94- m PUBLISHEDBY WATERSCHROMATOGRAPHY DIVISION, MILLIPORECORPORATION VOL. 1, NO. 1 SUMMER 1989 3-" l t New PowerLine Based Automated System For Dissolution Control: Waters 610/600E PowerLine System for Dissolution Testing ............................. • IN THIS ISSUE NewPowerCine Based Automated HPLC System for DissolutionControl... Introduction four solventgradient system(functionIn 1987Watersfirstintroducedan ally equivalentto the 600E)with the automateddissolution/HPLC(High addition of a gradientvalvestation, PerformanceLiquidChromatograph) ThesePowerLinemodulesare a new interfacebasedon the 840 Data and way to integratethe latesttechnoloChromatographyControl Station.This giesof today's HPLCsystemcomposystemimprovesthe efficiencyof the nentsintoa combinedcentralized dissolutionlab by transferringsamples controller,providingcompletesystem directly fromthe dissolutionteststation communicationand systemparameto the HPLC,analyzingthosesamples ter documentation.Thisnew approach and documentingthe results, fromWaters providespower beyond In March of thisyear,at the 40th that of stand-alone HPLCcompoAnnual PittsburghConferenceand nents.Additionally,PowerLineHPLC Exposition,Watersannounced the systemscan beconfiguredwith either incorporationof dissolutionacquisian integrator,or a chromatography tion software in the PowerLine 610 workstationusingWatersMaxima is.ocraticand the 600Egradientsol, or ExpertEase'" software. • vent deliveryand systemcontrol ThePowerLineAutomated modules.The610isocraticPowerLine Dissolution/HPLCsystemconsistsof a systemmodulecan be upgraded to a 600Eor 610PowerLinemodule (Rev. ' . I 2.0 Software),a 484TunableUV/VIS Absorbance detector,a WISP'" 712D TM TM • • • " _'-" _ i , z i DissolutionTestingRequiring HighSensitivity Analysis... TheWaters845ChromatographyDataStation... .WatersExtendsitsFamilyof HPLCAutoInjectors... - .................................... 1-3 4-5 5-6 7 sampleprocessor,and a Hanson TransferController (TC)*which transfersdissolutionsamplesfromthe dissolutiontestapparatus directlyintothe active712Dsampleprocessor. Howthc PowcrI,ineSvstcmfor :Dissolution "l_sting _,brks Thefollowing operational description isapplicable to eitherthe 610or the 600EPowerLinesolventdelivery and systemcontrolmodules. Dissolutioncontrolisaccessedfrom the AutoinjectorSetupscreenof the 600Econtroller.When enabled,the 600Ecoordinatesthe transferof dissolutionsamplesintothe active712D autosamplercarouselwhile performingchromatographyon previously collecteddissolutionsamplesor °1 Ic_30n ResecrcP (_'Cl_O'<:hOr: calibration _,;8"0 Vor,oI Ave"ul! standards. Chcrswor'h, CA "2"3: WatersPo;verLineD_ssolutionSystemincluding Waters600EGradient PowerLineModule, Waters484 detector,the WISP712D,and the Hanson TransferControlle_(HansonSR-2 DissolutionBothcourtesyof ttanson Research Corporation.) The Hanson Transfer Controller Through intelligent software rou- (TC), activated by the 600E, removes samples from the dissolution bath at times specified in the Dissolution Table editor. Autoprobes, equipped with 5-10 micron filters, are lowered simul- tines and advanced sample handling techniques, the Waters Automated Dissolution/HPLC system provides greater confidence in sampling and analytical results. Should verification taneously into each of the dissolution vessels at the specified sampling time. Excess filtered sample iswithdrawn into six volumetric cylinders under vacuum. Each of the six cylinders have a liquid leveling rod to determine the Waters600EPowerLine ModuleAutoinjector SerupScreen Editor allowsyoutoactivate dissolution software. Inthepharmaceutical indus- of specific sample data be necessary, identified samples can be re-injected, since all dissolution samples are retained until the analyst approves all the data. sample volume that will be dispensed into the 712D autosam#er carousel, try,it is commonto periodicallyruncalibration standardsaftera predetermined numberof Combined l)i£_olL_d(_n "rc_(ingand Canton( t:niformit)' for Quail[)' Control Using positive pressure, the TC delivers the excess sample back into the dissolution vessels, simultaneously backflushing the autoprobe filters, The remaining measured and filtered sample isthen dispensed into empty, capped one ml vials inside the active 712D sample processor, Dissolution sampling times are entered in the Dissolution Table editor of the 600E Controller according to the featureofhthe600E samples avebeenanalyzedA PowerLine control convenient module istheAutoStandards routine,whichallows theusertoprogramstandardvialinjections betweena setnumberofsamplevials, In addition to dissolution analyses, content uniformity determinations are also routinely performed in the pharmaceutical laboratory Content uniformity determinations are usually performed separately by HPLC techniques, while dissolution results are more often based on spectrophotometric techniques. Waters Automated Dissolution System lets the pharmaceutical chemist combine dissolution protocol specified by the dissolution test. The injection sequence consisting of samples and standards, and chro- testing and content uniformity determinations using a single automated method. matography run times are entered in the Program Methods editor, The 600E dissolution software Waters600EDissolution Tablescreeneditor allowsentryofsamplingtimes_ Theautomated dissolution systemuses:he96violcarouselin In combining these two tests, three criteria must be met. First,two dissolution baths must be used. Most dissolu- determines whether an H PLCinjection can be made without interfering with the712Dautosampler toaccommodate the many samplestaken duringo dissolution test. The D_ssolution Table editorprovides theuser the acquisition of dissolution samples, If it can, the HPLC injection is campleted and the 712D sample carousel is rotated to the correct position to collect the dissolution samples. On the other hand, if the system determines that it cannot make the next scheduled injection before the next set of dissolution samples are acquired, it interrupts the injection sequence to rotate the 712D carousel to the correct dissolution sample acquisition position. Once dissolution samples are acquired, the carousel is rotated back to the position it was in before the sample transfer withdocumentation of sampletimesandvial locationofdissolution samples. Typically, the firstsixpositions of thecarousel arereservedfor multilevel calibration standards, tion tests require that only six samples be analyzed; however content uniformity determinations are usually performed on a minimum of ten samples, necessitating the use of a second dissolution bath (six additional samples). Second, content uniformity determinations must be performed when the active ingredient(s) are 100% dissolved. Third, the same HPLC method must be capable of analyzing dissolution samples as well as performing content uniformity determinations. If all three of these criteria are met, content uniformity determinations can be made utilizing either the same samples analyzed for the dissolution test (if actives are known to be completely dissolved), or samples withdrawn from the dissolution test vessels at a later took place and the injection sequence continues. In all cases, the 600E software prioritizes dissolution sample acquisition over that of a sample or standard injection, There are cases ,where it is possible that the samples scheduled to be injected into the H PLChave not yet been acquired. This may happen in a situation where the dissolution sampiing intervals are hours apart and the chromatography is programmed at minutes per injection. In this case, the dissolution software recognizes that the samples scheduled for analysis have not been acquired and will delay injections from these vials until they have been filled with dissolution samples. -2- time point (if the last dissolution sampiing time is known to occur before the actives are 100% dissolved). _l . ] Todemonstrate the feasibility of combining these two test procedures, identical HPLC methods (See Figure 1) were used for both the dissolution test samples and content uniformity deter- Figure 1: Waters 860 Netwo&ing Workstation Chromatography Sample Report of Timolide Separation ......................................... Header: minationsfor the antihypertensive productTimolide_.Timolidecontains hydrochlorothiazideandtimolol maleate at 25 mg and 10 mg per tablet respectively The Waters Automated Dissolution System was used to control the sampiing of 12 dissolution test vessels, i Acquisition method Units COMBINED3 Processing System number method COMBINED3 I Channel Injection 1 1 Vial Total 30 1 Run time Injection volume Acquisition version 9.00 50 uL rain 840/6.21 Sample rate Mode Processing version injections i i l.O0 per Analysis Vl.2 sec i i Description: M404@295NM ri rstPlot: _3E ,,.,i_ _.__,,..,o,4,_,., i ch_,,_,_., _ , .... • 55" transfer the samples to the active 712D autosampler, and analyze the dissolution sample and standard solutions. In this case, the Timolide actives were completely dissolved at the final disso- i _ i ,_,v s_" 4s4_ 3s" 30" _s ! f i _ mity determinations were based on the HPLC lution test time results point. obtained Content fromuniter10 of the 12 dissolution samples. Additional content uniformity samples prepared offline by accepted United States l_'s" _i _- Run time Injection Scale factor Integration Delay Relative composite drug assays on a single automated system providing them with the specificity and sensitivity demanded by today's drug formulations. rate amount standard amt Analysis Keyboardsof RemoteDevices Locked PeakDetect Threshold 2S The Waters PowerLine System for Dissolution testing brings a new dimension to dissolution analyses. Itsease of use, versatility precision and reliability surpass all other dissolution sample assay techniques. It is the only online dissolution HPLC system that performs a complete drug product release study. Now for the first time researchdissolution '8 " " ' Peak 0.00 sec Quantltation by Area Force Through Zero is Disabled 10% 7.22 Reject Absolute 693929 Area 25000 Peak window HeightInt 49907 BB 394845 18768 OFF Amount 0.028 Intercept 0.000e+00 SS 0.011 O.O00e+O0 Peak Name Slope Response HYDROCHLORTN. 2. 516e+07 6. 93929e+05 TIMOLOL 3.647e+073.94845e+05 .............................................................. HPLC separation of Timolide active ingredients, hydrochlorothiazide and tirnolol maleate art a Waters _Bondapok C_u column. The mobile phase consisted of water, ocetonitrile, methanol and pt t 3.0 phosphate buffer (38:8:2:2). The separation was monitored at 295 nm as specdied in USP_0{I. Table 1: Comparison of Manually Prepared Content Uniformity Sample Data to the Combined 10 Vessel Dissolution Data ............................................. Hydr0chlorothiazide Tablet mg/ml Tirnolol Maleate Auto Tablet mg/ml Manual Tablet Auto mg/ml Tablet mg/ml 9.93 9.92 1 2 25.05 25.38 11 12 24.53 24.58 1 2 9.88 10.01 11 12 The data in Table I indicates that the automated 3 25.10 13 24.62 3 9.80 13 9.78 method 45 6 7 8 24.93 25.03 24.83 25.23 25.05 14 15 16 17 18 25.23 24.78 25.34 24.61 25.50 4 5 6 7 8 9.81 9.97 9.74 9.98 9.85 14 15 16 17 18 9.99 10.00 10.20 9.63 9.79 outcome of the content uniformity test. From the analysis of the means and variances by T-tests and F-tests respectively, we conclude that the manual method o{ sample preparation {or content uniformity could be substituted with the 9 10 25.18 25.88 lg. 20 25.21 2525 g 10 9.89 10.30 19 20 10.06 AVG SDV 25.17 0.28 24.96 0.35 9.92 0.15 9.92 0.15 automated RSO 1.11 1.42 1.51 1.52 sample preparation method, but same chromotographic method of onolysis), gives a similar mean and variance indicating that this method does not significantly affect the dissolution method without see 20 sec Area Ret 4.78 Time TIMOLOL PeakWidth per min Window LC Results: PeakName HYDROCHLORTH. 1.00 3.00 Manual (different I , , Sample Sample Internal Mode Single Point Calibration Retention Time Offset profiles, content uniformity studies and ,-,, 9.00 min 50 uL volume (_OI]C_usJoI] extended /k '6 conditions: Pharmacopeia (USP)methodology were manually loaded into the WISP carousel for comparison with the dissolution samples (see Table 1). ers can execute ik -'4 _ M_,',_te-_ _'2 signifi- cant bias to the results........................................................ -3- Dissolution Testing Requiring Sensitivity Analysis High Thetrend in the pharmaceuticalindusFigure2: Waters860 Expert EaseSystemSuitabilityReport try istoward morecomplexformula................................................................ tions,lower dosages, and more System Suitability Report from line i of multi-method SYSS sophisticateddrug deliverysystems Acquisitionmethod SYSSUZT Processin9 method suchas controlledreleaseproducts. Units syst ..... bet Channel 1 Vial As a result,pharmaceuticalresearchInjection 2 Total injections Run time 15.00 min Sample rate ers are moving away from the tradiInjection volume i00 uL Mode tional analytical spectrophotometric Acquisition version 840/6.21 Processing Version techniquesand are turningto HPLCfor ...._ vi,_ _,-.,.:t2ch, _.... dissolutionanalysesand quality con__ trol.Theexcellentspecificityof HPLC ..... :2 _ makesit a moresensitiveand reliable _0 ¢ analytical tool than spectrophotometry Inaddition, the USPaccepts HPLCas a method of analysisfor dis.... solutionsamples. Dissolution _ _ _r-., :.'_ _'J ]*_-_ 4 0 by Papas,et el.) Theirmethod involved dispersingthe tablet in 10ml of a mixture of methanol and water (80%/20%) • Thiscorrespondedto working concentrationsof l00 p,g/ml and 5 I-,.g/mlfor a tabletcontaining 1 mg norethindroneand 0.05mg ethinyl estradiol respectively, Dissolutiontestingof oral contraceptiveswas modified by the USP2in per sec Calibration V1.2 ,j A ,.: "; w A 4 OI-alContraccptives very low levels, hove been described 9 1.00 2 Testing of Sensitiveand reproducibleHPLCseparations of 23 differentcommercial oral contraceptivetablets, containing norethindroneand ethinylestradiol at S_rSS 1 1 Table of _ System _ ..... _e Minute__ Suitability _2 _ Results Sample Na.... stdVial 1 ZnJZChartl Sasel*ne RMS noise 8.161671 Time UNKNOWN Ret 2.44 UNK_;OWN 2.81 4.71 N_me UNKNOWN NORETHINDRUNE ETRINYL ESTRAD. 9.11 10.28 _ame UNKNOWN (minutes) RetTime 2.44 UNKNOWN 2.81 drift (uV) -0.035435 (uV /sec) 4013Height 385 Plates -198Resolution 5235 107_ 39n 7805 254 643 1.z72 0.n5 0. 582 Area 56495 23481 4420 1645 9575 10551 Tailing 2.779 21.282 Factor 1. 069 1.078 (data points) (seconds) (seconds) widtb: baseline S-sigma half height 24 25. 924 9. 808 15 UNKNOWN 4.71 9.11 ET.I*_ EST_D.10.28 .............................. 13. 57 37 NURETHINDRONE 378 5.244 55.716 30.030 40 27.929 45.028 13.301 12.005 SystemSuitabilityReportof Oral ContraceptiveStandard Injection.Thesystemsuitabilityreport tara single vial injectionprintsthe chromatogram and reports the USPparameterssuchas resolution,plate counts,and tailing factors forall peaksidentified in the calibration table. In addition, a summary report isalso printed for multipleinjectionsof standard vials,reporting means,standard deviations, and relativestandard deviationsof area, height, retentiontime, tailing factor,resolutionand plate count. 1988to reflectthe useof HPLCas a method of analyzingdissolutionsampies.Usingthis method the tablets are dissolvedin 600 mlof dissolution estradiol.Thisrepresentsa 60-fold dilution over the me'hodology originellydescribed by Papas. Todemonstratethe sensitivityof programmable,offering the chromatographer the ability to optimizethe detectoroutput. A standard solutionwas prepared mediaand the samplesanalyzedby HPLC.Theformulationdescribed above would provide working concentrationsof 1.7I,zg/ml for norethindrone and 0.08_g/ml for ethinyl Waters484 TunableAbsorbance Detector,Waters Pharmaceuticallaboratoryperformed the dissolutiontest outlined in the USPfor norethindrone and ethinylestradio!tablets, Inthis experiment,the uniquecapebilitiesof the PowerLineDissolution/ HPLCinterfacewere coupledwith thoseof the484 absorbancedetector. representingthe concentrationof a typical oral contraceptivedissolution sample.TheHPLCanalysisrequired detection of thesecompounds at concentrationsbelow 1.7_g/ml and 0.08 i_g/mLfor norethindroneand ethinyl estradiol respectively The optics of the 484 detector are con- trolledby the PowerLine600Esystem controlmodule by meansof an lEEK488 camm unicotions bus.Wavelengthsand sensitivitysettingsare -4- :Popo:_, Andrew N, Marc} ...... Salvo'ore M .... d geloney. M<hQcl F. Rop_d Delermrnahon o{ Nore:hmdtorlu and F,_,,,,y, Es:r,";d,o' ,,, O,al phase I,qu,d _85) Contracept, ve foble/sby R....... Chrorno:ogrophy. LC, Vol 3 No .4. (Ap i _._o,_:_,_,o_e or.d Dh,nyl Severlh Suoplemenl Fs,rad;ol 2813 !1988) bble:s. IJSP XXl. d- j r TheUSPprotocol suggests100p,I injectionsizeswhich isequivalentto 170ng of norethindroneand 8 ng of ethinylestradiol injectedonto the HPLCcolumn.Given the poor spectral characteristicsof thesetwo campounds,analytical requirementsdictarean HPLCmethod that can provide a stable UVdetectorresponse,at high sensitivity,when operating at 200 nm. Dissolutiondata was reprocessed on the Waters860 networkingcampurersystem.Accurateand reproducible integrationwas obtained on peaks representingas little as a two millivoltresponse, Figure2 showsa systemsuitability reportgenerated by the Waters 860 Networking Computer System,illustrating a typical separationof a syntheticallypreparedoral contraceptive dissolutionsample usinga Waters484 absorbancedetector.Fivereplicate injectionsof a standard solution resultedin RSD'sof 2.5% for both actives,below the 3.0% specification setin the USPmonograph, In addition, the systemsuitability reportshowsthe theoreticalplatesfor ethinylestradiolat 10,551(USPstates 7,000minimum)and the resolution between the two activesat 2.8 (USP states"i.5minimum).Thetailingfactor forboth activeswas well below the maximumof 2.0 describedin the USP monograph. The Waters 845 Chromatography TheIdealChromatography _brkstation for the Pharn_aceutical l.al:x)ratory TheWaters845 Chromatography Workstationisone of the mostpowerful laboratory data handlingand control systemsavailable today It isa quantitativetool that enablesthe pharmaceuticalchemistto measurearea and/or peak height and to accurately calculatethe trueamount of material presentin samples, " Workstation TheWaters845 combinesthe power of the DigitalVAXstation2000 or the VAXstation3100DesktopCampurerwith the latestenhancementsof WatersExpertEaseSoftware.The 845'swindow environmentenhances the system'smultitaskingcapabilities, WatersExpert Ease Software WatersExpertEasesoftware isa directaccess,menudrivensoftware package optimizedfor gas, liquid,and ion chromatographyapplications. (Thereisan optional software package designed for gel permeationopplications.)It automaticallycollects data, integratespeaks,stores methods,plotsin real-time, , runsstatisticsand generates i graphics and custom reports. J i (;onclusion Analysisof dissolutionsamplesby chromatographictechniquescan put a burdenon the HPLCsystem.Dilute solutionsof activeingredientsmustbe capableof being reproduciblyanalyzed.TheWaters Automated DissolutionSystemiscapable of providing the userwith convenientmethodssetup, variablesampleinjectionvolumesand detectoroptions to meetthe demands for high sensitivityanalyses. ------------------._ - TheWaters ExpertEasesoftware offersinternalor external,singleor multi-pointcalibrationsup to 99levels. Itallows linear,quadratic, cubic, point-to-point and segmentedcurve fits.Calibration curvesand regression statisticscan be printed or plotted forpresentationor documentation purposes. The845 offersdefault integration parametersto meetthe needsof most pharmaceuticalapplications and integrationtimedeventsare provided for complexchromatographic separations. Waters Easy-Viewsoftwareallows simplemousedrivenmanual integrationand chromatogram zooming to seeand accuratelyintegrate eventhe smallestpeaks. Waters CustomReportManager allows the userto tailor chromatographic reportformats,reportsequencesand the reportprint location. Waters ExpertEaseCustomConvertcapability allows transferof valuable chromatographic resultsinto RS-1,Lotus1-2-3%20/20 or ASCII TM Waters 845 Chromatography Workstation featurincj the VAXstotion31 O0pezsonal computer from Digital Equipment Corporation. -5- formats for further statistical analysis ....... Complete methods documentation is available for GMP/GLP purposes. Waters I.AC/I-"lk'chnology Waters Laboratory Acquisition and Communications Environment (LAC/E") module provides advanced system control and data acquisition capabilities to increase your laboratory's productivity: The LAC/E module controls all Waters pumping modules including isocratic, single-pump gradient and multi-pump gradients. It also controls Waters family of WISP sample processors as well as Waters detectors (410 Differential Refractometer, 484 and 490 Absorbance Detectors). LAC/E serial boa rd options offer the ability to connect up to eight additional printers offering a more organized reporting structure for a multiple H PLC/GC laboratory Waters SAT/IN'" module provides tWOchannels of high resolution A/D with separate time bases to meet the needs of capillary GC, as well as LC, IC and GPC data collection. Introducing Waters Expert Ease System Suitability Software Option Waters Expert Ease System Suitability software option provides the pharmaceutical chemist with critical HPLC separation data needed to validate i" i "x \ Through advanced protocols, the Waters 845, Chromatography software Figure3: Enhanced GraphicDisplayof System Suitability Results .......................................................................... Syste,,, Suitability Results fo, Hydrochlorothiazide andTimolol Maleale '1.8 1.7 1.6 AREART = Retention= Areo Time '1 . 5 HOT = Height _1. 4 TR = Factor : Tailing Resolution N =. Plate Count 1 . 3 I. 2 % _. 1.0 R 0.9 S 0.8 D 0.7 mance. The software calculates plate count, resolution, and tailing factors, determines percent relative standard 0.5 0.4 0.3 deviation chromatographic for retention time, parlorarea, height, plate count,system resolution and tailing factors for single vials or entire multi-methods. 0.1 0.6 O.P 0.0 Waters Custom Convert manager transfers summary reports into third part,/software formats (RS-1,Lotus 1-2-3, 20/20 and ASCII) for further statistical analysis and graphic presentations as shown in Figure 3. communication provides unlimitedmethodsdevelopment optionsandsophisticated multipledetectorcontrol. LEGEND [-_ HYDROCH _ TINOLOL _ , RT AREA ,I1 HGT r- 1 _, 1:1 N ............................................................................... Analyticalresultscanbe converted to a thirdpartysoftwareformatforenhancedvisualdisplays.In thisexample,system suitabililyresultsfromtheanalysisof a calibrationstandardcontaininghydrochlorothiazide andtimololmaleatewereconvertedtoa formatcompatiblewithAccessTechnology's 20/20software,andthengraphedtodemonstrate advancedreportingcapabilities of the845system. -6- Waters Extends its Family of Sample Processors _;at__'rs 715 and 1700 WISP SampleProccs._rs for Advanced SampleHandling exclusive fluid path design that ensures accuracy and reproducibility for all injector volumes while eliminating Waters has introduced two new products to extend its line of auto injectors, the 715 Ultra WISP and the 1700 WISP cross-contamination and carry-over• A choice of sample injection volumes, from 1 microliter to 2000 micro- Sample Processors. The 715 offers full, flexible programmability with easy- liters and carousels with either 48 four-milliliter or 96 one-milliliter vials, to-use menus and CRTdisplay and an all-electric valve design. It is ideal for applications involving reagent additions and derivatization. It is compatible with automated workstations such are available. The Waters 1700 WISP Sample Processor provides automated sample preparation with Waters Millilab Workstation, plus direct sample injection for as the Millilab'" 1A Sample Preparation Workstation. on-line HPLC analysis. This innovative system performs filtration, solid phase The WISP 715 is compatible with Waters PowerLine HPLC system for single keyboard control and documentation of al) operating system parameters. In addition, it features an extraction, liquid-liquid extraction, mixing, addition of internal standards, heating, cooling, evaporation, trace enrichment, plus many other repetitive sample cleanup routines completed unattended, Pharmaceutical _"_c '._'_ _,cT_,c,__ ,j_ EDrj cJ_ r-'_:J _,_c.,_ : _: _ __1r_l ::_ i : " TheCRTinthe715WISPverifies system status samplesas neededwithoutin e "feting w;ththerunin prayross.Inaddition, system andsystem diagnostic routines makemessages troubleshooting and and conveniently programs priority ma;ntenance easy Notes Free Column Offer We would like to receive editorial contributions from you. If you have an HPLC, IC or GPC application or an improved sample preparation technique that you would like to share with our readership, we would like to include it in a future issue of Pharmaceutical Notes ..... In exchange, we will send you a free analytical reverse-phase C]8 column of your choice in a Nova-Pak TM, i_Bondapak'" or Resolve" chemistry if your contribution isaccepted. -7- -' Pleasesend your suggestions to: Carole Wade-Clark MC Waters Division of Millipore 34 Maple Street, Milford, MA 01757 How to receive future" U.S.A. Sales Offices New England Waters Ckromotograpny D,vis_ar,/ Mi!l;po,e Corpcralion / 34 Mop,e Street / Milord MA 01757 / Tel. 508-478-2000 Japan Nihon Mil ipore Ltd./Waters Chromatog'aphy Division/ No. 5 Kaike8;dg. 3-'2, Kita-Snir,agawc I-chome / Shinagawa-Ku, Tokyo 140 / Tel. 81-3-471-7)91 Chicago Waters Chromolography Division 1565 Mittd BoJleva,d / Wood Dole, IL 60191/ Tel.3;2-766-606'3 Telex 781-0232416; Mexico Mdipore S.A. de C.V. / Waters Chromatography Division/ Avenida Ingenieros Returnthe enclosedreply cord today New Jersey Waters Chromatography Division Militares 85 / Mexico II 230 D.F. to be sure you don't miss a future issue 35AirportPcrk/Suite230/Morrlstown, 07960/Tel. 201-285-?404 Tel. b76-9688/[elex1777442 The Netherlands Millipore B.V./ Waters of PharmaceuticalNotes. Forfurther North Carolina Waters Chromatography Chromatography Division/ Penr,ingweg 33 information concerning any of the arti- DMsion / i 17 kdi':burgn Drive/ Suite206 Carv, NC 275_', / Tel.919-469-250i San FranciscoWaters Chromatography Divis_or,c/o Millipore Corporation / 448 Grandv,ew Drive/ South San Francisco,CA 94080 Tel. 4;5-952-9200 Postoust66 / 4870 AD Etten-I_eur(N.B.) Tel. (0 ',6 08) 22000 / Tdex 54 840 milwc nl Norway Millipore AB/ Waters Chromatography Division/ Enebakkveien 133/ 0680 asia 6 Tel. (02) 67 82 53 PeoplesRepublic of China (Beijing) Nihon Southwest Wmers Chromatography Division 2626 SouthLoop West / Suite !70 Walers Service Center Rm.638, Xiyuan Hotel EriigouX;jiao, Beijing Houston, TX 77054/TeL .713-665 73;0 Tel. 8022018/Telex 22685 Washington Waters/ Chromatography 3702 PenderDrive Faidax, VA 22030Division Tel. 1-800-526-0771 Overseas Offices AustraLia Milllpo,e Pty. {.tc./Waters PeoplesChromatography Republic of ChinaDivision/ MilliporeHope China Ltd. Waters Sea IndustrialCentre / 26 LamHing Street / Suite 105 / Ist Floor / Kowloon Bay / Hang Kong/ Tel.3-796 3736 Telex 3-7967374 Chromatography D'vision / Locked Bog 18 87-89 Mars Road / Lane Cove, N.S.W 2066 Tel. Ir,wats (008) 222-I I I / In Sydney Area (02) 427-0511 / Telex25934 Austria Miil:po,e Ges.m.b.h. / Hielzinger Puerto Rico Mill:pore Corporation PuertoRico Branch / Waters Chromatography Division KM 7.7 / State Road 172/ Cidra, Puerto Rico 00639 / Tel.809-747.8444 Telex 385 72 76 Haupstrasse 145/A-II 30 Wier: Tel. (0222) 82 89 26 / Telex 13i464 Millv A Belgium and Luxembourg Mil!ipore-N.V. Waters Chromatography Division/ Raketstraat 60 / Ruede ',a Fus6e60 / 1130 8ruxelles-Brussels Tel. (02) 242. i 740 / Telex 2?798 milwat b Brazil Millipore Industria e Comercio Ltda. Waters Chromatography Division/ RuNPro( Spain Millipore Ib4rica S.A. /Waters Chromatography Division/ Avdo dd Llano Caslellano, )3 / 28034 Madrid Tel. (91) 729.03.00 / Telex 23545 milli e Spain Mil!ipore IberiaNS.A. / Walers Chromatography Division/ Entenza28 08015 Barcelona / Tel./93) 325.96.16 Telex 50524 wtrs e Campus de O;ivelra, 430 ,' CFP 04675-Caixa Postal 19065 / SaDPauio /Tei. (01I) 548-7011 Telex (I I) 25078 Canada Waters Chromatcg'ophy Division 3688 Nashua Drive/ Mississauga/ Ontario 1.4V-IM5/ Tel.416-678-2151 Toll-kee 1-800-268 488i / lelex 06-9685;3 Denmark Millipore AIS/Wate,s Chromatography D'vision / RoskiLdevei342 2630 T6strup/ Tel. (42) 52 88 ! ! Telex 33544 Sweden Millipore AB / Waters Chromatography Division/ Box 233 / L&r,gebergsgatan 30 421 23 V8stra Fr61unda/ Tel. (031) 28 98 60 Telex 21064 miqpors Sweden Millipore A8 / Waters Chromalography Division/ Grahamsv@ger,15 17246 Sundbyberg / Tel. (08) 98 89 60 Switzerland Mil!ipore AG / SteinackerstrasseI I 8302 Klok;n / Tel. 41-181.41-363/ lelex 56067 issues of Pharmaceutical Notes alesin thisissue,contact your nearest Waters technical representative or call 508-478-2000. Waters Chromatography Division Millipore Corporation 34 Maple Street Milford, MA 508-478-2000 01757 NJ Finland Mil!ipore Oy / Waters Chromatography Div:sior, / Ruukir:kuja! / 02320 Espoo Tel.90-801 90 77 / Telex _2-2123milpar si France Millioore S.A. / Waters Chromatography Div,sion/ Boite Postam307/78054 Saint Quenrn / YvelinesC6dex / Tel. {I) 30.45.70.00 Telex 698 371 F India Waters Instruments(India) Pvt. Ltd. 1_'Floor Shankar House / ) RajmahalVilas Extn. / The United Kingdom and Ireland Millipore (U.K.) Umlted / Walers Chromalography Division The Boulevard/ Ascot Rood Croxley G'een, Wafford HertfordshireWDI 8YW fel. 44-923 816375 Telex:851-24191MILLIPORG USSR,Middle East,EasternEurope and Africa Millipore Gesmb.h. / Hietzinger Haupstrasse )45 / A-1130W;en / Tel. {0222) 82 89 26 Te,'ex131464Miliv A Mekh5 Circle / Bcnga!ore 560 080 Tel. {812) 34-9203 / relax 8452028 SWPLIN Italy Mid;po'e S.p.A. / Waters Chromatography Division / Via Ach:lle Grande, 23/20090 West Germany Millipore GmbH / Waters Chromatography Division/ Hauptstrasse8)-85 6236 Eschborn/ Tel.06196-494-0 Telex4072704 raild Vimodrone MI/Tel. (2)250781 Telex 312284 milsoai/Telefax(2)2650324 Italy Miilpo-e S.p.A. / Wa'e,s Chromatography Division/Viale del Caravaggio 107/ 00147 Roma RM / Tel. (06) 5 _.32.452 / (06) 51.31.593 (06) 51.31.512/Telelax (06) 51.26.267 For all other countries: Waters Chromatography Division/ Millipore Corporation I 34 Maple Street M:ltord, MA 01757 / USA/ Tel.b08 478-2000 Telex _74166 Waters, Millipore, LACE/!:, SAT/IN, Expert Ease, WISP, t_Bondapak, Resolve, Nova-Pak, Maxima and MilliLab are trademarks of Millipore Corporation. Lotus 1-2-3 is a registered trademark of Lotus Corporation. 20/20 is a *rademark of Access Technology, Inc. Timolide is a registered trademark o{Merck, Sharp & Dohme. Pharmaceutical Notes is published by the Pharmaceutical Group of Waters. Waters D,v,s,on of ,NailI IPOR[ @,1989 Milhpore Corporation Primed in the U.S.A. 8/89 WN040 %._ ,i " "ql