .00 60 .00 60 .00 65 .00 50 .00 50 .00 60 .00 65 .00 70 .00 45 .00 50 .00 55 .00 55 .00 65 .00 60 .00 65 .00 60 .00 50 .00 65 .00 45 .00 45 .00 45 .00 50 .00 55 .00 60 .00 65 .00 60 .00 60 .00 70 .00 65 .00 65 .00 70 .00 70 .00 75 .00 55 .00 55 .00 55 .00 60 .00 70 .00 50 .00 60 .00 60 .00 65 .00 70 .00 40 .00 40 .00 45 .00 50 .00 65 .00 45 .00 45 .00 50 .00 50 .00 65 .00 65 .00 70 .00 70 .00 65 .00 75 .00 54 .50 57 .00 58 .50 59 .50 69 .75 8 .26 8 .01 7 .80 7 .05 5 .96 55.00 9 .03 1000 Hz 2000 Hz 4000-8000 Hz Hz Audiology Clinic of the Department of Speech and Hearing Sciences at the University of Hong Kong or the Phonak Hearing Centre Hong Kong Limited. All had bilateral sensorineural hearing loss . To be included in the study, the degree of loss in the better ear (i .e ., the test ear) must be in the moderate to moderately severe range (i .e .' the pure-tone average for 0.5, 1, and 2 kHz lies between 42 to 68 dB HL) and the configuration of the loss must be flat (i .e ., the difference in hearing level between 500 and 4 kHz was 15 dB or less). Table 2 shows the pure-tone hearing thresholds of the subjects' better ears (test ears). Average pure-tone thresholds ranged from 40 to 67 dB HL (mean = 56 .7, SD = 7.3). All subjects were experienced hearing aid users with hearing aid experience of 1 to 23 years (mean = 7.9, SD = 6.6). Only Cantonese speakers with good oral communication abilities were recruited. The profile of the subjects is shown in Table 3. Equipment and Procedure All testing was carried out in a sound-treated room at the Audiology Clinic of the Department of Speech and Hearing Sciences at the University of Hong Kong . Each subject was tested for about 2 hours in each session. The hearing aid 448 was programmed for the test ear (better ear) before the subjects arrived . Hearing re-evaluation, hearing aid verification, speech intelligibility measurement both in quiet and in noise, and subjective rating were done within the same session. Hearing Instrument A Phonak behind-the-ear instrument, model Piconet2 P2 AZ, was used as the experimental hearing aid. Maximally, three hearing programs could be set and the circuit type could be selected independently in each program. The programs set could be switched via a digital remote control . It had a Multi Dynamic Compression Control, which could be set to yield linear amplification or WDRC . In this study, Program 1 (Pl) was set to linear amplification and Program 2 (P2) was set to WDRC . Both programs exhibited compression limiting with adaptive release time . The static and dynamic characteristics of the circuits are shown in Table 4. Hearing Aid Programming Before the subjects arrived, the target 2-cc coupler gain based on the subjects' most recently obtained audiogram was calculated using the Table 3 Profile of Subjects (N = 20 ) Factor Characteristics Data Gender Age Type of hearing loss Degree of hearing loss 9 male, 11 female 16-70 years old (mean = 44 .6) 20 sensorineural hearing loss 6 moderate (pure-tone average of 0 .5, 1 kHz and 2 kHz = 41-55 dB HL) 14 moderately severe (puretone average of 0 .5, 1 kHz, and 2 kHz = 56-70 dB HL) 20 flat (<15 dB difference between 1 and 2 kHz) 3 congenital 1 head trauma 6 presbycusis 7 viral infection 3 unknown 10 monaural, 10 binaural behind-the-ear users 1-23 (mean = 7 .9) Audiogram configuration Cause of hearing loss Hearing aid experience Years worn hearing aids Previous hearing aid circuit type 2 WDRC with CL, 18 linear with CL WDRC = wide dynamic range compression, CL = compression limiting . Comparison of WDRC and Linear Amplification/Kam and Wong Table 4 Circuit Amplification Char acteristics of Different Circuits in the Experimental Hearing Aid Compression Limiting Compression Characteristics CT CR AT RT Linear + CL N/A N/A N/A N/A WDRC + CL 45 1 .1 :1 up 5 30 to 2.7 :1 1 CT CR 80 8 :1 AT 5 RT Adaptiv e CT = compression threshold, CR = compression ratio, AT = attack time (msec), RT = release time (msec), CL = compression limiting, WDRC = wide dynamic range compression, N/A = not applicable . FIG6 (3 .0, Rev L) program. P1 of the instrument was set to linear amplification and P2 was set to WDRC . The 2-cc coupler gain of P1 and P2 was matched to the prescribed gain for 65 dB SPL input. Because the Phonak aid is a singlechannel device, the compression ratio for WDRC setting was set to the average value of the prescribed ratios for both the high- and low-frequency bands. The mean compression ratio used for the group is 2.35 (SD = 0 .30) . FIG6 was selected since it is designed for prescribing gains and compression ratios for compression hearing aids . The prescribed gain for 65 dB SPL input was also used to prescribe gain for the linear amplification to ensure similar response between the two programs . As the OSPL 90 of P2 would be set automatically according to the selected compression ratio, no adjustment was made . OSPL 90 of P1 was set to match that of P2 . The volume control wheel and the on-off switch on the instrument was deactivated. Frequency Response Fine Adjustment With the loudspeaker positioned 1 meter from the subject at 0" azimuth, subjective feedback to 65 and 80 dB SPL (root mean square) continuous discourse was collected. Subjects were asked to rate the signals using a 7-point scale: "cannot hear," "very soft," "soft," "comfortably loud," "loud," "very loud," and "intolerably loud ." The objective was to ensure that conversational level (65 dB SPL) speech signals were perceived as "comfortably loud" and high input level (80 dB SPL) signals were not "intolerably loud ." It was found that no adjustment was necessary for these subjects . Hearing Aid Veriffcation At the beginning of the session, a routine re-evaluation of the subject's hearing was done . The hearing thresholds of all subjects across all of the tested octave frequencies were within ±5 dB of the previously obtained value. After confirmation of the subject's hearing status, hearing aid verification was done . Real-ear insertion response (REIR) was measured using the composite noise signal from the Fonix 6500 real-ear analyzer at three input levels : 50, 65, and 80 dB SPL. These steps were performed to verify the electroacoustic performance of the hearing aid and to verify that the frequency response actually matched the target REIR . These input levels were also used to assess speech intelligibility. Speech Material for Speech Intelligibility Measure The Monosyllabic Cantonese Word Lists (Lau and So, 1988) and the Hearing In Noise Test (HINT, Chinese version) (Nilsson et al, 1994 ; Wong et al, in preparation) were used . The test stimuli of both tests were recorded by a male native Cantonese speaker. The speech stimuli were recorded on a CD-ROM. The recordings were played via a personal computer through a GSI-16 audiometer to a loudspeaker located 1 meter from the subject at 0° azimuth. The monosyllabic word lists were used to measure word recognition ability. There were 10 lists of monosyllabic words and 1 list was presented in each testing condition to determine the speech recognition score (SRS), which was defined as the percent of syllables correctly repeated . The HINT (Chinese version) was used to determine SRT, which was defined as the presentation level necessary for a listener to recognize the speech materials correctly 50 percent of the time . Four 10-sentence lists, each with sentences of equal level of difficulty and phonemic content, were used . One list was presented in each testing condition. 449 Journal of the American Academy of Audiology/Volume 10, Number 8, September 1999 Speech Intelligibility in Quiet One monosyllabic word list was presented at each of three sound levels : 50, 65, and 80 dB SPL. The presentation levels were selected to evaluate the performance with sounds of low input level just above the compression threshold of WDRC (50 dB SPL), high input level above the compression threshold of compression limiting (80 dB SPL), and at everyday conversational speech level (65 dB SPL) . The subjects were required to repeat aloud what they have heard. SRSs were obtained for both hearing aid programs . For the HINT, one sentence list was presented with each hearing aid program. The subjects were instructed to listen and repeat aloud whatever was heard or understood. An adaptive up-down strategy described by Nilsson et al (1994) was used to adjust the sentence presentation levels . The first sentence of a list was presented below threshold and the level was increased in 2-dB steps until the sentence was repeated correctly. The subsequent sentences were presented once each, with the presentation level dependent upon the accuracy of the preceding response . Presentation levels were decreased by 2 dB after a correct response and raised by 2 dB after an incorrect response . SRT using each program was estimated as the mean presentation level calculated from the fifth to tenth sentences in the list . Speech Intelligibility in Noise One monosyllabic word list was presented in each testing condition. SRSs in noise with noise fixed at 65 dB SPL and at -9, -6, -3, 0, +3, +6, +9 dB SNRs were obtained. The levels were chosen to survey a range of performance across SNR. Afour-talker babble (two female, two male) was used as competition signal during tests for speech recognition in noise. The sentence lists of HINT were presented in a background of spectrally matched noise fixed at 65 dB SPL. SRTs (in terms of SNRs) were obtained by a similar procedure as that was used in quiet test . For tests in noise, both signals and noise were presented from the same loudspeaker. For both tests in quiet and in noise, the sequence of presentation of word lists and sentence lists and the order of the program being evaluated were randomized . The sequence of testing conditions or presentation levels was also randomized to counterbalance any prac450 tice and fatigue effect . The subject was blinded to the program in use and was not informed of the difference in the programs before finishing the experiment . Subjective Preference Measure Another set of 12 HINT sentences was used for sound quality rating. One sentence was presented in each testing condition and 12 listening conditions were evaluated. Preference of "loudness appropriateness," "sound clarity," and "sound pleasantness" was rated for the programs in quiet at 50, 65, and 80 dB SPL and at +6 dB SNR with the spectrally matched noise fixed at 65 dB SPL. The three rating categories were chosen as they were relatively more concrete and easier to define compared to other dimensions, such as brightness, sharpness, spaciousness, or fullness, which had been commonly measured in other studies (e .g., Balfour and Hawkins, 1992 ; Lundberg et al, 1992). This made the task simpler, especially for the older subjects . The SNR of +6 dB was selected as the test condition in noise because the speech intelligibility at this SNR had been rated as satisfactory to good by normal-hearing and hearing-impaired subjects in Lazarus's (1985) study (cited in Bachler and Vonlanthen, 1994). Sound quality judgment on unintelligible speech might be very difficult and unreliable . A paired-comparison procedure was used to obtain subject preferences for linear amplification (Pl) and WDRC (P2) . The same sentence was presented in a listening condition while the hearing aid was switched to either programs via the remote control held by the experimenter. In evaluating the preference for loudness appropriateness, the subject had to indicate which presentation of the sentence sounded more suitably loud . For clarity, the subject had to indicate which presentation of the sentence sounded clearer, that is, from which of the two he/she could extract the text more easily. Pleasantness is independent of intelligibility. Hence, the subject had to indicate which presentation sounded pleasanter regardless of the ease of understanding. In each listening condition, the subject was allowed to switch back and forth between presentations with P1 and P2 as many times as was necessary before a decision was made . After the subject had decided a preference for either program, he/she was asked to assign a strength to that preference . That is, the subject had to indicate whether the preferred program was (a) Comparison of WDRC and Linear Amplification/Kam and Wong t 50 dB WDRC - 65 dB WDRC t 80 dB WDRC t 50 dB SPL - .n-- 50 dB Lmea o - 65 dBr Linear 80 &B L- 65 dB SPL -+-80 dB SPL 0 750 500 1500 1000 2000 3000 6000 4000 FrequencYLHz; 5O Figure l Mean REIGs for the WDRC program at three input levels (N = 20). much better, (b) moderately better, or (c) just slightly better than the other program . The subject was also encouraged to explain the preferences . The subjects were blinded to the program in use to minimize any subject bias . The sequence of listening conditions and the order of programs being evaluated were randomized to counter any task order effect . RESULTS Real-Ear Measurement Mean real-ear insertion gains (REIGs) at three tested input levels are shown in Figures 1 and 2 . REIGs using the two amplification schemes are overlaid in Figure 3 . The gains across all tested octave frequencies at 50 dB SPL were significantly higher than that at 65 dB SPL (p < .01) with the WDRC program, whereas the gains at these two levels did not differ significantly with the linear program . Both programs provided significantly less gain at 80 dB SPL than that at 65 dB SPL (p < .01) across all tested frequencies . When REIGs between the programs were compared (see Fig . 3), significantly more gain 500 1000 1500 2000 3000 Frequency (FLz ; 4000 6000 Figure 3 Mean REIGs for the linear and WDRC programs at various input levels . was provided by the WDRC program to 50 dB SPL input at 250, 500, 1000, 1500, and 2000 Hz (F = 7.94, p < .05) . There was no significant difference in REIG at the input level of 65 dB SPL. At 80 dB SPL, the WDRC program provided significantly less gain at 250 and 500 Hz (F = 7 .94, p < .05) than the linear program . Speech Intelligibility In Quiet As shown in Figure 4, the mean SRSs obtained using the monosyllabic word test with WDRC were better than those obtained with linear amplification at all presentation levels in quiet. A two-way analysis of variance (ANOVA) with repeated measures on both factors (program and presentation level) was performed to investigate the effects of type of amplification and test condition. Results are presented in Table 5. Both factors and their interaction were significant, indicating that the effect of program is significantly different for distinct presentation levels . As shown in Table 6, a matched-pair t-test revealed a significant difference between programs only at the presentation level of 50 dB SPL (t = 4.54, p < .01) . Mean SRTs obtained using the HINT in quiet with the two programs are shown in Figure 5 . Results of a matched-pair t-test revealed a significantly better SRT with WDRC (t = 2 .36, p < .05) . t 50 dB SPL --~ -- 65 dB SPL --r--80 dB SPL 250 500 1000 1500 2000 Frequency (Hz', 3000 4000 6000 Figure 2 Mean REIGs for the linear program at three input levels (N = 20). Speech Intelligibility In Noise Mean SRSs obtained using the monosyllabic word test in noise are shown in Figure 6. Results of ANOVA of the data, as presented in Table 7, revealed a significant effect of test condition (SNR) and a significant interaction of program and test condition. This suggests that the effect of program is significant in certain test 451 Journal of the American Academy of Audiology/Volume 10, Number 8, September 1999 Table 5 Analysis of Variance for Speech Recognition Scores Obtained in Quiet with Both Programs Source df Effect MS Effect df Error MS Error F Program Presentation level Program x presentation level 1 2 2 1880 .21 9799 .38 473.96 19 38 38 136 .35 259 .02 51 .15 13 . 79" 37 . 83" 9 .27" MS = mean square .p < .01 . conditions only. A matched-pair t-test was performed to investigate any significant difference in performance in each test condition . Results are summarized in Table 8. Better SRSs were obtained with "RC at a SNR of +6 dB (t = 2 .49, p < .05) and with linear amplification at SNRs of 0and-6 dB (t=2 .87,p< .01 ;t=3 .24,p< .01) . Mean SRTs obtained using the HINT in noise are also shown in Figure 5. No significant difference was found with matched-pair t-tests . Subjective Preference Subjective preference for the two different hearing aid programs in the different test conditions is shown in Table 9. The strength of preference for loudness, clarity, and pleasantness under various test conditions is shown in Figures 7, 8, and 9, respectively. A sign test was carried out to investigate any significant difference in preference . Significant subjective preference was found for WDRC for loudness appropriateness to signals at 50 and 80 dB SPL and for pleasantness to signals at 80 dB SPL (p < .05) . No significant preference for clarity was found in any test condition. DISCUSSION quiet. That is, WDRC yielded better word recognition in quiet, at least to low-level signals. One possible reason for this finding is the better audibility ensured by WDRC . WDRC provided significantly more gain to input at 50 dB SPL than did linear amplification . As the signal level increased from 50 to 65 dB SPL, the gain provided by WDRC decreased to an amount approximately equivalent to that provided by linear amplification, accounting for equivalent performance between programs . When the signals were presented at a high level (80 dB SPL), the gain provided by WDRC decreased further. For linear amplification, the gain also decreased due to activation of compression limiting, but to a lesser extent than WDRC . Although signals were louder, speech intelligibility did not improve significantly with linear amplification. At such a high presentation level, audibility is no longer a dominating factor for speech intelligibility. Other factors, such as distortion in the auditory system or the hearing aid, contribute to the difficulties of speech perception experienced by those with moderate or greater cochlear losses (Moore, 1996). For sentence material, the SRT obtained in quiet with WDRC was significantly lower than Speech Intelligibility in Quiet The SRSs obtained in the monosyllabic word test with WDRC were significantly better than those with linear amplification at 50 dB SPL in 90 a o. bo ~ 50 ------------ ------------------------°------------1----------------------- . --- = Mean+SD Mean-SD ° Mean 40 ,~ 30 - --------- 2 to -------- " o 10 a; o .. .{. . .------~--- ..--;. . . .. . . . . 13 ~ -10 a ~ -20 ------------------- ---- - - ----- ' ----------------------- .. WDRC In Quiet Linear In Quiet WDRC N Nom T..--_-- . ; Lirrear N Noise Test Condition Figure 4 Mean SRS (%) obtained in quiet with WDRC and linear amplification. 452 Figure 5 Mean and SD of SRTs obtained in quiet and in noise with both programs . Comparison of "RC and Linear Amplification/Kam and Wong Table 6 Matched-Pair t-test Results for Speech Recognition Scores Obtained in Quiet with Both Programs Program Linear WDRC Presentation Level (dB SPL) Mean SD Mean SD df t 50 69 .50 19 .53 54 .50 23 .11 19 4.54* 88 .75 91 .75 65 80 10 .87 9 .36 11 .01 6 .67 83 .50 90 .50 1 .64 0 .63 19 19 WDRC = wide dynamic range compression *p < .01 , that obtained with linear amplification . That is, subjects were better able to repeat sentences at lower level in quiet with WDRC . The effect is congruent with increased gain provided by WDRC to low-level signals . Another possible reason is the increase in consonant-vowel-ratio (CVR) caused by WDRC, which ensures that the more intense vowel sounds receive less gain than the less intense consonant sounds (Hickson and Byrne, 1995 ; Kuk, 1996) . Speech intelligibility may be improved by increased CVR . Increasing the consonant level should serve to enhance the audibility of acoustic cues necessary for perception . At the same time, decreasing the vowel level should decrease the masking effects of the stronger components of speech on the weaker ones, thus also enhancing the audibility of consonant acoustic cues . The findings in quiet were consistent with those obtained in some other studies . For example, Dreschler (1988a) found better phoneme identification in quiet with a compression aid compared with a linear instrument in a group of 16 hearing-impaired subjects . Using a modified rhyme test, Nabelek (1983) showed better word recognition in quiet with WDRC in eight OSRS osingWDRC " SRS uvng Linear --- Flt Ilne for WDRC 628 .. Fttlinefo,Linear hearing-impaired subjects . Due to the large variation in methodology used in different studies, it is impossible to make any fair comparison between studies . However, as suggested by Kuk (1996), one common conclusion could be drawn from those studies that reported supportive evidence for WDRC : positive effect was observed in tests when the stimulus was presented in quiet at a fixed low level . Results from the present study give further support to this conclusion. Speech Intelligibility in Noise The overall SRSs obtained in the monosyllabic word test and the SRT obtained using HINT with WDRC were not significantly different from those obtained with linear amplification . In other words, WDRC did not provide significant improvement in speech intelligibility in noise. This finding is not uncommon . Many previous studies, which also employed a constant background noise in testing, reported similar results . For example, no significant difference in SRT obtained in noise in 12 hearing-impaired listeners was reported by Dreschler et al (1984) . Tyler and Kuk (1989) also failed to Table 7 Analysis of Variance for Speech Recognition Scores Obtained in Noise with Both Programs Source df Effect MS Effect Program SNR q 6 3 0 -3 Signal-to-poise Ran. (SNR) (dB SPL I -6 Figure 6 Mean speech SRS (%) obtained in noise with WDRC and linear amplification . Program x SNR df Error MS Error F 6 29162.38 114 262 .98 110.89* 167 .76 1 .38 6 620 .36 114 152 .16 4 .08* 1 232.23 19 MS = mean square SNR = signal-to-noise ratio. *p < .01 . 453 Journal of the American Academy of Audiology/Volume 10, Number 8, September 1999 Table 8 Matched-Pair t-test Results for Speech Recognition Scores Obtained in Noise with Both Programs Program WDRC Linear SNR (dB) Mean SD Mean SD df t +9 +6 +3 0 -3 -6 79 .25 75 .50 68 .50 50 .00 39 .00 15 .00 13.79 15 .30 19 .27 19 .74 24 .47 16 .86 77 .50 66 .75 63 .50 62 .75 47 .00 23 .00 7 .00 12 .41 13 .79 20 .72 18 .95 21 .97 21 .67 12 .92 19 19 19 19 19 19 19 0 .50 2 .49* 0 .98 2 .88* 1 .51 3 .24** 0 .31 -9 7.50 13 .91 SNR = signal-to-noise ratio, WDRC = wide dynamic range compression . *p< .05;**p< .01 . demonstrate any significant improvement in consonant recognition with WDRC over linear amplification in a multitalker babble background noise for 11 listeners. Reasons for poor performance with WDRC in noise were not well understood and have not been verified. There are some proposed explanations . Tyler and Kuk (1989) suggested that the temporal information contained in the speech signal may be disrupted by the dynamic amplification . Another possible reason is that compression amplification may increase the level of the background noise in the gaps of the speech signal (if the noise is present at a lower level than the speech) and hence cause masking of the speech signal (Hickson et al, 1995). The relationship between circuit type and speech intelligibility requires further clarifica- tion . Although WDRC did not improve speech intelligibility in noise, it did not degrade the performance in comparison to linear amplification. In the monosyllabic word test, at favorable SNRs, the mean SRSs obtained with WDRC were better than those with linear amplification. In more adverse listening conditions, linear amplification provided slightly better speech intelligibility. For the sentence test, the SRT obtained with WDRC was lower (i .e ., better speech intelligibility) than that with linear amplification, although the amount was not statistically significant . The findings in the present study do not support the notion that WDRC has a negative effect on speech intelligibility in noise when compared to linear amplification. However, WDRC may be slightly more vulnerable to noise Table 9 Subjective Preference for Program in Various Test Conditions (N = 20) Quality Signal Level (dB SPL) Number of Subjects Preferring WDRC Number of Subjects Preferring Linear Amplification Loudness 80 in quiet 65 in quiet 50 in quiet In noise at SNR = +6 15* 11 15* 11 5 g 5 80 in quiet 65 in quiet 50 in quiet In noise at SNR = +6 7 10 11 9 13 10 9 11 15* 13 12 8 5 Clarity Pleasantness 80 in quiet 65 in quiet 50 in quiet In noise at SNR = +6 SNR = signal-to-noise ratio, WDRC = wide dynamic range compression. *Sign test significant, p < .05. 454 9 7 8 12 Comparison of WDRC and Linear Amplification/Kam and Wong IS 15 M WDRC much better M WDRC much better 0 WDRC moderately better E3 WDRC moderately better p WDRC slightly better p WDRC slightly better M Linear much better Linear moderately better 0 Linear slightly better 0 Linear much better ~, t0 0 Linear moderately better U M Linear slightly better w O N E S z z 0 I a- 80 dB SK In Quiet 65 dB SPL U Quiet In Noise 50 dB SPL hi Quiet Test Condition Figure 7 Strength of subjective preference for "loudness" in various test conditions . than linear amplification . From Figure 6, it can be seen that the regression line for WDRC is steeper than for linear amplification . That is, a shift from a higher SRS to a lower SRS takes a smaller change in SNR for WDRC than for the linear program . This finding could not be compared to related studies in the area as other studies seldom varied the SNR . The implication of this observation is that when performance comparison between WDRC and linear amplification is made, the use of a single SNR is undesirable . A range of SNR should be included to give a fair evaluation of both systems . This issue had been discussed in Yund and Buckles's (1995) study, which evaluated the perfor- IS M WDRC much better WDRC moderately better F71 WDRC slightly better O, Linear much better Linear moderately better Linear slightly better I 80 dB SPL In Quiet 65 dB SPL In Quiet 50 dB SPL In Quiet In Noise Test Condition Figure S Strength of subjective preference for clarity in various test conditions . 80 dB SPL In Quiet 65 dB SPL In Quiet 50 dB SPL In Quiet InNoise Test Condition Figure 9 Strength of subjective preference for pleasantness in various test conditions . mance of multichannel systems instead of single-channel compression. Subjective Preference Significant subjective preference was found for WDRC for loudness appropriateness to signals at 50 and 80 dB SPL and for pleasantness to signals at 80 dB SPL. The explanations for these findings are straightforward . At 50 dB SPL, the gain provided by WDRC was significantly higher than that provided by linear program . Better audibility gained more votes for better loudness . In fact, most subjects complained that the sound with the linear program was too soft . Better audibility also resulted in significantly better speech intelligibility to 50 dB SPL signals. At 80 dB SPL, significantly more gain was provided by linear amplification. However, the extra gain seemed to be too much for most subjects in this study Although the stimuli at 80 dB SPL was not so pleasant, subjects' word recognition scores, compared to WDRC condition, were not degraded . It seems that although the signal was not pleasant, it was not distorted enough to reduce speech intelligibility. Interestingly, subjects did not show a significant preference for clarity in all test conditions. One possible reason is that once the stimuli were audible or clear enough using either program, it was difficult for the subjects to tell which program sounded clearer . Preference of clarity does not seem to be affected by preference of pleasantness . Alarge intersubject variability was observed in the preference judgment . From the strength 455 Journal of the American Academy of Audiology/ Volume 10, Number 8, September 1999 of preference shown in Figures 7, 8, and 9, it can be seen that in one condition, different subjects preferred different programs . Even when voting for the same program, some subjects found it much better while some found it only slightly better than the other program. Clinical Implications The results obtained in this study showed better speech intelligibility and listening comfort using WDRC in certain situations . This means that there is at least some support for the selection of single-band WDRC for clients with moderate to moderately severe flat sensorineural hearing loss . However, the usefulness of WDRC in noisy situations, where it is often advertised as being of great value, was not found . In this study, the performance of hearing aid programs varied with stimulus presentation levels and SNRs . This indicates the need to employ multiple testing levels when fitting hearing aids, especially nonlinear instruments. We recommend that, in quiet, the hearing instruments should be evaluated with at least three stimulus levels : high, average, and low. In noise, SNRs of +3 and +6 dB are recommended as the test conditions . It has been found that the average SNR of conversational speech is about +4 .8 dB in noisy environments and substantially less in automobiles (Teder, 1990). Subjective preference for clarity did not yield significant findings in this study. In clinical practice, subjective evaluation of clarity may be omitted. However, the inclusion of subjective judgments of other dimensions such as loudness appropriateness and pleasantness may provide some valuable information for circuit selection. A simple paired-comparison procedure may be adequate and easy to administer in the clinical evaluation of hearing aid performance . It is interesting to note that results using materials in Cantonese, a tone language with its own phonologic system and language structure, yielded similar results to those obtained using English materials . It would be interesting to compare the results to studies using other language materials to determine whether generalizations can be made . This would have significant implications for audiologists serving multicultural populations . Acknowledgment . We are grateful to the Phonak Hearing Centre Hong Kong Limited for assistance in subject recruitment and sponsorship. We thank the technical staff in the Department of Speech and Hearing Sciences 456 for their support. Thanks also to Anthony Yuan, Polly Lau, and Kevin Yuen for their assistance in preparing the multitalker babble . We are indebted to the subjects whose cooperative spirit made the research possible and enjoyable . Appreciation is extended to Dr. Bradley McPherson for his helpful suggestions during the preparation of this manuscript. Kam ACS, Wong LLN. (April 4, 1998). Comparison of Performance with Wide Dynamic Range Compression and Linear Amplification . Poster presentation at the 10th Annual Convention of the American Academy of Audiology [Abstract p. 1191, Los Angeles, CA . This study was submitted by the first author for the degree of Master of Science in Audiology at the University of Hong Kong in May 1998. REFERENCES American National Standards Institute . (1989) . Specification for Audiometers . (ANSI S3 .6-1989) . New York : ANSI . Bachler H, VonlanthenA . (1994) . PiCS comfort programs : signal processing tools to support your manner of communication . Phonak Focus 17 :2-25. Balfour PB, Hawkins DB . (1992) . Acomparison of sound quality judgments for monaural and binaural hearing aid processed stimuli. Ear Hear 13 :331-339 . Boothroyd A, Springer N, Smith L, Schulman J . (1988). Amplitude compression and profound hearing loss . J Speech Hear Res 31 :362-376 . Byrne D. (1996) . Hearing aid selection for the 1990s : where to? JAm Acad Audiol 7 :377-395 . Dillon HC . (1988) . Compression in hearing aids . In : Sandlin RE, ed. Handbook ofHearingAidAmplification: Vol. 1. Boston: College-Hill, 121-146 . Dillon HC . (1996) . Compression? Yes, but for low or high frequencies, for low or high intensities, and with what response times? Ear Hear 17 :287-307 . Dreschler WA. (1988a). Dynamic range reduction by peak clipping or compression and its effects on phoneme perception in hearing-impaired listeners. Scand Audiol 17 :45-51. Dreschler WA. (1988b) . The effect of specific compression settings on phoneme identification in hearing-impaired subjects . Scand Audiol 17 :35-43 . Dreschler WA, Eberhardt D, Melk PW (1984) . The use of single-channel compression for the improvement of speech intelligibility. Scand Audiol 13 :231-236. Fortune T. (1996) . Amplifiers and circuit algorithms of contemporary hearing aids . In : Valente M, ed . Hearing Aids : Standards, Options and Limitations . New York : Thieme Medical, 157-209. Fromkin V, Rodman R. (1988) . An Introduction to Language . Fort Worth, TX : Harcourt Brace Jovanovich . Gioannini K, Franzen R. (1978) . Comparison ofthe effects of hearing aid harmonic distortion on performance scores for the MRHT and PB-50 test. JAudiol Res 18 :203-208 . Hawkins DB, Naidoo SV (1993) . Comparison of sound quality and clarity with asymmetrical peak clipping and Comparison of WDRC and Linear Amplification/Kam and Wong output limiting compression . J Am Acad Audiol 4:221-237 . pression hearing aid: paired-comparison judgments of quality. JAcoust Soc Am 96 :1471-1478 . Hickson LMH. (1994) . Compression amplification in hearing aids . Am J Audiol 3:51-65 . Nilsson M, Soli SD, Sullivan JA. (1994) . Development of the Hearing In Noise Test for the measurement of speech reception thresholds in quiet and in noise. JAcoust Soc Am 95 :1085-1099 . Hickson L, Byrne D. (1995) . Acoustic analysis of speech through a hearing aid: effects of linear vs compression amplification . Aust J Audiol 17(1) :1-13. Hickson L, Dodd B, Byrne D. (1995). Consonant perception with linear and compression amplification . Scand Audiol 24 :175-184 . Hohmann V, Kollmeier B. (1995) . The effect of multichannel dynamic compression on speech intelligibility. J Acoust Soc Am 97 :1191-1195 . Kuk FK. (1996) . Theoretical and practical considerations in compression hearing aids . Trends in Amplification 1:1-39. Ladefoged P. (1993) . A Course in Phonetics . Fort Worth, TX : Harcourt Brace Jovanovich . Lau CC, So KW. (1988) . Material for Cantonese speech audiometry constructed by appropriate phonetic principles . Br J Audiol 22 :297-304 . Lundberg G, Ovegard A, Hagerman B, Gabrielsson A, Brandstrom U. (1992) . Perceived sound quality in ahearing aid with vented and closed earmould equalized in frequency response . Scand Audiol 21 :87-92 . Moore BCJ. (1996) . Perceptual consequences of cochlear hearing loss and their implications for the design of hearing aids . Ear Hear 17 :133-160 . Moore B, Glasberg B, Stone M. (1991) . Optimization of a slow-acting automatic gain control system for use in hearing aids . Br JAudiol 25 :171-182 . Moore BCJ, Johnson JS, Clark TM, Pluvinage V (1992) . Evaluation of a dual-channel full dynamic range compression system for people with sensorineural hearing loss . Ear Hear 13 :349-370 . Nabelek IV (1983) . Performance of Loaring-impaired listeners under various types of amplitude compression. J Acoust Soc Am 74 :776-791 . Neuman AC, Bakke MH, Hellman S, Levitt H. (1994) . The effect of compression ratio in a slow-acting com- Peterson M, Feeney M, Yantis P. (1990) . The effect of automatic gain control in hearing-impaired listeners with different dynamic ranges . Ear Hear 11 :185-194 . Plomp R. (1988). The negative effect of amplitude compression in multi-channel hearing aids in the light of the modulation-transfer function . J Acoust Soc Am 83 :2322-2327 . Preves DA . (1991) . Output limiting and speech enhancement . In : Studebaker GA, Bess FH, Beck LB, eds. The Vanderbilt Hearing Aid Report II. Parkton, MD : York Press, 35-51. Steinberg JC, Gardner MB . (1937) . The dependency of hearing impairment on sound intensity. JAcoust Soc Am 9:11-23 . Teder H. (1990) . Noise and speech levels in noisy environments . Hear Instr 41(4) :32-33 . Tyler RS, Kuk FK. (1989) . The effects of "noise suppression" hearing aids on consonant recognition in speech-babble and low-frequency noise. Ear Hear 10 :243-249 . Verschuure J, Dreschler WA, de Haan EH, van Cappellen M, Hammerschlag R, Mare MJ, Hijmans AC . (1993) . Syllabic compression and speech intelligibility in hearing impaired listeners . Scand Audiol (Supp138):92-100. Wong LLN, Au DKK, Tong MCF, van Hasselt A. (1996) . Comparison of the performance of a programmable 2channel full dynamic range compression hearing aid with 1-channel conventional instruments in Hong Kong. In: Proceedings of the Fourth Asian-Pacific Congress on Deafness . Manila : Plantila Press, 216-224. Wong LLN, Nilsson M, Soli SF. (in preparation) . Deuelopment of Hearing In Noise Test-Chinese Version. Yund E, Buckles K. (1995) . Discrimination of multi-channel-compressed speech in noise: long term learning in hearing impaired subjects . Ear Hear 16 :417-427 .