Experiment 6 The Separation of Ferrocene, Acetylferrocene and 1, 1-Diacetylferrocene by Column Chromatography Background In 1951 Kealy and Pauson accidentally discovered ferrocene 3 by the attempted oxidation of cyclopentadienylmagnesium bromide 1 with anhydrous ferric chloride. Instead of the expected product, fulvene 2, a compound that was soluble in organic solvents and contained iron was isolated. Almost simultaneously Miller, Tebboth, and Tremaine isolated the same compound from the reaction of cyclopentadiene vapor with iron reduced at 300oC. The structure of ferrocene was shown to consist of two cyclopentadiene rings with an iron atom sandwiched between. Within five years of its discovery over 100 papers concerning the chemistry of ferrocene had been published. MgBr 2 1 FeCl3 Fe 3 2 ferrocene Of the many syntheses devised for ferrocene, the most satisfactory laboratory synthesis is the reaction between ferrous chloride and cylcopentadienylsodium in THF or 1,2-dimethyoxyethane as solvent. FeCl2 + 2 Fe Na + 2 NaCl 85-95% Similar sandwich compounds with other metals (metallocenes) as well as substituted ferrocenes may be prepared by this method. 92 Ferrocene is unusually stable for an organometallic compound (stable to 500oC). The stability is attributed to the sharing of the 12 -electrons of the cyclopentadienyl rings with the six outer shell electrons of iron (II) to afford a stable 18-electron noble gas configuration around iron. When other metals form similar compounds, known as metallocenes, they also react to achieve the 18-electron configuration. Because ferrocene is much more reactive than benzene (3.3 x 106 times in acetylation) it may be acetylated under milder conditions using acetic anhydride and phosphoric acid as catalyst. When acetic anhydride is used the major product is acetylferrocene with 1,1'-diacetylferrocene and ferrocene as impurities. O O CH3 Fe O O O CH3 H3PO4 CH3 Fe + Fe CH3 heat O major H3C minor 1,1'-Diacetylferrocene may be prepared under more vigorous reaction conditions: acetyl chloride/aluminum chloride; a mixture of mono- and diacetylferrocenes results. O O Fe 2 CH3 Cl Fe CH3 O + Fe CH3 AlCl3 CH2Cl2 O minor H3C major Thus, whichever procedure we use, there is always a mixture formed. We need to devise a way to separate small amounts of the acetylated ferrocenes. Recrystallization is a useful technique in this regard. However, we rarely ever isolate the minor components in a recrystallization. They constitute impurities in the mother liquors and as such we discard them. Column chromatography, on the other hand, is ideal for separating a small but macroscopic amount of a mixture into its components. Provided our observation of the conditions for separating a mixture by TLC is correct, we should obtain a comparable separation of a macroscopic sample using column chromatography. 93 Column Chromatography Column chromatography, as the name suggests, is carried out in a glass column, usually with a stopcock like that of a buret. The column "packing" is usually silica gel which is of a lower mesh size (i.e. more granular, providing less overall surface area) than that found on TLC plates. Hence, separation of compounds with very similar Rf’s on a column is often less complete than that achieved on a TLC plate. Just as in TLC, the solvent "elutes", only now from top to bottom. The objective is not only to separate the compounds on the column of silica gel but also to collect each purified sample as it exits the column through the stopcock. As long as fresh solvent is applied, the column keeps "running", unlike TLC where the eluting solvent is prevented from getting to the top of the plate. Therefore, we can separate macroscopic samples of solids (or sometimes liquids) from a mixture although not with the same purity as exhibited by the spots on a TLC plate The picture at the left shows a typical column. The column is packed with silica gel, usually introduced as a slurry in the eluting solvent. The solvent is allowed to run down to just above the level of the silica gel. Gentle tapping on the glass column while the solvent is running through the glass frit causes the silica gel to pack more uniformly and more rapidly. A concentrated solution of the mixture in the elution solvent is carefully applied to the top of the silica gel column and allowed to run down into the silica gel by opening the stopcock. Then eluting solvent is, at first, CAREFULLY applied to the silica gel, and afterwards filled up to the top of the column. The solvent then drains through the stopcock. If the compounds are colored the progress of the separation is evident as the bands separate, the higher Rf compounds travelling faster down the column than the lower Rf ones. The analogy of an upside-down TLC plate is evident from the progress of the "bands" of compounds. The solution containing the desired, purified component is collected in a flask, and the solvent is removed on a rotary evaporator leaving the purified solid behind. Another good way of introducing the mixture to be separated to the top of the column is by dissolving up the mixture in a round bottom flask in a minimum of a volatile solvent in which it is very soluble. Then an amount of silica gel, enough to soak up the solution, is added. The solvent is then removed to complete dryness on a rotary evaporator. This is really equivalent to applying a spot of the mixture on a TLC plate. This method is quite effective for most mixtures. Column chromatography is often used in place of recrystallization where complex mixtures are involved. It is adaptable to milligram quantities all the way up to several tens of grams, the amount of silica gel, and the size of the column being the only limitations. However, since the particle size of the silica gel used in columns (usually 70 to 230 mesh) is much larger than that found on most 94 TLC plates, the separations on columns are less resolved than those on TLC plates. This limitation is resolved by decreasing the amount of the polar component in the eluting solvent. For example, if a good separation on a TLC plate is achieved using 10% ethyl acetate in hexanes a good separation on a column may be achieved with 5% ethyl acetate in hexanes. A good rule-of-thumb is: halve the polar component of your TLC eluting solvent when you want a comparable separation on a column. Pressure Applied to the Column In flash chromatography pressure is applied to the column. This speeds up the operation and gives a better resolution compare to the slow elution observed without pressure. In the research lab compressed air or nitrogen is used in conjunction with a pressure regulator. In this lab balloons (double balloons) will be used to provide pressure. The balloons may be filled with compressed air from the dispensing hoods. Often, applying pressure to the top of the column can speed along a chromatography. When pressure is applied it is called flash chromatography. The Experiment Column Chromatographic Separation of Ferrocene, Acetylferrocene and 1,1'Diacetylferrocene WASTE: All used hexane and ethyl acetate should be poured into the waste collection bottle in hood Z. Check out a column from the organic prep room window-return it when finished. Read the procedure carefully before beginning. Tare the round bottomed flask in which you will collect the eluate. Pouring the column: Weigh 10 g of silica gel into a 150 mL beaker. Add approximately 30 mL 80:20 hexanes-ethyl acetate to make a slurry. Mix well with a glass rod. Add the slurry to the clamped chromatography column with the stopcock closed. Use two clamps on the column and make sure it is vertical. After all the silica gel has been added, let the solvent drain till the level is to the silica gel. Tap the column gently but frequently to settle the silica gel faster. Avoid air bubbles in the silica gel layer and do not let the column go dry. The solvent dripping through the column may be used to rinse the beaker containing the silica gel. At this point, the silica gel layer should be between 10 and 15 cm high. 95 There are several methods to prepare a sample to place on the column. Generally, the sample is dissolved in a small amount of the eluting solvent and the solution is applied to the column. In this experiment, the ferrocene mixture is too insoluble in ethyl acetate-hexanes. Instead, the ferrocene mixture will be dissolved in a methylene chloride, a more polar solvent. A small amount of silica gel is then added to form a slurry. The methylene chloride is evaporated leaving behind the silica gel containing the ferrocene mixture. The resulting silica gel is finally by applied to the top of the column. Place 0.2 g of the ferrocene mixture in a 25 mL round bottom flask. Add approximately 2 mL (1/2 pipette) methylene chloride to dissolve the sample. Add approximately 0.5 g of silica gel and mix well. Evaporate the methylene chloride on the rotary evaporator for about 5 minutes. Allow the flask to rotate for 20-30 seconds before lowering it into the warm water. The water in the bath should be slightly above room temperature otherwise the silica gel may shoot up into the rotary evaporator. Release the vacuum slowly and completely before removing the flask or air currents may enter the flask when you remove it. The air currents will cause the silica gel to be scattered and lost. Scrape the silica gel in the flask with a spatula to dislodge silica gel that may be adhering to the glass and transfer it to a piece of creased weighing paper. Applying the sample: Drain the solvent in the column so that it is level with the silica gel. Using a paper funnel, as demonstrated in the prelab, add the fereocene-silica gel to the top of the column distributing it evenly on the surface. Once the sample has been applied to the column the following operations should be done in succession without delays. At no point should the column go dry. Rinse the sides of the column with a small amount (1-2 mL) of the solvent using a 9-inch pipette. Attach the balloon (shown below) and apply pressure with the stopcock open until the solvent is level with the bed. Remove the balloon and add 2-3 mL solvent with the 9inch pipette and attached the balloon again (stopcock open). Repeat this 5-6 times so that the ferrocene mixture has moved into the silica gel bed. The developing solvent (hexanes-ethyl acetate 8:2) may then be added to halfway fill the column, the balloon attached with pinch clamp open, and the column stopcock opened. Collect the eluate in a flasks or beakers switching to tared round-bottomed flasks when the ferrocenes elute. The separation does not take place unless the eluting solvent is passing through the column. The least polar compound in the mixture, ferrocene, should elute almost with the solvent front. When the acetylferrocene has been completely eluted, switch to hexanes/ethyl acetate (4:6) to move the 1,1'-diacetylferrocene more quickly. The separation should take no more than 15 minutes-hence the name “flash chromatography”. Pour the hexanes/ethyl acetate eluate, collected between the colored fractions, into the hexanes/ethyl acetate recovery bottle. Returning to the elution operation now, you will notice, as each component elutes, some of the solvent evaporates at the tip, leaving solid behind. Wash this solid off the tip into the flask with ethyl acetate. When all three fractions have been collected, evaporate the solvent on the rotary evaporator in a dry 96 tared round bottom flask. It may be necessary to evaporate a solution portion-wise since a roundbottomed flask should not be filled more than 2/3. Residual solvent may be removed by attaching a vacuum adapter and applying a vacuum from the water aspirator while heating on the steam bath. Calculate the masses of ferrocene, acetylferrocene, and 1,1'-diacetylferrocene. Determine the percent composition of the mixture. Scrape out as much of the ferrocenes from the round-bottomed flasks as you can and place them into the labeled beakers on the lab desk. Flash Chromatography was introduced by W.C. Still and reported in the following article. The method is widely used in organic laboratories. Still, W.C. J. Org. Chem. 1978, 43, 2923. The Report should contain: 1. 2. 3. 4. 5. 6. Title Purpose Structures of compounds Procedure (reference and outline) Observations Results ferrocene (mg) acetylferrocene (mg) diacetylferrocene (mg) 7. Conclusion Possible Pre-Lab Quiz Questions: 1. What is the purpose of this experiment? 2. Compound A has an Rf of 0.75 and compound B has an Rf of 0.35. Which compound will elute first from a silica gel column using the same solvent system as in TLC? 3. Two of the substances used in this experiment, acetyl and diacetylferrocene, contain the acetyl group. What is an acetyl group? 4. Column chromatography is capable of yielding extremely pure substances. What accounts for this? [Hint: it is common practice to collect many fractions is test tubes when performing column chromatography on non-colored materials.] 5. During the chromatography the solvent is changed from hexanes/ ethyl acetate (8:2) to hexanes/ ethyl acetate (4:6). (a) What is the reason for doing this? 97 (b) Which is more polar? (c) How would you prepare 200 mL of hexanes/ ethyl acetate (8:2) 98