Where you find the city of Ulm Virtual Simulation and Verification in Total Body Irradiation using ACQSIM and an EPID with Extended Image Detection Unit Ulm J.E. Salk San Antonio University of Ulm Department of Radiotherapy University of Ulm Objectives Department of Radiotherapy • 2 Varian Clinac 2300C/D + 1 Clinac 600C – – – – 2 static + 1 dynamic MLC Enhanced Dynamic Wedge Varis R&V Portal Vision (Mark 2) • Picker CT PQ5000 – 2 VoxelQ & AcQSim • 2 CadPlan TPS Servers Objectives • Overview: Treatment Planning and Verification with AcQSim and CadPlan ______________________________________ • • • • • • TBI protocol at the University of Ulm TBI treatment facility EPID with extended Image Detection Unit for TBI Integration in VARiS Vision TBI Treatment Planning with AcQSim TBI Delivery and Verification • Overview: Treatment Planning and Verification with AcQSim and CadPlan ______________________________________ • • • • • • TBI protocol at the University of Ulm TBI treatment facility EPID with extended Image Detection Unit for TBI Integration in VARiS Vision TBI Treatment Planning with AcQSim TBI Delivery and Verification 3D Treatment Planning and Verification with CadPlan and AcQSim (Step 1) AcQSim CT Images Contours Ref.-Point (=Marked IC) Treatment Plan (optional) CadPlan CadPlan Dose Optimization CadPlan Dose Calculation BEV for adjustment of MLC 3D Treatment Planning and Verification with CadPlan and AcQSim (Step 2) axial coronal sagittal 3D Digitally Reconstructed Radiographs (DRR) CadPlan Field Parameters Monitor Units Isocenter Field Parameters MLC, etc. VARiS Server AcQSim Digitally Composited Radiographs (DCR) pa Beam Lateral Beam 3D Treatment Planning and Verification with CadPlan and AcQSim (Step 3) AcQSim DRR/DCR DRR DCR VARiS Vision Server DRR/DCR Film Multimodality Image Fusion Isocenter Verification with PortalVision Electronic Portal Image Digitally Reconstructed Radiograph MRT Image CT Image Interim Summary Total CNS Irradiation • AcQSim is an essential tool for many clinical situations • Virtual Simulation provides more information • CadPlan can be used more efficiently Verification Virtual Simulation (adjustment of beams) • Sagittal Reconstruction Provision of reference DRR DRR Total Body Irradiation Objectives • Overview: Treatment Planning and Verification with AcQSim and CadPlan á ______________________________________ • • • • • • TBI protocol at the University of Ulm TBI treatment facility EPID with extended Image Detection Unit for TBI Integration in VARiS Vision TBI Treatment Planning with AcQSim TBI Delivery and Verification (import beams from TPS) University of Ulm • Conditioning • Total patients with leukemia for BMT Dose: 12 Gy • 6 Sessions within 3 days • Lung Dose: 10 Gy AML and ALL: Radio-Immuno Therapy (Rh-188): • Kidney Dose: 10 Gy (RIT) + 7 Gy (TBI) TBI History and Progress Objectives University of Ulm • Overview: Treatment Planning and Verification with AcQSim and CadPlan á • 90 Patients (Oct 97 - Sep 99) 100 ______________________________________ Number of Patients (cumulative) 90 with RIT 80 without RIT 70 60 50 40 30 20 10 0 Oct 97 Dec 97 Mar 98 Jun 98 Sep 98 Dec 98 Mar 99 Jun 99 Sep 99 39 Patients with RIT + 51 Patients without RIT • • • • • • TBI protocol at the University of Ulm á TBI treatment facility EPID with extended Image Detection Unit for TBI Integration in VARiS Vision TBI Treatment Planning with AcQSim TBI Delivery and Verification Total Body Irradiation Objectives Machine and Treatment Room Layout ap beam: prone position pa beam: supine position ______________________________________ PV XL SSD = 5.10 m Treatment Room Clinac 600C • Overview: Treatment Planning and Verification with AcQSim and CadPlan á Treatment Room TBI Treatment Room Clinac 2300 C/D Portal Vision XL: An EPID with Extended Image Detection Unit • • • • • • TBI protocol at the University of Ulm á TBI treatment facility á EPID with extended Image Detection Unit for TBI Integration in VARiS Vision TBI Treatment Planning with AcQSim TBI Delivery and Verification Portal Vision XL: An EPID with Extended Image Detection Unit • 2 x 2 standard PortalVision IDUs Sensitive Area: 65 x 65 cm2 512 x 512 Electrodes = 262144 pixels Spatial Resolution: 1.27 x 1.27 mm2 Build Up Plates Optimized for 6 MV • Lead shielding for electronic parts Portal Vision XL: An EPID with Extended Image Detection Unit Portal Vision XL: An EPID with Extended Image Detection Unit • Fully closed housing • mounted on rails • Fully closed housing • mounted on rails Objectives VARiS Vision Network Layout • Overview: Treatment Planning and Verification with AcQSim and CadPlan á Additional VarisVision Workstations CadPlan Workstations Network printers VARiS Vision Server ______________________________________ • • • • • • TBI protocol at the University of Ulm á TBI treatment facility á EPID with extended Image Detection Unit á Integration in VARiS Vision TBI Treatment Planning with AcQSim TBI Delivery and Verification VoxelQ & AcQSim Workstations PVI Acquisition Stations Ximavision Workstation IAS 2 IAS 2 DRS PQ 5000 PortalVisionXL User Interface Ximatron Standard IDU Clinac with TBI option IDU XL Objectives • Overview: Treatment Planning and Verification with AcQSim and CadPlan á ______________________________________ • • • • • • TBI protocol at the University of Ulm á TBI treatment facility á EPID with extended Image Detection Unit á Integration in VARiS Vision á TBI Treatment Planning with AcQSim TBI Delivery and Verification TBI treatment planning: TBI treatment planning: Multiplanar Reconstruction Lung tissue density ρLung [g/cm3] = 1+HU/1000 (for HU < 0) TBI treatment planning: TBI treatment planning: Virtual Simulation Dose Calculation 3D Reconstruction Axial CT Digitally Reconstructed Radiograph Objectives TBI treatment setup • Overview: Treatment Planning and Verification with AcQSim and CadPlan á ______________________________________ • • • • • • TBI protocol at the University of Ulm á TBI treatment facility á EPID with extended Image Detection Unit á Integration in VARiS Vision á TBI Treatment Planning with AcQSim á TBI Delivery and Verification Patient prepared for treatment in supine position, radioopaque markers on anterior skin PortalVision XL in operating position Online Matching Clinac ready for Beam On Digitally Reconstructed Radiograph Electronic Portal Image • Gantry rotated • Couch retracted • TBI Accessory • TBI Flatness Filter Online Matching (cont’d) Digitally Reconstructed Radiograph Electronic Portal Image Final Summary and Conclusion • • • • • Extended SSD beam setups and improved localization AcQSim is smoothly integrated to VarisVision network EPID image acquisition time of approximately 2 seconds Image quality suitable for verification of shielding blocks Monitoring of patient during the treatment Combination of AcQSim and EPID is a good investment for fast and precise TBI The city of Ulm and the river Danube The city of Ulm and the river Danube